Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

Accessories - flexible tool for your help...

SHALYA
X111 Cord for patient return electrode
X121 Cord for Bipolar forceps
X126 Cord for Sealing Clamp BIP
CUT
X131 Cord for Endoscopy Equipment- 4mm
COAG
X135 HF cord for Single Stem Working Element
X141 Universal adapter
X143 Lure Lock Adaptor with RF connection
X151 Needle Electrode, 5 cm length
X152 Blade Electrode, 5 cm length
X153 Ball Electrode, 5 cm length
X155 Extended Blade Electrode, 8 cm
X158 Extended Blade Electrode, 12 cm
X161 Extended Blade Electrode, 15 cm
X191 Bipolar footswitch
X192 Monopolar footswitch
X201 Steel patient return electrode, Regular
X202 Steel patient return electrode, Pediatric
X211 Rubber patient return electrode, Regular
X216 Disposable Patient Plate- Single area, Regular, with cable
X221 Disposable Patient Plate- Dual area, Regular, with cable
X241 Foot switching chuck handle
X251 Monopolar Handswitching Pencil (Reusable)
X256 Monopolar Handswitching Pencil (Disposable)
X261 Bipolar Forceps, straight, 17 cm
X272 Bipolar Forceps, Bayonet, 20 cm
X281 Bipolar Forceps, straight, 17 cm, Nylon coated
X283 Bipolar Forceps, straight, 22 cm, Nylon coated
X293 Bipolar Forceps, Bayonet, 22 cm, Nylon coated
X301 Bipower Handle 001
X305 Bipower Tubing 36 cm
X311 Bipower Insert – Maryland -36 cm
X351 Sealer Handle 001
X355 Sealer Tubing 36 cm
X357 Sealer Tubing 18 cm
X361 Sealer Insert – Straight -36 cm
X362 Sealer Insert – Maryland -36 cm
X372 Sealer Insert – Maryland -18 cm
X381 Sealing Clamp Angled
X385 Sealing Clamp Angled- Coated
X405 Saline Loop for Single Stem Working Element- Storz type
X406 Saline Loop for Double Stem Working Element- Storz type

Add Hi-Tech Precision to your Surgical Expertise...

3057, Bhandup Indl. Estate, Pannalal Compound,


LBS Marg, Bhandup (W), Mumbai - 400 078. India Complete range of electro surgical solutions...
Tel : +91-22-2596 4645 / 3299 5113
Contact us at +91-9820329896
Shalya - Series of Electrosurgical Generators - the
Email : info@xcellancetech.com
most advanced electrosurgical system that delivers
Web : www.xcellancetech.com
optimum required electrosurgical effects through
all types of tissue - fibrous, muscle, vascular or fat.

Helping you make


the right choice..
advanced technological innovations... Effect Modes- plotted response as the tissue demands...
Cut Modes : Achieves consistent cutting in all types of tissues by 6SENSE Technology.
6SENSE Technology- more and more sensible... SP Cut / Low Cut : Offers precision cutting at lower voltages, hence suitable for delicate
cutting process where adjacent tissue damage should be minimum such as Neuro or ENT.
This also reduces capacitive coupling in endoscopy hence suitable for such procedures.
Shalya ESU Vs Conventional ESU in Bipolar mode Pure Cut : Suitable for clean & precisely controlled consistent cutting in all types of
This is a recent innovation in field of advance feedback controlled
Shalya ESU Vs Conventional ESU in Cut Mode
Electrosurgical Technology. System maintains the set power over tissues. Minimum Dragging effect, useful for TURP.
Excess Power results more tissue damage wider range of tissue by sensing -voltage, current, power, tissue Blended Cut : Consistent cutting power delivery with effective coagulation through a
density, patient return electrode monitoring and leakage RF Macro wide range of tissue types, reduces re-bleeding.
Standard
current- at “2000 TIMES” per second. Coag Modes : These modes offer every range of coagulation ability for modern surgeries.
6 SENSE ESU Micro
Conventional ESU Extra Power Desiccate : Specifically designed for contact coagulation at lower voltages, hence

Power in Walts
Conventional ESU Conventional ESU
Less power Fats Features
Power in Walts

Thermal spread >0.3 mm for blood rich suitable for Neuro, ENT & Endoscopy cases.
drags electrode Gall Bladder & saliva medium
in cutting • Minimizes dragging of electrodes in different tissues Fulgurate : Suitable for non contact coagulation with moderate sparking in variety of
Bowel • No need to adjust the power setting as the tissue changes applications such as cardiothoracic surgeries.
Liver • 50% less thermal damage than conventional ESU’s Spray Coag : Non touch coagulation for large areas with minimum depth of tissue
Prostate in conducing solutions • Minimizes the patient recovery time damage & necrosis. Useful in Liver Cancer & Trauma.
Neuro and conductive Tissue • Reduces the risk of collateral tissue damage Bipolar Modes : Delivers optimum Bipolar performance for improved coagulation.
• Minimum RF noise in other OT equipments Micro /Precise Bipolar : This mode offers clinical precision at lower voltages, prevent
• Reduces the risk of neuro - muscular stimulation Tissue Density in Ohms Sparking and reduce adjacent tissue damage used in Neuro, ENT, Ophthalmic surgery.
Tissue Density in Ohms
• Less charring and sparking hence precise & clean cutting 6 SENSE ESU
Thermal spread <0.15 mm
Standard Bipolar : Suitable for most of bipolar processes with optimized Coagulation.
Macro Bipolar / Bipolar Cutting : This mode offers smart control over consistent
CUTTING effects in all types of tissues. Output is specifically designed for today's new
generation of cutting instruments, bipolar needles and hooks.

Smart Pulse Cut- power on demand for Endo surgery... Randomized SPRAY COAG - the real spray...
Pulse Cut

Cut In conventional spray coag, beam tends to follow the same path left by previous spark
Coag resulting in non uniform coagulation over larger area. By randomizing spray frequency &
d1 d2 d3
For polypectomy & papillotomy, cutting is smartly divided in cut & pause, so that amplitude, spark tends to take multiple paths covering larger area.
coagulation occurs in each pause. The system is intelligent to change the pause interval as
tissue density changes, hence produces fine cutting with coagulation effect. Features
RF Power

• Starts the sparking from longer distance


• Uniform & homogeneous coagulation over large surface area at lower power settings
Smart Pulse Cut : Change in cut & pause as tissue demands. • Less tissue necrosis & carbonization
d1, d2, d3 are dependent on the tissue impedance. • Increases the speed of coagulation
• Reduces the output voltages
• Drastically improves the performance when combined with Argon enhanced ESUs

Time

Upgradebility - with the recent technology...


PREM- ultimate safety from burns...
Argon delivery along with electrosurgical RF current allows efficient non - contact,
uniform superficial and faster coagulation for better clinical results. Very effective for
superficial & high impedance tissue and reduces collateral tissue damage, hence suitable
to use with delicate tissue. Argon enhancement reduces smoke during open and
endoscopic surgeries

Features
Patient Return Electrode Monitoring – PREM - is the care & safety against the return • Uniform & faster coagulation.
electrode site burns. This ensures burn-free electrosurgery using DUAL AREA patient • Very effective for superficial & high impedance tissue such as bones, cancer
return electrode by continuous monitoring the contact quality with the patient. PREM fatty tissues
system gives an audio-visual alarm and stops the RF output ensuring the safe procedure. • Reduces collateral tissue damage, hence suitable to use with delicate tissue
• Reduces smoke
• System stops Argon delivery & ESU output if any of the system is not ready or fault
condition in handshake mode.

UEM - Ultrasound Electrosurigcal Module :


System can be connected to work with ultrasonic surgical Aspirator system in
Sp. Cut and Soft Coag modes.
• 2SiCoag : Two Simultaneous coagulation • 6 Bipolar Modes with 3 AutoBip Modes, Max Power – 70W
• 4 Cut Modes with PULSE CUT Mode for ERCP, Max Power - 300W • 4 Coag Modes with low CF Fulgurate, Max Power – 120W
• 6 SENSE Technology ESU • PREM - Patient Return Electrode Monitoring • Randomized SPRAY Coag • BiCoag Alarm

Q U A L I T Y • Bipolar Cut - in Macro Bip • AutoBip : Auto Start & Stop of Bipolar without foot switch • Argon Beam Coagulator up-gradable
• User Programmable : THIRTY Surgical Programs • Pulse Cut (Optional) for Polypectomy & Papillotomy • Self Diagnosis
• Remote Setting from surgical field

• 6 SENSE Technology ESU • 6 Bipolar Modes with 3 AutoBip Modes, Max Power – 70W
• 4 Cut Modes with PULSE CUT Mode for ERCP, Max Power - 300W
• 4 Coag Modes with low CF Fulgurate, Max Power – 120W
• PREM - Patient Return Electrode Monitoring • Randomized SPRAY Coag • BiCoag Alarm
• Bipolar Cut - in Macro Bip • AutoBip : Auto Start & Stop of Bipolar without foot switch
• Argon Beam Coagulator up-gradable • User Programmable : TWENTY Surgical Programs
• Pulse Cut (Optional) for Polypectomy & Papillotomy • Self Diagnosis
• Remote Setting from surgical field

S A F E T Y

D U R A B I L I T Y

• 2 Bipolar Modes, Max Power – 60W • 3 Cut Modes, Max Power - 250W
• 3 Coag Modes, Max Power – 100W • 6 SENSE Technology ESU • PREM - Patient Return Electrode Monitoring
• Randomized SPRAY Coag • Bipolar Cut - in Macro Bip • BiCoag Alarm (Optional)
• User Programmable : THREE Surgical Programs • Self Diagnosis • Argon Beam Coagulator up-gradable

• 2 Bipolar Modes, Max Power – 70W


• 4 Cut Modes with optional PULSE CUT Mode for ERCP, Max Power - 300W
• 3 Coag Modes, Max Power – 120W • 6 SENSE Technology ESU
• PREM - Patient Return Electrode Monitoring.
• Randomized SPRAY Coag • Bipolar Cut - in Macro Bip • BiCoag Alarm
• Pulse Cut (Optional) for Polypectomy & Papillotomy • Remote Setting (up-gradable)
• User Programmable : TEN Surgical Programs • Self Diagnosis • Argon Beam Coagulator up-gradable

• 2 Bipolar Modes, Max Power – 50W • 3 Cut Modes, Max Power - 200W
• 3 Coag Modes, Max Power – 80W • 6 SENSE Technology ESU
• PREM - Patient Return Electrode Monitoring • Randomized SPRAY Coag

E C O N O M Y •

Bipolar Cut - in Macro Bip • User Programmable - ONE Surgical Program
Self Diagnosis • Argon Beam Coagulator up-gradable
General Technical Specifications - comparison at a glance...
Smart Bipolar - auto bipolar & bicoag alarm for modern Monopolar Cut : Pure Sine wave 390 kHz Monopolar Coag : Damped Sine wave 460 kHz
Sp Cut - Programmable Pulse or continuous output, Soft - Repetition Frequency 62 kHz, 120 W at 500 Ù, CF 5.0
bipolar tools... 300 W at 300 Ù, CF 1.5 Fulgurate LCF - Repetition Frequency 56 kHz, 120 W at 500 Ù, CF 6.0
BiCoag Alarm : Audio feedback after completion of Bipolar Coag, reduces charring and sticking. Pure - Continuous output, 300 W at 300 Ù, CF 1.5 Fulgurate HCF- Repetition Frequency 34 kHz, 120 W at 500 Ù, CF 7.0
Blend - Bursts repeating at 29 kHz 55% ON cycle, Spray - Randomized Repetition Frequency 34 kHz < f < 50 kHz,
AutoBip : Automatic Start & Stop of Bipolar current by tissue sensing without foot switch activation. 200 W at 300 Ù, CF 2.5 120 W at 500 Ù, CF 8.0
Bipolar : Pure Sine wave 460 kHz DESIGN STANDARD
Micro - Lower output voltage, 70 W at 100 Ù, CF 1.5 Classification according to the EC- Directive 93/ 42/ EEC : II b
Standard - Medium output voltage, 70 W at 100 Ù, CF 1.5 Protective class according to EN 60 601-1 : I
Macro - Higher output voltage, 70 W at 100 Ù, CF 1.5 Type according to EN 60 601-1 : CF
2SiCoag - two coagulation generation in one...
LEAKAGE CURRENTS PREM SYSTEM
• RF leakage current • Monopolar : < 150 mA rms Measuring Frequency : 85 kHz ± 10 kHz
This is the unique feature of Shalya series, which system delivers two
• Bipolar : < 60 mA rms Measuring current : < 10 µA
simultaneous coagulation outputs at a time with independent control. Two
• Low Frequency leakage currents Acceptable Resistance Range.
surgeons can work at a time on different surgical area in a same case with • Normal polarity, intact ground : < 10 µA Dual Area PREM Patient Return Electrode : 10-135 Ù
flexibility as using seperate ESU. This makes the system more useful in • Normal polarity, ground open : < 50 µA Single Area Patient Return Electrode : < 10 Ù
cardiovascular & trauma surgeries. (Optional feature in Shalya TURoSeal)
FEEDBACK SYSTEM
INPUT POWER
• 3 Microprocessor Controlled 6SENSE Technology
Operating Voltage : 190-265Volt AC, (other ratings available on request)
with sampling rate above 2000 sense/second
• Maximum output power deviation < 15% or 5 watts, Max. Power input : 1000 VA,
Fuse : 4 A
Remote Power & Mode Settings - from surgical field.. whichever is greater at any setting with PER-97

Surgeon can change power and mode settings of ESU directly from sterile surgical field using Specification Shalya MX Shalya DX Shalya Easy+ Shalya Easy Shalya LX
conventional hand switching electrosurgical pencil.
BIPOLAR
Micro 70W at 100 Ù CF 1.5 70W at 100 Ù CF 1.5 70W at 100 Ù CF 1.5 60W at 100 Ù CF 1.5 50W at 100 Ù CF 1.5
Standard 70W at 100 Ù CF 1.5 70W at 100 Ù CF 1.5
Smart Surgical Assistant - to help you in need... Macro 70W at 100 Ù CF 1.5 70W at 100 Ù CF 1.5 70W at 100 Ù CF 1.5 60W at 100 Ù CF 1.5 50W at 100 Ù CF 1.5

User set able and Preset Programs : This helps you in setting up the system very fast after
CUT
turning ON. SP Cut (Low) 300W at 300 Ù CF 1.5 300W at 300 Ù CF 1.5 300W at 300 Ù CF 1.5 250W at 300 Ù CF 1.5 200W at 300 Ù CF 1.5
By pressing PROGRAM key, power setting display shows procedure’s name or number Pure 300W at 300 Ù CF 1.5 300W at 300 Ù CF 1.5 300W at 300 Ù CF 1.5 250W at 300 Ù CF 1.5 200W at 300 Ù CF 1.5
of program. After selection of required program, press RECALL just once to set system for Blend 200W at 300 Ù CF 2.5 200W at 300 Ù CF 2.5 200W at 300 Ù CF 2.5 150W at 300 Ù CF 2.5 120W at 300 Ù CF 2.5
selected application . Pulse Cut 300W, optional 300W, optional 300W, optional

COAGULATION
Desiccation 120W at 500 Ù CF 5.0 120W at 500 Ù CF 5.0 120W at 500 Ù CF 5.0 100W at 500 Ù CF 5.0 80W at 500 Ù CF 5.0
Self Diagnosis- Fulgurate 120W at 500 Ù CF 6.0 120W at 500 Ù CF 6.0 120W at 500 Ù CF 6.5 100W at 500 Ù CF 6.5 80W at 500 Ù CF 6.5
Fulgurate HCF 120W at 500 Ù CF 7.0 120W at 500 Ù CF 7.0
fastest error identification & solution...
Spray 120W at 500 Ù CF 8.0 120W at 500 Ù CF 8.0 120W at 500 Ù CF 8.0 100W at 500 Ù CF 8.0 80W at 500 Ù CF 8.0
System features the self diagnostic function during self test at the time of starting ON to ensure SAFETY
smooth functioning. As soon as fault is detected, it gives alarm and error code is displayed.
PREM Yes Yes Yes Yes Yes
Refer the Service manual may be referred to rectify the error.
Leakage RF current Yes Yes Yes Yes Yes

OUTPUT CHANNELS
- extended range... Monopolar 2 Foot / Hand 1 Foot / Hand 1 Foot / Hand 1 Foot / Hand 1 Foot / Hand
Bipolar 1 Foot / Auto / Hand 1 Foot / Auto / Hand 1 Foot 1 Foot 1 Foot

FEATURES
2SiCoag Yes
Inbuilt programs 30, (10 User Set) 20, (10 User Set) 10, (10 User Set) 3, (3 User Set) 1, (1 User Set)
Argon Enhancement Yes Yes Yes Yes Yes
BiCoag Alarm Yes Yes Yes Up -gradable
Auto Bipolar Yes Yes
Remote Setting Yes Yes Up -gradable

DIMENSIONS
In cm LxWxH 45.0 x 35.5 x 11.5 45.0 x 35.5 x 11.5 38.0 x 30.5 x 11.5 38.0 x 30.5 x 11.5 38.0 x 30.5 x 11.5
Weight in Kgs < 7.5 < 7.5 < 5.0 < 5.0 < 5.0
VesSeal CF : Crest factor indicates the ability of a waveform to coagulate without cutting.
Note : Specifications are subject to change without prior notice.

You might also like