Professional Documents
Culture Documents
Water Analysis Equipment COD BOD PH DO and Spectrophotometer Hotsale Catalog
Water Analysis Equipment COD BOD PH DO and Spectrophotometer Hotsale Catalog
Water Analysis Equipment COD BOD PH DO and Spectrophotometer Hotsale Catalog
Products Catalog
1, Chemical oxygen demand (COD rapid measurement)·········2
2, Digester / Reactor ·······················3
3, Biochemical Oxygen Demand/BOD meter············ ·4
4, Multi-parameter for COD/NH3-N/Total phosphorus/Turbity/metal and
nonmetal···························5
5, Turbidity, TSS (Total Suspended Solids) and residual chlorine····· 6
6, PH/DO/TDS/conductivity and multiparameter meter········ 7,8
7, UV/ VIS spectrophotometer··················9,10
1
Client Service: Elin Dai
Cell / Whatsapp: +86 133 7199 2597
Email: elin@shanghailianhua.com
elin@lianhuakeji.com
COD (chemical oxygen demand) Meter FEATURES
—Simple function design, users can easily get
on.
— Memory standard curve can be self-
corrected, and have power failure protection
function.
— With one key correction function, simple
operation.
— High quality material: Aviation thermal
insulation material, prevent scald effectively.
— Timing system, three timing switches,
automatic countdown reminder.
— Quick and accurate determination of
chemical oxygen demand (COD) in surface
water, reclaimed water, municipal waste water
and industrial waste water. Concentration data
can be read directly.
METER INCLUDES
Tube, Chemical Reagent, Cuvette, tube holder etc.
SPECIFICATION
Model 5B-3F (V10) 5B-3C (V8) LH-COD2M 5B-1B (V8)
Item COD COD and NH3-N Portable COD Reactor / Digester for
COD, phosphorus,
nitrogen
Range 50-800mg/L 20-10000mg/L 5-10000mg/L Digest temperature
(subsection) (subsection) 45℃~190℃
Accuracy ≤±10% ≤±5% ≤±5% Temperature accuracy
±0.5℃
Repeatability ≤±5% ≤±5% ≤±5% Testing samples: 30pcs
Anti-chlorine [CL-]<1000mg/L [CL-]<4000mg/L(optional)
interference
Colorimetric Rectangular Rectangular Round Tube Suit for tube inф16mm
method Cuvette Cuvette/round Tube
Lamp life 100 thousand 100 thousand hours 100 thousand Volume:0~10ml
hours hours
Display Digital tube LCD LCD 15 vials x2 double block
with different
temperature areas
2
Client Service: Elin Dai
Cell / Whatsapp: +86 133 7199 2597
Email: elin@shanghailianhua.com
elin@lianhuakeji.com
Digester / Reactor FEATURES
—Directly observe the condition of water sample
with safty.
—Aviation material, prevent scald effectively.
—LCD display, user-friendly menu design, operators
can
quickly master the instrument operation method.
—Digestion temperature and timing time can be
adjusted freely
in a large range.
—Automatically stops heating and power supply
after timing.
SPECIFICATION
Instrument name Digester / Reactor
Instrument model LH-9A 5B-1F(V8) 5B-1(V8) LH-25A
3
Client Service: Elin Dai
Cell / Whatsapp: +86 133 7199 2597
Email: elin@shanghailianhua.com
elin@lianhuakeji.com
BOD (Biochemical Oxygen Demand) Meter
FEATURES
You can measure 6 samples simultaneously without the need for
conversion, directly display the BOD concentration value;
2, automatically print data daily, automatically save this measurement
detail data;
3, colorful LCD display, each sample value is displayed in different colors,
easy to understand;
4, with a variety of stirring modes (intermittent, continuous), extend the
service life of the instrument;
5, the meter records the data automatically, and the recording frequency
can be automatically adjusted with the culture cycle;
6. You can view this experiment process data, as well as historical data
from this experiment.
METER INCLUDES
Chemical Reagent, culture bottle, measuring cap, etc.
SPECIFICATION
Model LH-BOD601 LH-BOD601L LH-BOD601SL
Item Biochemical Oxygen Demand Biochemical Oxygen Biochemical Oxygen
(BOD5) Demand (BOD5) Demand (BOD5)
Method 1-30 days culture by Mercury- 1-7 days culture by Mercury-free
free Manometric method
Manometric method
Range 0 ~ 4,000 mg/l (0-40, 0-80, 0-200, 0-400, 0-800, 0-2000, 0-4000 mg/l)
Accuracy ≤±5%
Record interval 6 minutes - 3 hours/time 5-7days 5-7days
Screen Big LCD on the meter Big LCD on the meter Small LCD on the cap
Measure amount 6
Data storage 20 years 5 years 7 days
Print With thermal printer With thermal printer Without printer
Culture bottle 580ml
volume
Culture temperature 20℃±1
Communication USB transmission, infrared transmission (optional)
4
Client Service: Elin Dai
Cell / Whatsapp: +86 133 7199 2597
Email: elin@shanghailianhua.com
elin@lianhuakeji.com
Multiparameter Water Quality Analyzer Meter
FEATURES
— Can measure COD, Ammonia Nitrogen, Total phosphorus
and Turbidity.
—High-definition color LCD, humanized operation and
simple to use.
— Intelligent data analysis.
— With date and time display function.
— With built-in curve function, you can modify the curve by
yourself. With one key correction function, simple to
operate.
— Can print current data and store historical data.
METER INCLUDES — Equipped with USB interface and infrared transmission
tube, powder Chemical Reagent, Cuvette, tube function, it can transfer stored historical data to the
holder computer.
SPECIFICATION
Model 5B-6C (V8) 5B-6C (V10) 5B-3B (V11) LH- MUP230 (V11S)
Item COD, NH3-N,Total phosphorus, turbidity COD, NH3-N, TP, total Portable COD, NH3-N, Total
nitrogen, turbidity, free phosphorus and
chlorine, chroma, total turbidity
suspended solid etc. 50
indicators
Range COD:5-5000mg/L COD:2-10000mg/L COD:(2~15000) mg/L COD: 2-10000mg/L
NH3-N:0.02-30mg/L NH3-N:0.02-100mg/L NH3-N:(0-160)mg/L (subsection)
TP:0.002-10mg/L TP:0.01-12mg/L TP:(0~100) mg/L Ammonia nitrogen:
(above subsection) (above subsection) Total nitrogen: (0~100) 0.02-140mg/L
Turbidity:0.5-400NTU Turbidity:1-300NTU mg/L (subsection)
Turbidity:(0.5~400) NTU Total phosphorus:
Chroma Range: (0~500) 0.002-10mg/L
Hazen (subsection)
TSS range: (0~1000) mg/L turbidity: 5-400 NTU
Free chlorine and total
chlorine range: (0~1.5)
mg/L
Accuracy COD≤± 5% NH3-N:≤±10% TP:≤±5% Turbidity:≤±8%
Repeatability ≤±5% ≤±2% ≤±5% ≤±5%
Batch process 12 12 16 16
Data save 12 thousands 16 thousands 12 thousands 21 thousand
Curve quantity 56 121 219 25
Colorimetric Rectangular Rectangular Cuvette/round Rectangular Cuvette/round round Tube
method Cuvette/round Tube Tube Tube
Lamp life 100 thousand hours 100 thousand hours 100 thousand hours 100 thousand hours
Display LCD LCD touch screen LCD touch screen LCD
Dimension 430*280*180mm 365*285*120mm 365*285*120mm 224*108*78mm
5
Client Service: Elin Dai
Cell / Whatsapp: +86 133 7199 2597
Email: elin@shanghailianhua.com
elin@lianhuakeji.com
6
Client Service: Elin Dai
Cell / Whatsapp: +86 133 7199 2597
Email: elin@shanghailianhua.com
elin@lianhuakeji.com
DO, PH, conductivity, TDS and multiparameter
FEATURES
Automatic temperature compensation ensures
accurate readings over the entire range.
— Auto-Read function senses and locks the
measurement endpoint.
— Calibration due alarm prompts user to
calibrate the meter regularly.
— Setup menu allows setting the number of
calibration points, stability criteria, temperature
unit, date and time, auto-power off, etc.
—Reset function automatically resumes all
settings back to the factory defaults.
— Expanded memory stores or recalls up to 500
data sets.
—USB communication interface for data
transfer and timed interval readings.
METER INCLUDES
electrode, chemical reagent, calibration liquid, membrane
SPECIFICATION
Model 810 901 900 900P
Item Benchtop Dissolved Benchtop pH Benchtop Portable
oxygen meter mV Multiparame multiparame
conductivity ter ter water
TDS testing
Meter
Range DO range: pH range: - pH range: - pH range: -
0.0~20.0mg/L, 2.000~20.000pH, 2.000~20.000pH, 2.000~20.000pH,
accuracy: ± accuracy: ± accuracy: ± accuracy: ±
0.5mg/L 0.002pH 0.002pH 0.002pH
%saturability mV range: - mV range: - mV range: -
range: 0~200%, 1999.9~1999.9mV, 1999.9~1999.9mV, 1999.9~1999.9mV,
accuracy: ± 2.0% accuracy: ± 0.2mV accuracy: ± 0.2mV accuracy: ± 0.2mV
conductivity range: Ion concentration Ion concentration
0~20.00mS/cm range: 0~19999 range: 0~19999
7
Client Service: Elin Dai
Cell / Whatsapp: +86 133 7199 2597
Email: elin@shanghailianhua.com
elin@lianhuakeji.com
(Max (ppm, mg/L, (ppm, mg/L,
200mS/cm), mol/L), accuracy:
mol/L), accuracy:
accuracy: ± 0.5% ± 1% F.S ± 1% F.S
F.S conductivity range:
conductivity range:
TDS range: 0~20.00mS/cm 0~20.00mS/cm
0~10.00ppt (Max (Max (Max
200ppt), accuracy: 200mS/cm), 200mS/cm),
± accuracy: ± 0.5%
accuracy: ± 0.5%
1% F.S F.S TDS range:
F.S TDS range:
0~10.00ppt (Max
0~10.00ppt (Max
200ppt), accuracy:
200ppt), accuracy:
± 1% F.S ± 1% F.S
salinity range:
Salinity range:
0~10.00ppt (Max
0~10.00ppt (Max
80ppt), accuracy:
80ppt), accuracy:
± 1% F.S ± 1% F.S
electrical resistivity
resistivity range:
range: 0~10MΩ0~10MΩ (Max
(Max 20MΩ), 20MΩ), accuracy:
accuracy: ± 1% F.S
± 1% F.S
DO concentration:
DO concentration:
0.00~20.00mg/L,
0.00~20.00mg/L,
accuracy: ± accuracy: ±
0.2mg/L 0.2mg/L
% Saturability
% saturability
range: 0~200.0%,
range: 0~200.0%,
accuracy: ± 2.0%
accuracy: ± 2.0%
Output Sensor Type: O100 USB USB USB
polarographic communication communicationcommunication
probe interface interface interface
Display Custom LCD (120 Custom LCD (125× Custom LCD (125×Custom LCD (80×
×60mm) 100mm) 100mm) 60mm)
Power Custom LCD DC5V, using AC DC5V, using AC 3×1.5V AA
(120×60mm) adapter, 220V/50Hz adapter, 220V/50Hz
batteries or DC5V
power adapter
Dimensions 210(L)×205(W)× 210(L)×188(W)×6 210(L)×188(W)×6 170(L)×85(W)×3
75(H)mm 0(H)mm 0(H)mm 0(H)mm
Weight 1.5kg 1.5kg 1.5kg 300g
8
Client Service: Elin Dai
Cell / Whatsapp: +86 133 7199 2597
Email: elin@shanghailianhua.com
elin@lianhuakeji.com
UV/VIS Spectrophotometer
EV-2200 / EU-2200
UV Visible / Visible Spectrophotometer
2200 series spectrophotometer is the ideal instrument for
education and QC laboratories. They are simple to fit a curve by
using your standard. They are widely used in colleges and
quantitative analysis lab.
※ Features
Small fine design, space saving
The excellent optical and electric system
Low noise and Low stray light
※ Function:
Large LCD display (128*64 Dots), can display curve
Photometric: T%, Abs
High quality grating, detector and lamps
Quantitative: Standard Curve
Data and Curve can be stored in real-time
System Utility
Auto setting WL, auto Blank
* WL Scan (Spectrum Scan) (PC)
Easy to change Pri-aligned lamps
* Time Scan (Kinetics) (PC)
Reinforced baseboard and bracket assures durability
* DNA/Protein Test (PC) * Multi-WL Test (PC)
※ Specifications:
Model EV-2200 EU-2200
Wavelength Range 320-1100nm 190-1100nm
Spectral Bandwidth 4.0nm/2.0nm
Optical System Single Beam, Blazed Holographic Grating (1200 lines/mm)
Wavelength Accuracy ±1.0nm
Wavelength Repeatability ≤0.3nm
Wavelength Setting Auto, Resolution 0.1nm
Photometric Range 0~200%T, -0.3-~3A, 0~9999C
Photometric Accuracy ±0.003A (0~0.5A), ±0.005A (0.5~1A), ±0.5%T (0~100%T)
Photometric Repeatability ≤0.002A (0~0.5A), ≤0.003A (0.5~1A), ≤0.3%T (0~100%T)
Stray Light ≤0.2%T(220/360nm)
Scan Speed High, Medium, Low. Max.1000nm/minute
Baseline Flatness ±0.004A
Stability ±0.002A/h (500nm,0A)
Noise ≤0.2%T/3min (250/500nm,0%T); ≤0.5%T/3min (250/500nm,100%T)
Sample Compartment Accommodate 5-100mm Pathlength Cuvette
Detector Silicon Photodiode
Lamps Tungsten Lamp & Deuterium Lamp (Pre-aligned)
Display Graphic LCD (128x64 Dots)
Keypad 11-key Membrance Keypad
Output Port USB Port & RS232 Port
Printer Mini Serial Printer; PC Printer
PC Software Optional PC Scanning Software
Power Requirements AC 90-250V,50/60Hz
Dimension 431x318x195mm
Weight 10kg 10kg
9
Client Service: Elin Dai
Cell / Whatsapp: +86 133 7199 2597
Email: elin@shanghailianhua.com
elin@lianhuakeji.com
EU-2600D
UV visible Spectrophotometer
EU-2600D spectrophotometer is double beam optical system
can restrain drift, compensates for blank changes, Suitable for
long time test. It is simple to fit a curve by using your standard
with single or dual WL. It have wavelength scanning function
using PC software. It is widely used in colleges and QC labs.
※ Features
Double beam optical system
Low noise and Low stray light
Large LCD display (128*64 Dots), can display curve
High quality grating, detector and lamps
※ Function: Data and Curve can be stored in real-time
Photometric: T%, Abs Auto setting WL, auto Blank
Quantitative: Standard Curve Lamps can be turned on/off individually
System Utility Easy to change Pri-aligned lamps
* WL Scan (Spectrum Scan) (PC) Reinforced baseboard and bracket assures durability
* Time Scan (Kinetics) (PC)
* DNA/Protein Test (PC)
* Multi-WL Test (PC)
※ Specifications:
Model EU-2600D
Wavelength Range 190-1100nm
Spectral Bandwidth 1.8nm
Optical System Double Beam, Blazed Holographic Grating (1200 lines/mm)
Wavelength Accuracy ±0.5nm
Wavelength Repeatability ≤0.2nm
Wavelength Setting Auto, Resolution 0.1nm
Photometric Range 0~200%T, -0.3-~3A, 0~9999C
Photometric Accuracy ±0.002A (0~0.5A), ±0.003A (0.5~1A), ±0.3%T (0~100%T)
Photometric Repeatability ≤0.001A (0~0.5A), ≤0.002A (0.5~1A), ≤0.2%T (0~100%T)
Stray Light ≤0.05%T(220/360nm)
Scan Speed High, Medium, Low. Max.2000nm/minute
Baseline Flatness ±0.0015A
Stability ±0.001A/h (500nm,0A)
Noise ≤0.2%T/3min (250/500nm,0%T); ≤0.5%T/3min (250/500nm,100%T)
Sample Compartment Accommodate 5-100mm Pathlength Cuvette
Detector Silicon Photodiode
Lamps Tungsten Lamp & Deuterium Lamp (Pre-aligned)
Display Graphic LCD (128x64 Dots)
Keypad 22-key Numeric Membrance Keypad
Output Port USB Port & RS232 Port
Printer Mini Serial Printer; PC Printer
PC Software Optional PC Scanning Software
Power Requirements AC 90-250V,50/60Hz
Dimension 490x390x280mm
Weight 16kg
10