Instruction Manual: Model ISC402G Inductive Conductivity Converter
Instruction Manual: Model ISC402G Inductive Conductivity Converter
Instruction Manual: Model ISC402G Inductive Conductivity Converter
Inductive Conductivity
Manual Converter
IM 12D8C2-E-H
4th Edition
OUTPUT
HOLD FAIL MENU POINTER
FLAG FLAG FLAGS
MENU FOR
MAINTENANCE
FUNCTIONS
MEASURED HOLD FAIL SEE CHAPTER 5
VALUE MODE
DISPLAY
∝S / c m
mS/cm
S3
>
>
ENT
FAIL/S4
SELECT MODE
MEASURE/COMMISSIONING
IM 12D8C2-E-H
Chapter Description Page Chapter Description Page
IM 12D8C2-E-H
1
1. INTRODUCTION 1-2. Measurement principle
1-1. Application Contrary to contact electrode conductivity,
The model ISC402G inductive conductivity highest quality engineering plastic known to- the EXA ISC Series analyses the conductivi-
mains powered converter is designed for day (Vitrex PEEK) for long service life in both ty without any contact between electrodes
use with the Model ISC40 sensor. This revo- abrasive and chemically corrosive process- and process fluid. The measurement is
lutionary conductivity measuring system fea- es. based on inductive coupling of 2 ring trans-
tures ± 0.5% accuracy over a wide range of formers (Toroids) by the liquid.
conductivity values (1 to 2,000,000 µS/cm) The stainless steel mounting thread and the
and process temperatures (-20 to 140 °C/0 Viton gasket allow safe and reliable installa-
to 280 °F) without changing cell constant or tion. The long insertion depth allows for in- Inductive conductivity measurement
recalibration. stallation of the sensor either by a bulkhead principle
V3 V1
mounting through a flange or tank wall, or by Receive coil Drive coil
The ISC402 transmitter is designed for ap- using one of the process adapter kits (e.g.
plication in which the concentration of acid, for flow-through or immersion service).
alkali or salt solutions is measured for proc- N3 N1
ess control purposes. The power supply can The large bore, greater than 17 mm (11/16
be from most AC or DC sources. The four in), gives long-term stability, preventing
separate supply options make it a universal measuring errors caused by coating or plug-
unit. As all operating parameters are stored ging. The large bore also offers fast re-
in non-volatile memory there is no danger sponse even at low flow rates.
"V2" = i x Rliquid
when power drops occur. The instrument
has a 'watchdog' timer which will always re-
turn it to the normal operating status after The transmitter supplies a reference voltage
power has been removed. The unit can lin- at a high frequency to the "drive coil". The
earise the output range in concentration core of this coil is of a high permeability
units and offers the potential for sufficient magnetic material, and a strong magnetic
"matrix" temperature coefficient depending field is generated in the toroid.
on the process composition. The liquid passes through the hole in the
toroid and can be considered as a "one
A self-tuning preamplifier provides a very ac- turn" secondary winding.
curate measurement over the full measuring
range. The ISC402 also features auto zero- The magnetic field will induce a voltage in
ing to eliminate traditional temperature and this secondary winding. The induced current
long term drift caused by magnetic offset in the liquid winding is proportional to this
between the two toriods. voltage and the conductance of the liquid
The Model ISC40 sensor is rugged steel- "one turn winding" is according to Ohm's
backed sensor encapsulated with the law.
The conductance (1/R) is proportional to the
IM 12D8C2-E-H
2
specific conductivity and a constant factor The software is designed with the user in 1-4. Instrument Check
that is determined by the geometry of the mind. It uses a simple 3 layer set-up to com- Upon delivery, unpack the instrument care-
sensor (length divided by surface area of the municate with the operator by giving mes- fully and inspect it to ensure that it was not
hole in the toroid) and the installation of the sages on the second line of the display area damaged during shipment. If damage is
sensor. and also indicates which keys should be found, retain the original packing materials
pressed. (including the outer box) and then immedi-
There are 2 toroids mounted in the "dough- The keys are scanned continuously and the ately notify the carrier and the relevant
nut" shaped sensor. The liquid also flows actions are taken immediately. An extensive Yokogawa sales office.
through the second toroid and therefore the system of checking values and parameters Make sure the model number on the
liquid turn can be considered as a primary is implemented. textplate affixed to the top of the display
winding of the second ring transformer. board of the instrument agrees with your
The current in the liquid will create a magnet- From the outside the ISC402G looks and order.
ic field in the second toroid. The induced volt- works just as a normal conductivity trans-
age being the result of this magnetic field can mitter. The operator has only access to NOTE:
be measured as an output. those functions that are needed for regular The textplate will also contain the serial
The output voltage of this "receive coil" is maintenance of the analyzer. Traditional number and power supply selection. Be
therefore proportional to the specific conduc- transmitters feature simplicity and ISC402G sure to apply correct power to the unit.
tivity of the process liquid. offers the same simplicity and the process- MODEL ISC402G-E-1-E
FD 020 034 ®
SERIAL NO.
ing power of the microprocessor makes it SUPPLY 110-120 VAC, 50/60 Hz, 10 VA EMA
EUR
1-3. Functional description possible to have a much higher level of func-
The EXA ISC402G is a real time micro-con- tionality and flexibility.
troller operated conductivity-analyzing sys- Check that all parts are present, including
tem. It uses a dedicated micro-controller to The power of the microprocessor is used mounting hardware, as specified in the op-
control all functions necessary in such a sys- for: tion codes at the end of the model number.
tem. - Diagnostic functions to increase the de- For a description of the model codes, refer
pendability of the instrument. to chapter 2 of this manual under General
The input and output functions are concen- - A self-tuning preamplifier to increase the Specifications.
trated in the analog section of the instrument. rangeability to cover almost all conductivi-
Even these functions are operated through ty applications. Basic parts list:
special interfaces designed to minimize inter- - Input/output flexibility to offer the user so- - Converter EXA ISC402G
ference with the digital functions. All functions lutions to compatibility problems and to - Packet with special cable grommet and
are executed separately. non-linearity characteristics of some elec- blanking pieces
trolytes. - Packet with 4 screws for mounting on a
The user-interface is limited to a basic set of 6 - Auto zeroing to insure long term stability panel (M6x8mm)
keys accessible through the flexible window - Sophisticated temperature compensation - Optional mounting hardware when speci-
cover. to achieve temperature independent read- fied.
ings for even the most difficult processes
like Sulphuric Acid and Sodium Hydroxide.
IM 12D8C2-E-H
3
2. SPECIFICATIONS
2-1. General specifications
A. Input specifications : Yokogawa Model ISC40 inductive Serial communication : Bi-directional according to EIA-485
conductivity sensor with integrated standard. Isolated from input and
temperature sensor YSI thermistor analog signals.
PT 1000 Ω.
E. Transmission range
B. Measuring range Conductivity : User programmable to any conduc-
Conductivity : 0 to 1999 mS/cm [at 25 °C (77 °F) tivity range within the indicating range.
reference temperature]. Configurable for linear and non-linear
Minimum conductivity signals.
at process Minimum span : 10 µS/cm.
temperature : 0.5 µS/cm. Maximum span : 1999 mS/cm.
Maximum conductivity Maximum zero
at process suppression : Up to 90 %. Topscale
temperature : 3000 mS/cm.
Temperature : -20 to 140 °C (0 to 280 °F). Temperature
Minimum span : 10 °C or 10 °F
C. Indicating range Maximum span : 160 °C or 280 °F
Main display : 0.0 µS/cm to 1999 mS/cm. Always linear and only on mA 2.
Message display : -20 to 140 °C (0 to 280 °F);
0 to 100.0 %. F. Contact Outputs
General : All contacts are SPDT contacts
D. Transmission signal : Two outputs, each 0/4-20 mA DC, LED indicators for all contacts (S1, S2
maximum load 600 Ohm, isolated and S3 LED on if the relay is powered)
from input and communication sig- (S4 LED on if the relay is not pow-
nals. A common negative line is used ered).
for both outputs. Switch capacity : 100 VA maximum at 250 VAC, 5 A
Output current on fault condition is 22 50 W maximum at 250 VDC, 5 A.
± 0.5 mA (user selectable). Both sig-
nals can be held (fixed or last value) Alarm contacts : 3 independent process alarm or con-
during maintenance trol contacts (S1, S2 and S3) for con-
mA 1 and mA 2 can be configured for ductivity or temperature value.
independent conductivity ranges. mA 1 contact S4 for conductivity or fail.
2 can alternatively be used to give a
temperature signal.
Status Alarm : High or Low process.
IM 12D8C2-E-H
4
2-2. Functional specifications
A. Performance
Conductivity alarm : Adjustable hysteresis and delay time Linearity : ± 0.3 % ± 0.5 µS/cm ± 0.02 mA.
Proportional duty cycle control with ad- Repeatability : ± 0.2 % ± 0.5 µS/cm ± 0.02 mA.
justable proportional range, pulse peri- Accuracy : ± 0.5 % ± 0.5 µS/cm ± 0.02 mA.
od and fixed minimum/maximum duty Influence of ambient
cycle. temperature changes : ± 0.1 % FS per °C.
Proportional frequency control with ad- Step response : Less then 3 seconds for 90% with-
justable proportional range and pulse in two decades.
frequency.
Temperature alarm : Adjustable hysteresis and delay time. B. Temperature
compensation : Automatic between -10 and 130 °C
Fail contact : Signals a fail of the measuring loop. (10 to 270 °F).
Specific process
liquids : User selectable matrix as a function
of concentration and temperature.
User programmable matrix as a
function of concentration and tem-
perature.
IM 12D8C2-E-H
5
2-3. Environment and operational conditions: 2-4. Construction
A. Ambient operating temperature A. Display : Custom liquid crystal display.
: -10 to + 55 °C (10 to 131 °F) Main display : 31/2 digit, 12.5 mm high.
Excursions to -30 °C (-20 °F) do not Message display : 6 alphanumeric characters, 7 mm
influence the current output func- high.
tion and excursions to +70 °C (160 Special fields : Flags for status indication
°F) are acceptable too. -HOLD, Hold output signal condition
-FAIL, Fault condition.
B. Storage Temperature : -30 to 70 °C (-20 to 160 °F). Measuring units: µS/cm; mS/cm.
YES, NO, ▲, , ENT. Menu pointer
▲
Key prompts :
▲
C. Relative humidity : 10 to 90 %. ( ).
D. Weather protection : Rain and dust tight to IP 65 B. Keys : 6 keys operated through flexible win-
(NEMA 4X). dow with tactile feedback and one
hidden key behind the front cover.
E. Data protection : Non volatile memory (EEPROM)
Backup for logbook C. Relay status
Battery powered Date/Time back- indication : 4 red LEDs.
up.
G. Automatic safeguard : Continues measurement if no key is Material : Cast aluminium with a chemically re-
pushed for 10 minutes. sistant coating. Cover fixing, stainless
steel screws in stainless steel inserts.
H. Voltage supply
interruption : At the nominal voltage Window : Flexible Poly-carbonate.
-Less than 50 ms, no effect
-Greater than 50 ms, reset. Colour : Cover, Moss green.
Case, Off-white.
I. Power down : No effect, reset to measurement.
Cable entries : Six polyamide glands 1/2"
J. Operation protection : 3 digit pass code.
Cable terminals : For maximum 2.5 mm2 cable (cable
finishing preferred).
IM 12D8C2-E-H
6
F. Shipping details
Dimensions 2-6. Model and suffix codes
(wxhxd) : 144 x 144 x 132 mm
(5.7 x 5.7 x 5.2 inch). Model Suffix Option Description
Package (wxhxd) : 290 x 225 x 170 mm code code
(11.5 x 8.9 x 6.7 inch). ISC402G .............................. Inductive Conductivity Transmitter
Weight : Approximately 2.5 kg (5 lb). -E ............................ Always E
Supply -1 ...................... 115 V, 50/60 Hz
voltage -2 ...................... 230 V, 50/60 Hz
-4 ...................... 24 V DC
-5 ...................... 100 V, 50/60 Hz
Instruction manual E ................ English *
Options /U ............ Pipe- and wall mounting hardware
/PM.......... Panel mounting hardware
/Q ............ Quality certificate
/SCT ........ Stainless steel tag
* For other languages, please contact your local sales organization
IM 12D8C2-E-H
7
3. INSTALLATION AND WIRING
3-1. Installation and dimensions
The EXA ISC402G should only be used in
conjunction with apparatus which meets the
relevant IEC, American or Canadian stand-
ards. Yokogawa can accept no responsibili- min.185
ty for the misuse of this equipment. (7.25)
144
3-1-1. Installation site (5.67)
The transmitter is a rain-tight type, and it can
be installed inside or outside. It should, how-
ever, be installed as close as possible to the min.195
(7.75)
sensors to avoide long cable lengths be-
tween sensors and transmitter.
(5.67)
Select an installation site where:
144
- Mechanical vibrations and shocks are
negligible
- No relay/power switches are in the direct Cut-Out = 138 x 138
environment (5.43 x 5.43)
- The transmitter is not mounted in direct
(1.0)
24
sunlight and severe weather conditions 138
- Maintenance activities are possible (no (5.43)
corrosive atmospheres).
(0.65)
The ambient temperature and humidity of
16.5
the installation environment must be within M6
the limits of the instrument specifications.
138
(4.55)
115.5
2x ø6.5
(0.26)
(7.87)
200
4x ø10
(0.4)
145
70 (5.70)
(2.75)
2" I.D. pipe
OPTION/U: Universal pipe/wall mounting kit
IM 12D8C2-E-H
9
3-1-3. Installation of the sensor
The ISC40G is a doughnut shaped sensor.
Ideally, the process flows through the hole of Unit: mm (inch)
the doughnut with the temperature com-
pensator up-stream. For minimal obstruc- Dimensions L=20000 (800) Installation instructions
tion of the flow and for accurate measure- bulk-head mounting
ment without the need for calibration of the
installation factor, the process will flow freely
around the doughnut, allowing a minimum 11
12
distance of 25 mm (1 in) between doughnut L= 5000 ( 200 )
17
and process piping (d). 13
15
The sensor is provided with a gasket and re- 16
taining nut. This allows "bulkhead mounting" 14
58 (2.28 )
cation on one flat aligns with the "up- D
Ø 27 ( 1.06 )
wrench opening
stream" position of the sensor. 32 ( 1.42 )
- For On-line mounting, adapters are availa- I
Ø 40 ( 1.57)
ble for standard 2 inch process connection
(Gas thread, NPT, ANSI-flange, DIN-
flange).
124 (4.88 )
d
- for by-pass measurement, flow fittings are
available in Polypropylene, Polyvinylidene
Fluoride and Stainless Steel.
- for measurements in open ducts or ves-
sels, immersion fittings in CPVC and Stain- 16.3 ( 0.64 )
47 ( 1.85 )
less Steel are available.
IM 12D8C2-E-H
10
3-2. System configuration
SENSORS POWER
0/4-20mA S1
OUTPUT CONTACT
SIGNALS OUTPUT
0/4-20mA S2
S3/WASH
CONTACT
RS485 OUTPUT
S4/FAIL
CONTACT INPUT
Figure 3-5.
IM 12D8C2-E-H
11
3-3. Wiring
3-3-1. Preparation
The relay contact terminals and power supply
connections are under the screening plate.
These should be connected first. Connect the
input, output and databus connections last.
The procedure to open the EXA ISC402G is
described below: Digital
Sensor Inputs mA Outputs Communications
1. Loosen the four captive screws and remove
the cover. 22 21 14 13 17 11 12 16 15 63 66 65 62 61 95 94 93 92 91
2. Use the rubber knob and swing open the
mA2 mA1
display board to the left. SCREEN
TL
SCREEN
TL
24 VDC 1 A
S4 S3 S2 S1
7. Connect the analog output, the sensor
input and if necessary the RS485 serial bus.
Refer to §3-3-4 and §3-3-5. Use the front
two glands, (refer to figs. 3-7 and 3-9).
8. Close the display board and switch on the High voltage compartment
power. Commission the instrument as re-
quired or use the default settings.
9. Replace the cover and secure with the four
screws.
IM 12D8C2-E-H
12
Option tag
IM 12D8C2-E-H
13
IM 12D8C2-E-H
14
IM 12D8C2-E-H
15
4. START-UP
The EXA ISC402G transmitter offers great 4-1-2. The secondary display 4-2. The Output (RANGE) functions
flexibility to let the user have one hardware The secondary display is a "message" dis- The following output functions are possible:
version for a multitude of different applica- play. It will indicate a description of the vari-
tions. able that can be adjusted during program- 4-2-1. Output linear to conductivity
ming, it allows display of a second variable This is the most common output function
The software has been programmed for a during normal measurement: and it is described in §6-2 "Output range
"general application". It will be necessary for - Type of Temperature Compensation (Out- adjustment".
the user to program the instrument to his put 1)
specific requirements or application. - Compensated SC (Output 2) 4-2-2. Output linear to concentration
- Type of Temperature Compensation (Out- When the conductivity system is used as
Although programming the transmitter is put 2) concentration analyzer, the ISC402G can be
simple, some preparation is necessary. This - % Concentration (Output 1) programmed to have the output current lin-
chapter will assist the user in recognizing the - % Concentration (Output 2) ear to concentration. This is described in
superb flexibility of the transmitter and mak- - Temperature §7-6 "Output table for non-linear range".
ing the best use of it's capabilities. - mA 1 The concentration can also be read as % by
- mA 2 weight on the LCD message display. This is
4-1. The DISPLAY functions - Installation Factor (I.F.) described in §7-3. "Percent by weight indi-
4-1-1. The primary display - Reference temperature cation on second display line".
The primary display is a 31/2 digit display - Software release.
that indicates the measured value. It is also 4-2-3. Output linear to temperature
used to display the variables that can be ad- mA 2 may be configured as a linear temper-
justed in the different programming steps. It will also, if relevant, display error codes. ature output. This is described in §6-2 and
Generally, the "SHIFT" > key changes the §7-5.
digit to be changed with the "UP" key. 4-1-3. Annunciators
When the period symbol (.) is flashing the The custom made display will also show rel- 4-2-4. Contact Input
"UP" key will adjust the value by a factor of evant messages as shown on the inside mA 1 and mA 2 will be swapped if the Input
10. cover of this manual. contact is closed.
NOTES:
1. The display indicates in either µS/cm or
in mS/cm (1 mS/cm = 1000 µS/cm).
Make sure that the correct unit is shown
during range adjustments.
2. If the prompt flag is flashing and no digit
nor period is flashing, then the "UP" key
will present the "1" and allows user to
change from 3 to 4 digits value. IM 12D8C2-E-H
16
4-3. Temperature (TEMP) 4-4. Calibration (CAL) 4-4-3. Calibration using a sample
compensation The factory calibration of the ISC402G al- Using an accurate reference standard such
The factory adjustment for automatic tem- lows the user to install the system without as the Yokogawa Model SC82 Personal
perature compensation is the standardized the need for calibration. In specific situations Pocket Conductivity Analyzer, the user can
NaCl algorithm according IEC 746-3. calibration may be necessary for high accu- perform a 1 point calibration according to
For concentrated electrolytes this may not racy. the instructions in §5-1.
be accurately compensate for process tem-
perature variations. NOTE: The factory calibration sets the I.F. 4-5. The contact output functions
The ISC402G gives the user 4 methods of (Installation factor) to a nominal Contact S1 and S2 defaults are set to high
optimizing the temperature compensation: 1.88. When a field calibration is per- and low alarm respectively.
formed it is the I.F. which is changed Contact S3 is not activated.
4-3-1.Field calibration of (linear) (see also sections § 4-4-2 & § 4-4- S4 is set to "FAIL"
temperature compensation 3.)
The procedure is described under §6-4. To change the allocation of the contacts
4-4-1. Air calibration/Air set refer to service codes 08-11.
4-3-2.Programming of calculated Every (new) ISC402G/ISC40 combination To change the characteristics of the con-
temperature coefficient has to be paired prior to installation of the tacts, refer to service codes 15 and, if pro-
This procedure is described in §7-13. sensor in process. portional control functions are required, to
Further the ‘Air Set’ routine enhances the service codes 12-14.
4-3-3.Selection of preprogrammed accuracy of the measurement at very low
electrolyte solutions conductivity values.
For Sulphuric Acid, Nitric Acid, Hydrochloric This procedure can be found in §6-5.
Acid and Sodium Hydroxide, Yokogawa has
defined standard ranges (see §7-23). 4-4-2.Programming of installation
factor
4-3-4.Programming of complex non- If the doughnut is surrounded by at least 25
linear, conductivity dependent mm (1 in) of process, the factory adjusted in-
temperature coefficients using stallation factor (cell constant) of 1.88 gives
an easy 5x5 matrix structure accurate results. In narrow piping systems,
This procedure is described in §7-24. the user can program different factors ac-
cording to the instructions in §7-2.
IM 12D8C2-E-H
17
NOTE: All three levels can be separately protected by a passcode. See §7-27
IM 12D8C2-E-H
18
4-7. Cleaning
If it is necessary to clean the outside of the
EXA housing, this should done using only a
soft cloth and household detergent. DO
NOT USE corrosive or abrasive cleaners.
4-8. Battery
The EXA instrument contains a lithium cell to
support the clock function when the power
is switched off. This cell needs to be
replaced at 5 yearly intervals (or when dis-
charged). Contact your nearest Yokogawa
service centre for spare parts and instruc-
tions.
IM 12D8C2-E-H
19
5. CONFIGURATION IN MAINTENANCE MODE 5-1. CALIBRATION
S3 S3 S3
> > >
>
>
>
ENT ENT ENT
FAIL/S4 FAIL/S4 FAIL/S4
CAL = Calibration
ESCAPE TO MEASURE can be used at any stage to abort operation. WARNING: If the HOLD function is activated, the instrument returns with
MODE the question HOLD (flashing); answer YES or NO or MODE again to return to measurement.
IM 12D8C2-E-H
20
1. When is calibration necessary? 5-1. CALIBRATION
Initially follow the procedure given in item
§4-4-1 for air calibration when setting up the Make sure the sensor does not touch the
A calibration solution can be prepared in the
instrument for the first time. sides of the container, refer to Fig. 5-1.
laboratory. A salt solution is prepared with a
known precise concentration.
Calibration of conductivity instruments is not Alternatively the instrument can be calibrat-
normally necessary as the conductivity cells ed using the process solution measured
The temperature is stabilized to the refer-
are manufactured to close tolerances and do with a standard instrument. Care must be
ence temperature of the instrument (usually
not alter in use. taken to make the measurement at the ref-
25 °C). The actual conductivity value of the
solution is taken from tables. To calibrate the erence temperature since differences in the
If the cell has severe fouling or been subject type of temperature compensation of the
instrument, the sensor is removed and sus-
to abrasion (possibly during cleaning) it may instruments may cause an error.
pended in the solution, the conductivity val-
be necessary to calibrate.
ue is then entered from the tables and the
calibration routine completed.
Alternatively calibration may be carried out
with a simulator to check the electronic cir-
cuits and the sensor.
IM 12D8C2-E-H
22
5-2. SELECTING A VALUE TO DISPLAY
S3 S3 S3
> > >
>
>
>
ENT ENT ENT
FAIL/S4 FAIL/S4 FAIL/S4
DISP.1 /DISP.2 = Display routine Display will show temperature compensation Display will show compensated conductivity
method value 2
MODE Press MODE key for access to
maintenance mode Press NO-key to change param- Press NO-key to change param-
NO NO eter to be displayed.
eter to be displayed.
Parameter Parameter
Select display DISP.1 or DISP.2 routine - Compensated SC type 2
- TC type 1
- % concentration 1 (if activated - TC type 2
NO Press this key until DISP.1 or - % concentration 2 (if activated
in §7-3)
DISP.2 is shown. in §7-3)
- Temperature
- Output signal 1 - Temperature
- Output signal 2 - Output signal 1
DISP=Display routine - Output signal 2
- Installation factor
- Reference temperature - Installation factor
Confirm selection of DISP.1 or - Reference temperature
- Software release
YES DISP.2 by pressing YES-key
- DISP.1 - Software release
when the required display is - DISP.2
shown. Press the YES-key to confirm
YES Press the YES-key to confirm
the required parameter to be
Continue with step 2 for DISP.1 parameters. YES the required parameter to be
displayed.
displayed.
Continue with step 3 for DISP.2 parameters. MODE Return to measure
MODE Return to measure
ESCAPE TO MEASURE can be used at any stage to abort operation. WARNING: If the HOLD function is activated, the instrument returns with
MODE the question HOLD (flashing); answer YES or NO or MODE again to return to measurement.
IM 12D8C2-E-H
23
5-2. SELECTING A VALUE TO DISPLAY
IM 12D8C2-E-H
23
5-3. USE OF HOLD FUNCTION
CONTACTS CONTACTS
YES NO MODE YES NO MODE
S1 * S1 *
S2 S2
S3 S3
> >
>
>
ENT ENT
FAIL/S4 FAIL/S4
NOTE: This function can only be used if activated Make selection the switch HOLD on or off
during commissioning (see §6-3).
ESCAPE TO MEASUREMENT can be used at any stage to abort operation. WARNING: If the HOLD function is activated the instrument returns
MODE
with the question HOLD (flashing); answer YES or NO or MODE again to return to measurement.
IM 12D8C2-E-H
25
5-3. USE OF HOLD FUNCTION
1. What is HOLD? 2. How does it work? 3. Example
Hold is a function which freezes the output From this level the HOLD function can only During the transfer of cleaning liquid into a
signal temporarily, it is normally used during be switched ON or OFF. batch reactor with a conductivity-control sys-
maintenance to prevent unwanted controller tem, the HOLD function is switched ON to
reaction when the cell is removed from the HOLD is switched on when you press YES prevent the controlling instruments from run-
solution being measured. when HOLD and YES/NO are blinking. ning wild.
When you press NO hold will be switched
The HOLD function must be commissioned off. After cleaning has ceased and the new batch
from the programming menu before it can has been started HOLD is switched OFF
be switched on or off. See commissioning A flag is kept in memory and an indication is again and conductivity control resumes.
the hold function for more details. made in upper left corner of the display field.
NOTE: AUTO RETURN will switch off The HOLD function only influences the out-
HOLD after 10 minutes, see §7-10. put signal, no other functions are influenced.
IM 12D8C2-E-H
26
5-4. ADJUSTING THE SET POINTS
1. ACCESS TO SET POINTS MENU 2. SELECT SET POINT TO ADJUST 3. ADJUST THE SET POINT VALUE
mS/cm
S3 S3 S3
>
>
>
>
ENT ENT
FAIL/S4 FAIL/S4 FAIL/S4
NOTE: This function can only be used if activated Display will show *SETP1 The display indicates the actual setting in engineer-
during commissioning (See §6-3 and §7-18). ing units. This setting can be changed using keys:
YES Press YES-key to adjust set
Press MODE key for access to point S1 or press NO-key to
MODE maintenance mode. NO change to required setpoint and > Select digit to adjust
after that the YES-key.
Select setpoints menu for contact outputs Adjust value
>
NOTE: Only those setpoints activated will be dis-
Press NO-key until SETP is indi- played, (refer to §7-18). ENT Confirm adjusted value
NO
cated
The message display will indicate the parameter to Continue routine to calibrate second point (repeat
SETP = Set points menu which the contact is allocated SC (conductivity) or step 2).
T °C/°F (temperature).
YES Confirm selection by pressing MODE Return to measurement mode.
YES-key.
See notes on page 27.
The display will indicate SETP1
ESCAPE TO MEASURE can be used at any stage to abort operation. WARNING: If the HOLD function is activated, the instrument returns with
MODE the question HOLD (flashing). Answer YES or NO or MODE again to return to measurement.
IM 12D8C2-E-H
27
5-4. ADJUSTING THE SET POINTS
IM 12D8C2-E-H
28
6. CONFIGURATION IN COMMISSIONING MODE
1. ACCESS TO SET POINTS MENU 2. SELECT SET POINT TO ADJUST 3. ADJUST THE SET POINT VALUE
mS/cm
S3 S3 S3
> > >
>
>
>
ENT ENT ENT
FAIL/S4 FAIL/S4 FAIL/S4
Remove cover by releasing 4 screws Display will show *SETP1 The display indicates the actual setting in engineer-
ing units. This setting can be changed using keys:
⊗ Press this key for access to
commissioning mode YES
Press YES-key to adjust set
point S1 or press NO-key to
NO change to required setpoint and > Select digit to adjust
Select setpoints menu for contact outputs after that the YES-key.
>
Adjust value
Press NO-key until SETP is indi- NOTE: Only those setpoints activated will be dis-
NO cated Confirm adjusted value
played, (refer to §7-18). ENT
SETP = Set points menu The message display will indicate the parameter to Continue routine to calibrate second point (repeat
which the contact is allocated SC (conductivity) or step 2).
Confirm selection by pressing T °C/°F (temperature).
YES YES-key. MODE Return to measurement mode.
ESCAPE TO MEASURE can be used at any stage to abort operation. WARNING: If the HOLD function is activated, the instrument returns with
MODE the question HOLD (flashing). Answer YES or NO or MODE again to return to measurement.
IM 12D8C2-E-H
29
6-1. ADJUSTING THE SET POINTS
IM 12D8C2-E-H
30
6-2. OUTPUT RANGE ADJUSTMENT
1. ACCESS TO OUTPUT MENU 2. ADJUST START VALUE OF THE RANGE 3. ADJUST THE END VALUE OF THE RANGE
µS/cm
mS/cm
S3
S3 S3
>
>
ENT
> >
>
>
ENT ENT
FAIL/S4
FAIL/S4 FAIL/S4
Remove cover by releasing 4 screws Display will show *RANG.1 Display will show 100%
Press YES-key to adjust
⊗ Press this key for access to com-
missioning mode
YES RANG.1 or press NO-key to
Adjust full-scale value of the output
change to RANG.2 and after
> Select digit to adjust
that YES-key.
Select RANGE menu
NO Display will show SC in RANG.1
Adjust value
>
Display will show SC or T in
Press NO-key until RANGE is in-
NO dicated
RANG.2
ENT Confirm adjusted value
(See §7-5 for selection of Temp
on mA 2)
RANGE = Output range adjustment MODE Return to measurement mode
YES Confirm selection
Confirm selection by pressing
Display will show 0%
YES YES-key.
Adjust start value of the output
The display will indicate RANG.1
> Select digit to adjust
NOTE:
Adjust value
>
ESCAPE TO MEASURE can be used at any stage to abort operation. WARNING: If the HOLD function is activated, the instrument returns with
MODE the question HOLD (flashing); answer YES or NO or MODE again to return to measurement.
IM 12D8C2-E-H
31
6-2. OUTPUT RANGE ADJUSTMENT
1. What is the output range? 2. Programming a linear output range 4. Related functions
The default value of the software defines the A linear output range is programmed by en- The EXA ISC402 has an output table which
output range of 4-20 mA = 0-1000 mS/cm. tering two values: can be programmed to give any non-linear
0% - the conductivity at the start of the output characteristic. Use of this table is de-
This display will always show the full range of output range which corresponds scribed in §7-5. After the table has been ac-
the instrument from 0-2000 mS/cm. Maxi- to 0.0 mA or 4.0 mA (see §7-4). tivated the linear RANGE routine will no
mum resolution is achieved by auto-ranging, 100% - the conductivity at the end of the longer have any effect and will not be acces-
where the position of the decimal point and output range which corresponds sible in the commissioning menu.
measuring units to 20 mA.
(microSiemens/milliSiemens) are moved to Programming of the 21 step table is de-
best fit the actual value. scribed in §7-6.
For control or recording the current output The non-linear output can be used to follow
can be ranged to a specific part of the total a concentration curve for a specific product.
measuring span. Care must be taken to ensure that the tem-
perature compensation is also set up for the
Linear ranges can be programmed in this measured fluid under representative condi-
routine. It is also possible to program non lin- tions. It should also be noted that conduc-
ear output ranges for specific applications, tivity measurements are not specific, they
these are programmed in the service level, can be influenced by impurities present or
see §7-5 and 7-6. changes in the fluid composition.
IM 12D8C2-E-H
32
6-3. SET UP HOLD FUNCTION
1. ACCESS HOLD ROUTINE 2. SELECT SET POINT TO ADJUST 3. SET VALUE TO HOLD
S3 S3 S3
> > >
>
>
>
ENT ENT ENT
FAIL/S4 FAIL/S4 FAIL/S4
Remove cover by releasing 4 screws Display shows actual status The display indicates the actual setting of
*H.OFF = Hold not activated *H.mA 1 in milliamps. This setting can be changed
⊗ Press this key for access to
commissioning mode
*H.ON = Hold activated using keys:
Activate (inhibit) HOLD function
> Select digit to adjust
Select HOLD menu
NO Change status
Adjust value
>
Confirm selection
Press NO-key until the display YES
NO indicates HOLD Confirm adjusted value and
Display shows current setting ENT
change to *H.mA 2.
*H.FIX = Hold fixed value
HOLD = HOLD menu
*H.LST = Hold last value
Adjust the value using the keys as above and
Confirm selection by pressing Select hold fixed or last press the ENT-key.
YES YES-key.
NO Change selection
MODE Return to measurement mode.
The display will indicate actual status Confirm selection. The display
YES
will indicate *H.mA 1 (*H.FIX) or
return to measurement mode
(*H.LST).
ESCAPE TO MEASURE can be used at any stage to abort operation. WARNING: If the HOLD function is activated, the instrument returns with
MODE the question HOLD (flashing); answer YES or NO or MODE again to return to measurement.
IM 12D8C2-E-H
33
6-3. SET-UP HOLD FUNCTION
1. What is HOLD? 2. How does it work? 3. Application example
HOLD is a function freezing the output sig- The HOLD-function has to be activated from In a storage tank the concentration of
nal temporary, during normal maintenance, this menu before it can be used. sodium hydroxide (caustic soda) has to
preventing unintended alarm or colour ac- be kept at 5%, the mixing process is
tions to occur. The instrument will keep the output frozen controlled by a conductivity transmitter
during the following events: and electromagnetic valves.
Two possibilities are generally used: - Access to the *-menu. During maintenance of the conductivity
- Keep the output at the LAST (H.LST) value - Access to the calibration mode. cell or the transmitter the HOLD function
just before the start of maintenance. This - Switching it from the MODE-menu. is activated to keep the mixing process
can only be used when a recorder is con- from wasting precious chemicals to the
nected. HOLD is signalled in the display by a special sewer.
- Keep the output at preset FIXED (H.FIX) field. The HOLD function is set for a fixed out-
value which will not cause any of the put so that the dosing valves will not be
alarms to go off or any controlling action to The operator is prompted to switch HOLD activated.
be taken. This is the preferred situation on or off before returning to normal meas-
when dealing with Conductivity-control urement. 4. Time out
systems. HOLD will be disabled after 10 minutes
if no key is pressed. To cancel this func-
In both situations the alarm contacts will tion refer to §7-10.
switch off and the proportional control (if ac-
tivated) will not be activated. The FAIL con-
tact will also not be activated.
IM 12D8C2-E-H
34
6-4. TEMPERATURE COMPENSATION
mS/cm
S3 S3 S3
> >
>
>
ENT > ENT
>
ENT
FAIL/S4 FAIL/S4 FAIL/S4
Remove cover by releasing 4 screws NaCl = Standard temperature compensation The display shows *WAIT for approximately 3 sec-
according to IEC tables. onds and then the coefficient can be set by adjust-
⊗ Press this key for access to
commissioning mode
MATRX = Selection or creation of programmed
compensation tables via §7-23 and
ing the conductivity value at process temperature
to the predetermined value at the reference
§7-24. temperature, normally 25 °C.
Select temperature compensa- T.C = Manually set temperature compensa-
tion menu tion factor by calibration or via §7-13. Adjustment of the value
NO Press NO-key until the display Select required temperature compensation > Select digit to adjust
indicates TEMP 1/TEMP 2
Change status
>
NO Adjust value
TEMP 1 = Temperature compensation menu mA 1. Press the YES-key to confirm
TEMP 2 = Temperature compensation menu mA 2. YES the selection ENT Confirm adjusted value.
NOTE: Only one output may have MATRX com- Error E2 indicates that the coefficient is outside the
Confirm selection by pressing
YES pensation selected, NaCl or T.C. are a permitted limits.
YES-key.
free choice.
The display will indicate actual status When T.C. is selected, it may be pro- MODE Return to measurement mode.
grammed in §7-13 or calibrated as fol-
lows.
ESCAPE TO MEASURE can be used at any stage to abort operation. WARNING: If the HOLD function is activated, the instrument returns with
MODE the question HOLD (flashing); answer YES or NO or MODE again to return to measurement.
IM 12D8C2-E-H
35
6-4. TEMPERATURE COMPENSATION
1. Why is temperature compensation 2. Standard temperature compensation 3. Manual setting of temperature com-
necessary? with EXA ISC402 pensation with EXA ISC402
The conductivity of a solution is very strongly From the factory EXA ISC402 instruments If the general compensation function is found
influenced by temperature. Typically for every are calibrated with a general temperature to be inaccurate for the sample in question,
1 °C change in temperature, the solution con- compensation function based on a sodium the instrument can be set manually for a lin-
ductivity will change by approximately 2%. chloride salt solution. This is suitable for ear factor on site to match the application.
The effect of temperature varies from one many applications and is compatible with
solution to another and is determined by sev- the compensation functions of typical labo- The procedure is as follows:
eral factors; solution composition, concentra- ratory or portable instruments. 1. Take a representative sample of the proc-
tion and temperature range. ess liquid during operation.
A coefficient (α) is used to express the amount A temperature compensation factor is de- 2. Heat or cool this sample to the reference
of temperature influence in % change in con- rived from the following equation: temperature of the instrument
ductivity/°C. (usually 25 °C).
In almost all applications this temperature in- Kt - Kref 100
fluence must be compensated before the α= X NOTE: See §7-14.
conductivity reading can be interpreted as an T - Tref Kref
accurate measure of concentration or purity. 3. Measure the conductivity of the solution
α = Temperature compensation factor with the ISC402 and note the reading of
(in %/°C) the display.
Table 6-1. NaCl compensation T = Measured temperature (°C) 4. Bring the sample to the typical process
according IEC 746-3 with Kt = Conductivity at T temperature. Check the temperature with
Tref = 25 °C Tref = Reference temperature the display routine.
Kref = Conductivity at Tref 5. Adjust the reading of the display to the
T Kt α T Kt α T Kt α previously noted value at the reference
temperature.
0 0.54 1.8 60 1.76 2.2 130 3.34 2.2
6. Check that the temperature compensa-
10 0.72 1.9 70 1.99 2.2 140 3.56 2.2 tion factor has been changed (from the
20 0.90 2.0 80 2.22 2.2 150 3.79 2.2 display routine).
7. Insert the conductivity cell into the proc-
25 1.0 -.— 90 2.45 2.2 160 4.03 2.3
ess again.
30 1.10 2.0 100 2.68 2.2 170 4.23 2.2
IM 12D8C2-E-H
36
5. Other possibilities
4. Calculating the approximate temper- 1. Enter calculated temperature coefficient,
ature coefficient Set the temperature coefficient to 1.30 on refer to §7-13.
If the temperature coefficient of a liquid to be the ISC402 converter (the meter can display 2. Enter matrix temperature compensation,
measured cannot be found, measure the to two places of decimals). refer to §7-24.
conductivity of the liquid at two tempera-
tures between 10 and 30°C with the tem- • Checking
perature coefficient set to 0.00 and use the When the temperature coefficient already
following equation to calculate an approxi- set is accurate, the conductivity to be dis-
mate temperature coefficient (α). played must be constant regardless of liq-
uid temperature. Make sure that the tem-
K2 - K1 perature coefficient already set is accurate.
Temp.coef.(α) = X100(%/°C) If, when the liquid temperature is lowered,
K1(t2 - 25) - K2 (t1 - 25)
a larger conductivity value is indicated, the
temperature coefficient already set is too
small.
where t1, t2 : liquid temperature (°C) The opposite also applies. If a smaller con-
K1 : conductivity at t1 (°C) ductivity value is indicated, the tempera-
K2 : conductivity at t2 (°C) ture coefficient already set is too large.
In this case, change the temperature coef-
Calculation example ficient so that the conductivity does not
Calculate the approximate temperature co- change.
efficient of a liquid from the following data.
Conductivity 124.5 µS/cm at a liquid tem- NOTE: When the ISC402 is used to deter-
perature of 18.0 °C and a conductivity of mine the temperature coefficient,
147.6 µS/cm at a liquid temperature of make sure that K1 and K2 are
31.0 °C. Substituting the data in the above measured with a temperature coef-
formula gives the following result. ficient value of 0.00
147.6 - 124.5
α= X100
124.5X(31.0-25) - 147.6x(18.0-25)
23.1
= X100
747.0 - (-1033.2)
= 1.298
IM 12D8C2-E-H
37
IM 12D8C2-E-H
38
6-5. AIR SET
S3 S3 S3
> > >
>
>
>
ENT ENT ENT
FAIL/S4 FAIL/S4 FAIL/S4
Remove cover by releasing 4 screws Display shows actual status The display shows calibrated zero
NO
⊗ Press this key for access to
commissioning mode YES
Press the YES-key to start the
AIR SET calibration
Repeat calibration.
Confirm calibration.
YES
Select AIR SET menu MODE Return to measurement mode.
Press NO-key until the display NOTE: If the sensor cannot be accurately calibrat-
NO
indicates AIR ed E9 will be displayed
ESCAPE TO MEASURE can be used at any stage to abort operation. WARNING: If the HOLD function is activated, the instrument returns with
MODE the question HOLD (flashing); answer YES or NO or MODE again to return to measurement.
IM 12D8C2-E-H
39
6-5. AIR SET
1. What is AIR SET procedure? 2. When is an AIR SET procedure nec- 3. How is the AIR SET procedure done?
The AIR SET procedure is a routine to com- essary? - Prepare the whole installation as de-
pensate for the small zero current due to the AIR SET is done only once during Start-Up, scribed in section 3.
inductive coupling of the sensor, cabling and and when a new sensor is connected to a - Do not install the sensor in the process,
transmitter. transmitter. AIR SET has to be done with a dry sen-
If the user wants to make full use of the sys- sor.
tems capabilities and requires accurate - Make sure that the sensor is in an EMI
measurement down to 0.5 µS/cm, this sim- and RFI free environment.
ple routine will optimize the performance. - Press "*" and 4 times "NO" until "AIR" is
indicated in the display.
- Press "YES" and after indication
"START", press "YES" again.
- Press "YES" after the indication "END".
- End of routine.
- If required, install the sensor.
IM 12D8C2-E-H
40
IM 12D8C2-E-H
41
7. CONFIGURATION IN SERVICE MODE
Code Function Use Default Section
Introduction 01 Temperature sensors Select sensor type and units 0.0 7-1
Usually there is no requirement to adjust the and units
settings of the service section. 02 Installation factor During start-up 1.88 cm-1 7-2
All parameters are pre-programmed to de- 03 Weight indication Read concentration directly 0 7-3
04 Output range Select output range 1.1 7-4
fault values which lets the instrument be put 05 Output function Conductivity (linear or table)
into service immediately. and/or temperature 0.0 7-5
06 mA table 1 Current output - 7-6
The advanced functions available through 07 mA table 2 Current output - 7-6
this section are only needed in some specif- 08 Contact S1 Program function 2.0 7-7
ic applications. This fine-tuning of the instru- 09 Contact S2 Program function 1.0 7-7
ment gives a superior performance over 10 Contact S3 Program function 0.0 7-7
conventional instruments. 11 Contact S4 Program function 3 7-7
12 Proportional range Set range 10 [%] 7-8
13 Proportional period Set period 10 [sec] 7-8
If it is necessary to adjust a function it is 14 Proportional frequency Set frequency 70 [pulses/min] 7-8
called-up with the appropriate code. After 15 Delay time and Delay time and 0.2 [sec] 7-9
selecting the code the function can be acti- hysteresis hysteresis 2.0 [%]
vated/(not activated) or the parameters 16 Auto return Time-out function 1 7-10
changed as required to suit the application 17 Burn function Signal fail on output 0 7-11
required. After any changes are made the 18 Temperature adjustment Adjust temperature offset 7-12
operator is returned to the entry point to 19 Temperature coefficient Adjust T.C. 2.1 [%] 7-13
20 Reference temperature Only if different from 25 °C 25/77 [°C/ °F] 7-14
make other adjustments or to return the in- 21 Alarm time-out Timer function 0 7-15
strument to the measure mode. 22 Set time Set current time - 7-16
23 Temperature hysteresis Set hysteresis temp. alarm 1.0/2.0 [°C/ °F] 7-17
If errors are made during the programming 24 Enable Setpoint adjust in Allow access to
process, these will be indicated, no action maintenance SETPOINT function 0 7-18
will be taken and the operator can start the 25 Deactivate Error E6 Disable low conductivity error 1 7-19
programming again. 26 Soft fail Programme errors to signal fail 1 [all codes] 7-20
27 Communication setup Set-up comm link to P.C. 0.1 7-21
28 Logbook Clear logbook 7-22
30 Select matrix Select matrix temperature
compensation - 7-23
31-36 Program matrix User program matrix - 7-24
50 Test Built in test - 7-25
55 Defaults Restore default values - 7-26
66 Passcodes Protect data 0.0.0 7-27
IM 12D8C2-E-H
42
7-0. ACCESS TO SERVICE SETTING
S3 S3 S3
> > >
>
>
>
ENT ENT ENT
FAIL/S4 FAIL/S4 FAIL/S4
Remove cover by releasing 4 screws Display indicates CODE Display indicates current setting.
Enter access code to select function Adjust the setting in accordance with the instruc-
⊗ Press this key for access to com-
missioning mode > Select digit to adjust
tions in the appropriate section §7-1 to §7-27.
>
cates SERV. ENT Confirm adjusted value
ENT Confirm adjusted value
SERV = Service menu
MODE Return to measurement mode
Confirm selection by pressing
YES YES-key.
ESCAPE TO MEASURE can be used at any stage to abort operation. WARNING: If the HOLD function is activated, the instrument returns with
MODE the question HOLD (flashing); answer YES or NO or MODE again to return to measurement.
IM 12D8C2-E-H
43
7-1. Temperature sensor & units 7-2. Installation factor adjustment constant of the contact electrode conductiv-
ACCESS CODE : 01 ACCESS CODE : 02 ity sensors. This factor is 1.88/cm for the
(see §7-0 for operation) (see §7-0 for operation) ISC40 sensor if installed with minimum of
DISPLAY : *T.CODE DISPLAY : *I.F. 25 mm of process fluid around the dough-
Adjustment : X.X nut.
Adjustment:
X.X Adjust the ratio between the measured con- Installed in stainless steel flow fitting (Model
X.0 Temperature in °C ductance of the sensor and the specific ISC40FF-S) the factor for ISC40 is 1.88.
X.1 Temperature in °F conductivity of the solution. Installed in a PP or PVDF flow fitting (Model
0.X NTC ISC40FF-P or F) the factor is 1.88.
1.X PT1000 Explanation: The factor is a function of the distance be-
The Installation factor for the ISC40 is the ra- tween sensor and the surrounding piping
Explanation: tio between the measured conductance of and is given in the figure below.
The code for the temperature units deter- the sensor and the specific conductivity of
mines the temperature indication on the dis- the solution. The unit is /cm just as the cell Default: 1.88/cm
play. The sensor selection chooses between
the integral (NTC) sensor, and external PT
1000 Ω RTD. 2.4
INSTALLATION FACTOR
Installation factor in cm-1
2.3
Default: NTC. °C.
2.2
* put PT 1000 Ω RTD on one line
non conductive
2.1 piping
D
2.0
1.9
1.88
1.8
conductive
piping
1.7
1.6
0 10 20 30 40 50
D in millimeters
Figure 7-1.
IM 12D8C2-E-H
44
* 0%: Adjust low span display value; Default: 1.1 = 4-20 mA for mA 1 and mA 2.
*100%: Adjust high span display value.
IM 12D8C2-E-H
45
7-5. Output function 7-6. Output table for non linear range
ACCESS CODE : 05 ACCESS CODE : 06 (mA 1) and Explanation:
(see §7-0 for operation) 07 (mA 2) (see NOTE) An example will be used to explain the pro-
(see §7-0 for opera- cedure.
DISPLAY : OUTP. F tion) Sulphuric acid in the range of 0-25% and a
reference temperature of 25 °C (see litera-
Adjustment : X.X DISPLAY : *0% to 100% ture for conductivity curves at 25 °C).
1. Define minimum value (0% sulphuric
X.X Adjustment : acid = 0 mS/cm).
0.X Linear range for mA 1. A 21 step table can be programmed from 2. Define maximum value (25% sulphuric
1.X Output table for mA 1. this entry ONLY if it was selected in code acid is 791 mS/cm).
X.0 Linear range for mA 2. 05. It is possible to set up a linear output to 3. Divide range by 20 and calculate con-
X.1 Output table for mA 2. concentration units. ductivity values (each point is 25/20 =
X.2 Temperature output for 1.25 % by weight).
mA 2. NOTE: Access to codes 06 & 07 is only 4. Fill in the table.
possible if "1" is programmed in
Explanation: code 05. NOTE: Sometimes not all of the 21 values
As a default, a linear output is set and only are known. If an unknown value is
0% and 100% can be adjusted from the Other possibilities: prompted, press the NO-key. The
commissioning output function. Programming "% by weight" indication on converter will step to NEXT value
When an output table is programmed by second display line, refer to §7-3. and find the unknown value by in-
setting a 1, the table can be programmed terpolation. Alternatively you can
through code 06. set up a bi-linear, logarithmic or
If 1 is not selected code 03 cannot be hyperbolic scale: examples to pro-
accessed. See also codes 06 an 07. gram these scales are on pages
The actual process temperature can be 46.
transmitted on mA 2.
IM 12D8C2-E-H
46
% Output mA Concentration Example Conductivity Example
OUTPUT LINEAR TO CONCENTRATION
EXAMPLE 0-25% SULFURIC ACID 0 4 0 0
5 4.8 1.25 60
Conductivity(mS/cm) Output in % 10 5.6 2.5 113
1000 100
15 6.4 3.75 180
800 80
20 7.2 5 218
600 60
Conductivity 25 8 6.25 290
400 40 % Output 30 8.8 7.5 335
200 20 35 9.6 8.75 383
0 0 40 10.4 10 424
0 2 4 6 8 10 12 14 16 18 20 22 24
Concentration(% by wt)
45 11.2 11.25 466
50 12 12.5 515
55 12.8 13.75 555
Figure 7-2.
60 13.6 15 590
65 14.4 16.25 625
70 15.2 17.5 655
75 16 18.75 685
80 16.8 20 718
85 17.6 21.25 735
90 18.4 22.5 755
95 19.2 23.75 775
100 20.0 25 791
IM 12D8C2-E-H
47
IM 12D8C2-E-H
48
7-8. Settings for proportional control % Controller output % Controller output
Explanation: Figure 7-3. Pulse frequency control Figure 7-4. Duty cycle control
It is only useful to set these values when one
or more contacts will be used for proportion- set when the pulse frequency control is pro- Defaults:
al control. The selection of the proportional grammed. The maximum number of pulses Proportional range 10%
control is described in §7-7. can be set between 50 to 120 pulses per Limits 0-100%
Whether you have selected proportional minute. The number of pulses is controlled Proportional period 10 seconds
duty cycle control or proportional pulse fre- between zero (0) and the programmed maxi- Limits 5-100 seconds
quency control, you have to program the mum by the deviation from the setpoint with- Proportional frequency 70 per minute
proportional range with access code 12 for in the proportional range, refer to Fig. 10-1. Limits 50-120 per minute
both. This determines the range in which
the control takes place and changes the NOTE: Proportional control only for contacts
control output between 0% and 100%. output allocated to conductivity.
IM 12D8C2-E-H
49
7-9. Setting the process alarm function 7-10. Auto return function
ACCESS CODE : 15 NOTE: For hysteresis on temperature ACCESS CODE : 16
(see §7-0 for operation) contact function, refer to service (see §7-0 for operation)
DISPLAY : *D.TIME and *HYST code 23 (§7-17). DISPLAY : *RET
t (sec)
delay delay
time time
IM 12D8C2-E-H
50
Default:
0 = OFF
IM 12D8C2-E-H
51
IM 12D8C2-E-H
52
7-14. Reference temperature 7-15. Timer on contact function 7-16. Setting the clock
ACCESS CODE : 20 ACCESS CODE : 21 ACCESS CODE : 22
(see §7-0 for operation) (see §7-0 for operation) (see §7-0 for operation)
DISPLAY : *T.R.°C or *°F DISPLAY : *EXPIR/*tE.min DISPLAY : *HOUR
Adjustment : Adjust the value for the refer- Adjustment : (X) Adjustment:
ence temperature used in the calculations. Set the internal clock to the current time.
Normally this value will be 25 °C. X Use the >, and ENT keys to program the
>
0 No timer current time. The sequence is Hours, Min-
Explanation: 1 Timer function activated utes, Seconds, Year, Month, Day. Press
All measured values are compensated for enter after each operation.
temperature to refer to comparable values at When 1 is selected, use the > and
25 °C. This temperature is easily produced keys to program the contact action ex-
>
Explanation:
at laboratory conditions and most literature piry time in minutes. This function is used as an event time out-
data use this temperature. put for the logbook.
In exceptional cases it is possible to use a
different reference temperature in combina- Explanation: NOTE:
tion with a programmed temperature coeffi- When the contacts are used for a control 1. A battery is incorporated in the instru-
cient (e.g.Sulphuric acid measurement). function contact action should have a result. ment to support the clock function if
power is removed from the instrument.
Default: 25 °C/77 °F If the result (measurement within a safe 2. The battery should be changed every 5
Limits: 0-100 °C / 30-225 °F band) does not take place within a set expi- years or if the time function does not
ry time, error E22 is triggered (see also (§8- keep correct time. Contact your local
4)). Yokogawa Service Centre for details.
The error code E22 can be cancelled by 3. The battery is a Lithium type, it should be
pushing the [YES] or [NO] keys. The timer is disposed of in accordance with the local
reset. health and safety regulations.
Default: 0 = No timer
IM 12D8C2-E-H
53
7-17. Settings for the temperature hys- 7-18. Enable Setpoint adjustment in 7-19. Deactivate Error E6
teresis maintenance ACCESS CODE : 25
ACCESS CODE : 23 ACCESS CODE : 24 (see §7-0 for operation)
(see §7-0 for operation) (see §7-0 for operation) DISPLAY : EN E6
DISPLAY : *T.HYST DISPLAY : *MODE
Adjustment: (X)
Adjustment: Adjustment: (X)
Settings the hysteresis for setpoint of con- 0 Error E6 disabled
tact allocated to process temperature. X 1 Error E6 enabled
0 Setpoints programmable at the
Explanation: commissioning level only ( -key)
When dealing with a high [low] alarm the re- 1
*
Setpoints also programmable at
Explanation:
Error E6 signals loss of sample (i.e. cell not
lay will switch back after the process value the maintenance level wetted). As this is a normal state in some
has passed the setpoint minus [plus] hyster- (MODE key) batch processes, E6 may be used to signal
esis. The hysteresis (dead-band) can be ad- a fault condition in the absence of sample,
justed in absolute units for temperature Explanation: or not, to suit the process.
measurement. Normally setpoints of the contact outputs
A specific setting of the hysteresis is valid for can only be adjusted on commissioning lev-
all contacts allocated to that specific param- el to get access, the cover has to be
eter. removed. For convenience the setpoints
adjustment routine can be made available
Default Limits on maintenance level; access through cover
1.0 °C 0.1-5.0 °C window.
2.0 °F 0.2-10.0 °F
Default: 0 = Setpoint setting not available
NOTE: The default is dependent on the set- on maintenance level.
ting in Service Code 01.
IM 12D8C2-E-H
54
IM 12D8C2-E-H
55
IM 12D8C2-E-H
56
Adjustment:
Adjust this value to the upper limit of the re-
IM 12D8C2-E-H
57
mS/cm
programming easier for duplicate sys- 600
tems or in case of data loss.
x
400
NOTES:
1. E17 occurs when a temperature range of 200 x
less than 50 °C/ 90 °F is entered, or if the
output span is <90% of the maximum 0
range. 0°C 25°C 50°C 75°C 100°C
2. Each matrix column has to increase in
conductivity value (for example L1xTx < Temperature (Celsius)
L2xTx < L3xTx < L4xTx < L5xTx). x 15%
3. E4 occurs when two standard solutions
1% 3% 6% 10%
have identical conductivity values at the Figure 7-6.
same temperature within the tempera-
ture range.
Matrix example actual example actual example actual example actual example actual
IM 12D8C2-E-H
58
gled and the LED will switch state. THORITY AS ALL SETTINGS, PROGRAMMED FUNC-
Use the > key to select * S2, S3 and S4 TIONS AND TABLES WILL BE LOST. IT IS NOT POSSI-
and repeat the test. BLE TO RETRIEVE THE DATA AFTER YES HAS BEEN
After testing S4, Press ENT. This *COMM PRESSED.
tests the communications function.
Press ENT to exit TEST mode. *SERV The purpose of this function is to reset the in-
strument to be moved to a new location or be
used for a different process with the factory
defaults being used as a starting point.
CAUTION
Do not enter service codes that are not listed
in this manual.
Entering invalid service codes can result in a
loss of factory set data.
IM 12D8C2-E-H
59
IM 12D8C2-E-H
60
8. TROUBLE SHOOTING
8-1. Introduction 8-2. Self diagnostics of the electronics 8-3. Checking during operation
The EXA ISC402G microprocessor based The microprocessor operation is checked Whenever the instrument is being pro-
conductivity analyzer continuously monitors by a watchdog which initiates an electronic grammed or calibrated, data is checked and
the condition of all key components of the reset if the normal functions suffer severe in- an error is shown when appropriate. Should
measuring system to ensure that measure- terference. During reset the instrument this occur the new data is rejected and the
ment is reliable. If a fault is detected it is im- checks the program and all stored data. If a instrument continues to work with the previ-
mediately signalled. Errors are shown on the fault is then detected an alarm is given. ous settings.
display with a code. The following table
shows the errors which can be detected
and gives information to help locate the fault
or identify the error. Faults detected while
the instrument is on line can also be sig-
nalled by a (temporary) high output (22 mA)
signal, refer to §7-20.
IM 12D8C2-E-H
61
8-4. Error messages and explanation
E2 Temperature coefficient incorrect. Calculated TC not between 0.00 and 3.50%/ºC. See §7-13
TC too high for measured temperature.
E3 Installation Factor out of range. Wrong unit (mS or µS) See §7-2
Defective sensor Replace sensor
Standard error Check standard
E4 Impossible program for Temperature Incorrect in 5 x 5 matrix (See codes 32-36) See §7-24
Compensation
E15 Impossible adjustment temperature Corroded contacts-wrong sensor programmed Clean and reterminate-Reprogram see §7-13
E18 Impossible program for output table Incorrect data in code See §7-6
E19 Programmed values not accepted Values exceed pre-set limits Try again, read instructions
IM 12D8C2-E-H
62
E20 Data integrity not good. Unauthorized programming The unit has to be re-initialized
Software problem Call Yokogawa
E21 Corrupted EPROM Software failure Call Yokogawa
E22* Expired contact action interval Ineffective process control Check control loop
NOTE: If abnormal values are displayed, usually in combination with E5, E6, E7 or E8 messages, then check with your local Yokogawa
Service Centre.
NOTE: If abnormal readings are displayed the process can be simulated by a loop through the dry doughnut to a resistance box (Yokogawa
QT40 sensor simulator). The reading should agree with IF. 1000/R in µS/cm in which R = Resistance in kOhm and IF = programmed
Installation Factor, typically 1.88
IM 12D8C2-E-H
63
SPARE PARTS
Housing K1541JJ 7
Fuse 230 V AC K1543AK
115 V AC K1543AL
100 V AC K1543AL
24 V DC K1543AM 8
Hingepin K1543KS 12
IM 12D8C2-E-H
64
EXPLODED VIEW
11
2
12
10
5 7
1 8 9
13 6
ERROR MESSAGES
CODE DESCRIPTION
E2 TEMPERATURE COEFFICIENT INCORRECT
E3 INSTALLATION FACTOR OUT OF RANGE
E4 IMPOSSIBLE PROGRAM FOR TEMPERATURE COMPENSATION
E5 CONDUCTIVITY TOO HIGH
E6 CONDUCTIVITY TOO LOW
E7 TEMPERATURE TOO HIGH
E8 TEMPERATURE TOO LOW
E9 NO AIR SET POSSIBLE
E10 WRITING TO EEPROM FAILED
E15 TEMPERATURE CORR. OUT OF LIMITS
E17 SPAN ERROR
E18 TABLE VALUES MAKE NO SENSE
E19 PROGRAMMED VALUES NOT ACCEPTABLE
E20 ALL PROGRAMMED VALUES ARE LOST
E21 CORRUPTED EEPROM
E22 EXPIRED CONTACT ACTION INTERVAL
IM 12D8C2-E-H
CODED SERVICE SETTINGS (DEFAULTS)
CODE DISPLAY FUNCTION & USE DEFAULT SECTION SETTING #1 SETTING #2 SETTING #3
01 T.CODE Select temperature sensor and units NTC °C 7-1
02 I.F. Set installation factor (calibrate) 1.88 cm-1 7-2
03 % Set for display in Wt % units 0.0 7-3
04 mA Set 0-20 or 4-20 for mA 1 & mA 2 1.1 7-4
05 TABLE Output selection - linear SC / table - %
concentration/temperature 0.0 7-5
06 0% / 100% Set mA output linearization mA 1 7-6
07 0% / 100% Set mA output linearization mA 2 7-6
08 S1 Set relay alarm/control operation 2.0 7-7
09 S2 Set relay alarm/control operation 1.0 7-7
10 S3 Set relay alarm/control operation 0.0 7-7
11 S4 Set relay FAIL/alarm/control operation 3 7-7
12 RANGE Set proportional range for relay control 10% 7-8
13 PER Set proportional control period 10 Sec 7-8
14 FREQ Set proportional control frequency 70 pulses/min 7-8
15 D.TIME Set relay delay time 0.2 Sec 7-9
HYST Set relay hysteresis 2.0%
16 RET Select auto return on/off 1 7-10
17 BURN Select for 22mA on FAIL alarm 0 7-11
18 T. ADJ Calibrate temperature sensor 7-12
19 T.C. 1 / T.C. 2 Set TC 1 and/or TC 2 2.1% 7-13
20 T.R. °C / °F Set reference temperature if different from 25°C 25°C/77°F 7-14
21 EXPIR Set alarm control action time out, and period 0 7-15
22 HOUR/MINUTE/ Set date and time
SECOND/YEAR/
MONTH/DAY 7-16
23 T.HYST Set hysteresis for the temperature alarm. 1.0 °C/2.0°F 7-17
24 MODE Enable setpoint adjustment in maintenance mode 0 7-18
25 EN. E6 Enable E6 error signal for empty cell. 1 7-19
26 EN. E(5/6/7/8/22) Set errors that will trigger FAIL contact 1 (all codes) 7-20
*SOFT Disable FAIL Contact for soft fail 0
27 COMM Set up RS485 link to PC 0.1 7-21
28 LOG.CLR Clear logbook 7-22
30 MATRIX Select which matrix temperature compensation 7-23
31 T1....T5 °C / °F Set matrix temperature range 7-24
32 L1XT1......L1XT5 Set matrix conductivity values (1) 7-24
33 L2XT1......L2XT5 Set matrix conductivity values (2) 7-24
34 L3XT1......L3XT5 Set matrix conductivity values (3) 7-24
35 L4XT1......L4XT5 Set matrix conductivity values (4) 7-24
36 L5XT1......L5XT5 Set matrix conductivity values (5) 7-24
50 TEST Built -in test function 7-25
55 ERASE Restore default settings 7-26
66 PASS Set password protection codes. 0.0.0 7-27
EUROPEAN HEADQUARTERS BELGIUM HUNGARY AUSTRALIA UNITED STATES OF AMERICA
Yokogawa Europe B.V. Yokogawa Belgium N.V./S.A. Yokogawa Hungaria Ltd. Yokogawa Australia Pty Ltd. Yokogawa Corporation of
Databankweg 20 Minervastraat 16 Alkotas Center 39 C Private mail bag 24 America
3821 AL AMERSFOORT 1930 ZAVENTEM 1123 BUDAPEST Centre Court D3 2 Dart Road
The Netherlands Tel. +32-2-719 55 11 Tel. +36-1-355 3938 25-27 Paul Street North NEWNAN, GA 30265-1040
Tel. +31-33-4641 611 Fax +32-2-725 34 99 Fax +36-1-355 3897 NORTH RYDE, N.S.W. 2113 Tel. +1-770-253 70 00
Fax +31-33-4641 610 Tel. +61-2-805 0699 Fax +1-770-251 28 00
E-mail: info@yokogawa.nl FRANCE ITALY Fax +61-2-888 1844
www.yokogawa-europe.com Yokogawa France S.A. Yokogawa Italia S.r.l. ISO 9001
Vélizy Valley Vicolo D. Pantaleoni, 4 SINGAPORE
THE NETHERLANDS 18-20 Rue Grange Dame Rose 20161 MILANO Yokogawa Engineering Asia Pte. Ltd.
Yokogawa Nederland B.V. 78140 VELIZY VILLACOUBLAY Tel. +39-02-66 24 11 11, Tampines Street 92
Hoofdveste 11 Tel. +33-1-39 26 10 00 Fax +39-02-645 57 02 SINGAPORE, 528872
3992 DH HOUTEN Fax +33-1-39 26 10 30 Tel. +65-783 9537 CERTIFICAT-
Tel. +31-30-635 77 77 SPAIN/PORTUGAL Fax +65-786 2606 ED
Fax +31-30-635 77 70 GERMANY Yokogawa Iberia S.A.
Yokogawa Deutschland GmbH C/Lezama, 22 JAPAN CENTRAL/EAST REGION
AUSTRIA Berliner Strasse 101-103 28034 MADRID Yokogawa Electric Corporation Via Yokogawa Ges.m.b.H.
Yokogawa Austria Ges.m.b.H. D-40880 RATINGEN Tel. +34-91-771 31 50 2-9-32, Nakacho, Musashino-shi Czechia, Slovakia, Poland,
Franzensbrückenstrasse 26 Tel. +49-2102-4983 0 TOKYO, 180 Croatia, Slovenia, Jugoslavia,
A-1021 WIEN Fax +49-2102-4983 22 UNITED KINGDOM Tel. +81-422 52 5617 Bulgaria, Romania, Macedonia,
Tel. +43-1-2165 043 0 Yokogawa United Kingdom Ltd. Fax +81-422 52 0622 Bosnia & Herzegovina
Fax +43-1-2165 043 33 NORTHERN EUROPE Stuart Road, Manor Park,
Yokogawa Nordic A.B. RUNCORN SOUTH AFRICA Distributors in:
Finlandsgatan 52, 2fl Cheshire WA7 1TR Yokogawa South Africa (Pty) ltd. Denmark, Finland, Greece,
SE-164 74 Kista Tel. +44-1-928 597100 67 Port Road, Robertsham Norway, Portugal, Russian
STOCKHOLM Fax +44-1-928 597101 Southdale 2135, JOHANNESBURG
YOKOGAWA Tel. +46-8-477-1900
Fax +46-8-477-1999
Tel. +27-11-680-5420
Fax +27-11-680-2922
Federation, Sweden,
Switzerland and Turkey.