Professional Documents
Culture Documents
DKE291 Ch04
DKE291 Ch04
PRINCIPLES OF OPERATION
Thermogravimetric analysis (TGA) is defined as the study of the change in
mass as a function of temperature, time, and/or atmosphere. TGA consists
of two essential components, mass and temperature detection.
The mass is measured using a balance. A balance uses the physical
principle of leverage: a sample is placed at the end of the arm while placing
a reference weight in the opposite end and the weight is recorded as the
change in force. In the TGA, the mass detection is based on the principle of
substitution weighing introduced by Mettler in 1945 (Fig. 1). In the TGA
shown in Fig. 1, the calibration weights and the sample are suspended from
a single arm. The calibration weight is substituted as the sample mass sub-
stitutes for it. The weighing principle is still leverage.
Most modern TGAs, however, in addition incorporate the principle of
electromagnetic balances that have relatively little dependence on vibration
(one of the common problems for weight measurements), have high sensi-
tivity, and display little thermal drift. The beam position is monitored by a
photodetection scheme. This concept was introduced by Cahn Instruments.
As shown in Fig. 2, after taring the sample, the balance is assumed to be in
equilibrium. Addition of the sample to the left side of the beam will cause
the right side of the beam to be displaced upward. Sufficient current is
CALIBRATION TECHNIQUES
All TGA instruments have the capability of being calibrated for mass and
temperature. The mass calibration usually consists of placing a standard
with a specific traceable NIST (National Institute of Standards, formerly
the NBS) weight, ranging between 50 and 100 mg. The scale then measures
the weight of the standard and by placing the expected weight it then adjusts
the tension in the weighing arm.
The recommended calibration procedure for a TGA is as follows:
CaC2 O4 H2 O!CaC2 O4 þ H2 O
CaC2 O4 !CaCO3 þ CO
Alumel 154.2
Nickel 355.2
Perkalloy 596
Iron 780
Hisat-50 1000
The procedure is to place a sample of CaC2O4 H2O and set the instru-
ment to run between 50 and 850C at a rate of 158C/min (Table 3 and Fig. 5).
APPLICATIONS
Application 1: Determination of the Degradation Temperature
The degradation temperature is the temperature at which a plastic reverts to
its original component(s). This is also known as the ceiling temperature. The
importance of the degradation temperature is that it helps to determine the
maximum limits at which the plastic will no longer be a plastic.
There are two TGA techniques to determine the degradation tempera-
ture of a plastic. The first is called the isothermal method and is based on
setting the sample on a temperature that is between 50 and 150C above its
melting temperature and letting the instrument measure the weight loss with
time. This is one technique used to determine physical aging of a sample
under extreme conditions. A typical decomposition curve is shown in Fig. 6.
The second technique is based on running the sample under an increas-
ing temperature pattern while measuring the weight loss. This is known as a
nonisothermal degradation test. This technique is used when the decomposi-
tion temperature of the sample is not known. Figure 7 shows the use of this
1 12.33
2 21.88
3 44.00
Overall 61.64
dry nitrogen atmosphere to run the test for a period between 1 and 2 h.
Ensure that the calibration procedure has been done before starting this
measurement. A typical example is shown in Fig. 9.
the quantities of the original components are determined but also the types
of components. Figure 10 shows a typical instrument setup to run the TGA/
GC-MS technique, Fig. 11 shows the connection between the TGA and GC-
MS unit, and Fig. 12 shows typical GC-MS scans done on TGA polymer
samples.
create a material whose properties are better than those of the individual
components.
The procedure is to run the sample under a heating rate of 20–308C
from around room temperature to 100C above the component having the
highest decomposition temperature. With the help of Table 4, the compo-
nents can be identified. This technique is approximate, and the best way to
determine the types of components present in a sample is by using DSC and
FTIR testing along with the TGA (Fig. 15).
ACKNOWLEDGMENTS
We are grateful to Mr. James Creedon and Dr. Roger Blaine, TA
Instruments; Mr. Jon Foreman, Mettler Corp.; Ms. Karen Gillette, Perkin
Elmer Instruments; and Mr. Tom Wampler, CDS Analytical.
Hyphenated Methods
Charsley, E. L., Warrington, S. B., Jones, G. K., and McGhie, A. R., TGA-MS
Using a Simple Capillary Interface. ‘‘Am. Lab., 1990.
Mettler Toledo Technical Notes.
Nicolet Instruments Technical Notes.
Perkin Elmer Technical Notes.
TA Instruments Technical Notes.