Download as pdf or txt
Download as pdf or txt
You are on page 1of 112

Go To Table Of Contents

SpeedClave®
Steam Sterilizers
Model Numbers: Serial Number Prefixes:
Service and
M7 -011 thru -016 MH, MJ, MK, ML, MM, MN, V Parts Manual
M7 -020 thru -022 V
To purchase a printed copy of this manual,
click on the "Place Order" button below.

Place Order

FOR USE BY MIDMARK TRAINED TECHNICIANS ONLY


SF-1855 Part No. 004-0454-00 Rev. D (01/05/11)

Style B
Back Next
GENERAL INFORMATION ACCESS PROCEDURES

General Information

Section C
Symbols .................................... i Removing & Installing:
Ordering Parts .......................... i Covers / Panels ........................ C-2
Model / Serial Number Tray Plate / Rack ...................... C-3

Table Of Contents
Location .................................. i Draining / Filling Reservoir ..... C-4
Weights, Dimensions,
Electrical Specifications ....... ii
Model Identification /
Compliance Chart .................. iii
Special Tools ............................ iv WIRING DIAGRAMS & SCHEMATICS

Section D
Warranty Information ............... iv 115 VAC models:
M7 (-011 / -013 / -014 / -015) ........ D-2
M7 (-020 / -022) ............................ D-3

OPERATION & 230 VAC models:


TROUBLESHOOTING M7 (-012 / -016) ............................ D-4
Section A

M7 (-021) ...................................... D-5


Electrical System:
M7 (-011 thru -016) ....................... A-2
M7 (-020 thru -022) ....................... A-4
Filling the Chamber ................. A-8
Heat-Up / Sterilization .............. A-14 EXPLODED VIEWS / PARTS LISTS

Section E
Venting the Chamber ............... A-20 M7 (-011 thru -016) .................... E-2
M7 (-020 thru -022): ................... E-3

TESTING & REPAIR


Checking for
Section B

Pressure Leaks ...................... B-2


Fuse ........................................... B-3
Bellows ..................................... B-4
Fill / Vent Valve:

Section E
Manual .......................................... B-6
Electronic ...................................... B-8
Temperature Regulator Assy .. B-12
Heating Element ....................... B-18
Overheat Thermostats ............. B-22
Pressure Relief Valve ............... B-25
Timer ......................................... B-26
Timer Buzzer ............................ B-30
Temperature Gauge ................. B-31
Door Assembly ......................... B-32
Reservoir Tank .......................... B-34
Chamber Assembly ................. B-36
Back Go To Table Of Contents Next

General Information

General Information
Symbols Model / Serial Number Location

Caution
Indicates a potentially hazardous situation
which could result in injury if not avoided.

Equipment Alert
Indicates a potentially hazardous situation
which could result in equipment damage if not avoided. R

Note
Amplifies a procedure, practice, or condition.
Serial Number
(all models)
Indicates that the component the check mark appears

3 beside should be tested before replacing it.


In Section A, test the components in the order indicated.
(ex. 1st 3 then, 2nd 3 )

Refer to Section B for component testing procedures.

Model & Serial Number


Ordering Parts

The following information is required when ordering parts:


• Serial number & model number
• Part number for desired part.
[Refer to Section E: Exploded Views / Parts Lists]
M7 (-011 thru -016) M7 (-020 thru -022)
Non-warranty parts orders may be faxed to Midmark using Location: bottom of unit Location: back of unit
the Fax Order Form in the back of this manual.

For warranty parts orders, call Midmark's Technical Service MA511502i


Department with the required information.
Hours: 8:00 am until 5:00 pm EST [Monday - Friday]
Phone: 1-(800)-Midmark

© Midmark Corporation 2004 SF-1854


i
Back Go To Table Of Contents Next

General Information
Weights, Dimensions, Electrical Specifications
R

A
ATTENTION
A separate (dedicated) electrical circuit is
recommended for all models.
Do not connect to a circuit with other devices,
unless the circuit is rated for the additional load.

C
B
M7 (-011 thru -016)
Dimensions [Refer to illustration]:
Height (A) ........................................... 12.8 in. (32.5 cm)
Width (B) ............................................ 13.5 in. (34.3 cm) M7 (-020 thru -022)
Depth (C) ........................................... 18.1 in. (46 cm)
Dimensions [Refer to illustration]:
Chamber Size: ..................................... Diameter: 7.5 in. (19 cm) Height (A) ........................................... 13 in. (33 cm)
Depth: 14.25 in. (36.2 cm) Width (B) ............................................ 14 in. (35.6 cm)
Shipping Carton: Depth (C) ........................................... 19 in. (48.3 cm)
(Length x Width x Height) .................. 24 in. x 16 in. x 16 in.
(61 cm x 40.6 cm x 40.6 cm) Chamber Size: ..................................... Diameter: 7.5 in. (19 cm)
Weight: Depth: 14.25 in. (36.2 cm)
Shipping Weight ................................. 39 lbs (17.7 kg) Shipping Carton:
w/reservoir empty .............................. 30 lbs (13.6 kg) (Length x Width x Height) .................. 24 in. x 16 in. x 16 in.
w/reservoir full ................................... 41.8 lbs (19 kg) (61 cm x 40.6 cm x 40.6 cm)
Weight:
Reservoir Capacity: ............................ Approx. 1.3 gallon (4.9 liters) Shipping Weight ................................. 39 lbs (17.7 kg)
at FULL mark w/reservoir empty .............................. 30 lbs (13.6 kg)
Pressure Relief Valve: w/reservoir full ................................... 41.8 lbs (19 kg)
opens at approximately: .................... 34 psi (234 kPa)
Reservoir Capacity: ............................ Approx. 1.3 gallon (4.9 liters)
Electrical Requirements: .................... [See Model Identification / at FULL mark
Compliance Chart] Pressure Relief Valve:
Power Consumption: opens at approximately: .................... 34 psi (234 kPa)
100 VAC models ................................ 1150 watts, 12 amps @ 100 VAC
115 VAC models ................................ 1150 watts, 10 amps @ 120 VAC Electrical Requirements: .................... [See Model Identification /
230 VAC models ................................ 1150 watts, 5 amps @ 240 VAC Compliance Chart]
Power Consumption:
115 VAC models ................................ 1300 watts, 10 amps @ 115 VAC
230 VAC models ................................ 1300 watts, 5 amps @ 230 VAC

Fuse (back of unit):


115 VAC models ................................ 12 amp, 250 V, Fast-Acting, 1/4" x 1-1/4"
230 VAC models ................................ 8 amp, 250 V, Fast-Acting, 5 x 20 mm
ii © Midmark Corporation 2004 SF-1854 Rev.12/04
Back Go To Table Of Contents Next

General Information
Model Identification / Compliance Chart

Complies To: Electrical Ratings:


Serial
Model Description Number UL CAN/CSA
CAN/CSA
Prefixes UL C22.2,
61010A-1 C22.2,
544 #1010 Cycles
61010-2-041 #151
#1010.2-041-96 VAC Amps (Hz)
Ritter M7 Sterilizer
M7-011
(115 VAC)
MH & V x x 115 10 60

Midmark M7 Sterilizer
M7-012
(230 VAC)
MJ & V x x 220 / 240 5 50

Midmark M7 Sterilizer
M7-013
(100 VAC)
MK & V x x 100 12 60

Midmark M7 Sterilizer
M7-014
(115 VAC)
ML & V x x 115 10 60

Dabi Alante M7 Sterilizer


M7-015
(115 VAC)
MM & V x x 115 10 60

Dabi Alante M7 Sterilizer


M7-016
(230 VAC)
MN & V x x 230 5 60

Midmark M7 Sterilizer
M7-020
(115 VAC)
V x x 115 10 60

Midmark M7 Sterilizer
M7-021
(230 VAC)
V x x 230 5 50

Ritter M7 Sterilizer
M7-022
(115 VAC)
V x x 115 10 60

© Midmark Corporation 2004 SF-1854


iii
Back Go To Table Of Contents Next

General Information
Special Tools
This table lists all special tools needed to diagnose and repair the sterilizer.

Special Tool Manfacturer Part Number Purpose of Tool

Digital Multimeter Commercially available any type To perform continuity / voltage checks

Digital Thermometer Commercially available any type To verify chamber temperature

Warranty Information
SCOPE OF WARRANTY ADDITIONAL INFORMATION
Midmark Corporation (“Midmark”) warrants to the original purchaser its new Alternate Care Failure to follow the guidelines listed below will void the
products and components (except for components not warranted under “Exclusions”) manufac- warranty and/or render the table unsafe for use.
tured by Midmark to be free from defects in material and workmanship under normal use and
service. Midmark’s obligation under this warranty is limited to the repair or replacement, at • If a malfunction is detected, do not use the table
Midmark’s option, of the parts or the products the defects of which are reported to Midmark until necessary repairs are made.
within the applicable warranty period and which, upon examination by Midmark, prove to be
defective.
APPLICABLE WARRANTY PERIOD
• Do not attempt to disassemble table, replace
The applicable warranty period, measured from the date of delivery to the original user, shall be
components, or perform adjustments unless you are
one (1) year for all warranted products and components. a Midmark authorized service technician.
EXCLUSIONS
This warranty does not cover and Midmark shall not be liable for the following: (1) repairs and • Do not use another manufacturer's parts to replace
replacements because of misuse, abuse, negligence, alteration, accident, freight damage, or malfunctioning components. Use only Midmark
tampering; (2) products which are not installed, used, and properly cleaned as required in the replacement parts
Midmark “Installation” and or “Installation / Operation Manual for this applicable product. (3)
products considered to be of a consumable nature; (4) accessories or parts not manufactured
by Midmark; (5) charges by anyone for adjustments, repairs, replacement parts, installation, or THIS WARRANTY IS MIDMARK’S ONLY WARRANTY
other work performed upon or in connection with such products which is not expressly AND IS IN LIEU OF ALL OTHER WARRANTIES, EX-
authorized in writing in advance by Midmark. PRESS OR IMPLIED. MIDMARK MAKES NO IMPLIED
EXCLUSIVE REMEDY WARRANTIES OF ANY KIND INCLUDING ANY WAR-
Midmark’s only obligation under this warranty is the repair or replacement of defective parts. RANTIES OF MERCHANTABILITY OR FITNESS FOR
Midmark shall not be liable for any direct, special, indirect, incidental, exemplary, or consequen ANY PARTICULAR PURPOSE. THIS WARRANTY IS
tial damages or delay, including, but not limited to, damages for loss of profits or loss of use. LIMITED TO THE REPAIR OR REPLACEMENT OF
NO AUTHORIZATION DEFECTIVE PARTS.
No person or firm is authorized to create for Midmark any other obligation or liability in
SF-1487 REV. A1
connection with the products.

iv © Midmark Corporation 2004 SF-1854


Back Go To Table Of Contents Next
Operation & Troubleshooting
Operation & Mode
Electrical System:
Page

Troubleshooting M7 (-011 thru -016) ........................... A-2

Section A M7 (-020 thru -022) ........................... A-4


Filling the Chamber .............................. A-8
Heat Up / Sterilization ........................... A-14
Venting the Chamber ............................ A-20

Models:
Serial Numbers:
© Midmark Corporation 2004 SF-1854
A-1
Back Go To Table Of Contents Next
Operation & Troubleshooting
Troubleshooting
[Electrical System]
Electrical System - [M7 (-011 thru -016)]
The illustration shows all of the electrical components of the sterilzer. Problem: Page
Refer to the following page for a detailed description of current flow. Heating element does not turn ON:
- Heater light is OFF ............................... A-6
- Heater light is ON ............................. A-7

Overheat
Sterilizer shuts down before
Thermostat timer setting expires ......................... A-17
(Manual-Reset)
Timer buzzer does not function ........... A-19

Overheat
Thermostat
(Auto-Reset)
m.
Co

Heating Timer
Element Buzzer

Heater
Light

Timer
Temperature
Regulator
Relay

Co
m.

MA670901i

Electrical System Models: M7 (-011 thru -016)


Serial Numbers: all
A-2 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next
Operation & Troubleshooting
Electrical System - [M7 (-011 thru -016)] When the timer is turned ON (continued)...
Heater Light & Heating Element
When the contacts of the temperature regulator relay are closed,
With the power cord properly connected...
current is supplied to the heater light and the heating element.
Overheat Thermostats
Current (115 / 230 VAC) continuously flows thru the two (normally
As the relay contacts open and close, the heating element cycles ON / OFF.
closed) overheat thermostats. This current supplies power to the timer.
This continues until the timer setting expires.
If either thermostat opens (overheat or malfunction), voltage is removed
The heater light is illuminated whenever the heating element is ON.
from the timer until the thermostat is reset or replaced.

NOTE
The Manual-Reset Thermostat contacts open at approximately 285°F (140°C).
To reset, allow unit to cool, then press RESET button on front of unit. When the timer setting expires...
Timer & Timer Buzzer
The Auto-Reset Thermostat contacts open at approximately 295°F (146°C). The contacts to the temperature regulator relay open, stopping the current
This thermostat automatically resets when the unit cools to approx. 265°F (129°C). flow to the heater light & heating element.
Timer The contacts to the timer buzzer close and current flows to the timer buzzer.
Current is supplied to the timer thru the two overheat thermostats. When voltage is applied, the buzzer emits an audible signal.

The contacts to the timer motor remain closed for one minute. After one
minute the contacts to the timer motor & the timer buzzer open, stopping the
When the timer is turned ON... current flow to these two components.
Timer
The (normally open) timer contacts close, and voltage is supplied to the
timer motor and the temperature regulator relay. The timer motor runs, and
begins to count down the time it was set for.
(The contacts to the timer buzzer remain open).

Temperature Regulator Relay


Current is supplied to the temperature regulator relay thru the timer. If the
chamber temperature is lower than the temperature knob setting*, the relay
contacts are closed. When these contacts are closed, current flows thru the
relay to the heating element and the heater light.

The diaphragm cup of the relay expands as the temperature & pressure
inside the chamber increase. When the chamber temperature reaches the
temperature knob setting, the relay contacts open, and voltage is removed
from the heating element & heater light.

[* The minimum temperature knob setting is approx. 220°F (104°C)]

Models: M7 (-011 thru -016) Electrical System


Serial Numbers: all

© Midmark Corporation 2004 SF-1854


A-3
Back Go To Table Of Contents Next
Operation & Troubleshooting
Troubleshooting
[Electrical System]
Electrical System - [M7 (-020 thru -022)]
The illustration shows all of the electrical components of the sterilzer. Problem: Page
Refer to the following page for a detailed description of current flow. When Fill/Vent Switch is pressed:
- Chamber does not FILL ....................... A-11
- Chamber does not VENT ..................... A-23
Fuse Heating element does not turn ON:
Overheat - Heater light is OFF ............................... A-6
Thermostat
(Manual-Reset) - Heater light is ON ............................. A-7
Sterilizer shuts down before
timer setting expires ......................... A-17
Timer buzzer does not function ........... A-19

Overheat
m.
Co Thermostat
(Auto-Reset)
Timer
Heating Buzzer
Element
Fill / Vent
Switch

Heater
Light
Fill / Vent
Valve

Temperature
Regulator
Relay Timer

Co
m.

Com. MA670900i

Electrical System Models: M7 (-020 thru -022)


Serial Numbers: all
A-4 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next
Operation & Troubleshooting
Electrical System - [M7 (-020 thru -022)] When the timer is turned ON (continued)...
Temperature Regulator Relay
Current is supplied to the temperature regulator relay thru the timer. If the
With the power cord properly connected...
chamber temperature is lower than the temperature knob setting*, the relay
Fuse
contacts are closed. When these contacts are closed, current flows thru the
Current (115 / 230 VAC) continuously flows thru the fuse located in the
relay to the heating element and the heater light.
back of the unit. This current supplies power to the fill / vent switch and
[* The minimum temperature knob setting is approx. 220°F (104°C)]
the overheat thermostats.
The diaphragm cup of the relay expands as the temperature & pressure
Fill / Vent Switch
inside the chamber increase. When the chamber temperature reaches the
Current is supplied to the fill / vent switch thru the fuse.
temperature knob setting, the relay contacts open, and voltage is removed
from the heating element & heater light.
Overheat Thermostats & Timer
Current is supplied to the two overheat thermostats thru the fuse.
Heater Light & Heating Element
Current continuously flows thru the thermostats to the timer.
When the contacts of the temperature regulator relay are closed,
current is supplied to the heater light and the heating element.
If either thermostat opens (overheat or malfunction), voltage is removed
As the relay contacts open and close, the heating element cycles
from the timer until the thermostat is reset or replaced.
ON / OFF. This continues until the timer setting expires.
The heater light is illuminated whenever the heating element is ON.
NOTE
The Manual-Reset Thermostat contacts open at approximately 285°F (140°C).
To reset, allow unit to cool, then press RESET button on front of unit. When the timer setting expires...
Timer & Timer Buzzer
The Auto-Reset Thermostat contacts open at approximately 295°F (146°C). The contacts to the temperature regulator relay open, stopping the current
This thermostat automatically resets when the unit cools to approx. 265°F (129°C). flow to the heater light & heating element.

When filling the chamber (pressing the fill/vent switch)... The contacts to the timer buzzer close and current flows to the timer buzzer.
Fill / Vent Switch When voltage is applied, the buzzer emits an audible signal.
The contacts of the (normally open) switch close. When the contacts of the
switch are closed, current is supplied to the fill / vent valve. The contacts to the timer motor remain closed for one minute. After one
minute the contacts to the timer motor & the timer buzzer open, stopping the
Fill / Vent Valve current flow to these two components.
When current is applied to the (normally closed) valve, the valve opens.
When the valve is open, water flows into the chamber. When pressing the Fill / Vent Switch (to VENT the chamber)...
Fill / Vent Switch
When the Timer is turned ON... The contacts of the (normally open) switch close. When the contacts of the
Timer switch are closed, current is supplied to the fill / vent valve.
The (normally open) timer contacts close, and voltage is supplied to the
timer motor and the temperature regulator relay. The timer motor runs, and Fill / Vent Valve
begins to count down the time it was set for. When current is applied to the (normally closed) valve, the valve opens.
(The contacts to the timer buzzer remain open). When the valve is open, steam is released thru the condensing coil & the
water is returned to the reservoir .

Models: M7 (-020 thru -022) Electrical System


Serial Numbers: all

© Midmark Corporation 2004 SF-1854


A-5
Back Go To Table Of Contents Next
Operation & Troubleshooting Problem: Heating element does not turn ON.
[Heater light is OFF]

Refer To: Page


Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
Access Procedures .............................. C-1
Wiring Diagrams ................................... D-1
Exploded Views / Part Numbers ........... E-1

Loose / Damaged Wire Connections 4th


Check all wiring connections.
3
(Power cord, overheat thermostats, etc.)
M7 (-020 thru -022) only 3rd
Fuse
3

Timer 7th 3

R
Temperature
Regulator Assy.
6th
3

Check supply voltage


2nd
3
(A dedicated circuit is recommended)

Press RESET button


Allow unit to cool for 15-20 minutes
1st
3
before pressing RESET button.

Overheat Thermostats 5th 3 MA670800i

Electrical System Models: M7 (-011 thru -016) M7 (-020 thru -022)


Serial Numbers: all all
A-6 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Problem: Heating element does not turn ON.


Operation & Troubleshooting
[Heater light is ON]

Refer To: Page


Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
Access Procedures .............................. C-1
Wiring Diagrams ................................... D-1
Exploded Views / Part Numbers ........... E-1

Heating 1st
Element
3

MA672300i

Models: M7 (-011 thru -016) M7 (-020 thru -022) Electrical System


Serial Numbers: all all
© Midmark Corporation 2004 SF-1854
A-7
Back Go To Table Of Contents Next
Operation & Troubleshooting
Troubleshooting
[Filling the Chamber]
Filling the Chamber
The illustrations show the water flow when filling the chamber. Problem: Page
Refer to the following page for a detailed description of this process. Chamber does not fill:
- M7 (-011 thru -016) .............................. A-10
- M7 (-020 thru -022) .............................. A-11
Water continuously flows into chamber:
M7 (-011 thru -016)
- M7 (-011 thru -016) .............................. A-12
Reservoir - M7 (-020 thru -022) ........................... A-13

M7 (-020 thru -022)

Reservoir

Fill / Vent Valve


(open)
Filter Screen

Fill / Vent Valve MA671500i

= Water (open)

Filling the Chamber Models: M7 (-011 thru -016) M7 (-020 thru -022)
Serial Numbers: all all
A-8 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next
Operation & Troubleshooting
Filling the Chamber

M7 (-011 thru -016) M7 (-020 thru -022)


When the Fill / Vent Lever is pressed and held... When the Fill / Vent Switch is pressed and held...
Fill Vent Valve Fill / Vent Switch & Valve
The (normally closed - manual) valve opens. When the valve is open, water Current (line voltage) flows thru the fill/vent switch to the fill/vent valve. When
from the reservoir flows thru the fill / vent valve into the chamber. voltage is applied, the (normally closed) fill / vent valve opens. When the valve
is open, water from the reservoir flows into the chamber thru the valve and filter
screen.
When the Fill / Vent Lever is released...
Fill / Vent Valve When the Fill / Vent Switch is released...
The valve closes, and stops the flow of water into the chamber. Fill / Vent Switch & Valve
The fill/vent switch opens, stopping the current flow to the fill/vent valve.
When voltage is removed, the valve closes. When the valve closes, water
stops flowing into the chamber.

R R

Fill / Vent Fill / Vent


Lever Switch MA512903i

Models: M7 (-011 thru -016) M7 (-020 thru -022) Filling the Chamber
Serial Numbers: all all
© Midmark Corporation 2004 SF-1854
A-9
Back Go To Table Of Contents Next
Operation & Troubleshooting Problem: Chamber does not fill.

Refer To: Page


Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
Access Procedures .............................. C-1
Wiring Diagrams ................................... D-1
Exploded Views / Part Numbers ........... E-1

Low water in reservoir? 1st


Add distilled water if necessary.
3

Fill / Vent Valve & Tubing 2nd


Clean / adjust / replace as necessary.
3

MA672400i

Filling the Chamber Models: M7 (-011 thru -016)


Serial Numbers: all
A-10 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Problem: Chamber does not fill.


Operation & Troubleshooting

Refer To: Page


Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
Access Procedures .............................. C-1
Wiring Diagrams ................................... D-1
Exploded Views / Part Numbers ........... E-1

Fuse 4th
3

Low water in reservoir? 1st


Add distilled water if necessary.
3
Filter Screen 2nd
Clean / replace as necessary.
3 R

Check supply voltage


3rd
3
(A dedicated circuit is recommended)

Fill / Vent Switch 6th 3


Fill / Vent Valve & Tubing 5th
Clean / adjust / replace as necessary.
3
MA672500i

Models: M7 (-020 thru -022) Filling the Chamber


Serial Numbers: all
© Midmark Corporation 2004
A-11
SF-1854
Back Go To Table Of Contents Next
Operation & Troubleshooting Problem: Water continuously flows into chamber.

Refer To: Page


Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
Access Procedures .............................. C-1
Wiring Diagrams ................................... D-1
Exploded Views / Part Numbers ........... E-1

Fill / Vent Valve & Tubing 1st


Clean / adjust / replace as necessary.
3

MA672400i

Filling the Chamber Models: M7 (-011 thru -016)


Serial Numbers: all
A-12 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Problem: Water continuously flows into chamber.


Operation & Troubleshooting

Refer To: Page


Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
Access Procedures .............................. C-1
Wiring Diagrams ................................... D-1
Exploded Views / Part Numbers ........... E-1

Fill / Vent Valve & Tubing 1st


Clean / adjust / replace as necessary.
3
Fill / Vent Switch 2nd 3

MA672600i

Models: M7 (-020 thru -022) Filling the Chamber


Serial Numbers: all
© Midmark Corporation 2004
A-13
SF-1854
Back Go To Table Of Contents Next
Operation & Troubleshooting
Troubleshooting
[Heat-Up / Sterilization]
Heat Up / Sterilization
The illustrations show the water / steam flow during heat up & sterilization. Problem: Page
Refer to the following page for a detailed description of this process. Heating element does not turn ON:
- Heater light is OFF ............................... A-6
- Heater light is ON ............................. A-7
Heat-Up Heating element turns ON, but does not
reach required temperature .............. A-16
Sterilizer shuts down before
timer setting expires ......................... A-17
Biological test strips indicate items
are not sterile ................................... A-18
Timer buzzer does not function ........... A-19
Bellows
(open)

Sterilization

Temperature
Regulator Relay
Pressure Relief Valve
(closed)
(closed)

Fill / Vent Valve


(closed)
Note: Electronic valve is shown.
Manual valve is also closed.

Bellows
(closed)

= Water
Temperature
Regulator Relay
= Heated Water (closed / open) MA672000i

Fill / Vent Valve


= Steam (closed)
Note: Electronic valve is shown.
Manual valve is also closed.

Heat-Up / Sterilization Models: M7 (-011 thru -016) M7 (-020 thru -022)


Serial Numbers: all all
A-14 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next
Operation & Troubleshooting
Heat-Up / Sterilization
When the timer is turned ON... When the timer setting expires...
Timer Timer & Timer Buzzer
The (normally open) timer contacts close, and voltage is supplied to the The contacts to the temperature regulator relay open, stopping the current
timer motor and the temperature regulator relay. The timer motor runs, and flow to the heater light & heating element.
begins to count down the time it was set for.
(The contacts to the timer buzzer remain open). The contacts to the timer buzzer close and current flows to the timer buzzer.
When voltage is applied, the buzzer emits an audible signal.
Temperature Regulator Relay
Current is supplied to the temperature regulator relay thru the timer. If the The contacts to the timer motor remain closed for one minute. After one
chamber temperature is lower than the temperature knob setting*, the relay minute the contacts to the timer motor & the timer buzzer open, stopping the
contacts are closed. When these contacts are closed, current flows thru the current flow to these two components.
relay to the heating element and the heater light.
[* The minimum temperature knob setting is approx. 220°F (104°C)]

The diaphragm cup of the relay expands as the temperature & pressure
inside the chamber increase. When the chamber temperature reaches the
temperature knob setting, the relay contacts open, and voltage is removed
from the heating element & heater light.

Heater Light & Heating Element


When the contacts of the temperature regulator relay are closed,
current is supplied to the heater light and the heating element.
R

As the relay contacts open and close, the heating element cycles ON / OFF.
This continues until the timer setting expires.

The heater light is illuminated whenever the heating element is ON.

Bellows & Pressure Relief Valve


Heat-Up:
Temperature Knob
As the water in the chamber begins to boil, air is forced out of the chamber.
This air passes thru the bellows into the reservoir.

Timer Knob MA672100i


Sterilization:
When pure steam begins to flow thru the bellows, the bellows closes
allowing pressure to build in the chamber. If the pressure in the chamber
exceeds 34 psi (234 kPa), the pressure relief valve opens to prevent unsafe
conditions.

Models: M7 (-011 thru -016) M7 (-020 thru -022) Heat-Up / Sterilization


Serial Numbers: all all
© Midmark Corporation 2004
A-15
SF-1854
Back Go To Table Of Contents Next
Operation & Troubleshooting Problem: Heating element turns ON, but does
not reach required temperature.
[Heater light is ON]
Refer To: Page
Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
Access Procedures .............................. C-1
Wiring Diagrams ................................... D-1
Exploded Views / Part Numbers ........... E-1

Check for pressure leaks.


Refer to Section B for checkpoints.
1st
3

R
Temperature 2nd
Regulator Assy.
3

Heating Element 3rd 3

MA672700i

Heat-Up / Sterilization Models: M7 (-011 thru -016) M7 (-020 thru -022)


Serial Numbers: all all
A-16 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Problem: Sterilizer shuts down before timer setting expires.


Operation & Troubleshooting

Refer To: Page


Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
Access Procedures .............................. C-1
Wiring Diagrams ................................... D-1
Exploded Views / Part Numbers ........... E-1

Check for pressure leaks.


Refer to Section B for checkpoints.
1st
3

Timer 5th 3
Temperature
Regulator Assy.
4th
3

Overheat Thermostats 3rd 3

Press RESET button 2nd


Allow unit to cool for 15-20 minutes
3
before pressing RESET button.

MA672800i

Models: M7 (-011 thru -016) M7 (-020 thru -022) Heat-Up / Sterilization


Serial Numbers: all all
© Midmark Corporation 2004
A-17
SF-1854
Back Go To Table Of Contents Next
Operation & Troubleshooting Problem: Biological test strips indicate items are not sterile.

Refer To: Page


Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
Access Procedures .............................. C-1
Wiring Diagrams ................................... D-1
Exploded Views / Part Numbers ........... E-1
Type / condition of indicator strips 1st 3
This unit requires test strips rated for:
Gravity Displacement Steam Sterilizers
Test strips must be stored in a cool, dry location.
Failure to do so will result in faulty readings.
(Follow all instructions provided with test strips)

Is the sterlizer overloaded? 2nd 3 Check for pressure leaks. 4th


Refer to Section B for checkpoints.
3
Large loads or heavy linen packs may
prevent strips from changing.

Are the correct trays being used?


3rd
3
Some trays may prevent proper air flow.
Be sure trays are designed for this sterilizer. R

Temperature
Regulator Assy.
5th
3

MA672900i

Heat-Up / Sterilization Models: M7 (-011 thru -016) M7 (-020 thru -022)


Serial Numbers: all all
A-18 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Problem: Timer buzzer does not function.


Operation & Troubleshooting

Refer To: Page


Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
Access Procedures .............................. C-1
Wiring Diagrams ................................... D-1
Exploded Views / Part Numbers ........... E-1

Timer Buzzer 1st


3
R

MA672900i

Models: M7 (-011 thru -016) M7 (-020 thru -022) Heat-Up / Sterilization


Serial Numbers: all all
© Midmark Corporation 2004
A-19
SF-1854
Back Go To Table Of Contents Next
Operation & Troubleshooting
Troubleshooting
[Venting the Chamber]
Venting the Chamber
The illustrations show the steam / water flow when venting the chamber. Problem: Page
Refer to the following page for a detailed description of this process. Chamber will not vent:
- M7 (-011 thru -016) .............................. A-22
- M7 (-020 thru -022) .............................. A-23

M7 (-011 thru -016)

Condensing Coil

M7 (-020 thru -022)

Fill / Vent Valve


(open)

Condensing
Filter Screen Coil

= Water

= Steam
MA672200i

Fill / Vent Valve


(open)

Venting the Chamber Models: M7 (-011 thru -016) M7 (-020 thru -022)
Serial Numbers: all all
A-20 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next
Operation & Troubleshooting
Venting the Chamber

Attention
M7 (-020 thru -022)
The door handle must be moved to the VENT position
before pressing the Fill / Vent Lever or Switch. When the Fill / Vent Switch is pressed and held...
Fill / Vent Switch & Valve
Current (line voltage) flows thru the fill/vent switch to the fill/vent valve. When
M7 (-011 thru -016) voltage is applied, the (normally closed) fill / vent valve opens. Pressure forces
water and steam back into the reservoir thru the valve and the condensing coil.
When the Fill / Vent Lever is pressed and held... When all of the pressure has been vented, the door will "pop".
Fill / Vent Valve
The (normally closed - manual) valve opens. Pressure forces water Note: Release the lever when the door "pops".
and steam back into the reservoir thru the valve and the condensing coil. If the lever is held too long, the chamber will begin to fill.
When all of the pressure has been vented, the door will "pop".
When the Fill / Vent Switch is released...
Note: Release the lever when the door "pops". Fill / Vent Switch & Valve
If the lever is held too long, the chamber will begin to fill. The fill/vent switch opens, stopping the current flow to the fill/vent valve.
When voltage is removed, the valve closes.
When the Fill / Vent Lever is released...
Fill / Vent Valve
The valve closes.

R R

Fill / Vent Fill / Vent


Lever Switch MA512903i

Models: M7 (-011 thru -016) M7 (-020 thru -022) Venting the Chamber
Serial Numbers: all all
© Midmark Corporation 2004
A-21
SF-1854
Back Go To Table Of Contents Next
Operation & Troubleshooting Problem: Chamber will not vent.

Refer To: Page


Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
Access Procedures .............................. C-1
Wiring Diagrams ................................... D-1
Exploded Views / Part Numbers ........... E-1

Fill / Vent Valve & Tubing 1st


Clean / adjust / replace as necessary.
3

MA672400i

Venting the Chamber Models: M7 (-011 thru -016)


Serial Numbers: all
A-22 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Problem: Chamber will not vent.


Operation & Troubleshooting

Refer To: Page


Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
Access Procedures .............................. C-1
Wiring Diagrams ................................... D-1
Exploded Views / Part Numbers ........... E-1

Filter Screen 1st


Clean / replace as necessary.
3 R

Fill / Vent Valve & Tubing 2nd


Clean / adjust / replace as necessary.
3
Fill / Vent Switch 3rd 3

MA673000i

Models: M7 (-020 thru -022) Venting the Chamber


Serial Numbers: all
© Midmark Corporation 2004
A-23
SF-1854
Back Go To Table Of Contents Next
Back Go To Table Of Contents Next

Component Testing & Repair


Testing & Repair Component / Procedure Page
Checking For Pressure Leaks .............. B-2
Fuse [M7 (-020 thru -022) only] ............ B-3

Section B Bellows ............................................... B-4


Fill / Vent Valve:
Manual .............................................. B-6
Electronic .......................................... B-8
Temperature Regulator Assy. ............... B-12
Heating Element ................................... B-18
Overheat Thermostats .......................... B-22
Pressure Relief Valve ........................... B-25
Timer Assembly.................................... B-26
Timer Buzzer ........................................ B-30
Temperature Gauge ............................. B-31
Door Assembly ..................................... B-32
Reservoir Tank ..................................... B-34
Chamber Assembly .............................. B-36

Models:
Serial Numbers:
© Midmark Corporation 2004 SF-1854
B-1
Back Go To Table Of Contents Next

Component Testing & Repair


Checking for Pressure Leaks
This illustration shows the areas to check for pressure leaks. Components Page
Bellows ............................................... B-4
Fill / Vent Valve:
Manual .............................................. B-6
Electronic .......................................... B-8
Pressure Relief Valve ........................... B-25
Door Assembly ..................................... B-32

Water leaking from fitting(s)? Steam from pressure relief valve?


Tighten / replace fittings if necessary. If YES, replace the pressure relief valve.

Steam from bellows tubing?


If YES, replace the bellows.
R

Note: The bellows should close at approx. 215°F


(101°C). Periodic "hissing' or "spitting" is
normal.

Bubbles from condensing coil?


If YES, clean or replace the fill / vent valve.

MA673100i

Water leaking from door?


If YES, replace door gasket.

Checking for Models: M7 (-011 thru -016) M7 (-020 thru -022)


Pressure Leaks Serial Numbers: all
B-2 © Midmark Corporation 2004 SF-1854
all
Back Go To Table Of Contents Next

Component Testing & Repair


Fuse
Fuses
Fuses Page
Page
Location Location
Location ............................................... B-3
............................................... B-2
Fuse Test ............................................. B-3
Fuse Test ............................................. B-2
Wiring
Wiring Diagrams
Diagrams ...................................
................................... D-1
D-1
Part Numbers .......................................
Fuse Rating ............................... 6.3A, 250V E-14.1
Fuse Type T, 5 x 20mm
Holder
Part Number ............................ 015-0346-20

Fuse Test
Fuse Test
Step 1: Place meter probes on ends of fuse.
Acceptable Range [Set meter to 200 ohms (Ω)]

less than 5 Ω

MA511501i

Fuse Ratings:
115V models ...... 12 amp, 250 V, Fast-Acting, 1/4" x 1-1/4"
230V models ...... 8 amp, 250 V, Fast-Acting, 5mm x 20mm

FuseTest
If reading is OL...
Replace fuse.
If reading is within acceptable range...
Fuse is OK.

Models: M7 (-020 thru -022) Fuse


all
Serial Numbers:
© Midmark Corporation 2004 SF-1854
B-3
Back Go To Table Of Contents Next

Component Testing & Repair


Bellows
Bellows Page
Location & Function .............................. B-4
Location & Function
Testing - refer to:
Checking for Pressure Leaks ............ B-2
Replacement ........................................ B-5
Exploded View / Part Numbers ............. E-11

The arrow indicates the direction


of flow toward the reservoir.

The bellows must be oriented correctly


to ensure proper operation.

As the water in the chamber begins to boil...


Air & steam are forced out of the chamber,
thru the open bellows, and back into the reservoir.

R
When pure steam begins flowing thru bellows...
The bellows closes allowing pressure to build in the chamber.
[Note: The bellows will periodically "hiss" or "spit", this is normal.

MA673301i

Bellows Models: M7 (-011 thru -016) M7 (-020 thru -022)


Serial Numbers: all all
B-4 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Component Testing & Repair


Bellows - continued
Refer to: Page
Cover Removal ..................................... C-2
Replacement

NOTE:
The arrow on the bellows indicates the direction of flow toward the reservoir.
The bellows must be oriented correctly to ensure proper operation.

Arrow must point


toward the reservoir
as shown
Removal
Step 1: Loosen two compression nuts.
Remove bellows.

Installation
(See NOTE for proper orientation of bellows)
Step 1: Install bellows.
Tighten compression nuts.

MA673401i

Models: M7 (-011 thru -016) M7 (-020 thru -022) Bellows


Serial Numbers: all all
© Midmark Corporation 2004 SF-1854
B-5
Back Go To Table Of Contents Next

Component Testing & Repair


Fill / Vent Valve (manual)
Fill / Vent Valve Page
Location & Function .............................. B-6
Location & Function
Testing - refer to:
Checking for Pressure Leaks ............ B-2
Early Models Later Models Removal / Installation / Adjustment ...... B-7
Exploded View / Part Numbers ............. E-9

When the fill / vent lever is pressed (no pressure in chamber)...


The (normally closed) fill / vent valve opens.
Water from the reservoir flows thru the open valve into the chamber.
The valve closes when the lever is released.

When the fill / vent lever is pressed (chamber is pressurized)...


The (normally closed) fill / vent valve opens.
Water and steam from the chamber are forced thru the open valve back into the
reservoir. When all of the pressure has been released, the door will "pop".
The valve closes when the lever is released.
R

Fill / Vent Lever MA673500i

Fill / Vent Valve Models: M7 (-011 thru -016)


Serial Numbers: all
B-6 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Component Testing & Repair


Fill / Vent Valve (manual) - continued
Refer to: Page
Cover Removal ..................................... C-2
Removal / Installation / Adjustment
Draining / Filling the Reservoir ............. C-4

Removal Installation
Step 1: Drain water from reservoir. Step 2: Align tubing with valve.
Tighten two compression nuts.
Adjust lever bracket. (See Adjustment)

Removal
Step 2: Loosen two compression nuts.

DISTANCE "A"

Lever Bracket

Lever Bracket

Removal
Step 3: Disengage lever bracket from lever. Adjustment
Remove valve. Adjust lever bracket so that Distance "A"
is approx. 1/4" (0.63 cm). To secure position,
tighten nut against lever bracket.
Lever

Installation
Step 1: Engage lever bracket with lever.
Press valve into mtg. brackets.

MA673600i

Models: M7 (-011 thru -016) Fill / Vent Valve


Serial Numbers: all
© Midmark Corporation 2004 SF-1854
B-7
Back Go To Table Of Contents Next

Component Testing & Repair


Fill / Vent Valve (electronic)
Fill / Vent Valve Page
Location & Function .............................. B-8
Location & Function
Testing - refer to:
Checking for Pressure Leaks ............ B-2
Electrical Test .................................... B-9
Replacement ........................................ B-10
Disassembly / Cleaning ........................ B-11
Wiring Diagrams ................................... D-1
Exploded View / Part Numbers ............. E-9.3

When the fill / vent switch is pressed (no pressure in chamber)...


Current (line voltage) flows thru the fill/vent switch to the fill / vent valve.
When voltage is applied, the (normally closed) fill/vent valve opens.
Water from the reservoir flows thru the open valve into the chamber.
The valve closes when the switch is released.

When the fill / vent lever is pressed (chamber is pressurized)...


R
Current (line voltage) flows thru the fill/vent switch to the fill/vent valve.
When voltage is applied, the (normally closed) fill / vent valve opens.
Water and steam from the chamber are forced thru the open valve back into
the reservoir. When all of the pressure has been released, the door will "pop".
The valve closes when the lever is released.

Fill / Vent Switch


MA673800i

Fill / Vent Valve Models: M7 (-020 thru -022)


Serial Numbers: all
B-8 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Component Testing & Repair


Fill / Vent Valve (electronic) - continued
Refer to: Page
Cover Removal ..................................... C-2
Electrical Test

Electrical Test
Step 1: Disconnect wires from valve terminals.

Electrical Test
Step 2: Place meter probes on terminals.
[Set meter to M ohms (Ω)]

Acceptable Range

115 VAC models: ........ 3.24 to 3.96


230 VAC models: ....... 3.24 to 3.96

Electrical Test
If reading is out of acceptable range...
Replace valve.

If reading is within acceptable range...


Electrical component of valve is OK.

MA674000i

Models: M7 (-020 thru -022) Fill / Vent Valve


Serial Numbers: all
© Midmark Corporation 2004 SF-1854
B-9
Back Go To Table Of Contents Next

Component Testing & Repair


Fill / Vent Valve (electronic) - continued
Refer to: Page
Cover Removal ..................................... C-2
Replacement

Removal
Step 1: Tag and disconnect wires Installation
from valve terminals. Step 3: Connect compression fittings.

Installation Removal
Step 4: Connect wires to valve terminals. Step 2: Disconnect compression fittings.

Installation
Step 1: Install valve onto bracket.
Locate "IN" port and direct flow into
the reservoir as shown.

Removal
Step 4: Remove valve from bracket.

Installation
Step 2: Secure bracket / valve to base plate.

Removal
Step 3: Remove bracket / valve from base plate.

MA677901i

Fill / Vent Valve Models: M7 (-020 thru -022)


Serial Numbers: all
B-10 © Midmark Corporation 2004 SF-1854 Rev. 7/07
Back Go To Table Of Contents Next

Component Testing & Repair


Fill / Vent Valve (electronic) - continued
Refer to: Page
Valve Removal ..................................... B-10
Disassembly / Cleaning

Disassembly / Cleaning
Step 1: Remove nut.
Disassemble valve.

Disassembly / Cleaning
Step 2: Remove any debris.
Inspect components for damage.

Models: M7 (-020 thru -022) Fill / Vent Valve


all
Serial Numbers:
© Midmark Corporation 2004 SF-1854
B-11
Back Go To Table Of Contents Next

Component Testing & Repair Temp. Regulator Assy. Page


Location & Function .............................. B-12
Temperature Regulator Assembly Troubleshooting .................................... B-13
Temperature Relay:
Location & Function Adjustment ........................................ B-14
Temperature
Relay Removal ............................................ B-15
Diaphragm Cup
Installation ......................................... B-16
Diaphragm Cup Replacement .............. B-17
Wiring Diagrams ................................... D-1
Exploded View / Part Numbers ............. E-7

Flexible Shaft

When the Temperature Control knob is adjusted...


The flexible shaft rotates, increasing or decreasing the distance
between the relay contacts. This adjusts the point (i.e. temperature)
at which the relay contacts will open & close*.

As the temperature & pressure inside the chamber increase...


The diaphragm cup expands, pushing the relay contacts apart.
When the relay contacts are open, the heating element is de-energized.

As the temperature & pressure inside the chamber decrease...


R
The diaphragm cup contracts, allowing the relay contacts to close.
When the relay contacts are closed, the heating element is energized.

MA674200i

Temperature Regulator Models: M7 (-011 thru -016) M7 (-020 thru -022)


Assembly Serial Numbers: all all
B-12 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Component Testing & Repair


Temperature Regulator Assembly - continued
Refer to: Page
Relay Adjustment ................................. B-14
Troubleshooting
Relay Removal ..................................... B-15
Diaphragm Cup Replacement .............. B-17

Sterilizer shuts down before timer setting expires...


Required action: Inspect relay.
(If contacts are corroded or "fused" together - replace relay)
Inspect diaphragm cup.
(If water / steam is leaking from this area, replace diaphragm cup) Temperature
Relay

Diaphragm Cup

Sterilizer does not reach desired temperature...


Required action: Perform Relay Adjustment.
Replace relay if necessary.

MA674201i

Models: M7 (-011 thru -016) M7 (-020 thru -022) Temperature Regulator


Assembly
Serial Numbers: all all
© Midmark Corporation 2004 SF-1854
B-13
Back Go To Table Of Contents Next

Component Testing & Repair


Temperature Regulator Assembly - continued
Refer to: Page
Cover Removal ..................................... C-2
Relay Adjustment

Relay Adjustment
Step 4: Allow chamber to reach its max.
temperature [>270°F (132°C)].

Relay Adjustment
Step 5: Adjust screw until temperature gauge
reads slightly above 270°F (132°C).

Relay Adjustment
Step 1: Fill chamber with water.
Close & latch door.
Set Timer knob to 30 minutes.

Relay Adjustment
Step 3: Loosen setscrew 2-3 turns.

Relay Adjustment
Step 6: Adjust setscrew until temperature
Relay Adjustment gauge reads 270-271°F (131-132°C).
Step 2: Turn Temperature knob to max.
MA674400i

Temperature Regulator Models: M7 (-011 thru -016) M7 (-020 thru -022)


Assembly Serial Numbers: all all
B-14 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Component Testing & Repair


Temperature Regulator Assembly - continued
Refer to: Page
Cover Removal ..................................... C-2
Relay Removal

Removal
Step 4: Remove timer buzzer.
Remove relay & spacer.

Removal
Step 2: Rotate Temperature knob until
flexible shaft unscrews from relay.

Removal Removal
Step 3: Disconnect two relay wires. Step 1: Remove setscrew from flexible shaft.
(one from heating element &
one from timer - not shown)

MA674500i

Models: M7 (-011 thru -016) M7 (-020 thru -022) Temperature Regulator


Assembly
Serial Numbers: all all
© Midmark Corporation 2004 SF-1854
B-15
Back Go To Table Of Contents Next

Component Testing & Repair


Temperature Regulator Assembly - continued
Refer to: Page
Relay Removal ..................................... B-15
Relay Installation
Relay Adjustment ................................. B-14

Installation
Step 1: Install relay & spacer.
R
Install timer buzzer.
Note: Do not tighten Adjustment Screw.
Fine Adjustment Setscrew should
be flush with bracket.

Adjustment Screw

Fine Adjustment
Setscrew

Installation
(Apply hi-temp lubricant to threads of flexible shaft)
Step 3: Turn the Temperature knob until flexible
shaft screws all the way into relay.
Rotate knob back (counter-clockwise)
1 1/2 turns, then install Shaft Setscrew.
Shaft Setscrew

Knob Setscrew

Installation
Installation Step 4: Turn the Temperature knob counter-clockwise MA674501i

Step 2: Connect two relay wires. until it hits the stop.


(one to heating element & Hold knob in place, then loosen Knob Setscrew.
one to timer - not shown) Rotate knob so that white mark is at 3:00 position, Installation
then tighten knob setscrew. Step 5: Perform Relay Adjustment.

Temperature Regulator Models: M7 (-011 thru -016) M7 (-020 thru -022)


Assembly Serial Numbers: all all
B-16 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Component Testing & Repair


Temperature Regulator Assembly - continued
Refer to: Page
Relay Removal ..................................... B-15
Diaphragm Cup Replacement
Relay Installation .................................. B-16

Removal
Step 1: Remove relay.

Installation
Step 3: Install relay.
Removal
Step 2: Remove nut & lockwasher.

Installation
Step 2: Install lockwasher & nut.

Installation
Step 1: Install gasket onto diaphragm cup.
Install diaphragm cup.

Removal
Step 3: Remove diaphragm cup
Remove gasket from diaphragm cup.

MA674700i

Models: M7 (-011 thru -016) M7 (-020 thru -022) Temperature Regulator


all all Assembly
Serial Numbers:
© Midmark Corporation 2004 SF-1854
B-17
Back Go To Table Of Contents Next

Component Testing & Repair


Heating Element
Heating Element Page
Location & Function .............................. B-18
Location & Function
Resistance Test .................................... B-19
Replacement ........................................ B-20
Wiring Diagrams ................................... D-1
Exploded View / Part Numbers ............. E-13

When the timer is turned ON...


The timer supplies current to the temperature relay. If the
chamber temperature is lower than the temperature knob setting*,
the relay contacts are closed. When these contacts are closed, current
flows thru the relay to energize the heating element and the heater light.

When the chamber temperature reaches the temperature knob setting,


Heater Light the relay contacts open, and voltage is removed from the heating element
& heater light.

[* The minimum temperature knob setting is approx. 220°F (104°C)]

When the timer is OFF...


Timer contacts to the temperature relay open, stopping the current flow
to the heater light & heating element.

MA674800i

Heating Element Models: M7 (-011 thru -016) M7 (-020 thru -022)


Serial Numbers: all all
B-18 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Component Testing & Repair


Heating Element - continued
Refer to: Page
Cover Removal ..................................... C-2
Resistance Test

Resistance Test
Step 1: Disconnect wires from heating element terminals.

Acceptable Range

100 VAC models ........ 8 to 10


115 VAC models: ....... 11 to 13
230 VAC models: ....... 45 to 51
Resistance Test
Step 2: Place meter probes on terminals.
[Set meter to 200 ohms (Ω)] Resistance Test
If reading is out of acceptable range...
Replace heating element.
If reading is within acceptable range...
Heating element is OK.

MA674900i

Models: M7 (-011 thru -016) M7 (-020 thru -022) Heating Element


Serial Numbers: all all
© Midmark Corporation 2004 SF-1854
B-19
Back Go To Table Of Contents Next

Component Testing & Repair


Heating Element - continued
Refer to: Page
Cover Removal ..................................... C-2
Replacement
Wiring Diagrams ................................... D-1

Removal
Step 1: Disconnect power to sterilizer.
Drain all water from reservoir. Removal
Step 2: Remove inspection cover.
Installation
Step 4: Install inspection cover.

Installation
Step 2: Install two flat washers,
lockwashers, & nuts.
Note: Hold heating element in
place when tightening nuts.

Removal Installation
Step 4: Remove two nuts , Step 3: Connect wires to heating element terminals.
lockwashers & flat washers. (Refer to Section D for wiring diagrams)

Removal
Step 3: Tag & disconnect wires from
heating element terminals.

MA675000i

Heating Element Models: M7 (-011 thru -016) M7 (-020 thru -022)


Serial Numbers: all all
B-20 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Component Testing & Repair


Heating Element - continued
Refer to: Page
Cover Removal ..................................... C-2
Replacement - continued

Removal
Step 5: Remove heating element.
Remove gaskets from terminal posts. (Later Models)
Spacers

Installation
Step 1: Install gaskets onto terminal posts.
Install spacers or spacer bracket*.
Insert terminal posts thru chamber
& thermostat bracket.

(Early Models)
Spacer Bracket*

* Spacer Bracket must be installed above gaskets to prevent leaking.

MA675200i

Models: M7 (-011 thru -016) M7 (-020 thru -022) Heating Element


Serial Numbers: all all
© Midmark Corporation 2004 SF-1854
B-21
Back Go To Table Of Contents Next

Component Testing & Repair


Overheat Thermostats
Overheat Thermostats Page
Location & Function .............................. B-22
Location & Function
Resistance Test .................................... B-23
Replacement ........................................ B-24
Manual-Reset Auto-Reset Wiring Diagrams ................................... D-1
Thermostat Thermostat
Exploded View / Part Numbers ............. E-13

With the power cord properly connected...


Overheat Thermostats
Current (115 / 230 VAC) continuously flows thru the two (normally closed)
overheat thermostats. This current supplies power to the timer.

If either thermostat opens (overheat or malfunction), voltage is removed


from the timer until the thermostat is reset or replaced.

NOTE
The Manual-Reset Thermostat contacts open at approximately 285°F (140°C).
To reset, allow unit to cool, then press RESET button on front of unit.

The Auto-Reset Thermostat contacts open at approximately 295°F (146°C).


This thermostat automatically resets when the unit cools to approx. 265°F (129°C).

MA675300i

Overheat Models: M7 (-011 thru -016) M7 (-020 thru -022)


Thermostats Serial Numbers:
B-22 © Midmark Corporation 2004 SF-1854
all all
Back Go To Table Of Contents Next

Component Testing & Repair


Overheat Thermostats - continued
Refer to: Page
Cover Removal ..................................... C-2
Resistance Test

Attention!
Inspect thermostat for physical damage (ex. cracked plastic).
If damage is apparent, replace thermostat

Resistance Test
Step 1: Disconnect wires from thermostat terminals.

Acceptable Reading

0.00 (approximately)

ResistanceTest
If reading is (approximately) 0.00 ...
Thermostat is good.
Resistance Test
Step 2: Place meter probes on terminals. If reading is OL...
[Set meter to 200 ohms (Ω)] Replace thermostat.

Models: M7 (-011 thru -016) M7 (-020 thru -022) Overheat


Thermostats
Serial Numbers: all all
© Midmark Corporation 2004 SF-1854
B-23
Back Go To Table Of Contents Next

Component Testing & Repair


Overheat Thermostats - continued
Refer to: Page
Wiring Diagrams ................................... D-1
Replacement
Installation
Step 1: Install thermostat.
Removal
Step 1: Disconnect power to sterilizer.
Drain all water from reservoir.

Removal
Step 4: Disconnect wires.
Remove thermostat.
Removal
Installation Step 3: Loosen two nuts on
heating element posts.
Step 3: Connect wires.

Installation
Step 2: Tighten two nuts.

Removal
Step 2: Remove inspection cover.

Installation
Step 4: Install inspection cover.

MA675500i

Overheat Models: M7 (-011 thru -016) M7 (-020 thru -022)


Thermostats Serial Numbers: all all
B-24 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Component Testing & Repair


Pressure Relief Valve
Pressure Relief Valve Page
Location & Function .............................. B-25
Location & Function
Testing - refer to:
Checking for Pressure Leaks ............ B-2
If the pressure in the chamber exceeds 34 psi (234kPa)... Replacement ........................................ B-25
The pressure relief valve opens to prevent unsafe conditions. Exploded View / Part Numbers ............. E-8

Pressure Relief
Replacement
Valve

Installation
Step 1: Apply hi-temp sealant to valve threads.
Install valve.

Removal
Step 1: Unscrew pressure relief valve.

MA675700i

Models: M7 (-011 thru -016) M7 (-020 thru -022) Pressure Relief Valve
Serial Numbers: all all
© Midmark Corporation 2004 SF-1854
B-25
Back Go To Table Of Contents Next

Component Testing & Repair


Timer
Timer Page
Location & Function .............................. B-26
Location & Function
Supply Voltage Test ............................. B-27
Output Voltage Test.............................. B-28
Replacement ........................................ B-29
Wiring Diagrams ................................... D-1
Exploded View / Part Numbers ............. E-12

NOTE
Current is supplied to the timer thru the two overheat thermostats.

When the timer is turned ON...


The timer contacts to the timer motor & the temperature relay
close, and voltage is supplied to these components. When
voltage is applied to the timer motor, the time setting counts
down.
(The contacts to the timer buzzer remain open).

When the timer setting expires...


The timer contacts to the temperature relay open, stopping
the current flow to the relay.
R

The timer contacts to the buzzer close for one minute.


Current flows to the buzzer, resulting in a audible signal. After
one minute, the contacts to the timer motor & the buzzer
open, stopping the current flow to these two components.

MA675800i

Timer Models: M7 (-011 thru -016) M7 (-020 thru -022)


Serial Numbers:
B-26 © Midmark Corporation 2004 SF-1854
all all
Back Go To Table Of Contents Next

Component Testing & Repair


Timer - continued
Refer to: Page
Supply Voltage Test Cover Removal ..................................... C-2
Output Voltage Test.............................. B-28
Caution
This test must be performed with the power cord connected.

Supply Voltage Test


Step 1: Place meter probes on terminals as shown.
[Set meter to VAC]

Acceptable Range

line voltage (115 / 230 VAC +10%)

Supply VoltageTest
If reading is within range...
Perform Output Voltage Test.
If reading is out of range...
Check voltage supply.
(overheat thermostats, fuse, etc.)

MA676000i

Models: M7 (-011 thru -016) M7 (-020 thru -022) Timer


Serial Numbers: all all
© Midmark Corporation 2004 SF-1854
B-27
Back Go To Table Of Contents Next

Component Testing & Repair


Timer - continued
Refer to: Page
Output Voltage Test (perform Supply Voltage Test first) Cover Removal ..................................... C-2
Supply Voltage Test ............................. B-27
Caution
This test must be performed with the power cord connected.

Voltage to Temp. Relay Voltage to Timer Motor


Step 1: Turn timer knob to 10 minutes. Step 1: Turn timer knob to 10 minutes.
Place meter probes on terminals as shown. Place meter probes on terminals as shown.
[Set meter to VAC] [Set meter to VAC]

Acceptable Range

line voltage (115 / 230 VAC +10%)

Output VoltageTest
If reading is within range...
Timer is functioning properly.
If reading is out of range...
Replace timer.

Voltage to Timer Buzzer


Step 1: Turn timer knob to 1 minute (or less).
Place meter probes on terminals as shown.
[Set meter to VAC]

MA675900i

Timer Models: M7 (-011 thru -016) M7 (-020 thru -022)


Serial Numbers: all all
B-28 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Component Testing & Repair


Timer - continued
Refer to: Page
Replacement Cover Removal ..................................... C-2
Wiring Diagrams ................................... D-1

Installation
Step 2: Connect wires.
(refer to Wiring Diagrams)

Removal
Step 2: Tag and disconnect wires.

Installation
Step 3: Install timer knob.
Tighten set screw.

Installation
Removal Step 1: Install timer.
Step 1: Loosen set screw. Install two screws.
Remove timer knob. Removal
Step 3: Remove two screws.
Remove timer. MA521501i

Models: M7 (-011 thru -016) M7 (-020 thru -022) Timer


all all
Serial Numbers:
© Midmark Corporation 2004 SF-1854
B-29
Back Go To Table Of Contents Next

Component Testing & Repair


Timer Buzzer
Timer Buzzer Page
Location & Function .............................. B-30
Location & Function
Testing - refer to:
Timer ............................................... B-28
Replacement & Volume Adjustment ..... B-30
When the timer setting expires...
Exploded View / Part Numbers ............. E-12
Timer contacts to the buzzer close for one minute.
Current flows to the buzzer, causing an audible signal.

Replacement & Volume Adjustment

Timer
Buzzer
Installation
Step 1: Install buzzer.

Removal
Step 2: Remove buzzer.
Volume Adjustment
To increase volume...
Loosen adjustment screw.

To decrease volume...
Tighten adjustment screw. Removal
R

Step 1: Disconnect buzzer wires.

Installation
Step 2: Connect buzzer wires.

Adjustment Screw
(present on new-style buzzers only)

MA676100i

Timer Buzzer Models: M7 (-011 thru -016) M7 (-020 thru -022)


Serial Numbers: all all
B-30 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Component Testing & Repair


Temperature Gauge
Temperature Gauge Page
Replacement ........................................ B-31
Replacement
Exploded View / Part Numbers ............. E-5

Installation
Step 3: Slide rubber sleeve back against fitting.
Tighten compression nut. (Do not overtighten).
Installation
Step 2: Orient gauge face properly.
Tighten nut.

Removal
Step 1: Loosen nut.
Slide retainer bracket back.

Removal
Step 2: Unscrew compressionnut.
Remove temperature gauge.

Installation
Step 1: Install retainer bracket, nut, compression nut,
& rubber sleeve onto temperature gauge.

MA252301i

Models: M7 (-011 thru -016) M7 (-020 thru -022) Temperature Gauge


Serial Numbers: all all
© Midmark Corporation 2004 SF-1854
B-31
threadlocking adhesive.
Back Go To Table Of Contents Next

Component Testing & Repair


Door Assembly
Door Assembly Page
Door Replacement Testing - refer to:
Checking for Pressure Leaks ............ B-2
Door Replacement ................................ B-32
Gasket Replacement ............................ B-32
Disassembly / Assembly ...................... B-33
Door Removal Exploded View / Part Numbers ............. E-6
Step 1: Move door handle to Door Installation
unlatched position. Step 1: Install door stop & screw.

Door Removal
Step 2: Remove nut, Gasket Replacement
screw, & door stop.
Gasket Installation
Step 1: Using a brush, clean all
R

debris from gasket channel.

Door Installation
Step 2: Install nut* (Do not overtighten)
Door
*apply removeable threadlocking adhesive

Gasket
Channel

Gasket
Gasket Installation
Step 2: Lubricate gasket w/soapy water.
Insert gasket into gasket channel.
MA676200i

Door Assembly Models: M7 (-011 thru -016) M7 (-020 thru -022)


Serial Numbers: all all
B-32 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Component Testing & Repair


Door Assembly - continued

Disassembly / Assembly

Disassembly
Assembly Step 1: Remove cap nut, washer, & door.
Remove short spring & long spring.
Step 2: Install long spring & short spring.
Install door w/bowed edges at top & bottom.
Assembly
Step 3: Apply removable threadlocking adhesive
to T-bolt threads & face of washer.
Install washer & cap nut.

Disassembly
Step 3: Remove T-bolt from U-bracket.

Assembly
Step 1: Install T-bolt into U-bracket.

Assembly
Step 4: Install cross arm & handle.
Secure with pin & push nut.
MA522701i

Disassembly
Step 2: Remove push nut, pin,
handle, & cross arm.

Models: M7 (-011 thru -016) M7 (-020 thru -022) Door Assembly


Serial Numbers: all all
© Midmark Corporation 2004 SF-1854
B-33
Back Go To Table Of Contents Next

Component Testing & Repair


Reservoir Tank
Reservoir Tank Page
Removal ............................................... B-34
Removal
Installation ............................................ B-35
Exploded View / Part Numbers ............. E-8

Removal
Step 1: Drain all water from reservoir.

Removal
Step 3: Remove bulkhead fitting & washer.

Removal
Step 2: Loosen nut.
Removal
Step 6: Remove condensing coil, & washer.

Removal
Step 7: Pull bellows tube out of tank.
R

Removal
Step 5: Loosen nut.

Removal MA523001i
Step 4: Remove clamp & drain hose. Removal
Step 8: Remove four screws & tank.

Reservoir Tank Models: M7 (-011 thru -016) M7 (-020 thru -022)


Serial Numbers: all all
B-34 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Component Testing & Repair


Reservoir Tank

Installation

Note
When reconnecting plumbing, apply teflon tape or sealant
to threads - except where compression fittings are used.

Installation
Step 5: Install bulkhead fitting & washer.
Partially install nut. Installation
Step 8: Tighten all fittings
before filling reservoir.

Installation Installation
Step 6: Reconnect plumbing. Step 1: Install condensing coil & washer.
Tighten nut. Secure with nut.

Installation
Step 3: Insert bellows tubing into tank.
R

Installation
Step 2: Reconnect plumbing.

Installation
Installation Step 4: Install four screws.
Step 7: Install drain tube & clamp. MA523001i

Models: M7 (-011 thru -016) M7 (-020 thru -022) Reservoir Tank


all all
Serial Numbers:
© Midmark Corporation 2004 SF-1854
B-35
Back Go To Table Of Contents Next

Component Testing & Repair


Chamber Assembly
Chamber Assembly Page
Removal Removal ............................................... B-36
Disassembly / Assembly ...................... B-37
Installation ............................................ B-38
Removal Exploded View / Part Numbers ............. E-10
Step 1: Drain all water from reservoir.
Refer to (Removing):
Removal Covers ............................................... C-2
Step 2: Remove the following components:
Covers Tray Plate / Rack .................................. C-3
Removal
Tray Plate / Rack Door Assembly / Gasket ....................... B-32
Door Assembly & Gasket Step 3: Disconnect compression
Temperature Regulator Assy. fitting from top of chamber. Temperature Regulator Assembly ........ B-15
Heating Element Heating Element ................................... B-20
Overheat Thermostats
Overheat Thermostats .......................... B-24

Removal
Step 5: Remove screw(s) securing bracket* to base.
Remove chamber.
* Bracket design may vary.

Removal
Step 4: Disconnect compression
fitting from bottom of chamber.

MA523101i

Chamber Assembly Models: M7 (-011 thru -016) M7 (-020 thru -022)


Serial Numbers: all all
B-36 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Component Testing & Repair


Chamber Assembly - continued

Disassembly / Assembly

Disassembly
Step 1: Remove banding & insulation.

Assembly
Step 4: Install insulation & banding.

Assembly
Step 2: Install two fittings* into chamber.
*apply teflon tape / hi-temp sealant

Disassembly
Step 3: Remove two fittings from chamber.

Assembly
Step 1: Position door hinge & washer. Assembly
Install screw & nut*. Step 3: (If applicable) Install bracket.
Connect spring.
Rear
*apply removeable threadlocking adhesive Spring Hole Disassembly
Step 2: Remove bracket from chamber.
Note: On later models, the bracket is welded
to the chamber - skip this step.
Disassembly
Step 3: Disconnect spring.
Remove door hinge. MA253301i

Models: M7 (-011 thru -016) M7 (-020 thru -022) Chamber Assembly


Serial Numbers: all all
© Midmark Corporation 2004 SF-1854
B-37
Back Go To Table Of Contents Next

Component Testing & Repair


Chamber Assembly- continued
Refer to (Installing):
Installation Overheat Thermostats .......................... B-24
Heating Element ................................... B-20
Temperature Regulator Assembly ........ B-16
Note
Replace compression fittings if damage is apparent. Door Assembly / Gasket ....................... B-32
Tray Plate / Rack .................................. C-3
Covers ............................................... C-2

Assembly
Step 3: Connect compression fitting / plumbing
to fitting on top of chamber.

Assembly
Step 4: Install the following components:
Overheat Thermostats Installation
Heating Element Step 1: Secure bracket* to base.
Temperature Regulator Assy.
Door Assembly & Gasket *Bracket design may vary.
Tray Plate / Rack
Covers
R

Assembly
Step 2: Connect compression fitting / plumbing
to fitting on bottom of chamber.

MA523101i

Chamber Assembly Models: M7 (-011 thru -016) M7 (-020 thru -022)


Serial Numbers: all all
B-38 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Access Procedures
Access Procedures Removing & Installing: Page
Covers / Panels ................................... C-2
Tray Plate / Rack .................................. C-3

Section C Draining / Filling the Reservoir ............. C-4

Models:
Serial Numbers:
© Midmark Corporation 2004 SF-1854
C-1
Back Go To Table Of Contents Next

Access Procedures
Covers / Panels
Refer To: Page
Removal / Installation Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
Access Procedures .............................. C-1
Caution
Always unplug power cord before Wiring Diagrams ................................... D-1
removing any covers / panels. Exploded Views / Part Numbers ........... E-1

Installation: M7 (-011 thru -016)


Removal Step 1: Slide cover down over
Step 2: Remove cover. edge of front panel.

Installation: M7 (-020 thru -022)


Step 1: Slide cover down over edge of front
panel & back panel (not shown).

Removal
Step 1: Remove six screws.

Installation: M7 (all)
Step 2: Install six screws.
MA521401i

Covers / Panels Models:


Serial Numbers:
C-2 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Access Procedures
Tray Plate / Rack
Refer To: Page
Removal / Installation Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
Caution Access Procedures .............................. C-1
Always allow unit to cool before
removing trays or rack. R Wiring Diagrams ................................... D-1
Exploded Views / Part Numbers ........... E-1

Bottom Wire

Removal
Step 1: Lift up on left side of tray plate
until it "pops" free of rack.
Remove tray plate. Installation
Step 3: Push left side of plate down until it
engages w/ offset ends of rack.
Installation
Step 2: Insert the right side of tray plate
under bottom wire of rack.

Removal
Step 2: Squeeze bottom of rack together.
Pull rack out of chamber.

Installation
Step 1: Position rack w/ offset ends to the left.
Squeeze bottom of rack together.
Push rack into chamber.

Offset Ends Straight Ends


(left) (right)

Models: ALL Tray Plate / Rack


Serial Numbers:
© Midmark Corporation 2004 SF-1854
C-3
Back Go To Table Of Contents Next

Access Procedures
Draining / Filling the Reservoir
Refer To: Page
Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
Note: To remove drain hose, press release lever. Access Procedures .............................. C-1
Wiring Diagrams ................................... D-1
Exploded Views / Part Numbers ........... E-1

Filling
Step 1: Pour distilled water into reservoir until
the water level reaches the "full mark".
Do not overfill!

Draining
Step 1: Hold hose over a drain or suitable container.
Insert drain hose into coupling on front of unit.
Note: The max. reservoir capacity is: 1.3 gallon (4.9 liters)

Full
Mark

MA514002

Draining / Filling Models: ALL


Reservoir Serial Numbers:
C-4 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Wiring Diagrams
Wiring Diagrams Model
115 VAC models:
Page

& Schematics M7 (-011 / -013 / -014 / -015) ............ D-2

Section D M7 (-020 / -022) ................................ D-3


230 VAC models:
M7 (-012 / -016) ................................ D-4
M7 (-021)........................................... D-5

Models:
Serial Numbers:
© Midmark Corporation 2004 SF-1854
D-1
Back Go To Table Of Contents Next

Wiring Diagrams

Refer To: Page


Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
Access Procedures .............................. C-1
Wiring Diagrams ................................... D-1
POWER CORD
Exploded Views / Part Numbers ........... E-1

WHITE BLACK
BLACK WHITE

GREEN

GREEN
MANUAL-RESET MANUAL-RESET
OVERHEAT OVERHEAT
THERMOSTAT THERMOSTAT
( LOW WATER ) ( LOW WATER )
GROUND

RED
NC

AUTO-RESET
OVERHEAT
THERMOSTAT AUTO-RESET
( LOW WATER ) OVERHEAT
NC THERMOSTAT
( LOW WATER )

YELLOW

RED
TIMER WHITE
SWITCH

"A" "B" "C"

WHITE
HEATER LIGHT

YELLOW
HEATING ELEMENT

WHITE
BLACK
WHITE

TIMER MOTOR 015-1815-00


WHITE
WIRING HARNESS

TEMPERATURE
YELLOW REGULATOR TIMER
RELAY BUZZER

YELLOW
TIMER BUZZER
TEMPERATURE TM 1 3
REGULATOR
RELAY B
NC YELLOW
2 YELLOW
WHITE YELLOW
WHITE

HEATER LIGHT
TIMER
ASSEMBLY

YELLOW
HEATING ELEMENT

Wiring Diagrams Models: M7 (-011) M7 (-013) M7 (-014) M7 (-015)


(115 VAC Models) Serial Numbers: all all all all
D-2 © Midmark Corporation 2004 SF-1854 Rev. 5/09
Back Go To Table Of Contents Next

Wiring Diagrams

Refer To: Page


Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
POWER
RED
Access Procedures .............................. C-1
CORD BLACK MANUAL- RESET
POWER
OVERHEAT
THERMOSTAT
Wiring Diagrams ................................... D-1
CORD
F1 FUSE : (LOW WATER)
115V UNIT; 12A 250V FAST ACTING
o
Open 285 F (+/-12 F)
o
o Exploded Views / Part Numbers ........... E-1
BLACK BLACK (140.5 C)
o o
Closes 200 F (93 C)
7
WH ITE
9 WH ITE
F1 FUSE:

GREEN
RED 12A, 250V FAST ACTING
MANUAL- RESET 8 RED
BLACK 9
OVERHEAT
THERMOSTAT 7 BL
GREEN

AC
(LOW WATER) K
o o
Open 285 F (+/-12 F)
o
GROUND
(140.5 C)
o o
Closes 200 F (93 C) NC AUTO- RESET
OVERHEAT
8
8 RED GROUND THERMOSTAT
AUTO- RESET (LOW WATER)
OVERHEAT o o 3
Open 295 F (+/-12 F)
THERMOSTAT o
(146 C)
(LOW WATER) FILL / VENT o o
Closes 265 F (129 C)
o o
Open 295 F (+/-12 F) NC SWITCH FILL / VENT
o (Switch Contacts Open
(146 C)
o o
VALVE
Closes 265 F (129 C) Until Lever is pressed)

3 RED 7 6 5
NO BLACK
HEATER ELEMENT

RED

RED
1 115 VAC = 12 Ohms
YELLOW
TIMER
SWITCH YELLOW
TIMER MOTOR 2
WH ITE
3 B TM 4

WH ITE FILL / VENT


BLACK BLACK
SWITCH
PILOT
TM TEMPERATURE LIGHT 1
4 (Switch contacts OPEN
until lever is pressed)
REGULATOR WH ITE (Lit when heater is ON)
WHITE

RELAY
7

YELLOW 6
TIMER BUZZER 015-1790-00
YELLOW WIRING
YELLOW YELLOW HARNESS
FILL / VENT
VALVE

BLACK
TEMPERATURE 2
1
TIMER
REGULATOR BUZZER
RELAY HEATER LIGHT YL YL
NC TIMER YL
(Lit When Heater
ASSEMBLY 5
3

Is On) 6
WH ITE
1

WH ITE 4 1 YELLOW YL
5
TM

WH
3
RD

HEATER ELEMENT RED


WH ITE 2 YELLOW
YELLOW
115 VAC = 12 Ohms MA676900i

Models: M7 (-020) M7 (-022) Wiring Diagrams


(115 VAC Models)
Serial Numbers: all all
© Midmark Corporation 2004 SF-1854
D-3
Back Go To Table Of Contents Next

Wiring Diagrams

Refer To: Page


Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
Access Procedures .............................. C-1
Wiring Diagrams ................................... D-1
POWER CORD
Exploded Views / Part Numbers ........... E-1
BLUE BROWN
BROWN BLUE

GREEN / YELLOW
GREEN / YELLOW
MANUAL-RESET MANUAL-RESET
OVERHEAT OVERHEAT
THERMOSTAT THERMOSTAT
( LOW WATER ) ( LOW WATER )

RED
NC
RED
AUTO-RESET
OVERHEAT GROUND
THERMOSTAT AUTO-RESET
( LOW WATER ) OVERHEAT
NC THERMOSTAT
( LOW WATER )
RED
YELLOW

RED
TIMER WHITE
SWITCH

"A" "B" "C"

WHITE
HEATER LIGHT

YELLOW
HEATING ELEMENT

WHITE
BLACK
WHITE

TIMER MOTOR 015-1815-00


WHITE
WIRING HARNESS

TEMPERATURE
YELLOW REGULATOR TIMER
RELAY BUZZER

YELLOW
TIMER BUZZER
TEMPERATURE TM 1 3
REGULATOR
RELAY B
NC YELLOW
2 YELLOW
WHITE YELLOW
WHITE

HEATER LIGHT
TIMER
ASSEMBLY

YELLOW
HEATING ELEMENT

Wiring Diagrams Models: M7 (-012) M7 (-016)


(230 VAC Models) Serial Numbers: all all
D-4 © Midmark Corporation 2004 SF-1854 Rev. 5/09
Back Go To Table Of Contents Next

Wiring Diagrams

Refer To: Page


Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
RED Access Procedures .............................. C-1
POWER
CORD BROWN MANUAL- RESET
OVERHEAT
Wiring Diagrams ................................... D-1
POWER
THERMOSTAT
CORD
F1 FUSE : (LOW WATER)
o o
Exploded Views / Part Numbers ........... E-1
230V UNIT; 8A 250V FAST ACTING Open 285 F (+/-12 F)
o
BLACK BROWN (140.5 C)
o o
Closes 200 F (93 C)
BLUE 7
9 BLUE
F1 FUSE:

GREEN / YELLOW
RED GREEN / YELLOW
MANUAL- RESET 8A, 250V FAST ACTING 8 RED
BLACK 9
OVERHEAT
THERMOSTAT 7 BL
AC
(LOW WATER) K
o o
Open 285 F (+/-12 F)
o
(140.5 C)
o o
Closes 200 F (93 C) NC AUTO- RESET
GROUND OVERHEAT
8
8 RED GROUND THERMOSTAT
AUTO- RESET (LOW WATER)
OVERHEAT o o 3
Open 295 F (+/-12 F)
THERMOSTAT o
(146 C)
(LOW WATER) FILL / VENT o o
Closes 265 F (129 C)
o o
Open 295 F (+/-12 F) NC SWITCH FILL / VENT
o (Switch contacts Open
(146 C)
o o
VALVE
Closes 265 F (129 C) until leveris pressed)

3 RED 7 6 5
NO BLACK
HEATER ELEMENT

RED

RED
1 230 VAC = 49 Ohms
YELLOW
TIMER
SWITCH YELLOW
TIMER MOTOR 2
WH ITE
3 B TM 4

WH ITE FILL / VENT


BLACK BLACK
SWITCH
PILOT
TM TEMPERATURE LIGHT 1
4 (Switch contacts OPEN
until lever is pressed)
REGULATOR WH ITE (Lit when heater is ON)
WHITE

RELAY
7

YELLOW 6
TIMER BUZZER 015-1790-00
YELLOW WIRING
YELLOW YELLOW HARNESS
FILL / VENT
VALVE

BLACK
TEMPERATURE 2
1
TIMER
REGULATOR BUZZER
RELAY HEATER LIGHT YL YL
NC TIMER YL
(Lit when heater is ON) ASSEMBLY 5
3

6
WH ITE
1

WH ITE 4 1 YELLOW YL
5
TM

WH
3
RD

HEATER ELEMENT RED


WH ITE 2 YELLOW
YELLOW
230 VAC = 49 Ohms MA677100i

Models: M7 (-021) Wiring Diagrams


all (230 VAC Models)
Serial Numbers:
© Midmark Corporation 2004 SF-1854
D-5
Back Go To Table Of Contents Next
Back Go To Table Of Contents Next

Model Page

Exploded Views & M7 (-011 thru -016) ...................... E-2


M7 (-020 thru -022) ...................... E-3

Parts Lists
Section E

Always Specify Model & Serial Number

Models:
Serial Numbers:
© Midmark Corporation 2004 SF-1854
E-1
Back Go To Table Of Contents Next

M7(-011 thru -016)


Fuses .............. E-15
Items not shown:
Heating Element
& Thermostats .......... E-13
Power Cord ................. E-14
Rack & Trays ............... E-15
Packaging .................... E-16 Labels & Decals .......... E-17

Temperature Regulator
Components ............... E-7
Covers ......................E-4
includes:
includes: temp. relay & diaphragm cup
main cover, base plate,
leveling feet, fill cap, etc.

Reservoir ..................... E-8


includes:
Chamber Fuses ..............
condensing E-15
coil & drain tube
Components ............... E-10
includes:
fittings & mounting bracket

Bellows & Plumbing ... E-11*


Door Components ...... E-6
includes:
hinge, & door gasket Timer / Buzzer ............. E-12
Timer & Buzzer ........... E-12

Front Panel Fill / Vent Valve


Components ............E-5 & Plumbing ................. E-9*
includes:
heater light, thermometer,
knobs, & drain coupling MA677800i

E-2 * Indicates multiple pages due to a serial number break for the parts illustration
Back Go To Table Of Contents Next

M7(-020 thru -022)


Items not shown:
Heating Element
& Thermostats .......... E-13.1
Power Cord & Fuse ..... E-14.1
Rack & Trays ............... E-15 Labels & Decals .......... E-17.1
Packaging .................... E-16

Temperature Regulator
Components ............... E-7
Covers ......................E-4.1 includes:
temp. relay & diaphragm cup
includes:
main cover, base plate,
leveling feet, fill cap, etc.

Reservoir Tank ............ E-8.1


includes:
Chamber condensing coil & drain tube
Components ............... E-10.1
includes:
fittings & mounting bracket

Door Components ...... E-6 Bellows & Plumbing ... E-11.2

includes:
hinge, & door gasket
Timer & Buzzer ........... E-12

Fill / Vent Valve


Front Panel & Plumbing ................. E-9.3
Components ............E-5.1
includes:
includes: fill/vent switch
heater light, thermometer,
knobs, & drain coupling MA677801i

* Indicates multiple pages due to a serial number break for the parts illustration E-3
Back Go To Table Of Contents Next

Refer To: Page


Operation & Troubleshooting ....... A-1
H98009 5
6 H98135 Component Testing / Repair ........ B-1
Access Procedures...................... C-1
Wiring Diagrams .......................... D-1
053-1037-00 2 Exploded Views / Part Numbers .. E-1

3 030-1198-00

002-0643-00 1

7 045-0001-31

H98011 4

122357 9
8 040-0010-125
050-4286-00 11

Item Description Qty.


1 Cover Kit (Includes items 2 thru 4) .............. 1
2 • Cooling Rail ............................................... 2
3 • Main Cover ................................................ 1
4 • Push Nut .................................................... 6
5 Fill Cap ......................................................... 1
6 Rubber Seal ................................................. 1
7 Lockwasher .................................................. 6
10 002-0822-00 8 Screw (#10 x 1/2", self-tapping) ................... 6
9 Keps Nut ....................................................... 4
10 Rubber Foot Kit (includes nut) ..................... 4
8 11 Base Plate .................................................... 1
12 050-4423-00 12 Inspection Cover .......................................... 1
Always Specify Model & Serial Number
MA674101i

Covers Models: M7 (-011 thru -016)


Serial Numbers: all
E-4 © Midmark Corporation 2004 SF-1854 Rev. 8/05
Back Go To Table Of Contents Next

Refer To: Page


H98009 Operation & Troubleshooting ....... A-1
Component Testing / Repair ........ B-1
H98135 Access Procedures ...................... C-1
Wiring Diagrams .......................... D-1
053-1037-00
Exploded Views / Part Numbers .. E-1

002-0840-00

045-0001-31

H98011

040-0010-163
029-3408-00 10
041-0010-04
Item Description Qty.
1 Fill Cap ......................................................... 1
2 Rubber Seal ................................................. 1
3 Main Cover w/ labels
(includes items 4 & 5) ................................ 1
4 Push Nut ....................................................... 6
5 Cooling Rail .................................................. 2
6 Lockwasher .................................................. 1
7 Screw (#10 x 5/8", self-drilling/tapping) ....... 7
002-0822-00 8 Nut (used only on units built prior to serial
number V411489) ........................................ 4
9 Rubber Foot Kit (includes nut) .................... 4
10 Base Assembly ............................................ 1
11 Inspection Cover ......................................... 1
11 050-4423-00
Always Specify Model & Serial Number
MA674101i

Models: M7 (-020 thru -022) Covers


Serial Numbers: all
© Midmark Corporation 2004 SF-1854 Rev. 4/07
E-4.1
Back Go To Table Of Contents Next

Refer To: Page


Operation & Troubleshooting ....... A-1
Ritter: 002-0688-00
Midmark: 002-0688-01 Component Testing / Repair ........ B-1
Access Procedures ...................... C-1
050-4285-10 040-0010-75 Wiring Diagrams .......................... D-1
045-0001-08
Exploded Views / Part Numbers .. E-1
Ritter/Midmark: 061-0774-00
Dabi Alante: 061-0774-02 H224033

041-0250-07

014-0183-01

053-0404-07

057-0461-00 045-0001-21

014-0192-00

100 / 120 VAC: 015-1304-00


230 VAC: 015-1304-01 Item Description Qty.
1 Front Panel Assy. (includes items 2 & 3) ..... 1
002-0242-00 2 • Front Panel (w/o label) .............................. 1
3 • Label .......................................................... 1
4 Alignment Pin ............................................... 1
5 Lockwasher .................................................. 1
6 Screw ........................................................... 1
7 Thermometer Bracket .................................. 1
8 Jam Nut ........................................................ 1
9 Refer to: Temp. Regulator Components E-7
10 Refer to: Fill / Vent Valve ......................... E-9
11 Shut-Off Coupling ........................................ 1
053-1008-00 053-0990-00 12 Lever Knob .................................................. 1
13 Timer Knob (includes set screw) ................ 1
14 Temperature Knob (includes set screw) .... 1
15 Refer to: Heating Element /
053-0987-00
Overheat Thermostats ....................... E-13
053-1008-01 16 Heater Light ................................................. 1
17 Thermometer Kit (incl. items 18 thru 21) .... 1
18 • Compression Nut ...................................... 1
19 • Washer (#6 - 5/16" O.D.) ......................... 1
20 • Rubber Sleeve .......................................... 1
21 • Male Connector ........................................ 1
Always Specify Model & Serial Number
MA671907i

Front Panel Models: M7 (-011 thru -016)


Components
E-5 © Midmark Corporation 2004 SF-1854 Rev. 6/09
Serial Numbers: all
Back Go To Table Of Contents Next

Refer To: Page


Operation & Troubleshooting ....... A-1
Component Testing / Repair ........ B-1
Access Procedures ...................... C-1
045-0001-08 040-0010-75 Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1
050-6034-10

Ritter: 061-0900-00 041-0250-20


Midmark: 061-0900-01

014-0183-01
057-0461-00
053-0404-07

045-0001-21

014-0192-00
100 / 120 VAC: 015-1304-00
230 VAC: 015-1304-01

002-0242-00
Item Description Qty.
1 Front Panel (w/o label) ................................ 1
2 Label ............................................................. 1
3 Alignment Pin ............................................... 1
4 Lockwasher ................................................. 1
5 Screw ........................................................... 1
6 Jam Nut ........................................................ 1
7 Refer to: Temp. Regulator Components E-7
8 Shut-Off Coupling ........................................ 1
053-1008-00 9 Refer to: Fill / Vent Valve ......................... E-9
10 Timer Knob (includes setscrew) ................. 1
053-0990-00 11 Temperature Knob (includes setscrew) ..... 1
12 Refer to: Heating Element /
Overheat Thermostats ....................... E-13
053-1008-01 13 Heater Light ................................................. 1
14 Thermometer Kit (incl. items 15 thru 18) .... 1
15 • Compression Nut ...................................... 1
16 • Washer ...................................................... 1
17 • Rubber Sleeve .......................................... 1
18 • Male Connector ........................................ 1
Always Specify Model & Serial Number
MA671908i

Models: M7 (-020 thru -022) Front Panel


Components
Serial Numbers: all
© Midmark Corporation 2004 SF-1854 Rev. 5/09
E-5.1
Back Go To Table Of Contents Next

Refer To: Page


Operation & Troubleshooting ....... A-1
040-0010-103 16 Component Testing / Repair ........ B-1
Access Procedures ...................... C-1
H97991 15 Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1
17 002-0243-00
H96773 14
18 040-0250-69
H96771 13
053-1579-00 12 19
H106313 11
002-0249-00 1 20 H106330
H96774 10

H96775 9
Item Description Qty.
H96776 8 1 Door Assembly(incl. items 2 thru 13) ........... 1
2 • Handle ....................................................... 1
H96772 7 3 • Push Nut .................................................... 1
4 • Pin ............................................................. 1
5 5 • Refer to Label Location ......................... Ref
6 • Cross Arm .................................................. 1
042-0005-16 4 7 • T-bolt ......................................................... 1
8 • U-bracket ................................................... 1
9 • Long Spring ............................................... 1
21 041-0250-15 10 • Short Spring .............................................. 1
11 • Door ........................................................... 1
041-0010-10 24 12 • Gasket ....................................................... 1
22 045-0000-04 13 • Washer ................................................... 1
14 • Cap Nut (apply adhesive #042-0024-02) .. 1
050-4783-00 23 15 Door Stop ..................................................... 1
16 Bolt ............................................................... 1
17 Door Gasket ................................................. 1
050-2359-00 6 18 Bolt ............................................................... 1
053-0513-00 2 19 Refer to Chamber Components ............... Ref
20 Spring ........................................................... 1
3 K906034 21 Torque Nut ................................................... 1
22 Washer ......................................................... 1
23 Hinge ............................................................ 1
24 Torque Nut ................................................... 1
Always Specify Model & Serial Number
MA511703i

Door Components Models: ALL


Serial Numbers:
E-6 © Midmark Corporation 2004 SF-1854 Rev. 5/05
Back Go To Table Of Contents Next

Refer To: Page


Operation & Troubleshooting ....... A-1
Component Testing / Repair ........ B-1
H97075 Access Procedures ...................... C-1
Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1

H97076

H97966

H97948

040-0008-47

040-0006-38

040-0006-44

Item Description Qty.


1 Temperature Relay Kit
002-0838-00 H98064 (incl. items 2 & 3) ...................................... 1
2 • Spacer ....................................................... 1
3 • Screw (#6-32 x 7/16") ................................ 3
H97951 4 Flexible Shaft Assembly ............................... 1
5 Set Screw ..................................................... 1
6 Diaphragm Cup ............................................ 1
7 Gasket .......................................................... 1
8 Nut ................................................................ 1
9 Lockwasher .................................................. 1

Always Specify Model & Serial Number


MA511902i

Models: ALL Temperature Regulator


Serial Numbers: Components
© Midmark Corporation 2004 SF-1854 Rev 1/08
E-7
Back Go To Table Of Contents Next

1 053-0944-00 Refer To: Page


Operation & Troubleshooting ....... A-1
Component Testing / Repair ........ B-1
Access Procedures ...................... C-1
Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1
H98089 2

014-0324-00 3

053-1036-01 4

5 029-2291-00
7

053-0391-02 10 6 014-0183-02

Item Description Qty.


7 014-0209-00
053-0990-01 13 11 1 Tank with Lid (not available separately) ...... 1
2 Pressure Relief Valve .................................. 1
8 053-1036-00 3 Bulkhead Fitting (includes nut) .................... 1
4 Neoprene Washer ........................................ 1
5 Condensing Coil Assy. (incl. items 6 & 7) ... 1
15 040-0010-75 6 • Compression Fitting .................................. 1
7 • Bulkhead Fitting (includes nut) ................. 1
9 8 Neoprene Washer ........................................ 1
016-0748-01
14 053-0391-03 9 Hose Clamp ................................................. 2
10 Tank Drain Tube .......................................... 1
11 Refer to: Front Panel Components .......... Ref
12 Drain Hose Kit (incl. items 13 thru 15) ......... 1
13 • Barbed Fitting ............................................ 1
14 • Removeable Drain Tube ........................... 1
15 Screw ........................................................... 2
12 002-0653-00
Always Specify Model & Serial Number
MA675105i

Reservoir Models: M7 (-011 thru -016)


E-8 © Midmark Corporation 2004 SF-1854 Rev. 1/09
Serial Numbers: all
Back Go To Table Of Contents Next

Refer To: Page


Operation & Troubleshooting ....... A-1
Component Testing / Repair ........ B-1
053-0944-00 2 Access Procedures ...................... C-1
Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1

H98089 3

014-0324-00 4
H97966 5
029-3302-00 1

6 014-0432-00
8

053-0391-02 11
7 014-0485-00
Item Description Qty.
12 014-0209-00 1 Reservoir Assembly (incl. items 2 thru 11) 1
8 2 • Tank with Lid (not available separately) .. 1
053-0990-01 3 • Pressure Relief Valve ............................... 1
14 9 053-1036-01 4 • Bulkhead Fitting (includes nut) ................. 1
5 • Washer ...................................................... 2
6 • Condensing Coil ........................................ 1
16 040-0010-75
7 • Compression Fitting .................................. 1
016-0748-01 8 • Bulkhead Fitting (includes nut) ................. 1
10 9 • Neoprene Washer .................................... 1
15 053-0391-03 10 • Hose Clamp .............................................. 2
11 • Tank Drain Tube ....................................... 1
12 Refer to: Front Panel Components ......... Ref
13 Drain Hose Kit (incl. items 13 thru 15) ........ 1
14 • Barbed Fitting ............................................ 1
15 • Removeable Drain Tube .......................... 1
002-0653-00
13 16 Screw ........................................................... 2

Always Specify Model & Serial Number


MA675106i

Models: M7 (-020 thru -022) Reservoir


Serial Numbers: all
© Midmark Corporation 2004 SF-1854 Rev. 1/09
E-8.1
Back Go To Table Of Contents Next

Refer To: Page


Operation & Troubleshooting ....... A-1
Component Testing / Repair ........ B-1
Access Procedures ...................... C-1
014-0183-02 Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1
H98136
002-0247-00
H98069

H98138
H98070

18
041-0010-09
16 052-0332-00
H98077
17 052-0331-00

H98068
18 014-0184-01
041-0010-10 10
13 040-0010-96

12 050-1926-00 Item Description Qty.


1 Fill/Vent Valve Kit (incl. items 2 thru 9) ........ 1
2 • Compression Fitting .................................. 2
050-4315-00 11 3 • O-ring (apply hi-temp lube) ....................... 1
4 • Plunger ...................................................... 1
5 • O-ring (apply hi-temp lube) ....................... 1
15 H98072 6 • Valve Body ................................................ 1
7 • Spring ........................................................ 1
8 • Nut (#10-32, brass) ................................... 1
9 • Bracket ...................................................... 1
10 Torque Nut ................................................... 1
11 Lever ............................................................ 1
14 042-0010-07 12 Lever Bracket ............................................... 1
13 Screw (#10-32 x 3/4") ................................. 1
14 Pop Rivet ..................................................... 3
15 Clip ............................................................... 3
16 Tubing (Fill) .................................................. 1
17 Tubing (Vent) ............................................... 1
18 Elbow Fitting ................................................. 2

Always Specify Model & Serial Number


MA674600i

Fill / Vent Valve Models: M7 (-011) M7 (-012) M7 (-014)


& Plumbing MH1000 thru MH1802 ML1000 thru ML2133
E-9 © Midmark Corporation 2004 SF-1854 Rev. 4/09
Serial Numbers: MJ1000 thru MJ1091
Back Go To Table Of Contents Next

Refer To: Page


Operation & Troubleshooting ....... A-1
Component Testing / Repair ........ B-1
Access Procedures ...................... C-1
Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1

014-0183-02 002-0648-00

050-4793-00 13
11 052-0332-00
040-0010-96

041-0010-10 12 052-0331-00
014-0369-00
13 014-0184-01

8 050-1926-00

050-4315-00
016-0843-00 Item Description Qty.
1 Fill/Vent Valve Kit (incl. items 2 thru 6) ........ 1
2 • Compression Fitting .................................. 2
3 • Bracket ...................................................... 1
4 • Valve Body ................................................ 1
042-0010-15 5 • Clip ............................................................ 3
6 • Pop Rivet ................................................... 3
7 Lever ............................................................ 1
8 Lever Bracket ............................................... 1
9 Torque Nut ................................................... 1
10 Screw (#10-32 x 3/4") ................................. 1
11 Tubing (Fill) .................................................. 1
12 Tubing (Vent) ............................................... 1
13 Elbow Fitting ................................................. 2

Always Specify Model & Serial Number


MA674601i

Models: M7 (-011) M7 (-012) M7 (-014) Fill / Vent Valve


MH1803 thru MH3104 MJ1092 thru MJ1241 ML2132 thru ML4209 & Plumbing
Serial Numbers:
© Midmark Corporation 2004 SF-1854 Rev. 4/09
E-9.1
Back Go To Table Of Contents Next

Refer To: Page


Operation & Troubleshooting ....... A-1
Component Testing / Repair ........ B-1
Access Procedures ...................... C-1
Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1

014-0402-00 12
002-0649-00
10 052-0332-00

11 052-0331-00
H98077
041-0010-10
12 014-0184-01
040-0010-96

7 050-1926-00

050-4315-00 016-0843-00

Item Description Qty.


1 Fill/Vent Valve Kit (incl. items 2 thru 5) ........ 1
042-0010-15 2 • Valve Body (includes nut) ......................... 1
3 • Bracket ...................................................... 1
4 • Clip ............................................................ 3
5 • Pop Rivet ................................................... 3
6 Lever ............................................................ 1
7 Lever Bracket ............................................... 1
8 Torque Nut ................................................... 1
9 Screw (#10-32 x 3/4") ................................. 1
10 Tubing (Fill) .................................................. 1
11 Tubing (Vent) ............................................... 1
12 Elbow Fitting ................................................. 2
Always Specify Model & Serial Number
MA674602i

Fill / Vent Valve Models: M7 (-011) M7 (-012) M7 (-014) M7 (-013/-015 /-016)


MH3105 thru present MJ1242 thru present ML4210 thru present all
& Plumbing Serial Numbers:
E-9.2 © Midmark Corporation 2004 SF-1854 Rev. 4/09
V2200 thru present V2200 thru present V2200 thru present
Back Go To Table Of Contents Next

Refer To: Page


Operation & Troubleshooting ....... A-1
Component Testing / Repair ........ B-1
Access Procedures...................... C-1
Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1

014-0433-00

115 VAC: 014-0236-10

115 VAC: 014-0236-00


230 VAC: 014-0236-01

014-0183-02
014-0236-11 014-0184-01

Item Description Qty.


015-1791-00
1 Fill/Vent Valve .............................................. 1
2 Mounting Bracket ......................................... 1
3 Screw (#10-32 x 3/8") .................................. 2
4 Bolt (1/4-20 x 1/2") ....................................... 2
5 Elbow Fitting ................................................. 1
6 Male Fitting (3/8") ......................................... 1
050-6057-00 7 Tubing .......................................................... 1
040-0250-157 8 Fill / Vent Switch ........................................... 1
9 Replacement Coil ........................................ 1
10 Replacement Kit (Includes plunger, guide,
040-0010-74 and seals) .................................................... 1
Always Specify Model & Serial Number
MA674605i

Models: M7 (-020 thru -022) Fill / Vent Valve


Serial Numbers: all & Plumbing
© Midmark Corporation 2004 SF-1854 Rev 2/08
E-9.3
Back Go To Table Of Contents Next

Refer To: Page


Operation & Troubleshooting ....... A-1
1 115 VAC: 029-2304-00
Component Testing / Repair ........ B-1
230 VAC: 029-2304-01
100 VAC: 029-2304-02 Access Procedures ...................... C-1
12 H99387
Wiring Diagrams .......................... D-1
H99388-29 11
2 014-0183-01 Exploded Views / Part Numbers .. E-1

3 030-0976-00

4 H99388-27

5 051-0721-00

6 040-0250-47

H97980 10

The chamber bracket (item #6) on some models may be attached to the base plate w/ pop rivets.
When installing a new bracket, replace rivets w/ screws (item #7).
041-0250-15 9
014-0184-01 8 Item Description Qty.
1 Chamber Assembly - complete
includes items 2 & 3 and the following: ..... 1
• Heating Element & Thermostats (2)
• Temperature Regulator Assembly
• Door Hinge & Door Gasket
2 • Compression Fitting .................................. 1
3 • Chamber Shell only .................................. 2
4 Banding Strap (end) ..................................... 1
5 Chamber Bracket ......................................... 1
6 Screw ........................................................... 1
7 Screw ........................................................... 2
8 Elbow Fitting ................................................. 1
9 Nut ................................................................ 1
7 040-0010-74 10 Insulation ...................................................... 1
11 Banding Strap .............................................. 2
12 Banding Clip ................................................. 2
Always Specify Model & Serial Number
MA676600i

Chamber Components Models: M7 (-011 thru -016)


Serial Numbers:
E-10 © Midmark Corporation 2004 SF-1854
all
Back Go To Table Of Contents Next

Refer To: Page


Operation & Troubleshooting ....... A-1
Component Testing / Repair ........ B-1
H99388-29 7
8 H99387 Access Procedures ...................... C-1
Wiring Diagrams .......................... D-1
2 014-0183-01
Exploded Views / Part Numbers .. E-1
1 115 VAC: 029-3278-00
230 VAC: 029-3278-01

3 002-0360-00

4 014-0184-01

H97980 6

Item Description Qty.


1 Chamber Assembly - complete
includes items 2 thru 4 & the following: .... 1
• Heating Element & Thermostats (2)
• Temperature Regulator Assembly
• Door Hinge & Door Gasket
• Timer Buzzer
2 • Compression Fitting .................................. 1
5 040-0250-157 3 • Filter Screen .............................................. 1
4 • Elbow Fitting .............................................. 1
5 Screw ........................................................... 1
6 Insulation ...................................................... 1
7 Banding Strap .............................................. 2
8 Banding Clip ................................................. 2

Always Specify Model & Serial Number


MA676601i

Models: M7 (-020 thru -022) Chamber Components


Serial Numbers: all
© Midmark Corporation 2004 SF-1854 Rev. 1/11
E-10.1
Back Go To Table Of Contents Next

Refer To: Page


Older Models w/ rubber Operation & Troubleshooting ....... A-1
tubing MUST use Item #6
002-0679-00 6 Component Testing / Repair ........ B-1
Access Procedures ...................... C-1
8 H335943 Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1
7
13
9 014-0184-00

10

014-0058-00 12
10
Item Description Qty.
1 1 Refer to: Reservoir ................................... Ref
2 Tube ............................................................. 1
11 014-0183-01 3 Compression Nut ......................................... 2
4 Bellows ......................................................... 1
014-0241-00 5 5 Tube ............................................................. 1
6 Chamber Manifold Tube Kit
(includes items 7 thru 9)
7 • Refer to: Chamber Assembly ................ Ref
3 8 • Tube .......................................................... 1
9 • Elbow Fitting .............................................. 1
10 Refer to: Front Panel Components .......... Ref
002-0375-00 4 11 Compression Fitting ..................................... 1
12 Street Tee Fitting ......................................... 2
13 Refer to: Chamber Assembly .................. Ref
2 014-0240-00
Always Specify Model & Serial Number
014-0392-00 3 MA675600i

Bellows & Plumbing Models: M7 (-011) M7 (-012) M7 (-014)


E-11 Serial Numbers: MH1000 thru MH6044 MJ1000 thru MJ1696 ML1000 thru ML7297
© Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Refer To: Page


Operation & Troubleshooting ....... A-1
Component Testing / Repair ........ B-1
Access Procedures ...................... C-1
Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1

12 H335943
014-0183-01 13
11
10 014-0184-00

Original Plumbing Reconfigured Plumbing


Configuration Configuration

(If plumbing has been


reconfigured, refer to next
page for replacement parts)

8
9

Item Description Qty.


014-0058-00 8 1 1 Refer to: Reservoir ................................... Ref
2 Bellows Kit (includes items 3 thru 7) ........... 1
3 Tube ............................................................. 1
3 4 Compression Nut ......................................... 3
5 Bellows ......................................................... 1
7 4 6 Tube ............................................................. 1
7 Compression Fitting ..................................... 1
6 2 002-0375-01 8 Street Tee Fitting ......................................... 2
5 9 Refer to: Front Panel Components .......... Ref
Note: Item #2 will reconfigure 10 Elbow Fitting ................................................. 1
Bellows plumbing to match 11 Refer to: Chamber Assembly .................. Ref
4 current models. 12 Tube ............................................................. 1
13 Compression Fitting ..................................... 1

Always Specify Model & Serial Number


MA675601i

Models: M7 (-011) M7 (-012) M7 (-014) M7 (-013/-015 /-016)


MH6045 thru present MJ1697 thru present ML7298 thru present Bellows & Plumbing
Serial Numbers: all
V2200 thru present V2200 thru present V2200 thru present
© Midmark Corporation 2004 SF-1854
E-11.1
Back Go To Table Of Contents Next

10 9 014-0480-01
Refer To: Page
Operation & Troubleshooting ....... A-1
Component Testing / Repair ........ B-1
Access Procedures ...................... C-1
Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1

1
7 014-0484-00

014-0481-00 6

4
Item Description Qty.
1 Refer to: Reservoir ................................... Ref
5 2 Bellows Kit .................................................... 1
002-0375-00
002-0375-01 2 3 • Tube .......................................................... 1
4 • Compression Nut ....................................... 2
5 • Bellows ...................................................... 1
6 • Tube .......................................................... 1
7 • Compression Fitting .................................. 2
014-0392-00 4 8 Refer to: Front Panel Components .......... Ref
9 Tube ............................................................. 1
10 Refer to: Chamber Components .............. Ref
014-0482-00 3 Always Specify Model & Serial Number
MA675602i

Bellows & Plumbing Models: M7 (-020 thru -022)


Serial Numbers: all
E-11.2 © Midmark Corporation 2004 SF-1854 Rev 10/05
Back Go To Table Of Contents Next

Refer To: Page


Operation & Troubleshooting ....... A-1
Component Testing / Repair ........ B-1
Access Procedures...................... C-1
Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1

3 115 VAC: 015-1813-00


230 VAC: 015-1813-01

040-0006-44 4

5 015-1815-00
[Ext. Wiring Harness - not shown]

Item Description Qty.


1 115 VAC: 002-0373-00 1 Timer Assembly Kit (incl. items 2 thru 5) ..... 1
230 VAC: 002-0373-01 2 • Screw (#8-32 x 5/16") ................................ 1
230 VAC -016: Not available 3 • Buzzer ....................................................... 1
4 • Screw (#6-32 x 7/16") ................................ 1
5 • Extended Wiring Harness - not shown
2 040-0008-29
(required for older units only) ................. 1
Always Specify Model & Serial Number
MA671802i

MA671801i
Models: M7 (-011 thru -015) M7 (-020 thru -022) Timer / Buzzer
Serial Numbers: all all

© Midmark Corporation 2004 SF-1854 Rev. 8/05


E-12
Back Go To Table Of Contents Next

115 VAC: 002-0839-00 Refer To: Page


230 VAC: 015-0839-01 Operation & Troubleshooting ....... A-1
100 VAC: 015-0894-02
Component Testing / Repair ........ B-1
Access Procedures ...................... C-1
H225913 9 Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1

H98137

H225639

10 041-0010-10

002-0370-00

030-0929-00
11 030-1484-00
H97971
Old-style rocker assembly is not available.
045-0001-56 Replace with item #11.
Item Description Qty.
1 Heating Element (includes items 2 & 3) ...... 1
H97949 2 • Nut ............................................................. 2
3 • Gasket ....................................................... 2
4 Overheat Thermostat (manual-reset) .......... 1
H225620 12 5 Overheat Thermostat (auto-reset) ............. 1
6 Bracket ......................................................... 1
7 Washer ......................................................... 2
057-0520-00 16 8 Lockwasher ................................................. 2
9 Heater Spacer ............................................. 1
10 Nut ................................................................ 1
053-1035-00 15 11 Rocker Assembly ........................................ 1
13 040-0010-74
12 Spacer .......................................................... 1
H225619 14 13 Screw ........................................................... 1
14 Bracket ......................................................... 1
15 Reset Button ................................................ 1
16 Reset Rod .................................................... 1
Always Specify Model & Serial Number
MA673901i

Heating Element & Models: M7 (-011 thru -016)


Thermostats Serial Numbers: all
E-13 © Midmark Corporation 2004 SF-1854 Rev. 6/09
Back Go To Table Of Contents Next

Refer To: Page


115 VAC: 002-0839-00
230 VAC: 002-0839-01 Operation & Troubleshooting ....... A-1
Component Testing / Repair ........ B-1
Access Procedures ...................... C-1
Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1

H98137

H225639

11 053-1461-00

002-0370-00

030-1521-00

053-0068-13
10 H97971
025-0078-00
Item Description Qty.
1 Heating Element Kit
9 045-0001-56 (includes nuts, e-clips & item 2) ............... 1
2 • Gasket ...................................................... 2
3 Overheat Thermostat (manual-reset) ........ 1
4 Overheat Thermostat (auto-reset) ............. 1
5 Bracket ......................................................... 1
6 Bushing ........................................................ 1
053-1460-00
7 Spring ........................................................... 1
8 Reset Rod .................................................... 1
9 Lockwasher ................................................. 2
10 Washer ......................................................... 2
11 Reset Button Actuator ................................. 1
Always Specify Model & Serial Number
MA673903i

Models: M7 (-020 thru -022) Heating Element &


Serial Numbers: all Thermostats
© Midmark Corporation 2004 SF-1854 Rev 6/09
E-13.1
Back Go To Table Of Contents Next

Refer To: Page


Operation & Troubleshooting ....... A-1
Component Testing / Repair ........ B-1
Access Procedures ...................... C-1
Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1

Attention
These models do not have a fuse.

1 115 VAC: H98159


230 VAC: 015-0688-13

122357 4

2 H98160

040-0010-74 3
Item Description Qty.
1 Power Cord .................................................. 1
2 Strain Relief .................................................. 1
3 Screw (#10-32 x 3/8") .................................. 1
4 Keps Nut ....................................................... 1
Always Specify Model & Serial Number
MA677200i

Power Cord Models: M7 (-011 thru -016)


Serial Numbers: all
E-14 © Midmark Corporation 2004 SF-1854
Back Go To Table Of Contents Next

Refer To: Page


Operation & Troubleshooting ....... A-1
Component Testing / Repair ........ B-1
Access Procedures ...................... C-1
Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1
115 VAC: 015-1259-04 3
230 VAC: 015-1259-05

122357 4 2 a.) 115 VAC: 015-0346-35


b.) 230 VAC: 015-0346-11

1 115 VAC: H98159


230 VAC: 015-0688-13

6 H98160

Item Description Qty.


1 Power Cord .................................................. 1
040-0010-74 5 2 Fuse:
a) 12A, 250V, Fast-Acting, 1/4" x 1-1/4" ... 1
b) 8A, 250V, Fast-Acting, 5mm x 20mm ... 1
3 Fuse Holder .................................................. 1
4 Screw (#10-32 x 3/8") .................................. 1
5 Keps Nut ....................................................... 1
6 Strain Relief .................................................. 1
Always Specify Model & Serial Number
MA677201i

Models: M7 (-020 thru -022) Power Cord / Fuse


Serial Numbers: all
© Midmark Corporation 2004 SF-1854 Rev. 7/05
E-14.1
Back Go To Table Of Contents Next

Refer To: Page


Operation & Troubleshooting ....... A-1
Component Testing / Repair ........ B-1
Access Procedures...................... C-1
Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1

1 002-0251-00

2 002-0252-00

3 002-0253-00

Item Description Qty.


1 Tray Rack ..................................................... 1
2 4 inch Tray ................................................... 1
3 5 inch Tray ................................................... 2
Always Specify Model & Serial Number
SA102200i

Rack & Trays Models: ALL


Serial Numbers:
E-15 © Midmark Corporation 2004 SF-1854 Rev. 1/07
Back Go To Table Of Contents Next

Refer To: Page


Operation & Troubleshooting ....... A-1
054-0326-00 Component Testing / Repair ........ B-1
2
Access Procedures ...................... C-1
3 054-0327-00
Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1

4 054-0325-00
027-1040-00 1 054-0324-00 5

6 066-0857-00

Item Description Qty.


1 M7 Packaging Kit - Generic
(includes items 2 thru 8) ............................ 1
2 • Top Front Pad ........................................... 1
3 • Top Rear Pad ............................................ 1
4 • Bottom Rear Pad ....................................... 1
5 • Bottom Front Pad ...................................... 1
6 • Carton ........................................................ 1
7 • Sterilizer Cleaner (not shown) ................... 1
8 • MSDS Sheet (not shown) ......................... 1
Always Specify Model & Serial Number
MA514703i

Models: ALL Packaging


Serial Numbers:
© Midmark Corporation 2004 SF-1854 Rev. 01/10
E-16
Back Go To Table Of Contents Next
Ritter/Midmark: 061-0443-00 1
Dabi Alante: 061-0594-00 Refer To: Page
Operation & Troubleshooting ....... A-1
Component Testing / Repair ........ B-1
Ritter/Midmark: H284693 2 Access Procedures ...................... C-1
Dabi Alante: 061-0593-00
Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1

Ritter/Midmark: 061-0649-00 9
R

Dabi Alante: 061-0599-00

3 Ritter/Midmark: H335669
Dabi Alante: 061-0598-00

10

4 Ritter/Midmark: 061-0203-00
Dabi Alante: 061-0600-00

Item Description Qty.


M7 (-011/-013/-014): 061-0519-00 8 1 Operating Instructions Plate ........................ 1
M7 (-012): 061-0527-00
5 2 Danger Plate (Dabi Alante only) .................. 1
M7 (-015): 061-0601-00 3 Warning Label .............................................. 1
M7 (-016): 061-0607-00 4 Caution Label ............................................... 1
6 061-0301-00 5 Serial Number Label (large- n/a) ................. 1
6 UL Label (applicable units only) ................... 1
7 CSA Label (applicable units only) ................ 1
7 061-0509-00 8 Wiring Diagram Label .................................. 1
9 Caution HOT Label ...................................... 1
10 Serial Number Label (small - n/a) ................ 1
Always Specify Model & Serial Number
MA674300i

Labels & Decals Models: M7 (-011 thru -016)


Serial Numbers: all
E-17 © Midmark Corporation 2004 SF-1854 Rev. 5/07
Back Go To Table Of Contents Next

Refer To: Page


Operation & Troubleshooting ....... A-1
Component Testing / Repair ........ B-1
Access Procedures ...................... C-1
061-0752-00 Wiring Diagrams .......................... D-1
Exploded Views / Part Numbers .. E-1

061-0901-00

061-0903-00
H335669
R

MPM60455

Item Description Qty.


1 Operating Instructions Plate ........................ 1
2 Wiring Diagram Label .................................. 1
3 Warning Label .............................................. 1
4 Caution Label ............................................... 1
5 Serial Number Label (large- n/a) ................. 1
6 Caution HOT Label ...................................... 1
7 Serial Number Label (small - n/a) ................ 1

Always Specify Model & Serial Number


MA674302i

Models: M7 (-020 thru -022) Labels & Decals


Serial Numbers: all
© Midmark Corporation 2004 SF-1854
E-17.1
Back Go To Table Of Contents Next
Back Go To Table Of Contents Next

COMMENTS
The Technical Publications Department of Midmark Corporation takes pride in its manuals. We are sure that our manuals will fill all your needs when you are perform-
ing scheduled maintenance, servicing, or repairs on a Midmark product. However, if you find any errors or feel there should be a change, addition, or deletion to the
manuals, please let us know. We will correct any errors that we are made aware of and we will review requests for changes, additions, or deletions to the manuals and
incorporate those requests deemed appropriate. If you see something in one of our manuals that you like or dislike, please let us know. Also, if there is something you
feel we could do to produce a better manual, please let us know.

If an error is found, please list the page and paragraph/figure in which the error was found along with a brief description of what the error is. If the correction to the error
is known, please include that information also. If a change, addition, or deletion is being requested, please list the page and paragraph/figure needing the change, along
with a brief description of how you feel the paragraph/figure should be changed.

Please fax or mail a copy of this completed comment sheet to: Midmark Corporation
ATTN: Technical Publications Dept.
60 Vista Drive
Versailles, Ohio 45380
Fax: (937) 526-5542

Page Number Paragraph/Figure Description


Back Go To Table Of Contents Next

IMPORTANT NOTES:
SERVICE PARTS FAX ORDERING FORM
(Do not tear out this page. Photo copy this page for use only.)
1) Use this form for all non-warranty
am
orders only. Warranty orders must
DATE: / / TIME: pm
be telephoned in (1-800-643-6275).
METHOD OF SHIPMENT:
2) FAX number to send order to:
877-249-1793 PRIORITY: NON-EMERGENCY ORDER {to ship
within 72 hours if part(s) are in stock.}
3) All emergency orders must be DEALER P.O. #:
ATTENTION: CUSTOMER SERVICE DEPARTMENT

received @ Midmark by 1:00 pm EST. EMERGENCY ORDER {to ship within


24 hours if part(s) are in stock.} ACCOUNT #:
[see note 3]
4) All underlined headings should be
filled in prior to submittal.
MODEL #: SERIAL #: SALES ORDER #
(if applicable)

NAME: SHIP TO:

ADDRESS:

CITY: STATE: ZIP:

CONTACT:

PHONE:

FAX #:

LINE # PART NUMBER QTY. DESCRIPTION COLOR


(if applicable)
ADDITIONAL COMMENTS:

CREDIT CARD INFORMATION


CARD TYPE __________________________ _ CARD #_________-_________-_________-_________ EXP. DATE ______ / ______ / ________

NAME ON CARD ___________________________________________ SIGNATURE_______________________________________________________


Back Go To Table Of Contents Next
Back Go To Table Of Contents

Subject to change without notice.


Refer to www.Documark.com for latest revision.

Midmark Corporation
60 Vista Drive
P.O. Box 286
Versailles, OH 45380-0286
Phone: 937-526-3662
Fax: 937-526-5542
www.midmark.com