Professional Documents
Culture Documents
Polaris 600
Polaris 600
Polaris 600
Polaris 600
Contents
Polaris 600
Disposal.......................................................... 81
Disposing of the medical device...................... 81
Disposing of non-rechargeable batteries......... 81
Disposal of accessories................................... 81
Technical data................................................ 82
EMC declaration.............................................. 82
Ambient conditions .......................................... 84
Technical light data ......................................... 84
Electrical data.................................................. 86
Mechanical data .............................................. 87
MedView camera............................................. 87
Polaris 600 remote control for MedView
camera ............................................................ 88
Wireless video transmission ............................ 88
Combination with system components ............ 90
Operating options ............................................ 90
Classifications ................................................. 90
List of accessories ........................................ 92
List of accessories for Polaris light and
MedView camera............................................. 92
Index ............................................................... 93
1 Consecutive numbers indicate steps of action, (A) Letters in parentheses refer to elements in the
with the numbering restarting with "1" for each related illustration.
new sequence of actions.
A Letters in illustrations denote elements
Bullet points indicate individual actions or dif- referred to in the text.
ferent options for action.
Bold, italicized text indicates labels on the
– Dashes indicate the listing of data, options, or device.
objects.
Regional availability
Trademarks
Trademark
Polaris®
WARNING CAUTION
A WARNING statement provides important A CAUTION statement provides important
information about a potentially hazardous information about a potentially hazardous
situation which, if not avoided, could result in situation which, if not avoided, may result in
death or serious injury. minor or moderate injury to the user or patient or
in damage to the medical device or other
property.
NOTE
A NOTE provides additional information intended
to avoid inconvenience during operation.
Target group
The tasks described in this document specify the This target group includes healthcare
requirements that have to be met by each professionals such as surgeons and OR nurses.
respective target group.
Task Requirement
The operating organization of this product must
ensure the following: Use of the product in Training in a relevant
accordance with its healthcare field or in a
– The target group has the required qualifications
intended use medical discipline and
(e.g., has undergone specialist training or
knowledge of the use of the
acquired specialist knowledge through
product
experience).
– The target group has been trained to perform
the task.
Reprocessing personnel
– The target group has read and understood the
chapters required to perform the task. This target group includes persons who are
responsible for the reprocessing of medical
devices.
Description of target groups
Task Requirement
The target groups may only perform the following Reprocessing Specialist knowledge in the
tasks if they meet the corresponding requirements. reprocessing of medical
devices
Safety-related information
The following WARNING and CAUTION Strictly follow the instructions for use of
statements apply to general operation of the other Dräger devices
medical device.
WARNING and CAUTION statements specific to
subsystems or particular features of the medical WARNING
device appear in the respective sections of these
instructions for use or in the instructions for use of Risk of incorrect operation and of incorrect
another product being used with this device. use
For correct use of a Polaris 100 or 200 light in
combination with the Polaris 600 light, read
Strictly follow these instructions for use and observe the Polaris 100/200 instructions
for use.
WARNING
WARNING
Risk of incorrect operation and of incorrect
use Risk of incorrect operation and of incorrect
use
Any use of the medical device requires full
understanding and strict observation of all For correct use of a display mount in
sections of these instructions for use. The combination with a Polaris light, read and
medical device must only be used for the observe the Polaris Multimedia instructions
purpose specified under "Intended use" (see for use.
page 15). Strictly observe all WARNING and
CAUTION statements throughout these
instructions for use and all statements on Strictly follow the assembly instructions
medical device labels.
Failure to observe these safety information
WARNING
statements constitutes a use of the medical
device that is inconsistent with its intended Risk of damage to the device
use.
The light system must only be installed by
assembly personnel. The installation must be
carried out in accordance with the assembly
instructions.
WARNING WARNING
Risk if service is not performed regularly Risk of electric shock and of device
malfunction
If service is not performed regularly,
malfunctions may occur, which can result in Any connected devices or device
personal injury and property damage. combinations not complying with the
requirements in these instructions for use
Perform the service in accordance with the
may compromise the correct functioning of
chapter "Maintenance".
the medical device.
Before using the medical device, refer to and
strictly comply with the instructions for use
Accessories of all connected devices and device
combinations.
WARNING
Risk due to incompatible accessories Safe coupling with electrical equipment
The use of incompatible accessories may
adversely affect the functional integrity of the
product. Personal injury and property CAUTION
damage may occur as a consequence.
Risk of patient injury
Use only compatible accessories. The
Coupling with electrical equipment that is not
accessories that are compatible with this
mentioned in these instructions for use or
product are listed in the list of accessories
assembly instructions may only be done with the
supplied with the product.
respective device manufacturer.
Or:
WARNING
– IEC 60601-1, 2nd edition (general safety
requirements) Risk due to electrostatic discharge
– IEC 60601-1-1 (device combinations) Malfunctions that endanger the patient may
– IEC 60601-1-2 (electromagnetic occur if no protective measures against
compatibility) electrostatic discharge are employed in the
– IEC 60601-1-4 (software-controlled following situations:
functions) – When touching the pins of connectors
that carry the ESD warning symbol.
– When establishing connections with these
Patient safety connectors.
To prevent malfunctions, observe the
The design of the medical device, the
following measures and train the relevant
accompanying documentation, and the labeling on
personnel:
the medical device are based on the assumption
that the purchase and the use of the medical – Observe the ESD protective measures.
device are restricted to persons familiar with the Such measures may include wearing
most important inherent characteristics of the antistatic clothing and shoes, touching a
medical device. potential equalization pin before and while
making the connection, or using
Instructions and WARNING and CAUTION electrically insulating and antistatic
statements are therefore largely limited to the gloves.
specifics of the Dräger medical device.
– Observe the requirements for the
The instructions for use do not contain any electromagnetic environment. Observe
information on the following points: the following section: "Electromagnetic
– Risks that are obvious to users environment" (page 82).
– Consequences of obvious improper use of the
medical device WARNING
– Potentially negative effects on patients with Risk due to electromagnetic disturbance
different underlying diseases
Wireless communication devices (e.g.,
Medical device modification or misuse can be cellular phones) and medical electrical
dangerous. equipment (e.g., defibrillators, electrosurgical
devices) emit electromagnetic radiation.
When such devices are operated too close to
Electromagnetic compatibility (EMC) this device or its cables, the functional
integrity of this device may be compromised
Medical electrical equipment is subject to special by electromagnetic disturbances. As a result,
precautionary measures concerning the patient could be put at risk.
electromagnetic compatibility. During installation – Maintain a distance of at least 0.3 m (1.0 ft)
and before initial operation, follow the information between this device and wireless
in section: "EMC declaration" (page 82). communication devices, to ensure that the
essential performance of this device is
This device can be affected by other electrical fulfilled.
devices. – Maintain an adequate distance between
this device and other medical electrical
equipment.
WARNING
Risk of device malfunction Obligation to report incidents
The light system can cause radio-frequency Serious incidents involving this medical device
interference. must be reported to Dräger and the competent
Shielding of components or reorientation of authorities.
the light system may be necessary.
Installing accessories
CAUTION
Risk of device failure
Install the accessory on the basic device in
accordance with the instructions of the basic
device.
Check for secure connection to the basic device.
CAUTION
Risk of incorrect use
Instructions for use must be kept accessible to
the user.
Training
Functional safety
Do not use non-sterile handles Exercise caution when working with the
system
WARNING
WARNING
Risk of infection
Personnel and patient may become infected Risk of personal injury and property damage
by a non-sterile handle. Objects can fall down into the operating field.
Use only sterile handles. Avoid collisions of the light heads or of parts
of the arm system with other objects.
Direct view of the operating field Observe the maximum load of the
display mount
WARNING
WARNING
Risk of personal injury
A camera is not intended to be used as the Risk of personal injury and property damage
sole visualization medium during an The specification for the maximum load on
operation. the display mount must not be exceeded.
Make sure that the OR personnel have a Comply with the maximum load of the
direct view of the operating field at all times. display mount.
Cautious positioning of the arm system Read and observe the instructions for
use of the displays
CAUTION
Risk of personal injury and property damage WARNING
The stops on the arm system may become Risk of device malfunction
damaged.
To properly use the light system, read and
When positioning the swivel arms and the comply with the instructions for use of the
attached devices, take care not to use force to respective display.
pull the arm system beyond the stops.
OR server
Risk of the light head overheating
WARNING
WARNING
Risk of patient injury
Risk of property damage
The OR servers used must conform to Safety
The light head may overheat during Class B of the IEC 62304 standard.
operation.
Do not cover the light head completely or WARNING
partially.
Risk of patient injury
The OR severs used must conform to the
Compatibility of displays IEC 60601-1:1988 (or later) or IEC 60950-1
standards.
WARNING
Risk due to incompatible displays
The displays used in combination with a
Polaris light system must be approved for
use in the patient environment according to
IEC 60601-1.
Application
Intended use
Environments of use
Description
The Polaris 600 light is certified and approved as a The color temperature can be adjusted in 4 steps,
surgical light. The Polaris 600 light uses LED which the surgeon can select depending on tissue
bulbs. This ensures low energy consumption and a structure, operating field, and individual
long service life of the bulb. The small size of the perception.
LED bulbs makes it possible to use a modern
Cold white light increases concentration, while
design which facilitates ease of handling and
warm white light reduces dazzle effects and
cleaning while also maintaining good flow behavior
increases the contrast.
under laminar flow ceilings.
OR light mode
Overview
26588
A
B
C
F G
E
E M
D
H H
N
I I P
Q
J
Q L
J K O
R S
C Ceiling tube
D Central axis
E Light swivel arm
F Display swivel arm
G Display spring arm
H Light spring arm
I Gimbal mounting
J Light control panel
K Polaris 600 light
L Polaris 600 light with MedView camera and
radio transmitter
M Display mount (24 inch to 32 inch) with radio
receiver
N Display
O Sterilizable handle E
P Sterilizable handle
Q Integrated handle
R Wall-mounted control panel
S Polaris 600 remote control for MedView
camera
Light system with a Polaris 600 and a Polaris 100 light (configuration
example)
26890
A
B
D
D
C
E E
F
F K K
J
G
I
G H H
A Gimbal mounting
26592
B Integrated handle
C Light control panel
D Sterilizable handle E
A
E Lens with LED
B F Bottom glass
F E
D
A Gimbal mounting
26591
B Integrated handle
C Light control panel
A D Sterilizable handle E
E Lens with LED
G B F Bottom glass
G MedView camera
A
F E
D
26793
A
B C D E F G
B C
D
E F
G
H I
26980
/LJKWFRQWURODUHD &DPHUDFRQWURODUHD
$ +
% & ' ( ) , -
/ 0
.
2 3
1
* 4
OR panel (option)
Abbreviations
Abbreviation Explanation
EMC Electromagnetic compatibility
IR Infrared
LED Light-emitting diode
Sterilizable han-
Sterilizable handle, ergonomic
dle E
Symbols
Quantity
Storage temperature
Relative humidity
Ambient pressure
Serial number
Revision index
Operating concept
Sterilizable handle E
NOTE
Correct operation of the sterilizable handle E Operating signal will not be implemented
The touch-sensitive part of the handle is designed
26659
so that, for correct operation, only one finger must
touch the operating section at any one time. If
several fingers touch the operating section
simultaneously, the operating signal will not be
implemented.
Correct operation
26658
26564
Risk of property damage When positioning the swivel arms and the
attached devices, take care not to use force to
The light head may overheat during pull the arm system beyond the stops.
operation.
Do not cover the light head completely or NOTE
partially.
Sterilizable handles are supplied by Dräger in a
non-sterile state.
WARNING
Risk of infection Checking the Polaris 600 remote control for
MedView camera
Personnel and patient may become infected
by a non-sterile handle. Press any key on the remote control.
– Status LED lights up green: the remote control
Use only sterile handles. is functioning.
– Status LED flashes green: the battery charge is
low. The remote control can only continue to be
used for a short time. To maintain the function
of the remote control, the battery must be – Status LED does not light up: the battery is
replaced immediately, see "Replacing the completely discharged or is defective. To
battery in the Polaris 600 remote control for reestablish the function of the remote control,
MedView camera" on page 58. the battery must be replaced immediately, see
"Replacing the battery in the Polaris 600
remote control for MedView camera" on page
58.
26562
Immediately before an operation, fit the handle C
under sterile conditions as follows:
26561
– To remove the handle, see "Disassembly" on Fitting the Dräger disposable handle
page 67. (optional)
28175
not exhibit any cracks, holes, or other damage.
– The inner handle must be free from damage.
– To prevent soiling of the disposable handle, do
not unpack it until immediately before an
operation.
– Do not use the disposable handle if the
packaging is damaged.
– Do not use a knife to open the packaging for
the disposable handle.
WARNING
Risk of infection
WARNING
Risk of patient injury
The sterilizable handle can fall down into the
operating field if it is not engaged properly.
The handle must be checked for secure
attachment after it has been fitted to the inner
handle.
36006
handle under sterile conditions as follows:
C
36005
A
B
Operation
Positioning a light
WARNING CAUTION
Risk of patient injury Risk of personal injury and property damage
The proper functioning of the system must be The stops on the arm system may become
checked each time before use. damaged.
When positioning the swivel arms and the
WARNING attached devices, take care not to use force to
pull the arm system beyond the stops.
Risk of patient injury
The system may only be used in accordance Sterile positioning of a light
with the information in the chapter, "Intended
The light can be rotated by 360° around the
use".
horizontal axis at the central joint.
26915
If the light fields from several lights overlap,
the heat input to the wound site may be
unacceptably high. This can lead to drying
out of the wound site.
If necessary, the total irradiance must be
adjusted by reducing the illuminance of
individual lights, see "Setting the
illuminance" on page 42. A
WARNING
Risk of personal injury and property damage
Do not place objects on the light heads or A
hang them on the arm system. These objects
can fall down into the operating field. 1 Take hold of the handle (A) by the handle
Do not hang heavy loads on the arm system section and use it to position the light.
(e.g., leaning body weight). The mechanism The light head is held at its new position.
of the arm system may be damaged, which
may compromise the correct positioning of
the arm system.
Sterile positioning of a light with a short gimbal 1 Take hold of the integrated handle (A) and use
mounting it to position the light.
The short gimbal mounting has a shorter upper The light head is held at its new position.
bow. The light can be rotated by 330° around the Non-sterile positioning of a light with a short
horizontal axis at the central joint. The rotational gimbal mounting
movement is limited in both directions by stops.
The short gimbal mounting has a shorter upper
The sterile handle is located on the bottom glass of bow. The light can be rotated by 330° around the
the light. horizontal axis at the central joint. The rotational
movement is limited in both directions by stops.
35334
The non-sterile integrated handle is located at the
edge of the light head.
CAUTION
A
Risk of crushing
When positioning the light with short gimbal
mounting, hand injuries may occur.
Keep sufficient space between the integrated
A handle and the upper bow of the short gimbal
mounting and between the integrated handle and
1 Take hold of the handle (A) by the handle the spring arm.
section and use it to position the light.
35335
The light head is held at its new position.
Non-sterile positioning of a light A
The light can be rotated by 360° around the
horizontal axis at the central joint. A
It is recommended that the lights are positioned so – The second light (C) is positioned at the foot of
that the surgeon and the sterile OR personnel can the operating table.
easily reach the sterile handle.
26959
When the lights are positioned in this way, the non-
sterile OR personnel can reach the integrated D
handle in order to make further adjustments if
required.
With central axes with more than one swivel arm, H
make sure that the different swivel arms and their
devices are not all positioned on one side.
Positioning several swivel arms on one side can E
cause tilting or unintended drifting of the swivel
arms.
F
NOTE
Positioning several swivel arms on one side can G
cause tilting or unintended drifting of individual
swivel arms. Take care that the different swivel
arms and their devices are not all positioned on
one side. The figure shows an example of how a
combination of 3 lights and a display can be
positioned to achieve suitable illumination of the
26958
operating field.
– The main surgical light (D) is positioned
laterally above the surgeon's shoulder to
A ensure that the operating field (E) is fully
illuminated.
– The second light (F) is positioned at the foot of
B the operating table.
C
– The third light (G) is positioned at the side of
the operating table.
– The display (H) is positioned at the head of the
operating table.
WARNING
Risk of patient injury
The figure shows an example of how a The patient's unprotected eyes can be
combination of 2 lights can be positioned to damaged by direct exposure to light.
achieve suitable illumination of the operating field.
Position the light so that the light field is not
– The main surgical light (A) is positioned directed towards the unprotected eyes of the
laterally above the surgeon's shoulder to patient.
ensure that the operating field (B) is fully
illuminated.
CAUTION NOTE
Risk of personal injury A distance of 100 cm (39.37 in) from the bottom
glass of the light to the operating field is
The user's eyes can be damaged by direct
recommended to achieve the optimum
exposure to light.
illumination of the operating field.
Do not look directly into the light field.
When using two or more lights to illuminate the
CAUTION operating field, observe the following to ensure
that the total permissible irradiance is not
Risk of property damage exceeded:
The light heads may overheat during operation. – Set the lights to the minimum illuminance
required for appropriate illumination of the
Position the light heads so that their light fields do operating field. If a higher illuminance is only
not illuminate other light heads. required temporarily, reduce the illuminance as
soon as it is no longer needed.
– If the maximum or nearly maximum illuminance
of a light is required, make sure that the
illuminance of the other lights is reduced and
that they do not illuminate the operating field.
26976
A
A
B
The MedView camera (A) is permanently Position the light head so that the MedView
installed in the light head. It is located in a camera has an open field of view (B) at all
segment next to the LED bulbs. times.
The optimum distance between the MedView
camera (from the bottom glass of the light) and
the operating field is 90 cm to 130 cm (35.43 in
to 51.18 in).
Status LEDs
The description of this function applies to both the Status LED for monitoring the power supply
light control panel and the wall-mounted control
The status LED (A) for monitoring the power
panel. Only the light control panel is shown in the
supply for the light system is located to the left of
illustrations.
the key.
Status LED for monitoring a light
26477
The status LED (A) for monitoring the light is
situated directly next to the key.
A
26399
The description of this function applies to both the – Even if the Ambient light mode was the last to
light control panel and the wall-mounted control be active, the light will operate in the OR light
panel. Only the light control panel is shown in the mode after switch-on, with the illuminance (C)
illustrations. that was set the last time the light was used.
Switching the light on Switching the light off
26415
26983
B C B
A A
If there is a light system fault, the status LED The light is in standby mode.
(B) lights up orange, see "Fault – Cause – The status LED (B) is not lit.
Remedy" on page 60. If there is a light system fault, the status LED
– The illuminance (C) when the light is switched (B) lights up orange, see "Fault – Cause –
on corresponds to the last illuminance set Remedy" on page 60.
before the light was switched off.
Synchronization function
26515
the same color temperature and the same
illuminance.
When the Synchronization function is activated
on the wall-mounted control panel, the panel B
sends a request to all Polaris 600 lights to set the
color temperature and the illuminance to specific A
values. The color temperature and the illuminance
are then set on the Polaris 600 lights in
accordance with the request. All Polaris 600 lights
now have the same color temperature and the
same illuminance.
1 Press the key (A).
When the Synchronization function is activated,
the Polaris 600 lights can be individually switched The Synchronization function is activated.
to standby mode. This deactivates the The status LED (B) lights up.
Synchronization function on a light. When a
Polaris 600 light in the light system is in standby Switching the Synchronization function off
mode, the key on this light is deactivated. The
26515
Synchronization function will not be activated
when the key is pressed. When all the
Polaris 600 lights in the light system are in standby
mode, the key on the wall-mounted control B
panel is also deactivated.
When the Synchronization function is activated, A
the Polaris 600 lights can be individually switched
to the Ambient light mode. This deactivates the
Synchronization function on a light. When a
Polaris 600 light in the light system is in the
Ambient light mode, the key on this light is
deactivated. The Synchronization function will 1 Press the key (A).
not be activated when the key is pressed.
When all the Polaris 600 lights in the light system The status LED (B) is not lit.
are in the Ambient light mode, the key on the The Synchronization function is deactivated.
wall-mounted control panel is also deactivated.
Settings made during synchronization are
Switching the Synchronization function on preserved.
The light must be switched on before this function All Polaris 600 lights can be individually set
can be activated, see "Switching the light on and once more.
off" on page 40. Technical background
The transmitted request is what is known as a
"broadcast command", which either the selected
Polaris 600 light or the wall-mounted control panel
transmits to the Polaris 600 lights in the light
system. The broadcast command contains the
values for illuminance and color temperature.
In a network of lights, each of the Polaris 600 lights
receives the same broadcast command. The
Polaris 600 lights receiving the broadcast
command set their respective illuminance and
color temperature to the requested values. With
each change in color temperature or illuminance, a wall-mounted control panel or from the selected
new broadcast command is transmitted to the Polaris 600 light without the receiving Polaris 600
Polaris 600 lights. The Polaris 600 lights receiving lights being able to identify the sender.
the broadcast command cannot use the command
Restriction when a Polaris 600 light is
to trace back whether the command originated
combined with Polaris 100/200 lights
from the selected Polaris 600 light or from the wall-
mounted control panel. In particular, the Polaris The Polaris 100/200 light does not receive any
600 lights cannot use the broadcast command to requests from a Polaris 600 light or from the wall-
trace back which Polaris 600 light transmitted the mounted control panel. Illuminance can always be
command. Thus a request can be sent from the set individually on the Polaris 100/200 light.
26418
panel
41287
The illuminance can be set in 12 steps. These 12
steps are visualized by the 6 LEDs of the LED C
indicator, which light up accordingly with half or full
brightness.
26512
B
A
41285
– At minimum illuminance, the LED (B) of the
LED indicator is lit at half brightness.
26513
NOTE
If several fingers touch the operating section of
the sterilizable handle E simultaneously, the
operating signal will not be implemented. Only
one finger at a time must touch the operating
A section.
Increasing illuminance
NOTE
26606
The operating section of the sterilizable handle E
must be dry to guarantee proper operability.
Reducing illuminance
26607
A
28240
(A) across the surface of the operating section.
The illuminance will be reduced by one step.
B
28240
WARNING
Risk of patient injury
If the light fields from several lights overlap,
the heat input to the wound site may be
unacceptably high. This can lead to drying
out of the wound site.
If necessary, the total irradiance must be
adjusted by reducing the illuminance of
individual lights, see "Setting the
illuminance" on page 42.
Light combina- Maximum Recommended adjustment of the total irradi- Thermal radiation in
tion total irradi- ance2) by reducing the illuminance the wound site after
ance1) adjustment of the
total irradiance
[W/m2]
(A)
(A) (B)
(A) (B)
(A) (B)
(A) (B)
Light combina- Maximum Recommended adjustment of the total irradi- Thermal radiation in
tion total irradi- ance2) by reducing the illuminance the wound site after
ance1) adjustment of the
total irradiance
[W/m2]
The description of this function applies to both the 2 Press the key (A) again.
light control panel and the wall-mounted control The color temperature is increased by a further
panel. Only the light control panel is shown in the step.
illustrations.
3 The change in color temperature is shown by
The light must be switched on before this function the LED indicator (B).
can be activated, see "Switching the light on and
off" on page 40. 4 If the highest color temperature has been
reached, pressing the key (A) again will
There is an indicator with 3 LEDs for the color reset the color temperature to its lowest value.
temperature. These LEDs indicate the color
– For further information on color temperature,
temperature set for the light.
see "Technical light data" on page 84.
LED indicator Color temperature
3800 K
4400 K
5000 K
5600 K
Setting the light field diameter on the 2 Press the key (A) again.
light control panel and on the wall- The light field diameter is increased by a
mounted control panel further step.
3 The change in the light field diameter is shown
The description of this function applies to both the by the LED indicator (B).
light control panel and the wall-mounted control If the largest light field diameter has been
panel. Only the light control panel is shown in the reached, pressing the key (A) again will
illustrations. reset the light field diameter to its smallest size.
The light must be switched on before this function – For further information on the light field
can be activated, see "Switching the light on and diameter, see "Technical light data" on page
off" on page 40. 84.
LED indicator Light field diameter
Small Setting the light field diameter on the
sterilizable handle E
Large NOTE
If several fingers touch the operating section of
the sterilizable handle E simultaneously, the
– The light field diameter when the light is operating signal will not be implemented. Only
switched on is the same as the light field one finger at a time must touch the operating
diameter set before the light was switched off. section.
NOTE
26608
26614
diameter, see "Technical light data" on page
84.
When the Synchronization function is activated, Switching the ambient light mode off
the Polaris 600 lights can be individually switched
to the Ambient light mode. This deactivates the
26681
Synchronization function on a light. When a
Polaris 600 light in the light system is in the
Ambient light mode, the key on this light is B
deactivated. The Synchronization function will
not be activated when the key is pressed.
When all the Polaris 600 lights in the light system
are in the Ambient light mode, the key on the A
wall-mounted control panel is also deactivated.
Switching the ambient light mode on
The description of this function applies to both the
light control panel and the wall-mounted control Press the key (A).
panel. Only the light control panel is shown in the
illustrations. The Ambient light mode is deactivated.
The light must be switched on before this function The light switches to the OR light mode.
can be activated, see "Switching the light on and The status LED (B) is not lit.
off" on page 40. – For further information on the ambient light
mode (Endo light), see "Technical light data" on
26681
page 84.
For a Polaris 600 light with MedView camera, the Switching the MedView camera off
camera control area is on the right-hand side of the
shared wall-mounted control panel.
26358
A
Switching the MedView camera on B
28134
A
B
C
C
D
26361
26445
C
A B
1 Press the key (A). When the Manual exposure function is activated,
The image detail is reduced. it is possible to switch back to the Automatic
exposure function.
If the key is kept pressed, the image is reduced
continuously. Press the key (A).
The image detail is magnified. The status LED (B) lights up.
If the key is kept pressed, the image is When the Automatic exposure function is
magnified continuously. activated, it is possible to switch to the Manual
exposure function.
The camera enables up to 120x magnification
(10x optical / 12x digital). Press the key (A).
3 The change in the image detail is shown by the The Manual exposure function is activated.
LED indicator (C). The status LED (B) is not lit.
The larger the image detail becomes, the more
LEDs light up.
The smaller the image detail becomes, the
fewer LEDs light up.
26389
default.
26362
B A
A B
– To achieve optimum image illumination, the
zoom function can be used to match the size of
the image detail (A) to the light field diameter
(B).
– The camera can only record optimally when the
illumination is uniform.
26377
26446
B A B
26380
B A
NOTE
The MedView camera adjusts itself automatically
to the color temperature selected for the Polaris
600 light.
26392
keys in the camera control area on the right-hand B
side of the wall-mounted control panel.
26391
A
Replacing the battery in the Polaris 600 remote control for MedView
camera
27992
type 6LR61, 6F22, or 1604D.
The operating time of the remote control depends
on the capacity of the battery used and is
approximately 1 year for the factory-supplied
battery. A
WARNING
Risk of damage to the device
Batteries must be stored in accordance with
the manufacturer's specifications. Batteries
must not be disassembled, opened or 1 Using light pressure, raise the housing flap (A)
crushed. Batteries must not be short- at the hinge at the bottom while at the same
circuited. Batteries must not be exposed to time folding it out at the top.
either heat or fire. Batteries must not be
27993
stored in direct sunlight.
CAUTION
Risk of damage to the device B
Only 6LR61, 6F22, or 1604D battery types may
be used, as otherwise the operation of the
remote control may be compromised.
CAUTION
Risk of damage to the device 2 Unlock the cover (B) with a ballpoint pen or
screwdriver and fold it out.
Non-rechargeable batteries must not be charged.
27994
CAUTION
Risk of damage to the device C
The battery must be removed if the remote D
control will not be used for a certain period of
time. If there is fluid escaping from the battery,
the fluid must not come into contact with the skin
or get into the eyes. If there has been contact,
rinse the affected area with plenty of water and
seek medical help immediately.
27995
Troubleshooting
Supply mains
20
Emergency power
21
Output voltage
22
Initialize in progress
23 24
Contact service
25
30
Initialize system
90 1
78
System number A
31
2 34
56
1
EF0 2
CD
345
32
System number B
AB
67
89
B A D
C
Reprocessing
Safety information
WARNING
Risk of patient injury and of device
malfunction
The device and accessories must be cleaned
and disinfected using a validated procedure.
WARNING
Risk of patient injury and of device
malfunction
Always observe the hygiene regulations of
the hospital when handling devices which are
contaminated with body fluids.
NOTE
To avoid the risk of infections for hospital staff
and patients, always clean and disinfect the
device after each use.
NOTE
Do not use abrasive cleaners for cleaning.
NOTE
Strictly follow the disinfectant manufacturer's
instructions for use!
Disassembly
26647
The handle is located on the bottom glass of the
light and is removed from the inner handle by
pulling.
26638
A
B
3 Pull the handle (D) down to remove it.
Removing the sterilizable handle
The handle is located on the bottom glass of the
light and is removed from the inner handle by
1 Take hold of the handle section and press the pulling.
locking mechanism (A).
28274
2 Keep the locking mechanism pressed and turn
the handle counterclockwise (B) until the
markers (C) line up.
A
B
36026
end of the handle and pull off the handle.
Removing the Dräger disposable handle
D
The disposable handle is located on the bottom
glass of the light and is removed from the inner C
handle by pulling.
36025
A
A
3 Hold the disposable handle by the recessed
grips (C) and turn it counterclockwise (D).
36027
1 Take hold of the disposable handle (A).
2 Pull the locking mechanism (B) on the
disposable handle until the entire perforation is
ripped open.
The locking mechanism is intended to ensure
single use of the disposable handle and is
unusable after it has been pulled.
Components:
NOTE
– Sterilizable handle E
– Sterilizable handle Use and frequent sterilization cause wear to the
handle. In the event of material fatigue such as
There is an identification mark on the upper side of cracking and/or discoloration, the handle must be
the handle which shows the year and month of replaced.
manufacture of the handle.
28426
12 1 2
9 1 11
18 5
34 5
0
67
The identification mark here shows January 2015
as an example of the date of manufacture.
The number in the inner circle corresponds to the
year of manufacture.
15 = 2015
The arrow points to the month of manufacture.
1 = January
The date of manufacture has no effect on the
service life of the handle.
The handle can undergo up to 300 reprocessing
cycles. After that, it must be disposed of.
NOTE
Cleaning methods based on alkaline agents (pH
values >10) can increase material wear to the
handle and thus shorten its service life.
Manual cleaning: 5 Check all items for damage and external signs
of wear, such as cracking, embrittlement, or
– Neodisher® LM 2 by Dr. Weigert pronounced hardening, and residual dirt.
Concentration: 2 % Replace parts if necessary.
Contact time: 10 minutes
CAUTION
Manual disinfection:
Risk of damage to the device
– Korsolex® extra by BODE Chemie
The device consists in part of materials that are
Concentration: 3 %
not resistant to certain components of surface
Contact time: 15 minutes disinfectants.
Machine reprocessing It is essential to follow the instructions in this
Machine cleaning: chapter.
Cleaning and disinfection of the bottom glass 3 Check items for visible soiling and damage. If
– For every new patient necessary, repeat manual disinfection.
4 Thoroughly shake out residual water. Allow
items to dry thoroughly.
27176
Machine cleaning and disinfection
Perform machine cleaning and disinfection using a
washer-disinfector in accordance with EN
ISO 15883, preferably with a cart for anesthesia
accessories and ventilation accessories.
Procedure:
1 Observe instructions for use for the washer-
disinfector.
2 Securely position items in the basket. Make
sure that all interior spaces and surfaces are
completely flushed and that water can drain off
freely.
3 Use suitable cleaning agent.
1 Wipe the bottom glass with a soft, lint-free cloth 4 Select suitable program, preferably anesthesia
radially from the inside outwards. program.
Critical medical devices – Thermal disinfection must be carried out at
93 °C (199 °F) for 10 minutes.
Manual cleaning
5 Carry out final rinsing with deionized water.
Perform manual cleaning preferably under running
6 Immediately remove items from the washer-
water and with commercially available cleaning
disinfector.
agents (pH value ≥10).
7 Check items for visible soiling and damage. If
Procedure: necessary, repeat program or perform manual
1 Wash off superficial soiling under running cleaning and disinfection.
water. 8 Allow items to dry thoroughly.
2 Use cleaning agents in accordance with the
Visual inspection
manufacturer’s instructions. Make sure that all
surfaces and interior spaces which must be Check all items for damage and external signs of
cleaned are reached. If necessary, use suitable wear, such as cracking, embrittlement, or
brushes. pronounced hardening, and residual dirt.
3 Rinse items thoroughly under running water
until cleaning agent residues are no longer CAUTION
discernible.
Risk from faulty accessories
4 Check items for visible soiling and damage. If
necessary, repeat manual cleaning. Even reusable accessories have a limited
maximum period of use, e.g., residues from
Manual disinfection disinfectants can attack the material in the
autoclave. Signs of external wear can show up,
Procedure:
e.g., cracks, deformation, discoloration, or
1 Immerse items in disinfectant. delamination).
2 After the contact time has elapsed, rinse items If signs of external wear occur, replace the
thoroughly under running water until affected accessory.
disinfectant residues are no longer discernible.
Sterilization
WARNING
For sterilization, use a vacuum steam sterilizer
(conforming to ISO 17665), preferably with Risk of infection
fractionated vacuum. If the sterilizable handle exhibits cracks or
Only sterilize cleaned and disinfected items. deformation as a result of frequent
sterilization, the handle may fall into the
– Sterilize with a fractionated vacuum steam
operation field and/or may not satisfy the
procedure at 132 °C (270 °F) for 4 minutes.
infection prevention policy.
– The sterilizable handle must not be subjected
to mechanical load during sterilization. Make sure that the handle is not used in the
Otherwise, it may be permanently deformed. event of such damage.
– The sterilizable handle must not be reattached
until immediately before an operation.
Maintenance
Overview
Term Definition
Maintenance All measures (inspection, preventive maintenance, repair) intended to maintain
and restore the functional condition of a medical device
Inspection Measures intended to determine and assess the actual state of a medical device
Preventive mainte- Recurrent specified measures intended to maintain the functional condition of a
nance medical device
Repair Measures intended to restore the functional condition of a medical device after a
device malfunction
Inspection
WARNING
Risk of electric shock
Disconnect all electrical connections from the
power supplies before carrying out any work.
Preventive maintenance
WARNING WARNING
Risk of faulty components Risk of electric shock
Device failure is possible due to wear or Disconnect all electrical connections from the
material fatigue of the components. power supplies before carrying out any work.
To maintain the function of all components,
this device must be inspected and serviced at
the intervals specified by the manufacturer.
Repair
The spring arm must hold the mounted light in any Adjusting the spring force
height position.
The spring force of the spring arm must be
Adjusting the height stop adjusted if the light does not remain at the set
height.
Vertical movement of the light is limited by a fixed
lower and adjustable upper stop. The adjustable When adjusting the spring arm, lift it approx. 10o
upper stop (height stop) can be set to any position above the horizontal position. Only in this position
between the positions (A) and (B). The factory is it possible to turn the set screw.
setting for the spring arm is position (B).
If necessary, set the upper stop (height stop) to a
higher position.
27202
27203
A
B C O
10
– If the adjustment has been carried out Turn the set screw clockwise to increase the
correctly, the light can be positioned at any braking force.
height and will remain there. Turn the set screw counterclockwise to
decrease the braking force.
The gimbal mounting is set correctly when the light Set the braking force as low as possible and as
head is stable in any position and does not move high as necessary.
on its own.
Adjusting the braking force at the intermediate
joint
27204
27205
A
A
B
NOTE
A
Set the braking force as low as possible and as
high as necessary.
27208
PP
0LQLPXPOHQJWK
0LQLPXPPDWHULDOWKLFNQHVV
PP
PP
0LQLPXPOHQJWK
0LQLPXPPDWHULDOWKLFNQHVV
PP
Disposal
WARNING
For countries subject to the EU
Directive 2002/96/EC
Risk of infection
The device and its components must be This device is subject to EU Directive 2002/96/EC
disinfected and cleaned before disposal! (WEEE). In order to comply with its registration
according to this directive, this device may not be
This product must be disposed of in accordance disposed of at municipal collection points for waste
with national regulations. electrical and electronic equipment. Dräger has
authorized a company to collect and dispose of
this device. To initiate collection or for further
information, visit Dräger on the Internet at
www.draeger.com. Use the search function with
the keyword "WEEE" to find the relevant
information. If access to the Dräger website is not
possible, contact the local Dräger organization.
Disposal of accessories
Technical data
EMC declaration
Electromagnetic environment
Emissions Compliance
Radiated emis- Class A, Group 1 (30 MHz to
sions 1 GHz)
Conducted emis- Class A, Group 1 (150 kHz to
sions 30 MHz)
CAUTION
The emissions characteristics of this equipment
make it suitable for use in industrial areas and
hospitals (CISPR 11, Class A). If it is used in a
residential environment (for which CISPR 11
Class B is normally required), this equipment
might not offer adequate protection to radio-
frequency communication services.
The user might need to take mitigation
measures, such as relocating or reorienting the
equipment.
Ambient conditions
During operation
Temperature For the light system:
10 °C to 40 °C
(50 °F to 104 °F)
For the radio receiver in the display mount:
10 °C to 30 °C
(50 °F to 86 °F)
Ambient pressure 700 hPa to 1060 hPa
(10.2 psi to 15.4 psi)
Relative humidity 30 % to 75 % (non-condensing)
1)
During storage and transport (up to 15 weeks)
Temperature –20 °C to 60 °C
(–4 °F to 140 °F)
Ambient pressure 700 hPa to 1060 hPa
(10.2 psi to 15.4 psi)
Relative humidity 10 % to 95 % (non-condensing)
1) Storage only indoors or in covered spaces
Depth of illumination L1+L2 (60 %) 700 mm (27.56 in) 700 mm (27.56 in)
Color rendering index Ra (at 3800 K) 95 95
Color rendering index R9 (at 3800 K) 94 94
Color rendering index R13 (at 3800 K) 93 93
Recommended working area (distance from 90 cm to 120 cm 90 cm to 120 cm
the bottom glass of the light to the operating (35.43 in to 47.24 in) (35.43 in to 47.24 in)
field)
Central irradiance Ee2) 600 W/m2 490 W/m2
Ee/Ec ratio 3.6 (mW/m2) / lx 3.8 (mW/m2) / lx
Setting the illuminance 12 steps, 20000 lx to 12 steps, 20000 lx to
160000 lx, 12.5 % to 120000 lx, 16.7 % to
100 % (2 % Endo light) 100 % (2 % Endo light)
Minimum illuminance 20000 lx 20000 lx
Setting the diameter of the light field 3 steps, 190 mm / 3 steps, 240 mm /
230 mm / 280 mm 280 mm / 320 mm
(7.48 in / 9.06 in / (9.45 in / 11.02 in /
11.02 in) 12.60 in)
Setting the color temperature 4 steps, 3800 K / 4 steps, 3800 K /
4400 K / 5000 K / 4400 K / 5000 K /
5600 K 5600 K
Bulbs LEDs LEDs
Number of LEDs 92 92
Number of LED modules 18 18
Service life of the LED bulbs 50000 hours 50000 hours
Replacement of LED bulbs possible possible
Electrical data
Mechanical data
MedView camera
Radio receiver
Operating frequency (European Union) U-NII band @ 5 GHz (5.180 GHz to 5.320 GHz
and 5.500 GHz to 5.825 GHz)
Transmitter power & modulation TPC (transmit-
ter power control) limited to 18 dBm EIRP /
16 dBm ERP max.
Transmitter modulation OFDM/QAM
Video outputs on each radio receiver 2x HD-SDI (with HDMI/HD-SDI converter)
1x DVI-D (with HDMI to DVI-D cable)
Maximum number of video receivers per Med- 4
View camera
Power supply 12 V, with associated 100 V to 240 V (AC) /
50 Hz/60 Hz power supply unit
Mains plug type Country-specific type supplied with device
Dimensions L x W x H 146 mm x 117 mm x 28 mm
(5.75 in x 4.60 in x 1.10 in) (without power plug)
146 mm x 130 mm x 83 mm
(5.75 in x 5.12 in x 3.27 in) (with power plug)
160 mm x 140 mm x 83 mm
(6.30 in x 5.51 in x 3.27 in) (with radio receiver
housing (wall or ceiling))
Weight including power supply unit 250 g (0.55 lb)
Surface mounting possible
Mounting in the display mount possible
Ceiling mounting possible
Radio transmitter
Operating frequency (European Union) U-NII band @ 5 GHz (5.180 GHz to 5.320 GHz
and 5.500 GHz to 5.825 GHz)
Bandwidth for each modulation/frequency
40 MHz
Operating options
Sterilizable handle E (to set illuminance and light field diameter) Sterile
Sterilizable handle Sterile
Dräger disposable handle Sterile
Disposable sleeve for handle adapter (e.g., manufactured by Covidien) Sterile
Integrated handle (at the light head) Non-sterile
Light control panel (on the gimbal mounting) Non-sterile
Wall-mounted control panel Non-sterile
Polaris 600 remote control for MedView camera Non-sterile
Classifications
Classifications (continued)
List of accessories
WARNING
Risk of patient injury
Device malfunctions may occur if
incompatible accessories are used with the
light system.
Only accessories from this list of accessories
may be used on the light system. Follow
these instructions for use when fitting the
accessories.
Index
A I
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 Illuminance . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Image illumination (MedView camera) . . . . . . . 54
Ambient light. . . . . . . . . . . . . . . . . . . . . . . . . 16, 51 Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Autofocus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55 Integrated handle . . . . . . . . . . . . . . . . . 18, 20, 20
Automatic exposure . . . . . . . . . . . . . . . . . . . . . . 53 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
B L
Battery replacement (Polaris 600 remote control) LED indicator. . . . . . . . . . . . . . . . . . . . . . . . . . . 43
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Light control area. . . . . . . . . . . . . . . . . . . . . . . . 23
Braking force . . . . . . . . . . . . . . . . . . . . . . . . . . . 78 Light control panel . . . . . . . . . . . . . . . . . 20, 20, 21
Light field diameter . . . . . . . . . . . . . . . . . . . . . . 49
C Light mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
List of accessories. . . . . . . . . . . . . . . . . . . . . . . 92
Camera control area. . . . . . . . . . . . . . . . . . . . . . 23
Color temperature. . . . . . . . . . . . . . . . . . . . . 16, 48 M
Counterweight . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Mains power . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
D Manual exposure. . . . . . . . . . . . . . . . . . . . . . . . 54
Manual focus . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Device combinations . . . . . . . . . . . . . . . . . . . . . . 9 MedView camera. . . . . . . . . . . . . . . . . . . . . 20, 22
Disposable handle . . . . . . . . . . . . . . . . . . . . . . . 33
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81 O
Drifting of swivel arms . . . . . . . . . . . . . . . . . . . . 37
Operating time (Polaris 600 remote control) . . . 58
E OR light . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16, 40
OR panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Electromagnetic compatibility . . . . . . . . . . . . . . . 10
EMC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 P
EMC declaration . . . . . . . . . . . . . . . . . . . . . . . . . 82
Emergency power. . . . . . . . . . . . . . . . . . . . . . . . 39 Patient safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Endo light . . . . . . . . . . . . . . . . . . . . . . . . . . . 16, 51 Polaris 600 control interface . . . . . . . . . . . . . . . 24
Polaris 600 remote control . . . . . . . . . . . . . . . . 22
F Positioning a light, non-sterile . . . . . . . . . . . . . . 36
Positioning a light, sterile. . . . . . . . . . . . . . . . . . 35
Fault – Cause – Remedy . . . . . . . . . . . . . . . . . . 60 Positioning of a light with short gimbal mounting,
Freeze frame . . . . . . . . . . . . . . . . . . . . . . . . . . . 56 non-sterile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Positioning of a light with short gimbal mounting,
G sterile. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Preventive maintenance . . . . . . . . . . . . . . . . . . 76
Gimbal mounting . . . . . . . . . . . . . . . . . . . . . . . . 78 Preventive maintenance intervals . . . . . . . . . . . 76
H R
Handle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Height stop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77 Reprocessing personnel . . . . . . . . . . . . . . . . . . . 6
S
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Service personnel. . . . . . . . . . . . . . . . . . . . . . . . . 7
Specialized . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Spring arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Standby mode . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Status LED (light) . . . . . . . . . . . . . . . . . . . . . . . . 39
Status LED (light system). . . . . . . . . . . . . . . . . . 39
Sterilizable handle E. . . . . . . . . . . . . . . . . . . . . . 28
Switching the light off . . . . . . . . . . . . . . . . . . . . . 40
Switching the light on . . . . . . . . . . . . . . . . . . . . . 40
Switching the MedView camera off . . . . . . . . . . 52
Switching the MedView camera on . . . . . . . . . . 52
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
T
Target group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Total irradiance, maximum . . . . . . . . . . . . . . . . . 45
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . 60
U
Users . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
W
Wall-mounted control panel . . . . . . . . . . . . . 18, 23
White balance (automatic) . . . . . . . . . . . . . . . . . 56
Z
Zoom . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
MD MD
(Polaris 600 light system) (Dräger Polaris disposable handle)
(Radio transmitter)
Manufacturer
9054591 – GA 6932.155 en
Á9054591È
© Drägerwerk AG & Co. KGaA
Edition: 7 – 2019-10
(Edition: 1 – 2014-08)
Dräger reserves the right to make modifications
to the device without prior notice.
ision]_0000047853_No.0411