Polaris 600

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 96

Instructions for use

Polaris 600

WARNING Surgical light


To properly use this medical device, Polaris 600

read and comply with these


instructions for use.
This page has been left blank intentionally.

2 Instructions for use Polaris 600


Contents

Contents
Polaris 600

Information regarding the instructions for Assembly and preparation............................ 31


use .................................................................. 5 Checking for operational readiness before
Regional availability......................................... 5 each use.......................................................... 31
Trademarks ..................................................... 5 Fitting the handle............................................. 32
Safety information definitions .......................... 6
Operation ........................................................ 35
Target group .................................................... 6
Abbreviations and symbols.............................. 7 Positioning a light ............................................ 35
Positioning a light system................................ 37
Safety-related information ............................ 8 Positioning the MedView camera .................... 38
General safety information .............................. 8 Status LEDs .................................................... 39
Product-specific safety information.................. 12 Switching the light on and off .......................... 40
Synchronization function ................................. 40
Application ..................................................... 15
Setting the illuminance .................................... 42
Intended use.................................................... 15 Color temperature function.............................. 48
Environments of use........................................ 16 Light field diameter function ............................ 49
Description....................................................... 16 Ambient light mode (Endo light) ...................... 51
Overview......................................................... 17 Operating the MedView camera using the wall-
mounted control panel..................................... 52
Multimedia system with 2 Polaris 600 lights Operating the MedView camera by means of
(configuration example) ................................... 17 the Polaris 600 remote control for MedView
Light system with a Polaris 600 and a camera ............................................................ 57
Polaris 100 light (configuration example) ........ 19 Replacing the battery in the Polaris 600 remote
Polaris 600 light with sterilizable handle E ...... 20 control for MedView camera............................ 58
Polaris 600 light with sterilizable handle E and
MedView camera............................................. 20 Troubleshooting............................................. 60
Light control panel ........................................... 21 Fault – Cause – Remedy................................. 60
MedView camera (option)................................ 22
Wall-mounted control panel for 2 Polaris 600 Reprocessing ................................................. 66
lights and a MedView camera (configuration Safety information ........................................... 66
example).......................................................... 23 Disassembly .................................................... 67
OR panel (option) ............................................ 24 Service life of the handle ................................. 69
Polaris 600 control interface (option)............... 24 Overview of reprocessing procedures............. 70
Abbreviations................................................... 24 Materials used in the light................................ 72
Symbols........................................................... 25 Assembling the parts....................................... 72
Operating concept ......................................... 26 Maintenance ................................................... 73
Operating controls and their functions on the Overview ......................................................... 73
Polaris 600 light ............................................... 26 Inspection ........................................................ 73
Operating controls and their functions on the Preventive maintenance.................................. 76
Polaris 600 and Polaris 100/200 light system.. 27 Repair.............................................................. 76
Sterilizable handle E........................................ 28 Adjusting the arm system ................................ 77
Wall-mounted control panel (option)................ 29 Locking segment of the spring arm ................. 79
Color assignment of the lights to the wall- Acceptance and handover............................... 80
mounted control panel ..................................... 30

Instructions for use Polaris 600 3


Contents

Disposal.......................................................... 81
Disposing of the medical device...................... 81
Disposing of non-rechargeable batteries......... 81
Disposal of accessories................................... 81
Technical data................................................ 82
EMC declaration.............................................. 82
Ambient conditions .......................................... 84
Technical light data ......................................... 84
Electrical data.................................................. 86
Mechanical data .............................................. 87
MedView camera............................................. 87
Polaris 600 remote control for MedView
camera ............................................................ 88
Wireless video transmission ............................ 88
Combination with system components ............ 90
Operating options ............................................ 90
Classifications ................................................. 90
List of accessories ........................................ 92
List of accessories for Polaris light and
MedView camera............................................. 92
Index ............................................................... 93

4 Instructions for use Polaris 600


Information regarding the instructions for use

Information regarding the instructions for use

1 Consecutive numbers indicate steps of action, (A) Letters in parentheses refer to elements in the
with the numbering restarting with "1" for each related illustration.
new sequence of actions.
A Letters in illustrations denote elements
 Bullet points indicate individual actions or dif- referred to in the text.
ferent options for action.
Bold, italicized text indicates labels on the
– Dashes indicate the listing of data, options, or device.
objects.

Regional availability

Some components may not be available in every


country. Please contact your local contact person
for further information.
The following web page lists the local contact
persons: www.draeger.com

Trademarks

Trademarks owned by Dräger

Trademark
Polaris®

The following web page provides a list of the


countries in which the trademarks are registered:
www.draeger.com/trademarks

Instructions for use Polaris 600 5


Information regarding the instructions for use

Safety information definitions

WARNING CAUTION
A WARNING statement provides important A CAUTION statement provides important
information about a potentially hazardous information about a potentially hazardous
situation which, if not avoided, could result in situation which, if not avoided, may result in
death or serious injury. minor or moderate injury to the user or patient or
in damage to the medical device or other
property.

NOTE
A NOTE provides additional information intended
to avoid inconvenience during operation.

Target group

Duties of the operating organization Users

The tasks described in this document specify the This target group includes healthcare
requirements that have to be met by each professionals such as surgeons and OR nurses.
respective target group.
Task Requirement
The operating organization of this product must
ensure the following: Use of the product in Training in a relevant
accordance with its healthcare field or in a
– The target group has the required qualifications
intended use medical discipline and
(e.g., has undergone specialist training or
knowledge of the use of the
acquired specialist knowledge through
product
experience).
– The target group has been trained to perform
the task.
Reprocessing personnel
– The target group has read and understood the
chapters required to perform the task. This target group includes persons who are
responsible for the reprocessing of medical
devices.
Description of target groups
Task Requirement
The target groups may only perform the following Reprocessing Specialist knowledge in the
tasks if they meet the corresponding requirements. reprocessing of medical
devices

6 Instructions for use Polaris 600


Information regarding the instructions for use

Service personnel Specialized service personnel

Task Requirement Task Requirement


Installation Specialist knowledge in Installation Specialist knowledge in
electrical engineering and electrical engineering and
Basic service activi- Basic and complex
mechanics mechanics
ties (inspection, service activities
Experience in the servicing Experience in the servicing
maintenance accord- (inspection, mainte-
of medical devices of medical devices
ing to the "Mainte- nance, repair)
Training in service activities
nance" section)
on this product

Dräger recommends arranging a service contract


with DrägerService.

Abbreviations and symbols

Explanations can be found in the sections


"Abbreviations" and "Symbols" in chapter
"Overview".

Instructions for use Polaris 600 7


Safety-related information

Safety-related information

General safety information

The following WARNING and CAUTION Strictly follow the instructions for use of
statements apply to general operation of the other Dräger devices
medical device.
WARNING and CAUTION statements specific to
subsystems or particular features of the medical WARNING
device appear in the respective sections of these
instructions for use or in the instructions for use of Risk of incorrect operation and of incorrect
another product being used with this device. use
For correct use of a Polaris 100 or 200 light in
combination with the Polaris 600 light, read
Strictly follow these instructions for use and observe the Polaris 100/200 instructions
for use.
WARNING
WARNING
Risk of incorrect operation and of incorrect
use Risk of incorrect operation and of incorrect
use
Any use of the medical device requires full
understanding and strict observation of all For correct use of a display mount in
sections of these instructions for use. The combination with a Polaris light, read and
medical device must only be used for the observe the Polaris Multimedia instructions
purpose specified under "Intended use" (see for use.
page 15). Strictly observe all WARNING and
CAUTION statements throughout these
instructions for use and all statements on Strictly follow the assembly instructions
medical device labels.
Failure to observe these safety information
WARNING
statements constitutes a use of the medical
device that is inconsistent with its intended Risk of damage to the device
use.
The light system must only be installed by
assembly personnel. The installation must be
carried out in accordance with the assembly
instructions.

8 Instructions for use Polaris 600


Safety-related information

Service Connected devices

WARNING WARNING
Risk if service is not performed regularly Risk of electric shock and of device
malfunction
If service is not performed regularly,
malfunctions may occur, which can result in Any connected devices or device
personal injury and property damage. combinations not complying with the
requirements in these instructions for use
Perform the service in accordance with the
may compromise the correct functioning of
chapter "Maintenance".
the medical device.
Before using the medical device, refer to and
strictly comply with the instructions for use
Accessories of all connected devices and device
combinations.
WARNING
Risk due to incompatible accessories Safe coupling with electrical equipment
The use of incompatible accessories may
adversely affect the functional integrity of the
product. Personal injury and property CAUTION
damage may occur as a consequence.
Risk of patient injury
Use only compatible accessories. The
Coupling with electrical equipment that is not
accessories that are compatible with this
mentioned in these instructions for use or
product are listed in the list of accessories
assembly instructions may only be done with the
supplied with the product.
respective device manufacturer.

Not for use in areas of explosion hazard Device combinations

WARNING This device may be operated in combination with


other Dräger devices or with devices from third-
Risk of fire party manufacturers. Observe the accompanying
The medical device is not approved for use in documents of the individual devices.
areas where combustible or explosive gas If a device combination is not approved by Dräger,
mixtures are likely to occur. the safety and functional integrity of the individual
devices may be compromised. The operator must
ensure that the device combination conforms to
the applicable editions of the relevant standards
for medical devices.
Device combinations that are approved by Dräger
meet the requirements of the following standards:
– IEC 60601-1, 3rd edition (general safety
requirements, device combinations, software-
controlled functions)
– IEC 60601-1-2 (electromagnetic
compatibility)

Instructions for use Polaris 600 9


Safety-related information

Or:
WARNING
– IEC 60601-1, 2nd edition (general safety
requirements) Risk due to electrostatic discharge
– IEC 60601-1-1 (device combinations) Malfunctions that endanger the patient may
– IEC 60601-1-2 (electromagnetic occur if no protective measures against
compatibility) electrostatic discharge are employed in the
– IEC 60601-1-4 (software-controlled following situations:
functions) – When touching the pins of connectors
that carry the ESD warning symbol.
– When establishing connections with these
Patient safety connectors.
To prevent malfunctions, observe the
The design of the medical device, the
following measures and train the relevant
accompanying documentation, and the labeling on
personnel:
the medical device are based on the assumption
that the purchase and the use of the medical – Observe the ESD protective measures.
device are restricted to persons familiar with the Such measures may include wearing
most important inherent characteristics of the antistatic clothing and shoes, touching a
medical device. potential equalization pin before and while
making the connection, or using
Instructions and WARNING and CAUTION electrically insulating and antistatic
statements are therefore largely limited to the gloves.
specifics of the Dräger medical device.
– Observe the requirements for the
The instructions for use do not contain any electromagnetic environment. Observe
information on the following points: the following section: "Electromagnetic
– Risks that are obvious to users environment" (page 82).
– Consequences of obvious improper use of the
medical device WARNING
– Potentially negative effects on patients with Risk due to electromagnetic disturbance
different underlying diseases
Wireless communication devices (e.g.,
Medical device modification or misuse can be cellular phones) and medical electrical
dangerous. equipment (e.g., defibrillators, electrosurgical
devices) emit electromagnetic radiation.
When such devices are operated too close to
Electromagnetic compatibility (EMC) this device or its cables, the functional
integrity of this device may be compromised
Medical electrical equipment is subject to special by electromagnetic disturbances. As a result,
precautionary measures concerning the patient could be put at risk.
electromagnetic compatibility. During installation – Maintain a distance of at least 0.3 m (1.0 ft)
and before initial operation, follow the information between this device and wireless
in section: "EMC declaration" (page 82). communication devices, to ensure that the
essential performance of this device is
This device can be affected by other electrical fulfilled.
devices. – Maintain an adequate distance between
this device and other medical electrical
equipment.

10 Instructions for use Polaris 600


Safety-related information

Radio-frequency interference – reduction of thermal radiation into the operating


field.

WARNING
Risk of device malfunction Obligation to report incidents
The light system can cause radio-frequency Serious incidents involving this medical device
interference. must be reported to Dräger and the competent
Shielding of components or reorientation of authorities.
the light system may be necessary.

Installing accessories

CAUTION
Risk of device failure
Install the accessory on the basic device in
accordance with the instructions of the basic
device.
Check for secure connection to the basic device.

Strictly observe the instructions for use and


assembly instructions.

Storing the instructions for use

CAUTION
Risk of incorrect use
Instructions for use must be kept accessible to
the user.

Training

Training for users is available via the Dräger


organization responsible (see www.draeger.com).

Functional safety

The essential performance of lights consists in:


– illumination of the operating field,

Instructions for use Polaris 600 11


Safety-related information

Product-specific safety information

Do not use non-sterile handles Exercise caution when working with the
system
WARNING
WARNING
Risk of infection
Personnel and patient may become infected Risk of personal injury and property damage
by a non-sterile handle. Objects can fall down into the operating field.
Use only sterile handles. Avoid collisions of the light heads or of parts
of the arm system with other objects.

Use of disposable sleeves


Incorrect use
WARNING
WARNING
Risk of infection
Disposable sleeves can fall down into the Risk of personal injury and property damage
operating field. Do not place objects on the light heads or
Dräger recommends using disposable hang them on the arm system. These objects
sleeves from another manufacturer (e.g., can fall down into the operating field.
Covidien) only with the corresponding handle Do not hang heavy loads on the arm system
adapter for disposable sleeves from the same (e.g., leaning body weight). The mechanism
manufacturer, see list of accessories. of the arm system may be damaged, which
may compromise the correct positioning of
the arm system.
Thermal radiation from the lights

Risk of electric shock


WARNING
Risk of patient injury
WARNING
If the light fields from several lights overlap,
the heat input to the wound site may be Risk of electric shock
unacceptably high. This can lead to drying Persons may receive an electric shock if all
out of the wound site. poles are not disconnected during
If necessary, the total irradiance must be maintenance work.
adjusted by reducing the illuminance of Make sure that a switching element to
individual lights, see "Setting the disconnect all poles of the supply is fitted
illuminance" on page 42. before installing the light. The required
switching element is part of the Dräger mains
connection component.

12 Instructions for use Polaris 600


Safety-related information

Direct view of the operating field Observe the maximum load of the
display mount
WARNING
WARNING
Risk of personal injury
A camera is not intended to be used as the Risk of personal injury and property damage
sole visualization medium during an The specification for the maximum load on
operation. the display mount must not be exceeded.
Make sure that the OR personnel have a Comply with the maximum load of the
direct view of the operating field at all times. display mount.

Cautious positioning of the arm system Read and observe the instructions for
use of the displays
CAUTION
Risk of personal injury and property damage WARNING
The stops on the arm system may become Risk of device malfunction
damaged.
To properly use the light system, read and
When positioning the swivel arms and the comply with the instructions for use of the
attached devices, take care not to use force to respective display.
pull the arm system beyond the stops.

OR server
Risk of the light head overheating
WARNING
WARNING
Risk of patient injury
Risk of property damage
The OR servers used must conform to Safety
The light head may overheat during Class B of the IEC 62304 standard.
operation.
Do not cover the light head completely or WARNING
partially.
Risk of patient injury
The OR severs used must conform to the
Compatibility of displays IEC 60601-1:1988 (or later) or IEC 60950-1
standards.

WARNING
Risk due to incompatible displays
The displays used in combination with a
Polaris light system must be approved for
use in the patient environment according to
IEC 60601-1.

Instructions for use Polaris 600 13


Safety-related information

Emission of high-frequency energy

This medical device is optionally equipped with a


radio transmitter.
This medical device has been designed and
manufactured to comply, amongst others, with
emission limit values for high-frequency energy.
These limit values are incorporated in international
safety standards such as IEC 60601-1-2 and in
radio equipment standards such as EN 301 893
which have been accepted by regulation
authorities.
The radio transmitter module of this device also
complies with Part 15 of the FCC Rules. Operation
is subject to the following two conditions:
1 this device may not cause harmful interference,
and
2 this device must accept any interference
received, including interference that may cause
undesired operation of the device.
Changes or modifications not expressly approved
by Dräger could void the user’s authority to
operate the equipment.
This medical device includes radio equipment in
compliance with Directive 2014/53/EU. The full
text of the EU declaration of conformity is available
at the following internet address:
www.draeger.com/doc-radio

14 Instructions for use Polaris 600


Application

Application

Intended use

Intended use of Polaris 600 light

Polaris product family Ceiling-mounted versions


The Polaris 600 light system is classified as a Single light
surgical light system in accordance with IEC
As a single light, the Polaris 600 light is intended to
60601-2-41 and is intended to be used for the local
be used as a surgical or examination light in
illumination of the operating field and examination
operating and treatment rooms for diagnostic and
field on the patient in operating and treatment
treatment purposes which can be interrupted
rooms.
without posing a hazard to the patient if the light
The intended uses remain unchanged when the fails.
Polaris 600 light is combined with a Polaris 100 or
As a single light, the Polaris 600 light is classified
200 light.
as a "small surgical light" in accordance with
IEC60601-2-41.
Double or triple light combination
The combination of 2 or more Polaris 600 lights as
a surgical light system enables use in operating
and treatment rooms and meets the requirements
for a fail-safe surgical light system.

Intended use of the Dräger disposable handle

The Dräger disposable handle provides a sterile


cover for the associated inner handle of Polaris
systems. The Dräger disposable handle is
intended solely for single use.

Intended use of the MedView camera

The MedView camera is intended for recording the


operating field for the purpose of patient
documentation, teaching, and training.
In the Polaris 600 light system, the MedView
camera is used together with a medical display as
a video system. The MedView camera must only
be used in conjunction with the Polaris 600 light.
The recordings must only be used for non-
diagnostic purposes.

Instructions for use Polaris 600 15


Application

Environments of use

Polaris 600 lights are intended for use in rooms


used for medical purposes, especially operating
rooms and treatment rooms.

Description

Light Color temperature

The Polaris 600 light is certified and approved as a The color temperature can be adjusted in 4 steps,
surgical light. The Polaris 600 light uses LED which the surgeon can select depending on tissue
bulbs. This ensures low energy consumption and a structure, operating field, and individual
long service life of the bulb. The small size of the perception.
LED bulbs makes it possible to use a modern
Cold white light increases concentration, while
design which facilitates ease of handling and
warm white light reduces dazzle effects and
cleaning while also maintaining good flow behavior
increases the contrast.
under laminar flow ceilings.

OR light mode

The Polaris 600 light combines the illumination


from 92 white LED bulbs to create a homogeneous
light column with deep cavity illumination and
minimal shadowing. The adjustable color
temperature, good color reproduction, and
adjustable illuminance ensure optimum working
conditions for demanding procedures within the
operating field and the examination field. The
illuminance can be adjusted in the OR light mode.

Ambient light mode (Endo light)

In the ambient light mode, the surroundings can be


illuminated for minimally invasive procedures. In
this light mode, the illuminance is significantly
lower than in the OR light mode. The illuminance
cannot be adjusted in the ambient light mode.

16 Instructions for use Polaris 600


Overview

Overview

Multimedia system with 2 Polaris 600 lights (configuration example)

26588
A
B
C
F G
E
E M

D
H H
N

I I P
Q

J
Q L
J K O

R S

A Ceiling cover B IR receiver for Polaris 600 remote control for


MedView camera

Instructions for use Polaris 600 17


Overview

C Ceiling tube
D Central axis
E Light swivel arm
F Display swivel arm
G Display spring arm
H Light spring arm
I Gimbal mounting
J Light control panel
K Polaris 600 light
L Polaris 600 light with MedView camera and
radio transmitter
M Display mount (24 inch to 32 inch) with radio
receiver
N Display
O Sterilizable handle E
P Sterilizable handle
Q Integrated handle
R Wall-mounted control panel
S Polaris 600 remote control for MedView
camera

18 Instructions for use Polaris 600


Overview

Light system with a Polaris 600 and a Polaris 100 light (configuration
example)

26890
A

B
D
D

C
E E

F
F K K
J
G
I

G H H

A Ceiling cover G Light control panel


B Ceiling tube H Sterilizable handle
C Central axis I Polaris 600 light
D Light swivel arm J Polaris 100 light
E Light spring arm K Integrated handle
F Gimbal mounting L Wall-mounted control panel

Instructions for use Polaris 600 19


Overview

Polaris 600 light with sterilizable handle E

A Gimbal mounting

26592
B Integrated handle
C Light control panel
D Sterilizable handle E
A
E Lens with LED
B F Bottom glass

F E
D

Polaris 600 light with sterilizable handle E and MedView camera

A Gimbal mounting
26591

B Integrated handle
C Light control panel

A D Sterilizable handle E
E Lens with LED
G B F Bottom glass
G MedView camera
A

F E
D

20 Instructions for use Polaris 600


Overview

Light control panel

26793
A
B C D E F G

A key: Light on/standby


B key: Switch synchronization on/off
C key: Reduce illuminance
D key: Increase illuminance
E key: Set color temperature
F key: Set light field diameter
G key: Switch ambient light (Endo light)
on/off
– For further information on controlling the
lighting, see "Operation" on page 35.

Instructions for use Polaris 600 21


Overview

MedView camera (option)

The MedView camera is permanently installed in


the light head. It is located in a segment next to the
LED bulbs.

Polaris 600 remote control for MedView


camera 26679

B C
D
E F
G
H I

A key: MedView camera on/standby


B key: Zoom out video image (Wide)
C key: Zoom in video image (Tele)
D key: Switch automatic exposure on/off
E key: Reduce exposure (manually)
F key: Increase exposure (manually)
G key: Switch autofocus on/off
H key: Focus on distant object (manually)
I key: Focus on near object (manually)
J key: Switch freeze frame on/off

22 Instructions for use Polaris 600


Overview

Wall-mounted control panel for 2 Polaris 600 lights and a MedView


camera (configuration example)

For a Polaris 600 light with MedView camera, the


camera control area is on the right-hand side of the
shared wall-mounted control panel.

26980
/LJKWFRQWURODUHD &DPHUDFRQWURODUHD
$ +
% & ' ( ) , -

/ 0
.

2 3
1

* 4

Light control area Camera control area


A key: Light on/standby H key: MedView camera on/standby
B key: Reduce illuminance I key: Zoom out video image (Wide)
C key: Increase illuminance J key: Zoom in video image (Tele)
D key: Set color temperature K key: Switch automatic exposure on/off
E key: Set light field diameter L key: Reduce exposure (manually)
F key: Switch ambient light (Endo light) M key: Increase exposure (manually)
on/off N key: Switch autofocus on/off
G key: Switch synchronization on/off O key: Focus on distant object (manually)
P key: Focus on near object (manually)
Q key: Switch freeze frame on/off

Instructions for use Polaris 600 23


Overview

OR panel (option)

The OR panel is a messaging and operating panel.


It is used for operating a Polaris 600 light system in
operating rooms and for displaying operational and
error messages.

Polaris 600 control interface (option)

The Polaris 600 control interface is an interface


which enables the control of the Polaris 600 light
with MedView camera via an OR server
(integrated OR system).
Information about the Polaris 600 control interface
is available from Dräger.
Restriction when a Polaris 600 light is
combined with Polaris 100/200 lights
The functions of the Polaris 100/200 light cannot
be controlled via the Polaris 600 control interface.

Abbreviations

Abbreviation Explanation
EMC Electromagnetic compatibility
IR Infrared
LED Light-emitting diode
Sterilizable han-
Sterilizable handle, ergonomic
dle E

24 Instructions for use Polaris 600


Overview

Symbols

Symbol Explanation Symbol Explanation


Observe the instructions for use Take care when touching hot
surfaces
Warning! Strictly follow these
instructions for use Warning! Risk of hand injury
Caution
WEEE marking
2 Not for reuse
Manufacturer
2
STERILIZE
Do not re-sterilize
XXXX Date of manufacture
Product sterile; sterilized by irra-
STERILE R diation Radio transmitter
NON
STERILE Non-sterile
On/off (radio receiver)
Do not use if package is dam-
aged

Use by: YYYY-MM-DD


Expiration date

The product is a medical device


(CE conformity assessment pro-
cedure)

Quantity

Storage temperature

Relative humidity

Ambient pressure

REF Part number

LOT Lot number

Serial number

Revision index

Protect from moisture

Instructions for use Polaris 600 25


Operating concept

Operating concept

Operating controls and their functions on the Polaris 600 light

This table is applicable to light systems consisting


of Polaris 600 lights, with or without a MedView
camera.
This table shows which control elements can be
used to make which settings for the functions.
What is used to set what?

Sterilizable handle E Dräger disposable Handle adapter with


handle sterile disposable
sleeve (e.g., manufac-
tured by Covidien)
Light field diameter X X –
Illuminance X X –
Color temperature – – –
Synchronization – – –
Ambient light (Endo light) – – –
MedView camera – – –

Light controlpanel Wall-mounted control OR panel (option)


panel
Light field diameter X X X
Illuminance X X X
Color temperature X X X
Synchronization X X X
Ambient light (Endo light) X X X
MedView camera – X X

Polaris 600 remote control for Polaris 600 control interface


MedView camera (option)
Light field diameter – X
Illuminance – X
Color temperature – X
Synchronization – X
Ambient light (Endo light) – X
MedView camera X X

26 Instructions for use Polaris 600


Operating concept

Operating controls and their functions on the Polaris 600 and


Polaris 100/200 light system

This table is applicable to light systems consisting


of Polaris 600 lights, with or without a MedView
camera, and Polaris 100/200 lights.
This table shows which control elements can be
used to make which settings for the functions.
What is used to set what?

Sterilizable handle E Dräger disposable handle


Polaris 100/200 Polaris 600 Polaris 100/200 Polaris 600
Light field diameter – X – X
Illuminance – X – X
Color temperature – – – –
Synchronization – – – –
Ambient light (Endo light) – – – –
MedView camera – – – –

Handle adapter with sterile dispos- Light control panel


able sleeve (e.g., manufactured by
Covidien)
Polaris 100/200 Polaris 600 Polaris 100/200 Polaris 600
Light field diameter – – – X
Illuminance – – X X
Color temperature – – – X
Synchronization – – – –
Ambient light (Endo light) – – X X
MedView camera – – – –

Wall-mounted control panel OR panel (option)


Polaris 100/200 Polaris 600 Polaris 100/200 Polaris 600
Light field diameter – X – X
Illuminance X X X X
Color temperature – X – X
Synchronization – – – –
Ambient light (Endo light) X X X X
MedView camera – X – X

Instructions for use Polaris 600 27


Operating concept

Polaris 600 remote control for MedView camera


Light field diameter –
Illuminance –
Color temperature –
Synchronization –
Ambient light (Endo light) –
MedView camera X

Sterilizable handle E

illuminance and the light field diameter.


26546

NOTE

A The sterilizable handle E will lose its functionality


if disposable handles and the associated handle
adapter for disposable handles from
manufacturers other than Dräger are used. It will
no longer be possible to set the illuminance or
the light field diameter in the operating section.

The operating section of the handle is touch-


sensitive, i.e., it is sufficient to lay a finger on the
surface. It is not necessary to exert any pressure.
B The better the contact between finger and surface,
the easier it is to adjust the light.
The operating section of the handle can be
operated when surgical gloves are worn.
 A wiping motion with one finger across the
surface of the operating section at any location
will adjust the illuminance, see "Setting
illuminance on the sterilizable handle E" on
page 43.
 Double-tapping with one finger on the surface
of the operating section at any location will
adjust the light field diameter, see "Setting
illuminance on the sterilizable handle E" on
The handle is used both for positioning and for page 43.
adjusting the light.
It consists of the operating section (A) and the
handle section (B).
The handle section is used for sterile positioning.
The operating section is used for setting the

28 Instructions for use Polaris 600


Operating concept

Correct operation of the sterilizable handle E Operating signal will not be implemented
The touch-sensitive part of the handle is designed

26659
so that, for correct operation, only one finger must
touch the operating section at any one time. If
several fingers touch the operating section
simultaneously, the operating signal will not be
implemented.
Correct operation

26658

– Several fingers are touching the operating


section at the same time

– One finger at a time is touching the operating


section

Wall-mounted control panel (option)

The wall-mounted control panel can incorporate up


to 3 light control panels. Optionally it can include a
camera control panel.
The keys on the wall-mounted control panel are
pressure-sensitive, i.e., a certain pressure must be
applied by the finger.

Instructions for use Polaris 600 29


Operating concept

Color assignment of the lights to the wall-mounted control panel

The individual light control panels on the wall-


mounted control panel are identified by different
colors on the right-hand side. These colors are
also be found in the colored rings on the gimbal
mounting of the associated light.

26564

– The uppermost light control panel on the wall-


mounted control panel and the associated
color are always assigned to the uppermost
light on the arm system.

30 Instructions for use Polaris 600


Assembly and preparation

Assembly and preparation

Checking for operational readiness before each use

Checking the system


WARNING
WARNING Risk of infection
Risk of patient injury The Dräger disposable handle must be
checked for secure attachment after it has
Patients may be endangered if the system is
been fitted to the inner handle. The inner
put into operation without first being
handle must be free from damage.
checked.
Make sure that the system is checked for WARNING
operational readiness.
Risk of infection
1 Check that the device and the arm system are
Disposable sleeves can fall down into the
free of visible damage.
operating field.
2 Check that the power supply to the device is
correct, see "Status LEDs" on page 39. Dräger recommends using disposable
sleeves from another manufacturer (e.g.,
3 Check that the device has been reprocessed in Covidien) only with the corresponding handle
accordance with the hospital's regulations, see adapter for disposable sleeves from the same
"Reprocessing" on page 66. manufacturer, see list of accessories.
4 Check that the locking mechanism on the
handle is engaged. CAUTION
The system is only ready for use when the checks Risk of personal injury and property damage
have been carried out successfully.
The stops on the arm system may become
WARNING damaged.

Risk of property damage When positioning the swivel arms and the
attached devices, take care not to use force to
The light head may overheat during pull the arm system beyond the stops.
operation.
Do not cover the light head completely or NOTE
partially.
Sterilizable handles are supplied by Dräger in a
non-sterile state.
WARNING
Risk of infection Checking the Polaris 600 remote control for
MedView camera
Personnel and patient may become infected
by a non-sterile handle.  Press any key on the remote control.
– Status LED lights up green: the remote control
Use only sterile handles. is functioning.
– Status LED flashes green: the battery charge is
low. The remote control can only continue to be
used for a short time. To maintain the function

Instructions for use Polaris 600 31


Assembly and preparation

of the remote control, the battery must be – Status LED does not light up: the battery is
replaced immediately, see "Replacing the completely discharged or is defective. To
battery in the Polaris 600 remote control for reestablish the function of the remote control,
MedView camera" on page 58. the battery must be replaced immediately, see
"Replacing the battery in the Polaris 600
remote control for MedView camera" on page
58.

Fitting the handle

Fitting the sterilizable handle E

26562
Immediately before an operation, fit the handle C
under sterile conditions as follows:
26561

A 2 Turn the handle (A) clockwise until the locking


mechanism (C) engages audibly.
3 Check that the handle is securely attached.
Turn the handle counterclockwise to do this.
1 Fit the handle (A) on the inner handle so that
the markers (B) line up. WARNING
Risk of patient injury
The sterilizable handle E can fall down into
the operating field if it is not engaged
properly.
The handle must be checked for secure
attachment after it has been fitted to the inner
handle.

32 Instructions for use Polaris 600


Assembly and preparation

– To remove the handle, see "Disassembly" on Fitting the Dräger disposable handle
page 67. (optional)

Safety information concerning the Dräger


Fitting the sterilizable handle disposable handle
– The disposable handle is intended solely for
Immediately before an operation, fit the handle single use. The disposable handle must not be
under sterile conditions as follows: reused, reprocessed, or sterilized.
– The membrane on the disposable handle must

28175
not exhibit any cracks, holes, or other damage.
– The inner handle must be free from damage.
– To prevent soiling of the disposable handle, do
not unpack it until immediately before an
operation.
– Do not use the disposable handle if the
packaging is damaged.
– Do not use a knife to open the packaging for
the disposable handle.

WARNING
Risk of infection

A Reuse of the disposable handle can lead to


cross-infection.
The disposable handle is intended solely for
1 Push the handle (A) onto the inner handle. The single use. The disposable handle must not
handle must engage audibly. If necessary, turn be reused.
the handle slightly to the right or left.
2 Check that the handle is securely attached. To
do so, pull in the direction opposite to that used
to attach it.

WARNING
Risk of patient injury
The sterilizable handle can fall down into the
operating field if it is not engaged properly.
The handle must be checked for secure
attachment after it has been fitted to the inner
handle.

– To remove the handle, see "Disassembly" on


page 67.

Instructions for use Polaris 600 33


Assembly and preparation

Immediately before an operation, fit the disposable

36006
handle under sterile conditions as follows:
C

36005

A
B

2 Turn the disposable handle clockwise (C) until


it can be felt to engage.
3 Check that the disposable handle is securely
attached. To do this, hold the disposable
1 Hold the disposable handle (A) by the recessed
handle by the recessed grips and turn it
grips (B) and fit it onto the inner handle.
counterclockwise.
NOTE
WARNING
To facilitate fitting the disposable handle, non-
Risk of patient injury
sterile OR personnel may hold the light steady.
The disposable handle can fall down into the
WARNING operating field if it is not engaged properly.
The disposable handle must be checked for
Risk of infection
secure attachment after it has been fitted to
If the inner handle is not free from damage, it the inner handle.
may damage the disposable handle.
– To remove the disposable handle, see
The inner handle must be free from damage.
"Disassembly" on page 67.

34 Instructions for use Polaris 600


Operation

Operation

Positioning a light

WARNING CAUTION
Risk of patient injury Risk of personal injury and property damage
The proper functioning of the system must be The stops on the arm system may become
checked each time before use. damaged.
When positioning the swivel arms and the
WARNING attached devices, take care not to use force to
pull the arm system beyond the stops.
Risk of patient injury
The system may only be used in accordance Sterile positioning of a light
with the information in the chapter, "Intended
The light can be rotated by 360° around the
use".
horizontal axis at the central joint.

WARNING The sterile handle is located on the bottom glass of


the light.
Risk of patient injury

26915
If the light fields from several lights overlap,
the heat input to the wound site may be
unacceptably high. This can lead to drying
out of the wound site.
If necessary, the total irradiance must be
adjusted by reducing the illuminance of
individual lights, see "Setting the
illuminance" on page 42. A

WARNING
Risk of personal injury and property damage
Do not place objects on the light heads or A
hang them on the arm system. These objects
can fall down into the operating field. 1 Take hold of the handle (A) by the handle
Do not hang heavy loads on the arm system section and use it to position the light.
(e.g., leaning body weight). The mechanism The light head is held at its new position.
of the arm system may be damaged, which
may compromise the correct positioning of
the arm system.

Instructions for use Polaris 600 35


Operation

Sterile positioning of a light with a short gimbal 1 Take hold of the integrated handle (A) and use
mounting it to position the light.
The short gimbal mounting has a shorter upper The light head is held at its new position.
bow. The light can be rotated by 330° around the Non-sterile positioning of a light with a short
horizontal axis at the central joint. The rotational gimbal mounting
movement is limited in both directions by stops.
The short gimbal mounting has a shorter upper
The sterile handle is located on the bottom glass of bow. The light can be rotated by 330° around the
the light. horizontal axis at the central joint. The rotational
movement is limited in both directions by stops.
35334
The non-sterile integrated handle is located at the
edge of the light head.

CAUTION
A

Risk of crushing
When positioning the light with short gimbal
mounting, hand injuries may occur.
Keep sufficient space between the integrated
A handle and the upper bow of the short gimbal
mounting and between the integrated handle and
1 Take hold of the handle (A) by the handle the spring arm.
section and use it to position the light.

35335
The light head is held at its new position.
Non-sterile positioning of a light A
The light can be rotated by 360° around the
horizontal axis at the central joint. A

The non-sterile integrated handle is located at the


edge of the light head.
26924

1 Take hold of the integrated handle (A) and use


it to position the light.
A The light head is held at its new position.

36 Instructions for use Polaris 600


Operation

Positioning a light system

It is recommended that the lights are positioned so – The second light (C) is positioned at the foot of
that the surgeon and the sterile OR personnel can the operating table.
easily reach the sterile handle.

26959
When the lights are positioned in this way, the non-
sterile OR personnel can reach the integrated D
handle in order to make further adjustments if
required.
With central axes with more than one swivel arm, H
make sure that the different swivel arms and their
devices are not all positioned on one side.
Positioning several swivel arms on one side can E
cause tilting or unintended drifting of the swivel
arms.
F
NOTE
Positioning several swivel arms on one side can G
cause tilting or unintended drifting of individual
swivel arms. Take care that the different swivel
arms and their devices are not all positioned on
one side. The figure shows an example of how a
combination of 3 lights and a display can be
positioned to achieve suitable illumination of the
26958

operating field.
– The main surgical light (D) is positioned
laterally above the surgeon's shoulder to
A ensure that the operating field (E) is fully
illuminated.
– The second light (F) is positioned at the foot of
B the operating table.
C
– The third light (G) is positioned at the side of
the operating table.
– The display (H) is positioned at the head of the
operating table.

WARNING
Risk of patient injury
The figure shows an example of how a The patient's unprotected eyes can be
combination of 2 lights can be positioned to damaged by direct exposure to light.
achieve suitable illumination of the operating field.
Position the light so that the light field is not
– The main surgical light (A) is positioned directed towards the unprotected eyes of the
laterally above the surgeon's shoulder to patient.
ensure that the operating field (B) is fully
illuminated.

Instructions for use Polaris 600 37


Operation

CAUTION NOTE
Risk of personal injury A distance of 100 cm (39.37 in) from the bottom
glass of the light to the operating field is
The user's eyes can be damaged by direct
recommended to achieve the optimum
exposure to light.
illumination of the operating field.
Do not look directly into the light field.
When using two or more lights to illuminate the
CAUTION operating field, observe the following to ensure
that the total permissible irradiance is not
Risk of property damage exceeded:
The light heads may overheat during operation. – Set the lights to the minimum illuminance
required for appropriate illumination of the
Position the light heads so that their light fields do operating field. If a higher illuminance is only
not illuminate other light heads. required temporarily, reduce the illuminance as
soon as it is no longer needed.
– If the maximum or nearly maximum illuminance
of a light is required, make sure that the
illuminance of the other lights is reduced and
that they do not illuminate the operating field.

Positioning the MedView camera


26932

26976
A

A
B

 The MedView camera (A) is permanently  Position the light head so that the MedView
installed in the light head. It is located in a camera has an open field of view (B) at all
segment next to the LED bulbs. times.
 The optimum distance between the MedView
camera (from the bottom glass of the light) and
the operating field is 90 cm to 130 cm (35.43 in
to 51.18 in).

38 Instructions for use Polaris 600


Operation

Status LEDs

The description of this function applies to both the Status LED for monitoring the power supply
light control panel and the wall-mounted control
The status LED (A) for monitoring the power
panel. Only the light control panel is shown in the
supply for the light system is located to the left of
illustrations.
the key.
Status LED for monitoring a light

26477
The status LED (A) for monitoring the light is
situated directly next to the key.

A
26399

– The status LED (A) is lit green.


The light is being supplied with mains power.
– The status LED (A) is lit orange.
– The status LED (A) is lit green.
The light is being supplied with emergency
The light is in operation. power.
– The status LED (A) is not lit. – The status LED (A) is not lit.
The light is in standby mode. The light is not being supplied with power.
– The status LED (A) is lit orange.
There is a light system fault, see "Fault –
Cause – Remedy" on page 60.

Instructions for use Polaris 600 39


Operation

Switching the light on and off

The description of this function applies to both the – Even if the Ambient light mode was the last to
light control panel and the wall-mounted control be active, the light will operate in the OR light
panel. Only the light control panel is shown in the mode after switch-on, with the illuminance (C)
illustrations. that was set the last time the light was used.
Switching the light on Switching the light off

26415

26983
B C B

A A

1 Press the key (A). 1 Press the key (A).


The light switches on in the OR light mode. The status LED (B) flashes green to confirm
The status LED (B) lights up. the switch-off procedure.

If there is a light system fault, the status LED The light is in standby mode.
(B) lights up orange, see "Fault – Cause – The status LED (B) is not lit.
Remedy" on page 60. If there is a light system fault, the status LED
– The illuminance (C) when the light is switched (B) lights up orange, see "Fault – Cause –
on corresponds to the last illuminance set Remedy" on page 60.
before the light was switched off.

Synchronization function

Within a light system, the Synchronization


CAUTION function can be activated at the light control panel
Risk of patient injury of each Polaris 600 light. When the
Synchronization function is activated, the
When the Synchronization function is activated, selected Polaris 600 light sends a request to all the
the wound site may be heated to an other Polaris 600 lights in the light system. The
unacceptable level if the light fields overlap. This request contains the value of the color temperature
can lead to drying out of the wound site. and the value of the illuminance of the selected
If necessary, the total irradiance must be Polaris 600 light. The other Polaris 600 lights then
adjusted by reducing the illuminance of individual set their color temperature and illuminance to the
lights, see "Setting the illuminance" on page 42.

40 Instructions for use Polaris 600


Operation

requested values. All Polaris 600 lights now have

26515
the same color temperature and the same
illuminance.
When the Synchronization function is activated
on the wall-mounted control panel, the panel B
sends a request to all Polaris 600 lights to set the
color temperature and the illuminance to specific A
values. The color temperature and the illuminance
are then set on the Polaris 600 lights in
accordance with the request. All Polaris 600 lights
now have the same color temperature and the
same illuminance.
1 Press the key (A).
When the Synchronization function is activated,
the Polaris 600 lights can be individually switched The Synchronization function is activated.
to standby mode. This deactivates the The status LED (B) lights up.
Synchronization function on a light. When a
Polaris 600 light in the light system is in standby Switching the Synchronization function off
mode, the key on this light is deactivated. The

26515
Synchronization function will not be activated
when the key is pressed. When all the
Polaris 600 lights in the light system are in standby
mode, the key on the wall-mounted control B
panel is also deactivated.
When the Synchronization function is activated, A
the Polaris 600 lights can be individually switched
to the Ambient light mode. This deactivates the
Synchronization function on a light. When a
Polaris 600 light in the light system is in the
Ambient light mode, the key on this light is
deactivated. The Synchronization function will 1 Press the key (A).
not be activated when the key is pressed.
When all the Polaris 600 lights in the light system The status LED (B) is not lit.
are in the Ambient light mode, the key on the The Synchronization function is deactivated.
wall-mounted control panel is also deactivated.
Settings made during synchronization are
Switching the Synchronization function on preserved.
The light must be switched on before this function All Polaris 600 lights can be individually set
can be activated, see "Switching the light on and once more.
off" on page 40. Technical background
The transmitted request is what is known as a
"broadcast command", which either the selected
Polaris 600 light or the wall-mounted control panel
transmits to the Polaris 600 lights in the light
system. The broadcast command contains the
values for illuminance and color temperature.
In a network of lights, each of the Polaris 600 lights
receives the same broadcast command. The
Polaris 600 lights receiving the broadcast
command set their respective illuminance and
color temperature to the requested values. With

Instructions for use Polaris 600 41


Operation

each change in color temperature or illuminance, a wall-mounted control panel or from the selected
new broadcast command is transmitted to the Polaris 600 light without the receiving Polaris 600
Polaris 600 lights. The Polaris 600 lights receiving lights being able to identify the sender.
the broadcast command cannot use the command
Restriction when a Polaris 600 light is
to trace back whether the command originated
combined with Polaris 100/200 lights
from the selected Polaris 600 light or from the wall-
mounted control panel. In particular, the Polaris The Polaris 100/200 light does not receive any
600 lights cannot use the broadcast command to requests from a Polaris 600 light or from the wall-
trace back which Polaris 600 light transmitted the mounted control panel. Illuminance can always be
command. Thus a request can be sent from the set individually on the Polaris 100/200 light.

Setting the illuminance

Setting illuminance on the light control Increasing illuminance


panel and on the wall-mounted control

26418
panel

The description of this function applies to both the C


light control panel and the wall-mounted control
panel. Only the light control panel is shown in the
illustrations.
The light must be switched on before this function A B
can be activated, see "Switching the light on and
off" on page 40.
Reducing illuminance
 Press the key (B).
26418

The illuminance is increased.


If the key is kept pressed, the illuminance is
C increased continuously.
The change in the illuminance is shown by the
LED indicator (C).
A B – For further information on illuminance, see
"Technical light data" on page 84.

 Press the key (A).


The illuminance is reduced.
If the key is kept pressed, the illuminance is
reduced continuously.
The change in the illuminance is shown by the
LED indicator (C).

42 Instructions for use Polaris 600


Operation

Functional principle of the LED indicator

41287
The illuminance can be set in 12 steps. These 12
steps are visualized by the 6 LEDs of the LED C
indicator, which light up accordingly with half or full
brightness.

26512
B
A

3 When the key (A) is pressed again, the LED


(C) of the LED indicator lights up with full
brightness.

41285
– At minimum illuminance, the LED (B) of the
LED indicator is lit at half brightness.
26513

– At maximum illuminance, all the LEDs of the


LED indicator are lit with full brightness.
A
Setting illuminance on the sterilizable
1 When the key (A) is pressed once, the
LED (B) of the LED indicator lights up with full
handle E
brightness.
A wiping motion with one finger across the surface
of the operating section at any location will adjust
26558

the illuminance. Each wiping motion changes the


C illuminance by one step.

NOTE
If several fingers touch the operating section of
the sterilizable handle E simultaneously, the
operating signal will not be implemented. Only
one finger at a time must touch the operating
A section.

2 When the key (A) is pressed twice, the


next LED (C) of the LED indicator lights up with
half brightness.

Instructions for use Polaris 600 43


Operation

Increasing illuminance
NOTE

26606
The operating section of the sterilizable handle E
must be dry to guarantee proper operability.

Reducing illuminance

26607
A

 Make a wiping motion with one finger to the


right (A) across the surface of the operating
section.
The illuminance will be increased by one step.

 Make a wiping motion with one finger to the left

28240
(A) across the surface of the operating section.
The illuminance will be reduced by one step.
B
28240

 The change in the illuminance is shown by the


LED indicator (B).
– For further information on illuminance, see
"Technical light data" on page 84.
 The change in the illuminance is shown by the
LED indicator (B).

44 Instructions for use Polaris 600


Operation

Maximum total irradiance for


combinations of lights

WARNING
Risk of patient injury
If the light fields from several lights overlap,
the heat input to the wound site may be
unacceptably high. This can lead to drying
out of the wound site.
If necessary, the total irradiance must be
adjusted by reducing the illuminance of
individual lights, see "Setting the
illuminance" on page 42.

When using 2 or more lights to illuminate the


operating field, the permissible standard total
irradiance limit may be exceeded if the light fields
from several lights overlap. This can lead to
increased drying out of the wound site. The
thermal radiation from each light increases linearly
with increasing illuminance. A single light cannot
exceed the permissible total irradiance limit, but it
is possible when a combination of lights is used. To
avoid increased thermal radiation, observe the
following recommendations:
– Set the lights to the minimum illuminance
required for appropriate illumination of the
operating field. If a high level of illuminance is
temporarily required, it should be reduced
again as quickly as possible.
– If the maximum level of illuminance from one
light is required, the illuminance from the other
lights must be reduced. The light fields from the
individual lights should not overlap if possible.

Instructions for use Polaris 600 45


Operation

The following table lists examples of illumination


settings for different combinations of lights. With
these illumination settings, there is no
impermissibly high thermal radiation into the
wound site if the light fields overlap.

Light combina- Maximum Recommended adjustment of the total irradi- Thermal radiation in
tion total irradi- ance2) by reducing the illuminance the wound site after
ance1) adjustment of the
total irradiance
[W/m2]

Polaris 600 (A) 600 W/m2 Acceptable

(A)

Polaris 600 (A)


Polaris 600 (B) 1200 W/m2 Acceptable

(A) (B)

Polaris 200 (A)


Polaris 200 (B) 1120 W/m2 Acceptable

(A) (B)

Polaris 600 (A)


Polaris 200 (B) 1160 W/m2 Acceptable

(A) (B)

Polaris 600 (A)


Polaris 100 (B) 1060 W/m2 Acceptable

(A) (B)

Polaris 600 (A)


Polaris 600 (B) 1800 W/m2 Acceptable
Polaris 600 (C)
(A) (B) (C)

46 Instructions for use Polaris 600


Operation

Light combina- Maximum Recommended adjustment of the total irradi- Thermal radiation in
tion total irradi- ance2) by reducing the illuminance the wound site after
ance1) adjustment of the
total irradiance
[W/m2]

Polaris 200 (A)


Polaris 200 (B) 1740 W/m2 Acceptable
Polaris 200 (C)
(A) (B) (C)

Polaris 600 (A)


Polaris 600 (B) 1780 W/m2 Acceptable
Polaris 200 (C)
(A) (B) (C)

Polaris 600 (A)


Polaris 200 (B) 1720 W/m2 Acceptable
Polaris 200 (C)
(A) (B) (C)

Polaris 600 (A)


Polaris 600 (B) 1660 W/m2 Acceptable
Polaris 100 (C)
(A) (B) (C)

Polaris 600 (A)


Polaris 100 (B) 1520 W/m2 Acceptable
Polaris 100 (C)
(A) (B) (C)

Polaris 600 (A)


Polaris 200 (B) 1620 W/m2 Acceptable
Polaris 100 (C)
(A) (B) (C)
1) When all lights in the light combination are set to maximum illuminance
2) Maximum permissible total irradiance according to IEC 60601-2-41 corresponds to 1000 W/m2

Instructions for use Polaris 600 47


Operation

Color temperature function

The description of this function applies to both the 2 Press the key (A) again.
light control panel and the wall-mounted control The color temperature is increased by a further
panel. Only the light control panel is shown in the step.
illustrations.
3 The change in color temperature is shown by
The light must be switched on before this function the LED indicator (B).
can be activated, see "Switching the light on and
off" on page 40. 4 If the highest color temperature has been
reached, pressing the key (A) again will
There is an indicator with 3 LEDs for the color reset the color temperature to its lowest value.
temperature. These LEDs indicate the color
– For further information on color temperature,
temperature set for the light.
see "Technical light data" on page 84.
LED indicator Color temperature
3800 K

4400 K

5000 K

5600 K

– The color temperature when the light is


switched on is the same as the color
temperature set before the light was switched
off.
26566

1 Press the key (A).


The color temperature is increased.

48 Instructions for use Polaris 600


Operation

Light field diameter function

Setting the light field diameter on the 2 Press the key (A) again.
light control panel and on the wall- The light field diameter is increased by a
mounted control panel further step.
3 The change in the light field diameter is shown
The description of this function applies to both the by the LED indicator (B).
light control panel and the wall-mounted control  If the largest light field diameter has been
panel. Only the light control panel is shown in the reached, pressing the key (A) again will
illustrations. reset the light field diameter to its smallest size.
The light must be switched on before this function – For further information on the light field
can be activated, see "Switching the light on and diameter, see "Technical light data" on page
off" on page 40. 84.
LED indicator Light field diameter
Small Setting the light field diameter on the
sterilizable handle E

Medium Double-tapping with one finger on the surface of


the operating section at any location will adjust the
light field diameter.

Large NOTE
If several fingers touch the operating section of
the sterilizable handle E simultaneously, the
– The light field diameter when the light is operating signal will not be implemented. Only
switched on is the same as the light field one finger at a time must touch the operating
diameter set before the light was switched off. section.

NOTE
26608

The operating section of the sterilizable handle E


B must be dry to guarantee proper operability.

1 Press the key (A).


The light field diameter is increased by one
step.

Instructions for use Polaris 600 49


Operation

– For further information on the light field

26614
diameter, see "Technical light data" on page
84.

1 Double-tap with one finger (A).


The light field diameter is increased by one
step.
2 Double-tap again with one finger (A).
The light field diameter is increased by a
further step.
26620

3 The change in the light field diameter is shown


by the LED indicator (B).
 If the largest light field diameter has been
reached, double-tapping again will reset the
light field diameter to its smallest size.

50 Instructions for use Polaris 600


Operation

Ambient light mode (Endo light)

When the Synchronization function is activated, Switching the ambient light mode off
the Polaris 600 lights can be individually switched
to the Ambient light mode. This deactivates the

26681
Synchronization function on a light. When a
Polaris 600 light in the light system is in the
Ambient light mode, the key on this light is B
deactivated. The Synchronization function will
not be activated when the key is pressed.
When all the Polaris 600 lights in the light system
are in the Ambient light mode, the key on the A
wall-mounted control panel is also deactivated.
Switching the ambient light mode on
The description of this function applies to both the
light control panel and the wall-mounted control  Press the key (A).
panel. Only the light control panel is shown in the
illustrations. The Ambient light mode is deactivated.
The light must be switched on before this function The light switches to the OR light mode.
can be activated, see "Switching the light on and The status LED (B) is not lit.
off" on page 40. – For further information on the ambient light
mode (Endo light), see "Technical light data" on
26681

page 84.

 Press the key (A) for 1.5 seconds.


The light switches to the Ambient light mode.
The status LED (B) lights up.

Instructions for use Polaris 600 51


Operation

Operating the MedView camera using the wall-mounted control panel

For a Polaris 600 light with MedView camera, the Switching the MedView camera off
camera control area is on the right-hand side of the
shared wall-mounted control panel.

26358
A
Switching the MedView camera on B

28134
A
B
C

C
D

 Press the key (A).


The camera is in standby mode.
The status LED (B) is not lit.
 Press the key (A).
– The key does not disconnect the camera
The status LED (B) lights up. from the mains power supply (camera in
– After switch-on, the camera runs through an standby mode).
initialization procedure. This procedure takes
about 15 seconds.
– The Automatic exposure and Autofocus
functions are activated by default.
– The camera can also be operated when the
Polaris 600 light is in standby mode.

52 Instructions for use Polaris 600


Operation

Zoom function Automatic exposure function


The image detail can be magnified or reduced with The Automatic exposure function is activated by
the zoom keys. default.

26361

26445
C

A B

1 Press the key (A). When the Manual exposure function is activated,
The image detail is reduced. it is possible to switch back to the Automatic
exposure function.
If the key is kept pressed, the image is reduced
continuously.  Press the key (A).

2 Press the key (B). The Automatic exposure function is activated.

The image detail is magnified. The status LED (B) lights up.

If the key is kept pressed, the image is When the Automatic exposure function is
magnified continuously. activated, it is possible to switch to the Manual
exposure function.
The camera enables up to 120x magnification
(10x optical / 12x digital).  Press the key (A).

3 The change in the image detail is shown by the The Manual exposure function is activated.
LED indicator (C). The status LED (B) is not lit.
The larger the image detail becomes, the more
LEDs light up.
The smaller the image detail becomes, the
fewer LEDs light up.

Instructions for use Polaris 600 53


Operation

Manual exposure function Optimum image illumination


The Automatic exposure function is activated by

26389
default.

26362
B A

A B
– To achieve optimum image illumination, the
zoom function can be used to match the size of
the image detail (A) to the light field diameter
(B).
– The camera can only record optimally when the
illumination is uniform.

1 Press the key (A) or the key (B).


The Manual exposure function is activated.
2 Press the key (A).
The video image becomes darker.
If the key is kept pressed, the video image
darkens continuously.
3 Press the key (B).
The video image becomes lighter.
If the key is kept pressed, the video image
lightens continuously.
4 The change in the exposure is shown by the
LED indicator (C).
The lighter the video image becomes, the more
LEDs light up.
The darker the video image becomes, the
fewer LEDs light up.

54 Instructions for use Polaris 600


Operation

Autofocus function Manual focus function


The Autofocus function is activated by default. With the Manual focus function, focusing on an
object is carried out manually.
With the Autofocus function, continuous
automatic focusing of the video image takes place.

26377
26446

B A B

When the Autofocus function is activated, it is


possible to switch to the Manual focus function.
When the Manual focus function is activated, it is
possible to switch back to the Autofocus function. 1 Press the key (A) or the key (B).
 Press the key (A). The Manual focus function is activated.
The Autofocus function is activated. 2 Press the key (A).
The status LED (B) lights up. The camera focuses on a more distant object.
3 Press the key (B).
NOTE
The camera focuses on a closer object.
The Manual focus function must be activated if
the current focus setting is to be maintained
without continuous automatic focusing taking
place.

Instructions for use Polaris 600 55


Operation

Freeze frame function


The Freeze frame function is used to switch
between a real-time image and a freeze frame.

26380

B A

1 Press the key (A).


A freeze frame is generated.
The status LED (B) lights up.
2 Press the key (A) again.
Real-time images are shown.
The status LED (B) is not lit.
Automatic white balance
The white balance for the camera takes place
automatically, i.e., the camera adjusts itself
automatically to the color temperature of the
Polaris 600 light.

NOTE
The MedView camera adjusts itself automatically
to the color temperature selected for the Polaris
600 light.

56 Instructions for use Polaris 600


Operation

Operating the MedView camera by means of the Polaris 600 remote


control for MedView camera

Operating the remote control Positioning the remote control


The keys on the remote control are identical to the

26392
keys in the camera control area on the right-hand B
side of the wall-mounted control panel.

26391
A

1 Select the desired function for the MedView


camera by pressing the corresponding key on
the remote control.
– Status LED lights up green: The remote control 1 The remote control (A) must be pointed at the
is functioning. IR receiver (B) on the ceiling cover.
– Status LED flashes green: The battery charge – The distance (C) between the remote control
is low. The remote control can only continue to and the IR receiver must not exceed 7 meters
be used for a short time. To maintain the (276 in).
function of the remote control, the battery must – Up to 2 IR receivers can be installed (option).
be replaced immediately, see "Replacing the
battery in the Polaris 600 remote control for NOTE
MedView camera" on page 58.
– Status LED does not light up: The battery is The operation of the remote control can be
completely discharged or is defective. To disturbed by strong incidence of extraneous light
reestablish the function of the remote control, on the IR receiver.If there are objects between
the battery must be replaced immediately, see the remote control and the IR receiver, this can
"Replacing the battery in the Polaris 600 also interfere with the operation of the remote
remote control for MedView camera" on page control.
58.

Instructions for use Polaris 600 57


Operation

Replacing the battery in the Polaris 600 remote control for MedView
camera

The remote control requires 9-volt batteries of the

27992
type 6LR61, 6F22, or 1604D.
The operating time of the remote control depends
on the capacity of the battery used and is
approximately 1 year for the factory-supplied
battery. A

WARNING
Risk of damage to the device
Batteries must be stored in accordance with
the manufacturer's specifications. Batteries
must not be disassembled, opened or 1 Using light pressure, raise the housing flap (A)
crushed. Batteries must not be short- at the hinge at the bottom while at the same
circuited. Batteries must not be exposed to time folding it out at the top.
either heat or fire. Batteries must not be

27993
stored in direct sunlight.

CAUTION
Risk of damage to the device B
Only 6LR61, 6F22, or 1604D battery types may
be used, as otherwise the operation of the
remote control may be compromised.

CAUTION
Risk of damage to the device 2 Unlock the cover (B) with a ballpoint pen or
screwdriver and fold it out.
Non-rechargeable batteries must not be charged.
27994

CAUTION
Risk of damage to the device C
The battery must be removed if the remote D
control will not be used for a certain period of
time. If there is fluid escaping from the battery,
the fluid must not come into contact with the skin
or get into the eyes. If there has been contact,
rinse the affected area with plenty of water and
seek medical help immediately.

3 Withdraw the battery (C) together with its clip-


on connector (D) from the compartment.

58 Instructions for use Polaris 600


Operation

4 Remove the clip-on connector from the battery.


Do not pull on the wires when doing this.

27995

5 Fit the clip-on connector to the new battery,


observing the correct polarity (E).
6 Insert the new battery into the compartment.
7 Close the cover.
8 Close the housing flap.
9 Perform a functional test, see "Checking for
operational readiness before each use" on
page 31.
10 Dispose of the old battery appropriately.
11 Charge rechargeable batteries as required.

Instructions for use Polaris 600 59


Troubleshooting

Troubleshooting

Fault – Cause – Remedy

LEDs on the mains connection


component

The LEDs on the mains connection component are


numbered and indicate the status of the light
system.
27270

Supply mains
20

Emergency power
21

Output voltage
22

Initialize in progress
23 24

Contact service
25
30

Initialize system

90 1
78

System number A
31
2 34

56

1
EF0 2
CD

345

32

System number B
AB

67
89

60 Instructions for use Polaris 600


Troubleshooting

No. LED Status Cause Remedy


Off No mains power available Notify the in-
house technician
Mains power sup-
Continuously yel- Mains power available No remedial
20 ply (Supply
low action necessary
mains)
Flashing yellow See "LEDs on the mains connection component (signal
combinations)" on page 62
Off No emergency power available Notify the in-
house technician
Emergency power
Continuously yel- Emergency power available No remedial
21 supply (Emer-
low action necessary
gency power)
Flashing yellow See "LEDs on the mains connection component (signal
combinations)" on page 62
Off Light is not supplied with power Notify Dräger-
Service
Continuously yel- Light is supplied with power No remedial
Output voltage to low Components of the light are in working action necessary
22 light (Output volt- order
age) Communication with the light is func-
tioning
Flashing yellow See "LEDs on the mains connection component (signal
combinations)" on page 62
Off Light system is initialized No remedial
action necessary
Continuously yel- Light system is initializing (during ini- No remedial
low tialization, do not press any key and do action necessary
not initialize another light system at the
System initializa-
same time)
23 tion (Initialize in
progress) Flashing yellow Light system is not initialized Perform initial-
ization according
to assembly
instructions or
notify Dräger-
Service
Off Light system is working correctly No remedial
action necessary
Continuously yel- Electronics fault Notify Dräger-
low Communication with the light system is Service
Notify DrägerSer- disrupted
24 vice (Contact ser- Flashing yellow Light system is not initialized Perform initial-
vice) ization according
to assembly
instructions or
notify Dräger-
Service

Instructions for use Polaris 600 61


Troubleshooting

No. LED Status Cause Remedy


Off Processor of the mains connection Notify Dräger-
component is faulty Service
25 LED
Pulsating yellow Processor of the mains connection No remedial
component is working correctly action necessary

LEDs on the mains connection


component (signal combinations)

No. LED Status Cause Remedy


Mains power Flashing yellow
20 supply (Sup-
ply mains)
Emergency Flashing yellow
power supply
21
(Emergency Internal fault in the mains connection
power) component Notify Dräger-
Output volt- Flashing yellow Input source is unknown Service
age to light Output voltage level is unknown
22
(Output volt-
age)
Notify Dräger- Continuously yellow
24 Service (Con-
tact service)

LEDs on the light control panel and on


the wall-mounted control panel

The description of this function applies to both the


light control panel and the wall-mounted control
panel. Only the light control panel is shown in the
illustrations.
27795

B A D
C

62 Instructions for use Polaris 600


Troubleshooting

LED Status Cause Remedy


Off Light is in standby mode No remedial action nec-
Ambient light mode (Endo light) essary
is switched on
Continuously white Indicates the illuminance in OR No remedial action nec-
LED indicator (A) on the light control light mode (the more LEDs are essary
panel lit, the stronger the illuminance)
Continuously orange
on the wall-mounted
control panel
Off Light is not supplied with power Notify the in-house
technician or Dräger-
Service
Off Light is not on No remedial action nec-
Status LED (B) Light is in standby mode essary
Continuously green Light is in operation No remedial action nec-
essary
Continuously orange Electronics fault Notify DrägerService
Off Light system is not supplied with Notify the in-house
power technician or Dräger-
Service
Continuously green Mains power is available No remedial action nec-
Status LED (C)
essary
Continuously orange Mains power is interrupted (light Notify the in-house
system is being supplied with technician
emergency power)
Off Ambient light mode (Endo light) No remedial action nec-
is not switched on (OR light essary
mode is switched on)
Light is in standby mode
Status LED (D) Continuously white Light is in ambient light mode No remedial action nec-
on the light control (Endo light) essary
panel
Continuously orange
on the wall-mounted
control panel

Instructions for use Polaris 600 63


Troubleshooting

Polaris 600 remote control for MedView


camera

Fault Cause Remedy


Remote control is not within the reception range of the IR Aim the remote control at
receiver the IR receiver, see
"Operating the MedView
camera by means of the
Polaris 600 remote con-
trol for MedView camera"
on page 57
Remote control is Hold the remote control
not responding closer to the IR receiver,
(status LED lights see "Operating the Med-
up green when View camera by means
key is pressed) of the Polaris 600 remote
control for MedView
camera" on page 57
Room address of the remote control or of the MedView Remote control and
camera is incorrectly set MedView camera must
be set to the same room
address.
Notify DrägerService
Battery is discharged Fit a new battery, see
"Replacing the battery in
the Polaris 600 remote
Remote control is control for MedView
not functioning camera" on page 58
(status LED does
not light up when Battery is defective Fit a new battery, see
key is pressed) "Replacing the battery in
the Polaris 600 remote
control for MedView
camera" on page 58
Remote control is Membrane keypad is damaged Replace defective
not functioning remote control
(status LED is lit
green continu-
ously)
Remote control is Battery charge is exhausted Fit a new battery, see
functioning but "Replacing the battery in
status LED the Polaris 600 remote
flashes green control for MedView
when key is camera" on page 58
pressed

64 Instructions for use Polaris 600


Troubleshooting

Wireless video transmission

Fault Cause Remedy


MedView camera is carrying out an initialization after No remedial action nec-
standby mode. essary. This procedure
takes up to 30 seconds
Connection between the radio transmitter and the radio Switch the MedView
receiver is interrupted camera to standby mode.
Switch the MedView
camera on again after a
few seconds
Connection between the radio transmitter and the radio Reestablish the connec-
receiver is interrupted tion:
1 Push up the locking
plate on the radio
receiver housing (dis-
Video image is
play mount).
not displayed
2 Set the on/off switch
on the radio receiver
to 0.
3 Wait for approx. 5
seconds.
4 Set the on/off switch
on the radio receiver
to I.
 The connection will
be reestablished.
This procedure takes
up to 30 seconds
Video image is Unknown Notify DrägerService
not displayed and
cannot be rees-
tablished

Instructions for use Polaris 600 65


Reprocessing

Reprocessing

Safety information

WARNING
Risk of patient injury and of device
malfunction
The device and accessories must be cleaned
and disinfected using a validated procedure.

WARNING
Risk of patient injury and of device
malfunction
Always observe the hygiene regulations of
the hospital when handling devices which are
contaminated with body fluids.

NOTE
To avoid the risk of infections for hospital staff
and patients, always clean and disinfect the
device after each use.

NOTE
Do not use abrasive cleaners for cleaning.

NOTE
Strictly follow the disinfectant manufacturer's
instructions for use!

66 Instructions for use Polaris 600


Reprocessing

Disassembly

Removing the sterilizable handle E

26647
The handle is located on the bottom glass of the
light and is removed from the inner handle by
pulling.

26638
A

B
3 Pull the handle (D) down to remove it.
Removing the sterilizable handle
The handle is located on the bottom glass of the
light and is removed from the inner handle by
1 Take hold of the handle section and press the pulling.
locking mechanism (A).

28274
2 Keep the locking mechanism pressed and turn
the handle counterclockwise (B) until the
markers (C) line up.

A
B

1 Take hold of the handle (A).

Instructions for use Polaris 600 67


Reprocessing

2 Push and hold down the button (B) at the lower

36026
end of the handle and pull off the handle.
Removing the Dräger disposable handle
D
The disposable handle is located on the bottom
glass of the light and is removed from the inner C
handle by pulling.

36025
A

A
3 Hold the disposable handle by the recessed
grips (C) and turn it counterclockwise (D).

36027
1 Take hold of the disposable handle (A).
2 Pull the locking mechanism (B) on the
disposable handle until the entire perforation is
ripped open.
The locking mechanism is intended to ensure
single use of the disposable handle and is
unusable after it has been pulled.

4 Pull the disposable handle (E) down to remove


it.

68 Instructions for use Polaris 600


Reprocessing

 Dispose of the disposable handle in


accordance with the applicable infection
prevention policy and disposal regulations of
the hospital.

Service life of the handle

Components:
NOTE
– Sterilizable handle E
– Sterilizable handle Use and frequent sterilization cause wear to the
handle. In the event of material fatigue such as
There is an identification mark on the upper side of cracking and/or discoloration, the handle must be
the handle which shows the year and month of replaced.
manufacture of the handle.
28426

12 1 2
9 1 11

18 5
34 5
0

67
The identification mark here shows January 2015
as an example of the date of manufacture.
The number in the inner circle corresponds to the
year of manufacture.
15 = 2015
The arrow points to the month of manufacture.
1 = January
The date of manufacture has no effect on the
service life of the handle.
The handle can undergo up to 300 reprocessing
cycles. After that, it must be disposed of.

NOTE
Cleaning methods based on alkaline agents (pH
values >10) can increase material wear to the
handle and thus shorten its service life.

Instructions for use Polaris 600 69


Reprocessing

Overview of reprocessing procedures

Classification of medical devices – Neodisher® MediClean by Dr. Weigert


For reprocessing, medical devices and their Machine disinfection (thermal):
components are classified by their type of – 93 °C (199.4 °F) for 10 minutes
application and the risk resulting from it:
Steam sterilization:
– Non-critical medical devices: surfaces
accessible to users and patients, e.g., surfaces – 132 °C (270 °F) for 4 minutes
of devices
Non-critical medical devices
– Critical medical devices: e.g., sterilizable
handles Manual cleaning and disinfection

Testing of procedures and agents Procedure:


1 Remove dirt immediately with a wipe soaked in
The cleaning, disinfection, and sterilization of
disinfectant.
medical devices has been tested using the
following procedures and agents. At the time of
testing, the following procedures and agents WARNING
showed good material compatibility and Risk of electric shock or device malfunction
effectiveness:
Liquid that enters into the device can cause
Non-critical medical devices the device to malfunction or may damage the
Manual disinfection with simultaneous cleaning: device and endanger the patient.
– Buraton® 10F by Schülke & Mayr Only scrub-and-wipe-disinfect device
surfaces and cables and make sure no liquids
Concentration: 1 %
penetrate into the device.
Contact time: 30 minutes
– Dismozon® pur by BODE Chemie 2 Perform surface disinfection (scrub-and-wipe
disinfection).
Concentration: 1.5 %
3 After the contact time has elapsed, remove
Contact time: 15 minutes disinfectant residues.
Critical medical devices 4 Wipe with a cloth moistened with water and
Manual reprocessing allow to dry.

Manual cleaning: 5 Check all items for damage and external signs
of wear, such as cracking, embrittlement, or
– Neodisher® LM 2 by Dr. Weigert pronounced hardening, and residual dirt.
Concentration: 2 % Replace parts if necessary.
Contact time: 10 minutes
CAUTION
Manual disinfection:
Risk of damage to the device
– Korsolex® extra by BODE Chemie
The device consists in part of materials that are
Concentration: 3 %
not resistant to certain components of surface
Contact time: 15 minutes disinfectants.
Machine reprocessing It is essential to follow the instructions in this
Machine cleaning: chapter.

– Neodisher® MediClean forte by Dr. Weigert

70 Instructions for use Polaris 600


Reprocessing

Cleaning and disinfection of the bottom glass 3 Check items for visible soiling and damage. If
– For every new patient necessary, repeat manual disinfection.
4 Thoroughly shake out residual water. Allow
items to dry thoroughly.

27176
Machine cleaning and disinfection
Perform machine cleaning and disinfection using a
washer-disinfector in accordance with EN
ISO 15883, preferably with a cart for anesthesia
accessories and ventilation accessories.
Procedure:
1 Observe instructions for use for the washer-
disinfector.
2 Securely position items in the basket. Make
sure that all interior spaces and surfaces are
completely flushed and that water can drain off
freely.
3 Use suitable cleaning agent.
1 Wipe the bottom glass with a soft, lint-free cloth 4 Select suitable program, preferably anesthesia
radially from the inside outwards. program.
Critical medical devices – Thermal disinfection must be carried out at
93 °C (199 °F) for 10 minutes.
Manual cleaning
5 Carry out final rinsing with deionized water.
Perform manual cleaning preferably under running
6 Immediately remove items from the washer-
water and with commercially available cleaning
disinfector.
agents (pH value ≥10).
7 Check items for visible soiling and damage. If
Procedure: necessary, repeat program or perform manual
1 Wash off superficial soiling under running cleaning and disinfection.
water. 8 Allow items to dry thoroughly.
2 Use cleaning agents in accordance with the
Visual inspection
manufacturer’s instructions. Make sure that all
surfaces and interior spaces which must be Check all items for damage and external signs of
cleaned are reached. If necessary, use suitable wear, such as cracking, embrittlement, or
brushes. pronounced hardening, and residual dirt.
3 Rinse items thoroughly under running water
until cleaning agent residues are no longer CAUTION
discernible.
Risk from faulty accessories
4 Check items for visible soiling and damage. If
necessary, repeat manual cleaning. Even reusable accessories have a limited
maximum period of use, e.g., residues from
Manual disinfection disinfectants can attack the material in the
autoclave. Signs of external wear can show up,
Procedure:
e.g., cracks, deformation, discoloration, or
1 Immerse items in disinfectant. delamination).
2 After the contact time has elapsed, rinse items If signs of external wear occur, replace the
thoroughly under running water until affected accessory.
disinfectant residues are no longer discernible.

Instructions for use Polaris 600 71


Reprocessing

Sterilization
WARNING
For sterilization, use a vacuum steam sterilizer
(conforming to ISO 17665), preferably with Risk of infection
fractionated vacuum. If the sterilizable handle exhibits cracks or
 Only sterilize cleaned and disinfected items. deformation as a result of frequent
sterilization, the handle may fall into the
– Sterilize with a fractionated vacuum steam
operation field and/or may not satisfy the
procedure at 132 °C (270 °F) for 4 minutes.
infection prevention policy.
– The sterilizable handle must not be subjected
to mechanical load during sterilization. Make sure that the handle is not used in the
Otherwise, it may be permanently deformed. event of such damage.
– The sterilizable handle must not be reattached
until immediately before an operation.

Materials used in the light

The following materials are used in the light: Components Material


Components Material Gimbal mounting Steel, powder
coating
Light head with housing for light Aluminum, pow-
control panel der coating Integrated handle Polyamide (PA)
Seals Silicone Light control panel Polyamide (PA),
polyester foil
Bottom glass Polycarbonate
(PC) Polaris 600 remote control for Aluminum, pow-
MedView camera der coating,
Sterilizable handle E Polyetherimide
polyester foil
(PEI)
Sterilizable handle Polyetherimide
(PEI)

Assembling the parts

Fitting the sterilizable handle E


– see "Fitting the sterilizable handle E" on page
32.
Fitting the sterilizable handle
– see "Fitting the sterilizable handle" on page 33.

72 Instructions for use Polaris 600


Maintenance

Maintenance

Overview

This chapter describes the maintenance measures


required to maintain the proper functioning of the WARNING
medical device. Maintenance measures must be Risk of electric shock
performed by the personnel responsible.
There are conducting components under the
WARNING housing cover.
– Do not remove the housing cover.
Risk of infection – Maintenance measures must be
performed by the personnel responsible.
The responsible personnel may be infected
Dräger recommends DrägerService for
by pathogenic germs.
repairs and complex maintenance tasks.
Disinfect and clean device or device parts
before any maintenance measures and also
before returning the medical device for repair.

Definitions of maintenance terms

Term Definition
Maintenance All measures (inspection, preventive maintenance, repair) intended to maintain
and restore the functional condition of a medical device
Inspection Measures intended to determine and assess the actual state of a medical device
Preventive mainte- Recurrent specified measures intended to maintain the functional condition of a
nance medical device
Repair Measures intended to restore the functional condition of a medical device after a
device malfunction

Inspection

Regular inspections must be carried out according


to the following specifications and at the stated
intervals. Technical documentation is available on
request.

WARNING
Risk of electric shock
Disconnect all electrical connections from the
power supplies before carrying out any work.

Instructions for use Polaris 600 73


Maintenance

Component Interval Task Personnel responsible


Complete light and arm Before each use Check for operational Users
system readiness, see "Checking
for operational readiness
before each use" on page
31.
Check accompanying doc-
uments: instructions for
use available.
Check equipment of the
medical device for com-
pleteness according to the
instructions for use.
Perform functional and
Complete light and arm visual inspection.
Every 12 months Service personnel
system Check for deformations of
the arm system and
whether parts have
become loose.
Check for paint damage
and cracks in plastic parts.
Check whether all plastic
parts are present and in
the correct position.
Check electrical safety of
the light system in accor-
dance with IEC 62353.
Replace components if
necessary.
Check free rotation and
stops of the arm system.
Complete light and arm Specialized service per-
Every 2 years Check the horizontal and sonnel
system
vertical joints of the arm
system for free movement
and regrease them if nec-
essary.
Check the height stop of
the arm system and read-
just if necessary.

74 Instructions for use Polaris 600


Maintenance

Component Interval Task Personnel responsible


Check the position and
form of the circlip of the
arm system.
Check the load distribution
and spring force of the
spring arm and readjust if
necessary.
Check the arm system
(central axis, swivel arm,
and spring arm) for colli-
sion damage (welding
Complete light and arm seams free of cracks). Specialized service per-
Every 2 years
system sonnel
Check for deformations of
the arm system and
whether parts have
become loose.
Check for paint damage
and cracks in plastic parts.
Check whether all plastic
parts are present and in
the correct position.
Perform functional and
visual inspection.

Instructions for use Polaris 600 75


Maintenance

Preventive maintenance

WARNING WARNING
Risk of faulty components Risk of electric shock
Device failure is possible due to wear or Disconnect all electrical connections from the
material fatigue of the components. power supplies before carrying out any work.
To maintain the function of all components,
this device must be inspected and serviced at
the intervals specified by the manufacturer.

Table of preventive maintenance intervals

Component Interval Task Personnel responsible


Locking segment of the Every 2 years Check the dimensions of Specialized service per-
spring arms the locking segment sonnel
(minimum material thick-
ness and minimum
length) and replace if
necessary.

Repair

It is recommended that only original parts from


Dräger are used and that the parts are replaced by
Dräger.

76 Instructions for use Polaris 600


Maintenance

Adjusting the arm system

Adjusting the spring arm

The spring arm must hold the mounted light in any Adjusting the spring force
height position.
The spring force of the spring arm must be
Adjusting the height stop adjusted if the light does not remain at the set
height.
Vertical movement of the light is limited by a fixed
lower and adjustable upper stop. The adjustable When adjusting the spring arm, lift it approx. 10o
upper stop (height stop) can be set to any position above the horizontal position. Only in this position
between the positions (A) and (B). The factory is it possible to turn the set screw.
setting for the spring arm is position (B).
If necessary, set the upper stop (height stop) to a
higher position.
27202

27203
A
B C O
10

1 Press down the seal inside the joint.


2 Adjust the height stop for the spring arm using 1 Raise the spring arm approx. 10o above the
the set screw (C). horizontal position.
Use an Allen key (size 5 mm) for this. 2 Press down the seal inside the joint.
 Turn the set screw counterclockwise to adjust 3 Adjust the spring force for the spring arm using
the stop upwards. the set screw (A).
Pay attention to the distance to the ceiling. Use an Allen key (size 5 mm) for this.
 Turn the set screw clockwise to adjust the stop  Turn the set screw clockwise to increase the
downwards. spring force.
 Turn the set screw counterclockwise to
decrease the spring force.

Instructions for use Polaris 600 77


Maintenance

– If the adjustment has been carried out  Turn the set screw clockwise to increase the
correctly, the light can be positioned at any braking force.
height and will remain there.  Turn the set screw counterclockwise to
decrease the braking force.

Adjusting the gimbal mounting NOTE

The gimbal mounting is set correctly when the light Set the braking force as low as possible and as
head is stable in any position and does not move high as necessary.
on its own.
Adjusting the braking force at the intermediate
joint
27204

27205
A

A
B

1 Adjust the braking force of the joint using the


set screw (A).
Use an Allen key (size 4 mm).
 To achieve this, the braking force can be 2 Adjust the braking force of the joint using the
adjusted on the joint (A) and on the set screw (A).
intermediate joint (B).  Turn the set screw clockwise to increase the
Adjusting the braking force at the joint braking force.
 Turn the set screw counterclockwise to
decrease the braking force.
28275

NOTE
A
Set the braking force as low as possible and as
high as necessary.

1 Adjust the braking force of the joint using the


set screw (A).
Use an Allen key (size 4 mm).

78 Instructions for use Polaris 600


Maintenance

Locking segment of the spring arm

The dimensions (minimum material thickness and


minimum length) of the locking segment must be
inspected every 2 years. If the dimensions are
lower than specified, the locking segment must be
replaced. The locking segment must be greased
before insertion.

27208
PP
0LQLPXPOHQJWK

0LQLPXPPDWHULDOWKLFNQHVV
PP
PP
0LQLPXPOHQJWK

0LQLPXPPDWHULDOWKLFNQHVV
PP

1 Remove the device from the spring arm.


2 Check the dimensions of the locking segment
(minimum material thickness and minimum
length) and replace if necessary.
3 Grease the locking segment and the gimbal
mounting journal.
4 Fit the device to the spring arm.

Instructions for use Polaris 600 79


Maintenance

Acceptance and handover

Polaris 600 light


Upon conclusion of the installation or maintenance
work, the system must be tested and accepted by
specialized service personnel.
Commissioning DrägerService is recommended.
These tests determine
– whether the safety requirements necessary to
protect the patient and personnel have been
met, and
– whether the entire functional range of the
systems is met.
The inspection results must be documented in
writing.
After acceptance, the operational system is
handed over to its owner along with the associated
documentation. A written record of the handover is
filed for reference.
The users are instructed on the use of the system.
Polaris 600 remote control for MedView camera
It may be necessary to change the room address if
MedView cameras are operated by remote control
in adjacent operating rooms which are only
separated by a pane of glass.
The room address of the remote control and of the
remote control interface can only be changed by
DrägerService.

80 Instructions for use Polaris 600


Disposal

Disposal

Disposing of the medical device

WARNING
For countries subject to the EU
Directive 2002/96/EC
Risk of infection
The device and its components must be This device is subject to EU Directive 2002/96/EC
disinfected and cleaned before disposal! (WEEE). In order to comply with its registration
according to this directive, this device may not be
This product must be disposed of in accordance disposed of at municipal collection points for waste
with national regulations. electrical and electronic equipment. Dräger has
authorized a company to collect and dispose of
this device. To initiate collection or for further
information, visit Dräger on the Internet at
www.draeger.com. Use the search function with
the keyword "WEEE" to find the relevant
information. If access to the Dräger website is not
possible, contact the local Dräger organization.

Disposing of non-rechargeable batteries

The medical device battery contains pollutant


WARNING substances.
Risk of explosion! Risk of chemical burn! Only applicable in the Federal Republic of
Do not throw the batteries into fire or try to Germany: In accordance with the Battery Act
open them by force. (BattG), the end user is obliged to return batteries
that contain pollutants to the place of purchase or
 Do not recharge the batteries. dispose of them in public recycling centers. The
battery contained in the device therefore must be
removed by specialist personnel before disposing
of the device. In other countries than the Federal
Republic of Germany observe the applicable
national laws and regulations for battery disposal.

Disposal of accessories

Dispose of the following articles in accordance with – Sterilizable handle


the hospital's hygiene regulations: – Dräger disposable handle
– Sterilizable handle E

Instructions for use Polaris 600 81


Technical data

Technical data

EMC declaration

General information accessories may result in increased


electromagnetic emissions or decreased
This device was tested for electromagnetic electromagnetic immunity of the device.
compatibility using accessories from the list of This device may be used in the direct vicinity of
accessories. Other accessories may only be used other devices only if Dräger has approved this
if they do not compromise the electromagnetic device arrangement. If no approval has been given
compatibility. The use of non-compliant by Dräger, it must be ensured that this device
functions correctly in the desired arrangement
before use. The instructions for use for the other
devices must be followed.

Electromagnetic environment

This device may only be used in environments


specified in section "Environments of use" on
page 16.

Emissions Compliance
Radiated emis- Class A, Group 1 (30 MHz to
sions 1 GHz)
Conducted emis- Class A, Group 1 (150 kHz to
sions 30 MHz)

CAUTION
The emissions characteristics of this equipment
make it suitable for use in industrial areas and
hospitals (CISPR 11, Class A). If it is used in a
residential environment (for which CISPR 11
Class B is normally required), this equipment
might not offer adequate protection to radio-
frequency communication services.
The user might need to take mitigation
measures, such as relocating or reorienting the
equipment.

82 Instructions for use Polaris 600


Technical data

Immunity against Test level and required electromagnetic environ-


ment
Electrostatic discharge (ESD) (IEC 61000-4-2) Contact discharge: ±8 kV
Air discharge: ±15 kV
Electrical fast transient disturbances (bursts) Power cable: ±2 kV
(IEC 61000-4-4)
Longer signal input lines/signal output lines: ±1 kV
Impulse voltages (surges) (IEC 61000-4-5) Voltage phase conductor – phase conductor: ±1 kV
Voltage phase conductor – protective ground conduc-
tor: ±2 kV
Magnetic fields at mains frequency 50 Hz: 30 A/m
(IEC 61000-4-8)
Voltage dips and short interruptions of the sup- Voltage dips of 30 % to 100 %, 8.3 ms to 5 s, different
ply voltage (IEC 61000-4-11) phase angles
Radiated high-frequency disturbances 80 MHz to 2.7 GHz: 3 V/m
(IEC 61000-4-3)
Conducted high-frequency disturbances 150 kHz to 80 MHz: 3 V, ISM bands: 6 V
(IEC 61000-4-6)
Electromagnetic fields in the vicinity of wireless Various frequencies from 385 MHz to 5785 MHz:
communication devices 9 V/m to 28 V/m

Recommended separation distances


from wireless communication devices

To ensure that the full functional integrity of this


device is not compromised, there must be a
separation distance of at least 1.0 m (3.3 ft)
between this device and wireless high-frequency
communication equipment.

Instructions for use Polaris 600 83


Technical data

Ambient conditions

During operation
Temperature For the light system:
10 °C to 40 °C
(50 °F to 104 °F)
For the radio receiver in the display mount:
10 °C to 30 °C
(50 °F to 86 °F)
Ambient pressure 700 hPa to 1060 hPa
(10.2 psi to 15.4 psi)
Relative humidity 30 % to 75 % (non-condensing)
1)
During storage and transport (up to 15 weeks)
Temperature –20 °C to 60 °C
(–4 °F to 140 °F)
Ambient pressure 700 hPa to 1060 hPa
(10.2 psi to 15.4 psi)
Relative humidity 10 % to 95 % (non-condensing)
1) Storage only indoors or in covered spaces

Technical light data

Polaris 600 Polaris 600 with anti-


reflective bottom
glass
OR light mode
Maximum central illuminance Ec1) 160000 lx 120000 lx
Central remaining illuminance (% of Ec)
with tube 100 % 100 %
with 1 mask 45 % 45 %
with 2 masks 45 % 45 %
with 1 mask and tube 45 % 45 %
with 2 masks and tube 45 % 45 %
Light field diameter at a distance of 1 m
(39.37 in)
d10 190 mm (7.48 in) 240 mm (9.45 in)
d50 120 mm (7.72 in) 135 mm (5.31 in)
Depth of illumination L1+L2 (20 %) 1300 mm (51.18 in) 1300 mm (51.18 in)

84 Instructions for use Polaris 600


Technical data

Technical light data (continued)

Depth of illumination L1+L2 (60 %) 700 mm (27.56 in) 700 mm (27.56 in)
Color rendering index Ra (at 3800 K) 95 95
Color rendering index R9 (at 3800 K) 94 94
Color rendering index R13 (at 3800 K) 93 93
Recommended working area (distance from 90 cm to 120 cm 90 cm to 120 cm
the bottom glass of the light to the operating (35.43 in to 47.24 in) (35.43 in to 47.24 in)
field)
Central irradiance Ee2) 600 W/m2 490 W/m2
Ee/Ec ratio 3.6 (mW/m2) / lx 3.8 (mW/m2) / lx
Setting the illuminance 12 steps, 20000 lx to 12 steps, 20000 lx to
160000 lx, 12.5 % to 120000 lx, 16.7 % to
100 % (2 % Endo light) 100 % (2 % Endo light)
Minimum illuminance 20000 lx 20000 lx
Setting the diameter of the light field 3 steps, 190 mm / 3 steps, 240 mm /
230 mm / 280 mm 280 mm / 320 mm
(7.48 in / 9.06 in / (9.45 in / 11.02 in /
11.02 in) 12.60 in)
Setting the color temperature 4 steps, 3800 K / 4 steps, 3800 K /
4400 K / 5000 K / 4400 K / 5000 K /
5600 K 5600 K
Bulbs LEDs LEDs
Number of LEDs 92 92
Number of LED modules 18 18
Service life of the LED bulbs 50000 hours 50000 hours
Replacement of LED bulbs possible possible

Ambient light mode (Endo light)


Illuminance (1 m (39.37 in) distance to light 3000 lx (corresponds 2000 lx (corresponds
emission surface) to 2 % of the maximum to 2 % of the maximum
central illuminance) central illuminance)
1) Light settings (1 m (39.37 in) distance to light-emitting surface): Central color temperature set to 4400 K and light field
diameter set to "Small".
2) The maximum total irradiance is attained at a distance of 91 cm (35.83 in).

Instructions for use Polaris 600 85


Technical data

Electrical data

Power supply with 100 V to 240 V mains connection


component
Mains power 100 V to 240 V (AC) / 50 Hz/60 Hz
I max= 2.4 A
P max = 240 VA
Emergency power 24 V (AC) / 50 Hz/60 Hz
I max= 7.9 A
P max = 190 VA
24 V (DC)
I max= 5.8 A
P max = 140 W
Output voltage ≤ 32 V (DC)
I max= 5 A
P max = 120 W
Power supply with 24 V mains connection compo-
nent
Mains power 24 V (AC) / 50 Hz/60 Hz
I max= 7.9 A
P max = 190 VA
Emergency power 24 V (DC)
I max= 5.8 A
P max = 140 W
Output voltage ≤ 32 V (DC)
I max= 5 A
P max = 120 W
Typical power consumption (incl. mains connection
component) at 100 V to 240 V (AC) / 50 Hz/60 Hz
Light at 100 % illuminance 115 VA
Light at 50 % illuminance 70 VA
Light in standby mode 35 VA
Maximum connected load
Light head Max. 240 VA
Display spring arm Max. 800 VA
Total rated power of all LED bulbs 70 W
Power supply for the light head 24 V

86 Instructions for use Polaris 600


Technical data

Mechanical data

Diameter of light head 620 mm (24.41 in)


Height of light head
Light head 90 mm (3.54 in)
Light head with sterilizable handle E 245 mm (9.66 in)
Light head with gimbal mounting and sterilizable 650 mm (25.59 in)
handle E
Weight of light head (incl. gimbal mounting, Med- 15.5 kg (34.17 lb)
View camera, and sterilizable handle E)
Weight of light head (incl. gimbal mounting and 15.5 kg (34.17 lb)
sterilizable handle E)
Brakes for gimbal mounting Friction brakes

All details are subject to manufacturing tolerances.

MedView camera

Camera retrofittable in light head Yes


Video image transmission by means of radio WHDI
standard
Encrypted video transmission 128 bit AES
Image sensor 1/2.8" CMOS
Resolution Full HD (1920 x 1080 pixels)
Zoom 120x (10x optical / 12x digital)
Objective width F 3.8 mm (Wide) to 38 mm (Tele),
1.8 to 3.4
Auto shutter Yes
Minimum illuminance 100 lx
Signal/Noise ratio 50 dB
White balance Automatic adjustment to the color temperature
of the Polaris 600 light
Freeze frame Yes
Focus Autofocus (AF)
Manual focus
Power supply 12 V, from the light head
Power consumption <15 W
Optimum distance of the camera to the object 80 cm to 130 cm (31.50 in to 51.18 in)

Instructions for use Polaris 600 87


Technical data

MedView camera (continued)

Dimensions L x W x H 78 mm x 46 mm x 49 mm (3.07 in x 1.81 in x


1.93 in)
Weight <250 g (0.55 lb)

All details are subject to manufacturing tolerances.

Polaris 600 remote control for MedView camera

Signal transmission Infrared radiation


Range Max. 7 m (275.60 in)
Wavelength of infrared radiation 950 nm
Radiated power 18 mW
Power supply 9 V, from battery
Size of battery 9-volt battery
Battery type 6LR61, 6F22, or 1604D
Dimensions L x W x H 170 mm x 60 mm x 35 mm
(6.69 in x 2.36 in x 1.38 in)
Weight 300 g (0.66 lb)
Holder for remote control Mounted on horizontal standard rail conforming
to ISO 19054

All details are subject to manufacturing tolerances.

Wireless video transmission

Range 10 m (393.70 in)


Modulation OFDM/QAM
Encryption 128 bit AES
Radio standard WHDI
Frequency band 5.180 GHz to 5.825 GHz, DFS
Transmitter power Max. 21 dBm

Radio receiver

88 Instructions for use Polaris 600


Technical data

Wireless video transmission (continued)

Operating frequency (European Union) U-NII band @ 5 GHz (5.180 GHz to 5.320 GHz
and 5.500 GHz to 5.825 GHz)
Transmitter power & modulation TPC (transmit-
ter power control) limited to 18 dBm EIRP /
16 dBm ERP max.
Transmitter modulation OFDM/QAM
Video outputs on each radio receiver 2x HD-SDI (with HDMI/HD-SDI converter)
1x DVI-D (with HDMI to DVI-D cable)
Maximum number of video receivers per Med- 4
View camera
Power supply 12 V, with associated 100 V to 240 V (AC) /
50 Hz/60 Hz power supply unit
Mains plug type Country-specific type supplied with device
Dimensions L x W x H 146 mm x 117 mm x 28 mm
(5.75 in x 4.60 in x 1.10 in) (without power plug)
146 mm x 130 mm x 83 mm
(5.75 in x 5.12 in x 3.27 in) (with power plug)
160 mm x 140 mm x 83 mm
(6.30 in x 5.51 in x 3.27 in) (with radio receiver
housing (wall or ceiling))
Weight including power supply unit 250 g (0.55 lb)
Surface mounting possible
Mounting in the display mount possible
Ceiling mounting possible

Radio transmitter
Operating frequency (European Union) U-NII band @ 5 GHz (5.180 GHz to 5.320 GHz
and 5.500 GHz to 5.825 GHz)
Bandwidth for each modulation/frequency
40 MHz

Signal converter (from HDMI to HD-SDI)


Power supply 12 V, with associated 100 V to 240 V (AC) /
50 Hz/60 Hz power supply unit
Mains plug type Country-specific type supplied with device
Dimensions L x W x H 115 mm x 93 mm x 23 mm
(4.53 in x 3.66 in x 0.91 in) (without power plug)
130 mm x 93 mm x 23 mm
(5.12 in x 3.66 in x 0.91 in) (with power plug)
Weight including power supply unit 300 g (0.66 lb)

Instructions for use Polaris 600 89


Technical data

Wireless video transmission (continued)

Surface mounting possible


Ceiling mounting possible

All details are subject to manufacturing tolerances.

Combination with system components

Polaris 100/200 surgical light possible


MedView camera with Polaris 600 light possible
Attaching a display mount possible
Display mount (24 inch to 32 inch) possible
Single display mount (19 inch to 24 inch) possible
Double display mount (19 inch to 24 inch) possible

Operating options

Sterilizable handle E (to set illuminance and light field diameter) Sterile
Sterilizable handle Sterile
Dräger disposable handle Sterile
Disposable sleeve for handle adapter (e.g., manufactured by Covidien) Sterile
Integrated handle (at the light head) Non-sterile
Light control panel (on the gimbal mounting) Non-sterile
Wall-mounted control panel Non-sterile
Polaris 600 remote control for MedView camera Non-sterile

Classifications

Protection class according to IEC 60601-1


Mains connection component Protection class 1
Light head Touchable secondary circuit according to
IEC 60601-1 (3rd edition)
Observed standards
IEC 60601-1 Medical electrical equipment – Part 1-1: Gen-
eral requirements for safety; Collateral stan-
dard: Safety requirements for medical
electrical systems

90 Instructions for use Polaris 600


Technical data

Classifications (continued)

IEC 60601-2-41 Medical electrical equipment – Part 2-41:


Particular requirements for basic safety and
essential performance of surgical luminaires
and luminaires for diagnosis
IEC 60601-1-2 Medical electrical equipment – Part 1-2: Gen-
eral requirements for basic safety and essen-
tial performance - Collateral standard:
Electromagnetic compatibility - requirements
and tests
IEC 60364-7-710/VDE 0100-710 Low-voltage electrical installations - Part 710:
Requirements for special installations or
locations - Medical locations

European medical device classification for the Class I


Polaris 600 light system
UMDNS code for the Polaris 600 light system 12-282

European medical device classification for the Class Is


Dräger Polaris disposable handle
UMDNS code for the Dräger Polaris disposable 17-977
handle
Universal Medical Device Nomenclature System
(nomenclature for medical devices)

Instructions for use Polaris 600 91


List of accessories

List of accessories

List of accessories for Polaris light and MedView camera

WARNING
Risk of patient injury
Device malfunctions may occur if
incompatible accessories are used with the
light system.
Only accessories from this list of accessories
may be used on the light system. Follow
these instructions for use when fitting the
accessories.

Designation Order no.


For all light versions
Sterilizable handle E (set of 2 handles) G36993
Sterilizable handle (set of 2 handles) G92099
Dräger Polaris disposable handle1) (optional), Solo (set of 120 packages, each con- MP05701
taining 1 piece)
Dräger Polaris disposable handle1) (optional), Duo (set of 60 packages, each contain- MP05700
ing 2 pieces)
Handle adapter2)3) (optional) for disposable sleeves for lights, manufacturer Covidien 31146895
For MedView camera
Polaris 600 remote control for MedView camera G36182
1) The Dräger Polaris disposable handle is expected to be available in selected countries from the beginning of 2019.
2) Not covered by IEC 60601-2-41 standard.
3) If the handle adapter for disposable sleeves is used, the corresponding disposable sleeves must be ordered directly from
Covidien.

92 Instructions for use Polaris 600


Index

Index

A I
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 Illuminance . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Image illumination (MedView camera) . . . . . . . 54
Ambient light. . . . . . . . . . . . . . . . . . . . . . . . . 16, 51 Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Autofocus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55 Integrated handle . . . . . . . . . . . . . . . . . 18, 20, 20
Automatic exposure . . . . . . . . . . . . . . . . . . . . . . 53 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

B L
Battery replacement (Polaris 600 remote control) LED indicator. . . . . . . . . . . . . . . . . . . . . . . . . . . 43
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Light control area. . . . . . . . . . . . . . . . . . . . . . . . 23
Braking force . . . . . . . . . . . . . . . . . . . . . . . . . . . 78 Light control panel . . . . . . . . . . . . . . . . . 20, 20, 21
Light field diameter . . . . . . . . . . . . . . . . . . . . . . 49
C Light mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
List of accessories. . . . . . . . . . . . . . . . . . . . . . . 92
Camera control area. . . . . . . . . . . . . . . . . . . . . . 23
Color temperature. . . . . . . . . . . . . . . . . . . . . 16, 48 M
Counterweight . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Mains power . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
D Manual exposure. . . . . . . . . . . . . . . . . . . . . . . . 54
Manual focus . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Device combinations . . . . . . . . . . . . . . . . . . . . . . 9 MedView camera. . . . . . . . . . . . . . . . . . . . . 20, 22
Disposable handle . . . . . . . . . . . . . . . . . . . . . . . 33
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81 O
Drifting of swivel arms . . . . . . . . . . . . . . . . . . . . 37
Operating time (Polaris 600 remote control) . . . 58
E OR light . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16, 40
OR panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Electromagnetic compatibility . . . . . . . . . . . . . . . 10
EMC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 P
EMC declaration . . . . . . . . . . . . . . . . . . . . . . . . . 82
Emergency power. . . . . . . . . . . . . . . . . . . . . . . . 39 Patient safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Endo light . . . . . . . . . . . . . . . . . . . . . . . . . . . 16, 51 Polaris 600 control interface . . . . . . . . . . . . . . . 24
Polaris 600 remote control . . . . . . . . . . . . . . . . 22
F Positioning a light, non-sterile . . . . . . . . . . . . . . 36
Positioning a light, sterile. . . . . . . . . . . . . . . . . . 35
Fault – Cause – Remedy . . . . . . . . . . . . . . . . . . 60 Positioning of a light with short gimbal mounting,
Freeze frame . . . . . . . . . . . . . . . . . . . . . . . . . . . 56 non-sterile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Positioning of a light with short gimbal mounting,
G sterile. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Preventive maintenance . . . . . . . . . . . . . . . . . . 76
Gimbal mounting . . . . . . . . . . . . . . . . . . . . . . . . 78 Preventive maintenance intervals . . . . . . . . . . . 76

H R
Handle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Height stop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77 Reprocessing personnel . . . . . . . . . . . . . . . . . . . 6

Instructions for use Polaris 600 93


Index

S
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Service personnel. . . . . . . . . . . . . . . . . . . . . . . . . 7
Specialized . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Spring arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Standby mode . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Status LED (light) . . . . . . . . . . . . . . . . . . . . . . . . 39
Status LED (light system). . . . . . . . . . . . . . . . . . 39
Sterilizable handle E. . . . . . . . . . . . . . . . . . . . . . 28
Switching the light off . . . . . . . . . . . . . . . . . . . . . 40
Switching the light on . . . . . . . . . . . . . . . . . . . . . 40
Switching the MedView camera off . . . . . . . . . . 52
Switching the MedView camera on . . . . . . . . . . 52
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

T
Target group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Total irradiance, maximum . . . . . . . . . . . . . . . . . 45
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . 60

U
Users . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

W
Wall-mounted control panel . . . . . . . . . . . . . 18, 23
White balance (automatic) . . . . . . . . . . . . . . . . . 56

Z
Zoom . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53

94 Instructions for use Polaris 600


This page has been left blank intentionally.

Instructions for use Polaris 600 95


These instructions for use apply only to
Polaris 600
with Serial No.:
If no Serial No. has been filled in by Dräger,
these instructions for use are provided for
general information only and are not intended
for use with any specific device.
These instructions for use are provided for
customer information only and will only be
updated or exchanged upon customer request.

MD MD
(Polaris 600 light system) (Dräger Polaris disposable handle)

(Radio transmitter)

Manufacturer

Drägerwerk AG & Co. KGaA


Moislinger Allee 53 – 55
23542 Lübeck
Germany
+49 451 8 82-0
FAX +49 451 8 82-2080
http://www.draeger.com

9054591 – GA 6932.155 en

Á9054591È
© Drägerwerk AG & Co. KGaA
Edition: 7 – 2019-10
(Edition: 1 – 2014-08)
Dräger reserves the right to make modifications
to the device without prior notice.
ision]_0000047853_No.0411

You might also like