CALIBRATION CERTIFICATE

Date of Issue Calibration Date Request No Certificate No Due Date


28-07-2021 28-07-2021 RQ 1056 QMC/1056-01/2021 27-07-2022
Customer Name & Address
One Branch National Cateering Company
DIP 1, Dubai, United Arab Emirates
Unit Under Calibration
Equipment Name : Probe Thermometer Serial No : 1056-PT-01
Make :EXTECH Range : -40 to 200 ºC
Model : 39240 Resolution : 0.1ºC

Environmental Condition
Temperature : 25±5ºC Humidity : 50±15% RH
Standard Used for Calibration
: Infrared Calibrator, Liquid Bath,
Equipment Name Certificate No : CU-600002543/ 22871
Reference Thermometer
Make : DELTA OHM, CEM, ISOTECH Calibration Procedure : QMC/WI/18
Traceability
All Measuring equipments used for calibration purposes is traceable back to international/national Standards
through UKAS or Other recognized bodies which provides traceability of measurement to national physical
laboratory or other recognized standards laboratory
Uncertainty Statement
The uncertainty reported is based on a standard uncertainty multiplied by a coverage factor k=2, providing a level
of confidence of approximately 95%.
Results Obtained
Reference Value (Master)(⁰C) Indicated Value (UUC)(⁰C) Deviation (⁰C)
-20 -19.6 0.399999999999999
30 28.9 -1.1
60 58.7 -1.3
100 98.7 -1.3
150 148.9 -1.09999999999999
200 198.7 -1.30000000000001
Maximum Deviation: -1.3⁰C

Calibrated by : Nithin Approved by : Ajeesh

QMC/CF/01 Rev. 1.0 Issue Date:10/05/2017 Page 1 of 1


QMC is not liable for any damage caused by or resulting from improper or negligent use. Above results only on the specified object and conditions.
This Certificate may not be reproduced other than in full, except with the prior written permission of Laboratory.

QMC/CF/01 Rev. 1.0 Issue Date:10/05/2017 Page 1 of 1


CALIBRATION CERTIFICATE
Date of Issue Calibration Date Request No Certificate No Due Date
28-07-2021 28-07-2021 RQ 1056 QMC/1056-02/2021 27-07-2022
Customer Name & Address
One Branch National Cateering Company
DIP 1, Dubai, United Arab Emirates
Unit Under Calibration
Equipment Name : Probe Thermometer Serial No : 1056-PT-02
Make : ETI Range : -40 to 200 ºC
Model : 810-275 Resolution : 0.1ºC

Environmental Condition
Temperature : 25±5ºC Humidity : 50±15% RH
Standard Used for Calibration
: Infrared Calibrator, Liquid Bath,
Equipment Name Certificate No : CU-600002543/ 22871
Reference Thermometer
Make : DELTA OHM, CEM, ISOTECH Calibration Procedure : QMC/WI/18
Traceability
All Measuring equipments used for calibration purposes is traceable back to international/national Standards
through UKAS or Other recognized bodies which provides traceability of measurement to national physical
laboratory or other recognized standards laboratory
Uncertainty Statement
The uncertainty reported is based on a standard uncertainty multiplied by a coverage factor k=2, providing a level
of confidence of approximately 95%.
Results Obtained
Reference Value (Master)(⁰C) Indicated Value (UUC)(⁰C) Deviation (⁰C)
-20 -19.4 0.600000000000002
30 29.1 -0.899999999999999
60 59.2 -0.799999999999997
100 98.9 -1.09999999999999
150 149.4 -0.599999999999994
200 199.2 -0.800000000000011
Maximum Deviation: -1.1⁰C

Calibrated by : Nithin Approved by : Ajeesh

QMC/CF/01 Rev. 1.0 Issue Date:10/05/2017 Page 1 of 1


QMC is not liable for any damage caused by or resulting from improper or negligent use. Above results only on the specified object and conditions.
This Certificate may not be reproduced other than in full, except with the prior written permission of Laboratory.