Download as pdf or txt
Download as pdf or txt
You are on page 1of 9

Effect of Denture Cleansers on Cobalt-Chromium Alloy

Surface: A Simulated Period of 5 Years’ Use


Patrı́cia Almeida Curylofo, DDS, MSc, Priscilla Neves Raile, DDS, MSc,
Glenda Lara Lopes Vasconcellos, DDS, MSc, PhD, Ana Paula Macedo, BSEE, MSc, PhD, &
Valéria Oliveira Pagnano, DDS, MSc, PhD
Department of Dental Materials and Prosthetics, Dental School of Ribeirão Preto, University of São Paulo (USP), Ribeirão Preto, SP, Brazil

Keywords Abstract
Confocal laser scanning microscopy;
removable partial denture; scanning electron
Purpose: To compare the effect of solutions of effervescent tablets (ET), cetylpyri-
microscopy; surface roughness. dinium chloride (CPC), and experimental solutions of Ricinus communis on the surface
of cobalt-chromium (Co-Cr) alloys.
Correspondence Materials and Methods: Fifty-five specimens of Co-Cr were prepared by the lost-
Dr Valéria Oliveira Pagnano, Department of wax casting method using circular patterns (࢝12 × 3 mm). The specimens were
Dental Materials and Prosthetics, Dental randomly divided into 5 groups: deionized water (control); 2% R. communis; 10%
School of Ribeirão Preto, University of São R. communis; ET, and CPC. The surface roughness of specimens (n = 10) was
Paulo, Avenida do Café, s/n, Monte Alegre, evaluated before immersion (baseline), and at simulated times of ½, 1, 2, 3, 4, and
14040–904, Ribeirão Preto, SP, Brazil. 5 years, by laser confocal microscope (Sa, μm) and profilometer (Ra, μm). The
E-mail: valpag@forp.usp.br. surface topography and chemical composition (n = 1) was qualitatively analyzed
with scanning electron microscopy (SEM), and energy dispersive X-ray spectrometry
This study was funded in part by the (EDS). Data were subjected to Kruskal–Wallis followed by Dunn tests, and Friedman
Coordenação de Aperfeiçoamento de Pessoal followed by Wilcoxon tests (α = 0.05).
de Nı́vel Superior - Brasil (CAPES) - Finance
Results: For Sa, there was no difference for the solution factor. For the time factor a
Code 001.
significant difference was found with 2% R. communis solution among baseline and
The authors deny any conflicts of interest in ½, 2, 3, and 5 years (p < 0.001) and with 10% R. communis solution between 1 and
regards to the present study. 2 years (p = 0.007), with decreasing roughness over time. For Ra, cetylpyridinium
chloride exhibited less roughness than 10% R. communis solution in ½ (p = 0.048)
Accepted October 11, 2018 and 5 years (p = 0.013). In the SEM and EDS analysis the solutions did not present
deleterious effects or changes in the chemical composition on the surfaces.
doi: 10.1111/jopr.12996 Conclusions: Although a significant difference was found for the roughness, the
results, below 0.2 μm, are clinically acceptable. Thus, all solutions can be used safely
in removable partial denture cleaning for a period of 5 years.

Removable partial dentures (RPDs) have a complex structure is easy to manipulate and allows access to the more retentive
and are made up of different types of materials, which tends areas that accumulate biofilm.5
to make them difficult to clean.1 Acrylic resin, although sus- Chemical solutions have proven antimicrobial action and are
ceptible to surface changes,2 has been extensively studied with effective in the removal of biofilm,7-9 having special importance
various cleaning methods; however, in relation to the metal in the remission of denture stomatitis.10 However, it has been
surface, the literature is scarce, and there is still no consensus reported that some cleansers promote adverse effects on the
regarding the most advisable cleanser.3 Thus, there is no clean- metallic surface, making them impracticable for use in RPD
ing protocol that can be proven safe and indicated for RPDs cleaning, since compatibility with the constituent materials of
that are widely used by the population. the denture is an important requirement for denture cleansers.11
The mechanical method of brushing, although the most com- Effervescent cleansing tablets are commonly indicated and,
mon and accessible, may not be as effective for the unmotivated their use is often recommended in the literature.1,12 There are
elderly, or for people who are physically incapable of promoting reports that they do not provide deleterious effects in relation
adequate cleaning of their prostheses, due to either manual or to the hardness and surface roughness of the metal alloys,11,13
visual limitations.4 Thus, chemical immersion is often used to besides being effective in the removal of biofilm.14 However,
accompany or replace the mechanical method of cleaning, pro- although these compounds are indicated by manufacturers for
viding satisfactory results.5,6 In addition, chemical immersion denture cleaning, undesirable effects such as oxidation of the

142 Journal of Prosthodontics 29 (2020) 142–150 


C 2018 by the American College of Prosthodontists
Curylofo et al Effect of Denture Cleansers on Co-Cr

Table 1 Experimental groups

Commercial name Total immersion time


(Group) Manufacturer Composition Immersion time to simulate 5 years

Deionized water NA H2 O 20 minutes daily2,15,16 609 hours


(H2 O/Control)
Ricinus communis Instituto de Quı́mica R. Communis oil (Diluted in water) 20 minutes 609 hours
(2% Rc/ 10% Rc) EESC-USP, São daily9,20,21,23
Carlos, Brazil
Cepacol (CPC) Sanofi-Aventis CPC, disodium EDTA, sodium 10 minutes daily2,15,16 304 hours
Farmacêutica Ltda. saccharin, sodium phosphate,
Suzano, Brazil disodium phosphate, eucalyptol,
menthol, methyl salicylate, flavoring,
perfume, Cl 19140 (tartrazine),
alcohol, and water (without dilution)
NitrAdine (ET) Bonyf AG Vaduz, Citric acid, sodium lauryl sulfate, 15 minutes 2x per 130 hours
Liechtenstein lactose monohydrate, sodium week
bicarbonate, sodium chloride, (recommended by
potassium monopersulfate, sodium manufacturer)
carbonate, mint flavoring, PVP
(diluted in water)

metal surface15 and release of ions11,16 have been found. In Materials and methods
addition, they have a relatively high cost, which makes ac-
cess difficult for some. NitrAdine effervescent tablet has been Table 1 presents the experimental groups used in the study
reported as having effective antimicrobial action against Can- and their respective immersion times. Fifty-five wax patterns
dida albicans,14,17 unlike other tablets.17 In addition, there are were obtained (GEO; Renfert GmbH, Hilzingen, Germany) in
reports that it does not cause adverse effects on the surface of disk shape (࢝12 × 3 mm) from a metal mold. They were
the acrylic resin.2,15,18 incrusted and used to cast the Co-Cr alloy (64.8% Co, 28.5%
Mouthwashes are also among the products most used by the Cr, 5.3% Mo, 0.5% Si, 0.5% Mn, 0.4% C - DeguDent; Dentsply
population for cleaning dentures. Among them is cetylpyri- Ind. e Com. Ltda., Petropolis, Brazil) by lost-wax technique as
dinium chloride (CPC), which has been reported as effective described by Bezzon et al.26
against C. albicans in complete dentures immersed for a pe- The metal discs were polished with 220-, 400-, 600-, and
riod of 8 hours,8 besides not causing change of roughness in 1200-grit sandpaper (Norton Abrasivos, Guarulhos, Brazil)
cobalt-chromium (Co-Cr) alloy for 6 months of use with daily in a polishing machine (Aropol-E; Arotec, Wellington, New
immersions of 10 minutes.15 However, because mouthwashes Zealand), and by a lathe (Kota, Cotia, Brazil) with diamond
are not specific for denture immersion, adequate guidelines for polishing paste with granulation of 2 to 3 μm (Saphir; Renfert
this use are not provided, and they may cause damage to the GmbH). A homogeneous surface was obtained with an initial
components of a prosthesis when used for a long period (RPD surface roughness between 0.04 and 0.10 μm. All metal discs
lifetime has been reported to be 5 years).1,19 were cleaned by ultrasound (Ultra Cleaner USC 750; Unique,
For complete dentures, castor oil (Ricinus communis) has São Paulo, Brazil) with isopropyl alcohol for 10 minutes, for
been evaluated as a cleanser7,9,11,20-22 due to its antifungal and surface washing and removal of any residue.
antibacterial characteristics, confirmed by studies related to The metal disks were randomly distributed in 5 experimental
endodontics.23 According to Pisani et al,21 R. communis can groups (n = 11) according to the cleanser solutions used (Fig 1).
be seen as an alternative, inexpensive, and safe substance to be Initial surface topography analysis and qualitative evaluation of
used by the population. Although there are studies related to the surface composition of the Co-Cr alloy was performed on
the proven antimicrobial action of castor oil7,9,24 and its adverse one specimen of each group by scanning electron microscopy
effects on acrylic resin,20,25 its action on dental alloys and the (SEM) and energy dispersive X-ray spectroscopy (EDS)11 and
best concentration to be used have not yet been evaluated. measurement of the surface roughness values (n = 10). The
Thus, the aim of this study was to evaluate the effect of specimens were then placed in plastic containers with 150 mL
the experimental cleaning aqueous solution of R. communis in of cleanser solution. All solutions were maintained at room
different concentrations (2%, 10%), of the effervescent tablet temperature, except the solution with the effervescent tablet in
(NitrAdine), and of CPC on the surface of Co-Cr alloy, for the which the distilled water was heated to 37°C,13 according to
simulated period of 5 years. The null hypothesis was that there the manufacturer’s instructions, in an oven (DeLeo Laboratory
would be no clinically relevant alteration of the Co-Cr alloy Equipment, Porto Alegre, Brazil). The solution was then kept
surface due to the use of different cleaning solutions tested in the oven to maintain this temperature during the 8 hours of
during a simulated period of 5 years’ use. action of the solution, after which the solution was exchanged.27

Journal of Prosthodontics 29 (2020) 142–150 


C 2018 by the American College of Prosthodontists 143
Effect of Denture Cleansers on Co-Cr Curylofo et al

Figure 1 Flowchart of experimental design.

focal microscope (LEXT OLS4000; Olympus, Tokyo, Japan),


which allowed high-resolution 3D imaging of the surfaces of
the metallic specimens (n = 10). Measurements were taken
1500 μm from the reference marking at the lateral edge of the
specimen.

Profilometer
To evaluate the roughness by profilometer (Surftest SJ-201P;
Mitutoyo Corp., Kawasaki, Japan), 3 readings were obtained on
the polished surface of each specimen to obtain an arithmetic
mean of the profile roughness deviations (Ra, μm). The surface
analyzer was calibrated at a 0.8-mm sample length, 4.0-mm
Figure 2 Matrix and positioning of the specimen. percussion of measure, and 0.5-mm/sec speed.

Considering that over the 5-year lifetime of the RPD, it would


Surface topography and chemical composition
be subject to 1825 daily immersions,19 this experiment was
performed with consecutive immersions respecting the daily One specimen from each experimental group was analyzed
immersion time of each group (Table 1). After each immersion by SEM (EVO 50; CARL ZEISS, Oberkochen, Germany) for
period to be evaluated, the specimens were washed in distilled qualitative evaluation of the surface in relation to changes and
water for 10 seconds. oxidation of the metallic surface, and with EDS in relation to
the chemical composition of areas of interest (500 Digital Pro-
Surface roughness cessing; IXRF Systems, Austin, TX). Analyses were conducted
at a magnification of 300x in the spot mode (EDS2004 Version
The surface roughness analyses, both by laser confocal micro-
1.3 Software User´s Guide - IXRF Systems). The direct analy-
scope and by profilometer, were performed before the start of
sis of the light elements was performed taking into account the
the immersions and at periods corresponding to ½, 1, 2, 3, 4,
flatness and cleanness of the specimens, and also the proces-
and 5 years. A Teflon matrix was used (Precision workshop;
sor characteristic (digital pulse). The ZAF correction factors
Dental Materials and Prosthesis Department of FORP-USP)
were used for element quantification.28 The acronym ‘ZAF’
(Fig 2) to standardize the positioning of the specimen. The
describes a procedure in which corrections for atomic number
Teflon matrix had a position index that, at the time of analysis,
effects (Z), absorption (A), and fluorescence (F) are calculated
coincided with the lateral marking of the specimen, such that it
separately from suitable physical models.
was always analyzed in the same position, making it possible
to divide the area of the specimen by means of its markings,
so that the confocal microscope readings were not impaired by Statistical analysis
the profilometer analyses.
The data were analyzed with statistical software (IBM SPSS 20;
SPSS Inc, Chicago, IL). The data showed a non-normal distribu-
Confocal microscopy
tion (Shapiro-Wilk test), and were subjected to Kruskal–Wallis
The average surface roughness (Sa, μm) of the polished sur- followed by Dunn tests to compare solutions, and Friedman
face of the Co-Cr specimens was measured by microscopic followed by Wilcoxon tests to compare times. Multiple com-
analysis of the specimens (5× objective lens) with a laser con- parisons were adjusted by Bonferroni correction (α = 0.05).

144 Journal of Prosthodontics 29 (2020) 142–150 


C 2018 by the American College of Prosthodontists
Curylofo et al Effect of Denture Cleansers on Co-Cr

integrity of the metal surfaces was maintained after the 5-year


immersion simulations in the different solutions used.
Table 2 shows the median and confidence interval (μm) and
results of the statistical analysis of surface roughness obtained
by confocal microscopy (Sa) before and after the immersion
simulations. Comparing the solutions at each time, there was
no statistically significant difference between any solution. In
the comparison of the times, there was a statistical difference
in surface roughness (Sa – median [confidence interval]) in the
2% R. communis solution from baseline (0.66 [0.59; 0.73]) to ½
(0.53 [0.46; 0.63]), 2 (0.55 [0.46; 0.63]), 3 (0.50 [0.45; 0.55]),
and 5 years (0.54 [0.47; 0.61]) (p < 0.001), with a tendency
to decrease over time. In the 10% R. communis solution there
was a significant decrease in the roughness from 1 (0.53 [0.47;
0.69)] to 2 years (0.50 [0.44; 0.65]) (p = 0.007); the other
times presented intermediate values. For CPC, a significant dif-
ference (p = 0.03) was observed among the times but when the
post-test was used, with adjustment for the number of pairwise
comparisons, no significant difference was found for surface
roughness (Sa).
Table 3 presents the results of the statistical analysis of sur-
face roughness obtained with the surface profilometer (Ra),
before and after the immersion simulation corresponding to
5 years. Comparing the solutions at each time, there was a
significant difference at baseline (p = 0.028); however, after
adjusting for the number of pairwise comparisons, no signifi-
cant difference was found for any pair. The CPC exhibited less
roughness (p = 0.048) than 10% R. communis solution at ½
(0.05 [0.05; 0.06]; 0.08 [0.07; 0.09]) and 5 years (0.05 [0.04;
0.06]; 0.08 [0.06; 0.09]) (p = 0.013). The other solutions pre-
sented intermediate values at ½ and 5 years. No difference was
observed among the solutions at 1, 2, 3, and 4 years. In the
comparison of the times, water showed a decrease in surface
roughness (Ra) between baseline (0.08 [0.06; 0.09]) and 1 year
(0.06 [0.05; 0.07]). In the other solutions, there was no sig-
nificant difference between surface roughness at the evaluated
times.

Surface topography and chemical composition


The analysis of the surfaces obtained through the SEM can be
visualized in Figure 4. There is an absence of apparent surface
changes, as well as no evidence of corrosion before or after
immersion in cleanser solutions.
The results of the chemical composition analyses of the Co-
Figure 3 Images obtained through the confocal microscope: I. Before Cr alloy obtained by EDS before and after immersion in the
immersion; II. After simulated immersion of 5 years in the following so- cleanser solutions are presented in Table 4. According to the
lutions: A, Deionized water; B, 2% R. communis; C, 10% R. communis; quantitative analysis of the chemical elements, the chemical
D, Effervescent tablet; E, CPC. elements retained final mass values close to the initial ones.

Discussion
Results
In search of an RPD cleaning method that preserves the metallic
Surface roughness
surface, this study evaluated the surface roughness of the Co-Cr
Micrographs of the initial and final surface aspects of the Co-Cr metal alloy commonly used in the construction of RPDs against
alloys immersed in different solutions were obtained by laser the use of chemical cleansers, such as 2% and 10% R. commu-
confocal microscopy (Fig 3). The initial appearance of the Co- nis, effervescent tablet, and CPC. Since a roughness below 0.2
Cr alloy surface was preserved in the final appearance, that μm is considered clinically acceptable because it does not pro-
is, there were no significant alterations of topography, and the mote microbial adhesion,29,30 the null hypothesis of this study

Journal of Prosthodontics 29 (2020) 142–150 


C 2018 by the American College of Prosthodontists 145
146
Table 2 Median (95% confidence interval) of the surface roughness (Sa, μm) for the different times / cleanser solutions and statistical comparisons between the solutions

Baseline ½ year 1 year 2 years 3 years 4 years 5 years P∗∗

H2 O 0.55(0.52; 0.62)a,A 0.56(0.50; 0.65)a,A 0.57(0.49; 0.68)a,A 0.57(0.50; 0.67)a,A 0.60(0.53; 0.70)a,A 0.56(0.49; 0.68)a,A 0.57(0.47; 0.65)a,A 0.572
2% Rc 0.66(0.59; 0.73)a,A 0.53(0.46; 0.63)a,B 0.52(0.47; 0.56)a,AB 0.55(0.46; 0.63)a,B 0.50(0.45; 0.55)a,B 0.57(0.49; 0.59)a,AB 0.54(0.47; 0.61)a,B <0.001
10% Rc 0.52(0.43; 0.66)a,AB 0.52(0.45; 0.67)a,AB 0.53(0.47; 0.69)a,B 0.50(0.44; 0.65)a,A 0.52(0.44; 0.71)a,AB 0.54(0.47; 0.69)a,AB 0.52(0.44; 0.63)a,AB 0.007
ET 0.63(0.51; 0.69)a,A 0.59(0.52; 0.68)a,A 0.62(0.53; 0.75)a,A 0.60(0.52; 0.67)a,A 0.58(0.50; 0.69)a,A 0.62(0.51; 0.73)a,A 0.59(0.52; 0.74)a,A 0.489
CPC 0.51(0.48; 0.69)a,A 0.53(0.47; 0.63)a,A 0.57(0.49; 0.65)a,A 0.57(0.45; 0.63)a,A 0.56(0.46; 0.67)a,A 0.56(0.48; 0.65)a,A 0.51(0.44; 0.60)a,A 0.030
Effect of Denture Cleansers on Co-Cr

P∗ 0.246 0.788 0.482 0.682 0.417 0.756 0.476


*
Kruskal-Wallis test, comparison of solutions at each time
ab
equal lowercase letters indicate statistical similarities between solutions
**
Friedman test, comparison of the times for each solution
AB
equal capital letters indicate statistical similarities between times
H2 O - Deionized water (control); 2% Rc - 2% R. communis;10% Rc - 10% R. communis; ET- Effervescent tablet (NitrAdine) e; CPC - Cetylpyridinium chloride

Table 3 Median (95% confidence interval) of the surface roughness (Ra, μm) for the different times / cleanser solutions

Baseline ½ year 1 year 2 years 3 years 4 years 5 years P∗∗

H2 O 0.08(0.06; 0.09)a,A 0.07(0.06; 0.07)ab,AB 0.06(0.05; 0.07)a,B 0.06(0.06; 0.07)a,AB 0.07(0.06; 0.07)a,AB 0.07(0.05; 0.07)a,AB 0.07(0.06; 0.08)ab,AB 0.013
2% Rc 0.08(0.06; 0.08)a,A 0.06(0.05; 0.08)ab,A 0.06(0.05; 0.07)a,A 0.07(0.05; 0.07)a,A 0.06(0.05; 0.07)a,A 0.07(0.06; 0.08)a,A 0.06(0.05; 0.07)ab,A 0.271

Journal of Prosthodontics 29 (2020) 142–150 


10% Rc 0.08(0.06; 0.09)a,A 0.08(0.07; 0.09)a,A 0.06(0.05; 0.07)a,A 0.08(0.06; 0.09)a,A 0.07(0.06; 0.08)a,A 0.06(0.05; 0.08)a,A 0.08(0.06; 0.09)a,A 0.233
ET 0.08(0.06; 0.09)a,A 0.07(0.05; 0.09)ab,A 0.07(0.06; 0.07)a,A 0.07(0.06; 0.09)a,A 0.07(0.06; 0.08)a,A 0.07(0.06; 0.08)a,A 0.07(0.06; 0.08)ab,A 0.415
CPC 0.06(0.05; 0.07)a,A 0.05(0.05; 0.06)b,A 0.05(0.04; 0.06)a,A 0.06(0.05; 0.06)a,A 0.05(0.05; 0.06)a,A 0.06(0.05; 0.07)a,A 0.05(0.04; 0.06)b,A 0.252
P∗ 0.028 0.048 0.084 0.084 0.066 0.491 0.013
*
Kruskal-Wallis test, comparison of solutions at each time
ab
equal lowercase letters indicate statistical similarities between solutions
**
Friedman test, comparison of the times for each solution
AB
equal capital letters indicate statistical similarities between times
H2 O - Deionized water (control); 2% Rc - 2% R. communis;10% Rc - 10% R. communis; ET- Effervescent tablet (NitrAdine) e; CPC - Cetylpyridinium chloride

C 2018 by the American College of Prosthodontists


Curylofo et al
Curylofo et al Effect of Denture Cleansers on Co-Cr

Surface roughness favors the formation and accumulation of


biofilm,32,33 causing it to be strongly adhered to the base of the
denture33 and promoting the development of an inflammatory
process in the underlying mucosa.29 According to Teughels
et al,32 the initial adhesion of microorganisms in rough areas
occurs due to a lower incidence of shearing forces under these
conditions and, additionally, it makes the daily cleaning of the
denture, especially of RPDs, more complex.
In this study, immersion simulations were performed in the
cleansers from ½ to 5 years, corresponding to the RPD lifes-
pan noted in the literature.1,19 According to the results obtained
through the confocal microscope, there was no significant dif-
ference between the cleaning solutions in the 6 evaluated times
(Table 2); however, with the profilometer, the results indicated
a higher Ra in 10% R. communis when compared to CPC, both
at ½ and 5 years (Table 3). This result can be attributed to the
initial Ra of the 10% R. communis and CPC groups, which
presented the highest and lowest initial values, respectively,
but without significant difference when compared to the initial
roughness values of the other groups. Thus, the small rough-
ness changes observed at ½ and 5 years promoted significantly
different results between them; however, such roughness values
were considered within normal ranges, since they are clinically
acceptable (below 0.2 μm).29,30
The profilometer (Ra) calculates the mean deviation of a pro-
file, and the confocal microscope (Sa) calculates the arithmetic
mean of the surface deviation.34 The profilometer is a quanti-
tative method only, whereas the laser confocal microscope al-
lows high-resolution 3D images to be obtained for subsequent
quantitative and qualitative analysis of surface roughness. Re-
cently, studies related to several areas have used this apparatus
for surface analyses;34,35 however, none was performed with
a similar purpose to this study. Therefore, these two differ-
ent methods were used to evaluate, in addition to the surface
changes, the sensitivity of each technique, in order to help future
studies.
The confocal microscope was sensitive to surface changes,
allowed a greater area of analysis, better distributed the er-
ror, and did not induce different results, such as the difference
between 10% R. communis and CPC obtained when using the
profilometer. In this case, it is possible to consider that while the
Ra value was very close to the recommended value, there was
no significant surface change, as well as in all other evaluated
solutions.
The results found in this study for the effervescent tablet
corroborate Davi et al’s2 and Paranhos et al’s findings,18
Figure 4 Electromicrographs: I. Before immersion; II. After simulated
that when assessing metallic surfaces of different dental al-
immersion of 5 years in the following solutions: A, Deionized water; B,
loys, such as pure titanium, nickel-chromium-molybdenum-
2% R. communis; C, 10% R. communis; D, Effervescent tablet; E, CPC.
titanium, nickel-chromium, and nickel-chromium-beryllium,
that there was no influence of the effervescent tablet cleanser on
the surface roughness, and furthermore that it did not change the
was accepted, since there were no clinically relevant differ- brightness of the metallic surfaces. However, both studies sim-
ences between the chemical cleansers tested on the Co-Cr metal ulated only 6 months of use. Nevertheless, in previous studies
surface during the simulated 5-year period. Several chemical with similar methodology, which also used NitrAdine effer-
cleaning methods have proven to be beneficial, especially re- vescent tablet, adverse effects such as oxidation15 and greater
garding antimicrobial action and removal of biofilm.7-9,17 How- release of ions16 were found in Co-Cr alloys. Such divergent
ever, at the same time that chemical cleaning is important for results may be related not only to the simulated immersion
the removal of biofilm, some studies imply that the surface of period reproduced in these studies, but also to the change in
the acrylic resin changes when using these cleansers.15,31 composition and mode of use of the effervescent tablet.

Journal of Prosthodontics 29 (2020) 142–150 


C 2018 by the American College of Prosthodontists 147
Effect of Denture Cleansers on Co-Cr Curylofo et al

Table 4 Quantitative analysis (% by mass) of the chemical elements before and after the 5-year immersion simulation

Solutions Phases Co Cr Mo O

H2 O Initial 64.274 28.006 5.318 0.950


Final 65.152 28.104 5.357 0.570
2% Rc Initial 63.663 27.936 5.819 1.083
Final 64.429 27.970 5.889 0.563
10% Rc Initial 64.537 28.163 5.231 0.794
Final 64.500 27.842 5.956 0.549
ET Initial 64.040 28.059 5.765 0.854
Final 65.258 28.078 5.196 0.565
CPC Initial 64.085 28.326 5.686 0.934
Final 64.758 28.051 5.625 0.548

H2 O - Deionized water (control); 2% Rc - 2% R. communis; 10% Rc - 10% R. communis; ET- Effervescent tablet (NitrAdine) e; CPC- Cetylpyridinium chloride.

Currently, the manufacturer recommends immersion only 2 tivity in complete dentures,7 and to 0.5% sodium hypochlorite
times a week, that is, every 3.5 days, a fact that decreases the when used in root canals with pulp necrosis.23 R. communis
exposure of the metal to the action of the constituent elements presents benefits such as low cost, biocompatibility, and deter-
of the effervescent tablet. Moreover, the new formulation has gent properties.38 In this way, the solution of R. communis can
the addition of sodium carbonate (neutral salt), which may be safely indicated in the concentrations evaluated as an RPD
contribute toward preserving or decreasing the amount of cleanser. It is cultivated in many countries, thus its production
oxygen in the surfaces of metallic specimens, which may be and commercialization is feasible for denture users.
why the oxygen content did not increase as Felipucci et al’s The cleanser 0.5 mg CPC is classified as a cationic surfactant
study.15 Furthermore, unlike most commercially available having antibacterial action.39 The results obtained in this study
effervescent tablets, which include sodium perborate or sodium corroborate Felipucci et al16 and Davi et al’s findings2 that the
percarbonate as an oxygen-releasing agent, and an alkaline surface of the dental alloys used in the manufacture of partial
detergent such as trisodium phosphate,36 NitrAdine contains dentures showed no change due to the use of this solution. In
citric acid, responsible for breaking bridges of calcium ions addition, this solution does not cause oxidation in the Co-Cr
that act as chemical binding sites for the extracellular polymer alloy or change the roughness of the acrylic resin,15 and can
chains, contributing to its antimicrobial action,37 and sodium thus be seen as an alternative for denture cleaning.
lauryl sulfate, which is a detergent used in the solubilization The simulated immersion periods varied according
of proteins. Thus, it is possible that the different components to the manufacturers’ recommendations and literature
that contribute to its effectiveness14,36 do not change the Co-Cr findings,2,9,16,18,20,25 and the solutions tested were selected
surface. according to their properties. Our findings show the effects
The solutions of 2% R. communis, and 10% R. communis of these cleansers over the lifetime of the denture, bring-
promoted significant changes of the surface roughness (Sa) as ing information on the adverse effects not yet reported in
a function of time, when the metallic specimens were evalu- the literature, since the studies found were of short-term
ated by the laser confocal microscope. Experimental solutions evaluations.2,13,15,16,18
presented a tendency to decrease roughness in ½, 2, 3, and Complementary qualitative analysis confirmed the findings
5 years in 2% R. communis, and 1 and 2 years in 10% R. of the quantitative analysis; there were no changes in texture,
communis. According to Pisani et al,20 the loss of soluble com- staining, oxidation, or any other deleterious effect promoted
ponents during immersions may cause the formation of voids by the solutions. From the EDS analysis, it was observed that
or bubbles with a configuration of peaks and valleys responsi- there was a tendency for the oxygen levels to decrease over
ble for the roughness, and when the solubilization happens in time immersed, confirming the absence of oxidation of the
the peaks, the final average roughness decreases, making the metal surface. As a control solution, deionized water was used,
specimen smoother. However, as previously mentioned, the dif- as in other works of the group,2,15,16,18 to avoid any interference
ferences between times were occasional and not for all groups. to the surface roughness due to the release of ions which would
Additionally, there is no way to compare these findings with occur when using tap or distilled water. According to the ob-
other authors because none evaluated the action of R. com- tained results, from different complementary methodologies,
munis in dental alloy, this being a pioneering study in this the tested solutions did not promote adverse effects that could
evaluation. harm the Co-Cr surface. It is expected that these results will be
These results demonstrate that solutions of both 2% and 10% associated with other variables, so that there is a consensus for
R. communis (Table 3) provided clinically acceptable values of indication of RPD cleaning methods, considering its limitations
roughness,29,30 and did not cause deleterious effects on the and means of action. Thus, it will be possible for the dentist to
Co-Cr surface. R. communis, despite its mechanism of action safely indicate the best cleaning method for RPD users.
still being unknown, is indicated to have an action compara- The focus of this study was to analyze the chemical im-
ble to that of alkaline peroxide toward the antimicrobial ac- mersion method alone, that is, without associating it with the

148 Journal of Prosthodontics 29 (2020) 142–150 


C 2018 by the American College of Prosthodontists
Curylofo et al Effect of Denture Cleansers on Co-Cr

mechanical method, seeking an objective evaluation of the 12. Pero AC, Scavassin PM, Nunes EM, et al: Bond strength of
chemical methods of cleaning. Future in vitro studies should artificial teeth attached to a microwave-polymerized denture base
be performed simulating oral cavity conditions such as biofilm resin after immersion in disinfectant solutions. J Prosthodont
presence, temperature changes, and salivary pH, and in vivo, 2016;25:556-579
to evaluate the effects and the acceptance of cleansers by RPD 13. Rodrigues Garcia RC, Joane Augusto S Jr, Rached RN, et al:
Effect of denture cleansers on the surface roughness and
users, to indicate their use as a viable alternative.
hardness of a microwave-cured acrylic resin and dental alloys.
J Prosthodont 2004;13:173-178
Conclusions 14. Silva-Lovato CH, Wever BD, Adriaens E, et al: Clinical and
antimicrobial efficacy of NitrAdine TM -based disinfecting
Based on the experimental conditions and methods used in the cleaning tablets in complete denture wearers. J Appl Oral Sci
present study, it was concluded that: 2010;18:560-565
15. Felipucci DNB, Davi LR, Paranhos HFO, et al: Effect of
1. There was no change in surface roughness of the cobalt- different cleansers on the surface of removable partial denture.
chromium alloy immersed in the solutions, for the simu- Braz Dent J 2011;22:392-397
lated period of 5 years’ use. 16. Felipucci DNB, Davi LR, Paranhos HFO, et al: Effect of
2. According to the images obtained by SEM and confo- different cleansers on the weight ion release of removable partial
cal microscopy, the cleaning solutions evaluated did not denture: an in vitro study. J Appl Oral Sci 2011;19:483-487
cause changes in the cobalt-chromium surface suggestive 17. Coimbra FC, Salles MM, De Oliveira VC, et al: Antimicrobial
efficacy of complete denture cleansers. Am J Dent
of oxidation.
2016;29:149-153
3. The cleaning solutions evaluated did not promote ad- 18. Paranhos HFO, Bezzon OL, Davi LR, et al: Effect of cleanser
verse effects on the cobalt-chromium surface and could solutions on the color of acrylic resins associated with titanium
be safely used under recommended use conditions over and nickel-chromium alloys. Braz Oral Res 2014;28:1-7
a period of 5 years. 19. Knezović-Zlatarić D, Celebić A, Valentić-Peruzović M, et al: A
survey of treatment outcomes with removable partial dentures.
J Oral Rehabil 2003;30:847-854
References 20. Pisani MX, Silva-Lovato CH, Paranhos HFO, et al: Evaluation of
1. Cakan U, Yuzbasioglu E, Kurt H, et al: Assessment of hygiene experimental cleanser solution of Ricinus communis: effect on
habits and attitudes among removable partial denture wearers in soft denture liner properties. Gerodontology 2012;29:179-185
a university hospital. Niger J Clin Pract 2015;18:511-515 21. Pisani MX, Macedo AP, Paranhos HFO, et al: Effect of
2. Davi LR, Felipucci DNB, Souza RF, et al: Effect of denture experimental Ricinus communis solution for denture cleaning on
cleansers on metal ion release and surface roughness of denture the properties of acrylic resin teeth. Braz Dent J 2012;23:15-21
base materials. Braz Dent J 2012;23:387-393 22. Badaró MM, Prates TP, Leite-Fernandes VMF, et al: In vitro
3. Souza RF, Paranhos HFO, Lovato-Silva CH, et al: Interventions evaluation of resilient liner after brushing with conventional and
for cleaning dentures in adults. Cochrane Database Syst Rev experimental Ricinus communis-based dentifrices.
2009;7:CD007395 J Prosthodont. 2017 Dec 13. https://doi.10.1111/jopr.
4. Padilha DM, Hugo FN, Hilgert JB, et al: Hand function and oral 12680. [Epub ahead of print]
hygiene in older institutionalized Brazilians. J Am Geriatr Soc 23. Meneghin MP, Nomelini SMB, Souza-Neto MD, et al:
2007;55:1333-1338 Morphologic and morphometric analysis of the root canal apical
5. Paranhos HFO, Silva-Lovato CH, Souza RF, et al: Effects of third cleaning after biomechanical preparation using 3.3%
mechanical and chemical methods on denture biofilm Ricinus communis detergent and 1% NaOCl as irrigating
accumulation. J Oral Rehabil 2007;34:606-612 solutions. J App Oral Sci 2006;14:178-82
6. Papadiochou S, Polyzois G: Hygiene practices in removable 24. Arruda CNF, Salles MM, Badaró MM, et al: Effect of sodium
prosthodontics: a systematic review. Int J Dent Hyg hypochlorite and Ricinus communis solutions on control of
2017;16:179-201 denture biofilm: a randomized crossover clinical trial. J Prosthet
7. Andrade IM, Andrade KM, Pisani MX, et al: Trial of an Dent 2017;117:729-34.
experimental castor oil solution for cleaning dentures. Braz Dent 25. Badaró MM, Salles MM, Arruda CNF, et al: In vitro analysis of
J 2014;25:43-47 surface roughness of acrylic resin exposed to the combined
8. Aoun G, Cassia A, Berberi A: Effectiveness of a chlorhexidine hygiene method of brushing and immersion in Ricinus communis
digluconate 0.12% and cetylpyridinium chloride 0.05% solution and sodium hypochlorite. J Prosthodont 2017;26:516-521
in eliminating Candida albicans colonizing dentures: a 26. Bezzon OL, Pedrazzi H, Zaniquelli O, et al: Effect of casting
Randomized Clinical in vivo Study. J Contemp Dent Pract technique on surface roughness and consequent mass loss after
2015;16:433-436 polishing of NiCr and CoCr base metal alloys: a comparative
9. Salles MM, Oliveira VC, Souza RF, et al: Antimicrobial action study with titanium. J Prosthet Dent 2004;92:274-277
of sodium hypochlorite and castor oil solutions for denture 27. Unlü A, Altay OT, Sahmali S: The role of denture cleansers on
cleaning—in vitro evaluation. Braz Oral Res 2015;29:1-6 the whitening of acrylic resins. Int J Prosthodont 1996;9:266-270
10. Badaró MM, Salles MM, Leite VMF, et al: Clinical trial for 28. Moy A, Merlet C, Dugne O: Standardless quantification of heavy
evaluation of Ricinus communis and sodium hypochlorite as elements by electron probe microanalysis. Anal Chem
denture cleanser. J Appl Oral Sci 2017a;25:324-34 2015;87:7779-7786
11. Vasconcelos GLL, Curylofo PA, Raile PN, et al: Effect of 29. Quirynen M, Bollen CM: The influence of surface roughness and
alkaline peroxides on the surface of cobalt chrome alloy: an in surface-free energy on supra- and subgingival plaque formation
vitro study. J Prosthodont 2018 Mar 24. in man. A review of the literature. J Clin Periodontol
https://doi.org/10.1111/jopr.12789. [Epub ahead of print] 1995;22:1-14

Journal of Prosthodontics 29 (2020) 142–150 


C 2018 by the American College of Prosthodontists 149
Effect of Denture Cleansers on Co-Cr Curylofo et al

30. Bollen CM, Lambrechts P, Quirynen M: Comparison of surface 36. Dhamande MM, Pakhan AJ, Thombare RU, et al: Evaluation
roughness of oral hard materials to the threshold surface of efficacy of commercial denture cleansing agents to reduce
roughness for bacterial plaque retention: a review of the the fungal biofilm activity from heat polymerized denture
literature. Dent Mater 1997;13:258-269 acrylic resin: an in vitro study. Contemp Clin Dent 2012;3:
31. Schwindling FS, Rammelsberg P, Stober T: Effect of chemical 168-172
disinfection on the surface roughness of hard denture base 37. Desrosiers M, Myntti M, James G: Methods for removing
materials: a systematic literature review. Int J Prosthodont bacterial biofilms: in vitro study using clinical chronic
2014;27:215-225 rhinosinusitis specimens. Am J Rhinol 2007;21:527-532
32. Teughels W, Van Assche N, Sliepen I, et al: Effect of material 38. Nemudzivhadi V, Masoko P: In vitro assessment of cytotoxicity,
characteristics and/or surface topography on biofilm antioxidant, and anti-inflammatory activities of Ricinus
development. Clin Oral Implants Res 2006;17:68-81 communis (Euphorbiaceae) leaf extracts. Evid Based
33. Wu T, Hu W, Guo L, et al: Development of a new model system Complement Alternat Med 2014.
to study microbial colonization on dentures. J Prosthodont https://doi:10.1155/2014/625961
2013;22:344-350 39. Watanabe E, Nascimento AP, Guerreiro-Tanomaru JM, et al:
34. Park JB, Kim N, Ko Y: Effects of ultrasonic scaler tips and Antiseptic mouthwashes: in vitro antibacterial activity. Acta
toothbrush on titanium disc surfaces evaluated with confocal Odontol Latinoam 2015;28:180-184
microscopy. J Craniofac Surg 2012;23:1552-1558 40. Zoidis P, Polychronakis N, Lagouvardos P, et al: Evaluation of a
35. Evans AR, Harper IS, Sanson GD: Confocal imaging, realistic cleansing protocol for preventing discoloration of
visualization and 3-D surface measurement of small mammalian denture resins. J Prosthodont 2018 Feb 22. https://doi.10.1111/
teeth. J Microsc 2001;204:108-118 jopr.12763. [Epub ahead of print]

150 Journal of Prosthodontics 29 (2020) 142–150 


C 2018 by the American College of Prosthodontists

You might also like