Professional Documents
Culture Documents
Applikat 102 Standby Power Rev 1.5 e
Applikat 102 Standby Power Rev 1.5 e
5)
ZES ZIMMER Electronic Systems GmbH Tel. +49 (0) 6171/6287-50 sales@zes.com www.zes.com
ZES ZIMMER Inc. Phone. + 1 760 550 9371 usa@zes.com
Application Note 102 (Rev. 1.5)
ZES ZIMMER Electronic Systems GmbH Tel. +49 (0) 6171/6287-50 sales@zes.com www.zes.com
ZES ZIMMER Inc. Phone. + 1 760 550 9371 usa@zes.com
Application Note 102 (Rev. 1.5)
The situation is slightly different if the and hence not a subject for common mode
signal becomes too low for a range. signals. This could particularly be a
Imagine a relatively long measuring cycle. problem for cheap instruments because
Almost immediately after the start of a their common mode rejection of only
new running cycle the consuming device 60 dB-80 dB might not be sufficient!
switches from normal mode down to However such a recommendation may lead
standby. This would justify a lower range. to wrong measurement results. Fig. 3
However, at the end of this measuring shows a simplified circuit. To fulfil the
cycle, the instrument has correctly EMC requirements the circuit has a filter
registered, that the peak value of the with capacitors (CX and CY) and so the
measured signal was still within the actual device is actually not a 2 wire but a 3 wire
range. Only after the end of the following system because of PE. Only by measuring Il
cycle the instrument can detect, that it the entire current flowing through the
should switch down to a lower range using device is measured.
the algorithm already described above.
In brief it can be summarized here that
shifting the range down will result in a
signal measurement with low accuracy and
a loss of samples for about 2 cycles.
If your device has a constant input current
it doesn’t matter how to select a range. But Fig. 3: Where to measure?
in a worst case it could happen with With one power meter the current IL must be measured,
because the load has a 3 wire connection.
automatic range selection, that a pulsed
signal is measured completely wrong: The And obviously due to the fact that loads in
pulses occur in the gap when switching up, reality usually contain reactive and non-
and the signal between them is measured in linear components it should be clear that
a range which is too high. only measuring the current and multiplying
Thus, whenever it is possible it is it with the nominal voltage in order to
recommended to use manual range calculate the power consumption cannot
settings. In many cases the higher error suffice.
caused by a range which is too high for Gapless Measurements
some signal parts is not as fatal as missing As written in section “Range selection”, a
signal parts like peaks. Please remember, gap in the measuring algorithm can end up
that rms values are measured according a in completely useless results, especially
square law: under conditions where the input current is
not constant. This gap occurs with every
1 T
T ∫t =0
I TRMS = i (t ) 2 dt instrument for the reasons explained
above.
This shows that a signal part with let’s say
a 100 times higher amplitude will influence But there are also other causes for gaps
the result by a factor of 10000! which can be avoided: Some cheap
instruments have to work in two modes
Where to measure the current due to its limited processing capacity. At
For single phase applications it is usually first they measure for several periods. Then
recommended to measure the current in the data is processed and displayed. Data
the neutral wire. By doing so the current acquisition is halted during the latter
channel of the instrument is not floating phase.
3/8
ZES ZIMMER Electronic Systems GmbH Tel. +49 (0) 6171/6287-50 sales@zes.com www.zes.com
ZES ZIMMER Inc. Phone. + 1 760 550 9371 usa@zes.com
Application Note 102 (Rev. 1.5)
ZES ZIMMER Electronic Systems GmbH Tel. +49 (0) 6171/6287-50 sales@zes.com www.zes.com
ZES ZIMMER Inc. Phone. + 1 760 550 9371 usa@zes.com
Application Note 102 (Rev. 1.5)
On the other hand the power source – 0.5 W shall be made with an uncertainty of
especially if it is a switching power supply less than or equal to 0.01 W at the 95 %
– might have a remaining ripple on the confidence level.”
voltage. A value of 1 V and frequencies in
Seemingly the two phrases do not contra-
the range of 40 kHz are common. This
dict to each other but the devil is in the
voltage can drive appreciable currents,
details.
especially through filter capacitors (CX, CY,
see fig. 3) but also in a conventional Here at ZES ZIMMER we manufacture
switched power supply, when the diodes of probably the most precise single phase
the rectifier are in conductive mode. In power meter on the market, the LMG95. So
such cases further active power might be let’s do the error calculation for this
consumed. device:
For energy saving intentions, a bandwidth The standard accuracy at 50 Hz for a power
of 2 kHz should be sufficient. Estimates measurement is specified as
about the thermal effects need to take the ± (0.015 % of reading + 0.01 % of range)
active power above 2 kHz being generated
and physically consumed into account. Along with the error of the shunt SHxxx-P
Otherwise the calculation might be we get a maximum total uncertainty of
completely wrong. ΔP = ±(0.165% of reading + 0.01% of range)
In any case a rich selection of filters is which is about 1/3 of the recommendation
good to have for determining easily in of the standard.
which frequency band the power is being
Example 1a
consumed.
A 10 W load, power factor is 1.0 (i.e. a
Alternatively, the harmonic analysis fully ohmic load). The current is 43.48 mA,
function - if the power meter has one - can used is the range with 50 mA nominal
also calculate the active power produced at value and a peak value of 156.3 mA
each frequency.
ΔP = ±0.023W
Accuracy
This is a relative error of 0.23 %, which is
When taking a look at EN 62301, Annex B.5
less than the standard’s requirement of
you’ll find the following sentence:
2 %.
“Generally, a digital power analyser with a
fundamental power accuracy of 0.5 % or
better will comfortably meet the instrument Example 1b
specification and measurement uncertainty A 10 W load, the power factor is 0.3 and
required in this standard.” the crest factor is 3, typically for a
At first glance is this great. For 0.5 % switching power supply in this power
fundamental power accuracy you find a lot range. The current is 144.9 mA with
of cheap instruments. about 435 mA peak value. We have to use
But, the above is an informative annex and a range with 250 mA nominal value and a
the real requirements are specified in peak value of 781.5 mA
chapter 4.5: ΔP = ±0.048W
“Measurements of power of 0.5 W or greater
This is a relative error of 0.48 %, which is
shall be made with an uncertainty of less
lower than the standard’s requirement of
than or equal to 2 % at the 95 % confidence
2%.
level. Measurements of power of less than
5/8
ZES ZIMMER Electronic Systems GmbH Tel. +49 (0) 6171/6287-50 sales@zes.com www.zes.com
ZES ZIMMER Inc. Phone. + 1 760 550 9371 usa@zes.com
Application Note 102 (Rev. 1.5)
These two examples do not look like a measured with any existing instrument
challenge, at least not for such a precise within the required uncertainty of 2 %.
instrument. For an instrument with a lower
specification of 0.5 % fundamental power So it is not reputable if a manufacturer
accuracy this would also not be a problem. advertises that the instruments meet the
But this was at a fairly high power. Let’s requirements or are certified concerning EN
therefore repeat the calculation with a 62301. A general statement like this is
typical standby power of 0.5 W: untrustworthy for the reasons explained
above.
Example 2
Examples:
Manufacturer Y attained under almost ideal
A 0.5 W load with a power factor of 0.1 conditions (range fully used, crest factor <
and a crest factor of 6 which is typical 3, ...) already a usage of 70 % of the
for a switch mode power supply in this allowed uncertainty only by regarding the
power range. The current then is 21.74 instrument without the environment Such
mA with a 130.43 mA peak value. The
an instrument can hardly be used in
chosen range is 50 mA nominal value practice, because such conditions cannot
and a peak value of 156.3 mA be found in reality. It seems wry that this
ΔP = ±7.077 mW instrument is advertised as certified for
measuring standby power.
This is a relative error of 1.42%, which
is lower than the standards requirement Likewise also manufacturer V achieves the
of 2 % and the absolute error is less requirements of the EN 62301 according to
than 10 mW. his own calculation under almost ideal
conditions (power factor 1, ideal external
As it can be seen the required tolerance of shunt …). 85 % of the allowed uncertainty
2 % resp. 10 mW is already utilised by is used by the instrument itself, again
70 %. Clearly does a much lower specified without any external components which
instrument according annex B.5 not are essential. Likewise, this instrument as
“comfortably meet” the requirements of the well can hardly be used in practice.
standard. Nevertheless the manufacturer guarantees
to measure according to the requirements
But please bear keep in mind that of EN 62301 if some technical hints and
EN 62301 does not limit the uncertainty of the terms expressed in the manual are met.
the instrument but the uncertainty of the
complete measurement setup. Such delusive certificates and guarantees,
valid only under ideal conditions, will last
No instrument will fulfil the standard until the first uncertainty calculation with
under all circumstances! a real world EUT takes place. If the daily
In the above examples values were used grind shows that these instruments will
which are today the edge of the not be able to keep their certificates or
technology: A power factor of 0.1 and a guarantees promises, the bill will be paid
crest factor of 6. Sophisticated „0 W PCs“ by the customer, not by the manufacturer.
already have crest factors of 14. It is hard Frankly spoken: Our products cannot
to imagine what else will be developed in measure everything under all
the future or already exists, but it is circumstances!
certain that an EUT with a power factor of
0.01 and a crest factor of 100 cannot be
6/8
ZES ZIMMER Electronic Systems GmbH Tel. +49 (0) 6171/6287-50 sales@zes.com www.zes.com
ZES ZIMMER Inc. Phone. + 1 760 550 9371 usa@zes.com
Application Note 102 (Rev. 1.5)
On the other hand we at ZES ZIMMER time a lot happens. Drift, ageing, chemical
won’t hide behind fancy guarantees and reactions… they all influence the accuracy
certificates. Serious certificates and over time and cannot be contained in an
guarantees have to show the limits of an one-time calibration. But the
instrument clearly, because global manufacturer’s specification takes care for
statements are not possible as shown it. So if the above 10 % instrument has 0
above. % error according to the calibration
protocol, it is quite likely that this
This leads to the fact that users can’t trust
instrument will have much more than 2 %
to the certification of an instrument only
after one year. Had it been designed for a
but has to make themself familiar with the
better specification, the documentation
contained limits. So we will use the two
would tell it!
foregoing examples to calculate our
uncertainties to be comparable: So it is as much important to look for a
good specification as to have a valid
For the “Y” example (230 V, 20 mA, CF<3,
calibration protocol. A calibration with a
0.46 W) we use a range with nominal value
good result is never a replacement for a
of 24 mA and peak value of 78.12 mA. Our
bad specification.
product achieves an uncertainty of 0.84 %
which is the usage of only 42 % of the How to build up a reasonable testing
requirements of the standard. So our system?
instrument is roughly 2 times better than The provocative message of last chapter
the competitor and the requirements. that no instrument will completely fulfil
And for the “V” example (230 V, 4.3 mA, the standard is unfortunately true.
CF<3, 1 W) we apply a range with nominal Nevertheless there are some strategies to
value 5 mA and peak value 15.63 mA. An perform these kinds of tests seriously:
uncertainty of 0.23 % is the worst case, • Use an instrument with smallest
which is the utilization of 12 % of the possible uncertainty. A 0.025 %
allowed 2 % uncertainty - roughly 7 times instrument can be used for way more
better than the competitor and the critical EUTs than a 0.2 % instrument.
requirements. Certificates aren’t necessary • Calculate the (instrument) uncertainty
at all. for each measure point. This is the
only way to get an impression about
Can I test if a measuring instrument is
the reliability of the results and to
suitable for this kind of measurement?
exploit the limits of the instrument.
Yes and no!
Yes, you can measure if an instrument has If one complies with these two rules, one
a sufficient accuracy at the time of will always receive a reliable testimony
calibration. But this is not really relevant, about the measurement. This might be
because even an instrument with 10 % important in cases where regulatory
specified uncertainty might have an error authorities question the results.
at calibration time of 0 % by coincidence! Additional influencing factors
No, calibration is a singular event. Every The EC directives as well as EN 62301
manufacturer tries to specify its require an uncertainty of 2 % at 95 %
instruments as accurate as possible. confidence level. Not only the instrument
Additionally, they are designed to meet uncertainty but all other influencing
the specification for a certain period at factors are included in this marginal value.
minimum, typically 12 months. During that This requirement is extremely important to
7/8
ZES ZIMMER Electronic Systems GmbH Tel. +49 (0) 6171/6287-50 sales@zes.com www.zes.com
ZES ZIMMER Inc. Phone. + 1 760 550 9371 usa@zes.com
Application Note 102 (Rev. 1.5)
8/8
ZES ZIMMER Electronic Systems GmbH Tel. +49 (0) 6171/6287-50 sales@zes.com www.zes.com
ZES ZIMMER Inc. Phone. + 1 760 550 9371 usa@zes.com