Professional Documents
Culture Documents
SpectroScan 80D IM
SpectroScan 80D IM
SpectroScan 80D IM
SpectroScan 80D/80DV
UV-VIS SPECTROPHOTOMETER
SpectroScan 80D/80DV Instruction manual
1- CONTENT
Chapter 1 Installation………………………………………………………………………………………..1
1.1 Inspection of Parts………………………………………………………………………………1
1.2 General Description of the Spectrophotometer………………………………………………………1
1.2.1 Front side of the instrument……………………………………………………………………1
1.2.2 Rear side of the instrument…………………………………………………………………….…2
1.2.3 Display……………………………………………………………………………………………2
1.2.4 Keyboard………………………………………………………………………………………….3
1.3 Site Requirement…………………………………………………………………………………….…4
1.4 Installation………………………………………………………………………………………...…4
1.5 Turning on the instrument…………………………………………………………………………...…4
1.6 Performance Check at Installation……………………………………………………………………..5
1.6.1 Wavelength accuracy and reproducibility…...……………………………………………………5
1.6.2 Baseline flatness…………………………..……………………………………………………5
ii
Chapter 7 Utilities………………………………………………………………………………35
7.1 Information Display and Function Keys…………………….………………………...………………. 35
7.2 Setting Lamp Changeover Wavelength…………………………………………………………………35
7.3 Setting Spectral Bandwidth……………………………………………………………………………..36
7.4 Setting W Lamp On/ Off………………………………………………………………………………36
7.5 Setting D2 Lamp On/Off………………………………………………………………………………..36
7.6 Setting time……………………………………..……………………………………………………….36
7.7 Dark Current Adjustment………………………………………………………………………………..37
7.8 Wavelength Calibration (Wavelength Adj)……………………………………………………………...37
7.9 Quitting Utilities Mode………………………………………………………………………………….37
Chapter 8 Maintenance……………………………………………………………………………………………38
8.1 Points for attention…………………………………………………………………………….…………38
8.2 Daily Care……………………………………………………………………………………….……….38
8.3 Trouble shooting and maintenance……………………………………………………………….……...38
8.3.1 Diagnoses of initialization…………………………………………………………………….……38
8.4 Light Source Replacement………………………………………………………………………….…...39
8.4.1 Replacement of the W lamp…………………………………………………………………….…..40
8.4.2 Replacement of the D2 lamp……………………………………………………………………..…40
iii
1.Installation
The 320×240 dots high resolution LCD, which allows users to adjust the brightness (the brightness adjustor is
on the top left corner of the keyboard) , is used to display measurement information, parameters, data and
spectrograms.
(2) Keyboard
The 27 soft-touch keys on the keyboard are used for instrument control and operation.
(3) Sample compartment
Automatic 8-cell holder to hold the samples.
light source housing. The lamps switch over automatically according to the wavelength range while measuring.
The lamps will turn hot when they are lit. Be careful not to be burnt when lamp replacement is made.
(2) Printer interface
Used to connect with HP 600 series of jet printer.
(3) RS-232 interface
RS-232 interface is for serial communication.
(4) Power Switch
AC power outlet is used to connect to the 110-240V, 50-60Hz main power.
Fuse holder is to hold a 1A-fuse tube of 5mm.
Power switch is to turn on/off the power of the instrument.
(5) Light source housing cover
It is for access to the light source for the replacement. Be careful not to be burned by the heated lamps when
replacing the light sources.
(6) Ventilating fan
The opening of the fan is the outlet for ventilation and dissipating heat. Please place the main unit at least
20-30cm away from the wall in case that the outlet is blocked.
1.2.3 Display
The display of SpectroScan 80D is the 320×240 dots high resolution LCD that can not only display the
measurement data, parameters, mode and man-machine interaction information, but also the spectrum measured.
The display is shown as follows:
Indicating Area For
Mode Line
Function Keys
Information
Indicating Area
Operating
Time
Area
The first line is the mode line, which is used to indicate current working mode. The system provides four working
modes:
Photometric Measurement – Used to make measurement at fixed wavelengths. K-Factor calculation is
possible.
Spectrum Measurement – Used to provide a wavelength scan and data processing in the set wavelength
range.
Quantitative measurement – Used to provide one wavelength quantitative determination and data
processing.
DNA/Protein analysis – Used to provide 3 measurement methods for the analysis of DNA/Protein.
Utilities – Used to set the parameters for the instrument system.
The column on the right of the screen is the indicating area for function keys (F1~F4). The function of each
function key may vary from page to page.
In the middle of the screen is the operating area.
At the bottom of the screen is the information indicating area, which allows you to enter parameter values
there.
1.2.4 Keyboard
There are altogether 27 soft-touch keys on the keyboard of SpectroScan 80D, which are divided into 4 groups.
The functions of the keys are described as follows:
Cursor keys (▲ ,▼ ): Used to shift the cursor up and down while browsing data and increase or decrease
Return key [RETURN]: Used to quit the current function and return to the preceding page.
Auto zero key [AUTOZERO]: Used to set ABS zero (transmittance 100%). In photometric mode and the
quantitative mode, [AUTOZERO] set ABS zero only at current wavelength. In spectrum mode,
[AUTOZERO] performs baseline calibration in the set wavelength range.
e. Power requirements: AC110V±10%, 60Hz±1Hz, single phase. A regulated power supply with at least
500VA capacity is recommended. There should be a ground line in the laboratory to ensure the
instrument to be well grounded.
1.4 Installation
SpectroScan 80D Spectrophotometer consists of the spectrophotometer main unit and a HP jet-printer (optional).
The connections between them are shown in the following figure. Item (1) and (2) are the power cords which
should not be connected to the main power until other connections are made, to avoid instrument damage. Item (3)
is the printing cable, one end of which with a 25-pin plug is connected to the main unit, the other end with a 36-pin
plug is connected to the printer.
Connect all the cables firmly by locking the screw (if there is any) on the plugs. After checking to ensure that the
power supply is AC220V, 50Hz, plug in the power cords to connect the spectrophotometer and the printer to the
main power.
Note: The spectrophotometer is a precision electronic instrument. Don’t pull out or insert the power cable or
tie cable when the instrument is on, otherwise it will destroy the instrument and the computer.
After confirming that there is no light shield in the sample compartment, turn on the power switch of the main unit
and then the printer (if there is). The initializing page will be displayed on the LCD. The instrument starts
self-checking and initializing. This process will take about 3 minutes. After initializing, the system displays the
instrument main menu.
Users can select a working mode by depressing the corresponding numeric keys (1-5). It usually takes 15~30
minutes for the instrument to warm up before measurement can be started.
After measurement, you should first turn off the main unit, then the printer, and at last other equipment (such as the
regulated power supply).
The wavelength accuracy should be within ±0.5nm. The wavelength reproducibility should be within 0.2nm.
Check with the two spectral lines of the D2 lamp according to the following procedure.
0) Confirm the spectral bandwidth is set to 2nm. The initialized spectral bandwidth is 2nm. Please refer to
THE UTILITIES FUNCTION for details of setting the spectral bandwidth.
1) Press [2] key to select mode “Spectrum Measurement” on the main menu.
2) In Abs or %T mode, press [GOλ ] to set the wavelength to 330nm, to turn to the D2 lamp working mode.
3) Press [F1] key to select parameter setting function. Set the following parameters: Sampling interval: 0.2nm;
Scanning speed: middle; Wavelength range: 660~480nm; Recording range: 0~100; Mode: Ess.
4) Press [START/STOP] key and set the energy gain value to 5~7, to start scanning.
5) Press [F2] key and set the min. peak/valley value to 1~10, to perform the peak-picking. Press [F4] to print out.
Repeat scanning for three times and perform peak-picking and printing out. Compare the picked wavelengths with
the standard wavelengths of the two peaks of D2 lamp: 656.1nm and 486.0nm.
spectrum mode:
0) Confirm that spectral bandwidth is set to 2nm. The initialized spectral bandwidth is 2nm. Please refer to
THE UTILITIES FUNTION for details of setting the spectral bandwidth.
1) Press [2] key to select mode “Spectrum Measurement” on the main menu.
2) Press [F1] key to select parameter set function. Set the following parameters: Sampling interval: 1nm;
Scanning speed: middle; Wavelength range: 1100~190nm; Recording range: -0.1~0.1; Photometric mode:
Abs.
3) Press [AUTOZERO] to start baseline correction.
4) Press [START/STOP] to start scanning.
5) Press [F4] key to print out the spectrum.
Compare the max. Abs value on the spectrum with the value required for the baseline flatness. (Sudden change in
Abs value is allowed at the point of light source changeover.)
2.Instrument Specifications
3.Photometric Measurement
In photometric mode, one-wavelength measurements of absorbance and transmittance can be made. K-Factor
calculation and printing out are possible.
On this page, you can also set the wavelength and execute auto-zero by depressing [GOλ ] key and [AUTOZERO]
key respectively.
On this page, users can set the photometric mode, wavelength and the value of factor K by pressing the
corresponding item No.. After finishing parameter setting, press [RETURN] key to return to the preceding page.
At the bottom of this page, enter desired wavelength value by pressing the numeric keys (0~9) and decimal point
(.). The CE key is used to clear off the preceding entry. Press [ENTER] key to confirm the numeric entry and quit.
If nothing is to be entered, press [RETURN] key to quit this page directly. A 5-digit figure entry is allowed. The
range is from 190 to 1100. The input value keeps one-digit floating number after the decimal point If the figure
entered exceeds five digits, the system will delete what has been entered automatically and you should start a new
entry again.
entered, press [RETURN] key to quit this page directly. A 4-digit figure (0~±9999) entry is allowed. If the figure
entered exceeds four digits, the system will delete what has been entered automatically. You should start a new
entry.
On the main menu of photometric measurement, press [GOTO-λ ] key to display the following page (Fig. 3-6) and
10
11
12
measurement. “Abs” is the measured value Column. “K·Abs” is the value of the measurement multiplied by
K-Factor. In this page, press [START/STOP] key to start measurement. Press [F2] key to delete all the measuring
data. Press [F4] to print out the measuring data. After finishing all the measurement operation, press [RETURN]
key to return to the main menu of operation.
3.6.1 Setting measuring wavelength
Press [GOλ ] key to set measuring wavelength. Please see Section 3.3 for the details of measuring wavelength
setting.
3.6.2 Auto zeroing (100%T calibration)
Press [AUTOZERO] key to set Abs zero at the current wavelength. Please see Section 3.4 for the details of
auto-zeroing (100%T calibration).
samples (refer to§3.5.2). Then the measured value will be calculated, saved and displayed on LCD one by
one.
(2) When Reagent Blank Correction in Cell No. 1 (§3.5.3) is set to “YES”, the system will first measure the set N
samples (refer to §3.5.2). Then the measured values of samples No.2~No.N will be deducted (corrected) by
13
that of the blank in sample cell No. 1 following the formula below. The corrected results will be recorded,
calculated and displayed on LCD one by one.
Formula of deduction (correction):
The absorbance of sample No.n = measured value of sample No.n – measured value of sample No.1(blank)
The transmittance of sample No.n = measured value of sample No.n / measured value of sample No.1(blank)
× 100%
During measurement, users can look up the measured results line by line, by pressing the cursor keys [▲ ],[▼ ] .
14
lost. Users can press [RETURN] key to return to the main menu of photometric measurement.
4.Spectrum Scanning
This mode provides a wavelength scan and print out for absorbance, transmittance and energy of the measured
sample.
On this page, wavelength setting, baseline correction and spectrum scanning can be done by depressing [GOλ ]
15
On this page, users can set the photometric mode, scanning speed, scanning wavelength range, ordinate full scale
range and sampling interval by pressing the corresponding numeric keys. After parameter setting is finished, press
[RETURN] key to return to the preceding page.
16
At the bottom of the page, first set the starting wavelength (long wave) by pressing numeric keys (0~9). [CE] key
is used to clear off the preceding entry. After setting the starting wavelength, the system indicates user to set the
ending wavelength (short wavelength). Press [ENTER] key to confirm the numeric entry and quit. If nothing is to
be entered, press [RETURN] key to quit directly. A 5-digit figure entry is allowed. The range is from 190 to 1100.
If the figure entered exceeds five digits, the system will delete what has been entered. You should start a new entry.
If the ending wavelength value entered is larger than that of the starting wavelength, an error message will be
displayed. You should enter it again.
At the bottom of the page, the system indicates user to set the upper limit of the ordinate. Press numeric keys (0~9),
(.) and (-) to enter the desired value. [CE] key is used to clear off the preceding entry. After setting the upper limit,
the system indicates user to set the lower limit. Press [ENTER] key to confirm the numeric entry and quit. If
nothing is to be entered, press [RETURN] key to quit directly. A 4-digit figure entry is allowed. The range is from
–9999~9999. If the figure entered exceeds four digits, the system will delete what has been entered. You should
enter it again.
17
On the main menu of spectrum scanning, press [GOλ ] key to display the following page (Fig. 4-6) and set the
At the bottom of this page, enter the desired wavelength value by pressing the numeric keys (0~9), [.]. [CE] key is
used to clear off the preceding entry. Press [ENTER] key to confirm the numeric entry and quit. If nothing is to be
entered, press [RETURN] key to quit directly. A 5-digit figure entry is allowed. The range is from 190 to 1100. If
the figure entered exceeds five digits, the system will delete what has been entered. You should start a new entry.
18
range. Before starting baseline correction, put the blank sample into the sample compartment. [START/STOP] key
can be pressed to stop baseline correction operation. If the wavelength range has been changed, the baseline
calibration should be done again.
4.6 Peak-picking
On the main menu of spectrum scanning, press [F2] key to display peak-picking operating page as shown in Fig. (4-8).
19
result will be displayed on the screen. The peaks and valleys on the spectrum are marked. At the same time, an
arrow indicates the current peak/valley point. Below the spectrum, the wavelength and the measured value of the
pointed peak/valley are displayed. By pressing the [ ]or [ ] key, the arrow will be shifted to look up other
peak/valley points. Press [F4] key to print out the result. Press [RETURN] key to return to the main menu of the
spectrum scanning after all the operations are ended.
20
5. Quantitative Measurement
In this mode one-wavelength determination can be done. Both K-Factor Method and Calibration Curve Method are
used in quantitative measurement.
The first line on the menu is the mode line, which shows that the current operation mode is quantitative
measurement.
The upper part of the operating area shows the current measuring wavelength (500.0nm) and the measured value
(0.000Abs), the lower part is a table in which the measured data are displayed.
The right part of the page is the function key indicating area, which shows the functions of function keys [F1]
~[F4].
[F1] – parameters setting
[F2] – clearing off the measurement data
[F3] – sample control
[F4] – printing
On this page, the measuring wavelength setting, auto-zeroing and sample measurement can be carried out by
21
At the bottom of this page, enter the desired wavelength value by pressing the numeric keys (0~9), [.]. [CE] key is
used to clear off the preceding entry. Press [ENTER] key to confirm the numeric entry and quit. If nothing is to be
entered, press [RETURN] key to quit directly. A 5-digit figure entry is allowed. The range is from 190 to 1100. If
the figure entered exceeds five digits, the system will delete what has been entered. You should start a new entry.
22
Fig. (5-4) Calibration curve parameter setting page for K-Factor method
Press [1] key and [2] key to enter slope and intercept values of the calibration curve on this page. Press [F4] key to
print out the parameters currently displayed on the screen. After all parameter settings are completed, press
[RETURN] key to return to the preceding page.
At the bottom of this page, enter the slope value by pressing the numeric keys (0~9), [.]. [CE] key is used to clear
off the preceding entry. Press [ENTER] key to confirm the numeric entry and quit. If nothing is to be entered, press
[RETURN] key to quit directly. A 4-digit figure entry is allowed. The range is from 0 to 9999. If the figure entered
exceeds four digits, the system will delete what has been entered. You should start a new entry. When a decimal
number is entered, a “0” should be put before the decimal point, such as 0.12.
23
At the bottom of this page, enter the intercept value by pressing the numeric keys (0~9), [.] and [-]. [CE] key is
used to clear off the preceding entry. Press [ENTER] key to confirm the numeric entry and quit. If nothing is to be
entered, press [RETURN] key to quit directly. A 4-digit figure entry is allowed. The range is from 0 to 9999. If the
figure entered exceeds four digits, the system will delete what has been entered. You should start a new entry.
Fig.(5-7) Calibration curve parameter setting page for calibration curve method
Press numeric keys [1]~ [3] to enter the number of standard samples, the concentration and absorbance values (if
they are known) of the standard, and to carry out absorbance measurement to the standard. After finishing
24
parameter setting, press [F3] key to set up calibration curve. After all parameter settings are completed, press
[RETURN] key to return to the preceding page.
At the bottom of this page, enter the number of standards by pressing the numeric keys [1]~[8]. [CE] key is used to
clear off the preceding entry. Press [ENTER] key to confirm the numeric entry and quit. A 1-digit figure entry is
allowed. The range is from 1 to 8. If the figure entered exceeds 1 digit, the system will delete what has been
entered. You should start a new entry.
At the bottom of this page, enter the concentration value by pressing the numeric keys (0~9), [.]. [CE] key is used
to clear off the preceding entry. After the concentration is entered, the system will indicate users to enter the
absorbance values. Press [ENTER] key to confirm the numeric entry and quit. If nothing is to be entered, press
[RETURN] key to quit directly. A 4-digit figure entry is allowed. The range is from 0 to 9999. If the figure entered
25
exceeds four digits, the system will delete what has been entered. You should start a new entry. When a decimal
number is entered, a “0” should be put before the decimal point, such as 0.5.
Press [2] again and repeat the above procedure if concentration and absorbance of other standards are to be entered.
26
Press [GOλ ] key to set the measuring wavelength on the main menu of quantitative measurement, which is shown
in Fig. (5-12).
At the bottom of this page, enter the desired wavelength value by pressing the numeric keys (0~9), [.]. [CE] key is
used to clear off the preceding entry. Press [ENTER] key to confirm the numeric entry and quit. If nothing is to be
entered, press [RETURN] key to quit this page directly. A 4-digit figure entry is allowed. The range is from 190 to
1100. If the figure entered exceeds four digits, the system will delete what has been entered automatically. You
should start a new entry.
27
28
samples (refer to§5.5.2). Then the measured value will be calculated, saved and displayed on LCD one by one.
(2) When Reagent Blank Correction in Cell No. 1 (§5.5.3) is set to “YES”, the system will first measure the set N
samples (refer to §5.5.2). Then the measured values of samples No.2~No.N will be deducted (corrected) by
that of the blank in sample cell No. 1 following the formula below. The corrected results will be recorded,
calculated and displayed on LCD one by one.
Formula of deduction (correction):
The absorbance of sample No.n = measured value of sample No.n – measured value of sample No.1 (blank)
The transmittance of sample No.n = measured value of sample No.n / measured value of sample No.1 (blank)
× 100%
29
During measurement, users can look up the measured results line by line, by pressing the cursor keys [▲ ],[▼ ] .
30
6. DNA/Protein Analysis
In this mode DNA/Protein analysis can be done. There are three methods: Method 1, Method 2 and Self-defining.
31
Press [ ] [ ] to select the measuring methods. Method 1 and Method 2 are the system defaulted parameters
that can not be changed by users. Users can set parameters according to their requirements under the self-defining
mode. After finishing all parameter settings, press [RETURN] key to return to the preceding page.
At the bottom of this page, enter the desired wavelength value by pressing the numeric keys (0~9), [.]. [CE] key is
used to clear off the preceding entry. The system will indicate users to input the second wavelength after inputting
the first one. Press [ENTER] key to confirm the numeric entry. If nothing is to be entered, press [RETURN] key to
32
quit directly. A 5-digit figure entry is allowed. The range is from 190 to 1100. If the figure entered exceeds five
digits, the system will delete what has been entered. You should start a new entry.
6.2.3 Setting DNA factor
Under the self-defining mode, press [2] key to display the DNA factor setting page. The system will indicate users
to input DNA factor, as shown in Figure 6-5.
33
6.4 Auto-zeroing
On the main menu of DNA/Protein, press [AUTOZERO] key to set Abs zero automatically at current measuring
wavelengths. Before starting auto zeroing, put the blank sample into the cell.
During measurement, users can look up the measured results line by line, by pressing the cursor keys [▲ ],[ ▼ ] .
34
7. Utilities
In this mode, the operating parameters can be set, such as lamp changeover wavelength, spectral bandwidth,
tungsten lamp On/Off, deuterium lamp On/Off and wavelength calibration etc..
this page, [GOλ ], [AUTOZERO] and [START/STOP] keys are not to be used.
35
entered, press [RETURN] key to quit this page directly. A 4-digit figure entry is allowed. The range is from 190 to
1100. If the figure entered exceeds four digits, the system will delete what has been entered automatically. You
should start a new entry.
7.3 Setting Spectral Bandwidth
Press [2] key to set the spectral bandwidth of the instrument. Four options are available: 2.0, 5.0, 0.5, 1.0nm. They
can be selected in turn when [2] key is pressed each time.
Press [6] key to display time setting page. Press [▲ ] and [▼ ] to set time. Press [ ][ ] to set minute, hour,
36
37
8. Maintenance
SpectroScan 80D Spectrophotometer is a precision instrument. Careful assembling and final test have been done
before delivery. Appropriate routine attention and maintenance will not only ensure the reliability and stability of
the instrument but also extend its operating lifetime.
4. Insert the W lamp holder into the W lamp socket in the light source housing and tighten the two holding
screws (1).
5. If the performance of the instrument changes greatly after the replacement, the height of W lamp should be
adjusted (The horizontal position of W lamp is adjusted automatically when the self-testing is done upon
switching on the power.). The adjusting procedure is as follows:
a) Remove the W lamp holder.
b) Loosen the three screws (2) on the W lamp holder with a watch screwdriver.
c) Adjust the height of W lamp.
d) Tighten the three screws (2) .
e) Insert the W lamp holder back to the socket.
PRECAUTIONS:
• The light source and light source housing are considerably heated
during operation. Therefore, be sure to turn off the power switch so as
to cool down the light source sufficiently before replacing the light
source.
• The reflective mirror (G) should be covered with paper bag to avoid
being contaminated by fingers.
40