CALIBRATION OF pH METER

1) Switch on the Instrument from backside .

Display shows Note:- Before Puting pH


2) Press Cal Botton Dsiplay Shows Press Enter
put 7.0 pH solution, dry electrode
Std Y /N
carefully with tissue
paper.

Put 7.0 pH sol.

Press Enter

Display shows put Display Shows


4.0 pH 40 Second

Put 9.2 pH sol. Note:- Before Puting pH


To diplay 9.2 pH in Display
solution, dry electrode
place of 4.0 pH shows put
carefully with tissue
press cal botton. 9.2 pH
paper.

Press Enter

Display Shows
40 Second

Calibration of pH meter is Over.


3) Press pH botton and Enter for measure

Note:- 1) If proper solution & procedure is not used then display Shows ERR1 and calibration will fail.
2) For successful calibration, used proper solution & procedure.
CALIBRATION OF CONDUCTIVITY METER

1) Switch on the Instrument from backside .

Display shows Note:- Before Puting


2) Press Cal Botton Dsiplay Shows Press Enter
put solution solution, dry electrode
Std Y /N
carefully with tissue
paper.

Put 0.01 N
(KCL) sol.

Press
calibration
botton

Calibration of
Press Conductivity
Cond. meter is
botton.
Over.

Meter Shows cond.


Of 0.01N KCL
solution (1418 µS)

Note:- 1) If proper solution & procedure is not used then display Shows ERR1 and calibration will fail.
2) For successful calibration, used proper solution & procedure.
3) The 0.01N reading may be different from 1418.