Professional Documents
Culture Documents
"Quality Control Manual" Template For Manufacturers of Displays and Workstations Labeled For Final Interpretation in Full-Field Digital Mammography
"Quality Control Manual" Template For Manufacturers of Displays and Workstations Labeled For Final Interpretation in Full-Field Digital Mammography
"Quality Control Manual" Template For Manufacturers of Displays and Workstations Labeled For Final Interpretation in Full-Field Digital Mammography
Published by:
www.nema.org
© Copyright 2006 by the National Electrical Manufacturers Association. All rights including translation into
other languages, reserved under the Universal Copyright Convention, the Berne Convention for the
Protection of Literary and Artistic Works, and the International and Pan American Copyright Conventions.
NOTICE AND DISCLAIMER
The information in this publication was considered technically sound by the consensus of persons
engaged in the development and approval of the document at the time it was developed.
Consensus does not necessarily mean that there is unanimous agreement among every person
participating in the development of this document.
The National Electrical Manufacturers Association (NEMA) standards and guideline publications, of
which the document contained herein is one, are developed through a voluntary consensus
standards development process. This process brings together volunteers and/or seeks out the
views of persons who have an interest in the topic covered by this publication. While NEMA
administers the process and establishes rules to promote fairness in the development of
consensus, it does not write the document and it does not independently test, evaluate, or verify
the accuracy or completeness of any information or the soundness of any judgments contained in
its standards and guideline publications.
NEMA disclaims liability for any personal injury, property, or other damages of any nature
whatsoever, whether special, indirect, consequential, or compensatory, directly or indirectly
resulting from the publication, use of, application, or reliance on this document. NEMA disclaims
and makes no guaranty or warranty, express or implied, as to the accuracy or completeness of any
information published herein, and disclaims and makes no warranty that the information in this
document will fulfill any of your particular purposes or needs. NEMA does not undertake to
guarantee the performance of any individual manufacturer or seller’s products or services by virtue
of this standard or guide.
In publishing and making this document available, NEMA is not undertaking to render professional
or other services for or on behalf of any person or entity, nor is NEMA undertaking to perform any
duty owed by any person or entity to someone else. Anyone using this document should rely on
his or her own independent judgment or, as appropriate, seek the advice of a competent
professional in determining the exercise of reasonable care in any given circumstances.
Information and other standards on the topic covered by this publication may be available from
other sources, which the user may wish to consult for additional views or information not covered
by this publication.
NEMA has no power, nor does it undertake to police or enforce compliance with the contents of
this document. NEMA does not certify, test, or inspect products, designs, or installations for safety
or health purposes. Any certification or other statement of compliance with any health or safety–
related information in this document shall not be attributable to NEMA and is solely the
responsibility of the certifier or maker of the statement.
XR 22-2006
Page i
CONTENTS
Page
Foreword to the Device Manufacturer ..................................................................................................... iii
Foreword ...................................................................................................................................................... 1
Section 1 OVERVIEW OF THE QUALITY CONTROL MANUAL ............................................................... 2
1.1 Scope......................................................................................................................................... 2
1.2 Regulatory Considerations ........................................................................................................ 2
1.3 Structure of the Document......................................................................................................... 2
1.4 Alternative Standard on Use of Test Results ............................................................................ 2
Section 2 QC TESTS FOR THE RADIOLOGIC TECHNOLOGIST ............................................................ 4
2.1 Introduction ................................................................................................................................ 4
2.2 QC Intervals............................................................................................................................... 4
2.3 Display Screen Cleanliness....................................................................................................... 4
2.3.1 Frequency .....................................................................................................................4
2.3.2 Objective .......................................................................................................................4
2.3.3 Equipment Required .....................................................................................................4
2.3.4 Procedure......................................................................................................................4
2.3.5 Action Limit....................................................................................................................5
2.3.6 Use of Test Results.......................................................................................................5
2.4 Viewing Conditions Check for the Display................................................................................. 5
2.4.1 Frequency .....................................................................................................................5
2.4.2 Objective .......................................................................................................................5
2.4.3 Equipment Required .....................................................................................................5
2.4.4 Procedure......................................................................................................................5
2.4.5 Action Limit....................................................................................................................5
2.4.6 Use of Test Results.......................................................................................................5
2.5 Image Quality Check ................................................................................................................. 5
2.5.1 Frequency .....................................................................................................................5
2.5.2 Objective .......................................................................................................................6
2.5.3 Equipment Required .....................................................................................................6
2.5.4 Procedure......................................................................................................................6
2.5.5 Action Limit....................................................................................................................6
2.5.6 Use of Test Results.......................................................................................................7
2.6 Radiologic Technologist QC Test Results Record Forms......................................................... 7
2.6.1 Chart 1 – Daily Tests.....................................................................................................7
2.6.2 Chart 2 – Image Quality Check .....................................................................................7
Section 3 QC TESTS FOR THE MEDICAL PHYSICIST............................................................................. 8
3.1 Introduction ................................................................................................................................ 8
3.2 Viewing Conditions Check and Setting ..................................................................................... 8
3.2.1 Frequency .....................................................................................................................8
3.2.2 Objective .......................................................................................................................8
3.2.3 Equipment Required .....................................................................................................8
3.2.4 Procedure......................................................................................................................8
3.2.5 Action Limit....................................................................................................................9
3.2.6 Use of Test Results.......................................................................................................9
3.3 Display Luminance and Gray-Scale Calibration Check ............................................................ 9
3.3.1 Frequency .....................................................................................................................9
3.3.2 Objective .......................................................................................................................9
3.3.3 Equipment Required .....................................................................................................9
3.3.4 Procedure....................................................................................................................10
3.3.5 Action Limits ................................................................................................................10
3.3.6 Use of Test Results.....................................................................................................10
3.4 Image Quality Check ............................................................................................................... 10
When full-field digital mammography (FFDM) systems were first introduced, all components (e.g., image
receptor, acquisition workstation, diagnostic workstation, hardcopy output device) were provided or
qualified for use by the image receptor manufacturer (IRM). The IRM also provided a comprehensive
quality control (QC) plan to enable mammography facilities to meet their responsibilities under the Quality
Mammography Standards of the Mammography Quality Standards Act (MQSA). Subsequently the U.S.
Food and Drug Administration (FDA) approved manufacturers other than the IRMs to market hardcopy
and softcopy displays for FFDM images. This has made system QC more difficult since, under MQSA
regulations, the facility is required to follow a quality assurance program substantially the same as the
one recommended by the IRM. However, the QC plan of the IRM might not adequately address the
needs of components developed by other manufacturers.
This increasing heterogeneity of FFDM systems has created a desire to delegate the responsibility for
developing QC procedures for the individual system components to the manufacturers of those
components. However, it is also desirable, especially for the end-user, to retain some consistency in the
QC documentation that accompanies new FFDM components. The Mammography Group of the NEMA
X-ray Imaging Section has developed a template that provides both a consistent presentation format and
a minimum set of QC tests that should be included as part of the quality assurance plan of a display or
workstation labeled for final interpretation in an FFDM system.
Within this document, “display” refers to an electronic image presentation device with minimal image
processing capability. “Workstation” refers to an electronic device capable of sophisticated image
processing as well as image storage and retrieval. A workstation is expected to incorporate at least one
and most likely more than one display. The tests in this template primarily address the QC of the display
either as an independently operating device or as a component of a workstation.
Display and workstation manufacturers who follow this template when developing their quality assurance
programs and include at least the tests listed, will have incorporated the essential aspects of quality
assurance that the image receptor manufacturers have included in their recommended quality assurance
programs. This will ease the burden of the mammography facility to establish and maintain a quality
assurance program for the display or workstation substantially the same as the one recommended by the
IRM. The sections following this foreword to the device manufacturer constitute the content of the QC
template.
Each QC test includes a testing frequency, a set of procedure steps, action limits that define acceptable
outcomes for the test, and instructions on the use of the test results. Examples of each of these elements
are included in the template. In some cases, test conditions and action limits that have been found to be
appropriate and reasonable have been included in the QC tests.
However, it is recognized that it might not be practical or possible on all displays or workstations to
implement verbatim the tests as set forth in these examples. The content of this template may be
incorporated in whole or in part as deemed appropriate by the manufacturer of the display or workstation.
However, it is recommended that the manufacturer include in the QC plan a test for each concept
presented in this template. The manufacturer may alter the details of the tests to accommodate the
characteristics of a specific display or workstation or particular regulatory requirements.
Within the template are instructions and comments to the manufacturer using the template. These
instructions and comments are italicized and enclosed in brackets, e.g., [Manufacturer to …]. The
manufacturer of the display or workstation must insert the necessary information. In the context of this
document, the “manufacturer” is the entity seeking clearance from FDA to market the device. For
example, in the case of a workstation produced by one manufacturer but incorporating displays produced
by a second, the producer of the workstation would be considered the manufacturer referred to in the
template. However, if the display manufacturer also sells displays that can be incorporated into an
imaging network independent of a workstation, the display manufacturer would be the one to provide the
material for the template.
The manufacturer may also include additional tests beyond those specified in the template. Each such
test must include the elements and follow the form of the tests included in the template. The content of
the procedure must be written in regulatory language using auxiliary verbs such as “shall” or “must” or the
imperative mood. It may not include references to other documents or commercial products. Such
references as well as recommendations regarding test procedures may be included in a Guidance
section of the QC plan. An example of Guidance is included as the final section of the template. The
manufacturer should review the content of this Guidance section and provide material of a similar nature
appropriate to the manufacturer’s product.
Test patterns suitable for performing some of the tests in the template have been prepared by Task
Group #18 (TG 18) of the American Association of Physicists in Medicine and by the Society of Motion
Picture and Television Engineers (SMPTE). The TG 18 report and test patterns are available as “AAPM
Online Report #3: Assessment of Display Performance for Medical Imaging Systems” and Supplemental
Files, Imaging Informatics Subcommittee Task Group #18. These documents are available at
• http://www.aapm.org/pubs/reports
• http://www.aapm.org/pubs/reports/public/OR_03_Supplemental/
The SMPTE pattern is available as “SMPTE Recommended Practice RP 133-1991 - Specifications for
Medical Diagnostic Imaging Test Pattern for Television Monitors and Hard-Copy Recording Cameras” and
is available at
• http://www.smpte.org/smpte_store/standards/
and at
• http://brighamrad.harvard.edu/research/topics/vispercep/tutorial.html
It is recommended to include any required test patterns with the equipment for ease of use by the
operator and to ensure that the correct pattern is used for each test.
Foreword
Quality Control (QC) is important in any imaging system, but it is especially important in mammography.
MQSA1 regulations mandate that “each facility shall establish and maintain a quality assurance program
to ensure the safety, reliability, clarity, and accuracy of mammography services performed at the facility”
(21 CFR 900.12(d)). The steady movement from screen/film mammography to full-field digital
mammography (FFDM) makes system quality control more difficult because FFDM systems are
becoming increasingly heterogeneous. No longer are all of the system components (e.g., image receptor,
acquisition workstation, diagnostic workstation, displays, hardcopy output device) provided or qualified for
use by the image receptor manufacturer.
As a result, there is an increasing need for QC at the component level. This manual provides QC
procedures for the [Manufacturer to enter commercial name of device], a [Manufacturer to specify display
or workstation] that receives images from an FFDM system, and presents them for final interpretation.
The procedures in this manual represent a minimum set of QC tests that should be included as part of the
quality assurance plan of a display or workstation used for final interpretation in an FFDM system.
1
The United States Mammography Quality Standards Act, as amended by the Mammography Quality
Standards Reauthorization Acts of 1998 and 2004 (MQSRA).
Section 1
OVERVIEW OF THE QUALITY CONTROL MANUAL
1.1 SCOPE
This document defines the minimum set of Quality Control (QC) tests to be applied to the [Manufacturer
to enter commercial name], a [Manufacturer to specify display or workstation] labeled for final
interpretation of images acquired using a full-field digital mammography (FFDM) image acquisition
system. This document is intended to constitute an element of the Quality Assurance Plan (QAP) of the
mammographic facility.
Except for the Phantom Image Quality Test, Sec. 3.5, these tests are intended to monitor the
performance of the [Manufacturer to specify display or workstation] as a component of the full-field digital
mammography system and not the performance of the system as a whole. Hence, a characteristic such
as compatibility of the component with the communication protocols used in an imaging network is not an
element of these tests.
These tests are intended to monitor the consistency over time of the device’s performance. While they
may be included as elements of acceptance testing, they are not intended to represent the entirety of an
acceptance test.
The actions to be taken in regard to the QC plan for the mammography [Manufacturer to specify
display or workstation] are as follows:
(ii) If the test results for the image display system of the FDA-approved, full-field digital mammography
(FFDM) equipment fall outside of the action limits, the source of the problem shall be identified and
corrective actions shall be taken:
(A) Before any further mammographic images are reviewed using the component of the image display
system that failed any of the following tests:
(1) Display Screen Cleanliness
(2) Viewing Conditions Check for the Display
(3) Viewing Conditions Check and Setting
(4) Image Quality Check
(5) Phantom Image Quality Test
(B) Within 30 days of the test date for the following test:
(1) Display Luminance and Gray-Scale Calibration Check
Section 2
QC TESTS FOR THE RADIOLOGIC TECHNOLOGIST
2.1 INTRODUCTION
QC tests are simple checks which ensure that the [Manufacturer to specify display or workstation] is
operating to its design standards. They are intended to detect any changes in settings which might
compromise image quality, as well as any deterioration in equipment performance over time.
QC tests for the [Manufacturer to specify display or workstation] are described in the following sections:
• Section 2.3 Display Screen Cleanliness.
• Section 2.4 Viewing Conditions Check for the Display.
• Section 2.5 Image Quality Check.
Section 2.6 provides charts for use in recording the results of the various checks. It is recommended that
you use the chart pages in this manual to make copies for recording the results.
2.2 QC INTERVALS
The QC tests described in this section must be performed at least as frequently as specified in the test
descriptions and as summarized in Table 2-1.
Table 2-1
INTERVALS FOR PERFORMING QC TESTS
Refer to Section 4, Guidance, for additional information regarding application of the QC Tests.
2.3.2 Objective
To ensure good image review conditions by keeping the display screens free of dust, finger prints, and
other marks.
2.3.4 Procedure
1. Check the screen to verify that it is free from dust, finger prints, and other marks.
2. If dust, finger prints, or other marks are observed to be present, clean the display screens using a
cloth or cleaning tissue to remove dust, finger prints, and other marks.
3. Record completion of the procedure on Chart 1.
2.4.2 Objective
To ensure optimal viewing conditions.
2.4.4 Procedure
1. Review the data form “Viewing Conditions Check and Setting” posted in the reading room during the
most recent Medical Physicist’s evaluation. On this data form you will find the reading room
conditions established as optimal during this evaluation.
2. Compare the reading room configuration to the one specified in the data form.
3. If differences exist between the configuration specified and the actual one, adjust the room
configuration to match the configuration specified in the data form. For example, if desk lights were
specified to be turned off, turn them off, or, if curtains were specified to be closed, close them. For
additional information on maintaining the appropriate illuminance level, see Section 4.2.
4. Record completion of the check on Chart 1.
NOTE—On the “Viewing Conditions Check and Setting” data form posted by the Medical Physicist you will find the value of the
ambient illuminance. This is a record of the most recent measurement and provides a comparison value for the next Medical
Physicist survey. The Radiologic Technologist is not required to measure the ambient illuminance as part of the Viewing Conditions
Check.
2.5.2 Objective
To ensure that the display is calibrated for the reading of mammographic images.
2.5.4 Procedure
(It is recommended that the manufacturer base this test on comprehensive test patterns such as those
developed by the SMPTE and by TG-18 of the American Association of Physicists in Medicine. These
comprehensive patterns may be supplemented by other patterns specified by the equipment
manufacturer in order to test for all of the characteristics identified below. The manufacturer shall provide
a Procedure and Action Limits appropriate for the test patterns specified. Subsequent reference in this
section to “pattern” in the singular does not imply a constraint to use only a single test pattern to
accomplish the specified tests.)
1. View the test pattern on all displays used for final interpretation of mammograms.
NOTE—It is intended that the test pattern be viewed without modification of its size or gray scale. The pattern is displayed with
each pixel of the pattern represented by one pixel of the display. For the test to remain valid, magnification or zoom features of the
display must not be used when evaluating the appearance of the features described below. Additionally, the intensity levels used in
the pattern are intended to enable evaluation of the gray scale calibration of the display. Adjustment of contrast, brightness, window
width, window level, or other gray scale modifications must not be applied.
2. View the display from the position that would be used by a single viewer seated directly in front of the
display. The observer must be at least 50 cm from the display screen.
3. Examine each image on each display carefully for the following features.
a. Verify that there are no new artifacts or non-uniformities.
b. Verify that a continuous gray-scale ramp appears continuous and free from contrast reversal.
That is, along the length of the ramp, the brightness must not include dark-light-dark or light-dark-
light transitions.
c. Verify that borders and lines of the pattern are visible and straight and that the pattern appears to
be centered in the active area of the display.
d. Verify that the 0%-5% contrast pattern is visible.
e. Verify that the 95%-100% contrast pattern is visible.
f. Verify that each gray-level step of a discontinuous gray-scale ramp can be distinguished from the
adjacent steps.
g. Verify that the alphanumeric characters that appear on the pattern are sharp and in focus.
h. Verify that the high contrast line-pair images at the center and corners of the pattern are
distinguishable.
4. Record the results on Chart 2.
Year Month
Date 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
Initials
Display
cleanliness
Viewing
conditions
Year Month
Date 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
Initials
Display
cleanliness
Viewing
conditions
Date
Initials
Left Right
Display
(Pass/Fail) (Pass/Fail)
Artifacts, non-uniformities
Gray-scale continuity
0%-5% contrast
95%-100% contrast
Gray steps
Alphanumerics
Section 3
QC TESTS FOR THE MEDICAL PHYSICIST
3.1 INTRODUCTION
The QC tests listed in this section must be performed by the Medical Physicist to ensure that the
[Manufacturer to enter commercial name] [Manufacturer to specify display or workstation] provides a high
level of mammographic image quality. These tests also form the basis of the mammography equipment
evaluations (MEEs) that must be performed following the installation of a [Manufacturer to enter
commercial name] [Manufacturer to specify display or workstation], the repair or replacement of a major
component of the system, the relocation of the display to a new viewing environment, or substantial
change of the illumination of an existing viewing environment. Additional information regarding changes
of viewing conditions is included in Section 4.2. Additional information regarding the tests to be done for
MEEs can be found in Section 4.7.
NOTE—When performed as part of an MEE, all required tests must pass before the device may be used to interpret clinical images.
In addition to MEE testing, the QC tests described in this section must be performed at least as frequently
as specified in the test descriptions and as summarized in Table 3-1. These tests are described in
Sections 3.2, 3.3, 3.4, and 3.5. Charts for recording the results of the QC tests or MEEs are provided in
Section 3.6.
Table 3-1
INTERVALS FOR PERFORMING QC TESTS
3.2.2 Objective
To ensure optimal viewing conditions.
3.2.4 Procedure
1. Place the displays in optimal positions in the room. See Section 4.4.
2. Reduce any lighting coming from sources such as viewboxes, or overhead, or task lighting to the
minimum, to reproduce typical ambient lighting conditions that would be set for reading digital images
on the display.
3. Check that the ambient lighting set at Step 2 is stable (no flickering effect due to any of the lighting
sources that are turned on).
4. Make each display as dark as possible. For example, minimize the application windows or display a
black, flat field.
5. Obtain the illuminance reading at each display. Position the photometer with its sensitive area facing
away from the surface of the display and located at the center of the display and on the display
surface.
NOTE—The medical physicist must take care not to block any light that could be detected by the photometer.
6. Read the illuminance value and verify that it does not exceed the value specified in the Action Limit.
7. If the value is greater than the specified limit, adjust the room ambient light and/or the room
configuration in such a way that the illuminance does not exceed that limit.
8. Record a description of the room configuration established as optimal in the data form as detailed in
Steps 9 through 11.
9. Fill out a check-list of the condition of major elements in the reading room (for example, display
positions, room lights, desk lights, curtains).
10. Record the illuminance on the data form.
11. Draw a room layout on the data form sufficiently accurate to enable the Radiologic Technologist to
reproduce the illumination conditions as they were at the time the illuminance level was set.
12. Post the form in the reading room, so that it will be readily available for the Radiologic Technologist’s
daily “Viewing Conditions Check.”
13. Restore each display to its normal condition, i.e., undo any measures taken in Step 4 to darken the
display.
3.3.2 Objective
To ensure that each display is calibrated for the reading of mammographic images.
documentation the user must include in the QC record. This could include, for example, the date of the
last calibration or intercomparison.]
3.3.4 Procedure
[This test is intended to check the consistency of the black and white levels and the gray-scale display
function. Additional tests, e.g., luminance uniformity or artifact evaluation, may also be included. The
manufacturer must provide procedure steps and action limits appropriate for the display hardware and
software used in the display or workstation. The procedure must have the capability to achieve the
manufacturer’s specified level of accuracy for the luminance levels independent of confounding factors
such as variation of brightness uniformity or photometer calibration.]
3.4.2 Objective
To ensure the display is calibrated for the reading of mammographic images.
3.4.4 Procedure
Complete Section 3.2, Viewing Conditions Check and Setting, and Section 3.3, Display Luminance and
Gray-Scale Calibration Check, before performing this test.
The remainder of this test is the same as the Radiologic Technologist’s test “Image Quality Check,” Sec.
2.5.
NOTE—It is intended that the test pattern be viewed without modification of its size or gray scale. The pattern is displayed with
each pixel of the pattern represented by one pixel of the display. For the test to remain valid, magnification or zoom features of the
display must not be used when evaluating the appearance of the features described below. Additionally, the intensity levels used in
the pattern are intended to enable evaluation of the gray scale calibration of the display. Adjustment of contrast, brightness, window
width, window level, or other gray scale modifications must not be applied.
3.5.2 Objective
To ensure that the quality of images acquired at the acquisition work station remains adequate after the
images are transmitted to and presented on the display.
3.5.4 Procedure
1. Acquire an image of the mammographic phantom at the acquisition workstation using the procedure
specified in the quality assurance plan of the image receptor manufacturer.
2. Apply the same image processing to the image as applied to a phantom image that would be
analyzed at the acquisition workstation.
3. Send the image to the display and present it.
4. Score the phantom image using the procedures specified by the image receptor manufacturer for
scoring a phantom image displayed on the acquisition workstation.
Serial Numbers
Mammographic Unit: __________________________________________
Diagnostic Workstation: __________________________________________
Signature: __________________________________________________________
Facility: _________________________________________________________
Room: __________________________________ Date: ________________
Room description
Display positions: ________________________________________________
_____________________________________________________________
Room lights: _____________________________________________________
_____________________________________________________________
Desk Lights: _____________________________________________________
_____________________________________________________________
Others: __________________________________________________________
______________________________________________________________
Room layout
[Manufacturer to provide a table for recording the data acquired in the procedure and to indicate by
means of Pass/Fail checkboxes whether or not the results meet the action limits. An example follows.]
Facility: _____________________________________________________________
Room: ____________________________________ Date: _________________
Luminance Photometer: Model: ____________________ Ser. No. __________
NOTE—Use the display serial number if the photometer is internal to the display or does not have its own serial number.
[Manufacturer to provide a table for recording the data acquired in the procedure and to indicate by
means of Pass/Fail checkboxes whether or not the results meet the action limits. An example follows.]
Facility: _________________________________________________________
Room: __________________________________ Date: ________________
Gray-scale continuity
0%-5% contrast
95%-100% contrast
Gray steps
Alphanumerics
[Manufacturer to provide a table for recording the data acquired in the procedure and to indicate by
means of Pass/Fail checkboxes whether or not the results meet the action limits. An example follows.]
Facility: _________________________________________________________
Room: __________________________________ Date: ________________
Requirement: The score of the image presented on the display must meet the respective action limits set
by the image receptor manufacturer of each FFDM system that will send images to the display for final
interpretation.
Section 4
GUIDANCE
MQSA final regulations consider performance of the image receptor manufacturer’s QC Test procedures
to be a requirement for compliance with the regulations. This Guidance section is not considered an
element of those regulations. It is intended to provide guidance to mammography facilities and their
personnel. It represents the equipment manufacturer’s current thinking on appropriate procedures for
conducting the QC tests. Procedures described in guidance represent supplementary information on
acceptable ways of doing a test but, unlike regulations, do not bind the facility to using that way only. An
alternative procedure may be used if such a procedure satisfies the requirements of the applicable
statute, regulations, or both. Mandatory language, such as “shall,” “must,” and “require,” is used when
referring to statutory or regulatory requirements. Non-mandatory language, such as “should,” “may,”
“can,” and “recommend” is used when referring to guidance. It is the responsibility of the facility to read,
understand, and follow the final regulations.
Under its own authority, a state may impose more stringent requirements beyond those specified under
MQSA and its implementing regulations. A facility may want to check with the state or local authorities
regarding their requirements. Tests may be performed more frequently than specified in the QC tests if
required by local regulations or hospital requirements.
QC tests may be used as a check for correct operation, for example, after a change of operating
parameters.
The following sections provide specific guidance for sections in the QC tests.
obtain the optimum ambient illuminance, it is recommended to check artificial lighting, the adjustment of
variable lighting, natural lighting, and main lighting (curtains or door openings, environmental conditions).
If the system fails the Viewing Conditions Check and Setting, it is recommended that you contact your
field service engineer.
The MQSA Quality Mammography Standard on MEEs [21 CFR 900.12(e)(10)] states that when “major”
repairs are performed on existing equipment, an MEE must be performed by a medical physicist or under
the direct supervision of a medical physicist, i.e., the medical physicist must be in the room during the
performance of the individual equipment tests. In its guidance, the FDA has classified particular repairs
as being “major,” and for repairs not considered major (and thus not requiring an MEE) has identified the
level of medical physicist involvement required. Table 4-1 provides similar information for repairs to the
[Manufacturer to specify display or workstation]. The results of the MEE may be recorded in the charts in
Section 3.6.
[Manufacturer to review Table 4-1 and provide content appropriate to the display workstation.]
FDA Guidance states “there are adjustments, changes, or repairs for which verification ... can be
performed by other qualified personnel (e.g., radiologic technologist or service representative with
appropriate training/experience) without involving the medical physicist.” Generally, in the event of a
system repair or upgrade, if actions supplemental to the procedures in this manual are required to be
taken by “other qualified personnel,” those actions will be specified in documentation to be provided by
the field service engineer. However, the calibration of the display—setting the black and white levels and
performing the grayscale display function calibration—is intended to remain the responsibility of the field
service engineer.
Table 4-1
MEDICAL PHYSICIST INVOLVEMENT IN DISPLAY OR WORKSTATION REPAIRS