VPAP™ IV & VPAP™ IV ST

Quick Setup Guide for Clinicians


English

To access the Clinical Menu: To access the Review menu: To access the Mask Fit function:
Hold down and simultaneously Hold down and simultaneously Hold down for at least three
for at least three seconds. for at least three seconds. seconds.
Choose Exit to return to the Patient Press to return to the Patient
menu. menu.

Treatment Screens Patient Menu


After starting therapy an introductory treatment screen will
display. Press or to switch between views.
WELCOME
>>>>>>

Treatment Screen 1
Mode Set EPAP RAMP: 10min
menu
pressure
Trigger indicator S:RAMP 4.0-10.0
Standard Menu Advanced Menu
S>>C PS 6.0 Set IPAP/CPAP
Pressure bar pressure
graph MASK: ULTRA SETTINGS
change exit enter exit
Treatment Screen 2 MASK: ULTRA
change exit

Leak LK: 10L/min RR: 15


Respiratory USED: 00090hrs TUBE LENGTH: 2m

MV: 10.2 VT: 680 rate 00006/00012 days change exit

Minute ventilation SMARTSTART: OFF

Tidal volume SW: SX459-0200


exit
exit

LEAK ALERT: OFF


Treatment Screen 3 Cycle indicator change exit

Trigger indicator (TiMn/C/TiMx) RESULTS


enter exit
USED: 00090hrs
(S or T) S**TiMn TiMx 2.0 Set Ti Max 00003/00122days

Ti 1.5s 1: 1.6
Calculated SMART DATA

inspiration time Calculated I:E These menus only appear if


enter exit
MASK FIT: *****
ratio at least one Smart Data item
excelnt exit

is enabled in the Clinical PRESSURE: 8.0


Trigger indicator: S indicates a patient triggered breath. T Menu.
exit

USAGE: 4.00hrs
indicates a device tiggered breath. exit
OPTIONS
Cycle indicator: C indicates a patient cycled breath. TiMn enter exit
SMART DATA
indicates a TiMin cycled breath. TiMx indicates a TiMax enter exit
AUTO APPEAR: ON
change exit
cycled breath.
LANGUAGE:ENGLISH
change exit
SERVICING
TiControl (S and ST modes) enter exit
SN: 20080102356
exit

You can set the minimum and maximum time limits that the SW: SX459-0200

device spends in IPAP. exit

Patient Ttot = I:E = 1:2 I:E = 1:1 I:E = 1:3


Breath TiControl
60 ÷ BPM (Reference) (eg, restrictive) (eg, COPD)
(BPM) (sec) Ti Max
Ti Min Ti Max Ti Max Ti Min
Cycle
10 6 2 1.0 2.0 1.5 Window
15 4 1.3 1.0 2.0 1.3
20 3 1.0 0.8 1.5 1.0
25 2.4 0.8 0.7 1.2 0.8
Patient
30 2 0.7 0.6 1.0 0.7 Flow
35 1.7 0.6 0.5 0.8 0.7
40 1.5 0.5 0.5 0.7 0.7
Clinical Menu

WELCOME
>>>>>

RAMP: 10 min
menu

CLINICAL REVIEW
menu exit menu exit

SETTINGS SETTINGS
enter exit MODE: CPAP MODE: S MODE: ST MODE: T enter exit
change exit change exit change exit change exit

CPAP: 10.0 IPAP: 10.0 IPAP: 10.0 IPAP: 10.0


change exit change exit change exit change exit ADV SETTINGS
enter exit
SMARTSTART: ON EPAP: 4.0 EPAP: 4.0 EPAP: 4.0
change exit change exit change exit change exit

MASK: ULTRA RESP RATE: OFF RESP RATE: 10bpm RESP RATE: 10bpm
change exit change exit change exit change exit RESULTS
enter exit
RISE TIME: 300ms RISE TIME: 150ms RISE TIME: 150ms
change exit change exit change exit

Ti Max: 2.0s Ti Max: 2.0s Ti: 2.0s


change exit change exit change exit

Ti Min: 0.3s Ti Min: 0.3s SMARTSTART: ON


change exit change exit change exit

SMARTSTART: ON SMARTSTART: ON MASK: ULTRA


change exit change exit change exit

MASK: ULTRA MASK: ULTRA


change exit change exit
ADV SETTINGS
enter exit LEAK ALERT: OFF LEAK ALERT: OFF LEAK ALERT: OFF LEAK ALERT: OFF
change exit change exit change exit change exit

TUBE LENGTH: 2m TRIGGER: MED TUBE LENGTH:


TRIGGER: MED 2m TUBE LENGTH: 2m
change exit change exit change exit change exit

MAX RAMP: 45min CYCLE: MED MAX RAMP:


CYCLE: MED35min MAX RAMP: 45min
change exit change exit change exit change exit

START CPAP: 4.0 TUBE LENGTH: 2m TUBE LENGTH:


START 2m
EPAP: 4.0 START EPAP: 4.0
change exit change exit change exit change exit

MAX RAMP: 45min MAX RAMP: 45min


change exit change exit

START EPAP: 4.0 START EPAP: 4.0


RESULTS change exit change exit
enter exit
EFFICACY DATA LEAK: 12L/min LEAK: 13L/min LEAK: 14L/min LEAK: 13L/min
enter exit wk... exit mth... exit 6mths... exit yr... exit

VT: 0-245 VT: 0-249 VT: 0-248 VT: 0-249


wk... exit mth... exit 6mths... exit yr... exit

RESP RATE: 0-20 RESP RATE: 0-19 RESP RATE: 0-20 RESP RATE: 0-20
wk... exit mth... exit 6mths... exit yr... exit

MV: 0.0-11.7 MV: 0.0-11.8 MV: 0.0-11.7 MV: 0.0-11.8


wk... exit mth... exit 6mths... exit yr... exit

PRESS: 5.1 PRESS: 5.2 PRESS: 5.1 PRESS: 5.0


wk... exit mth... exit 6mths... exit yr... exit

AHI: 13 AHI: 12 AHI: 14 AHI: 13


wk... exit mth... exit 6mths... exit yr... exit

AI: 9 AI: 10 AI: 11 AI: 10


wk... exit mth... exit 6mths... exit yr... exit

%SPONT T: 70 %SPONT T: 70 %SPONT T: 70 %SPONT T: 70


wk... exit mth... exit 6mths... exit yr... exit

%SPONT C: 70 %SPONT C: 70 %SPONT C: 70 %SPONT C: 70


wk... exit mth... exit 6mths... exit yr... exit

USAGE DATA USAGE: 7:05hrs USAGE: 6.55hrs USAGE: 7:15hrs USAGE: 7:20hrs
enter exit wk... exit
OPTIONS mth... exit 6mths... exit yr... exit
enter exit
SMART DATA
MASK FIT: OFF
enter exit
change exit
REMINDERS PRESSURE: OFF
enter exit change exit
PRESSURE: cmH20
USAGE: OFF
change exit change exit
FACTORY DEFAULTS AUTO APPEAR: OFF
reset exit change exit
ERASE DATA
erase exit
DATE: 01 Jul 2008
change exit

TIME: 15:54
change exit
MENU TYPE: STD
change exit
LANGUAGE:ENGLISH
SERVICING
change exit
enter exit
RUN HRS: 12345
exit

SN: 20080102356
exit

SW: SX459-0200
exit

BAUD RATE: 9600


change exit

See the VPAP IV and VPAP IV ST User and Clinical Guides for full details.
©2008 ResMed Ltd. 268197/2 08 08