Professional Documents
Culture Documents
Hazardous Gas Monitors
Hazardous Gas Monitors
Chapter 1
Introduction
T
he detection of hazardous gases has always been
a complex subject and makes choosing an appro-
priate gas monitoring instrument a difficult task.
To address this problem, this book aims to provide
the following essential tools:
1
Hazardous Gas Monitors
2
Chapter 1 Introduction
Gas Sensors
A gas sensor is a transducer1 that detects gas mol- 1
Any device that converts input en-
ecules and which produces an electrical signal with a ergy of one form to output energy
magnitude proportional to the concentration of the gas. of another.
Unlike other types of measurement, types that are Gas Molecules
3
Hazardous Gas Monitors
4
Chapter 1 Introduction
5
Hazardous Gas Monitors
UEL
COMBUSTIBLE MIXTURE
LEL
0%
Fig. 1. A Window of Combustibility
6
Chapter 1 Introduction
7
Hazardous Gas Monitors
8
Chapter 1 Introduction
9
Hazardous Gas Monitors
CEILING
concentrations.
STEL Figure 4 illustrates the overall sum-
mary of the toxic gas safety concept.
PEL
TWA Combinations of Substances. When
0 1 2 3 4 5 6 7 8 9
two or more hazardous substances
HOURS are present in the area, the additive
Fig. 4. Basic Rules for Compliance: TWA for eight
effects should be considered. That
working hours does not exceed PEL/TLV. Observe the is, if the sum of the fractions of haz-
definition of STEL and never exceed the ceiling limit. ardous substance concentrations di-
vided by the respective PEL exceeds
unity, then the threshold limit of the mixture must
be considered excessive. This case is illustrated in
the following formula:
C1 C2 C C
+ + 3 + ... + n < 1
PEL1 PEL2 PEL3 PELn
If it is greater than one, then the PEL is exceeded.
C1,C2 and C3 represent the TWA of various hazardous
substances.
For example, if the ambient air contains 35 ppm
of carbon monoxide (PEL 50 ppm) and 350 ppm of
carbon dioxide (PEL 5000 ppm), the calculation is as
follows:
35 350
+ = 0.77
50 5000
The threshold limit is not exceeded.
This discussion is for conceptual purposes only. It does
not take into account the effect of the combining of
chemicals that can react with each other, resulting in
a final mixture that can be more toxic to humans than
the individual toxicities of each gas.
Performance Specifications
Accuracy: Webster’s dictionary defines accuracy as
10
Chapter 1 Introduction
11
Hazardous Gas Monitors
12
Chapter 1 Introduction
13
Hazardous Gas Monitors
14
Chapter 1 Introduction
100%
Scale
50%
0% Time
T80 T100
Fig.6. A Typical Sensor Response
15
Hazardous Gas Monitors
Hazardous Locations
Gas monitoring instruments are often installed in in-
dustrial process and production areas. These areas are
often classified as hazardous locations. Industrial fa-
cilities in which potentially explosive gas atmospheres
exist or may exist must utilize proper explosion proof
protection methods when using these types of instru-
ments. It is beyond the scope of this book to provide
full details; however, the following information may
be helpful. The reader is advised to consult with the
manufacturer for specific needs.
North America and other parts of the world that
16
Chapter 1 Introduction
17
Hazardous Gas Monitors
18
Chapter 1 Introduction
19
Hazardous Gas Monitors
20
Chapter 1 Introduction
21
Hazardous Gas Monitors
22
Chapter 1 Introduction
23
Hazardous Gas Monitors
Summary
There is no clear definition delineating a gas moni-
tor from an analytical analyzer. The distinction be-
tween the two is based largely upon their usage in the
actual application. A gas monitor is most frequently
used to monitor gases in toxic and combustible ranges
for area air quality and safety applications. For this
type of safety application, the concept of the units used
in the measurements and the definition of terminolo-
gies is somewhat unique. It is important to understand
the terms used. However, the examples of chemical
toxicity presented in this book do not consider the
complexity of actual toxicology and interreactions be-
tween contaminant chemicals. In critical applications,
a specialist on the subject needs to be consulted.
24
Chapter 2 Electrochemical Sensors
Chapter 2
Electrochemical
Sensors
T
he oldest electrochemical sensors date back to
the 1950s and were used for oxygen monitoring.
More recently, as the Occupational Safety and
Health Administration (OSHA) began requiring the
monitoring of toxic and combustible gases in confined
space applications, new and better electrochemical
sensors have been developed.
By the mid-1980s, miniaturized electrochemical
sensors became available for detection of many dif- Fig. 1. Electrochemical Sensors
ferent toxic gases in PEL ranges, with the sensors ex-
hibiting good sensitivity and selectivity. Currently, a
variety of electrochemical sensors are being used ex-
tensively in many stationary and portable applications
for personal safety. Figure 1 shows a small collection
of such electrochemical sensors.
The physical size, geometry, selection of various
components, and the construction of an electrochemi-
cal sensor usually depends on its intended use. Quite
often, the final design results in a compromise between
various performance parameters of the electrochemi-
cal sensor. The most common misconception about
electrochemical sensors is that they are all the same.
In fact, the appearance of the electrochemical sen-
sors used to detect various gases may be similar, but
their functions are markedly different. Consequently,
27
Hazardous Gas Monitors
Principle of Operation
Capillary
Diffusion Electrochemical sensors operate by reacting with
Barrier Hydrophobic the gas of interest and producing an elec-
Membrane
trical signal proportional to the gas concen-
Sensing Electrode
tration. A typical electrochemical sensor con-
Reference Electrode
sists of a sensing electrode (or working elec-
Counter Electrode
trode), and a counter electrode separated by a
Electrolyte thin layer of electrolyte, Figure 3.
Fig. 3 Typical Electrochemical Sensor Setup Gas that comes in contact with the sensor
28
Chapter 2 Electrochemical Sensors
29
Hazardous Gas Monitors
In basic electrolyte
Oxidation at the anode: 2Pb + 4OH- → 2PbO + 2H2O + 4e-
Reduction at the cathode: O2 + 2H2O + 4e- → 4OH-
Major Components
An electrochemical sensor consists of the following ma-
jor components:
A. Gas Permeable Membrane (also called hydrophobic
membrane): This is used to cover the sensor’s sens-
ing (catalyst) electrode and, in some instances, to
control the amount of gas molecules reaching the
electrode surface. Such barriers are typically made
of thin, low-porosity Teflon membranes. Such sen-
sors are called membrane clad sensors. Alternatively,
the sensing electrode is covered with a high-poros-
ity Teflon and the amount of gas molecules reach-
ing the electrode surface is controlled by a capil-
30
Chapter 2 Electrochemical Sensors
31
Hazardous Gas Monitors
32
Chapter 2 Electrochemical Sensors
Characteristics
There are many different ways that electrochemi-
cal sensors are constructed, depending both on the
gas to be detected as well as the manufacturer. How-
ever, the main characteristics of the sensors are es-
sentially very similar. Following are some of the com-
mon characteristics of electrochemical sensors:
1. With a three-electrode sensor, there is nor-
mally a jumper which connects the working
and reference electrodes. If it is removed dur-
ing storage, it will take a long time for the sen-
sor to stabilize and be ready to be used. Some
sensors require a bias voltage between the elec-
trodes, and in such cases, the sensors are
shipped from the factory with a nine-volt bat-
tery powered electronic circuit. It takes any-
where from thirty minutes to twenty-four hours
for the sensor to stabilize, and it will continue
to stabilize over a three-week period.
When installed in a portable or stationary
instrument, the sensor cannot be removed
from power for an appreciable amount of time.
It is wise to double-check the instrument be-
fore use if batteries or power were removed at
some point. The portable instrument’s cir-
cuitry provides a small current needed to
maintain the sensor in the ready-to-use condi-
tion, even if the instrument is turned off.
Two-electrode sensors do not require any
bias voltage. For example, oxygen sensors do
not require a bias voltage.
2. Most of the toxic gas sensors require a small
amount of oxygen to function properly. There
is a vent hole on the side or back of the sensor
for this purpose. It is wise to double-check with
the manufacturer in applications that use non-
oxygen background gas.
33
Hazardous Gas Monitors
34
Chapter 2 Electrochemical Sensors
The higher the ratio, the less the effect of inter- Range: 2-10 times permissible exposure limit
ference gas on the sensor. The interference data are Life Expectancy: 12 to 24 months normal;
taken using relatively low gas concentrations. In ac- depends on manufacturer and sensor
Summary
Electrochemical sensors require very little power
to operate. In fact, their power consumption is the
lowest among all sensor types available for gas moni-
toring. For this reason, the sensors are widely used in
portable instruments that contain multiple sensors.
They are the most popular sensors in confined space
applications.
A sensor’s life expectancy is predicted by its manu-
facturer under conditions that are considered normal.
However, the life expectancy of the sensor is highly
dependent on the environmental contaminants, tem-
perature, and humidity to which it is exposed.
35
Chapter 3 Catalytic Combustible Gas Sensors
Chapter 3
Catalytic Combustible
Gas Sensors
C
atalytic bead sensors are used primarily to de-
tect combustible gases. They have been in use
for more than 50 years. Initially, these sensors
were used for monitoring gas in coal mines, where
they replaced canaries that had been used for a long
period of time.
The sensor itself is quite simple in design and is
easy to manufacture. In its simplest form, as used in
the original design, it was comprised of a single plati-
num wire. Catalytic bead sensors were produced all
over the world by a large number of different manu-
facturers, but the performance and reliability of these Fig. 1 A Catalytic Bead Sensor
sensors varied widely among these various manufac-
turers. A catalytic bead sensor is shown in Figure 1.
Principle of Operation
Combustible gas mixtures will not burn until they
reach an ignition temperature. However, in the pres-
ence of certain chemical media, the gas will start to
burn or ignite at lower temperatures. This phenom-
enon is known as a catalytic combustion. Most metal
oxides and their compounds have these catalytic prop-
erties. For instance, volcanic rock, which is comprised
of various metal oxides, is often placed in gas burning
fireplaces. This is not only decorative, but it also helps
37
Hazardous Gas Monitors
D.C. POWER
RB
nc
Fig. 2 A catalytic bead sensor of English physicist and inventor Sir Charles Wheat-
Wheatstone bridge–a circuit for
measuring an unknown resist- stone (1802-75). Wheatstone bridges are commonly
ance by comparing it with known used in many electrical measurement circuits. As
resistances. shown in Figure 2, four circuit branches are arranged
38
Chapter 3 Catalytic Combustible Gas Sensors
39
Hazardous Gas Monitors
PLATINUM WIRE
CATALYTICALLY-TREATED
METAL OXIDE
40
Chapter 3 Catalytic Combustible Gas Sensors
Characteristics
The sensor’s output is directly in proportion to
the rate of oxidation. The maximum output of the
signal occurs at about the stoichiometric1 mixture of 1
Pertaining to substances that are in the
the gas, or it is based on the theoretical combustion exact proportions required for a given re-
reaction formula. Methane, for example: action.
41
Hazardous Gas Monitors
LEL UEL
10
8
Sensor Output
6
in volts
0
5 10 15 20 40 60 80 100
% Methane Gas Concentration
42
Chapter 3 Catalytic Combustible Gas Sensors
43
Hazardous Gas Monitors
44
Chapter 3 Catalytic Combustible Gas Sensors
45
Chapter 4 Solid-State Gas Sensors
Chapter 4
W
hen scientists were doing research work re-
lated to semiconductor p-n junctions,1 they
discovered that these junctions were sensitive
to environmental background gases.
At that time, such a behavior was considered a
problem. This problem, however, was solved by en-
capsulating the semiconductor chip so that it was no
longer exposed to the outside environment. Subse-
quently, unsuccessful attempts were made to utilize
the sensitivity of the semiconductor junction as a gas
detection device.
It wasn’t until 1968 that Mr. N. Taguchi marketed Fig. 1 A solid-state sensor
a simple semiconductor, or a solid-state sensor, for the used for detecting more than a
detection of hydrocarbons in LEL combustible ranges. hundred toxic gases.
The intention was to provide an alternative to the
1
popular catalytic bead sensor, which suffered from The positive and negative junctions
several problems, including loss of sensitivity with time in a semiconductor.
due to poisoning and burning out when exposed to
high gas concentrations.
In 1972, International Sensor Technology (IST) in
Irvine, California introduced a solid-state sensor for
the detection of hydrogen sulfide in a range of 0-10
ppm. A few years later, IST developed solid-state sen-
sors for the detection of more than 100 different haz-
ardous gases at low ppm levels. This was a significant
development, since OSHA was being formed at about
47
Hazardous Gas Monitors
Principle of Operation
A solid-state sensor consists of one or more metal
48
Chapter 4 Solid-State Gas Sensors
49
Hazardous Gas Monitors
Characteristics
Silicon Substrate Solid-state sensors are among the
Heater Resistor most versatile of all sensors, as they de-
tect a wide variety of gases, and can be
Fig. 3 Chip-type Sensor
used in many different applications. Dif-
ferent response characteristics are achieved by vary-
ing the semiconductor materials, processing tech-
niques, and sensor operating temperature.
Among the unique attributes of the solid-state sen-
sor are the abilities of the sensor to detect both low
ppm levels of gases, as well as high combustible levels.
Longevity. The main strength of the solid-state
sensor is its long life expectancy, as the sensor typi-
cally lasts 10 years or more in clean applications. This
is a major advantage compared to other sensor types,
such as catalytic bead or electrochemical sensors,
which typically last only one to two years.
However, while solid-state sensors have a longer
life expectancy, they are also more susceptible to in-
terference gases than the other types of sensors. Thus,
in applications where other background gases are
present, solid-state sensors may trigger false alarms.
In certain instances, the interferences from other
gases are minimized by using appropriate filtering ma-
terials that absorb all other gases except the gas to be
detected.
For example, a solid-state sensor for monitoring
carbon monoxide and hydrogen can be equipped with
a charcoal filter which eliminates the majority of in-
terfering gases. This way the sensor performs very well
and becomes very selective for those two gases.
Versatility. The versatility of the solid-state sensor
50
Chapter 4 Solid-State Gas Sensors
51
Hazardous Gas Monitors
52
Chapter 4 Solid-State Gas Sensors
53
Chapter 5 Infrared Gas Sensors
Chapter 5
I
nfrared (IR) gas detection is a well-developed
measurement technology. Infrared gas analyzers
have a reputation for being complicated, cumber-
some, and expensive. However, recent technical ad-
vancements, including the availability of powerful am-
plifiers and associated electronic components, have
opened a new frontier for infrared gas analysis. These
advancements have resulted from an increase in de-
mand in the commercial sector, and these demands
will likely continue to nourish the advancement of this Gas Cell
technology.
Gases to be detected are often corrosive and reac-
Fig. 1 An example of an IR
tive. With most sensor types, the sensor itself is directly gas monitor with the gas cell
exposed to the gas, often causing the sensor to drift assembly exposed.
or die prematurely.
The main advantage of IR instruments is that the
detector does not directly interact with the gas (or
gases) to be detected. The major functional compo-
nents of the analyzer are protected with optical parts.
In other words, gas molecules interact only with a light
beam. Only the sample cell and related components
are directly exposed to the gas sample stream. These
components can be treated, making them resistant to
corrosion, and can be designed such that they are eas-
ily removable for maintenance or replacement.
Today, many IR instruments are available for a wide
variety of applications. Many of them offer simple,
55
Hazardous Gas Monitors
Principle of Operation
The infrared detection principle incorporates only a
small portion of the very wide electromagnetic spectrum.
The portion used is that which we can feel as heat. This
is the region close to the visible region of the spectrum
to which our eyes are sensitive. Electromagnetic radia-
tion travels at close to 3 x 108 m/sec and has a wave-like
profile. Let’s review the basic physics of electromagnetic
radiation by defining the terminology involved with it.
Wave: Similar to a wave in the ocean, the electro-
magnetic radiation waves oscillate, one wave followed
by another. There are both electromagnetic and me-
chanical waves, with mechanical waves having a much
longer wavelength. Figure 2 illustrates a mechanical wave.
1 cm 2 cm
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
1 sec 2 sec
Frequency = 10 Hz. Wavelength = 0.1 cm, Wave Number = 10 cm -1
56
Chapter 5 Infrared Gas Sensors
57
Hazardous Gas Monitors
Ultra-
Radio Television Microwave Infrared
violet
X-rays Gamma rays
6 9 12 15 18 20
10 10 10 10 10 10
4 x 1014 8 x 1014
Frequency (cycles/sec.)
Fig. 3 Location of Infrared in the Electromagnetic Spectrum
58
Chapter 5 Infrared Gas Sensors
METHANE CH4
CH4
59
Hazardous Gas Monitors
60
Chapter 5 Infrared Gas Sensors
61
Hazardous Gas Monitors
62
Chapter 5 Infrared Gas Sensors
63
Hazardous Gas Monitors
64
Chapter 5 Infrared Gas Sensors
100%
Peak Wavelength
40% Bandwidth
Center Wavelength
20%
0%
3.0 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 4.0
Wavelength (µm)
Fig. 7 Methane Filter Specifications
65
Hazardous Gas Monitors
Configuration
There are a number of ways by which various IR
components can be arranged to produce a gas ana-
lyzer. The design may be relatively simple, or very
complicated, using many different optical components
depending on the type of analyzers for the applica-
tions. For applications that require high sensitivity, se-
lectivity, and stability, the design of the analyzer is more
complex. For applications where the selectivity and
66
Chapter 5 Infrared Gas Sensors
MIRROR
flashing IR sources are re- IR Source
flected back to the detec-
Reference
tors. The active detector Filter/Detector
has a filter for the target
Fig. 9 A Two-Detector Layout
gas, while the reference de-
tector has a filter with a different wavelength. In other
words, the active detector is used to detect the target
gas and the reference detector is used to ignore the
target gas. In actual operation, the reference detec-
tor provides a base point value or zero point while the
active detector is used to provide the signal; with the
differential between the two detectors providing the
actual span value of the instrument.
This arrangement offers the advantage of compen-
sating for the changes that occur in the detector’s sen-
sitivity with time. For instance, the intensity of the light
source can change with time due to contamination,
which will create a zero drift. The two-detector arrange-
ment minimizes this type of drift. Also, in this arrange-
ment, the path length is doubled which leads to higher
signal strength.
Figure 10 illustrates another popular design. This
design uses two tubes or cells. One is a reference cell
67
Hazardous Gas Monitors
MIRROR
SAMPLE CELL
MIRROR
REFERENCE CELL
IR Source
Characteristics
1. Temperature: An IR detector is essentially a
temperature sensor and is, therefore, potentially very
sensitive to changes in the ambient temperature. How-
ever, a properly designed detector can be operated
between -40°C to 60°C without being susceptible to
ambient temperature fluctuations. Most detectors do
not react well to sudden temperature variations. The
instrument typically requires 10 to 20 minutes to
achieve temperature equilibrium. For outdoor appli-
cations, this usually is not a serious problem because
the ambient temperature changes only slowly. Gener-
ally, the detector unit is operated at a temperature
slightly above the surrounding temperature to prevent
condensation. Water vapor condensation at the op-
tics and/or on the detector can seriously affect the
68
Chapter 5 Infrared Gas Sensors
69
Hazardous Gas Monitors
Application
As shown in Figure 11, the IR detector responds to
Span
radiation by generating a constant signal, which is con-
sidered the “zero” point for the source. Once the zero
0 point is established and maintained, the span calibra-
Gas Concentration
tion is automatically taken care of. This is due to the
Fig. 11 Advantage of an IR
Detector. Zero point has the
fact that the absorption of radiation by the gas is always
highest output making it easy to in the same proportion, regardless of its initial source
detect any abnormal condition. intensity. Therefore, as long as the zero point is main-
tained, the accuracy of the detector remains intact. This
is one of the biggest advantages of IR technology. How-
ever, a routine calibration check is an invaluable safety
check and should not be eliminated from any periodic
maintenance.
For gas monitoring applications, the design of the
unit should be relatively compact. Sampling should be
done by diffusion. The extractive type methods that
require a pump to move the sample into the detector
are more troublesome due to the limited electrical
motor life expectancy and the maintenance required
on the pump.
The IR instruments used for this monitoring appli-
cation are typically limited to the detection of higher
concentrations (1% and above) of hydrocarbons and
carbon monoxide. Carbon dioxide absorbs infrared ra-
diation very strongly, and many monitors are available
that can detect carbon dioxide in concentration ranges
of 0.1% and higher.
With the filter used for the %LEL combustible gases
application, the center wavelength is typically at 3.4
microns. This is the wavelength of most hydrocarbons
and it is also where most of the hydrocarbon derivative
gases have a strong absorption.
70
Chapter 5 Infrared Gas Sensors
Summary
Detection of hydrocarbons using IR has been available
for a few years. However, because of the many design op-
tions, the features and implementation of the technology
vary greatly from one manufacturer to another. Neverthe-
less, IR detection has been well received by many indus-
tries, including the petrochemical industry.
For the purpose of area air quality and safety applica-
71
Hazardous Gas Monitors
72
Chapter 6 Photoionization Detectors
Chapter 6
Photoionization
Detectors
T
he photoionization detector (PID) utilizes ultra-
violet light to ionize gas molecules, and is com-
monly employed in the detection of volatile or-
ganic compounds (VOCs). This technique originally
found use in bench top laboratory instruments, but
its complexity limited its use elsewhere. The heart of
the photoionization detector is an ultraviolet source, Fig. 1 A Pocket PID Monitor
which is essentially a lamp. Early versions of this lamp
used electrodes inside the lamp similar to those used
in the early days of the vacuum tube and were quite
costly to manufacture.
The lamps used today do not contain electrodes
and are both less costly and have a longer life expect-
ancy. In the 1980s, with the advent of integrated cir-
cuit (IC) technology, the electronics that were de-
signed were better able to process the small signal from
PID sensors into useful and reliable data.
Additionally, requirements for the monitoring
of underground storage tanks to help prevent
ground water contamination began to emerge,
which required monitoring of VOCs.
These events led to the design of small, portable
PIDs (see Figure 1) that have proven to be both prac-
tical and reliable, and which offer fast response and
the ability to detect low gas concentrations. To this
73
Hazardous Gas Monitors
day, PID sensors are the preferred choice for the de-
tection of VOCs.
Principle of Operation
Ultraviolet (UV) light is the radiation group that
has frequencies that are higher and are directly above
visible light in the electromagnetic radiation spectrum.
UV wavelengths, being in the nanometer (ηm = 10-9
meters) region, are much shorter than IR wavelengths.
The wavelengths of infrared light used for gas analy-
sis are in the micron region, 10- 6 meters.
Because shorter wavelengths have higher frequen-
cies, and hence more energy, UV radiation has more
energy than IR radiation. UV radiation energy levels
are commonly described in terms of electron volts, or
1
eV = energy gained when a particle of eV.1 Wavelength is related to eV through Planck’s Con-
one electronic charge is accelerated stant2 which is 4.135 x 10-5 eV. [An electron volt equals
through a potential of one volt. 1.2395 x 10-6/wavelength (ηm). The term eV is used
for convenience, to give a simple numerical expres-
2
Planck’s Constant is a fundamental con-
stant in quantum physics. The radiant
sion of the radiant strength.]
energy only changes in “quanta” or The PID Lamp. The heart of the detector is the
discrete steps. PID lamp which has an exterior shape resembling a
medical ampule and comes in different sizes and di-
mensions, depending on the manufacturer. The lamp
is filled with a low-pressure inert gas. When this gas is
energized with energy in resonance with the natural
frequency of the gas molecules, an ultraviolet spec-
tral radiation is produced.
The wavelength of the UV light emitted depends
on the type of gas in the lamp. For instance, krypton,
when excited, will emit 123.9 ηm and 116.9 ηm radia-
tion, or the equivalent of 10 eV and 10.6 eV. This 10.6
eV lamp happens to be the most popular lamp cur-
rently used in PID instruments. Various other gases
are also used. Figure 2 on the following page shows
wavelengths emitted by argon, krypton, and xenon.
Many different techniques are employed to power
the PID lamp. A high voltage is required for the lamp
74
Chapter 6 Photoionization Detectors
105.9ηm
116.9ηm
129.1ηm
75
Hazardous Gas Monitors
Gas Molecules
Signal i
+
Insulation V
-
Electrode
Window
Electrode Plate
for Lamp Illumination
Lamp
Fig. 3 A Typical Photoionization Detector Configuration
76
Chapter 6 Photoionization Detectors
77
Hazardous Gas Monitors
Characteristics
PID sensors offer fast response for detection of
many volatile organic compounds. They will respond
to all gases that have ionization potentials equal to or
less than the eV output of their lamps.
Correction Factor. PIDs are typically calibrated
with isobutylene, a stable gas with a slightly pungent
odor. This gas is easy to handle and can be stored at
high pressure, allowing calibration bottles to provide
many calibrations. PID instruments typically detect
gases at low concentrations, and most of these gases
are normally liquid solvents or other gases that are
not easy to calibrate. Thus, it is much easier to cali-
brate these instruments using isobutylene as the cali-
bration gas. Readings for other gases are obtained by
multiplying the reading by a correction factor. For ex-
ample, benzene has a correction factor of 0.5, which
means that a reading of 100 ppm isobutylene on the
sensor indicates a concentration of 50 ppm benzene.
For ammonia, which has a response factor of 10, a
reading of 100 ppm on the sensor would indicate an
ammonia concentration of 1000 ppm. Instrument
manufacturers typically supply a list of correction fac-
tors with their product.
An important point to remember is that correc-
tion factors are not absolute, and data can be slightly
different from manufacturer to manufacturer. In fact,
these factors can even vary somewhat from lamp to
lamp, and the results may vary depending on the qual-
ity of the UV output of the lamp at the time of mea-
surement. Thus, for the most accurate readings on
specific gases, it is necessary to individually calibrate
the gas of interest.
PID instruments are very sensitive to the composi-
tion of gases in the sample. The results will differ vastly
if one uses gas with dry nitrogen as the balance to
calibrate, from those results obtained when one uses
air as the balance. Therefore, it is important to use
78
Chapter 6 Photoionization Detectors
79
Hazardous Gas Monitors
80
Chapter 6 Photoionization Detectors
6000 hours.
Typical Specifications
The 11.7 eV lamps have lithium fluoride win-
for PID Instruments
dows that transmit the shorter UV wavelength light
at 11.7 eV, which allows for the detection of many Detector: 10.6 eV electrodeless discharge
lamp. 11.7 eV lamp optional.
more gases when compared to lower eV lamps.
However, the intensity of the light they emit is Response Time: 3 secs. to 90% reading.
weaker than the light emitted from lower eV Ranges: 100 ppm, 1000 ppm, and 2000
lamps. The weaker energy emitted means less sig- ppm isobutylene equivalent, auto-ranging.
nal, which results in less stable and more tempera- Sensitivity: 0.1 ppm isobutylene.
ture-sensitive instruments. Thus, 11.7 eV lamps are Accuracy: 1-10%, depending on range.
not typically used for general applications. Further-
Temperature Range: –20oC to +50oC.
more, lithium fluoride is hygroscopic, which means
that lithium fluoride crystals attract and absorb Humidity: 0–95% relative humidity,
moisture from the air, causing the window to de- noncondensing.
grade. They also cannot be exposed to cleaning sol-
vents because most solvents contain small amounts
of water. Therefore, a special, fine solid aluminum
oxide powder is needed to clean the lamp.
81
Hazardous Gas Monitors
Monitors Analyzers
Monitor Type: Target Gas: Analyzer Type:
Electrochemical O3 UV Photometers
Toxic
Gases
Solid-State SO2 UV Fluorescence
NOx Chemiluminescence
Catalytic Air
Quality THC*
Monitor Flame Ionization
Combustible Solid-State
Gases
Single CO Infrared
Infrared Gas
(Online)
CO2
Continuous
Volatile Emission SO2
Organic Photoionization Monitor
Compounds
O2 Paramagnetic;
Zirconium Oxide;
Electrochemical
Total
Hydro- Flame Ionization Most UV/IR
Carbons UV/IR Spectrometer
Absorbing Gases
Multi- Most IR
Gas Absorbing Gases Fourier Transform IR
Thermal (Laboratory)
Other Conductivity
Gases Most Gases Mass Spectrometer
Colorimetric
Gas Chromatograph
*Total hydrocarbon
Chapter 8 Sensor Selection Guide
Chapter 8
E
ach of the following sensors—electrochemical,
catalytic bead, solid state, infrared and photo-
ionization detectors—must meet certain criteria
to be practical for use in area air quality and safety
applications. Some of the basic requirements are:
1. The sensor should be designed for a housing
that is small and rugged. The sensor should be suit-
able for use in hazardous locations and harsh envi-
ronments, and, it should also be explosion-proof. The
sensor should be cost-effective, designed for installa-
tion and use in industrial production areas, and
installable at a reasonable cost.
2. For portable applications, the instruments
should have reasonable energy consumption and the op-
tion of powering the instruments with batteries should
be easily available. The instruments should be small
and portable so they can be carried easily. They should
be safe for use in industrial environments. Preferably,
the instruments should be certified as intrinsically safe
for use in a hazardous area.
3. The operation and maintenance of the instru-
ments should be easily performed by regular plant
personnel with minimal special training require-
ments.
4. In stationary installations, the sensors should
be able to function continuously and reliably for a period
of time, preferably longer than 30 days. The sensor
103
Hazardous Gas Monitors
104
Chapter 8 Sensor Selection Guide
105
Hazardous Gas Monitors
106
Chapter 8 Sensor Selection Guide
Summary
Electrochemical Sensors. Except for oxygen appli-
cations, electrochemical cell sensors are designed to
be used as toxic gas monitors. These sensors are only
suitable for low concentration ppm ranges. For por-
table applications, the electrochemical sensor has
many advantages: it has very low power consumption,
responds quickly to gas, and is not affected by humid-
ity. Also, the sensors are only exposed to gas periodi-
cally, which maximizes the sensor life.
Electrochemical sensors are therefore a good
choice for portable instruments. Electrochemical sen-
sor life expectancy is two years; however, depending
on the application, it can be much shorter. The cost
107
Hazardous Gas Monitors
108
Chapter 8 Sensor Selection Guide
Controller Specifications
1. Four-channel control unit, microprocessor-based.
2. Independent digital display of reading for each channel. Provides calibration func-
tions, three alarm settings, and diagnosis functions.
3. Three-level alarm set points (low, mid, high) with LED indicator for each alarm.
4. Relays: SPDT, 5 amp. 220 VAC max. resistive
Common relays for each alarm level (3 relays per channel)
Options: Individual relays for each alarm level ( 3 relays per channel)
“Fault” relays for each channel, indicating electrical fault of the sensor
5. 4-20 mA analog output signal for each channel.
6. Wall-mounted weatherproof NEMA 4X enclosure.
7. Provide 24 VDC, 500 mA for each sensor module, accepts and processes 4-20 mA
transmitter output signal.
109
Chapter 11 Gas Sensor Calibration
Chapter 11
G
as sensors need to be calibrated and periodi-
cally checked to ensure sensor accuracy and
system integrity. It is important to install station-
ary sensors in locations where the calibration can be
performed easily. The intervals between calibration
can be different from sensor to sensor. Generally, the
manufacturer of the sensor will recommend a time
interval between calibration. However, it is good gen-
eral practice to check the sensor more closely during
the first 30 days after installation. During this period,
it is possible to observe how well the sensor is adapt-
ing to its new environment.
Also, factors that were not accounted for in the
design of the system might surface and can affect the
sensor’s performance. If the sensor functions prop-
erly for 30 continuous days, this provides a good de-
gree of confidence about the installation. Any possible
problems can be identified and corrected during this
time. Experience indicates that a sensor surviving 30
days after the initial installation will have a good
chance of performing its function for the duration
expected. Most problems—such as an inappropriate
sensor location, interference from other gases, or the
loss of sensitivity—will surface during this time.
During the first 30 days, the sensor should be
checked weekly. Afterward, a maintenance schedule,
161
Hazardous Gas Monitors
162
Chapter 11 Gas Sensor Calibration
163
Hazardous Gas Monitors
Calibration Gas
Flow Nafion
Meter Dryer Tube or To Sensor
Wetting Material
Regulator Fig. 1 Adding Moisture to Calibration Gases
164
Chapter 11 Gas Sensor Calibration
165
Hazardous Gas Monitors
Vmix = V . (P/Pa)
where
Vmix = the volume of the gas mixture
V = the volume of the cylinder
P = the pressure in the cylinder
Pa = the atmospheric pressure
166
Chapter 11 Gas Sensor Calibration
B. Permeation Devices
A permeation device is a sealed container that con-
tains chemicals in liquid and vapor phase equilib-
rium. The gas molecules are either permeated
through the permeable container wall or through
the end cap. The rate in which the gas molecules
permeate depends on the permeability of the mate-
rial and temperature. The rate of permeation is con-
stant over long periods of time. At a known rate of
permeation at a given temperature, a constant flow
rate of air mixed with the permeated chemicals forms
a constant stream of calibration gas. A calibrator with Examples of Permeation Tubes
constant temperature and flow regulation is needed.
However, the permeation tube continuously emits
chemicals at a constant rate thus creating a storage
and safety problem. Also, the rate of permeation for
a given gas of interest can be too high or too low for
a given application. For example, high vapor pres-
sure gases permeate too quickly while very low vapor
pressure chemicals have a permeation rate that is too
low to be of any use.
Permeation devices find most of their use in labo-
ratories and in applications using analytical analyzers.
For gas monitoring applications, the concentrations
needed to calibrate the sensor are typically too high
for the permeation device. Therefore, they have been
found to be of limited use.
C. Cross Calibration
Cross calibration takes advantage of the fact that
every sensor is subject to interference by other gases.
For example, for a sensor calibrated to 100% LEL hex-
ane, it is usually much easier to use 50% LEL meth-
ane gas to calibrate the sensor instead of using an ac-
tual hexane mixture. This is because hexane is a liq-
uid at room temperature and it has a low vapor pres-
sure. Therefore, it is more difficult to make an accu-
rate mixture and to keep it under high pressure.
167
Hazardous Gas Monitors
168
Chapter 11 Gas Sensor Calibration
169
Hazardous Gas Monitors
170
Chapter 11 Gas Sensor Calibration
171
Hazardous Gas Monitors
172
Chapter 11 Gas Sensor Calibration
173
Appendix I PID Correction Factors
Appendix I
Compound Name Formula 9.8 C 10.6 C 11.7 C IP (eV)
191
Hazardous Gas Monitors
192
Appendix I PID Correction Factors
193
Hazardous Gas Monitors
194
Appendix I PID Correction Factors
195
Hazardous Gas Monitors
196
Appendix I PID Correction Factors
197
Hazardous Gas Monitors
198
Thermocatalitic sensor in dynamic mode
1. General principle
Schematic diagram of sensor for methane concentration in dinamic mode is shown on the figure below:
C0
St
Figure 1
When the coil is powered, it will start to heat up, and gas into chamber will start to combust. At the
beguinning inflow of the gas through calibrated hole is smaller then the quantity of the combusted gas
onto the coil, concentration of gas C will be reduced. When the quantity of combusting gas becomes
equal to quantity of inflow gas, concentration stabilize with new value Ci (see diagram 3). It is easy to
show theoreticaly, and experiments certify, that gas concentration into chamber reduce with exponential
trend.
At the moment when concentration almost reach Ci, power is turned off and the chamber starts to fill
again, it fills up for about the same time. The power is turned on again and so on alternatively.
Transitive process in which we measure the signal in two consecutively points is repeated, and on that
base we determine concentration of the target gas, as we will see later. It is validated truth that
transitial process is much more reliable and much informational than the stationar.
Ss
Si
I II III IV
Figure 3
One short time period after energizing the coil (t=0) there is no respond from system and that is
inertial time of the system, a signal remains null (period I), after that time, coil starts to heat up
rapidly caused from current and also from gas combustion until the signal reach its maximal value Ss
(period II). In the period III, which is of the greatest interest for metrology, flux of outer gas inflowing
through the hole is smaller than the quantity of the combusted gas and the concentration decrease ‘till
reach equilibrium point when the signal becomes Si. Signal then stays constant on minimal value Si
(period 4).
The following trend of exponencial decreasing of the concentration in the period III is changing by
following formula:
C = Cs [ r + ( 1 - r ) e-t/T]
Where are : r = Ci/Cs factor of concentration reduction, and T — is time constant. The both of this
factors are characteristics specific for each. Time t is calculated from the end of the period II (t=0).
S = k Cs [ r + (1 - r ) e —t/T ]
This is, of course, only a part fo the signal and its origin is from the gas combustion. Other part of
the signal, parasitic signal, is the result in the first place of resistor heating caused by current flow but
its also caused by ambient temperature. It also consists of other relevant factors as is the air pressure,
humidity, resistive changes with time, etc.
St= S + Spar
Signal total St
Ss
L = St(t1)-St(t2)
Si
Spar
Figure 4
On this curve coordinate beguinning (t=0) is shifted at the end of period II. Total signal is measured in
two points t1 and t2 on transition curve.
The difference of the signals in those two points L (t1 , t2 ) = S (t1) - S (t2) is proportional with gas
concentration.
From theory stated above it can be seen that the number of parameters are before the constructor to
adapt them in the best way to his specific applications. We will comment only briefly some of them.
The diameter of the calibrated hole and the capacity of combustion of coil, are obviously in relation to
each other. It’s stated before that it is beter to have coil with as smaller thermal inertia as it is
possibile, but when it is achieved by minimizing the vire dimesions and coil itself, thermocatalitic
surface layer should achieve very intensive combustion and in that way increase sensitivity of sensor
itself. The diameter of calibrated hole considerably influence on the form of exponential curve,
determining its gradient, level of the Ci (Si) and the time for reaching that value (time constant).
Selection of the points t1 and t2 in which we measure the signal is of great importance. Point t1
should be chosen far enought from maximal value, to avoid shifting of that point to left side from the
maximum, because during time coil characteristic gets deformed and maximum point is sliding slowly
towards right side.
The distance between points t1 and t2 is from great importance. It won’t be same for stationary
devices in mines which are constantly exposed to target gas and devices in industry (stationary or
portable) where the presence of target gas is mostly incidently.
The proper choose of this two parameters enables construction of so called “universal” explosimeter.
As a matter of fact, classical explosimeters shows accurately only concentration of gas for which they
are calibrated and indicate it in percents of lower explosive limit (LEL). Any other explosive gas they
only indicate, and measuring of explosivity itself can deflect up to 300% in extreme cases. This makes
those exposimeters practically useless, in areas where we don’t know which gas is present, or when
there is a mixture of gases.
Dinamic mode enables to, with proper chose of parameters t1 and t2, realize explosimeter wich will
equaly acurate measure explosivity of mostly combustile gases, and for extreme cases deflection is
below few tenth of percent (referred to curently 300%).
gas
sinter
0.5 - 0.7mm
HC LC
Principle of opertion is as follows: first of all, micro-controller turns on thermoconductive sensor for
high methane concentrations HC with short current impulse which lasts for 0.35 sec., during that time
thermocatalitic sensor is turned off (no methane combustion).
As the concentration of methane is higher, the slope is under lower angle because the wire start to
cool more intensive (element is constantly energized) this is previously stated distinction of the
catarometric sensors.
If a methane concentration is above level of DGE (5%CH4), sensor for lower concentrations stays
turned off, otherwise it turns on, but only if the concentration is less than DGE. Time line of one
measuring cycle, in case that after initial turning on the DMS-01 has measured concentration below
DGE, is shown on following figure:
1 — short current impulse for thermoconductive sensor, lasting for 0,35 sec
2 — sensor response HC
3 — current impulse for heating the LC sensor, lasting for 2,5 sec (turns on at t=1s and turns off at
t=3.5s ), when proper temperature is reached (about 450°C) and combustion of methane starts.
4 — sensor response LC in lack of methane
5 — sensor response in presence of methane
Briefly, LC sensor is a tipical thermocatalitic sensor, it is installed in a chamber with calibrated hole
0.5 — 0.7mm. The gas pass thru this calibrated hole, sensor (LC) turns on and when wire reach proper
temperature the methane combustion starts. Quantity of the gas that combust is in proportion with
concentration of the gas that’s inside the chamber, and after some particular time, because the power
supply of the sensor is turned off, the balance renew. Transition process from the moment of turning
on ‘till stabilization is used for measuring.
Measuring is conducted in two points, first point is at 1.3 sec. and the second one at 2.3 sec. (after
turning on the sensor). Concentration is equal to difference of these two signals.
1.3s 2.3s
1s
After impulse of 2.5 sec., there is a pause of 5 sec. to fill the chamber again, and the whole process is
cyclicly repeating. Total cycle of measurment lasts for 7,5 sec. and is cyclicly repeating, untill to
moment when sensor of lower concentration exceed the DGE (5% methane) when the sensor of higher
concentration HC turns on. Farther work of this device is on the same principle as after turning on
(when the concentration falls below DGE, sensor HC is turned off, and sensor LC turns on).
- Power supply of the sensor via electronic circuitry with intrinsically safety with maximal
values of electrical parameters: Um=1.41V, Iib=116 mA.
- Limited heating of the thermoelement to safety limit (≤ 450°C).
- Thanks to exceptional low power consumption enough for accurate sensor operation (low wire
diameter), a high temperature coefficient of resistive change ( K=3.9 x 10-3), and rather low
thermal inertia (measuring cycle lasts for 7.5 sec., 2.5 sec for heating, and 5 sec. for cooling
down), its created the effect of developing a rather low amount of thermal energy, which leads
to low operating temperature of the sensor (Tmax=73°C), and which is a far lower from
maximal safety temperature tbez=295°C, derived with a safety factor of 1.5. Measurements in
stated example are achieved with power supply conditions U=1.41V, I=116 mA, in presence
of mixture of propane 5,3% and ambient air, at the temperature of 50°C (mixture is in comply
with group of gases IIA).
- Inlet of the target gas inside the sensor is thru sintered mass which eliminate possibility of
intruding of coal dust inside the sensor, which lead to higher permited temperature limit inside
the sensor from 1500 to 450°C.
Design of the sensor, that means the operational principle of dynamic mode, have allowed to designers
to construct the measuring intrument, with intrinsically safety conditions in compy with limit curves
(IEC 79-11) witout any problems, i.e. far from the limit curve.
4. Summary
Thermocatalitic sensors in dynamic mode of operation has following advantages compared to
“classical” thermocatalitic sensor:
1. With only one element, without reference, we determine the gas concentration, avoiding in that way
any inconvenience which reference element can cause (need for selection — complying of elements,
operative with the comparative, different aging of this two elements, which is inevitable, and which
lead to zero drift, etc.)
2. With dynamic mode, actually we compare element with itself, it is referent to itself, comparative, if
its characteristics change over time, those changes are the same in points t1 and t2. (figure No. 4). On
that way problem of zero drift is solved which is one of the biggest advantages of this sensor.
4. Unlike classical thermocatalitic sensors which has to be installed in flameproof housing, i.e. realised
in Ex d, Ex protection of this sensor is Ex ia, thus eases in a great way construction of detector itself.
Dynamic mode allows us to realize explosimeter, with proper selection of parameters (t1 and t2),
which will equaly acurate measure explosivity of most of common combustive gases, and for those
extreme cases deflection is around few tenth of percents (referred to curently 300%).
General conclusion is, that the despite of use of IR sensors, changed thermocatalitic sensor in dynamic
mode of operation has its distinctive position as well in mining as so in industry.