Professional Documents
Culture Documents
Philips: MX 4000 Dual MX 6000 Dual
Philips: MX 4000 Dual MX 6000 Dual
Philips: MX 4000 Dual MX 6000 Dual
MX 4000 Dual
MX 6000 Dual
4535 673 83741
PHILIPS
Contents
equipment.
• The “operators” are those persons who actually handle the equipment.
Before attempting to operate the equipment, you must read, note, and
strictly observe all DANGER notices and safety markings on the CT
System.
Before attempting to operate the equipment, you must read this manual
thoroughly, paying particular attention to all Warnings, Cautions and
Notes incorporated in it. You must pay special attention to all the
information given and procedures described in the SAFETY section.
Caution Directions, which if not followed, could cause damage to the equipment
described in this Instructions for Use and/or any other equipment or goods,
and/or cause environmental pollution.
Within this Instructions for Use, the most extensive configuration of the
system is described, with the maximum number of options and accessories.
Not every function described may be available on your system.
This volume explains how to use the CT scanning system. It also contains
information about safety, data security, system start-up, software
navigation, scanning protocols, networking, and calibration.
operators must only use and operate the equipment in such ways as do not
conflict with applicable laws, or regulations which have the force of law.
Caution In the United States, Federal law restricts this device to sale, distribution,
and use by or on the order of a physician.
1.3 Contraindications
The MX 4000 Dual and MX 6000 Dual CT systems should not be used if
any of the following contraindications exist or are thought to exist.
• The image performance quality assurance checks listed under the
heading, Maintenance, have not been satisfactorily completed.
• The preventative maintenance program is not up-to-date.
• If any part of the equipment or system is known
(or suspected to be) operating improperly.
1.4 Compatibility
Equipment described in this manual should not be used in combination
with other equipment or components unless such other equipment or
components are recognized as compatible.
Changes and/or additions to the equipment should only be carried out by
Philips and Neusoft Medical Systems or by third parties expressly
authorized by Philips and Neusoft Medical Systems to do so. Such changes
and/or additions must comply with all applicable laws and regulations that
have the force of law within the jurisdiction(s) concerned, and with best
engineering practice.
Changes and/or additions to the equipment that are carried out by persons
without the appropriate training and/or using unapproved spare parts may
lead to the PNMS warranty being voided. As with all complex technical
equipment, maintenance by persons not appropriately qualified and/or
using unapproved spare parts carries serious risks of damage to the
equipment and of personal injury.
1.5 Compliance
The MX 4000 Dual and MX 6000 Dual CT systems comply with relevant
international and national standards and laws. Information on compliance
will be supplied on request by your local PNMS representative, or by:
Philips and Neusoft Medical Systems Co. Ltd.
Hun Nan New District
110719 Shenyang
China
The MX 4000 Dual and MX 6000 Dual CT systems comply with relevant
international and national laws and standards on EMC (electro-magnetic
compatibility) for this type of equipment when used as intended. Such laws
and standards define both the permissible electromagnetic emission levels
from equipment and its required immunity to electromagnetic interference
from external sources.
1.7 Training
Operators of the MX 4000 Dual and MX 6000 Dual CT systems must
have received adequate training on its safe and effective use before
attempting to operate the equipment described in this Instructions for Use.
Users must also ensure that operators receive adequate training in
accordance with local laws or regulations which have the force of law.
Wa r n i n g The X-ray unit may be dangerous to patient and operator unless safe
exposure factors and operating instructions are observed.
Wa r n i n g Do not use the CT system for any application until you are sure that the
Image Performance Quality Assurance has been satisfactorily completed,
and that the Preventative Maintenance Program is up to date. If any part of
the equipment or system is known (or suspected) to be operating
improperly or wrongly-adjusted, DO NOT USE the system until a repair has
been made.
You can find information about the Image Performance Quality Assurance
and the Preventative Maintenance Program in the Maintenance section of
this Instructions for Use.
Wa r n i n g Do not use the CT system for any application until you have read,
understood and know all the safety information, safety procedures and
emergency procedures contained in this SAFETY section. Operation of the
CT system without a proper awareness of how to use it safely could lead to
fatal or other serious personal injury.
Wa r n i n g Do not use the CT system for any application until you have received
adequate and proper training in its safe and effective operation. If you are
unsure of your ability to operate this equipment safely and effectively DO
NOT USE IT. Operation of this equipment without proper and adequate
training could lead to fatal or other serious personal injury. It could also
lead to clinical misdiagnosis.
Wa r n i n g Do not use the CT system for any purpose other than those for which it is
intended. Operation of the CT system for unintended purposes, or with
incompatible equipment, could lead to fatal or other serious injury. It could
also lead to clinical misdiagnosis.
Wa r n i n g After the Stop button is pressed, the table is locked in place for two
seconds. Then it will be free floating with no up/down capabilities. Make
sure that you maintain control of the table so that it does not move.
Wa r n i n g Make sure that the motion of the table is in the direction that will ensure
that the patient can be easily released and will not get pressed against the
gantry covers.
Wa r n i n g Do not remove covers from this equipment. Removing covers could lead to
serious or fatal personal injury.
Wa r n i n g Do not subject the system to serious mechanical shock, as the cathode ray
Philips and Neusoft Medical Systems 4535 673 83741
tube (CRT) can fracture if struck or jarred. This may result in flying pieces
of glass and phosphor coating that can cause serious injury.
Fire regulations for the type of medical area being used should be fully
applied, observed and enforced. Fire extinguishers should be provided for
both electrical and non-electrical fires.
All operators of this medical electrical equipment should be fully aware of
and trained in the use of fire extinguishers and other fire-fighting
equipment, and in local fire procedures.
Wa r n i n g You should not allow portable radio transmitting devices (such as mobile
telephones) into the examination room - whether switched on or off. Such
devices could exceed EMC radiation standards could interfere with the
proper functioning of the CT system. This could, in extreme cases, lead to
fatal or other serious personal injury or to clinical misdiagnosis.
Failure to observe these warnings may cause serious or fatal bodily injuries
to the operator or those in the area.
Caution If oil leaks are detected, shut down the scanner and immediately contact
the nearest Philips and Neusoft field service office.
Philips and Neusoft Medical Systems 4535 673 83741
Wa r n i n g
• Do not stare into the laser beam and instruct the patient not to stare into
the beam.
• The use of optical instruments, such as eyeglasses with large diopter or
mirrors, with this product will increase eye hazard.
• Ensure that, for head examinations, the patient wears protective glasses
when the laser beams are on.
• Refer to the equipment labels for specific compliance and laser safety
information.
Caution Although the MX 4000 Dual and MX 6000 Dual CT systems operate on a
personal computer (PC) platform, the installation of PC software not
specified in the system documentation may adversely affect the operation
and security of the system, as well as the networks to which the system is
connected. These adverse effects may not be immediately apparent to the
user. Users should therefore not install unauthorized software onto their
system.
Network security
The MX 4000 Dual and MX 6000 Dual CT systems must be placed on a
secure local computer network that has protections against viruses and
other harmful computer system intruders. Make sure the equipment is
connected to a local network that uses appropriate protection, such as a
firewall and virus scanners.
Antivirus Updates
Virus definition files and updated virus scanning engines are included with
Philips and Neusoft Medical Systems 4535 673 83741
Caution Under no circumstances should updated virus definition files (or any type of
software) be installed on the CT system by anyone other than a Philips and
Neusoft Field Service Engineer or an authorized Philips and Neusoft agent.
Positioning of monitors
Unauthorized visual access to protected information can be minimized by
positioning the system’s monitor so it faces a wall, to prevent viewing from
doorways, hallways and other traffic areas.
To help in limiting unauthorized visual access, an unattended CT monitor
automatically goes blank after a set period of time.
Caution Whenever media is inserted into the MX 4000 Dual or MX 6000 Dual CT,
be sure that the media has not been exposed to potential viruses, worms
and trojans that infect desktop PCs.
Caution Removable media that contains images and/or other medical information
Philips and Neusoft Medical Systems 4535 673 83741
4.1.1 Monitor
A flat screen monitor displays images and operating the system.
An optional flat panel monitor is also available.
The power switch for the monitor is located on the front. The power
indicator LED lights when you turn on the monitor.
Caution To maintain optimal adjustment and correlation with the filmed images, do
not change the settings of the monitor.
1
STOP 2 3
6 7
4 5
X 8
9
Wa r n i n g Observe the condition of the patient when operating the Scan control box.
Recorder microphone
A recorder microphone is located in the scan control box. This microphone
is used to record messages that can be used during the scan process.
Internal CD writer
The internal CD writer is a CDR drive that stores DICOM images along
with the necessary viewing software on a CD. It provides an alternative for
archiving images or transferring patient images to referring physicians.
Note • Always use a blank CD for recording.
• Upon completion of the writing process, verify that all required
information has been written to the CD.
3 Laser on/off button serves a dual purpose. It turns on and off the laser
markers which are used for positioning the patient in the slice plane. It
also displays the maximum and minimum tilt and patient table
positions.
• Press the button once to turn on the laser. The system displays the
maximum positive tilt position, the maximum table height and the
maximum table in position.
• Press it a second time to turn off the lasers (the lasers turn off after five
seconds). The system displays the current table position.
• Press the button a third time to display the maximum negative tilt
position, the minimum table height, and the maximum table out
position (lasers turn on again).
• Press the button a fourth time to turn off the lasers (the lasers turn off
after five seconds). The system displays the current table position.
Note Lasers automatically turn off after 60 seconds.
Wa r n i n g Be sure to observe the condition of the patient when operating the control
panel.
Head holder
Philips and Neusoft Medical Systems 4535 673 83741
The head holder can be used for most routine child or adult CT head
exams. The angle of the holder positions the head naturally for a routine
brain scan and minimizes the required gantry angle to achieve optimal
results.
Patient pads
This complement of accessories includes head cushions, patient straps for
immobilization, and pads for patient comfort.
24 kW, (at 120 kV, 200 mA 36 kW, (at 120 kV, 300 mA
Nominal output power
and load time of 4s) and load time of 4s)
28 kW, (at 140 kV, 200 mA 42 kW, (at 140 kV, 300 mA
maximum output power
and load time of 54s) and load time of 38s)
Wa r n i n g Make sure the system and the scan room comply with operational
requirements before initiating the system.
5.1 Overview
This chapter provides a brief description of the features available in the
software as well as instructions on how to access them. Be sure to familiarize
yourself with this information before conducting any scan procedures.
Philips and Neusoft Medical Systems 4535 673 83741
Orange buttons are active (as in the all other buttons above). This stage of
the workflow is accessible, if you want to move to it.
Gray buttons are inactive. These functions cannot be accessed at this stage
in the workflow.
The Workflow bar consists of these buttons:
• Home to toggle to the Home window. This window features a directory
of completed studies including a series list, data list and a review
window.
• Schedule to access the patient catalog. This window features options for
scheduling new patients, removing patients from the catalog and a
comprehensive list of all patients previously scanned.
• Scan to start a new study or to continue the previous one. The window
varies depending on the current process:
• If you click this button during scan preparation, the current scan
parameters display.
• If you click this button when scan preparation is not in progress, it
displays the patient data window.
This step is active automatically after you log in. This button is always
enabled, except during image acquisition or between acquisitions in
timed scans. It is also enabled automatically after clicking End Study.
Search provides a dialog box where you can enter information to search for
a specific patient.
• Click Search.
• Enter the desired search parameters.
• Click Confirm.
Lock allows you to toggle the lock on and off. Locked patient data cannot
be removed from the system.
Fix patient data allows you to edit the details of a patient who was
previously entered in the system.
Wa r n i n g • Do not use the Fix patient data to change patient data entered from the
HIS/RIS system. This tool is solely for use with data entered manually
through the patient data form.
• The patient ID cannot be changed using the Fix patient data feature.
Delete allows you to delete the selected patient and their corresponding
data.
Copy To allows you to copy the currently selected items to another device.
A dialog box opens with lists of available Local and Remote archive devices.
Load to Viewer allows you to load the selected images for Review.
Philips and Neusoft Medical Systems 4535 673 83741
6.1 Overview
This chapter covers these scan operation procedures:
• login and daily checks
• operation of patient table
• use of the scan window
• scanning process
• ancillary functions
Review this information carefully before using the scanner.
Philips and Neusoft Medical Systems 4535 673 83741
6.2 Startup
Before using your system, confirm that the scan room meets the
appropriate conditions:
18° ~ 24° less than 5° per hour 30% ~ 60% (no condensing)
18° ~ 28° less than 5° per hour 20% ~ 80% (no condensing)
Use this procedure to start up your scanner when it has been completely
shut down.
1 Turn ON the wall power (if gantry power is off ).
2 Locate the power switch on the side of the gantry.
3 Turn on the gantry power.
4 Turn on the power to the computer (inside the cabinet), and monitor.
• Computer power-up takes about 1.5 minutes.
5 Double click the Spiral CT icon to activate the software.
6 Type the desired User name and password.
7 Click Login.
Caution • Blood and contrast medium are health risks. Take safety precautions
when removing blood or residual contrast medium.
• Use a commercial biocide, approved by your governing authority to clean
the surface of the system including the table, headholders and
accessories.
Table Up/Down
To vertically position the region to be scanned from the lowered
position (where the patient can sit and then lay down on the patient
table in the gantry opening), use the Up button on the gantry
panel. Use the Up and Down buttons to properly adjust the table
position.
Table In/Out
To bring the patient’s region of interest into the gantry opening, use
the In or Out buttons.
• “Tapping” an In or Out button will move the table 0.5 mm each time it
is tapped.
• When an In or Out button is continuously held down, movement
accelerates after about five seconds. For fine adjustments, press and
Philips and Neusoft Medical Systems 4535 673 83741
Note • The patient table cannot be moved in when it is below a certain height.
Raise the patient table up to enable moving it in.
• The patient table motion stops within 10 mm upon actuation of an
emergency stop control.
• If the couch is pushed in manually, be sure couch is high enough (above
210) to avoid colliding with the gantry.
Note Movement of the patient table is executed at two speeds. After pressing the
appropriate button, the table moves at a slow speed and after several
seconds, if the button remains pressed, it proceeds to a higher speed.
Caution • While moving the table and gantry, take care not to insert your feet
under the table side covers or between the gantry and patient table.
• Avoid inserting your fingers between the stretcher and the table carriage.
• Avoid placement of ancillary equipment (such as wheelchairs, IV pumps
or beds) under the table. The table could collide with these items during
movement.
The sample above shows the patient data form which is used to enter
patient data for a new study. These options are available:
• New opens a blank patient form.
• Anonymous fills a preset subset of generalized patient details for
unidentified patients.
• Current brings the details of the current patient (that was last entered)
to the parameter fields.
Note After the scan has finished, you may operate a utility for correcting
Anonymous patient data retroactively in the patient catalog and in archives.
• Operator Name
• Orientation
• Description
Note • Age Group - You must select the appropriate setting for the age of the
patient. If the patient is a child up to 18 months, the appropriate setting is
Infant.
• Gender - The Other option refers to Anonymous patients whose gender
is anatomically unidentifiable. It may also refer to anything other than a
living creature (for example, minerals and phantoms).
Note You have the option to configure your system so that, during a manual
patient entry, the first field to be completed when you start a study is either
the Patient ID or Accession number.
Should you manipulate the planned area, the system will automatically
update the corresponding protocol.
You can use this window to ensure the quality of the images, as well as send
specific images to print. Only window and level options operate in this
window. To see all images, you must load the study into the Viewer (see
Review mode for more information).
• The All Scanned tab displays those patients who have already been
scanned.
• Scheduled tab displays only those patients who are scheduled to be
scanned.
Add a patient
1 Click Schedule Patient to add a new patient to the list. The system
displays the patient data form.
2 Enter the patient information.
3 Click Ok to add the patient to the list.
Note Click Schedule Again to create multiple instances of the same patient.
Delete a patient
1 To delete a patient, select the desired name from the patient list.
2 Click Remove patient.
3 Confirm your selection to remove the patient.
Note When HIS/RIS is enabled you also have the option of displaying the medical
alerts stored in the database. Usually the Scheduled list is updated by the
HIS/RIS pushing patients into it.
Wa r n i n g The Fix patient name option should not be used with patients entered from
the HIS/RIS system.
You may also choose to add any of these options to your exam procedure:
• Plan on Surview—exam protocols usually include a Surview scan and
automatically launch Plan on Surview upon reconstruction of the
Surview image. See Plan on surview, on page 6-23, for more
information.
• Multi-reconstruction
The scanning process is set-up and initialized from the Scan control panel
and tabs on the screen. Table movement is controlled from the Scan control
Philips and Neusoft Medical Systems 4535 673 83741
box outside of the scan room or the gantry control panels inside of the scan
room. This section provides detailed steps to complete a typical exam
procedure, as well as descriptions of the available options.
Note In addition to the scanning options you also have the option to film and
conduct post-processing analysis.
2 Enter the patient data (mandatory fields are highlighted). You have
several options for entering patient data:
• For a new patient, click New. See Entering a new patient, on
page 6-19 for instructions on filling in patient data.
• For an anonymous patient, click Anonymous, the system fills in the
patient ID and the first name.
• For a current patient, click Current. The system fills the fields with
the information from the last patient entered.
Note You can also begin a scan by selecting a scheduled patient in the schedule
mode.
Caution Before proceeding to protocol selection, verify that the patient information
loaded into the patient data fields (from any source) is correct. Failure to do
so could result in scanning a patient with the wrong information thus
requiring another scan on the patient.
are highlighted. Use Tab, or the mouse to navigate through fields while
entering a new patient:
1 Click Scan.
2 Based on your system set-up, the Patient ID field automatically
contains a value. If you want to change it, type a new patient ID for the
patient.
3 In the Last Name field, type the last name of the patient.
4 In the First Name field, type the first name of the patient.
5 Click the appropriate Gender for the patient:
• Male
• Female
• Other
Note Your system may be configured to make the birthdate mandatory instead of
the age group.
The Protocol window displays exam protocol groups on the left, and the
corresponding protocols on the right:
Philips and Neusoft Medical Systems 4535 673 83741
Note The image above shows the Adult Exam Protocol Groups. When the
Classify protocols feature is active, an infant displays if the infant or Child
Age Group is selected.
Note After scanning a clinical series, the surview image is saved with the ES
(executed scan) lines.
1 If desired, adjust the planned scan box over the desired area.
• Click and drag the box to move it.
• Click on a point on the box to expand or contract it.
The system adjusts the parameters accordingly.
Philips and Neusoft Medical Systems 4535 673 83741
2 Click Go to execute the planned scan. The system displays the final
image in the Scan viewer window when the scan is complete.
3 Select an option:
• Edit and Rescan deletes the current surview and allows to re-create
the entire plan.
• Rescan deletes the current surview and rescans with the same
parameters.
• Plan allows you to accept the current plan.
4 The scan parameters for the next scan display. Click Go to execute the
next scan.
Note Once you scan the surview, the plan lines turn blue and are locked in place.
You can add scans, but you must start a new scan to create a new surview.
There are several customizable features included in the MX 4000 Dual and
MX 6000 Dual software packages. This chapter provides information and
procedures for setting up the system according to your needs. Make sure
you complete the system setup before scanning any patients.
Daily service functions include these items:
• Warm-up (see Scan operation)
• Air Calibration (see Scan operation)
• Protocol
• Setting
• Anti virus
• Log off user
Philips and Neusoft Medical Systems 4535 673 83741
Note When the temperature or humidity in the scanning room changes in a large
scale, ring artifact shadow or central shadow may be generated in the scan
image.
Note Make sure there is no object in the scan hole in the sensitivity calibration
operation.
Setting
Setting allows you to customize a variety of system options to best suit the
needs of your facility.
• Password modification allows you to create new passwords.
• ID generator allows you to select options for creating the patient ID.
• Record use to add new messages, or edit and delete existing messages
that may be used during the scan process.
Philips and Neusoft Medical Systems 4535 673 83741
Protocol
The protocol service option allows you to create, modify, and delete
protocols for scanning and viewing images. Use this option to configure the
protocols to best suit your needs.
1 Click Protocol to access the protocol edit function. The system displays
this window:
Note The Philips logo displays on factory set protocols. These protocols cannot
be edited or deleted. Use the Save As feature to create a new protocol
containing the changes to default settings.
4 If you selected Edit mode, the system displays the scan parameters.
Adjust the parameters as needed.
5 Click the back arrow in the upper left corner to return to the protocol
edit window.
6 Click Exit to return to the Daily Service menu.
Display edit
The Display edit feature allows you to alter the viewing protocols. You can
Philips and Neusoft Medical Systems 4535 673 83741
Film template
Allows you to create a new film template (layout). New templates are
selectable for the auto film feature of when you generate a new protocol.
When you scan, your film output automatically displays in the selected
layout.
Antivirus
The system includes built-in Antivirus software. Click this option to run a
virus check on the CT system.
8.1 Overview
This chapter covers the protocols used during the scan procedure as well as
the parameters available in each protocol. There are several scan modes:
• Surview scans are radiographic-like scans upon which the study is
planned.
• Axial scans are the CT mode of slice-by-slice scanning while the Patient
Table is motionless. The result is: n slice images (n is the product of the
number of scans, and the number of slices/scan in the specific scanner,
one to forty depending on the selected collimation).
Philips and Neusoft Medical Systems 4535 673 83741
• Helical (spiral) scans are multi-rotational scans while the Patient Table
is incrementing continuously. The result is a series of slice images,
reconstructed with flexible increment.
Caution Make sure that the correct scan parameters are entered to ensure correct
left/right orientations.
Length [mm]
The Length parameter gives the region covered. Usually, the value in this
box is copied from Plan on Surview but you can type any desired value in
the correct resolution and range. If the entered value is out of range (for
example, the time that is needed to execute is less than the scanner
limitation), a message displays.
Current [mA]
This parameter is used to set the X-ray current. Type the desired current in
the range of 10 mA to 230 mA in adult protocol and 30 mA in infant
protocol. You can also select from a list of frequently used values.
Low mA values are generally recommended for the surview scans.
Voltage [kV]
The Voltage parameter is used to set the voltage according to the absorption
characteristics of the scanned body part.
All Infant scans are only performable with Medium (120 kV) or Low
(80 kV) voltage. Low or Medium voltages improve contrast resolution in
small and medium objects or bodies, and therefore are preferred for
scanning infants and normal size patients respectively. On the other hand, a
High voltage (140 kV) scan provides greater penetration in large objects
and reduces the noise of the images.
Collimation [mm]
Tailored to different applications, several collimation apertures are
available:
• 0.8 (optional)
• 1
• 2.5
• 5
• 7
• 10
The minimum available thickness is always larger than the basic
collimation.
8.3 Surview
The image shows the Surview tab, which in addition to the common
parameters (see Common parameters, on page 8-3)includes these options:
Scan type
The Scan type setting determines the rotation of the tube. PA indicates a
180 degree scan angle. Lateral is a 90 degree scan.
Philips and Neusoft Medical Systems 4535 673 83741
Scan Mode
This parameter allows you to select from the Surview, Axial and Helical
scan modes.
Rot Time [sec]
The Rotation Time [sec] parameter defines the duration of one rotation of
the gantry. The table below shows the relationship between the rotation
time and Resolution.
Caution Increment of zero is allowed, but then the scanned area will receive an
increased amount of radiation. This mode will be used for biopsies and for
Bolus tests. It is suggested that the dose used in these cases should be as low
as allowed by the specific application.
Tilt [deg]
The Tilt value (in degrees) denotes the Gantry Tilt angle for the planned
scan on lateral (90 degree) Surview scan. The value in this box is copied
from Plan on Surview, where it is interactively set by the Rotate function.
The Gantry will tilt to the desired tilt angle before the scan start (while the
Enable button is pressed and held). The range for Axial scans is from -30°
to +30° depending on the couch height (see Operating the gantry tilt and
table, on page 6-6).
Note When the Surview scan angle is 180°, the Rotate button is grayed out.
Sample
This parameter is available with values of 512 and 1024.
Scan angle
The Scan Angle parameter is used to set the angle of the data acquisition.
At this time only 360° is available.
Philips and Neusoft Medical Systems 4535 673 83741
Trigger
The Trigger is used to generate a timing mechanism. You determine the
desired delay before the start of the scan, counting from the beginning of a
selected event. The initiation of contrast injection then serves as the trigger,
with radiation starting after the desired delay has elapsed.
Note • When using Trigger, always plan from a surview; complete the surview
before entering the desired time delays. The Timing Ruler displays at the
screen bottom while the scan is in progress, allowing the scan progress to
be monitored.
• When using timed sequences, the maximum delay for the first scanned
series is 180 seconds with voice time and the minimum time is 1 second
without voice time. Maximum time between series is 180 seconds with
voice time and the minimum time is 7seconds without voice time.
• There is no longest delay time when the time between series is less than
180s. You can add multiple series if desired.
• Pausing the injector will NOT pause the scanner once the injector delay is
reached.
The Trigger to start the scan can be started manually or automatically. For
Spiral scans the scan can be started with Auto Start or SAS.
• In manual mode operation, you must press the Auto scan button and
the Injection button at the same time.
• In automatic mode, after you press the Auto scan button, the scanner
continues to wait for a triggering signal from the contrast injector.
The countdown of the delay before scanning begins immediately following
the triggering signal from the contrast injector.
Wa r n i n g Keep the patient under continuous observation. Between the series, the
bed will move automatically until completion of the plan.
Auto Voice
The Auto voice parameter is used to select a pre-recorded message set:
before the scan (for example, “hold your breath”) and after it (for example,
“you can relax now”).
Philips and Neusoft Medical Systems 4535 673 83741
Thickness [mm]
Use the Thickness parameter to set the tomographic thickness, which
determines the spatial resolution in the axial direction (perpendicular to the
plane of the slice). Select the desired combination of collimation and
reconstruction from a string in the list:
• 2 x 0.8 (optional)
• 2x1
• 2 x 2.5
• 2x5
• 2x7
• 2 x 10
Increment [mm]
The Increment parameter is used to set the distance between two
consecutive scans in millimeters.
The default value of the Increment is equal to the selected collimation.
Philips and Neusoft Medical Systems 4535 673 83741
Caution Increment of zero is allowed, but then the scanned area will receive an
increased amount of radiation. This mode will be used for biopsies and for
Bolus tests. It is suggested that the dose used in these cases should be as low
as allowed by the specific application.
Matrix
The Image Matrix parameter sets the number of pixels that the
reconstructed image will contain. The matrix size is 5122. Understanding
the relationship between FOV, resolution mode and reconstruction will
help you make a matrix choice that produces the best image quality:
Standard resolution, matrix =1.6 x FOV (except filter D)
HR: matrix = 3.2 x FOV
For example, if the scan mode is HR and the FOV is 300 mm, then the
matrix is 3.2 x 300 = 960. The closest available matrix selection is 1024.
Window Width/Window
Window Center and Window Width are used to set the gray levels of the
displayed image.
Trigger Type
Toggles the timed scan on and off. A time bar appears along the bottom of
the window when timed scanning is enabled.
PID(s) (Post injection delay)
Philips and Neusoft Medical Systems 4535 673 83741
Language
To select a language for the auto voice option.
Pre-scan
To select a phrase before the scan.
Post-scan
To select a phrase after the scan.
9 Image reconstruction
9.1 Overview
Image reconstruction allows you to perform reconstructions of raw scan
data using one of these methods:
• On-line-reconstruction begins immediately after the scan is complete.
• Off-line-patient files are accessed in the directory and reconstructed.
You also have the option to automatically launch reconstruction results in
corresponding viewers.
Philips and Neusoft Medical Systems 4535 673 83741
The displayed parameters correspond with the type of scan that was
performed.
Note You have the option of selecting the patient after clicking Offline recon.
Leaf & select Images for Fast Leafing and also for deselecting the graphics,
zoom, and pan buttons, thus enabling selection of images.
Pan to center the feature of interest in the image frame by dragging the
image in the image window.
Windowing - allows you to quickly set the Center and Width of the image
windowing. These options are available:
• P.F.
• Brain
• IAC
• Spine
• Bone
• Lung
• Abdomen
• Liver
Philips and Neusoft Medical Systems 4535 673 83741
Using the mouse for Windowing - You can also adjust the window Level
and Width by clicking and dragging the middle mouse button. Vertical
dragging changes windowing Level and horizontal dragging changes
windowing Width.
Titles off - Turn the image titles off and on. If turned off, they come back
on when the image is filmed or saved.
Reset All - For resetting the study’s images to the state they were in upon
loading.
10.1.2 Film
Use the Film icon to send selected images, a window or series to the
Filming application. Alternatively, filming options are accessible from the
File menu. Filmed images are added to the images already in the Filming
application. Use this procedure to film images:
1 Click on the appropriate selection mode. In some applications, only the
displayed image can be filmed.
2 Launch the film option using one of these methods:
• Click the Film icon on the toolbar.
• Select Film Images from the File menu.
The system sends the selected images to the Film application.
10.1.3 Report
You have the option to send images to the Reporting application for use
during report creation:
1 Select the desired image.
2 Click Report.
Note There is no limit to the number of images sent to the reporting application.
10.1.4 Batch
The Batch set of tools is used to organize a set of images prior to saving,
filming, reporting, or viewing them in the Cine mode. These tools are
available in the 2D, planar and volume modes.
Note Before using the Batch mode, be sure to set up the images the way you
want them using the 2-D tools.
From allows you to define the starting image of the batch. This is the
default when you enter Batch & Cine.
To allows you to define the end image of the batch.
All selects all images for the batch.
Clear eliminates the start and end definitions you have made.
Images to Skip allows you to select every 2nd, 3rd, 4th, 5th image, and so
on, for the batch (2D mode only).
Step Size [mm] - defines the step size in mm between the first and last
Philips and Neusoft Medical Systems 4535 673 83741
2D tools
These tools are available in the 2D Viewer.
Compare Series - when you load a series, they display in tiled mode (see
Image Layout below). The series are sorted one after the other.
• Click Compare Series to enter the compare mode. The window
changes so that each view port contains a series. Each series can be
independently manipulated.
• Click a Series Layout button to select a vertical or horizontal layout of
the compared series.
Image Layout tools - Four arrangements are available for displaying images
(except in Compare mode). From left to right, they are 1x1 (which is used
to view images in a stack mode), 2x2, 3x3, and 4x4.
The layout in Compare mode can be 1x2 or 2x1 (the layout icons change
accordingly).
Flip/Rotate tools - The first two tools flip the axial image top-to-bottom
Philips and Neusoft Medical Systems 4535 673 83741
and left-to-right. The third tool rotates the image in 90 degree increments
clockwise.
Selection Mode - Use these tools to select individual images or all images in
order to leaf, zoom, pan, and change the Window or Level.
Note If you to want view images in larger arrangements, use the Film application.
MPR tools
MPR tools provide methods to alter the view of the images in order to
follow the path of a specific organ. These tools are available:
Show crosshair toggles crosshairs on or off.
Curved MPR icon toggles curve definition on or off.
Show Path displays the path drawn in curve definition mode.
Oblique displays the oblique path across the selected organ.
Thickness (not available in Normal mode) allows two methods to change
the thickness of the displayed slice:
• Type a value in the thickness box, and click Apply.
• Click the up or down arrows beside the box.
Threshold allows you to change maximum and minimum Hounsfield units
for the displayed image.
These options are available for changing the rendering technique.
• Normal (default, renders according to the average value along the ray;
thickness cannot be changed)
• MinP-minimum intensity projection
• AIP-average intensity projection
• MIP-maximum intensity projection
Above is the opening window after the patient image data is loaded into the
CT Endo application.
1 Tool panel contains the CT Endo tools, the Select Series tab, and the
common tool set.
2 Main viewport displays a rendered perspective image of the study. It is
blank until you define the anatomy to be rendered. The protocol is set
to the default for the application, but you can change it.
3 Reference viewports display axial, coronal, and sagittal images, with
yellow markings showing the camera direction and angle in the upper
and lower images, with the view plane in the center image. Green
crosshairs in each reference image indicate the camera’s target.
Step-by-step flythrough
The two middle buttons of the flythrough tool bar are for the step
forward and step back buttons. They allow you to go through the
structure one image at a time.
1 Use the combo box to change the Step size (between 1 and
20 mm per step).
2 Click either the step back or step forward button. The selected
button becomes active.
3 Position the mouse pointer on the main image onto the center
of the location you want to step toward.
4 Click the mouse. The image updates to your new camera
location.
5 Continue stepping as desired.
Note You can pass through a wall and leave the structure if you point at a wall.
Controlled flythrough
The two outer buttons are for moving through the structure in
automatic mode.
Note If you want to record your path, click the Record button before beginning.
1 Set the desired Step (from 1 to 50) and Step size parameters to
suit the structure you are viewing.
2 Click either the forward or backward button. The images will
begin updating automatically, advancing or receding according
to the speed and incremental distance you selected.
3 To stop the flythrough, click the left mouse button. To re-start,
click the left mouse button again.
4 When you are finished:
•stop the flythrough with the mouse button
•if you were recording the path, click the Record button off
•click the Stop button
Playback function
Note • If you want to save (record) the path of a flythrough, click the Record
button before you begin.
Philips and Neusoft Medical Systems 4535 673 83741
• Only one path will be saved in the Endo viewer’s memory. If you record a
second path, the previous one will be erased.
• If you close the Endo application, the path will be erased.
11.1 Overview
The Filming application is used for viewing, rearranging, windowing and
zooming images prior to sending them to be printed.
These are some of the advantages provided by the Filming application:
• Better organization and economy of films can be achieved by filming in
the Multiformat mode.
• Measurements and annotations (done with graphic elements) can be
added or deleted from the images.
Philips and Neusoft Medical Systems 4535 673 83741
Wa r n i n g Measurements on 3D images are in the screen plane and not on the three-
dimensional curved surfaces.
1 Workflow bar
2 Layout tools for printing and graphical notation and measurement.
3 Common tools
4 Menu bar includes additional filming options.
5 Main area displays the images and information selected for print.
6 Tabs allow you to toggle between the selected films.
Standard1×1
Standard4×5
Standard5×7
Location line form displays the first, last and a selected line. Alternately,
all lines display on the image.
Sort series allows you to restore images in the selected series to their
original order after they have been moved.
Film direction
Film can be printed in Landscape or Portrait format.
Flip/Rotate tools
The first two tools flip the axial image top-to-bottom and left-to-right. The
third tool rotates the image in 90 degree increments clockwise.
Selection modes
Image(s) to select one or several frames. These options allow you to make
Philips and Neusoft Medical Systems 4535 673 83741
Pan to center the feature of interest in the image frame by dragging the
image in the image window.
Gray bar to display a gray scale along the side of the image
Windowing - allows you to quickly set the Center and Width of the image
windowing. These options are available:
• P.F.
• Brain
• IAC
• Spine
• Bone
• Lung
• Abdomen
• Liver
Using the mouse for Windowing - You can also adjust the window Level
and Width by clicking and dragging the middle mouse button. Vertical
dragging changes windowing Level and horizontal dragging changes
windowing Width.
Titles off - Turn the image titles off and on. If turned off, they come back
on when the image is filmed or saved.
Philips and Neusoft Medical Systems 4535 673 83741
Reset All - For resetting the study’s images to the state they were in upon
loading.
11.2.6 Multiformat
In addition to standard format options, you have the option to combine
groups of images within a single film frame. Use the multiformat tools to
adjust the film view accordingly.
1 Turn on multiformat mode by clicking the Multiformat icon.
2 Hold Ctrl while clicking on the desired images.
3 Click a Multiformat icon. The system displays the new image frame in
the location of the first selected image.
1 When desired, you have two options to send the images from the
Filming application to the printer or imager:
• Click the DICOM Print icon (or select DICOM print from the file
menu).
• Click the Print icon.
2 From the dialog box, you can print the current (or all) film pages and
Philips and Neusoft Medical Systems 4535 673 83741
12.1 Overview
The Reporting package allows you to create customized reports using pre-
formatted templates. A template is a specially designed formatting
document that places the analytical information and images that you send
from an application into an organized report which can be printed and
saved.
Note Additional templates can be created. Contact your Philips representative
for more information on Report templates.
Philips and Neusoft Medical Systems 4535 673 83741
The system displays a new report with patient information from the current
patient. You can customize the report using the tools and options in the
tool panel.
New Report opens a new blank report in the default template format.
Add template allows you to modify the Select Report template list.
Image gallery
The Image gallery displays the images you’ve loaded to the reporting
application (see Report, on page 10-4 for more information on loading
images into the image gallery).
The Image gallery holds all loaded images until you remove them. Use this
procedure to remove images from the Image gallery:
1 Click on the desired image. A yellow box indentifies the selected image.
2 Press Delete on the keyboard to remove the image.
Philips and Neusoft Medical Systems 4535 673 83741
Wa r n i n g When you have completed a report, make sure you delete the images from
the image gallery.
Menu bar
The Menu bar includes these options:
File includes options for opening, printing and saving files.
Processing includes options for displaying the images.
View includes options for viewing the image and titles.
Windowing allows you to quickly set the Center and Width of the image
windowing using the standard windowing options.
Wa r n i n g Make sure that you have the correct patient information for the image you
have placed in your report.
13.1 Overview
Imaging performance of the scanner is checked by scanning head and body
system phantoms.
When testing image quality, the system should be properly calibrated.
This chapter covers information on these areas:
• Head and body phantom.
• Quality assurance checks-daily and monthly.
Read this section carefully and follow all instructions regarding scheduling
and performance of quality assurance checks.
Philips and Neusoft Medical Systems 4535 673 83741
Item Description
1 Head Phantom
2 Body Phantom
3 Physics Layer
4 Water Layer
5 Multi-Pin Layer
6 Copper wire
7 45 degree angle
Item Description
1 Physics Layer
4 Water Layer
5 Multi-pin layer
7 Lexan
8 Perspex
10 Teflon
11 Polyethylene
12 Body layer
13 Water hole
14 Teflon pin
Multi-pin layer using Head Std (B) protocol Water layer using Head Std (B) protocol
Monthly checks
Use your facility’s recommended schedule for monthly checks. These
procedures must be conducted at least once each month.
• Contrast scale and artifacts - on head phantom, multi-pin layer.
Advanced checks
Advanced checks are intended for use as advanced applications for
Physicists and your Philips and Neusoft Service Specialist.
• Impulse Response - on head phantom, physics layer.
• Slice thickness - on head phantom, physics layer for all slice
thicknesses.
Head scan
1 Position the water layer of the head phantom in the center of the scan
circle.
2 Perform a scan using the Head Std (B) protocol with scan and
reconstruction parameter values in the following tables:
Reconstruction parameters
Thickness 14
3 Check the four resulting images to ensure they are free of artifacts.
Note If the images display with artifacts, check that the water phantom is the
only item contained within the scan field, and repeat the procedure.
If there is still a problem, contact your Philips and Neusoft Service
Specialist.
Reconstruction parameters
Thickness 7
3 Check the four resulting images to ensure they are free of artifacts.
Note If the images display with artifacts, check that the body phantom is the only
item contained within the scan field, and repeat the procedure.
If there is still a problem, contact your Philips and Neusoft Service
Specialist.
4 From these images, select one to conduct the remainder of the checks.
5 Check the absorption reading for the Teflon pin. It should be as
follows:
Head scan
1 Position the multi-pin layer of the head phantom in the center of the
scan circle.
2 Perform a scan using the Head Std (B) protocol with scan and
reconstruction parameter values in the following tables:
Reconstruction parameters
Thickness 14
Image quality
Check the quality of one image according to the following criteria:
• Check with the LINE tool that the diameter of the large Perspex pin is
50 ±1.75 mm.
• All resolution holes (seven rows) in the Perspex pin should be visible.
• Five of the six low contrast pins in the Aculon body should be
detectable.
Note As the phantom ages, the low contrast pin is more difficult to visualize.
Absorption readings
The readings for the absorption of the different pins must be as follows
(values in CT numbers):
Water 0±4
Nylon +100 ± 15
Polyethylene -75 ± 15
Teflon +950 ± 50
Perspex +120 ± 15
Polycarbonate +110 ± 15
• You may have to adjust the gantry tilt so that the phantom axis is
perpendicular to the rotation plane.
4 Adjust In/Out position of phantom to obtain better precision.
5 Perform a scan using the Body bone HR (B) protocol, with the
parameter values listed in the Daily Checks section, except FOV
(in mm) = 250.
6 Click Service.
7 Select Advanced.
8 Click CT Scan Evaluation. The system displays a window.
9 Select FWHM.
10 Set the appropriate line number and FOV.
11 Click and drag across the image to display the plan lines.
12 Adjust the lines to the desired angle and position.
13 Click Evaluate to measure the thickness.
14 Repeat the procedure for these slice thicknesses and tolerances:
Caution When cleaning the buttons and the inside of the Gantry opening, take care
to avoid leaking liquid inside.
Blood and contrast medium are health risks. Take appropriate health and
safety precautions when removing blood or residual contrast medium.
Caution Do not use detergents or organic solvents to clean the system. Strong
detergents, alcohol and organic cleaners may damage the finish and also
cause structural weakening.
Voltage 120 kV
Slice Thickness 7 mm
mA 160 mA
View 250 mm
50mm
D 〈 z〉-
CTDI100 = ∫ ------------ dz
– 50mm N ×T
CTDI100 B1 B2 B3 B4 B5 CTDIw
10 mm 1.05 1.05
7 mm 1 1
1 mm 1.85 1.78
3s 1.8 2.0
1.5 s 1 1
1s 0.75 0.53
160 mA 1 1
30 mA 0.18 0.2
Philips and Neusoft Medical Systems 4535 673 83741
80 kV 0.25 0.27
120 kV 1 1
Note Only one parameter can be modified each time. The configuration of other
parameters is typical value.
Voltage 120 kV
Slice Thickness 7 mm
MA 240 mA
View 250 mm
CTDI100 B1 B2 B3 B4 B5 CTDIw
10 mm 1.05 1.05
7 mm 1 1
1 mm 1.85 1.78
3s 3.5 3.5
2s 2.1 2.3
1s 1 1
240 mA 1 1
10 mA 0.04 0.05
Philips and Neusoft Medical Systems 4535 673 83741
80 kV 0.27 0.28
120 kV 1 1
Note Only one parameter can be modified each time. The configuration of other
parameters is typical value.
7mm Slice
Philips and Neusoft Medical Systems 4535 673 83741
0.8 mm Slice
Royal Philips Electronics or in part, in any form or by any means electronic, mechanical or
otherwise, is prohibited without the prior written consent of the
copyright owner.
www.medical.philips.com
medical@philips.com X197
This Medical Device meets the provisions of the transposition of
the Medical Device Directive 93/42/EEC within the country of
the origin of the Notified Body concerned with the device.
Manufacturing address
Philips and Neusoft Medical Systems Co. Ltd. 4535 673 83741 * 06/2006
Hun Nan New District
110719 Shenyang
China
Copyright address
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
5684 PC Best
The Netherlands