Professional Documents
Culture Documents
Vita Troubleshooting Guide
Vita Troubleshooting Guide
Troubleshooting Guide
Confidential
0086
Use of Manual
CARESTREAM CR Systems are designed to meet international safety and performance standards. Personnel
operating the unit must have a thorough understanding of the proper operation of the system. This manual has been
prepared to aid medical and technical personnel to understand and operate the system. Do not operate the system
before reading this manual and gaining a clear understanding of the operation of the system. If any part of this manual
is not clear, please contact your Carestream Health representative for clarification.
Authorized Representatives
Carestream Health France
1, rue Galilée
93192 NOISY-LE-GRAND CEDEX
FRANCE
ii 2013-07-08 – 8H8443
Table of Contents
Description Page
iv 2013-07-08 – 8H8443
Safety and Regulatory Information
Introduction
The information contained herein is based on the experience and knowledge relating to the subject matter gained by
Carestream Health prior to publication. No patent license is granted by this information.
Carestream Health reserves the right to change this information without notice, and makes no warranty, express or
implied, with respect to this information. Carestream Health shall not be liable for any loss or damage, including
consequential or special damages, resulting from any use of this information, even if loss or damage is caused by
Carestream Health’s negligence or other fault.
Document Conventions
In this document the CARESTREAM CR Vita Family Systems will be referred to as “Vita CR”, “CR Scanner”, or
“System”.
Note
Notes provide additional information, such as expanded explanations, hints, or reminders.
Caution
Cautions point out procedures that you must follow precisely to avoid damage to the system or any of its components,
yourself or others, loss of data or corruption of files in software applications.
Important
Important highlights critical policy information that affects how you use this manual and this product.
8H8442 – 2013-07-08 1
SERVICE MANUAL
• The images provided by this system are intended as tools for the trained user. They are explicitly not to be
regarded as a sole incontrovertible basis for clinical diagnosis.
• Be aware of the product specifications and of system accuracy and stability limitations. Consider these limitations
before making any decision based on quantitative values. If you have any doubts, consult your Sales
Representative.
• This system is Class I continuous operated stationary equipment without applied parts and has one signal input/
output part.
• The appliance coupler of the flexible cord is used as a disconnecting device.
• In order to guarantee medical-grade leakage current, the computer that is connected to the system should be a
medical-grade computer or connected to the mains through a medical-grade isolation transformer.
• To achieve grounding reliability in USA installations, the equipment must be connected to an equivalent
receptacle marked “Hospital Only” or “Hospital Grade”.
Important
When the System is not in use, disconnect the power cable from the electrical outlet.
Electrical Hazards
Caution
• Do not remove or open system covers or plugs. Internal circuits use high voltage capable of causing serious
injury.
• Fuses blown within 36 hours of being replaced by a qualified technician may indicate malfunctioning electrical
circuits within the system. Have the system checked by qualified service personnel. Do not attempt to replace
any fuse.
• Fluids that seep into the active circuit components of the system may cause short circuits that can result in
electrical fires. Therefore, do not place any liquid or food on any part of the system.
• To avoid risk of electric shock, this equipment must only be connected to mains supply with protective earth.
Electromagnetic Emissions/Immunity
• Electromagnetic Compatibility Precautions
– Medical electrical equipment requires special precautions regarding electromagnetic compatibility (EMC).
Medical equipment must be installed and put into service according to the EMC information provided in the
following documentation.
• Communications Equipment
– Portable and mobile radio frequency (RF) communications equipment can affect medical electrical
equipment EMC performance.
• Replacement of Cables, Accessories or Transducers
– The use of cables, accessories or transducers other than those specified below with the exception of
transducers or cables sold by the manufacturer of the equipment as replacement parts for internal
components, may result in increased emissions or decreased immunity of the medical equipment.
• Other Equipment
– The Vita CR should not be used adjacent to or stacked on other equipment. If adjacent or stacked use is
necessary, verify normal operation in the configuration in which the scanner will be used.
• Cable, Accessory and Transducer Information for the Vita CR.
2 2013-07-08 – 8H8442
Safety and Regulatory Information
Port
Port Type Description From To Cable Type Cable Length
Power AC Power Vita CR AC Mains Unshielded 3 m (9.84 ft.)
Signal USB Vita CR PC Shielded 3 m (9.84 ft.)
8H8442 – 2013-07-08 3
SERVICE MANUAL
4 2013-07-08 – 8H8442
Safety and Regulatory Information
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a.Field from fixed transmitters, such as base station for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the scanner is used exceeds the applicable RF
compliance level above, the scanner should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the scanner.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 v/m.
Recommended separation distance between portable and mobile RF communications equipment and the
Vita CR
The Vita CR is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Vita CR can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communication equipment (transmitters) and the Vita CR as
recommended below, according to the maximum output of the communications equipment.
W Rated maximum output
power of transmitter m Separation distance according to frequency of transmitter
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
d = 1.17÷P d = 1.17÷P d = 2.33÷P
0.01 0.117 0.117 0.233
0.1 0.37 0.37 0.737
1 1.17 1.17 2.33
10 3.7 3.7 7.36
100 11.7 11.7 23.3
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
8H8442 – 2013-07-08 5
SERVICE MANUAL
Caution
• Do not operate the equipment in the presence of explosive liquids, vapors, or gases.
• Do not plug in or turn on the system if hazardous substances are detected in the environment. If these
substances are detected after the system has been turned on, do not attempt to turn off the unit or unplug it.
Evacuate and ventilate the area before turning off the system.
Overheating
Caution
Do not block the air circulation around the unit. Always maintain at least 15 cm (6 in.) clearance around the unit to
prevent overheating and damage to the system.
Laser Warning
• The Vita CR contains a Class 3 laser. During normal operation, always keep the unit enclosed in its protective
cover.
• Do not attempt to remove the cover. Only a qualified technician may remove the cover to service this product.
Complies with 21 CFR 1040.10 and 1040.11 except for deviations pursuant to Laser Notice No. 50, dated
June 24, 2007.
Class 1 Laser Product, and IEC/EN 60825-1.
Class 1 Equipment, and IEC/EN 60601-1.
Intended for continuous operation.
Product is provided with ordinary protection against the harmful ingress of water.
Not suitable for use in the presence of a flammable anesthetics mixture with air or with oxygen or with nitrous oxide.
The use of accessory equipment not complying with the equivalent safety requirements of this equipment may lead
to a reduced level of safety of the resulting system. Consideration relating to the choice shall include:
• Use of the accessory in the patient vicinity.
• Evidence that the safety certification of the accessory has been performed in accordance with IEC 60601-1 or
the system to IEC 60601-1-1 or local equivalent.
LIFTING HAZARD
The Vita CR weighs <36 kg (79 lb). Do not try to lift the unit by yourself.
Always seek assistance from another person. Lifting equipment that is
too heavy may result in serious injury and/or damage to equipment.
6 2013-07-08 – 8H8442
Safety and Regulatory Information
Safety
USA ANSI/AAMI ES6061-1:2005 -Part1: Medical Electrical Equipment General Requirements for safety
and essential performance
Canada CAN/CSA 22.2 No. 60601-1-08 - Medical Electrical Equipment-Part1:General Requirements for
safety and essential performance
European EN 60601-1: 2006 - Medical Electrical Equipment -Part1:General Requirements for safety and
Union essential performance
EN 60825-1: 2007 - Safety of Laser Products
International IEC 60601-1: 2005 - Medical Electrical Equipment -Part1:General Requirements for safety and
essential performance
IEC 60825-1: 2007 - Safety of Laser Products
EMC
IEC 60601-1-2: 2007 Medical Electrical Equipment - Electro-Magnetic Compatibility Requirements & Tests.
EN 60601-1-2: 2007 Medical Electrical Equipment - Electro-Magnetic Compatibility Requirements & Tests.
CE Conformity
This product conforms to the requirements of Council Directive 93/42/EEC. The system bears the following mark of
conformity.
0086
8H8442 – 2013-07-08 7
SERVICE MANUAL
Power On
Power Off
8 2013-07-08 – 8H8442
Safety and Regulatory Information
Labelling Summary
External Labels
Product data plate
- Vita/Vita LE
Laser-emitting product
High voltage
8H8442 – 2013-07-08 9
SERVICE MANUAL
In the European Union, this symbol indicates that when the last user wishes to discard this product, it must be sent
to appropriate facilities for recovery and recycling. Contact your local representative for additional information on the
collection and recovery programs available for this product.
10 2013-07-08 – 8H8442
How to Use the Vita CR System Troubleshooting Guide
Introduction
The purpose of this manual is to assist field and service engineers when providing service and supporting the Vita
CR Scanners.
This manual is a supplementary document to the Carestream Vita CR Service Manual 8H8442. Please review the
Service Manual before applying the procedures described in this manual.
The manual lists various types of problems and error messages and detailed explanations for suggested fix
directions.
General CR Failures
Provides information for general failures in the CR Scanner and how to fix them.
Image Artifacts
Provides images of all known artifacts, origin causes for each of them and what steps to take in order
to fix the problem.
Tests
Provides information for general tests that are used in some of the chapters above.
8H8443 – 2013-07-08 11
MISCELLANEOUS PUBLICATION
12 2013-07-08 – 8H8443
Vita CR Hardware Error Messages
8H8443 – 2013-07-08 13
MISCELLANEOUS PUBLICATION
14 2013-07-08 – 8H8443
Vita CR Hardware Error Messages
Recommended Actions
User Response
[1] Click OK.
The system resets itself.
[2] Insert the cassette again.
[3] Does the error repeat?
Yes No
Contact Service. Continue using the CR Scanner.
Service Response
[1] Install version V3.2 SP2 (V.3.2.2) or higher.
Note
This may be a faulty error message that has been solved by V3.2 SP2
[2] Remove the CR Scanner cover.
[3] Insert the technician key.
[4] Activate the linear motor by means of the Diagnostics screen.
8H8443 – 2013-07-08 15
MISCELLANEOUS PUBLICATION
Yes No
Check that the power supply is producing 24 V and Replace it according to
5 V according to the PS test, See “Controller Board “Replacing the Stall
Test Point - Voltages” on Page 67. Detector Wheel” section
in the service manual.
Yes No
Check the linear motor Replace the 24 V
connectors (J6) to the and 5 V power
controller board. OR supply unit.
Replace the controller board.
OR
Replace the linear board.
16 2013-07-08 – 8H8443
Vita CR Hardware Error Messages
Recommended Actions
User Response
[1] Reset the CR Scanner by turning it OFF and ON again.
[2] Does the error repeat?
Yes No
Contact Service. Continue using the CR Scanner.
Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Turn the CR Scanner OFF and ON.
[4] Activate the rotation motor by means of the Diagnostics screen.
[5] Does the rotation motor spin?
Yes No
a. Exit the Diagnostic a. Check all 5 cable connections on the laser driver board.
mode. b. Check that the orange flex rotation motor cable is connected to the optical head
b. Continue using the and all corresponding pins are in place (see the graphic following this table).
CR Scanner. c. Check the blue flex cable connector between the laser driver board (J1) and the
controller board.
d. Check if fuse F1 2.5 A has been burnt.
fuse F1
Yes No
Replace the laser driver board. Replace fuse F1.
In case the fuse burns again, replace the
blue flex cable.
8H8443 – 2013-07-08 17
MISCELLANEOUS PUBLICATION
Corresponding
Orange flex
pins
rotation motor cable
18 2013-07-08 – 8H8443
Vita CR Hardware Error Messages
Recommended Actions
User Response
[1] Click OK.
[2] Reset the CR Scanner by turning it OFF and ON again.
[3] Does the error repeat?
Yes No
Contact Service. Continue using the CR Scanner.
Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Check for an existing screen inside the drum that is blocking the W0 sensor.
[4] Clean the feeder rollers.
[5] Check that the W0 cable is connected to the erase LED board. If the cable is connected, proceed to the next
action.
[6] In some cases, the loader pin is not at the correct height and doesn’t pull out the screen from the cassette during
the loading process. Use Jig MP001520 to verify the loader pin height and adjust it if needed. See Loader Pin
Height Adjustment in the Service Guide.
8H8443 – 2013-07-08 19
MISCELLANEOUS PUBLICATION
[7] Check if the erase LED type is as shown in the picture below, replace the LED module.
[8] Check if the erase LED type is as shown in the picture below, replace the W0 sensor (Vita XE only).
20 2013-07-08 – 8H8443
Vita CR Hardware Error Messages
Recommended Actions
User Response
[1] Click OK to acknowledge the error message.
[2] Reset the CR Scanner by turning it OFF and ON again.
[3] Does the error repeat?
Yes No
Contact Service. Continue using the CR Scanner.
Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Check for an existing screen inside the drum that is blocking the W1 sensor.
[4] Replace the erase LED assembly.
[5] Replace the W1 bottom sensor.
8H8443 – 2013-07-08 21
MISCELLANEOUS PUBLICATION
Note
The error message will appear after 25 seconds of unsuccessful unloading sequence attempts.
Recommended Actions
User Response
[1] Click OK.
[2] Manually release the cassette using the release knob under the cassette tray.
[3] Remove the cassette.
[4] Reset the CR Scanner by turning it OFF and ON again.
[5] Insert the cassette.
[6] Does the error repeat?
Yes No
Contact Service. Continue using the CR Scanner.
Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Is the screen carriage stuck?
Yes No
a. Manually move the screen carriage to its homing position. Check/replace the Z0 sensor (“Sensors
b. Adjust the drum rails with Jig AS000723 according to the Tests” on page 65.
instructions in the Service Manual
c. Turn the CR Scanner ON.
d. Insert a cassette and see if there is any object interfering with
the screen carriage movement.
Yes No
Remove the object that is a. Replace the screen
interfering. carriage spring.
b. Replace the screen
carriage trolley.
22 2013-07-08 – 8H8443
Vita CR Hardware Error Messages
Recommended Action
User Response
[1] Click OK.
[2] Reset the CR Scanner by turning it OFF and ON.
[3] Does the error repeat?
Yes No
Contact Service. Continue using the CR Scanner.
Service Response
[1] Click OK to acknowledge the error message.
[2] Insert the technician key.
[3] Check/replace the Z4 sensor (see “Sensors Tests” on page 65).
8H8443 – 2013-07-08 23
MISCELLANEOUS PUBLICATION
Recommended Actions
User Response
[1] Reset the CR Scanner by turning it OFF and ON again.
[2] Insert the cassette again.
[3] Does the error repeat after unloading the screen?
Yes No
Contact Service. Continue using the CR Scanner.
Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Activate the stepper loader motor by means of the Diagnostics screen.
[4] Is there a mechanical noise while operating the motor?
Yes No
Check/replace the L1 sensor (see “Sensors Tests” a. Remove the CR Scanner base cover
on page 65). (refer to the Service Manual).
Or b. Verify that the loader stepper motor
Replace the controller board coupling is firmly fastened from both
sides.
c. Replace the loader assembly.
d. Perform leader assembly adjustment-
Adjust the pin height.
24 2013-07-08 – 8H8443
Vita CR Hardware Error Messages
Recommended Actions
User Response
[1] Click OK to acknowledge the error message.
[2] Reset the CR Scanner by turning it OFF and ON again.
[3] Does the error repeat?
Yes No
Contact Service. Continue using the CR Scanner.
Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Turn the CR Scanner OFF and ON.
[4] Activate the roller motor by means of the Diagnostics screen.
[5] Does the roller motor spin?
Yes No
a. Check the encoder reader cable a. Check the roller motor cable connector.
connection to the controller board b. Check that the power supply is producing 24 V and
(J1). 5 V.
b. Replace the roller encoder reader. c. Adjust the encoder and reader encoder of the
roller motor with Jig AS000703 and MP001890,
See Roller Reader Encoder Positioning in the
service guide.
Yes No
a. Replace the controller Replace the 24 V or 5 V
board. OR power supply.
b. Replace the roller
motor.
8H8443 – 2013-07-08 25
MISCELLANEOUS PUBLICATION
Note
The error message will appear if one of the above problems occurs.
Recommended Actions
User Response
[1] Check if the screen is in the cassette.
Yes No
a. Insert the cassette into the CR a. Insert the screen into the cassette. Make sure
Scanner. that you insert the screen right side up.
b. Go to Step 2. b. Insert the cassette into the CR Scanner.
c. Go to Step 2.
Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Insert the cassette into the CR Scanner.
26 2013-07-08 – 8H8443
Vita CR Hardware Error Messages
8H8443 – 2013-07-08 27
MISCELLANEOUS PUBLICATION
Recommended Actions
User Response
[1] Manually release the cassette using the release knob under the cassette tray.
[2] Reset the CR Scanner by turning it OFF and ON.
[3] Try to load the cassette again.
[4] Does the message appear again?
Yes No
a. Upload a different cassette. Continue working with the CR
b. Does the message re-appear? Scanner.
Yes No
Contact Service. a. There is a problem with the first
cassette. Inform Service.
b. Continue working with the CR
Scanner.
Service Response
Is it a cassette problem or a CR Scanner problem?
28 2013-07-08 – 8H8443
Vita CR Hardware Error Messages
Note
The error message will appear every 55 seconds after clicking OK.
Recommended Actions
User Response
[1] Reset the CR Scanner by turning it OFF and ON.
[2] Does the error repeat?
Yes No
Contact Service. Continue using the CR Scanner.
Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Go to the Settings>Diagnostics tab.
[4] Does the X1, X0 LED illuminate according to the linear position on the pictogram?
Yes No
Check/replace the right limit sensor. Check/replace the left limit sensor.
8H8443 – 2013-07-08 29
MISCELLANEOUS PUBLICATION
Recommended Actions
User Response
[1] Reset the CR Scanner by turning it OFF and ON again.
[2] Insert the cassette again.
[3] Does the error repeat after unloading the screen?
Yes No
Contact Service. Continue using the CR Scanner.
Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Activate the stepper loader motor by means of the Diagnostics screen.
[4] Does the stepper loader motor move?
Yes No
Is there any mechanical noise while operating the motor? a. Check the stepper motor connection J6
Yes No on the controller board.
Check/replace the L0 sensor a. Open the CR Scanner b. If it still does not work, replace the
(see “Sensors Tests” on base cover (refer to the loader assembly.
page 65). Carestream Vita CR
Service Manual).
b. Verify that the loader
stepper motor coupling is
firmly fastened from both
sides.
30 2013-07-08 – 8H8443
Vita CR Hardware Error Messages
“#69 The loaded cassette is not secured properly. Please insert the
cassette, making sure that the inserted cassette reached the end
Cassette size: <dimensions read>
Cassette SN: <serial number read>”
Scenario
When the cassette is inserted into the Scanner, the system will wait 1.5 seconds before starting the roller motor. This
will ensure the user pushes the cassette further into lock position (some users push the cassette slowly into the
reader slot and stop when the roller motor starts so the cassette might be still unlocked).
In Light-QC single-mode, the QC application will wait 5 seconds before closing, to ensure error #69 is displayed to
the user if needed. The message will indicate to the user to properly insert the cassette so the screen can unload
properly.
Recommended Actions
User Response
[1] Verify that the cassette you are attempting to insert into the system matches the same size and number reported
in the error message. Doing so will ensure normal operation.
Note
Clicking OK on the error message without verifying that the correct cassette is being used may result in:
• A problem while ejecting the cassette, and the possibility of error #41 occurring
• The screen will not fully return into the cassette
8H8443 – 2013-07-08 31
MISCELLANEOUS PUBLICATION
Recommended Actions
User Response
[1] Click OK.
The screen is returned to the cassette and the cassette is ejected.
[2] The CR Scanner operating status is changed to State stalled.
[3] Reset the CR Scanner by turning it OFF and ON again.
[4] Insert the cassette again.
[5] Does the error repeat after unloading the screen?
Yes No
Contact Service. Continue using the CR Scanner.
Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Activate the erase LED by means of the Diagnostics screen.
[4] Do any of the LED indicators light up red?
Yes No
Replace the erase
LED assembly.
[5] Check for 3.3 V at red wire of J7 connector with erase LED off.
No
Replace the controller board
[6] Check for 0 V at red wire of J7 connector with erase LED on.
No
Replace the controller board
32 2013-07-08 – 8H8443
Vita CR Hardware Error Messages
Recommended Actions
User Response
[1] Click OK.
[2] Turn the CR Scanner OFF and wait 45 minutes before turning it ON again.
Note
This is the time needed to reduce the temperature of the erase LED.
[3] Check room temperature according to site specifications.
[4] Insert the cassette again, after 45 minutes, and scan.
[5] Does the error repeat?
Yes No
Contact Service. Continue using the CR Scanner.
Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Activate the erase LED by means of the Diagnostics screen.
[4] Does the LED indicator light up red to the right of the word “Temperature”?
Yes No
Replace the LED assembly.
8H8443 – 2013-07-08 33
MISCELLANEOUS PUBLICATION
Recommended Actions
User Response
[1] Click OK.
[2] Manually erase the cassette after every scanning cycle.
[3] Contact Service.
Note
The erase LED assembly needs to be replaced.
Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Turn ON the CR Scanner.
[4] Activate the erase LED from the Diagnostic screen
[5] Does the LED indicator to the right of the word Current light up red.
Yes No
Replace the Erase LED assembly.
[6] Check for 24 V input at the F1 fuse on the erase LED assembly.
Yes No
Replace the Erase LED assembly. Replace the Erase 24 power supply.
34 2013-07-08 – 8H8443
Vita CR Hardware Error Messages
Recommended Actions
Service Response
[1] Click OK to acknowledge the error message.
[2] Insert the technician key.
[3] Operate the CR Scanner. Does the message appear:
Yes No
a. Check sensor cover is connected to Check the existing magnet on the system
J16 on the controller board. cover and its position
b. If it is, then replace the cover sensor
8H8443 – 2013-07-08 35
MISCELLANEOUS PUBLICATION
Recommended Actions
User Response
[1] Reset the CR Scanner by turning it OFF and ON again.
[2] Insert the cassette again.
[3] Does the error repeat after unloading the screen?
Yes No
Contact Service. Continue using the CR Scanner.
Service Response
[1] Go to the About screen.
[2] Check that the motion version is compatible with the CR Systems software.
Yes No
Replace the controller board. a. Update the motion version by means of the SW Update &
Backup screen.
b. Does the motion version update fail?
Yes No
Replace the controller board. Check the system operation.
36 2013-07-08 – 8H8443
Vita CR Hardware Error Messages
“No License Detected. Click OK for Settings Screen. Click Cancel to Exit”
Scenario
The error appears while trying login to the CR System software, and the proper license is not detected.
Recommended Actions
User Response
Verify that the proper licenses exist.
[1] In the CR System software, select Settings>Settings>About.
[2] Click License Manager.
[3] Check that the relevant licenses appear checked in the License Options window.
Service Response
[1] Open the iButton cover on the Scanner’s rear panel. It is next to the USB connector.
[2] Verify that there is an iButton inserted correctly into the socket.
[3] If the iButton is installed correctly, verify that the CR System software detects the iButton and the licenses. See
the User Response section directly above.
[4] If the CR System still does not detect the iButton or licenses.
(a) Replace the Internal USB cable.
(b) Replace the iButton electronic board.
8H8443 – 2013-07-08 37
MISCELLANEOUS PUBLICATION
(c) Replace the iButton - Verify that the iButton has the right licenses.
[5] Connect a WibuKey (license dongle) to the PC. Verify that a License is present on the System.
38 2013-07-08 – 8H8443
Software Error Messages
8H8443 – 2013-07-08 39
MISCELLANEOUS PUBLICATION
Error
No. Error Message Text Possible Reason and Solution
1 Cassette was ejected due to two minute time-out. User inserted a cassette, but didn’t scan within two
minutes. The SW is notified and generates an abort
command.
Insert the cassette again to perform a scan.
2 The following image(s) should be Accepted/ The user didn’t choose the Reject or Accept option.
Rejected first.
3 CR Scanner is ready for rollers cleaning. Please The technician/user needs to complete the rollers
insert cleaning tray and then insert cleaning cleaning process.
Plate.
4 Pull out the cleaning Plate and then release the The technician/user needs to complete the rollers
cleaning tray. cleaning process.
5 This patient was found in the Patient list. The ID number appears in the database.
6 The cassette does not match the selected sub- The cassette type is not compatible to the sub-
organ. Please either replace the cassette or re- organ or to BMD scanning type.
select sub-organ.
7 Can’t choose a sub-organ during scanning. No thumbnail is added.
Operation is cancelled.
8 Press Cassette Size button and select cassette. Barcode failure or unrecognized cassette barcode.
9 No available demo image for the selected sub- There is no existing demo image for the selected
organ. sub-organ in the database. Not all sub-organs have
demo images.
40 2013-07-08 – 8H8443
Software Error Messages
Error
No. Error Message Text Recommended Action
1 The machine is disconnected. Changes cannot 1. Check USB cable connections.
be saved, except the change to the scan mode. 2. Check power cord connections.
3. Restart the scanner. If error repeats call service.
8H8443 – 2013-07-08 41
MISCELLANEOUS PUBLICATION
Error
No. Error Message Text Possible Reason and Solution
1 Can’t choose BOTTOM before TOP was scanned Insert LLI 1—top before LLI 2—bottom or insert
or scan TOP twice. Operation is cancelled. LLI 2—bottom after scanning LLI 1—top.
2 The cassette is not an LLI cassette. Please either The cassette type is not compatible to the sub-
replace the cassette or re-select sub-organ. organ.
3 Can’t choose a sub-organ during LLI procedure. To choose a different sub-organ after the LLI scan,
Operation is cancelled. go to the patient list and choose the patient again
without a LLI procedure.
4 Can’t choose a sub-organ which is not for LLI If working within a LLI procedure, the user can’t
during an LLI study. Operation is cancelled. choose a sub-organ that is not part of a LLI
procedure.
5 Can’t choose LLI sub-organ after regular sub- To perform a LLI sub-organ, go to the patient list,
organ was chosen. Operation is cancelled. reselect the patient and perform only LLI sub-organ.
6 The cassette Barcode ID does not match the top Check that the Barcode ID label on the LLI 1
side. Operation is cancelled. cassette is readable.
42 2013-07-08 – 8H8443
Software Error Messages
Error
No. Error Message Text Recommended Action
1 Failed to store the DICOM send information for 1. Check the RIS network settings.
this image. 2. Repeat the RIS query.
2 Please note: Modality query has failed. 1. Check the RIS network settings.
Information retrieved from the last successful 2. Check the Modality Worklist settings.
query.
3. Repeat the RIS query.
3 Please note; Modality Query has failed. No access to RIS\HIS. Cannot present study list.
4 Note: Host name and Port number are invalid. DICOM Settings needs to be re configured.
MWL mode will be disabled until those values
configured via the DicomSetting tool.
8H8443 – 2013-07-08 43
MISCELLANEOUS PUBLICATION
EEPROM
Note
The error numbers in the list below are for user convenience only.
44 2013-07-08 – 8H8443
General Operation Failure
When the cassette is pushed, the CR Scanner does not load the screen
Suggested Solution
Is there a “Screen didn’t reach W0 Sensor” error message?
Yes No
Go to “#24 Screen didn’t reach W0 a. Open the cover.
sensor” on Page 26 b. Check Casette Presence Sensor (CPS) Sensors Tests on Page 65
Note
If one of the steps solves the problem, there is no need to continue with the following steps.
Barcode misreading
Workaround
Barcode misreading occurs when the system fails to read the cassette label to indicate cassette size.
To continue, use the manual cassette size setting as follows:
[1] Log in to Settings > User tab.
[2] Select the Display cassette size button check box.
[3] Load the screen. The cassette size list is displayed.
[4] Select the correct cassette size from the list.
Solution
[1] Remove the tray assembly (refer to the Service Manual).
[2] Check if the system has the most current software installed.
8H8443 – 2013-07-08 45
MISCELLANEOUS PUBLICATION
[3] Check if the flex cable is connected correctly to the electronic board as shown below.
[4] Check that the barcode grounding cable is attached to the barcode reader bracket.
Note
Record this update in the Mod label as Mod 2. In case one of the above updates is not implemented, update the
system according to AT000243.
[5] Make sure the barcode assembly is secured well.
[6] Adjust the barcode position with Jig MP001692, See Barcode Reader Adjustment in the Service Guide.
[7] Test the barcode reader is positioned correctly according to the “Barcode Reader Test” procedure in the service
manual.
[8] Replace the barcode assembly.
The Screen is left outside the cassette after the unloading cycle
Solution
[1] Does this failure occur with every case
Yes No
a. Check Loader Pin height and its travel with Jigs Replace Cassette
MP001689 and MP001691.
b. See Loader Pin Height Adjustment in the Service Guide.
[2] Check Last software version update.
[3] Clean the Rollers.
[4] Replace the Loader Assembly.
46 2013-07-08 – 8H8443
General Operation Failure
The Screen is not fully uploaded to the drum and has slipped between the
rollers
Solution
[1] Clean the screen.
[2] Clean the Rollers.
[3] Replace the Rollers.
[4] Replace the Roller springs.
The scanner works as Vita, at 40 cassettes per hour, when it should work
as Vita XE, at 60 cassettes per hour
Checking that the system is recognized as Vita XE
The system is recognized as Vita XE under the following criteria:
[1] Existing new Erasing LED module
[2] Existing cable RS232
[3] License Vita XE applied on iButton
[4] Carestream CR SW is V3.2. SP 2 (3.2.2.) or higher
[5] Firmware FPGA and Motion version
To verify that these criteria are correct, go to the About Tab and check as follow:
8H8443 – 2013-07-08 47
MISCELLANEOUS PUBLICATION
[9] If all the above criteria are as they “Should be”, then the Vita system should be recognized by the software as
Vita XE, Check in Scanner information – Machine Type Vita XE, if the system type is not Vita XE then recheck
all the above again.
48 2013-07-08 – 8H8443
Image Artifacts
Cause
These dots (streaks) are caused by an Electrostatic Discharge (ESD). These ESD dots generate very short sharp
and high light sparks that are coupled into the photo multiplier through the blue filter. The ESD effect is induced by
several factors, mainly environmental conditions such as dry air and low humidity, and the rubbing of the screen by
the mechanical CR Scanner (rollers).
Recommended Actions
Clean the screen with a screen solution that contains an anti-static agent. See the User Guide and/or Service Manual
for cleaning instructions.
8H8443 – 2013-07-08 49
MISCELLANEOUS PUBLICATION
Cause
White dots appear on the scanned image. This results from accumulated dust.
Recommended Actions
Clean the screen. See the Service Manual for screen cleaning instructions.
50 2013-07-08 – 8H8443
Image Artifacts
Shifted Image
Cause
Lines and/or distorted image because of incorrect computer specifications.
Recommended Actions
[1] Check that the computer’s USB Chipset is the INTEL model.
[2] Reinstall the computer’s USB Chipset driver.
Note
The USB Chipset driver can be downloaded from the computer model Web site.
[3] Verify that the issue is with the computer by activating the CR Scanner via another computer.
[4] Replace the controller board.
Note
If one of the steps solves the problem, there is no need to continue with the following steps.
8H8443 – 2013-07-08 51
MISCELLANEOUS PUBLICATION
Cause
The BSM algorithm failed to detect the collimators blades (in the above left sample image, the right hand side blade
was not detected).
• In cases where an apron is used to collimate the radiation, the curved line will not be detected by the BSM
automatically. It will leave a white collimation area (see sample image on the above right).
• Implants in the clinical body might be detected as collimation border and the clinical image will be cropped.
Recommended Actions
[1] Reject the image in order to prevent it from being sent to the PACS.
[2] Browse to the Image Storage screen.
[3] Locate the study/image and select Reprocess.
[4] Select the Edit Mask checkbox to set the BSM manually.
[5] Click and drag the BSM blades to redefine the Region of Interest (ROI).
[6] Once defined – click Process. The software will apply BSM the only to the area within the frame.
[7] Accept the image in order to send it to the PACS.
Note
For details on using Black Surround Mask, see the Vita CR System Software User Guide.
52 2013-07-08 – 8H8443
Image Artifacts
Note
The size of the grid lines varies with the magnification.
Cause
Grid lines on the image.
Recommended Actions
[1] Go to the Technician>Settings>Setup screen.
[2] Check that Grid Suppression Filter is selected.
[3] Check that the grid is 40 lines per cm (103 lines per in.).
8H8443 – 2013-07-08 53
MISCELLANEOUS PUBLICATION
Inverted Screen
Cause
The screen was inserted the wrong way in the cassette.
Recommended Actions
[1] With the Screen Extraction Tool, remove the screen from the cassette.
[2] Re-insert the screen with the correct side up.
54 2013-07-08 – 8H8443
Image Artifacts
Cause
The USB Chipset on the computer mother board is faulty.
Recommended Actions
[1] Check that the PC USB Chipset is the Intel model.
[2] Reinstall PC USB Chipset driver. It can be downloaded from the PC model Web site.
[3] Verify that the issue is with the PC by activating the CR Scanner via another PC.
Note
If one of the steps solves the problem, there is no need to continue to the following steps.
8H8443 – 2013-07-08 55
MISCELLANEOUS PUBLICATION
Cause
The image appears with random vertical lines all over the image. It may be caused by electrical interference such as
an unstable External power supply - 220v.
Recommended Action
[1] Connect the PC to UPS which will stabilize the electrical power supply.
[2] Replace the Blue flex cable.
Note
For more information, see Electrical Hazards in the CARESTREAM Vita Safety and Regulatory Guide (8H8439).
56 2013-07-08 – 8H8443
Image Artifacts
Interference from
radio therapy generator
Cause
A radio therapy generator may be near the Vita system and was influencing the system while it was switched On.
Recommended Action
Clear Vita system environment from any powerful generator. For more information, See Electrical Hazards in the
Safety and Regulatory Guide (8H8439)
8H8443 – 2013-07-08 57
MISCELLANEOUS PUBLICATION
Cause
The image appears black after image processing even though the raw image is fine. This can be caused by the
monitor having the incorrect resolution or the wrong DICOM tag.
Recommended Actions
[1] Check that the monitor settings are 1024 x 768 with a frequency of 74 Hz.
[2] For a high resolution monitor (1m-3m), select grayscale.
[3] Ensure that the DICOM tag (00028,00103) pixel representation value is “0” or “1”.
58 2013-07-08 – 8H8443
Image Artifacts
Example of Image
Cause
The image become with white spots even after exposure- no image at all was showed.
Recommended Action
Replace PM board
8H8443 – 2013-07-08 59
MISCELLANEOUS PUBLICATION
Cause
The image appears with a white band artifact across the image. It may be caused due to damage on the screen or
to an interfering object in the Optical laser head assembly.
Recommended Action
[1] Remove the screen from the cassette, check and clean the screen else replace it.
[2] Replace the optical laser head assembly
Note
After replacing the optical head assembly you must recalibrate the system: Offset and E-gain calibration.
60 2013-07-08 – 8H8443
Image Artifacts
Cause
The image appears with a white band artifact across the image. It may be caused due to an interfering object in the
Optical laser head assembly.
Recommended Actions
Replace the optical laser head assembly.
Note
After replacing the optical head assembly you must recalibrate the system: Offset and E-gain calibration.
8H8443 – 2013-07-08 61
MISCELLANEOUS PUBLICATION
Cause
The image appears with horizontal lines all over the image. It may be caused by electrical failure from one of the
system components.
Recommended Action
[1] Replace the PM tube.
[2] Replace the Blue flex cable.
Note
After replacing the PMT assembly, you must recalibrate the system, Offset and E-gain calibration.
62 2013-07-08 – 8H8443
Image Artifacts
Cause
Only part of the image was scanned due to an electrical problem that stopped from the system from completing the
scan.
The electrical malfunction was caused by damage or a missing grounding spring in the Tray:
Recommended Action
Replacing the tray's grounding springs.
8H8443 – 2013-07-08 63
MISCELLANEOUS PUBLICATION
Cause
The small marks on the image are caused by outside Electrical lamps of the light box system which was placed 15
cm. above the Vita and generated electromagnetic noise which interfered with the Vita optical system.
The image artifacts disappeared by just turning off the light box.
Recommended Action
Clear the Vita system environment from any powerful light generator.
Note
For more information, see Electrical Hazards in the CARESTREAM Vita Safety and Regulatory Guide (8H8439).
64 2013-07-08 – 8H8443
Vita CR Troubleshooting Tests
Sensors Tests
Introduction
Perform this test to check if a sensor is functioning. This test is valid for all sensors.
Procedure
[1] Remove the CR Scanner cover.
[2] Turn the CR Scanner ON.
[3] Go to Settings>Diagnostics tab.
[4] Insert a piece of paper into the sensor where the flag passes during the procedure. This blocks the sensor
signal. The green LED adjacent to that sensor on the pictogram should light up if the sensor is functioning
properly. If the specific sensor that is blocked does not light up, the sensor is not functioning properly.
[5] Remove the piece of paper to unblock the sensor.
The green LED should turn OFF.
8H8443 – 2013-07-08 65
MISCELLANEOUS PUBLICATION
Note
To perform this test on the W0 or W1 sensor: The sensor is divided into two parts; the top is the transmitter and the
bottom is the receiver. Insert a screen into the drum and check the W0 or W1 sensor LED position change. During
the screen upload, the LED should change from green to OFF.
F1
F2
F3
66 2013-07-08 – 8H8443
Vita CR Troubleshooting Tests
There are six test points located on the controller board that confirm the power supply voltages and the controller
converting voltage status. These points can be tested using a digital voltmeter (DVM).
8H8443 – 2013-07-08 67
MISCELLANEOUS PUBLICATION
TP2
TP9
TP13
TP10
TP14
TP11
In addition to the test points, the controller board indicates normal voltage status using LED indicators.
(D10)+15 V 5 V (D11)
(D13)3.3 V 24 V (D12)
-15 V (D17)
(D11)+5 V
If one of the voltages does not fall within the ±10% definition range listed in the Power Supply section in the
Diagnostic screen, the LED indicators on the controller board will not light up.
68 2013-07-08 – 8H8443
Vita CR Troubleshooting Tests
8H8443 – 2013-07-08 69
MISCELLANEOUS PUBLICATION
Important
The LED indicators should be turned ON using the DIP switch only when testing the controller board function.
Make sure to turn the LED indicator(s) OFF following the test to prevent any light from penetrating the system during
a scan cycle.
Note
To initialize the system, plug in the interlock key and then press the Reset button once.
70 2013-07-08 – 8H8443
Vita CR Troubleshooting Tests
D2
D6
SW1
SW2
SW3
D19
8H8443 – 2013-07-08 71
MISCELLANEOUS PUBLICATION
72 2013-07-08 – 8H8443
Electric Schematic Diagrams
8H8443 – 28JUL2012 73
MISCELLANEOUS PUBLICATION
74 28JUL2012 – 8H8443
Electric Schematic Diagrams
Sensors Schematics
8H8443 – 28JUL2012 75
MISCELLANEOUS PUBLICATION
76 28JUL2012 – 8H8443
Publication History
Publication History
2013-07-08 – 8H8443 77
CARESTREAM and VITA are trademarks of Carestream Health, Inc.
78 2013-07-08 – 8H8443
Publication History
Printed in U.S.A.
150 Verona Street
Carestream Health
Rochester, N.Y. 14608
United States
CARESTREAM is a trademark of Carestream Health.
© Carestream Health, Inc., 2010
2013-07-08 – 8H8443 79
80 2013-07-08 – 8H8443