1 Introduction
ar1200 view combines reliability,
®
modularity, versatility and capacity to be updated which characterise the latest
generation of cardioline electrocardiographs.
ar1200 view is a portable electrocardiograph with dual power supply, (mains and rechargeable internal batteries),
which in the basic configuration will:
record an ECG exam in automatic or manual mode;
reproduce the ECG signal on 120 mm paper in 3-6 channel format thanks to the high resolution thermal printer;
storage of the most recent recording in automatic mode and print additional copies.
"ECG signal interpretive option": a useful and dependable diagnostics support provided by the “HES” program;
"arrhythmia option": program enabling detection of arrhythmia events during continuous recording;
“HRV analysis option”: program enabling detection of variations in heart rate;
"L.E.M.S - PC archive option": to store the exam to archive on a Personal Computer running “L.E.M.S.” (Local ECG
Management System) software;
"R.T.E. - PC-ECG option": for real time display of the twelve leads on your computer screen to allow management of patient
medical records and archiving of exams in digital format using " R.T.E" (Real Time ECG) software. R.T.E has an optional
"HES" module for automatic interpretation of the ECG signal.
comply withcan
regulations safety regulations;
cause all other
physical harm to thedevices connected and
patient connected respond
to thelikewise to regulations.
person operating Non-compliance
the device. Should it bewith these
difficult to
obtain the necessary information for assessing the risk of the individual connections, apply directly to the manufacturers
concerned or avoid making the connections.
In the event of other equipment being connected directly or indirectly to the patient, check for the possible risks caused by the
sum of the leakage currents on the body of the patient.
The device is protected against defibrillation discharges in accordance with IEC standard 601-1-25; to ensure that the signal is
restored, use only original electrodes or electrodes responding to IEC and AAMI standards.
If an electrosurgical scalpel is in use, the patient cable should be disconnected from the device.
In any event, the greatest care should be taken when using defibrillators or high-frequency surgical devices at the same time
as the ECG. If you have any doubts while using such devices, disconnect the patient from the electrocardiograph temporarily.
The device recognizes the impulses generated by a pacemaker and does not interfere with its operation, as prescribed by
standards in use at the time of drafting this manual.
Avoid exposing the equipment to extreme temperatures, excessive dust or dirt, and very salty or damp environments; observe
the ambient conditions described in detail under the "Technical specifications” heading.
Periodically check the efficiency of all accessories and of the device itself. Use the built in test function to perform an initial
efficiency check. Contact the Authorized Assistance Centre whenever the device seems to be operating irregularly.
To prolong the life of your ar1200view, have it periodically checked at an Authorised Assistance Centre
Warning: do not use the device in the presence of anaesthetics or volatile gases!
Warning: the indications obtained using automatic interpreting programs or other diagnostic aids must be reviewed and
countersigned by a qualified medical person!
Warning: the device is provided with an IR interface for the transfer of data to other devices. The IR interface must not be
masked , even accidentally, as this will adversely affect its capability and its operation, interrupting and preventing the correct
flow of data.
Warning: Environmental protection: When no longer in use the device must be disposed of according to local regulation, do not
dispose as ordinary refuse.
The manufacturer will acknowledge liability for the safety, reliability and functional efficiency of the device only if:
• modifications and repairs are performed by the manufacturer or by an Authorized Assistance Centre;
• the a.c. mains power supply of the premises in which the device is used corresponds to current regulations;
• the device is operated according to user instructions;
• any accessories in use are those approved by the manufacturer.
2.4 Keyboard
The keyboard card consists of the device functional keys and LED messaging devices.
2.5 Printer unit
This consists of the thermal printer head support and the mechanical elements required to position it correctly.
2.6 IR Module (RS232)
The function of the IR module is to transmit and receive data from an external PC.
http://slidepdf.com/reader/full/cardioline-ar-1200-service-manual 7/38
8/12/2019 Cardioline AR 1200 - Service Manual
http://slidepdf.com/reader/full/cardioline-ar-1200-service-manual 8/38
8/12/2019 Cardioline AR 1200 - Service Manual
AR1200VIEW
7
: . * 2
+0 ** ** ** ** ** ** ** ** ** ** ** **
1 /-
*
Note: The 0.5 Hz pitch linear phase anti-drift filter is always on and cannot be switched off.
The 50 or 60 Hz filter eliminates modified notch digital type mains disturbances in the linear phase, with a frequency
response of 32 Hz - 3dB.
6 Functional blocks
6.1 Mother board
The motherboard is based on the following principal components:
6.1.1 Processor
32 bit Fujitsu MB91101 RISC microprocessor with 12 MHz clock
2 x 4 Mb static rams
2 x 8 Mb flash memories including the following software:
• boot code;
• operating system software;
• applications software;
• calibration data;
• ECG archive.
#++* 5
Warning: Changing mains voltage imply the replacement of the mentioned label. Only original ones can
be used to assure proper electrical isolation.
In order to change mains voltage move the 3 jumper as illustrated in the picture below.
The Manufacturer declines all responsibility for any damage caused as a result of tampering.
If the mains on led does not light up, check using the following procedure:
Disconnect the mains cable;
Check the externally accessible mains fuses;
Disconnect the battery from the device as indicated in chapter 8.4;
6.3 Battery
NiMH battery 12V-1800 mAh
Attention: In order to ensure a correct equipment operation and safety, replace the battery with original et
medical devices battery pack.
Warning: The battery must only be removed if the device is off and the mains supply cable disconnected.
Note: The operation of the device is not guaranteed if the battery is flat or missing.
Note: The configuration parameters and any stored ECGs are not lost when the battery is changed.
Note: Update the system date and time after replacing the battery.
Battery check
18 function keys;
10 alphanumeric keys;
1 LED for mains line indication;
CPU connection through bi-directional synchronous serial port.
Testing the keyboard card circuits
Execution of keyboard self-test;
6.6 IR Module
The function of the IR module is to transmit and receive data from an external PC. The IR module can be used to perform the
following functions:
http://slidepdf.com/reader/full/cardioline-ar-1200-service-manual 15/38
8/12/2019 Cardioline AR 1200 - Service Manual
http://slidepdf.com/reader/full/cardioline-ar-1200-service-manual 16/38
8/12/2019 Cardioline AR 1200 - Service Manual
AR1200VIEW
EC signal disturbed or anomalous. Error message on display, on printout. See user manual.
The device stops during use. No functions operational, no Press reset button on right side of
commands accepted. device.
Blank display. Backlit lamp blown or step up Check for brightness regulation
converter failed Replace display assembly
To reassemble and
subassemblies the device, perform
connections arethe operations
performed described below in reverse order, ensuring that all
correctly.
8.2 Opening and closing the device (table T1)
Remove battery pack as per § 8.4;
Remove the 6 fixing screws from the mobile base (T1 Ref. 1);
Lift the mobile base of the device;
Remove the lower mobile part;
To reassemble the equipment proceed in reverse order.
• keyboard flat;
• thermal head flat;
• motor flat;
• mark sensor card flat;
• display connector flat;
• inverter flat.
To reassemble the motherboard proceed in reverse order.
Warning: Replacing the motherboard means that the paper feed speed and mark presence sensor must
be recalibrated.
Warning: Respect the correct polarity of the battery connection as specified in Table T1 ref.7,8.
The paper
driven compartment
by the motor of thedoor system is
equipment, a mechanical
which transportsdevice, inserted in the
the heat-sensitive casing, accessible from the outside,
paper.
It is located above the paper compartment and covers it completely.
To remove this part, proceed as follows:
Insert a tool in the special hollow in the left-hand wall of the equipment where the paper comes out and exert pressure upwards
so as to release the paper compartment door.
This assembly must be disassembled before inserting a roll or pack of paper, when cleaning the roller and always before
opening the equipment.
To reassembly it, slip it into place with the rubber roller towards the inside of the equipment and holding it up on the other side.
As soon as it is inserted, press gently down, on the raised side, so as to snap it shut.
Note: If the roller is not clean and the paper compartment door is badly inserted or fastened, the paper
transport is faulty and the equipment functions incorrectly.
Memory: a non-destructive
following type is thentest is performed
printed on the memories (the data in the memory are not cancelled), and a report of the
(all tests OK).
Language : CRC OK
Ram Memory : OK
Info: The following items of information are printed: model identification, serial number of the device, details of software
installed, version and language code, reference of any options installed.
SERVICE
To enter the self-testing menu, switch on the device, press the MENU key and select and confirm the TOOLS,
SELF-TEST, SERVICE(2) submenu using the arrow key (DOWN). When the access code is requested, press the
following keys: FILTER – AMPLITUDES – SPEED – START and confirm.
the following tests are available in the service section:
9.1.2 Mark calibration
Confirm the submenu MARK CALIBRATION (1), position the mark at a distance from the sensor (visible near the
exit from the device), and confirm “enter”
The microprocessor automatically reads the sensor output voltages, voltage with white paper and black mark, and
then sets the digital trimmer that regulates the photodiode current to the optimal settings and then stores them.
The mark calibration voltages are shown on the display, as is the numerical position of the digital potentiometer.
9.1.3 Speed calibration
Testing and setting
Select and firm the SPEED CAL. (2) submenus, and the speed to calibrate
The following messages then appear on the display: Output PWM0 Output – Freq = xxxx Hz with the possibility
of selecting from: test – change – exit.
Confirming the “test” command enables the paper feed and the print unit emits a 1 impulse per second signal. If
the speed has to be calibrated, select “change” and increase or decrease the frequency to set the paper feed
speed entered. When the speed has been set correctly, select and confirm with “exit”.
The permitted tolerance values for the various paper feed speeds are as follows:
50 mm/s ± -5%
25 mm/s ± -5%
5 mm/s ± -10%
9.1.4 Metres of paper printed
Displays the number of metres of paper printed.
9.1.5 Firmware upgrades
Enables upgrading of the instrument’s internal firmware.
The firmware upgrade menu can be accessed from the main menu, pressing the Shift + S keys together.
Note: Refer to the “Loader” programme manual for further information.
9.1.6 Print Screen
Enable print screen ( for internal use only)
It is highly recommended to leave this function disabled.
9.2 Print head alignment
In case of print head adjustment remove mother board as per § 8.3, then follow instructions provided in table T4,
repeat adjustment until the best result is obtained.
10 Periodic maintenance
The ar1200 view electrocardiograph has been designed to assure a high degree of reliability during the life cycle
of the product.
Any incorrect conditions of use or anomalous operation will be signalled by messages on paper and on the
display.
10.1 Inspection frequency
To guarantee a long and safe duration, the instrument and its accessories must be periodically inspected and
checked.
Table 9 indicates the type and frequency of controls required, based on normal use of the electrocardiograph
(about 4000 ECG recordings per year).
Warning: Check immediately if there are events not referring to normal use.
Note: the device cannot be sterilized! The electrodes can be sterilized with ethylene oxide.
Proceedas follows:
open the paper feed;
pass the cloth over the dots of the thermal head without pressing too hard.
Warning: the printer thermal head is extremely sensitive to electrostatic potentials. Do not touch it for any
reason. If necessary, handle it after connecting to a ground via a suitable protective bracelet or belt.
List of spare parts for the ar2100 view device code 80600062
Spare part code Description Ref Note
69700788 FUSE 0,5 AT 5X20 MM (10PCS) Installed on mains plug
69701148 FUSE 5 A SMF SLO-BLO SMD (10PCS) Installed on mother board
69701307 BLACK MARK PHOTOSENSOR BOARD AR 600/AR 1200
69701352 BATTERY PACK (5+5) X 1,2V
69701354 MOTOR ASSEMBLY + GERAS AR 1200 II SERIE
69701358 PAPER SPINDLE + PAPER BOARD GUIDE AR 1200
69701360 MAINS PLUG SOCKET AR 1200
69701361 THERMAL PRINTER – WRITING SYSTEM AR 1200
69701363 MAINS POWER LABEL 115V E 230V
69701440 DISPLAY + GR. SUPPORTO AR 1200 VIEW
69701441 DISPLAY GLASS FOR AR 1200 VIEW
69701492 ELECTRONIC KEYBOARD BOARD AR 1200 VIEW EXT.
69701502 MOTHER BOARD+IRDA AR1200VIEW EXT C/LINE
Table 12.1
* The replacement motherboard is supplied complete with basic software in English and S/N 00000000.
To reinstall the configuration before the replacement of the motherboard, the following data must be supplied:
device code number (REF)
serial number (SN)
language
options purchased
et medical devices will send the corresponding firmware (binary file to load through the Loader application), which will install the correct S/N
and enable the options previously purchased.
Warning: Once the firmware has been reloaded with a specific S/N it cannot be further modified.
Note: The data requested can be printed out directly by the instrument, using the info function on the
self-test menu.
Note: Do not install the drivers on the cd rom supplied on the PC. If the drivers should have been
inadvertently installed, uninstall them.
http://slidepdf.com/reader/full/cardioline-ar-1200-service-manual 25/38
8/12/2019 Cardioline AR 1200 - Service Manual
AR1200VIEW
ECG interpretation
arrhythmia
HRV
PC archiving
PC ECG
Battery capacity Internal battery: 100 min. with printing in 3 channels mode
Recharging time Internal battery: 18 hours
Housing protection category IP 20
Environnemental conditions:
- operation Environment temperature: from +10 °C to +40°C
15 APPENDIX A
15.1 Procedures for manipulating and storing components (ESD)
All modern electronic components, particularly those based on CMOS technology, may be irreparably damaged
by even modest electrostatic discharges.
Precautions must be taken against electrostatic discharges when handling and working on electronic components
sensitive to electrostatic discharges. ELECTROSTATIC SENSITIVE DEVICE (ESD).
Staff involved in checking, warehousing, shipping and assembly operations must be earthed through a suitable
conductive bracelet compliant with safety standards. If this precaution cannot be taken, the operator must wear
suitable antistatic shoes or boots.
Work equipment must be earthed.
Tables, work surfaces and other surfaces on which components are handled must be covered in conductive
material and earthed.
All tables and work surfaces must be covered with a layer of conductive material and earthed.
The person performing the repair must be earthed with a suitable protective bracelet compliant with safety
regulations.The material must be contained in suitable antistatic bags or containers, and labelled according to MIL
STD 129J. The containers must guarantee adequate protection against impact and handling during transport.
All E.S.D. components must be stored in their original boxes and kept in special metal containers. While stored in
the warehouse, electronic components must be kept in their original packaging.
Any containers used must be exclusively in metal and/or conductive material. If handled directly, the personnel
must adopt the precautions specified above.
During handling cards must be stored in suitable antistatic containers.
Each component sensitive to electrostatic discharges will be identified by the abbreviation ESD.
16 APPENDIX B
16.1 Figures and illustrated tables
* 0=
*
* =
*"
* D !
* *
et me dical
de vic es
VISTA DA SOPRA
TOP VIEW
T1
ar v ie w
a r 120 0 view
SPORTELLO CARTA
PAP ER COMPARTMENT INGRESSO PAZIENTE TASTO RESET
PATIENT INPUT RESET KEY
VISTA DA SOTTO
BOTTOM VIEW
1 5 6 1
2 BATTERY PACK
1
1 3 7 8 9
1
http://slidepdf.com/reader/full/cardioline-ar-1200-service-manual 31/38
8/12/2019 Cardioline AR 1200 - Service Manual
AR1200VIEW
et me dical
de vic es
VISTA INTERNA - FISSAGGI
INSIDE VIEW - FIXING
T2b
6 KEYBOARD
2 2 3
5 2 5 2 5 4
et me dical
devices
VISTA PARTICOLARI
DETAILS VIEW
69701354
T3
CHIMO GRAFO
MOTOR ASSEMBLY
69701358
69701515 69701361
FLAT S TAMPANTE
FLAT THERMAL HEAD
et me dical
de vic es T4
A) con
Fissare
viti eil rondelle
gruppo stampante
. ( fornito già tarato )
B) Eseguire le connessioni elettriche,
chiudere l’apparecchio e verificare la qualità di stampa.
If the pri ntout is not un iform on all the paper width, do the following:
et me dical
d evic es
TEST POINT TESTA TERMICA
THERMAL HEAD TEST POINT
10 8 mm
T5
333 nSEC
CLOCK (3MHz)
PIN 6 DI J49
DATA
PIN 2 4 DI J49
LATCH
PIN 7 DI J49
1 msec.
470µsec.
STROBE 1 e 2
PIN 8 e 9 di J49
470µsec.
STROBE 3 e 4
PIN 22 e 23 di J49