Professional Documents
Culture Documents
Lab Report Experiment 10 CHM3100
Lab Report Experiment 10 CHM3100
Lab Report Experiment 10 CHM3100
Objectives :
Assignment 1
Answer:
�2 − �1
�=
�� – �1
0.401−0.380
� = 0.410–0.380 = 0.7
The suspicious measurement, 0.380 should not be rejected because it meets the
Answer :
�2 − �1
�=
�� – �1
0.400−0.380
� = 0.413–0.380 = 0.606
The suspicious value, 0.380 is rejectedbecause it doesnot comply the Q test at 90% confidence level.
Assignment 2
to one decimal place and added water until the water reached the 400
mL line. This was repeated 8 times and recorded the following masses of
water 403.2g, 402.7g, 399.6g, 401.4g, 397.3g, 400.1g, 395.6g and 398.3g.
�� − � 2
�2 =
n−1
S2 : sample variance
400− 399.8 2
�2 = 8−1
= 0.0057
�� − � 2
�=
n−1
�= 0.0057 = 0.0755
S
�������� ����� =
�
S : standard deviation
n : number of observations
0.0755
�������� ����� = 8
= 0.0267
�2 − �1
�=
�� – � 1
397.3 −395.6
� = 403.2 – 395.6 = 0.22
Assignment 3
You are to measure Ca2+ concentration associated with aerosol particles collected
detector. You prepare 8 standard solutions for your calibration curve. The
a. In Microsoft Excel, plot the calibration data as Absorbance versus Concentration (M units).
(Note : 1 ppm = 0.00002495 M)
b. Define the linear region of your calibration curve. Using the trendline function, generate
the equation and standard deviation of the calibration curve for the linear region.
[Based on the graph in (a)]
c. Using your calibration curve equation, calculate the calcium concentrations (M units) for the four
samples. Remember, your calibration curve has limitations.
y = 0.1782 x – 0.1038
Sample A : Sample C :
x = 2.08 M x = 6.70 M
Sample B : Sample D :
x = 6.87 M x = 4.84 M
Sample Concentration (M) Average Absorbance
(A)
A 2.08 0.266
B 6.87 1.121
C 6.70 1.09
D 4.84 0.759
d. For which samples are you able to determine the calcium concentration with your calibration
curve? Which ones cannot be determine? Explain.
Sample A, B and C can be determined with the calibration curve but sample D cannot. This is
because the value of sample D exceed the limitation of the calibration curve. For example, the
average absorbance of sample D is 0.759 therefore, the calcium concentration should be within
5.0 ppm to 6.0 ppm.
e. What modifications must you do to your samples or calibration curve so that you can accurately
measure the absorbance and calculate the correct calcium concentrations for all samples?
First, increase the accuracy by increasing the number of sampling calibration. Next,
the accuracy of some methods can be improved by adding a suitable internal standard to
both calibration standards and test samples and basing the regression on the ratio of the analyte
response to that of the internal standard. The use of an internal standard corrects for small
variations in the operating conditions. Other than that, the average absorbance and calcium
concentration need to be adjusted. For instance, the average absorbance should be within 47.74
to 57.74 hence we can get the calcium concentration within 5.0 ppm to 6.0 ppm.