Professional Documents
Culture Documents
Comparison: 5.1 Comparison With Previous Model
Comparison: 5.1 Comparison With Previous Model
Comparison
5.1 Comparison with previous model
Hitachi Hitachi Hitachi
Item
F-7100 F-7000 F-2700 (4J1-0008)
Model
11 Wavelength accuracy ±1 nm ±1 nm ±2 nm
Available Available
13 Micro Plate Not available
(option) (option)
Available Available Available
14 Integrating shpere
(option) (option) (option)
Spectral correction Available Available Available
15
for long wavelength range (option) (option) (option)
295, 320, 370, 395, 420nm 295, 320, 370, 395, 420nm
295, 320, 370, 395, 420nm
17 Cut-off filters (optional filter set) (optional filter set)
(optional filter set)
Automatic Filter option Automatic Filter option
20 Weight (kg) 41 41 41
1/4
Hitachi High-Tech Science Corporation
Sales Information
F-7100 Fluorescence Spectrophotometer
Model
1 Sensitivity 360 Peak to Peak 350 Peak to Peak 300 Peak to Peak
Sensitivity
2 20000 RMS (No information) 5000 RMS
(BG method)
3 Wavelength Scan speed 60,000 nm/min 60,000 nm/min 60,000 nm/min
2/4
Hitachi High-Tech Science Corporation
Sales Information
F-7100 Fluorescence Spectrophotometer
Thermo
Hitachi Edinburgh Horiba Perkin Elmer Agilent
Parameter Scientific
F-7100 FS5 Fluoromax-4 LS-55 Cary Eclipse
Lumina
Stand-alone
1 Not available Not available Not available Not available Not available Not available
control
2 PC control Standard Standard Standard Standard Standard Standard
Sensitivity
3 (BG 20000 RMS 4000 RMS 6000 RMS 16000 RMS 4000 RMS 700 RMS
method)
Sensitivity
(No (No (No (No (No
4 (Peak 1200 RMS
information) information) information) information) information)
method)
Wavelength
5 60,000 nm/min 6,000 nm/min 6,000 nm/min 4,800 nm/min 1,500 nm/min 24,000 nm/min
Scan speed
Wavelength
6 60,000 nm/min 20,000 nm/min 6,000 nm/min 4,800 nm/min 1,500 nm/min 24,000 nm/min
Drive speed
Dynamic (No (No
7 6 digits 4 digits 6 digits 6 digits
range information) information)
Mechanical
Czerny-Turner
ruling aplanatic concave
Czerny-Turner design with
concave gratings (No (No
8 Grating design with plane gratings
diffraction 20 cm focal information) information)
plane gratings 1200
grating: 900 length
groove/mm
lines/mm, F2.2
Wavelength
200 - 750nm EX:230 nm – 200 to 800nm
range
(standard) 1000 nm 200 nm – 870 on EX
9 with 190-1100 nm 200-900 nm
200 - 900nm EM:230 nm – nm 200 to 650nm
standard
(option) 870 nm on EM
detector
EX : 2.5 - 15nm
10 Band pass 1 - 20 nm 0.5 - 20 nm 0 to 30nm 0 to 30nm EM : 2.5 - 1.5 - 20 nm
20nm
Wavelength
11 ±1 nm ±0.5 nm ± 0.5 nm ± 0.5 nm ±1 nm ±1.5 nm
accuracy
Minimum
(No (No (No
12 Sample 0.6 mL 0.5 mL 0.5 mL
information) information) information)
volume
3/4
Hitachi High-Tech Science Corporation
Sales Information
F-7100 Fluorescence Spectrophotometer
EX: Bandpass
filters
(250–
395 nm, 335–
620 nm)
Cuttoff
295, 320, 370, filters
350 – 800 nm
395, 420nm (550
in 50 nm
Cut-off (optional filter 320, 435, 530 (No 290, 350, 390, nm, 695 nm)
16 increments
filters set) nm information) 430, 515nm EM: Bandpass
(optional filter
Automatic Filter filter
set)
option (250–
395)
Cutoff
filters
(295,
360, 430 and
550 nm)
4/4
Hitachi High-Tech Science Corporation