Method of Leak Compensation

From Patient
(Expiratory)

To Patient
(Inspiratory)
3L / min

Compare Insp to
Exp Base Flow

Flow
Automatic Leak Compensation L/min

• Determine a level of “Baseline”


support to be maintained. Baseline = 3L / min

• This will act as a buffer so that when (Bias Flow)

the amount of leak changes, the


“automatic” leak compensation system
has time to react.
e360AutoLeakPPT_WW Rev A 0312 3
Method of Leak Compensation
Leakage
(0.5L/min)
From Patient
(Expiratory)

2.5L/min
To Patient
(Inspiratory)
3L/min

Flow Leak Compensation


Automatic Leak Compensation GOAL L/min activated

• Recognize when the leak changes,


whether increasing or decreasing,
and adjust the leak compensation flow
accordingly, maintaining the 3 L/min
buffer. Baseline
dropped

e360AutoLeakPPT_WW Rev A 0312 4