Electrodes For PH Measurements: - When You Need To Be Sure..
Electrodes For PH Measurements: - When You Need To Be Sure..
measurements
Combined
Reference
Glass
Electrode working principle
Electrode body
Filling hole
Internal reference element
Reference element
Internal solution
Reference electrolyte
Glass membrane
Liquid junction
Sensitive parts
Electrode body
Filling hole
Internal and external
reference elements
Reference electrolyte
Internal solution
• Non-aqueous solutions or
with low ionic strength.
• Electrolyte includes
incompatible ions
How to select
a reference electrode
Interference from Double junction or
K+ or Cl- Hg/Hg2SO4 as reference element
Ag wire
Active part
External solution
(saturated KCl)
KCl crystals
Benefits of Red Rod
technology
Longer lifetime
6.8
pH 7.000 at 50°C
6.3
pH 4.005 at 25°C
5.8
pH
Traditional Technology
5.3
'Red Rod' Technology
4.8
4.3
3.8
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70
Time (mn)
Electrode maintenance
Electrode maintenance
Fast response
A longer lifetime
Electrode maintenance
• Crystals:
Present
Move freely
No air bubble
Every day
Check the state of the
electrode
Glass electrode
• Glass membrane:
Clean
No cracks or scratches
Every day
How to clean
Immerse electrode in RENOVO.X solution for
approx. 3 min. Rinse with deionised water.
Protein
Immerse electrode in KS400 solution (pepsin in
contamination HCl 0.1 M) for several hours and rinse with
deionised water.
Rinse with a solvent miscible with water (i.e.
Grease or oil acetone), then rinse with deionised water.
deposit Close filling hole and immerse in RENOVO.N for
12 to 16 hours.
Close the filling hole and immerse electrode in a
Silver or sulphide KS410 (Thiourea) solution for several hours.
contamination Renew the filling solution.
GK ANNEX
Drifting potential
Unstable readings
Low sensitivity/slow response
Remove bubbles:
Air bubbles • Tap the electrode with fingers
• Dip the tip in water at 60°C
Hydration:
•At least 1 hour in pH 4
Low sensitivity/slow response
Buffers:
Calibration error • Check date of validity
• Use fresh, quality buffers
Remove bubbles:
Air bubbles • Tap the electrode with finger
• Dip the tip in water at 60°C
Reproducible stirring
No stirring during calibration and
measurements
Unstable readings
Remove bubbles:
Air bubbles • Tap the electrode with finger
• Dip the tip in water at 60°C