Professional Documents
Culture Documents
Mpdct0248eab - AsteionSuper4 48kW TSX-021B5
Mpdct0248eab - AsteionSuper4 48kW TSX-021B5
Mpdct0248eab - AsteionSuper4 48kW TSX-021B5
Product Data
No. MPDCT0248EAB 48 kW
APPLICATION
The AsteionTM is a multislice Helical CT scanner that sup-
ports whole-body scanning.
The system generates 5.3 slices per second using a
Selectable Slice-thickness Multi-row Detector (SSMD).
Since it can also perform high speed reconstruction at a
maximum 4 images per second, every examination can be
processed at ultra-high speed. In combination with the
Continuous Imaging technology, it is possible to quickly
perform more accurate scanning and examinations.
FEATURES
• High-speed scanning
Four slices can be acquired simultaneously in a single
rotation. In the case of a run-off study from the shoulder
to the pelvis of an adult patient, 720 mm can be scanned
within 16.5 s with 5-mm data acquisition. • Tilt helical scanning
TCOT cone-beam reconstruction, which uses the
• Long scanning range Feldkamp method to precisely compensate for the angu-
The long scanning range of 1,800 mm and the high lation of the beam, enables helical scanning with the
scanning speed together facilitate whole-body trauma gantry tilted in the range from 30° forward to 30° back-
examinations without repositioning of the patient. ward. Angled helical acquisition is useful for many appli-
cations, including obtaining the correct scan plan for
• Guided mode routine brain scans. This ability makes it easy to avoid
When this function is selected, the operating procedures X-ray exposure to the patient's orbit.
are displayed on the monitor screen of the scan console.
Examinations can be performed by following the on- • High patient throughput
screen step-by-step instructions. This enables inexperi- The high cooling rate of the 4-MHU tube, which reduces
enced technologists to handle emergency cases. the cooling time required between scans; the high scan-
ning speed, which reduces the exposure time; and the
• Excellent image quality high reconstruction speed, which reduces the patient
The system achieves a low-contrast detectability of 2 mm scan cycle time, all increase the patient throughput sig-
at 0.3% and a high-contrast resolution of 0.35 mm in the nificantly.
x, y, and z directions. The multi-tasking capabilities of the system enable
Routine multislice helical CT examinations can be per- patient registration and protocol setting with reconstruc-
formed using slices as thin as 0.5 mm, enabling high- tion in the background and thus increase flexibility.
precision 3D and MPR images to be generated from the In addition, the system's ability to generate images of
fine isotropic voxel data. various slice thicknesses from a single data set permits
acquisition for routine examinations, biopsy procedures,
• Exposure dose reduction and 3D image generation to be performed in a single
The detector system has an excellent S/N ratio, which scan.
permits examinations with a reduced exposure dose. In For example, it is possible to reconstruct 5-mm images
helical scanning, the Real-EC function, which makes the for routine examinations, 1-mm images for detailed
amount of noise uniform for each slice, can be selected examinations, and 0.5-mm slice images for 3D image
in an eXam Plan, making the minimization of unneces- generation from a single data set acquired by helical
sary patient exposure easy. scanning with 0.5-mm slices.
48 kW
2
MPDCT0248EAB
3
48 kW
• The relationships between the slice thickness and the X-ray generation
imaging area scanned in 7.5 s (0.75 s, 10 rotations) in • X-ray exposure: Continuous
the longitudinal direction are shown for helical pitches • X-ray tube voltage: 80, 100, 120, and 135 kV
3.0, 4.5, and 6.0. • X-ray tube current: 10 mA to 400 mA (350 mA for
135 kV) (in increments of 10 mA)
Image slice Setting slice Helical pitch (Note)
• X-ray tube heat capacity:4.0 MHU
thickness thickness 3.0 4.5 6.0 • X-ray tube cooling rate: Max. 864 kHU/min
0.5 mm 0.5 mm 15.0 mm 22.5 mm 30.0 mm • Focal spot size
1 mm 1 mm 30.0 mm 45.0 mm 60.0 mm – IEC standard,
nominal : 0.9 mm × 0.7 mm (small)
2 mm 2 mm 60.0 mm 90.0 mm 120.0 mm 1.4 mm × 1.4 mm (large)
5 mm 5 mm 150.0 mm 225.0 mm 300.0 mm
X-ray detection
Note: The couch-top traveling distance per rotation is shown in
• Detection system: Solid-state detectors
proportion to the slice thickness.
• Main detector: 896 channels × 22 elements
• Data acquisition: 896 channels × 4 rows
Dynamic scan
• Reference detector: 1 set
• Scan time: 0.75, 1, 1.5 s/360°
• View rate: Max.1,200 views/s
• Programmable time: Max. 1 hour
This refers to the maximum time
Data processing
within which a series of scans
• Reconstruction matrix: 512 × 512 pixels
are performed following a pre-
• Picture element (pixel) size
determined eXam Plan.
– CT image: Unit: mm
• Number of
programmable scans: Max. 10 Scan field SS S M L LL
Maximum time of one continu- *to *to *to *to *to
Pixel size
ous scan is 100 s. 0.35 0.47 0.63 0.78 0.98
• Scan plan
– Scan interval: Min. interval: 1 s (if an interval is
– Scanogram: Unit: mm
set).
(Note that "None" (no interval) Enlargement
Standard 2× 4×
can be selected.) ratio (area)
Setting is possible in increments 1.00 (L) 0.71 0.50
Pixel size
of 0.1 s in a scan interval more 0.50 (S) – –
than 1 s.
*: Depending on the Vari-Area or Zoom factor
Note: When a scanning mode with patient couch movement is
used, the minimum scan interval is increased by the time
required for movement.
• Reconstruction filter functions
– Functions for the abdomen with BHC
• Scan start delay time: Min. 0.5 s – Functions for the abdomen without BHC
Setting is possible in increments – Functions for the brain with BHC
of 0.1 s. – Functions for the brain without BHC
• Scan rate: Max. 133 scans/100 s – Functions for the inner ear and bone
(0.75 s scan, 133 rotations) – Functions for the lung
• Image reconstruction – Functions for high-resolution test mode
– Number of images: Max. 4 images/scan – Functions for auditory ossicles and the spine/with high-
– Image interval: Reconstruction is possible in resolution processing
increments of 0.1s. – Functions for maintenance
• Reconstruction time: Min. 0.5 s • Image reconstruction time
• Real-time reconstruction – Image reconstruction
time: 12 images/s (0.083 s/image) time: Up to 4 images/s
– Scanoscopy: Real-time
– Real-time
reconstruction: 12 images/s (1 slice)
24 images/s (3 slices)
Note: Real-time reconstruction of 3 slices is only for CT fluo-
roscopy (option). For helical scan, dynamic scan, and
SURE
Start, real-time reconstruction is for 1 slice.
4
MPDCT0248EAB
5
48 kW
6
MPDCT0248EAB
Gantry Console
The scanner is composed of the gantry and the patient The console is provided with a hybrid keyboard, monitor,
couch. The scanner uses a fan-shaped continuous X-ray and mouse.
beam to scan the region to be examined. Transmitted X- • Functions of the console
rays are detected and converted into electrical signals by – Selection of scan parameters
the SSMD. – Scanoscope control
The gantry includes the main body and its support mecha- – Scan control
nism. The X-ray tube and the SSMD are mounted facing – Remote control of couch-top movement
each other on either side of the gantry aperture, and the X- – Remote control of gantry tilt
ray tube and detectors rotate continuously around the – Window level and window width adjustment
aperture of the gantry. A slipring is employed to transmit – Other mouse-operated image processing functions
power between the gantry and the rotating X-ray high-volt-
age generator assembly. Magneto-optical disk unit
The gantry can be tilted forward and backward in order to The magneto-optical disk unit permits repeated recording
perform tilt scanning. Three-dimensional alignment lights and retrieval of data, like a floppy disk drive. It is a com-
are provided for setting slice positions. Gantry and patient pact, externally mounted, large-capacity digital storage
couch operating controls are provided on both sides of the device that can be used to store a large amount of data.
front of the gantry housing. The patient guide display indi- Automatic image archiving to the magneto-optical disk unit
cates the scan status to the operator and the patient. The can be set in eXam Plans.
X-ray high-voltage generator is built into the gantry, and
the system employs a high-frequency inverter for generat-
ing and stabilizing the high voltage supplied to the X-ray
tube. The generator includes electronic circuits for control-
ling the speed of the rotating anode in the X-ray tube. Use
of a high-frequency inverter system results in high power
output combined with excellent stability. In addition, the
system is compact and light weight.
7
48 kW
8
MPDCT0248EAB
9
48 kW
(1,170*)
1,470
In case of In case of
3-phase, 200 V 3-phase, 400 V
Ground resistance:
As per applicable
1
5,530 (4,930*)
legal requirements.
NFB NFB NFB
150 A 50 A 100 A
7,030 (6,430*)
Ground bar 2
Gantry Gantry control/ System
Console transformer
(option)
Ambient conditions
Temperature(*1) Humidity Heat generation 3
1,500
Scan room
Gantry 20°C to 26°C 40% to 80% Approx.
Unit: mm
3,020
Tolerance: ±2°C No 9,720 kJ/h (*2)
condensation 32,070 kJ/h (*3) 1 Gantry 3 Console
2 Patient couch
Patient couch 20°C to 26°C 40% to 80% Approx. *: For the short patient couch version
10
MPDCT0248EAB
(REC)
6
Speaker 17
(scan room)
Console
Gantry Navibox
17
(CPU)
Patient
12 11
couch 18
Speaker
Breaker control
box room side
Unit: m
• With system transformer
Console
Navibox
(REC)
6
Speaker 17
(scan room)
Console
Gantry Navibox
17
(CPU)
System 11
Patient 12 18
couch transformer
12
5 18 Speaker
Breaker control
box room side
Unit: m
11
48 kW
OUTLINE DRAWINGS
12
MPDCT0248EAB
OUTLINE DRAWINGS
890 (35.0)
360 530
(14) (21)
30°
1,970 30° (TILT ANGLE)
(77.6)
(74.8)
1,900
(69.3)
1,760
(35.8)
910
320 530
(12.6) (21.0)
Gantry
Unit: mm (in)
13
48 kW
OUTLINE DRAWINGS
(18.5)
(24.8)
470
630
2,190 2,187
(86.2) (86.1)
Unit: mm (in)
14
MPDCT0248EAB
OUTLINE DRAWINGS
(18.5)
(24.8)
470
630
1,890 1,887
(74.4) (74.3)
Unit: mm (in)
15
48 kW MPDCT0248EAB
OUTLINE DRAWINGS
(32.1)
(32.1)
815
815
(29.5)
750
1,600
(63.0)
2,050
(80.7) 100 (3.9)
(17.9)
455
(49.4)
1,255
(31.5)
800
(27.6)
(27.6)
700
700
450 450
(17.7) (17.7)
Console
http://www.toshibamedicalsystems.com
©Toshiba Medical Systems Corporation 2004-2006 all rights reserved.
Design and specifications subject to change without notice.
"Made for Life" is a trademark of Toshiba Medical Systems Corporation.
Asteion, SUREFluoro, and SUREStart are trademarks of Toshiba Medical Systems Corporation. Toshiba Medical Systems Corporation meets Toshiba Medical Systems Corporation Nasu Works meets
internationally recognized standards for Quality the Environmental Management System standard, ISO 14001.
This document may include trademarks and registered trademarks of other companies. Management System ISO 9001, ISO 13485.
Model TSX-021B 2006-04 TME/KI Printed in Japan