Prevent Electrical Shock Hazard Switch off the Dialog/Dialog Advanced and disconnect unit
from mains before opening the machine!
Do not touch any exposed wiring or conductive surfaces
while the Dialog/Dialog Advanced is open. The voltages
present when electrical power is connected to the
Dialog/Dialog Advanced can cause serious injury or death!
If necessary open the machine for maintenance and repair
Fuses If fuses are replaced they must exactly match the type and
rating specified by the manufacturer. Where applicable: Fuses
must be approved by UL/CSA!
MT-MD-DE08C M.KAY
DialogAd_sm_Chapter 5_4-2002.doc/pdf <280602> ddmmyy B. Braun Medizintechnologie GmbH
Dialog/Dialog Advanced
5. Technical Safety Inspection and Preventive Maintenance 4/2002 5-2
Seite
5. Technical Safety Inspection and 5-3
Preventive Maintenance
5.1 Inspection Protocol for Automatic Blood Pressure 5-7
Measurement ABPM
MT-MD-DE08C M.KAY
DialogAd_sm_Chapter 5_4-2002.doc/pdf <280602> ddmmyy B. Braun Medizintechnologie GmbH
Dialog/Dialog Advanced
5. Technical Safety Inspection and Preventive Maintenance 4/2002 5-3
The technical safety inspection, according to the specified check list, must be performed and
documented every 12 months, with reference to the service manual and operating
instructions, which are subject to technical amendments.
The preventive maintenance with the exchange of wear and tear parts, is recommended
every 12 months and should be documented, according to the specified check list, with
reference to the service manual and operating instructions, which are subject to technical
amendments.
Art. No.: ....................................... Machine No.: .............................................
Year of Purchase: ........................ User: ............................................................
.................................................................................................................................
Operating Hours: ..................... h Inventory No.: ............................................
SW Version: ................................. HW Version: ...............................................
Manufacturer: B. Braun Medizintechnologie GmbH
2.
Visual Inspection with Maintenance Procedures
S 2.1
Machine: clean/complete; no damages/moisture influences; unit rollers are moveable; machine record
book present; no special incidents; type plates and inscriptions legible, options present
M 2.1.1 Clean interior space and exterior surfaces
S 2.2 Check mains supply (power supply line and connectors)
M 2.3 Suction tubing for disinfection and concentrate exchanged b
M 2.4 O-rings in suction rod exchanged b
M 2.5 Tight seat of the tubing connectors, clamps and couplings
M 2.6 Dialyser coupling membranes, o-rings and tubings exchanged b
M 2.7 Tight seat of tubings, clamps and couplings of inlet and outlet tubings
M 2.8 Tight seat of boards and connectors
M 2.9 Fan in power supply clean (if necessary clean)
2.9.1 Exchange dust filter of fan (rear door) b
M 2.10 PC motherboard:
2.10.1 CPU fan exchanged (depends on hardware status)
(CPU fan not included in the maintenance kit, if necessary order separately.)
2.10.2 Battery or rechargeable battery on PC motherboard exchanged (after operating period of 4 years)
(Battery or rechargeable battery are not included in the maintenance kit, if necessary order separately.)
M 2.11 Monitor: no restriction of motion of rotation (and fan); tight seat touch membrane/keyboard
membrane
S 2.11.1 Function of the keys, display illumination and touch
2.11.2 Image display, geometry
2.11.3 Alarm activation
M 2.12 No cracks in covers (rear door cover, front cover and mains suppression cover)
M 2.13 Tight seat of ground terminals in frame, front door, rear door and sub-racks (if necessary order separately for sub-racks)
M 2.14 Check/set pressure reducer (DMV) to 1.3 bar
2.14.1 Check pressure non-return valves (RVDA, DDE, RVFPE, RVFPA)
M 2.15 Conductivity Sensors
2.15.1 ENDLF sensor: Visual inspection
2.15.2 BICLF sensor: Visual inspection
MT-MD-DE08C M.KAY
DialogAd_sm_Chapter 5_4-2002.doc/pdf <280602> ddmmyy B. Braun Medizintechnologie GmbH
Dialog/Dialog Advanced
5. Technical Safety Inspection and Preventive Maintenance 4/2002 5-4
2.21.1 PE (permissable tolerance ±50 [mmHg]) Comparison measurement at: - 400 = .................. [mmHg]
2.21.2 PDA (permissable tolerance ±50 [mmHg]) Comparison measurement at: - 400 = .................. [mmHg]
0 = .................. [mmHg]
M 2.22 Heparin Pump
2.22.1 Heparin Pump Comfort: Syringe recognition passed
(TSM, LLC Manual Test, 1.4 Heparin Pump, with 20 ml and 50 ml template)
2.22.2 Heparin Pump Compact: Check display position (80 mm ± 15 mm) ....................... [mm]
(Drive fast forward to end position with 30 ml syringe. Check display position in mm.)
M 2.23 Filter (FVD, FB, FK, FBK1/2, FBIC) exchanged b
M 2.24 O-rings in BIC cartridge holder (option) exchanged b
M 2.25 Summer/Winter Time Setting
(TSM, Treatment Support, System Configuration)
M 2.26 HDF Online: Filter (HFB) exchanged
(Filter not included in maintenance kit, if necessary order separately.)
S 2.27 SAKV Curentless Closed (Option)
2.27.1 O-ring for magnetic plate exchanged (O-ring not included in maintenance kit, if necessary order separately.)
2.27.2 Gap for magnetic plate 1.6 (± 0.05 mm)
MT-MD-DE08C M.KAY
DialogAd_sm_Chapter 5_4-2002.doc/pdf <280602> ddmmyy B. Braun Medizintechnologie GmbH
Dialog/Dialog Advanced
5. Technical Safety Inspection and Preventive Maintenance 4/2002 5-5
S
3.6 Inlet Pressure PBE - Comparison measurement at: 0 = .................. [mmHg]
(permissable tolerance ±12 [mmHg]) + 400 = .................. [mmHg]
S
3.7 Inlet Pressure PBS (Option): - Comparison measurement at: 0 = .................. [mmHg]
(permissable tolerance ±12 [mmHg]) + 400 = .................. [mmHg]
S
3.8 Venous Pressure PV: - Comparison measurement at: 0 = .................. [mmHg]
(permissable tolerance ±12 [mmHg]) + 400 = .................. [mmHg]
S 4.3 Earth leakage current ≤ 0.5 [mA] - During heat-up phase ....................... [mA]
S 4.4 Patient leakage current < 0.1 [mA] - Under normal conditions ....................... [mA]
(Conductivity at 14 mS/cm)
MT-MD-DE08C M.KAY
DialogAd_sm_Chapter 5_4-2002.doc/pdf <280602> ddmmyy B. Braun Medizintechnologie GmbH
Dialog/Dialog Advanced
5. Technical Safety Inspection and Preventive Maintenance 4/2002 5-6
User:
......................................................................
Date / Signature
MT-MD-DE08C M.KAY
DialogAd_sm_Chapter 5_4-2002.doc/pdf <280602> ddmmyy B. Braun Medizintechnologie GmbH
Dialog/Dialog Advanced
5. Technical Safety Inspection and Preventive Maintenance 4/2002 5-7
Measurement Inspection The measurement inspection is recommended every 12 months and should be
documented. The measuring methods are described in the execution
procedures.
Dialysis Machine Dialog/Dialog Advanced Serial No.: ............................................ Art. No.: ............................................
Environmental Conditions The environmental conditions must be from 15 °C to 25 °C, the relative
humidity must be between 20 % and 85 % (normal room condition).
2. Limits of Error of the Pressure Admit a static pressure to the unit. The permissable tolerance is ± 5 mmHg for
each measurement.
Indication
Pressure Measurement Value Deviation
[mmHg] Dialog Reference [mmHg]
300
250
200
150
100
50
0
4. Rapid Exhaust Time for pressure reduction from >260 mmHg to <15 mmHg max. 10 s ........ s
5. Function Inspection The results of a measurement on a test person are plausible ----------- OK
6. Result of Inspection The blood pressure measurement module has passed the measurement
inspection
no yes
Comments: .............................................................................................................
.................................................................................................................................
MT-MD-DE08C M.KAY
DialogAd_sm_Chapter 5_4-2002.doc/pdf <280602> ddmmyy B. Braun Medizintechnologie GmbH
Dialog/Dialog Advanced
5. Technical Safety Inspection and Preventive Maintenance 4/2002 5-8
To Point 2: Limits of Error of the Pressure Indication Measurement according to DIN EN 1060-1:1995 and
DIN EN 1060-3:1997.
Reference Instrument
Referenzmeßgerät 1. Connect all components acc. to figure (stopcock
system acc. fig.).
Measurement Chamber
2. Switch on Dialog in TSM service program.
Meßkammer 3. Select in TSM Manual Test and Calibration, TTL
Dialog ABPM Manual Test and Calibration and then ABPM
500 ml
Maintenance.
4. Press P Test button.
Wait until a pressure build-up of approx. 170 mmHg
(start of pressure retention phase).
Pressure Tube 5. Open/close valve 4 and set a test pressure of 300
Druckschlauch
mmHg with the syringe (in ABPM Maintenance
1 2 3 4 menu) respectively. Enter displayed value from
50 ml
MT-MD-DE08C M.KAY
DialogAd_sm_Chapter 5_4-2002.doc/pdf <280602> ddmmyy B. Braun Medizintechnologie GmbH
Dialog/Dialog Advanced
5. Technical Safety Inspection and Preventive Maintenance 4/2002 5-9
MT-MD-DE08C M.KAY