CHAUDHRY MUHAMMAD AKRAM TEACHING &

RESEARCH HOSPITAL
Department Name: Molecular Pathology Document Number: CMA/MP-01/005
Title of the Policy & Procedure: Version Number: MP-001-A
Quality Work Instructions for COVID-19 Amplification Effective Date: 22-FEB-2021

Quality Work Instructions for COVID-19 SYSTAAQ Amplification

Note: Thaw all reagents at room temperature. Vortex and short spin it before use.

For each RNA sample prepare a reaction mix according to the table below: Include sufficient

reactions for positive and negative controls.

Reagent Required for each sample

RT Mix 13 µl

Internal control 0.5 µl

Final Volume 13.5 µl

Pipette 13.5μl of these mixes into each well according to your qPCR plate set up.

Pipette 7 μl of RNA template into each well, according to your plate setup.

For negative control wells use 7 μl of RNase/DNase free water. The final volume in each

well is 20.5 μl.

Seal the tubes.

Now sample is ready, Place in SYSTAAQ AB QuantGene and run the program.