Cross Calibration of Ionisation Chamber Iaea Dosimetry Protocol Trs 398 Aim
Cross Calibration of Ionisation Chamber Iaea Dosimetry Protocol Trs 398 Aim
Cross Calibration of Ionisation Chamber Iaea Dosimetry Protocol Trs 398 Aim
Result:
In this study, the dose measurement of electron beams using a parallel plate chamber. By
cross-calibration based on the cylindrical chamber .The maximum dose calculated by
applying the calculated correction factor. There was a difference of about less than
±1.0%. Dose measurement based on absorbed dose standards for water in IAEA TRS-
398.
Conclusion:
we recommend cross calibrating the Parallel plate chamber with user beam quality as
recommended by the TRS 398 Protocol since the chambers calibrated at laboratories have
Co-60 as their reference beam for which the N D,W will not account for perturbation of
wall. By cross calibrating the results are better improved and the same can be cross
verified with the results using Cylindrical Chambers.