Professional Documents
Culture Documents
Manual Roller 20 PN
Manual Roller 20 PN
TEST1(*)
ROLLER 20 LC(*)
ROLLER 10 PN
ROLLER 20 PN
ROLLER 20 MC
INTERFACING MANUAL
(*)
Updated for Fw 8.00x
The disclosure of this document or parts of it, including but not limited to copying, reproduction and transmission in any form or by any
means, such as electronic, mechanical, photocopied, recorded, or otherwise, direct or indirectly, is prohibited without the prior written
authorization of Alifax S.r.l.
REVISIONS
ROLLER10/20PN INTEGRATION
01 09/07/2008 601F 6.01C 1.00P INTERFACING LATEX
PROTOCOL
Production Site:
Via Merano 30 33045 Nimis (UD) Italy
Tel +39 0432 547454
Fax +39 0432 547378
INDEX
1. INTRODUCTION------------------------------------------------------------------------------------------------------------------------------ 4
2. CABLES ---------------------------------------------------------------------------------------------------------------------------------------- 4
3. USEFUL TOOLS FOR DEBUGGING --------------------------------------------------------------------------------------------------- 4
4. SYMBOLS -------------------------------------------------------------------------------------------------------------------------------------- 4
5. TEST1 RS232 PORTS features: -------------------------------------------------------------------------------------------------------- 5
6. SERIAL CABLE PIN CONNECTIONS TO CONNECT TEST1 TO PC-LAB: ------------------------------------------------- 5
7. TEST1 protocol options up to Sw 6.51x --------------------------------------------------------------------------------------------- 6
8. RS232 test function to check the communication of TEST1 valid for Sw up to 6.51x ------------------------------- 8
9. TEST1 protocol options from Sw 8.00x on ward ------------------------------------------------------------------------------- 10
10. RS232 test function to check the communication of TEST1 from Sw 8.00x ------------------------------------------ 11
11. QUERY PROTOCOL for TEST1 up to Sw 6.51x --------------------------------------------------------------------------------- 12
12. QUERY PROTOCOL for TEST1 from Sw 8.00x ---------------------------------------------------------------------------------- 13
13. Query Protocol description valid for all TEST1 Sw: --------------------------------------------------------------------------- 14
14. ROLLER20 LC RS232 PORTS features: ------------------------------------------------------------------------------------------- 17
15. SERIAL CABLE PIN CONNECTIONS TO CONNECT ROLLER20 LC TO PC-LAB:------------------------------------ 17
16. ROLLER 20LC protocol options up to Sw 6.51x -------------------------------------------------------------------------------- 18
17. RS232 test function to check the communication of ROLLER20 LC ---------------------------------------------------- 18
18. QUERY PROTOCOL ROLLER20LC. ------------------------------------------------------------------------------------------------- 19
19. Protocol description: -------------------------------------------------------------------------------------------------------------------- 23
20. ROLLER 10 - 10PN - 20PN and R20MC RS232 PORTS features: --------------------------------------------------------- 25
21. SERIAL CABLE PIN CONNECTIONS TO CONNECT ROLLER INSTRUMENTS TO PC-LAB: --------------------- 25
22. ROLLER INSTRUMENTS PROTOCOL OPTIONS:------------------------------------------------------------------------------- 26
23. RS232 test to check the communication of ROLLER INSTRUMENTS -------------------------------------------------- 27
24. QUERY PROTOCOL for ROLLER INSTRUMENTS------------------------------------------------------------------------------ 28
25. Protocol explanation (BCI Configuration): --------------------------------------------------------------------------------------- 28
26. TEST1, ROLLER20LC and ROLLER INSTRUMENTS Multiple connections ----------------------------------------- 31
27. INTERFACING LATEX PROTOCOL -------------------------------------------------------------------------------------------------- 32
Note: The paragraphs written in blue characters (as on this note) indicate a variation or an addition to the
previous version of the manual.
We reserve the right to make changes in the course of technical development without previous notice.
1. INTRODUCTION
This manual is focused to give detailed information about serial ports connections available on TEST1
ROLLER20LC, ROLLER10/10PN/20PN and ROLLER 20-MC instruments and communication protocol
features.
It should help Software House Companies during the modification or the installation of their program in the
Host Computer.
Please notice for easiness of explanations, we assume Roller10, Roller10PN, Roller20PN and Roller20MC from this
moment are referred as “Roller Instruments”
2. CABLES
A standard cable to connect two Personal Computers via serial port.
This cable must have:
D-type 25 pin female connector at one side and a D-type 9 pin female connector at the
opposite side for TEST1 and ROLLER20 (OLD MODEL).
D-type 9 pin female connector at one side and a D-type 9 pin female connector at the
opposite side for ROLLER10PN/20PN and R20-MC
4. SYMBOLS
The below symbols will be displayed on HyperTerminal if the program is configured with Terminal character.
☻= STX symbol
♥ = ETX symbol
♠ = ACK symbol
§ = NAK symbol
CHECKSUM = XOR calculation of part of R received message to check the correctness of it.
HyperTerminal program must have the same parameters to a correct data exchanging.
PORT2 connection.
PORT1 connection
DAT8 either DAT15 option can be choice accessing in the COMMUNICATION MENU passing through
TECH MENU. Press key 5 from MAIN MENU and keys 1 and subsequently 9 as password and then key 3.
If the internal bar code reader is installed in a TEST1 instrument, DAT15 will be forced automatically at powering
on.
In the TEST1 COMMUNICATION MENU, RS232 Protocol option can be recalled pressing key
8. RS232 test function to check the communication of TEST1 valid for Sw up to 6.51x
It simulates one transmission of 10 TEST1 results to host computer through RS232 serial port (SERIAL PORT 2).
If DAT8 is chosen, a message with 8 ID digits will be sent to the Host.
Selecting DAT15, the ID digit number to send to the Host is going to be 15.
ETX ♥ 03 1 char The tenth field is the same to all ten strings. same procedure
calculating checksum of the
when
Note : what above explained refers to the TEST strings received from the
Strings instrument send to host computer, this analyser.
string is displayed above this table.
To be more clear the previously called “DAT 15” of Sw 6.51 which was acting in bidirectional mode, now have
been renamed as BCI.
Independently TEST1 is set on DAT 15 or BCI the protocol description for the result string ( R ) sent to HOST is
exactly the same.
Please notice that being TEST1 equipped with an Internal Bar Code Reader, the BCI protocol is automatically
forced, independently DAT15 is set.
If the customer is not interested in working with Query Host, leave instrument set on BCI mode and configure
ACK option set to OFF.
DAT15 either BCI option can be choice accessing in the COMMUNICATION MENU passing through
TECH MENU. Press key 5 from MAIN MENU and keys 1 and subsequently 9 as password and then key 3.
If the internal bar code reader is installed in a TEST1 instrument, DAT15 will be forced automatically at powering
on.
In the TEST1 COMMUNICATION MENU, RS232 Protocol option can be recalled pressing key
Option (1) Enables the patient identification number length to 15 bytes unidirectional.
Option (2) Enables the patient identification number length to15 bytes but it enables the Query Host Mode
Please notice that being TEST1 equipped with an Internal Bar Code Reader, the BCI protocol is automatically
forced, independently DAT15 is set.
10. RS232 test function to check the communication of TEST1 from Sw 8.00x
It simulates one transmission of 10 TEST1 results to host computer through RS232 serial port (SERIAL PORT 2).
DAT15 and BCI test transmission and its explanation:
COMMUNICATION MENU in the TECH MENU, allows the configuration of some parameters for the correct communication
protocol. The accessing in it is allowed by typing key 5 from MAIN MENU, keys 1 and subsequently 9 for the password
requesting and key 3.
RS232 PROTOCOL= It has to be set to DAT15 (Working with an Internal Bar Code Reader it is forced automatically).
To access inside and select DAT15 press key 1 in the COMMUNICATION menu.
SETUP [ * ]= Press key 4 in the COMMUNICATION menu to access inside the option and 1 to enable the
asterisk character transmission to the host. It works when a analysed sample gives a low value of
hematocrit .
ENABLE BAYER Press key 5 in the COMMUNICATION menu to access Bayer option and 1 to enable it to
guarantee the Bayer protocol specifications (field Cycle/Bayer = 01).
If Bayer protocol is not requested this option must be kept to 0 (disable).
MEASURE MISSING ID= Press key 6 in the COMMUNICATION menu to access to the option and key 1 to allow the
analysis of samples with an auto generated ID in case IBCR is not able to read the barcode label.
Keeping it to 0, these samples won‟t be analysed.
TIME OUT UART= Press key 7 in the COMMUNICATION menu to access this option to set the waiting time (from 2 to
99 seconds) if the increasing of the waiting time to receive “T” message and the confirmation byte
(ACK) for having received “R” message is requested. The maximum time to receiving “T” message
and the confirmation byte (ACK) for having received “R” message remain set to 2 seconds keeping
it to OFF. If within this time no message comes in, the instrument will go ahead. This option works
if ACK option, COMMUNICATION menu is enabled.
TEST1 NUMBER= A number between 01 to 99 on each instrument can be assigned by this option if multiple
instruments connections to Host computer is required
Notice: following parameters MUST BE ALWAYS DISABLED: Change ID, Send ESR Limits, Send 6 Par, Send Curve
Parameters and VS
Messages.
„Q‟ message from TEST1 „T‟ message from HOST „R‟ message from TEST1
Lengt
Field Hex Lengt Field Hex Lengt Hex
Field name h
name value h name value h value
31
Total length
char
Note 1: Checksum value is the result of Exclusive OR (XOR) of all characters of the message, excluded STX, ETX and
Checksum. If the Checksum value is 03 (corresponding to ETX), 02 (corresponding to STX), 00 (corresponding to
NULL), the value will be replaced with 7F hex value. The software house should take into account these references
and modify its program to adapt such requisition. From software version 6.51E onwards, in the transmission protocol,
hexadecimal characters like 0x13 (DC3 character), 0x14 (DC4 character), and 0x17 (ETB character), are not
transmitted unless used as checksum
Note 2: From the 4.02E to 4.03J software versions, this field take care of the Cycle number. From the 4.03K onwards, this field
could be relating to Bayer parameter setting it by software. If Bayer is 0, the field deals with the number of Cycle
otherwise if Bayer is 1, the field should contain '01' (hex 30 and 31) to maintain the Bayer protocol compatibility.
Note 3: <STX> character ( ☻symbol looking at the hyper terminal program) in the serial transmission is represented by
0x02 hexadecimal bytes and <ETX> ( ♥ symbol looking at the hyper terminal program) by 0x03.
COMMUNICATION MENU in the TECH MENU, allows the configuration of some parameters for the correct communication
protocol. The accessing in it is allowed by typing key 5 from MAIN MENU, keys 1 and subsequently 9 for the password
requesting and key 3.
RS232 PROTOCOL= It has to be set to BCI (Working with an Internal Bar Code Reader it is forced automatically).
To access inside and select BCI press key 1 in the COMMUNICATION menu.
SETUP [ * ]= Press key 4 in the COMMUNICATION menu to access inside the option and 1 to enable the
asterisk character transmission to the host. It works when a analysed sample gives a low value of
hematocrit .
ENABLE BAYER Press key 5 in the COMMUNICATION menu to access Bayer option and 1 to enable it to
guarantee the Bayer protocol specifications (field Cycle/Bayer = 01).
If Bayer protocol is not requested this option must be kept to 0 (disable).
MEASURE MISSING ID= Press key 6 in the COMMUNICATION menu to access to the option and key 1 to allow the
analysis of samples with an auto generated ID in case IBCR is not able to read the barcode label.
Keeping it to 0, these samples won‟t be analysed.
TIME OUT UART= Press key 7 in the COMMUNICATION menu to access this option to set the waiting time (from 2 to
99 seconds) if the increasing of the waiting time to receive “T” message and the confirmation byte
(ACK) for having received “R” message is requested. The maximum time to receiving “T” message
and the confirmation byte (ACK) for having received “R” message remain set to 2 seconds keeping
it to OFF. If within this time no message comes in, the instrument will go ahead. This option works
if ACK option, COMMUNICATION menu is enabled.
TEST1 NUMBER= A number between 01 to 99 on each instrument can be assigned by this option if multiple
instruments connections to Host computer is required
Notice: following parameters MUST BE ALWAYS DISABLED: Change ID, Send ESR Limits, Send 6 Par, Send Curve
Parameters and VS
2. HOST computer verifies if the patient identity code is present in the working list and, accordingly, it replies with
"T" message. TEST1 waits to receiving it within two seconds and then it goes ahead.
If ACK option is off, TEST1 is going to remain waiting up to 2 seconds for receiving it, otherwise if it is on,
TEST1 is going to wait for all the set time (TIMEOUT from 2 to 99 seconds).
TIMEOUT UART time and ACK are selectable in the COMMUNICATION MENU.
“T” message is going to have the result field filled with “0001”, if
<STX>T01123456789 0001e<ETX>
the sample requires the analysis. The result field will be filled with
“0000” if the sample doesn‟t require the analysis.
<STX>T01123456789 0000d<ETX>
4. Host computer should send ACK confirmation byte (♠ symbol displayed on the HyperTerminal program) to
TEST1 for having accepted “R” message. TEST1 will ignore it, if ACK is off. Otherwise it will wait for all the set
time (TIMEOUT from 2 to 99 seconds) if ACK is On.
ACK = Hexadecimal code 06
Warnings:
1 If a patient identity (ID) read by the TEST1 Scanner doesn‟t find its own correspondent code in the Host
computer working list, “T” message will be replaced by <NAK> message (NAK Message) Hex code 21
(§ symbol displayed on the HyperTerminal program). The sample will be
however analysed and “R” message sent to host.
2 If a particular sample has a very low hematocrit (around 20% and less) and * option in the communication
menu is enabled, the result field of “R” message, sent
to the host, will contains an asterisk set in the first <STX>R01123456789 010305*015[ <ETX>
position on the left and its own ESR result
(i.e.
If the analysis of a particular sample generates an error and * option in the communication menu is enabled,
ESR result field will contain 0 in the first left position,
“-“ (minus) in the second and the error code (i.e. <STX>R01123456789 0103050-04\ <ETX>
3 If the internal Scanner cannot read the patient ID and “MISS ID” option in the Communication setup is 0,
“Q” message won‟t be created. The sample without its own patient identity won‟t be processed.
4 If the internal Scanner cannot read the patient ID and “MISS ID” option in the Communication setup is 1,
“Q” message won‟t be created. The sample without its own patient identity will be, however, processed and
“R” message sent to host with an identity code generated by the instrument itself.
The new patient ID, in this case, will be composed by 10 characters: cycle number (from 01 to 99)
(2 characters), instrument s/n (4 characters), rack number (from 01 to 04- 2 characters) sample position in
the rack (from 01 to 15- 2 characters).
Example:
0307920106
HyperTerminal program must have the same parameters to a correct data exchanging.
PORT2 connection.
PORT1 connection.
STD, DAT8 DAT15 or BCI protocol can be selected by software accessing in the SETUP menu.
In the SETUP MENU list, RS232 Protocol can be recalled by pressing or until the following
message is displayed:
SETUP
RS232 PROTOCOL Confirm the function by pressing
1 - STD 2 - DAT08
3 - DAT15 4 – BCI
The DAT8 and DAT 15 test stings sent out from R20LC are the same sent out from TEST1, please refer to chapter
8
In the TECH MENU, RS232 function can be recalled pressing or for many times until the
following message is displayed.
COMMUNICATION
RS232 TEST TX Press to run the test
RS232 PROTOCOL= It must be set to BCI (And an External Bar Code Reader must be installed).
SETUP
RS232 PROTOCOL
Confirm the function by pressing
KINET. SETUP= This function is available in the SETUP MENU. It can be recalled by pressing or
SETUP
KINET. SETUP Confirm the function by pressing
KINET. SETUP= 0
1 -Y 0-N _
It must be set to 0
SETUP [ * ]= This function is available in the TECH MENU. It can be recalled by pressing or
1- [ * ] [Y]
0- [ * ] [N]
The asterisk function activated allows the instrument to send the asterisk parameter to the
Host computer through the serial port. It will be displayed near the ESR value if the
sample analyzed has a haematocrit value low (< 20%).
IDENTITY ESR
Example of ESR result (045) with the * parameter enabled
and sent to the host computer with the patient identification
2367483493 *045 together.
IDENTITY ESR
Example of a sample analysed with a haematocrit value over
20% and * parameter enabled. The ESR result (5) will be
2367483493 0-05 sent to the host computer using this form.
Note:
Having disabled the asterisk function, the instrument cannot send the asterisk character to the host if the analysed
sample had a low haematocrit value, but it will be printed out next the ESR result however.
ENABLE BAYER Bayer communication protocol detects the Bayer field value (the previous field of the
ESR result) in the “R” message. If Bayer function in the COMMUNICATION menu is
enabled, the field will contain '01' (hex 30 and 31) to maintain the Bayer protocol
compatibility otherwise if it is disabled, the field will report the number of Cycle.
1- ENABLE [Y]
0- DISABLE [N]
Query setup= This function is available in the TECH MENU. It can be recalled by pressing or
1- Query [Y]
0- Query [N]
If Query is enabled (default), the instrument will be able to interrogate the Host computer
for the analysis requesting of each specific sample.
ACK= This function is available in the TECH MENU. It can be recalled by pressing or
TEC.CAL MENU 18
ACK SETUP Confirm the function by pressing
1 ACK [Y]
0 ACK [N]
This function allows to extend the waiting time for receiving “T” message (Query answer from the HOST computer)
and ACK (acknowledge message confirmation from the HOST computer for having received “R” message). To
modify the extension time (in seconds), it is available the TIMEOUT function explained in the Protocol description.
Notice: following parameters MUST BE ALWAYS DISABLED: Send ESR Limits, Send 6 Par, Send Curve Parameters
and VS
Parameters verification:
In the SETUP menu you will find FLAG LIST function. Press or until the following message is
displayed:
SETUP
FLAGS LIST
Confirm the function by pressing
Below is an example
ROLLER 20
UK ver. 6.xx
ROLLER20 SN. xxx
# AVAILABILITY xxxx
# GLOBAL COUNT. xxxx
GAIN 1.0000
MIXING CYCLES 180
MIXING SPEED 1
RS232 BCI
FLOW CONT. ON
OFFSET LEV. FLOW 800
OT OFFSET LEV. 700
T. 100 974
ROLLER n. 1
ACK setup 0
Query setup 1
Print setup 1
Calib. setup 1
Bayer. setup 1
[ * ] setup 0
Kinet.setup 0
Offset nee. 0
Vs 0
TimeOut 20
Warning SMART 1000
2. HOST computer will verify if the patient identity has been inserted in the working list and then it replies with
"T" message.
If ACK option (in the TECH MENU list) is 0, ROLLER 20 remains waiting up to 2 seconds for receiving it.
otherwise if ACK option is 1 it will wait for all the set time (TIMEOUT from 2 to 99 seconds).
TIMEOUT UART time option is located in the TECH MENU list. (Note 2)
th
The procedure above explained will be repeated for the first to the 18 sample tubes before starting
the analysis.
♠
4. Host computer should send an ACK confirmation byte ( symbol on hyper ACK Hex 06
terminal program) to ROLLER 20 for having accepted “R” message. The
instrument will ignore it, if ACK option is 0.
It will wait for all the set time (TIMEOUT from 2 to 99 seconds) if ACK option is 1.
Warnings:
1. If the patient identity (ID) read by the Scanner doesn‟t find its own correspondent in the Host computer
working list, “T” message will be replaced by <NAK> message Hexadecimal code 21
(§ symbol on hyper terminal program), the wheel will however rotate to the next
position to allow the operator to insert the tube. The sample will be analysed and “R” message sent to host.
2. If a particular sample has a very low haematocrit (less then 20%) and * option in the TECH MENU is
enabled, the result field of “R” message sent to the host will be composed by an asterisk in the first position
on the left and the rest its own ESR result (i.e. “ *015 ”).
If the analysis of a particular sample generates an error, ESR result field will have 0 in the first left position
“-“ (minus) for the second position and the rest of the field the error code (i.e. “ 0-04 “).
3. If no label has been stuck on the tube or the external Scanner cannot read the patient ID, “Q”
message won‟t be created. The wheel will rotate to accept the sample by pressing ENTER key
The sample will be however processed and “R” message sent to host with an identity code
generated by itself.
The new patient ID in this case will be composed by 10 characters:
cycle number (2 characters from 01 to 99), instrument s/n (4 characters), wheel number ( 2 characters)
sample position in the wheel (from 01 to 18- 2 characters).
Example:
Instrument s/n Wheel number
0307920106
Note 1: If Query Setup option is 0, no Query protocol will be created; the instrument is going to accept and analyse all
samples.
Note 2: Follow the instruction looking at the Chapter 13-8 of the technical manual for TIMEOUT UART activation.
Note 3: This field is related to Bayer parameter on SETUP. If Bayer is 0, the field contains the number of Cycle. If Bayer is 1,
the field contains '01' (hex 30 and 31) to keep compatibility with Bayer protocol.
Note 4: Checksum value is the result of Exclusive OR (XOR) of all characters of the message, excluded STX, ETX and
Checksum. If the Checksum value calculated by the analyser is 03 (corresponding to ETX), or 02 (corresponding to
STX), or 00 (corresponding to NULL), the value will be automatically replaced with 7F hex value.
Software House must adopt the same procedure when calculating checksum of the strings received from the
analyser.
PORT2 connection.
PORT1 connection
„R‟ 52 1 char R states for result message Note2: Bayer field is not used, in
Workstation Is the number identifying the instrument if more than this case show s the
2 char progressive analysis cycle
Number one is connected to Host computer
Patient ID
8 char Patient identification
Number
Wheel no. 2 char Indicate the progressive number of analysis made
Position 2 char Indicate the sample position inside the wheel
Bayer 2 char
(Note 2) See note 2
Result 4 char ESR result
Checksum 1 char See note 1
(Note 1)
Note1:
Field name Symbol Hex value Length Description
See note at the end of this
STX ☻ 02 1 char Start transmission
page
Checksum value is the result of Exclusive OR (XOR) of all characters of the message, excluded STX, ETX and Checksum. If the Checksum
value calculated by the analyser is 03 (corresponding to ETX), or 02 (corresponding to STX), or 00 (corresponding to NULL), the value will be
automatically replaced with 7F hex value.
Software House must adopt the same procedure when calculating checksum of the strings received from the analyser.
In the Communication menu is possible to choose the appropriate protocol and options:
A keyboard to type the password is going to be displayed. Type the password:1990 then press the “left arrow” to
confirm or “Clear” to exit without changing anything, in both cases, the instrument is going to display again the
previous image, then press “Back”.
A keyboard to type the password is going to be displayed. Type the password:1990 then press the “left arrow” to
confirm or “Clear” to exit without changing anything, in both cases, the instrument is going to display again the
previous image, then press “Back”.
Pressing Comm menu and TEST it is possible to send a communication test to host computer through RS232
serial port.
If DAT8 is chosen, the test Message If DAT15 or BCI is selected, the test
having 8 ID digits will be sent to Host Message having 15 ID digits is sent to
is: Host is:
<0x2>R01TESTR10A0001010002U<0x3> <0x2>R01TESTR10A 0101010002t<0x3>
<0x2>R01TESTR10B0001020015S<0x3> <0x2>R01TESTR10B 0102010015r<0x3>
<0x2>R01TESTR10C0001030031U<0x3> <0x2>R01TESTR10C 0103010031t<0x3>
<0x2>R01TESTR10D0001040090^<0x3> <0x2>R01TESTR10D 0104010090<0x7f><0x3>
<0x2>R01TESTR10E0001050100V<0x3> <0x2>R01TESTR10E 0105010100w<0x3>
<0x2>R01TESTR10F0001060120T<0x3> <0x2>R01TESTR10F 0106010120u<0x3>
<0x2>R01TESTR10G000107-004N<0x3> <0x2>R01TESTR10G 010701-004o<0x3>
<0x2>R01TESTR10H000108-004N<0x3> <0x2>R01TESTR10H 010801-004o<0x3>
<0x2>R01TESTR10I000109-002H<0x3> <0x2>R01TESTR10I 010901-002i<0x3>
<0x2>R01TESTR10J000110-001@<0x3> <0x2>R01TESTR10J 011001-001a<0x3>
If NO HOST is configured, the instrument will not communicate with Host Computer
to HOST computer through RS232 serial port (or to LAB PC if it is used like a bridge) if Query Setup option
is 1. (Note 3)
2. HOST computer verifies if the patient identity has been inserted in the working list and then it replies with "T"
message. Roller 10 remains waiting for receiving it on time (within 2 seconds), if ACK option is disabled.
It will wait for all the set time if ACK option is enabled. (Follow the instruction looking on page 22 of this manual.)
3. As soon as the instrument receives “T” message, it is checked. <STX>T01123456789 0001e<ETX>
4. The wheel is going to rotate to one position to allow the operator
to insert the sample tube, if “T” message has the result field <STX>T01123456789 0000d<ETX>
compiled with “0001” (that means the sample requires the
analysis). The wheel will remain stop, if the result field is “0000” (the analysis of that particular sample is not
requested).
th
Before starting the analysis, the explained above, from 1 to 2, procedure will be repeated for the further 10
th
(Roller 10) or the 20 (Roller 20) sample tubes.
In this case the instrument doesn‟t query the Host Computer, it just perform the analisis and send the result to host
computer as explaned at points 3 and 4.
Warnings:
1. If the patient identity (ID) read by the Scanner doesn‟t find its own correspondent in the Host computer
working list, “T” message will be replaced by <NAK> message (§ symbol Hexadecimal code 0x21
looking at the hyper terminal program). If MISS ID option is enabled, the wheel
will rotate to the next position to allow the operator to insert the tube. The sample, then will be analysed and
“R” message sent to host.
2. If no label has been stuck on the tube or the external Scanner cannot read the patient ID, “Q” message won‟t
be created. The wheel is going to rotate to accept the sample tube by pressing “Next” button on touch-screen
display.
3. The sample is going to be processed and “R” message sent to host with an identity code generated by itself.
The new patient ID in this case will be composed by 10 characters:
4. Cycle number (2 characters from 01 to 99), instrument s/n (4 characters), wheel number ( 2 characters)
sample position in the wheel (from 01 to 10- 2 characters).
Example:
Instrument s/n Wheel number
0307920106
Note 1: Checksum value is the result of Exclusive OR (XOR) of all characters of the message, excluded STX, ETX and
Checksum. If the Checksum value calculated by the analyser is 03 (corresponding to ETX), or 02 (corresponding
to STX), or 00 (corresponding to NULL), the value will be automatically replaced with 7F hex value.
Software House must adopt the same procedure when calculating checksum of the strings received from
the analyser.
Note 2: <STX> character (☻symbol looking at the hyper terminal program) in the serial transmission is represented by
0x02 hexadecimal bytes and <ETX> ( ♥ symbol looking at the hyper terminal program) by 0x03.
Note 3: If Query Setup option is 0, no Query protocol will be created; the instrument will accept and analyse all samples.
Note 4: Follow the instruction looking at the of the technical manual for TIMEOUT UART activation.
Note 5: This field is related to Bayer parameter on SETUP. If Bayer is 0, the field contains the number of Cycle. If Bayer is
1, the field contains '01' (hex 30 and 31) to keep compatibility with Bayer protocol.
Please notice for easiness of explanations, we assume Roller10, Roller10PN, Roller20PN and Roller20MC from this
moment are referred as “Roller Instruments”
In that case every instrument connected to the PC-LAB has to have an identification number different for each and
the PC-LAB the same number of serial ports as the instruments.
Our instruments have an option: TEST1 number in the COMMUNICATION menu for TEST1 models, ROLLER
NUMBER inside Tech Menu for ROLLER20LC models and STR N° inside COMM menu for the “Roller
Instruments” family” to assign a different identification number to each instrument,
“R” message should be different for each like in the following example:
LATEX PROTOCOL INTERFACE EXPLICATION VALID ONLY FOR TEST1 AND ROLLER20LC Up to Sw.
version 6.01C (Test 1), 6.01B (Roller20LC)
The three R Latex messages are sent, to the serial port, after having analysed the three referent Latex of the
Control kit and at the end of the washing, by withdrawing water from the fifth and sixth inserted tube, for rinsing the
capillary.
It should be understood R Latex messages are not available by using Calibration Latex kit.
Messages
„R‟ message for Latex
Hex Length
Field name Description
value
STX (Note 2) 02 1 char
„R‟ 52 1 char Means the instrument is sending a Result
Identifies the workstation if more instrument are
Workstation Number 2 char
connected, normally is 01
LATEXSTDL 9 char Is the identification of the latex tube
Can be 1, 2 or 3 and identifies level 1, 2 and 3 of
Field 1 1 char
latex
ESR Value 3 char This is the expected Latex ESR value
Space. 20 2 char Two spaces
Field 2 2 char Fixed to 01
Field 3 2 char The same as Field 1
Field 4 2 char The same as Field 2
ESR Value 4 char This is the detected Latex ESR value
Checksum (Note 1) 1 char
ETX 03 1 char
Note 1: Checksum value is the result of Exclusive OR (XOR) of all characters of the message, excluded STX, ETX and
Checksum. If the Checksum value calculated by the analyser is 03 (corresponding to ETX), or 02 (corresponding
to STX), or 00 (corresponding to NULL), the value will be automatically replaced with 7F hex value.
Software House must adopt the same procedure when calculating checksum of the strings received from
the analyser.
Note 2: <STX> character (☻symbol looking at the hyper terminal program) in the serial transmission is represented by 0x02 hexadecimal
bytes and <ETX> ( ♥ symbol looking at the hyper terminal program) by 0x03.
LATEX PROTOCOL INTERFACE EXPLICATION From Sw. version 6.01D on ward up to Sw 6.51x and from
Sw 8.00x on ward (Test 1), 6.01C up to Sw 6.51x and from Sw 8.00x on ward (Roller20LC) and from 1.00P on
ward ( Roller 10-10PN / 20PN / R20P2) and Roller 20 MC (from Sw. version 4.00A)
When the instrument runs latex controls, it is possible to send the results to host computer, in this case the instrument will
send a specific string here explained:
Please notice for easiness of explanations, we assume Roller10, Roller10PN, Roller20PN, Roller20PN2 and
Roller20MC from this moment are referred as “Roller Instruments”
Note 1: Checksum value is the result of Exclusive OR (XOR) of all characters of the message, excluded STX, ETX and
Checksum. If the Checksum value calculated by the analyser is 03 (corresponding to ETX), or 02 (corresponding
to STX), or 00 (corresponding to NULL), the value will be automatically replaced with 7F hex value.
Software House must adopt the same procedure when calculating checksum of the strings received from
the analyser.
<0x2>R01LATEX08451005 0101010002<0x19><0x3>
<0x2>R01LATEX08452036 0102010034<0x1c><0x3>
<0x2>R01LATEX08453083 0103010071<0x13><0x3>