Professional Documents
Culture Documents
Operator's Guide - Dimention
Operator's Guide - Dimention
02, 2019-03
Dimension EXL with LM/EXL 200 System Operator’s Guide
1
Dimension EXL with LM/EXL 200 System Operator’s Guide
2
Dimension EXL with LM/EXL 200 System Operator’s Guide
3
Dimension EXL with LM/EXL 200 System Operator’s Guide
Calibration and
Quality Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Calibration and Verification Overview . . . . . . . . . . . . . . . . . . . 4-2
Calibrating the IMT System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Calibrating Urine Drugs of Abuse Assays . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Calibrating HA1C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Calibrating HB1C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
When to Calibrate or Verify Photometric and LOCI Assays . . . . . . . . . . 4-3
Calibration Alert . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Group Calibration Alerts and Run Calibration Group . . . . . . . . 4-5
Calibrating Photometric and LOCI Assays . . . . . . . . . . . . . . . . 4-6
Setting Up and Running a Manual Calibration . . . . . . . . . . . . . . . . . . . . 4-6
Reviewing and Accepting Calibration Results . . . . . . . . . . . . . . . . . . . . 4-7
Cancelling a Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Viewing Calibration History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Restore Calibrations screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Restoring a Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
Calibration Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
Troubleshooting Precision of Calibration Results . . . . . . . . . . . . . . . . . 4-11
Troubleshooting Calibration Statistics . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
Troubleshooting Quality Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Verifying Photometric Assays . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
Setting Up and Running a Verification . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
Reviewing and Accepting Verification Results . . . . . . . . . . . . . . . . . . . . 4-14
Cancelling a Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
Verification Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
Troubleshooting Precision of Verification Results . . . . . . . . . . . . . . . . . 4-16
Troubleshooting Verification Statistics . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
Troubleshooting Quality Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
Calibrating the IMT System . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
Electrolyte Coefficients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
Using Calibration Status Lists . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
Create a List of Methods for Calibration . . . . . . . . . . . . . . . . . . . . . . . . 4-19
Deleting a Selected Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
Create a Reminder List for Expiring Calibrations . . . . . . . . . . . . . . . . . . 4-20
4
Dimension EXL with LM/EXL 200 System Operator’s Guide
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
General Instrument Care and Cleaning . . . . . . . . . . . . . . . . . . 5-2
Undiluted Bleach . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Bleach Solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Sodium Hydroxide Solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Opening the Reagent Lid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Opening the LOCI Access Lid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Removing the HM Incubation Wheel Cover . . . . . . . . . . . . . . . . . . . . . 5-4
Weekly Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Cleaning HM Wash Probes and the R2 Reagent Probe . . . . . . . . . . . . . 5-5
Cleaning the R2 Reagent Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Monthly Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Cleaning the Clot Check Drain on the IMT Port . . . . . . . . . . . . . . . . . . . 5-8
Replacing IMT Pump Tubing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Cleaning the IMT System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Replacing Instrument Air Filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
Styletting HM Wash Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
Replacing the HM Pump Heads on Wash Station . . . . . . . . . . . . . . . . . 5-17
5
Dimension EXL with LM/EXL 200 System Operator’s Guide
6
Dimension EXL with LM/EXL 200 System Operator’s Guide
Alignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
General Alignment Information . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Alignment Offsets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Printing and Viewing Alignment Offsets . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Barcode Scanner Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Cuvette Ring Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
HM Module Alignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Aligning the Incubate Wheel to the Transfer Opening . . . . . . . . . . . . . 6-7
Aligning the Wash Wheel to the Incubate Wheel . . . . . . . . . . . . . . . . . 6-8
Aligning the Wash Probes to the Wash Wheel . . . . . . . . . . . . . . . . . . . 6-9
Aligning the Vessel Transfer Shuttle to the
Incubate Wheel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
Aligning the Vessel Transfer Shuttle to the Wash Wheel . . . . . . . . . . . 6-11
Aligning the Vessel Transfer Shuttle to Load . . . . . . . . . . . . . . . . . . . . 6-12
LOCI Arm Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-15
LOCI Arm to Vessel Alignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-15
LOCI Arm to Detector Alignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
Photometer Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
R1 Reagent Probe Alignments . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
R1 Reagent Arm to Drain Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
R1 Reagent Arm to Target (Cuvette) Alignment . . . . . . . . . . . . . . . . . . 6-19
R1 Reagent Arm to Reagent Tray Alignment . . . . . . . . . . . . . . . . . . . . . 6-21
R2 Reagent Probe Alignments . . . . . . . . . . . . . . . . . . . . . . . . . 6-22
R2 Reagent Arm to Drain Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-22
If a Coarse Adjustment (R2 Home Flag) is Needed . . . . . . . . . . . . . . . . . . . . . . . . . . 6-23
R2 Reagent Arm to Target (Cuvette) Alignment . . . . . . . . . . . . . . . . . . 6-24
R2 Reagent Arm to Reagent Tray Alignment . . . . . . . . . . . . . . . . . . . . . 6-25
R2 Reagent Arm to HM Incubate Wheel Alignment . . . . . . . . . . . . . . . 6-26
Reagent Tray Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-27
Sample Probe Alignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-29
Sample Probe to Cuvette Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-30
Sample Probe to Cup Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-31
Sample Probe to Drain Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-33
Sample Probe to the HM Incubate Wheel Alignment . . . . . . . . . . . . . . 6-33
Sample Probe to the IMT Port Alignment . . . . . . . . . . . . . . . . . . . . . . . 6-35
Sample Probe Maximum Depth Alignment . . . . . . . . . . . . . . . . . . . . . . 6-36
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Troubleshooting Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Help Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Specifying the Problem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Solving the Problem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
7
Dimension EXL with LM/EXL 200 System Operator’s Guide
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Selecting Instrument Options . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Select or Change an Instrument Option . . . . . . . . . . . . . . . . . . . . . . . . 8-3
System Configuration Menu Screen Fields . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
System Configuration Menu Function Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5
Possible Dimension EXL System to LIS Communication Configurations . . . . . . . . . 8-6
Clot Check Configuration Menu screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6
Configuring Automated System Check Using CHK . . . . . . . . . 8-7
Automatic Cuvette Removal . . . . . . . . . . . . . . . . . . . . . . . . . . 8-8
Sample Auto Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
Automatic Reflex Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
Setting Up an Automatic Reflex Test . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
To Use Automatic Reflex Test Feature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-11
To Deactivate Reflex Testing for a Specific Assay . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-11
Automatic Panic Rerun . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-12
Setting Up an Automatic Panic Rerun . . . . . . . . . . . . . . . . . . . . . . . . . . 8-12
Automatic Reflex Testing and Panic Rerun Report Slips . . . . . 8-13
Using Calculated Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-15
Calibration Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-19
Define Calibration Product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-19
Scanning Barcode Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-19
Entering Data Manually . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-19
8
Dimension EXL with LM/EXL 200 System Operator’s Guide
9
Dimension EXL with LM/EXL 200 System Operator’s Guide
Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
General Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Biohazard and Probe Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Instrument Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Safety Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
Safety Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
Symbols Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-7
Instrument Labeling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-8
Performing Power Shutdowns and Startups . . . . . . . . . . . . . . 9-9
Types of Shutdowns . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-9
Important Power Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-9
Performing a Controlled Power Shutdown . . . . . . . . . . . . . . . . . . . . . . 9-10
Restoring Power After a Controlled Shutdown . . . . . . . . . . . . . . . . . . . 9-10
Performing an Emergency Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . 9-10
10
Dimension EXL with LM/EXL 200 System Operator’s Guide
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-1
11
Dimension EXL with LM/EXL 200 System Operator’s Guide
12
Dimension EXL with LM/EXL 200 System Operator’s Guide System Overview
1-1
System Overview Dimension EXL with LM/EXL 200 System Operator’s Guide
Intended Use
The Dimension® EXL™ with LM system and Dimension® EXL™ 200 system are in
vitro diagnostic devices that are intended to measure a variety of analytes in
human body fluids. The systems utilize photometric, turbidimetric,
chemiluminescence and integrated ion selective multisensor technology for
chemical and immunochemical applications for clinical use.
This guide is used to supplement a laboratory’s procedure manual. The College
of American Pathologists (CAP) has approved the use of a manufacturer’s
procedure manual as a component of the overall departmental procedures if it
complies with Clinical and Laboratory Standard Institute (CLSI, formerly
NCCLS) GP2-A5 guidelines.
1-2
Dimension EXL with LM/EXL 200 System Operator’s Guide System Overview
1-3
System Overview Dimension EXL with LM/EXL 200 System Operator’s Guide
Chapter Contains
4: Calibration & Quality Controls Procedures for reagent calibration and quality
control on the Dimension EXL system.
1-4
Dimension EXL with LM/EXL 200 System Operator’s Guide System Overview
1-5
System Overview Dimension EXL with LM/EXL 200 System Operator’s Guide
Features
Table 1-1: Features: Testing
Testing
Can perform up to 440 photometric chemistry tests and 187 IMT test per hour on serum, plasma, urine
and cerebrospinal fluid samples.
Can test for electrolytes using the Siemens exclusive QuikLYTE integrated multisensor.
Can process high-sensitivity immunoassays using the Siemens exclusive LOCI technology.
Consumable
Can automatically remove reagent cartridges when these cartridges are empty or their onboard life has
expired.
Can use the Flex reagent cartridge and onboard cuvette manufacturing system.
QC or Calibration
1-6
Dimension EXL with LM/EXL 200 System Operator’s Guide System Overview
Software
Has the ability to link with other Siemens analyzers (with or without barcode generation capability) or
with a Laboratory Information System (LIS).
Sample Tubes
Uses 1-mL SSC (Small Sample Container) on top of barcoded tubes for short volume samples.
1-7
System Overview Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
1 IMT Lid
2 Sample Area
3 LOCI Module
4 HM Area
5 Reagent Lid
6 Touchscreen
7 IMT Area
8 Keyboard
9 Printer Shelf
10 RMS
11 Control Boards
12 Computer
14 Pump Panel
18 Reagent Area
1-8
Dimension EXL with LM/EXL 200 System Operator’s Guide System Overview
# Description
3 IMT Port
5 Flush Solution
6 Standard A
7 Standard B
9 Sample Diluent
1-9
System Overview Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
2 Vessel Holder
3 Wash Probe 2
6 Wash Probe 1
7 Wash Pump 2
8 Wash Pump 1
9 Wash Wheel
10 Incubate Wheel
12 Vessel Shuttle
1 - 10
Dimension EXL with LM/EXL 200 System Operator’s Guide System Overview
# Description
1 R2 Reagent Arm
2 Reagent Tray
3 R1 Reagent Arm
1 - 11
System Overview Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
1 Thermal Chamber
4 Flow Meter
5 Compressor
1 - 12
Dimension EXL with LM/EXL 200 System Operator’s Guide System Overview
# Description
1 Printer
2 Safety Label
3 Shelf
4 Supports
1 - 13
System Overview Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
1 LOCI Arm
4 Power Supply
8 Vacuum Assembly
1 - 14
Dimension EXL with LM/EXL 200 System Operator’s Guide System Overview
Software Overview
Using the Keyboard
Figure 1-8: Keyboard Layout
The Dimension EXL system Keyboard has six areas which are discussed in the
following sections.
Note: Some of these keys can be used in combination. These are
referred to as keystroke combinations, which are also
described on the following pages.
Test Keys
Use the test keys to select tests and panels of tests.
The test keys labeled P1 through P10 are the panel keys. By programming
groups of frequently used tests to these ten panel keys, a panel of up to 20
tests can be requested with a single keystroke. (See “Creating Panel Keys” on
page 43 of Chapter 8: Configuration). Tests can also be assigned to specific
test keys. (See “Test Key Assignments” on page 76 of Chapter 8:
Configuration).
Figure 1-9: Test Key Panel Setup
1 - 15
System Overview Dimension EXL with LM/EXL 200 System Operator’s Guide
Keyboard Overlays
Additional tests may be shown on the overlay above or below the test keys.
To select these tests, press Alt and the key indicated by the overlay. For
example, the TSHL test is on an overlay above the test keys:
Figure 1-10: Keyboard Overlays
Action Keys
Table 1-6: Action Keys
Key Use
Stop The Stop key can be used only with the Control key. This key combination stops
all instrument operations in progress without damaging the instrument. All tests
in progress are aborted; however all scheduled tests are retained in instrument
memory. Press the Reset key to resume operations.
PAUSE The PAUSE key turns the sampler system off, preventing the photometric
sample arm from moving. Press the PAUSE key again to restart the sampler
system.
Reset The Reset key is used to clear error messages on the screen. It initializes any
modules that are not being used and causes the instrument to resume
processing. This key is also used to resume operations if the
Control/Stop key combination is used to halt processing.
Run The Run key causes the instrument to look for and process any new samples. It
is equivalent to the F4: RUN key on the Load List screen.
Function Key
The Function keys, labeled F1–F8, have different actions depending on which
screen is displayed. Keys are identified in the boxes at the bottom of the
screen.
1 - 16
Dimension EXL with LM/EXL 200 System Operator’s Guide System Overview
Keypad Keys
Table 1-7: Keypad Keys
Key Use
Delete When entering information in a field, hold down the Shift key and select DELETE
to remove all characters to the right of the cursor in a field.
Exit Press the Exit key to leave the screen that is currently on the display; the system
displays the previous screen. This key operates the same way as the Exit key on the
touchscreen.
Help Press the Help key at any time to get information about the screen that is on the
display or the functions of various keys on the keyboard that are used with that
screen.
Alarm Off When there is a problem with the system, an alarm sounds. Press the Alarm Off
key to turn off this alarm. Press Alarm Off again to enable the alarm.
Enter Press ENTER after entering data to store what was entered in the current field or to
activate a command.
Numbers Use to enter numerical data. Numerical data can also be entered using the
Keyboard keys.
Keyboard Keys
Table 1-8: Keyboard Keys
Key Use
BACKSPACE Deletes the character immediately to the left of the cursor when entering/
editing data in a field.
ENTER Stores what was entered in the current field or to activate a command.
\ Place the backslash key at the far left of a data field to delete the entire entry in
a data field.
Highlight a character in a data field and press the backslash key to delete all
characters to the right of the cursor.
ALT Used in combination with other keys, displays other screens or performs
routine functions. See “Keystroke Combinations” in the Appendix for a
description of the key combinations.
1 - 17
System Overview Dimension EXL with LM/EXL 200 System Operator’s Guide
Help Keys
The Help keys provide help on a specific screen's tasks, Function keys, and in some
cases procedures.
Table 1-9: Help Keys
Help The Quick Index. The Quick Index is a listing of specific tasks and
(From the Operating Menu) how to display the screens where help on performing that task is
available. The Quick Index includes the Function Key sequence to
display those screens from the Operating Menu.
Help Displays information on how to use the screen and the function keys
(From Any Other Screen) that are currently on the display.
SHIFT+Help If the cursor was in a field in which data can be entered (e.g. Patient
Name on the Enter Sample Data screen), information on acceptable
entries for that field appears.
ALT+Help Presents a screen that explains how Help works and a list of the Alt
key combinations. It contains much of the same information
provided in this table.
CTRL+Help Displays a list of information for all operating condition icons. Scroll
down on this page to get information for a particular icon.
1 - 18
Dimension EXL with LM/EXL 200 System Operator’s Guide System Overview
Keystroke Combinations
To use a keystroke combination, hold down the first key and press the second
key.
Table 1-10: Keystroke Combinations
CONTROL/Stop Stops all operations in progress without damaging the instrument. All tests in
progress are aborted. However, all scheduled tests are retained in instrument
memory.
SHIFT+DELETE When entering information in a field, press SHIFT +DELETE to delete all
characters to the right of the cursor in a field.
SHIFT+EXIT Moves from the current screen directly to the Operating Menu.
SHIFT/--> When entering information in a field, hold down SHIFT and press the right or
left arrow to move the cursor one space to the left or right.
ALT+L Displays the Load List – Short Samples screen. To display the All and New
Samples views, press F2: NEXT STATUS.
ALT+M Displays an explanation and troubleshooting information for the error message
that is on the screen.
ALT+R When the reagent manager icon appears in the Operating Conditions Status
area of the screen, press ALT+R to see information on why the icon appeared.
ALT+S Displays the Segment Status screen to view the status of either the segment
positions currently loaded on the instrument (On Board Segments view) or all
segments (All Segments view).
1 - 19
System Overview Dimension EXL with LM/EXL 200 System Operator’s Guide
Operating Passwords
Operating Passwords let the operator customize the instrument operation to
their work.
1. To start or stop using a password, display the System Configuration
Menu screen.
2. Select F7: PASSWORD and type the password exactly as it appears
below.
Table 1-11: Operating Password
SHOWCL A chloride test is automatically requested when the Na/K test key is pressed. This
does not require additional sample.
ignoredup The software will not process samples on the instrument that have duplicate
barcode IDs. However, an operator may need to run a load list with samples that
have the same barcode ID by using the password "ignoredup" to ensure those
samples will run.
DATA Turns on Method Diagnostic Data and allows the operator to print filter data for
test result troubleshooting, “Printing Filter Data for a Test Result” on page 7 of
Chapter 7: Troubleshooting
1 - 20
Dimension EXL with LM/EXL 200 System Operator’s Guide System Overview
1 - 21
System Overview Dimension EXL with LM/EXL 200 System Operator’s Guide
Status Meaning
Initializing The system is starting up. Samples cannot be processed and access to
the sample wheel is disabled until the system has finished initializing.
Processing The system is processing samples, the IMT system is being calibrated,
or photometric calibrations or reagent hydrations are being
performed.
Diagnostics Diagnostic software is being used. Samples cannot process when this
box displays “Diagnostics.” The instrument must be in “Standby” to
access the Diagnostic software.
Status Meaning
Sampler Busy The system is using the sample handler or the sample wheel.
Waiting... The PAUSE key was pressed while samples were processing.
The system is waiting for an appropriate place and time to stop the
photometric sampler system.
Waiting...60 The instrument is 60 seconds from pausing. This will count down the
number of seconds until the instrument is paused (“Waiting...59,”
“Waiting...58,” etc.).
Samplers Off Displays when PAUSE is selected. The photometric sampler system
has stopped and will not move until PAUSE is selected again.
Moving Wheel... The instrument is within five seconds of accessing samples in the
sample area.
1 - 22
Dimension EXL with LM/EXL 200 System Operator’s Guide System Overview
Status Meaning
IMT Paused All activity in the IMT system has been paused. The IMT system will not
move, prime, or calibrate unless the function key is pressed or the
screen is closed.
IMT Calibrating The IMT system is running a calibration. IMT tests can be scheduled
and will be run when status changes to “IMT OK.”
Na, K, Cl One or more of the individual sensors failed calibration or has been
overridden. The color indicates the status of each:
Green = calibrated
Yellow = IMT method has been overridden
Red = not calibrated
IMT Not Calib The IMT system is not calibrated. IMT tests cannot be processed until
the IMT system is calibrated.
IMT Not Config The IMT system has not been configured. IMT tests cannot be
processed.
Status Meaning
When Flashing "Local Instrument is in local access mode and is being controlled by on-site
Access" personnel.
1 - 23
System Overview Dimension EXL with LM/EXL 200 System Operator’s Guide
A segment box with a dash indicates that the instrument has scanned the
sample wheel and did not identify a lettered segment in that segment
position.
1 - 24
Dimension EXL with LM/EXL 200 System Operator’s Guide System Overview
Reagent Temperature
Table 1-17: Icon Symbol and Meaning
Icon Description
1 - 25
System Overview Dimension EXL with LM/EXL 200 System Operator’s Guide
Cuvette Temperature
Table 1-18: Icon Symbol and Meaning
Icon Description
1 - 26
Dimension EXL with LM/EXL 200 System Operator’s Guide System Overview
Icon Description
A red ON inside this icon indicates that power to the instrument has
just been interrupted and the UPS is in use. An intermittent alarm
also sounds, changing the word to LOW when the instrument is
preparing to shut down before the UPS loses all power. The UPS is
recharged when normal electric power is restored. If LOW appears
continuously while the instrument is connected to the wall outlet,
contact the Customer Care Center - Technical Solution: Inside U.S.:
1-800-441-9250; Outside U.S. Refer to the local support team.
Icon Description
The reaction vessel sensor, located near the top of the vessel feeder
chute, is not detecting a reaction vessel. Add reaction vessels to the
holder or check for a vessel jam in the vessel transfer system.
Icon Description
1 - 27
System Overview Dimension EXL with LM/EXL 200 System Operator’s Guide
Reagent Manager
Table 1-22: Icon Symbol and Meaning
Icon Description
Printer
Table 1-23: Icon Symbol and Meaning
Icon Description
Short Sample
Table 1-24: Icon Symbol and Meaning
Icon Description
One or more sample containers in the current Load List does not
have enough sample volume to run all its requested tests. Press
ALT+L to view a list of these short samples. See “Resolving a Short
Sample Detected” on page 13 of Chapter 3: Sample Reports.
1 - 28
Dimension EXL with LM/EXL 200 System Operator’s Guide System Overview
Check Needs
Table 1-25: Icon Symbol and Meaning
Icon Description
The yellow Needs Check icon appears when the system is checking
itself for any needs to process the Load List.
If the red Check Needs icon appears, the system has found needs
that require operator attention before samples can be processed.
Press ALT+N to display the System Needs screen to see which needs
are required. To fill these system needs, see “Responding to System
Needs” on page 9 of Chapter 2.
Alarm Status
Table 1-26: Icon Symbol and Meaning
Icon Description
The alarm has been turned off by the operator. If an error condition
occurs, the alarm will not sound; however, a message indicating the
reason for the alarm appears in the error message area.
Icon Description
Indicates that the Clot Check Error sample aborted error has
occurred. It shares the same location as the printer icon. If both a
printer icon and a clot check icon need to be displayed, the screen
will flash from one icon to the other.
Indicates that the Clot Check feature is turned off. It shares the same
location as the printer icon. If both a printer icon and a clot check
icon need to be displayed, the screen will flash from one icon to the
other.
1 - 29
System Overview Dimension EXL with LM/EXL 200 System Operator’s Guide
Applications Area
Below the error message area is the applications area, where sample data is
either entered or edited and test requests are made. The color of the words in
the Application Area of any screen indicate whether information in that field
can be changed:
• Black indicates a text field. Information cannot be changed.
• White indicates a data field. Information can be changed.
Message Area
The message area is below the applications area. This area can contain the
following kinds of messages:
Table 1-28: Message Area Type and Meaning
Prompt When the system needs additional information about a certain task, it
displays a message and waits for a response.
The function keys area is located at the bottom of the screen. It contains two
rows of four boxes each. The titles on these eight boxes correspond to the
eight function keys, F1 through F8, on the keyboard.
1 - 30
Dimension EXL with LM/EXL 200 System Operator’s Guide System Overview
Alert Keys
Five alert keys are displayed vertically on the left side of the screen. These keys
change color to alert situations that require attention. Select the key to find
out the reason for the alert.
Table 1-29: Alert Key Type and Purpose
Key Purpose
STAT Status Displays the STAT Samples screen. Shows the STAT requests that are
pending, in-process, and completed.
Sample Alert Displays the Sample Alert screen. Shows rerun status of sample
requests which encountered exceptions while processing (autodilute,
reflex, panic repeats, process errors)
Supplies Displays the Reagent Cartridge Alerts screen. Shows Flex reagent
cartridge lots which are nearing an operator-defined depletion
threshold. Warns that it is time to load cartridges for specific methods.
QC Alert Displays the QC Tests Out of Range screen. Shows the method, QC level
and high or low status of QC tests.
Calib Alert Displays the Calibration Alert screen. Shows onboard methods/lots and
the amount of time left to calibration expiration. Shows onboard lots
which have triggered configurable calibration alert criteria.
Key Purpose
Home Closes the currently displayed screen and returns to the Operating
Menu (Home) screen.
Enter Select Enter after entering data to store what was entered in the
current field or to activate a command.
Exit Closes the currently displayed screen and returns to the previously
displayed screen.
Arrow Keys Moves the cursor from field to field on the screen displayed in the
Applications Area.
1 - 31
System Overview Dimension EXL with LM/EXL 200 System Operator’s Guide
1 - 32
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
2-1
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
Daily Setup
Overview
The Daily Setup procedure includes four tasks:
• Cleaning the sample area and emptying cuvette waste
• Performing Daily Maintenance
• Running a System Check
• Recording the Daily Maintenance results
• Onboard Maintenance Logs provide the capability to document daily,
weekly, and monthly logs either on paper or in the on-board electronic logs
saved on the instrument.
Note: Systems with the RMS configured should refer to the RMS
Operator’s guide for additional Daily Setup tasks.
WARNING: The cuvettes and the contents of the cuvettes may present a
biohazard or chemical hazard. Follow standard laboratory
procedures for protection from biohazards and chemicals
when performing maintenance and troubleshooting
procedures
11. Update the Maintenance Log.
2-2
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
Daily Maintenance
1. Select F4: SYSTEM PREP > F8: DAILY MAINT.
Note: If all System Check Means are failing and ABS/CHK lots were
recently changed, enter the carton value from the ABS/CHK
carton end flap every time a new ABS/CHK lot is used.
2. Refer to the section “Understanding the System Counters Screens” on
page 7.
Note: Additional maintenance items appear in red. Check the RMS
Operator’s Guide for RMS Daily Maintenance requirements.
2-3
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
• * The temperature range listed here is of the HM ring itself, not the air bath or liquid in a reaction
vessel. In order to maintain a constant temperature inside the reaction vessel, the system
compensates for ambient temperature, thus varying the HM ring temperature. It is common for the
HM ring temperature to vary throughout the day in order to compensate for ambient temperature.
In this manner, the system is able to maintain the temperature of the liquid in the reaction vessel
between 36.5º C and 37.5º C.
• Numbers in brackets such as [37.0] mean that the temperature is not current. Dashes inside
brackets [--] mean that one or more requirements for monitoring temperature are not met, for
example, the lid may be open.
Mean SD
2-4
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
Mean SD
Note: If the System Check fails and tests are aborted, go to the Error
Log screen. Display the Error Log screen by navigating to: F5:
PROCESS CONTROL > F6: ERROR LOG, and resolve any error
messages and rerun the system check. See “System Check
Troubleshooting” on page 8 of Chapter 7: Troubleshooting.
2-5
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
MM/DD/YY form. Select F6: <== PREV LOG or F7: NEXT LOG ==> to scroll
through stored logs in the system.
Note: The previous 60 logs are stored on the instrument.
5. Select F8: PRINT.
Expiration: Refer to carton for expiration date of individual unopened reagent cartridges. Once
unwrapped, cartridges are stable for 30 days.
There are numerous distinct tests processed during a system check, each using
a different fluid delivery sub-system. For each test, the absorbance difference
between two wavelengths of Ponceau S diluted as sample or reagent are
measured by a two-filter (510, 600 nm) endpoint technique. The results of
these determinations must meet the specifications given in the specific
Operator’s Guide. If these conditions are not met, a system malfunction may
be indicated. Refer to the specific Operator’s Guide for troubleshooting
information.
Table 2-5: System Check Tests & Conditions
Test Condition
Temperature 37º C
2-6
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
2-7
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
2-8
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
Supplies
System Needs
Adding:
• Reagent Cartridges
• IMT Consumables
• Cuvette Film Cartridges
Performing:
• Calibration/Verification on photometric methods, LOCI methods, and the
IMT system
• Quality Control
Before beginning sample processing, the system checks to see if any reagents
or supplies are needed or if any process control functions (calibration or quality
control) need to be performed.
When either F2: PROCESS SINGLE from the Enter Sample Data screen,
F4: RUN from the Load List screen, or RUN is selected, the system checks its
needs (the yellow Needs Check icon appears) for processing the requested
tests.
• If no system needs are required, the system begins processing.
• If system needs are required, the red Check Needs icon appears. Complete
the system needs listed or ignore them (tests that require those needs are
not processed). Tests are processed and reported only if you ignore QC
needs.
• The System Needs screen:
• Shows which system needs must be completed to process all samples on
the Load List. Samples with no system needs are scheduled for processing.
• Shows function keys only for those categories of supplies and procedures
that are needed.
2-9
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
3. To fill a specific system need, select its function key to see what is
needed. Function keys are displayed only for system needs that must be
filled to run the Load List.
Table 2-6: Function Keys
Function Key Press the function key and follow this procedure
F3: CAL/VER See “When to Calibrate or Verify Photometric and LOCI Assays” on
page 3 of Chapter 4: Calibration and Quality Control.
F4: IMT CONSUMS See “Replacing IMT Consumables” on page 22 of this chapter.
F5: IGNORE NEEDS See “Understanding System Needs” on page 11 of this chapter.
F7: SYS COUNTERS To replace the cuvette film cartridge, press F3: FILM LOAD and see
“Replacing the Cuvette Film Cartridge” on page 41 of Chapter 5.
4. If after filling a system need, other system needs are still required, the
System Needs screen displays and allows the operator to press another
function key to fill another need. If no additional system needs are
required, the system begins processing the samples.
2 - 10
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
Cartridges Not enough reagent available to run all As many tests as possible are run for
the tests. the assay.
Cuvette Cartridge Not enough film in the cuvette film As many tests as possible are run.
cartridge to make all the necessary
cuvettes.
IMT Consumable The QuikLYTE® sensor or a required IMT As many Na, K, Cl assays as possible
fluid is not sufficient enough to run the are run.
remaining assays.
OR
The time remaining for the QuikLYTE All Na, K, Cl assays are run.
sensor or a required IMT fluid has
expired.
IMT System The IMT system is not calibrated. An IMT calibration is scheduled. If it
fails, no Na, K, Cl assays are run.
QC QC for these lots has not been run All tests are run.
within the specified time period.
Photo Method Calibration status is “Expired” for these All assays are run, but a test report
reagent cartridges. message will appear on the report
slip.
OR
Calibration status is “Never Calibrated” Assays are not run and a test report
for these reagent cartridges. message appears on the report slip.
2 - 11
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
Supplies Alert
When the Supplies alert key changes to yellow, select the key to display the
Reagent Cartridge Alerts screen.
Figure 2-1: Reagent Cartridge Alerts Screen
Scenario Description
ACP A single ACP Flex cartridge is on board and has only five tests remaining, five fewer
than the designated alert threshold of 10.
ALB Multiple ALB Flex cartridges are on board, all with the same lot number. The “Tests
Left” field shows the cumulative number of tests from all ALB reagent cartridges.
CHOL Multiple CHOL Flex cartridges from two different lots are on board. The threshold of
161 is compared to the cumulative number of “Tests Left” from all CHOL reagent
cartridges.
When a method displays on the Reagent Cartridge Alerts screen, load enough
reagent cartridges to exceed the “Alert At” threshold for the method (see
“Adding Reagent Cartridges” later in this section).
2 - 12
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
Field Information
Lot Number The six-character (two letters and four numbers) manufacturing lot number for
the cartridge.
Calib Exp Date The date that the calibration for that lot expires.
System Exp Date The date after which the cartridge will not be used by the system and must be
discarded.
In Use YES (red)The cartridge is currently being hydrated or is being used by the
system to process a Load List or a System Check.
NO (white) The cartridge is not being used.
2 - 13
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
2 - 14
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
If the automatic loader will not insert the cartridge into the instrument, the
loader light blinks and a blinking reagent manager icon displays. Remove the
cartridge from the loader. Press the Alt/R key combination to see why the
reagent did not load.
• If the reagent tray is full, remove a reagent cartridge from the instrument
to make room for an additional cartridge. See “Removing Reagent
Cartridges” on page 16 of this chapter.
• If the reagent cartridge barcode could not be read, see “Manually entering
a Reagent Cartridge Barcode” on page 15 of this chapter.
2 - 15
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
d. Type the last three numbers of the sequence number and select
ENTER.
5. Select F1: ACCEPT DATA to confirm the addition of the reagent
cartridge. The automatic loader then places the cartridge into the
system.
WARNING: Wait until the red loader light is blinking before placing hands
or fingers into the loader to remove a reagent cartridge.
2 - 16
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
2 - 17
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
Replacing HM Consumables
Replacing HM Fluids
The HM module requires three fluids:
• Chemistry Wash
• Reagent Probe Cleaner
• Sample Probe Cleaner
The volume of chemistry wash, reagent probe cleaner, and sample probe
cleaner is monitored by a low level sensor in each bottle.
2 - 18
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
2 - 19
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
5. Unscrew and remove the bottle cap/pump assembly on the empty bottle,
and transfer it to the new bottle.
6. If it is necessary to put down the assembly, place it on a clean paper
towel.
7. Screw the bottle cap/dip tube assembly onto the new bottle and place
the bottle in the instrument.
8. Move the cursor to each bottle that was replaced and select ENTER.
9. Select F1: STORE CHANGES.
10. Run QC on HM methods after changing the Chemistry Wash.
CAUTION: Do not overfill the holder. Overfilling may cause jams in the
feeder. Use the red line on the holder as a guide for the maximum
reaction vessel limit.
2 - 20
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
WARNING: The used reaction vessels and waste container liner are
biohazards; use your laboratory’s safe biohazard waste
disposal procedures when discarding.
Figure 2-5: Reaction Vessel Holder
3. Place a new biohazard bag in the HM waste container. Ensure the bag is
fully open and conforms to the inside walls and bottom of the container.
4. Reinsert the waste container.
5. Select F4: SYSTEM PREP > F6: SYSTEM COUNTERS > F6: HM
COUNTERS.
6. Move the cursor to the Vessels in Waste field, and select ENTER to
change the field to YES.
7. Select F1: STORE CHANGES.
2 - 21
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
Diluent 39
Standard A 300
Standard B 57
Flush 90
2 - 22
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
Flush 3000
4. Enter new lot number or press enter to retrieve the old lot number.
5. Select F8: STORE CHANGES.
2 - 23
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
WARNING: Do not pull the holder all the way forward or it may lock.
Forcing it back into position may cause damage.
3. Slide the used QuikLYTE sensor out of the holder.
4. Insert the new QuikLYTE sensor. Apply equal pressure on both sides of
the sensor holder to close it properly.
5. Select F7: CHANGE SENSOR. The screen updates with the maximum
assays and time for the new QuikLYTE sensor.
6. Enter new lot number for the new sensor.
7. Select F8: STORE CHANGE. The system automatically primes the Salt
Bridge Solution and the IMT Condition and Dilution Check screen
appears.
2 - 24
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
Run a dilution check as part Refrigerate bottles of Dilution Never run a dilution check on a
of the replacement Check solution sensor that has already been used
procedure for the QuikLYTE to run samples
sensor
Use the same bottle of Dilution Never use Dilution Check solution
Check solution when checking from a bottle that has less than
performance across multiple one inch of fluid remaining in it
instruments
Never rerun a dilution check using
the same sample cup. Always pour
a fresh cup
1. Select F4: SYSTEM PREP > F3: IMT > F4: COND /DILCHK.
2. At the IMT Condition and Dilution Check screen, type the segment
position for the conditioning sample in the Start at Position field.
Note: Select a segment that has two consecutive positions with
adapters available
3. Select ENTER.
4. Select F4: COND & DILCHK to update assigned positions for the
conditioning and dilution check fluids.
5. Fill a sample cup with conditioning sample (fresh plasma or serum) and
place it into its assigned position shown on the screen.
6. Remove the cap from the Dilution Check Solution bottle.
7. Fill a sample cup by pouring directly from the bottle and load this sample
cup into its assigned position shown on the screen.
8. Select F5: START. An IMT calibration is scheduled automatically with the
dilution check.
a. If the dilution check passed, continue with Step 9 below.
b. If the dilution check failed, refer to the section “Resolving a Failed
Dilution Check ” in this chapter. When the dilution check passes,
running an IMT calibration is not required.
c. If the IMT calibration fails, press ALT+M and follow the help
procedures to resolve why the IMT calibration failed.
9. Review the IMT results on the IMT log electronically, or record the
calibration and dilution check results from the printout on the QuikLYTE
Results Log sheet.
10. Run QC for Na, K, and Cl.
2 - 25
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
2 - 26
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
2 - 27
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
2 - 28
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
Types of Segments
There are three colors of segments (black, yellow, and orange) available from
Siemens for use in loading sample containers onto the sample wheel. All are
identical except for their color. It is useful to use the yellow and orange colored
segments to:
• Designate SSC segments.
• Differentiate sample cup and tube segments.
• Identify STAT segments.
• Identify segments with calibrators and verifiers.
• Identify instruments by segment colors (in multi-instrument laboratories).
• Separate work from satellite or other labs.
2 - 29
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
• Sample cups must be placed into one of the adapters supplied by Siemens
to load them into a segment. Push the sample cup completely down onto
the adapter.
Figure 2-7: Sample Cup Adapters
# Description
3 Beige 7 mL Adapter
2 - 30
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
Adapter* Adapter
D L
Size Color
2 - 31
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
WARNING: Do not place an SSC on a sample tube that presents the SSC at
a higher position than the tube that was used to perform the
maximum depth alignment for the SSC. Failure to follow
these instructions can result in operator injury, exposure to
biohazardous samples, or damage to the instrument.
Figure 2-8: SSC Tube
# Description
2 1 mL
2 - 32
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
2 - 33
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
2 - 34
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
Preparing Samples
Sample Preparation
Ensure the sample is free of clots, fibrin strands, and other impurities that may
affect metering fluids through the instrument before it is loaded into a
segment. There should be no air bubbles in the sample container.
The Dimension EXL System HIL feature can alert for undesirable hemolysis,
icterus and lipemia in a serum or plasma sample; proper preparation is very
important. Follow the sample container manufacturer’s instructions and
specifications on proper storage and handling techniques for mixing, timing,
and centrifugation.
A sample tube does not have a barcode on Pipette the fluid into a sample cup before
it or there is less than 1 mL of sample processing
available
Using a barcode on the sample tube and Pipette the fluid into an SSC and place the
there is at least 1 mL of sample available SSC on that same barcoded sample tube.
Place this SSC/tube combination onto a
segment that has been designated for SSC
containers
2 - 35
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
2 - 36
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
Field Enter
Patient ID Displays only if External Printer mode = Test Results Only and the LIS
Communications mode = OFF.
Dilution (if used) Use the arrow keys to move the cursor down to the Location field and then over to
the Dilution field.
Location (if Use the arrow keys to move the cursor to this field.
used)
2 - 37
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
5. A new Enter Sample Data screen displays with the Position field
automatically updated to the next available position in that segment.
6. Repeat Step 2 through and Step 4 to enter more samples.
Field Explanation
Mode To choose a different mode, select F7: NEXT MODE to view the type of sample container
used with the sample. Possible modes are:
Fluid To choose a different sample fluid, select F8: NEXT FLUID. There are nine sample fluid
types to choose from: 4 Sample Fluids, and 5 QC Fluids.
Sample Fluids
• Serum • Plasma
• Urine • CSF/Blood
QC Fluids
• SerumQC1 • UrineQC1
• SerumQC2 • UrineQC2
• SerumQC3
When a QC fluid is selected, the Priority field must be QC or XQC. When a urine drugs of
abuse test is selected, the fluid field must be set to Urine.
2 - 38
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
Field Explanation
Priority The system automatically sets the priority to Routine. To choose a different priority,
select F4: NEXT PRIORITY.
• Routine • QC
• ASAP • XQC (Crossover QC)
• STAT
When QC or XQC priority is selected, the sample fluid must be set to SerumQC1,
SerumQC2, SerumQC3, UrineQC1 or UrineQC2. If there are two Flex® reagent lots of
the method on board and they have been calibrated, QC will be performed on both
lots.
Volume For reference only. It shows only how many µL of sample are needed to run the tests
requested.
Do not use the volume field to determine the amount of sample needed for
processing.
Dilution Enter a dilution factor in this field if a dilution of the sample was made manually.
Enter the manually made dilution factor as a whole number.
When a dilution factor is entered, that sample will not be autodiluted by the
instrument, except in certain circumstances see “Automatic Dilutions” on page 36 of
Chapter 8. To move the cursor to the Dilution field, move it to the Location field and
press the right arrow key.
Location This is a six character optional field used to indicate types of sample identification
information.
Tests Select tests by pressing a test key or a predefined panel key [P1-P10]. Up to 36 tests
can be ordered for a sample.
2 - 39
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
To change Use
Note: All remaining sample positions for the batch are assigned
automatically to onboard segments alphabetically and
numerically beginning with the segment position entered in
the Position field.
4. Select F1: ASSIGN POS.
5. Repeat Step 2 through Step 4 to enter more batch samples.
Note: If onboard segments in the segment status area of the screen
are A, X, L, M, then the remaining samples are automatically
assigned to the next available positions in the segment.
6. Select F3: LOAD LIST and use the Load List screen to load samples in
their assigned segment positions.
7. Make a separate handwritten list of patient samples and their positions
on the segment wheel.
8. Select the F4: RUN or RUN.
Note: To add single (non-batch) samples from the Enter Batch
sample Data screen, select F4: ENTER DATA.
2 - 40
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
Loading Samples
Samples Without Barcode Labels
For samples without barcode labels, use the Load List screen to see the
segment position assigned to each sample.
1. Select F1: ENTER DATA > F3: LOAD LIST.
2. If "New Samples" does not display, select F2: NEXT STATUS.
Note: Load List segment positions: c = Sample Cup, l = Limited Cup -
No Level Sense, t = Sample Tube, x = SSC.
3. Ensure that any lids on sample cups are pressed down and that all
stoppers have been removed from sample tubes
WARNING: Do not touch the sample wheel area while the instrument is
initializing or processing. It could cause self -injury or damage
to the instrument.
WARNING: Test tubes must be fully seated to the bottom of the adapter.
Do not touch a tube or segment while the instrument is
operating.
Note: Barcode sample tubes will not appear on the Load List screen
until they are located on the instrument. LIS-downloaded,
non-barcoded samples do not appear on the Load List screen
until they are assigned a segment position. See “Processing
Samples Downloaded from an LIS” on page 44 of this chapter
to assign positions.
2 - 41
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
2. Completely seat the segment in its position on the sample wheel. Do not
remove segments when the Moving Wheel Light is on or when the
photometric sampler status box displays Moving Wheel.
2 - 42
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
Processing Samples
When the sample information is available and segments have been loaded into
the sample wheel, begin processing by selecting the appropriate option.
2 - 43
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
WARNING: Do not touch the sample wheel area while the instrument is
initializing or processing. It could cause self-injury or damage
the instrument.
4. Select Run.
Sample ID/ F6: SYSTEM CONFIG > F4: Sample Sample cup (or to the
Computer Menu COMPUTER Edit type of sample container
used most frequently).
2 - 44
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
WARNING:
6. Repeat Step 2 through Step 5 until all required asterisked (**) samples
have been assigned positions and loaded.
7. Select Run.
2 - 45
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
2 - 46
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
Status Meaning
Entered This sample’s data has been entered into a Load List, but the operator has not
requested the system to run this sample yet.
Ready This samples data has been entered into a Load List, the sample has been loaded into
the sample area, and the operator has requested the system to run this sample.
Begun The system has begun to process tests for the sample; some test results may be
available.
Done All tests for the sample have been processed; all test results are available.
Printed All tests for the sample have been processed; all test results have been sent to the
printer.
Report All tests for the sample have been processed; all test results have been sent to an
external host computer.
2 - 47
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
Entered but not yet processing Not Started (if alert is active)
Entered (if alert is not active)
There are four display modes for the STAT Samples screen, depending on the
configuration. See “Configuring the STAT Status Alert Key” on page 48 of
Chapter 8: Configuration.
To change the display, select F3: Next Display.
Table 2-24: Sample Field Modes
Mode Displays
Show Processing Displays all STAT samples currently processing. The When
Available field is updated every five seconds. When a sample is
completed, it is removed from this list.
STATS Not Started Displays STAT samples entered but not started within a specified
time period.
2 - 48
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
Additional function keys on the STAT Samples screen can save time by
displaying screens to research and resolve errors. To use the following function
keys, first move the cursor to an individual sample.
Table 2-25: STAT Sample Fields
F1: TEST RESULTS Displays the Test Results screen. For additional information, see
"Displaying Test Results" later in this section.
F2: EDIT/RERUN Displays the Enter Sample Data screen. For more information, see
"Editing and Rerunning a Sample" later in this section.
F5: CLEAR ALERT Removes the yellow or red alert status from the selected sample and
the stat key.
F8: COMPLETE ALERT When processing is complete for the highlighted sample, the alert is
activated.
Select the Sample Alert key to display the Sample Alert screen. Samples appear
on the screen at the time the initial test is completed and are removed after
the rerun is completed. The display is refreshed every 10 seconds.
While the sample is displayed, select it with the cursor and select F1: TEST
RESULTS to see the detail (see “Displaying Test Results” on page 7 of Chapter
3).
Table 2-27: Sample Results Field Terms
Field Information
Meth The assay that is being rerun or reflexed for the sample.
2 - 49
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
2 - 50
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
Use the Segment Status - All Segments screen to view the processing status of
segment positions for all segments. From the Segment Status - On Board
Segments screen, select F1: SEE ALL to view this screen. Use this screen when
pre-entering samples into segments that are not currently on the instrument,
or when the availability of segment positions is unknown.
Table 2-28: Letters and Colors Used for Sample Status
N No Data Red
E Entered White
R Ready Green
B Begun Yellow
P Printed Blue
2 - 51
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
P Processed Assay complete, vessel has been used and will be discarded.
The dot on the screen above status position 1 and the line above position 23
correspond to the white alignment dot and the white alignment mark on the
incubate wheel itself. If there is a vessel listed on this screen that needs to be
located, use these reference points and refer to the illustration below.
2 - 52
Dimension EXL with LM/EXL 200 System Operator’s Guide System Operation
# Description
2 - 53
System Operation Dimension EXL with LM/EXL 200 System Operator’s Guide
2 - 54
Dimension EXL with LM/EXL 200 System Operator’s Guide Sample Reports
3-1
Sample Reports Dimension EXL with LM/EXL 200 System Operator’s Guide
HIL Index
The HIL index alerts for potential interference from hemolysis, icterus, and
lipemia in a sample, where:
• H = hemoglobin resulting from lysis of red blood cells
• I = icterus resulting from endogenous bilirubin
• L = lipemia or turbidity caused by insoluble lipids
3-2
Dimension EXL with LM/EXL 200 System Operator’s Guide Sample Reports
When the HIL feature has been programmed (See “HIL Feature” on page 44 of
Chapter 8: Configuration), the instrument automatically pipets 20 µL of
sample into an empty cuvette along with system water. Spectral absorbance
measurements are used to generate a sample-specific HIL index. The HIL index
displays on the report slip as a three-digit value where:
• 1st digit = H index
• 2nd digit = I index
• 3rd digit = L index
Each index value correlates to an approximate concentration for each of the
potential interferents, as specified in this table:
Table 3-1: Concentration of Potential Interferents
H I L
Hemoglobin Bilirubin Intralipid ™
Index (mg/dL) (mg/dL) (mg/dL)
1 H ≤ 25 I ≤ 2 L ≤ 25
If the following conditions are met, the test report displays the “HIL Interf”
message next to the affected test result:
• The HIL Operating Mode has been set to an “AUTO-ON” mode.
• A method with entered Alert Index values between 2 and 6 has been
processed on the sample.
• Any of the measured HIL index values (H, I or L) is greater than or equal to
the corresponding Alert Index value entered for that method.
If the “HIL Interf” message is displayed, follow laboratory procedures for
reporting results when the sample is hemolyzed, icteric and/or lipemic.
If the maximum number of tests (36) has been ordered for a sample, HIL is not
processed.
3-3
Sample Reports Dimension EXL with LM/EXL 200 System Operator’s Guide
Clot Check
A test result with a "ClotCheckErr" report message will not include a numerical
result. Any successfully completed test results from the same sample cup will
be displayed. Any tests that have not yet aspirated sample from the same
sample cup will be aborted, and have an "Aborted Test" message associated
with them.
Figure 3-1: Test Report
3-4
Dimension EXL with LM/EXL 200 System Operator’s Guide Sample Reports
3-5
Sample Reports Dimension EXL with LM/EXL 200 System Operator’s Guide
The order for listing test results is configurable. Test results can be printed and
displayed on the screen in the order they were processed or a customized
order. See “Selecting a Test Result Order” on page 75 of Chapter 8:
Configuration.
3-6
Dimension EXL with LM/EXL 200 System Operator’s Guide Sample Reports
The top header of the system printer’s test report can be customized with a
laboratory name. See “Configuring the Printer” on page 51 of Chapter 8:
Configuration.
Table 3-2: Test Report
Test Description
HIL Indicates the three indices for HIL measurement in the sample.
CA This indicates that two CA tests were requested and run on this sample.
CREA These lowercase letters indicate that this method’s standard sample volume
was changed on the Method Parameters screen or the method was correlated.
See “Entering Method Parameters” on page 33 of Chapter 8: Configuration.
MG The result exceeded the assay range of the method. Although a result appears
with an "HI" test report message, it cannot be reported because of the "assay
range" non-reportable test report message.
VANC The result was above the assay range of the method and could not be
calculated. No result is reported with the "above asy rng" non-reportable test
report message.
CA The last line was automatically calculated by the instrument because more
than one CA test was run on the sample.
3-7
Sample Reports Dimension EXL with LM/EXL 200 System Operator’s Guide
3-8
Dimension EXL with LM/EXL 200 System Operator’s Guide Sample Reports
Search Examples
Table 3-3: Search Examples
Search Result
*Smith Samples that end with Smith. It would find: John Smith, 123-
Smith.
It would not find: John Smith-123, Smith-Jones.
123**9 Numbers that begin with 123, end with 9, and have two characters
between the 3 and 9.
It would find: 123649, 123AZ9, 123GB9, 1235W9.
3-9
Sample Reports Dimension EXL with LM/EXL 200 System Operator’s Guide
Screen Action
Load List Press Alt/L to display the Load List and select F2: NEXT STATUS to
access all. If the sample is still on the Load List, move the cursor to
the sample and select F8: EDIT SAMPLE.
Sample Status 1. Select F2: NEXT STATUS until the status category in the upper
right corner of the display changes to the status category of the
sample, if known or All.
2. Find the sample in this list by either moving the cursor to the
sample using the arrow keys or by selecting F3: SEARCH and
entering the required search pattern
2. The Enter Sample Data screen or the Add Tests screen will display.
Change or add information.
3. After editing the information:
a. Select F2: PROCESS SINGLE or select RUN to begin processing the
sample.
b. Select F3: LOAD LIST to enter the sample into the Load List.
c. Select F4: RUN or select RUN to begin processing all samples
shown on the Load List.
4. Select F1: NEW SAMPLE to enter a new sample.
Note: If the status of the sample to be edited is Begun, the Add
Tests screen appears and changing the sample’s priority or
adding new tests are the only available options. The sample’s
container cannot be changed until it finishes processing.
3 - 10
Dimension EXL with LM/EXL 200 System Operator’s Guide Sample Reports
Segment Selection
* Any segment currently appearing in the segment status box that has
samples with errors on it.
3 - 11
Sample Reports Dimension EXL with LM/EXL 200 System Operator’s Guide
Short Sample Error and Clot Check Error on load "Short Sample"
list
No Short Sample Error and a Clot Check Error on "Clot Check Error"
load list
3 - 12
Dimension EXL with LM/EXL 200 System Operator’s Guide Sample Reports
Sample Troubleshooting
Resolving a Short Sample Detected
When the short sample icon displays, the instrument sounds the alarm and
displays the error message “No Sample Found During Level Sense”. This
indicates that there is not enough sample in the sample container to perform
all the assays requested. Choose from the following options:
• Add more of the same sample to the sample container.
• Reduce the number of assays on the sample using F8: EDIT SAMPLE.
• Transfer the sample into a different sample container using the procedures
below.
1. Press ALT+L to display the Load List screen.
2. View the list of samples that do not have enough fluid to perform the
requested assays.
Sample container letter designations on the Load List screen are:
a. t = sample tube
b. c = sample cup
c. x = SSC
d. p = pediatric tube
e. l = limited cup - no level sense
Table 3-7: Sample Container Procedures
Tube (with bar code) Short Sample – Primary Tube with a barcode
Tube (no bar code) Short Sample – Primary Tube without a barcode
3 - 13
Sample Reports Dimension EXL with LM/EXL 200 System Operator’s Guide
WARNING: Failure to place the sample cup into the same segment
position that the sample tube was in will cause erroneous
patient identification and erroneous results.
4. Select PAUSE.
5. From the Load List short sample screen, move the cursor to the short
sample.
6. Select F8: EDIT SAMPLE to display the sample’s Enter Sample Data
screen.
7. Select F7: NEXT MODE until the Mode field changes to "sample cup".
8. Select F2: PROCESS SINGLE.
3 - 14
Dimension EXL with LM/EXL 200 System Operator’s Guide Sample Reports
WARNING: Failure to place the sample cup into the same segment
position that the sample tube was in will cause erroneous
patient identification and erroneous results.
4. Select PAUSE.
5. Move the cursor to the short sample.
6. Select F8: EDIT SAMPLE to go to the sample’s Enter Sample Data screen.
7. Select F7: NEXT MODE until the Mode field changes to "sample cup".
8. Select F2: PROCESS SINGLE.
3 - 15
Sample Reports Dimension EXL with LM/EXL 200 System Operator’s Guide
WARNING: Do not report an assay result that appears on the printed test
report with a reference range indicator if it also appears with
a test report message indicating that the result is NON-
REPORTABLE.
These reference range indicators appear next to a result according to the
ranges that are programmed for that assay in the Method Parameters screen.
See “Entering User-Defined Method Parameters” on page 90 of Chapter 8:
Configuration. Remember to follow laboratory procedures for lp and hp
indicators.
3 - 16
Dimension EXL with LM/EXL 200 System Operator’s Guide Sample Reports
3 - 17
Sample Reports Dimension EXL with LM/EXL 200 System Operator’s Guide
What to Do:
For non-HM assays:
This result cannot be reported. Align the sample and reagent probes, and then
rerun the sample.
For HM assays:
This result cannot be reported. If the sample is fibrinated, centrifuge the
sample and rerun. If the sample is not fibrinated or the error persists, contact
the Customer Care Center - Technical Solution: Inside U.S.: 1-800-441-9250;
Outside U.S. Refer to the local support team.
For LOCI assays:
This result cannot be reported:
If multiple samples are sequentially reporting with this flag, do not report any
results and please call Customer Care Center - Technical Solution as the results
require further investigation by Siemens personnel.
If only one flagged result is seen:
Verify the system operation by repeating the sample with fresh Quality Control
samples.
If the abnormal reaction on the sample persists and the Quality Control is
within expected limits, dilute the sample 1:2 with a patient/QC sample with a
known value as described in the section “Mixture of a Sample and a Known
Standard” .
If the abnormal reaction flag persists, the result should not be reported and the
specimen should be analyzed by an alternate method. Please contact the
Customer Care Center - Technical Solution in the event this occurs.
Aborted Test
Explanation:
An action by either the operator or instrument aborted this assay.
What to Do:
Rerun the assay.
Absorbance
Explanation:
The MAU at the measurement wavelength exceeded 2000.
What to Do:
Rerun the sample. If the error occurs on the rerun, it may be due to optical
interference.
3 - 18
Dimension EXL with LM/EXL 200 System Operator’s Guide Sample Reports
Arithmetic
Explanation:
This error is associated with nonlinear (logit) assays only. This error occurred in
result calculations because the change in absorbance was less than the C0 or
greater than C0 + C1.
What to Do:
If on a patient sample, the concentration is either very high or very low. Rerun
the sample or contact the Customer Care Center - Technical Solution: Inside
U.S.:
1-800-441-9250; Outside U.S. Refer to the local support team. If QC has also
shifted low, change the R2 reagent pump 2500 µL syringe, cancel the well in
use and prepare new reagent.
3 - 19
Sample Reports Dimension EXL with LM/EXL 200 System Operator’s Guide
Assay Range
Explanation:
The result is above or below the assay range listed in Method Parameters on
the instrument. Can be either low or high.
What to Do:
If the result is LOW, analyte values may be depressed below the assay range
because the patient sample has very little or no concentration of the analyte,
insufficient sample, is falsely depressed because of interfering substances, or
there was an instrument system failure. Each laboratory should establish its
own protocol for addressing the "Below Assay" test report message before
reporting the result as less than the established clinical reportable range. The
procedure should check:
• The sample container has sufficient usable sample for the assays ordered.
• The sample container was placed in the appropriate segment and cup
position.
• Verify instrument function.
If no obvious reason for the low analyte result is found, or the result is
inconsistent with the available clinical information and prior assays, confirm
the accuracy of the result with additional testing such as:
• If the result is a negative number, perform a 50% recovery of a known
standard or QC material to confirm that there was no activity in the sample,
that there was enough sample volume used by the instrument, and that
there was no system malfunction. See “50% Recovery of a Known Standard
Using a Sample” on page 28 of this Chapter.
• If the result is below the reportable assay range, prepare and run a mixture
of the sample with a known standard or QC material to confirm that there
was low activity in the sample below the assay range, that there was
enough sample volume used by the instrument, and that there was no
system malfunction. See “50% Recovery of a Known Standard Using a
Sample” on page 28 of this Chapter.
If the result is HIGH, check the assay IFU to determine if the sample can be
diluted. If yes, manually dilute the sample (see IFU for recommended diluent)
and rerun the test. Make the smallest dilution possible to bring the result down
into the assay range. Enter the dilution factor as a whole number on the Enter
Sample Data screen and the instrument will multiply the result by this factor.
See “Dilution of a Sample” on page 27 in this Chapter.
3 - 20
Dimension EXL with LM/EXL 200 System Operator’s Guide Sample Reports
3 - 21
Sample Reports Dimension EXL with LM/EXL 200 System Operator’s Guide
Diluted
Explanation:
The test has been autodiluted.
What to Do:
If the assay result is printed, the assay range was not exceeded and the result
may be reported.
If no assay result is printed, the result did exceed the assay range. Do the
following:
• Ensure that sufficient sample volume was available in the sample
container.
• Verify the sample quality (fibrin, air bubbles, etc.).
• Rerun the test.
3 - 22
Dimension EXL with LM/EXL 200 System Operator’s Guide Sample Reports
HI
Explanation:
Result is higher than the reference interval.
What to Do:
Result may be reported.
HIL Interf
Explanation:
One or more of the HIL indexes for the assay is equal to or greater than the HIL
Alert Index entered during HIL Setup.
What to Do:
Follow laboratory procedures for reporting results when the sample is
hemolyzed, icteric, and/or lipemic.
3 - 23
Sample Reports Dimension EXL with LM/EXL 200 System Operator’s Guide
hp
Explanation:
Result is higher than the panic interval.
What to Do:
Follow laboratory’procedure for panic values that are out of range.
LO
Explanation:
Result is lower than the reference interval.
What to Do:
Result may be reported.
lp
Explanation:
Result is lower than the panic interval.
What to Do:
Follow laboratory procedure for panic values that are out of range.
Measurement
Explanation:
During photometric measurement, the system detected an insignificant
software timing error.
What to Do:
Rerun sample.
• If this error recurs for a photometric assay, a problem may exist in the
instrument measurement system. Call the Technical Solutions Center.
• If this error recurs for an IMT assay, troubleshoot the IMT error message.
3 - 24
Dimension EXL with LM/EXL 200 System Operator’s Guide Sample Reports
No Reagent
Explanation:
There is no reagent cartridge on the reagent tray for this assay.
What to Do:
Load a new reagent cartridge into the instrument. If a reagent cartridge was
previously loaded but did not obtain any results, check for and troubleshoot
any reagent preparation errors listed on the Error Log screen.
Process Error
Explanation:
An error occurred that prevented the system from determining the result.
What to Do:
If this error message appears in the message field on the screen, press Alt/M
and follow the steps that appear. If it does not appear in the message field, go
to the Error Log screen. Move the cursor to the error, press F5: MORE INFO,
and follow the steps to resolve it. To navigate to the Error Log screen from the
Operating Menu, press F5: PROCESS CTRL > F6: ERROR LOG.
3 - 25
Sample Reports Dimension EXL with LM/EXL 200 System Operator’s Guide
Temperature
Explanation:
This message appears when one of the following modules is outside of
temperature range: cuvette, HM incubation wheel, and LOCI reader module.
What to Do:
Check the operating conditions status area for a temperature icon.
If a cuvette temperature icon is present in the status area, calibrate the cuvette
system temperature. If an HM temperature icon is present in the status area,
calibrate the HM temperature. If a LOCI temperature icon is present, call the
Customer Care Center - Technical Solution.
3 - 26
Dimension EXL with LM/EXL 200 System Operator’s Guide Sample Reports
Dilution Examples
This section contains three examples of dilutions and how they are used to
resolve samples with test report messages.
Dilution of a Sample
When a result exceeds the assay range for an assay, the Dimension EXL system
will automatically dilute the sample and rerun the assay.
If the test result still exceeds the assay range the printed test report will
contain the message “assy rng/dilu” indicating that the instrument diluted the
sample and the result is still above the assay range for the assay.
Make a manual dilution of the sample and rerun it. If the sample can be
diluted, make the smallest dilution possible to bring the result down into the
assay range. Make a dilution slightly greater than what was done by the
instrument.
To run this diluted sample, enter the number 5 in the Dilution field on the
Enter Sample Data screen and the instrument will multiply the result by this
factor.
3 - 27
Sample Reports Dimension EXL with LM/EXL 200 System Operator’s Guide
Since this calculated result matches the original sample result of 0.5 g/dL
report a result of “less than 2.0 g/dL” for the original sample.
3 - 28
Dimension EXL with LM/EXL 200 System Operator’s Guide Sample Reports
Process Error
Aborted Test
No Reagent
No Calib’n
Arithmetic
Abnl Assay
Subst Dplet’n
To Antigen Excess
Absorbance
Hemoglobin
Measurement
Assay Rng/Dilu
Assay Range
HIL Interf
Calib’n Exp’d
Diluted
3 - 29
Sample Reports Dimension EXL with LM/EXL 200 System Operator’s Guide
3 - 30
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
4-1
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
4-2
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
Calibrating HA1C
To calibrate the HA1C assay, refer to the “Calibrating the Hemoglobin A1c
(HA1C) Assay” procedure in the Hemoglobin A1c (HA1C) Kit Supplement.
Calibrating HB1C
To calibrate the HB1C assay, refer to the Hemoglobin A1c (HB1C) Kit
Supplement.
4-3
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
Calibration Alert
The Calib Alert key can be configured to change color when a calibrated lot
approaches or reaches its expiration date and when a Flex cartridge is in
inventory but not calibrated. The threshold used (1 to 240 hours before
expiration) is operator-assigned and is the same for each method. See
“Configuring Calibration Alerts” in Chapter 8.
1. When the Calib Alert key is yellow, select it to display the Calibration
Alert screen.
Table 4-1: Calibration Alert Fields
Field Information
2. From the Calibration Alert screen, the operator can perform several tasks
by selecting the appropriate function key.
Table 4-2: Function Keys
Key Task
F1: SET UP & RUN Highlight a method and press this key to display the Calibration
Setup screen.
(see “Setting Up and Running a Manual Calibration” section later in
this chapter).
F2: CALIB REVIEW Highlight a method and press this key to display the Calibration
Review screen when the alert indicates auto-accept.
F3: GROUP CALS Press this key to display assays with calibrator alerts grouped by
Calibration Product.
See the “Group Calibration Alerts and Run Calibration Group” section
later in this chapter.
F4: CONFIG ALERTS Press this key to display the Configure Calibration Alerts screen.
See the“Group Calibration Alerts and Run Calibration Group” section
later in this chapter
F5: DEF CAL PRODCT Press this key to enter a calibration product.
4-4
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
4-5
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
Field Information
Lot If this is not the correct reagent lot number, select F1: OTHER LOT.
Calibrator Lot QCC PowerPak: Lot number is automatically entered from previously defined
calibrator products.
Manual: Enter the calibrator lot number. The system uses the same calibrator
cup for assays which are in the Calibrator Product/Lot field. In this case, enter
the calibration lot number in only one set of cups for each calibration setup.
Start at Position Enter the segment position where the first cup will be placed. Do not use SSC
segments for calibrations.
Bottle Values QCC PowerPak: Bottle values are automatically entered from previously defined
calibrator products.
Manual: Enter the bottle values from the calibrator product IFU in ascending
order (lowest to highest).
Note: The status field for an assay lot can either Never Calibrated,
In Progress, Not Accepted, Calibrated, or Expired. A
calibration cannot be set up for an assay if its status is “Not
Accepted”. Accept or Reject the calibration before proceeding.
6. Select F8: QC YES/NO to change the QC Levels fields to Yes.
4-6
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
4-7
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
Linear Logit
Intercept (b) 0.0 or clinically insignificant Intercept (b): 0.0 or clinically insignificant
Correlation Coefficient (r) 0.990 - 1.000 Correlation Coefficient (r): 0.990 - 1.000
Cancelling a Calibration
A calibration can be cancelled only if it has been scheduled. After cancellation,
the Calibration Set Up screen status field changes to Not Accepted for that lot.
The lot will then need to be rejected and recalibrated.
1. From the Calibration Set Up screen, select F4: CANCEL CALIB.
2. Select EXIT.
3. Select F3: REVIEW DATA.
4. Select the test key for the assay being cancelled.
5. Select F8: REJECT DATA and press Y.
4-8
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
Key Function
F4: PLOT PRINT Presents the information in a plot format that can be printed.
F7: SEARCH CALIB Displays calibration history for a Calibration ID that was specified.
4-9
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
Restoring a Calibration
1. From the Home screen, select F5: PROCESS CTRL > F1: CALIBRATION.
2. Enter a password and press ENTER.
3. Select F7: CALIB HISTORY.
4. Select an assay (method).
5. Use F1: <== PREV CAL or F2: NEXT CAL ==> to scroll through all
available archived calibrations for that assay or search by calibration id
and date until the desired record is selected.
Note: The Dimension instrument can only store the 70 most recent
calibrations run per assay. Any additional calibrations will
overwrite existing assay records in the system.
6. Select F5: RESTORE CAL.
7. Press ENTER to confirm the calibration.
Note: If Exit or any other key is entered, the restore calibration is
canceled.
8. Enter initials and press ENTER. A calibration restore record is created.
Note: The calibration will not be restored if operator initials are not
entered.
When a calibration is successfully restored, the calibration data is loaded as the
active instrument calibration for that lot. The original calibration date (not the
restoration date) remains in effect. The restoration date and the initials
entered when the calibration was restored are preserved.
A new calibration history record (file) is created for the assay. Restored
calibration history records are grouped separately from standard calibrations.
They can be viewed by selecting F8: VIEW RESTORED. Restored calibrations
display the date of restoration and the user’s initials.
4 - 10
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
Calibration Troubleshooting
This topic describes troubleshooting for a calibration that has unacceptable
precision, calibration statistics or quality control.
4 - 11
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
4 - 12
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
Field Information
Lot If this is not the correct reagent lot number, press F1: OTHER LOT.
Status The status of this lot for the method. A verification for this lot cannot be set up
if its status is “Not Accepted”.
Calibrator Lot QCC PowerPak: Lot number is automatically entered from previously defined
verifier products.
Bottle Values QCC PowerPak: Bottle values are automatically entered from previously defined
verifier products.
Manual: Enter the bottle values from the verifier product IFU in ascending
order (lowest to highest).
4 - 13
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
4 - 14
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
• If the verification statistics are acceptable, select F2: ACCEPT DATA (the
Status field on the screen changes to CALIBRATED).
• If the verification statistics are not acceptable, reject this verification by
selecting F8: REJECT DATA. See “Verification Troubleshooting” on page 16
of this chapter to resolve the problem and rerun the verification.
Cancelling a Verification
A verification can be cancelled only if it has been scheduled. After cancelling
and rejecting a verification, the Calibration Set Up screen Status field changes
to Rejected for that lot.
1. From the Calibration Set Up screen, select F4: CANCEL CALIB.
2. Select EXIT.
3. Select F3: REVIEW DATA.
4. Press the method test key for the assay that was cancelled (be sure to
check the Lot number).
5. Select F8: REJECT DATA and enter Y.
4 - 15
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
Verification Troubleshooting
This topic describes troubleshooting for a verification that has unacceptable
precision, calibration statistics or quality control.
4 - 16
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
4 - 17
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
Electrolyte Coefficients
The following are the default coefficients used by the software when urine is
selected as the fluid type:
Table 4-8: Urine Default Coefficients
C0 C1
Na+ 0 1.0
K+ 0 1.0
Cl- 0 1.06
Indirect (diluted) C0 C1
K+ –0.2 1.05
4 - 18
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
List Contents
Recommended Shows a list of reagent lots that are within X hours of their expiration date.
Lots
Same Product Shows all reagent lots that use the same calibrator/verifier product.
All Lots Shows all reagent lots that are in system memory.
Selected Lots Shows all the selected reagent lots to be calibrated/verified from the three
status list screens listed above.
4 - 19
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
4 - 20
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
4 - 21
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
4 - 22
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
Quality Control
Quality control (QC) is a system of checks to ensure that the system and its
associated materials meet with the laboratory’s acceptable standards of
performance.
The laboratory is responsible for the proper functioning of the Dimension EXL
system after installation. Siemens assists in setting up the system and ensures
that the assays to be used initially are calibrated and verified.
Daily QC
Check the Dimension EXL system daily for proper performance by using one of
the System Check procedures found in Chapter 2: System Operation, and by
performing routine maintenance as described in Chapter 5: Maintenance.
Each day run two levels of quality control material of known activity or
concentration for each assay that will run during that day. If the results are
outside laboratory acceptable limits, investigate the cause immediately.
4 - 23
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
QC Materials
The quality control material used should not be the same material that was
used to calibrate or verify an assay. This could cause an undetected error and
lead to inaccurate results.
Carefully follow the procedures provided by the manufacturer of the quality
control material for handling and reconstruction.
It is best to use a quality control material with a concentration (or activity) at a
clinical decision-making level (as used by the laboratory).
Processing QC Samples
When processing a QC sample for an assay, the system runs a test using each
Flex reagent cartridge lot that is on the system for that assay. All QC samples
are processed like patient samples.
When the test results are complete, the QC status becomes valid and the QC
expiration time period begins. If another QC sample isn’t processed before the
QC expiration time period is up, the QC status for that method expires.
The QC expiration time period can be determined by using the Quality Control
Status List screen.
Crossover QC
Before using a new lot of QC material, you will want to perform crossover QC
tests on it. If you give a sample of the new lot of QC material a crossover QC
priority (XQC) on the Enter Sample Data screen, the system provides results for
the sample to use in calculating your new QC ranges for the new lot. However,
since the system does not compare these crossover QC results to the QC
ranges of the current lot number of QC material, these results do not affect the
QC status of the method. Crossover QC results are stored in the Method Review
files.
4 - 24
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
WARNING: For urine Drugs of Abuse assays, negative and positive quality
control samples must be run using the fluid types SerumQC1,
SerumQC2, UrineQC1, or UrineQC2. Processing QC samples as
SerumQC3 overrides the current calibration and could result
in an erroneous calibration of these methods.
Note: When processing QC Samples for Urine Drugs of Abuse
Assays, follow the QC procedure in the Urine Drugs of Abuse
Supplement.
4 - 25
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
Result Explanation
Error If an error occurs during QC processing which suppresses the result, the
assay is flagged on this screen.
High The QC result is above the upper limit of the stored QC range.
Low The QC result is below the lower limit of the stored QC range.
Table 4-12 explains the tasks that can be performed with the QC Alert function
keys:
Table 4-12: QC Alert Function Keys
Key Task
F1: TEST RESULTS To view QC results, highlight the assay and select this key. See “Editing and
Rerunning a Sample” on page 10 of Chapter 3: Sample Reports.
F2: EDIT/RERUN To rerun the QC, highlight the assay and select this key. See “Editing and
Rerunning a Sample” on page 10 of Chapter 3: Sample Reports.
F3: CLEAR ALERT To clear an alert (change key color from yellow to gray) without rerunning
the QC, highlight the assay and select this key.
F4: CONFIG ALERT To set parameters for QC alerts, select this key. See “Configuring QC Alerts”
on page 50 of Chapter 8: Configuration.
F5: DEFINE QC PRODUCT To display the Define QC Product screen, select this key. See “Defining QC
Products” on page 28 of this chapter.
F6: GROUP ALERTS To group alerts with the same product definition, select this key. See
“Grouping QC Alerts” on page 31 of this chapter.
F7: METHOD REVIEW To review the QC levels and applicable Westgard rule, highlight the assay
and select this key. See “Reviewing QC Results” on page 42 of Chapter 4:
Calibration and Quality Control.
F8: PRINT To print the information on the screen, select this key.
4 - 26
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
Entering QC Ranges
1. Select F5: PROCESS CTRL > F4: QC RANGES.
2. Select the test key for the assay that will be edited.
3. Select the Lytes test key, then select F1: NEXT METHOD to see the K and
Cl assays.
4. For each QC level, enter the QC range (low and high), expected mean, or
expected SD as appropriate. The number of QC levels used determines
how many days of QC data can be stored in the software for an assay.
Note: For the auto-acceptance option enter a minimum of two QC
ranges.
Table 4-13: QC Levels
QC Levels Set for Each Assay Days of QC Results Stored for Each Assay
2 270 Days
3 180 Days
4 135 Days
5 108 Days
4 - 27
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
Checking QC Status
An asterisk (*) to the left of the QC Expires column indicates that the QC for
that reagent lot has expired.
Defining QC Products
The Dimension EXL system will display up to 100 QC products. QC panels must
be defined using a previously defined QC product.
1. Select the QC Alert key.
2. Select F5: DEFINE QC PRODUCT.
3. Type the QC product name and select ENTER.
4. Type the product level and select ENTER.
5. Type the QC Lot number and select ENTER.
6. Use F3: NEXT QC FLD LEV to select a QC Level.
7. Use F4: NEXT FLUID to select the fluid appropriate for the product.
8. Change the Active QC and Cal fields with F5 and F6 if this product will be
active for daily QC and included with current calibrations and select
ENTER.
Note: For Auto Urine Dilute assays (BUN, CREA, CRE2, EZCR, PHOS,
URCA) toggle the F6 key to make the calibration fields
inactive.
9. Use the test keys to enter assays associated with the product.
10. Select F7: STORE.
a. To enter additional levels of the product, change the product level,
QC fluid level and select F7: STORE.
b. To define another QC product, select F1: NEW PRODUCT.
c. To view a list of defined products, display the Edit QC Products
screen.
4 - 28
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
Editing QC Products
1. Select: F5: PROCESS CTRL > F3: QC STATUS > F2: EDIT QC PRODUCT to
display the Edit QC Product screen.
2. Highlight the product to edit and use the appropriate function keys to
make changes.
Table 4-14: Edit QC Product Function Keys
Key Function
F2: DELETE PRODUCT Removes the highlighted product from the list.
F3: SORT BY PROD/QC FLD/FLUID Toggles among choices for sorting the QC products list by product
name, QC fluid level, and fluid.
F7: SET CAL ACTIVE/SET CAL Toggles between activating and deactivating calibration.
INACTV
Defining QC Panels
Use this screen to create up to 50 different QC panel definitions with up to
36 methods in each. Panels are designed to be run in SSC (Small Sample
Container) placed on barcoded tubes.
When an SSC with a panel ID is processed, the system completes the following
actions:
• Queries the LIS, if configured for the ID.
• Searches the QC panel database for a match.
1. Select F5: PROCESS CTRL > F3: QC STATUS > F6: DEFINE QC PANELS.
2. Select F1: NEW PANEL to define the new QC Panel.
4 - 29
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
Editing QC Panels
Use the Edit QC Panel screen to view a list of defined QC panels and to edit the
panel.
1. Select F5: PROCESS CTRL > F3: QC STATUS > F7: EDIT QC PANEL.
Table 4-15: Edit QC Panels Function Keys
Key Function
F2: DELETE PANEL Removes the highlighted panel from the list.
2. Highlight the panel to be edited and select F1: EDIT PANEL to make
changes.
3. Move the cursor to the field to be changed.
4. Type in the new QC product name or QC lot.
5. Use test keys to add tests or select F2: DELETE TEST to remove tests.
Note: If adding tests to a panel always add to the product first.
6. Use the function keys to change Sample ID, QC Fluid Level Types and
Priority. Priority options include QC Active (Quality Control) or XQC
(Crossover Quality Control).
7. Select F7: STORE.
4 - 30
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
Grouping QC Alerts
Use the QC Groups screen to combine assays with the same assigned QC
product and active QC alert into one sample ID and SSC. Using this feature
allows QC to be run on a maximum number of methods in a minimum number
of sample cups or SSCs in barcoded tubes.
1. Select the QC Alert key.
2. Select F6: GROUP ALERTS.
3. Move the cursor to a product.
4. Select F1: ASSIGN SAMPLE ID. Enter the ID by scanning the barcode or
by typing it.
5. Select ENTER.
6. Repeat Step 3 and Step 4 for additional products.
7. To remove assays that will not be processed, highlight the assay and
select F2: DELETE METHOD.
8. Select F5: Load One to select a sample for processing, or select F6: LOAD
ALL to process all samples listed.
Note: If this is a QC Rerun, move the sample to a new location.
9. To remove assays that will not be processed, highlight the assay and
select F2: DELETE METHOD.
10. Select F8: PRINT to create a report of the QC group.
4 - 31
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
(Specific Lot) Select F1: SET PERIOD/LOT and type an asterisk (*) followed by the lot ID
number and select ENTER. To return to viewing results for all lots of the
method, select F1: SET PERIOD/LOT, type an asterisk (*) and select ENTER.
4 - 32
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
Field Meaning
Result Test result (units are those appearing in the Units field).
Lot The lot ID of the Flex® reagent cartridge used to process the assay.
4 - 33
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
When a QC fluid is selected, the > and < signs indicate that the data point
exceeds the rule (>) or is lower than the rule (<) shown in the Rule field.
Table 4-18: Fluid Type Chart & Answer Message
Answer
Message
If the Fluid Type Is To Print Out
1 2
Patient The summary information (the top half of the screen). The Y N
summary information and supporting data for the assay and
fluid combination. N
Histogram Plot
To create a histogram plot, five results are required for the assay/fluid
combination. Each result must fall within the time period and the specific
reagent cartridge lot that appear on the top line of the screen.
Select: F3: SEE HISTOGRAM.
The X-axis shows the test result. The Y-axis shows the number of tests having that
test result.
For better viewing, use the left and right arrow keys on the keyboard to
change the scale of the X-axis as appropriate. The left arrow key expands the
range of the X-axis; the right arrow key compresses the range of the X-axis.
Printing Histogram Plot Information
1. To print only the data listed at the top of the view, select F8: PRINT.
2. To print the entire screen, select ALT+P.
Note: Both of these printouts can only be printed by the system
printer; they will not print to an external printer.
4 - 34
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
4 - 35
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
4 - 36
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
Field Meaning
Fluid The fluid for the patient or QC results shown on the screen. Select F5: QC/PATIENT to go
from patient to QC results and vice versa. Select F6: NEXT FLUID to display results for
another fluid.
[dates] Time period brackets. If there is nothing entered in the brackets, all results in the test result
buffer for the assay and fluid combination display on the screen.
If a time period is entered, all screens (patient and QC) will display only results that fall
within this time period. See “Method Review Screen—Function Keys”.
(lot ID) Lot ID or crossover QC parentheses. If there is nothing entered in the parentheses, all results
for all reagent cartridge lots in the test result buffer for the method and fluid combination
display. If the parentheses contain the letters XQC, only crossover QC results display.
If a reagent cartridge lot ID or crossover QC is entered, all screens (patient and QC) will
display only those results that use the specific lot ID or are crossover QC results. See
“Method Review Screen—Function Keys”.
Tests The total number of results in the test result buffer that match the assay and fluid on the
screen (and, if specified, are within the specific time frame or are for the specific reagent
cartridge lot ID or crossover QC).
(Tests) The number and percentage of results in the Tests field that are within the Interval field
In shown on the screen.
(Tests) The number and percentage of results in the Tests field that are outside the Interval field
Out shown on the screen.
4 - 37
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
Field Meaning
Interval The interval to which results are compared to determine the numbers and percentages
shown in the In and Out fields.
Units Results units. These units are taken from the Method Parameters screen for the method.
Expecte Value only appears when QC results are selected. The value that appears (if any) was
d Mean entered by the operator on the Quality Control Ranges screen.
Expecte Value only appears when QC results are selected. The value that appears (if any) was
d SD entered by the operator on the Quality Control Ranges screen.
Actual The actual mean is calculated using all the Tests field results.
Mean
Actual The actual standard deviation is calculated using all the Tests field results.
SD
CV % The coefficient of variation percentage is calculated using all the Tests field results.
Rule The rule or constraint against which the results are compared. Each result is compared to
the Interval field to determine whether it exceeds or is below the interval.
Marker: This field contains the result and the date of the result corresponding to the position of the
( ) marker on Levey-Jennings–like plots. See “Levey-Jennings Plots”.
4 - 38
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
F1: SET Used to enter a specific time period in Same as patient results. Also
PERIOD/LOT brackets [ ] and a specific reagent allows crossover QC data to be
cartridge lot ID in the parentheses ( ) selectively displayed inside the
fields in the upper right corner of the parentheses ( ).
screen. See “Method Review Screen —
Fields” on page 37 of this chapter.
F2: DELET RESULT Patient results cannot be deleted. This Deletes the QC result where the
(F2: UN-DELETE) function key is disabled when patient cursor is positioned on the Data
results are displayed. screen or where the marker is
(Function key F2 is positioned on the L-J like QC plot.
always Delete For more information on deleting
unless the cursor and un-deleting results using this
or marker is function key, see “Reviewing QC
positioned on a Results” on page 42 of this chapter.
data point that has
been deleted).
F3: See DATA Used to display the results as a list, a Same as patient results.
(F3: See histogram, or a Levey-Jennings like
HISTOGRAM) plot. These views must be displayed in
(F3: SEE L-J PLOT) sequential order.
F4: SET INTERVAL Used to select results within a specific Same as patient results. Also
user-defined interval/ To enter a new allows a decimal SD to be entered.
range in the Interval field, select F4: Set See “Levey-Jennings Plot Views” on
Interval and at the prompt enter the page 35 of this chapter.
low range for the new interval and then
the high range. Select ENTER after each
entry. The In and Out fields will update
automatically based on the new range.
Note: When Using
F4: SET INTERVAL, all
Method Review Screens
for this assay/fluid
combination will use
this interval.
Note: The Interval field shows
the new interval as the
result histogram and L-J
plot views are displayed
for the assay and fluid
combination. It reverts
to the appropriate
intervals from the
Method Parameters
screen upon exiting and
reentering the Review
screen or when another
patient fluid or assay is
selected.
4 - 39
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
F5: QC/PRINT Used to change displayed results from Same as patient results.
patient to QC and vice versa.
F6: NEXT FLUID Used to display results for the next Used to display results for the next
patient fluid type. QC fluid type.
F7: SHOW mAU Used to toggle result units between Does not appear on the QC display.
SHOW CONC mAUs and concentration units. When
mAU’s are selected, the Interval field
will always be
0.0 - 9999.9 (The mAU information is
typically used for troubleshooting).
F7: NEXT RULE Does not appear on the patient results Used to change the Rule field
display. selection. There are 7 possible rule
selections, 6 of which are
Shewhart Rules. See “Rules Field
Definitions” on page 40 of this
chapter.
F8: PRINT Used to print out results information Same as patient results.
appearing at the top of the screen using
the system printer. (This does not print
to an external printer).
1. Select F8: PRINT.
2. Answer the prompt with a Y to print
out only the information listed at the
top of the screen, or with an N to
print out a report that contains all the
individual results in that method’s
test results buffer on this instrument.
3. Select ENTER.
4 - 40
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Calibration and Quality Control
When a new rule is selected, the Interval, In, and Out fields are updated
automatically to reflect the new rule selected.
Table 4-21: Method Review Screen Function Keys
High/Low Interval Any result that falls outside the range listed on the Quality Control Ranges
screen for that fluid.
any point > 2 sd Any result that falls outside ±2 standard deviations from the mean. This is
termed a “warning” rule and requires further evaluation to judge whether
the data presents a problem.
any point > 3 sd Any result that falls outside ±3 standard deviations from the mean. A
violation of this rule suggests systematic error.
two consecutive > 2 sd Two consecutive results that fall outside ±2 standard deviations from the
mean. A violation of this rule suggests systematic error.
four consecutive > 1 sd Four consecutive results that fall outside ±1 standard deviation from the
mean. A violation of this rule suggests systematic error.
two consecutive > 4 sd The absolute difference between two consecutive results is greater than
or equal to 4 standard deviations. A violation of this rule suggests random
error.
ten consecutive Ten consecutive results above or below the mean. An above/below mean
violation of this rule suggests systematic error.
Note: These rules use the expected mean and SD values, unless
actual mean and SD values are entered. If entered, the actual
values are used to create the plots. These rules consider only
results “within” a given material and not across materials. To
review results across materials, print out a plot of both levels
and then visually examine them.
4 - 41
Calibration and Quality Control Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
Reviewing QC Results
Any deleted results are not included in or used to calculate the values in the
fields at the top of a Method Review QC screen. These results are not
permanently deleted from the QC test result buffer.
Note: Patient results cannot be deleted.
4 - 42
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
Chapter 5: Maintenance
5-1
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
Undiluted Bleach
All references to “undiluted bleach” in this guide refer to a product that meets
the following criteria:
• 5-12% sodium hypochlorite
• No additives
• No detergents
• No surfactants
• No fragrances
• Minimal impurities
Bleach Solutions
Note: All bleach solutions should be prepared fresh prior to use.
All references to a bleach solution, refer to a solution that is made by diluting
an undiluted bleach product that meets the criteria listed above with water.
Decontamination
Decontaminate surfaces in or on the instrument where biohazard material has
spilled.
1. With the system in standby, select PAUSE.
2. Lift the instrument lid.
3. Locate the spill.
4. Decontaminate the area using a 10% bleach solution.
5. Thoroughly dry the decontaminated area using a lint-free cloth.
6. Close the lid, if necessary.
7. Select PAUSE, and then select RESET.
5-2
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
WARNING: Ensure that all lids are fully open to avoid injury while
performing maintenance.
5-3
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
5-4
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
Weekly Maintenance
WARNING: Ensure that all lids are fully open to avoid injury while
performing maintenance.
3. Use a moist cotton swab dipped in water and beginning at the top of the
probe, wipe down the outside of both HM wash probes.
Figure 5-4: Cotton Swab
5-5
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
3. Use a cotton swab dipped in 0.1N sodium hydroxide and scrub the nut at
the top of the probe.
4. Beginning at the top, wipe down the outside of the R2 reagent probe.
5-6
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
5-7
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
Monthly Maintenance
WARNING: Ensure that all lids are fully open to avoid injury while
performing maintenance.
5-8
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
5. Dispense approximately 1mL of water into the Clot Check drain 10 times.
Note: Allow the port to empty itself before each subsequent
addition.
6. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
5-9
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
2 X2 Tubing
3 X Tubing
4 Pressure Plates
Note: Do not use the bottom tube position on the pump. The X
tubing goes around the top position of the pump and
connects to the straight (top) portion of the IMT pump.
8. Close the pressure plates.
5 - 10
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
9. Reinsert the pressure plates foot bar and turn it to lock in the pressure
plates.
10. Select PAUSE to restart the sampling system.
11. Select F1: PRIME SALT BRIDGE solution.
12. Prime with Salt Bridge solution until the solution exits the IMT sensor and
is visible in the X2 tubing.
13. Select HOME on the touchscreen or press SHIFT+ EXIT.
Note: The system performs an IMT pump alignment and IMT
calibration.
14. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
5 - 11
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
CAUTION: Replace the QuikLYTE sensor after bleaching the IMT system.
11. Replace the QuikLYTE Integrated Multisensor. See “Replacing the
QuikLYTE Integrated Multisensor” on page 24 of System Operation.
12. After changing the QuikLYTE sensor, close the IMT lid.
13. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
CAUTION: When installing these filters, make sure that the air flow arrows
on the frame of each filter are pointing as indicated in Table 5-1.
Table 5-1: Replacing Air Filters
5 - 12
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
Cuvette Ring (Non This filter is located in the front right of the instrument panel.
RMS) 1. Open the right instrument cabinet panel.
2. Slide the filter out of it’s holder and replace with the new filter.
5 - 13
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
Thermal Chamber This filter is located in the rear right corner of the instrument.
1. Pull the filter cover and filter off of the plastic frame. This filter does not
have any air flow arrows and cannot be installed backwards.
2. Discard the used filter.
2. Remove all dust from the used filter by washing with water and air drying
or by using a vacuum cleaner.
Note: All filters can be reused with the exception of the thermal
chamber filter, which must be replaced.
3. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screens” in Chapter 2.
5 - 14
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
4. Remove the plastic guard from the wash station by loosening the two
screws on the facing side until the guard can slide out from under the
screws.
5. Pass the IMT stylet through the aspirate probe of each wash station,
pulling the stylet out through the bottom of each probe to dislodge any
obstruction or accumulations.
5 - 15
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
1 Insert Stylet
2 Lip
3 Sleeve
4 Positioning Groove
5 Aspirate Probe
6 Dispense Probe
5 - 16
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
Wash Station The aspirate and dispense pump heads are located on the wash station frame.
# Description
5 - 17
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
CAUTION: The shipping plug (or “T”) must be inserted into the pump head
before installing onto the motor to prevent damage to the pump
rollers as the pump head is pushed onto the motor shaft.
6. Prime the HM wash pump.
a. Select F1: HM WASH PUMP.
b. Close all instrument lids.
c. Perform a System Check.
7. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
5 - 18
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
5. Remove the R1 Reagent Probe from the drain by turning the gold colored
splined shaft of the R1 Reagent arm.
6. Remove the R3 Reagent Probe from the drain by turning the gold colored
splined shaft of the R3 reagent arm.
7. Turn off vacuum to Dimension EXL system.
8. Select F7: DIAGNOSTICS > F1: ELECTRO MECHANICAL >
F6: WATERWORKS.
9. Select F4: VACUUM ON/OFF twice to turn vacuum off.
10. Clamp off non-labeled drain line directly at R1, R2 or R3 drains (if
installed) using hemostats.
11. Place a paper towel under the probe.
5 - 19
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
12. Insert the disposable pipette tip into the drain near to the bottom and
dispense the undiluted bleach into the drain unit until it is completely
filled with liquid.
Note: Sample probe cleaner can be used instead of bleach. Ensure
that a paper towel is around the drain to catch any overflow.
Note: Do not overflow the R1 drain. Significant overflow of liquid
from the R1 drain can destroy the photometer home sensor.
13. Remove the pipette.
14. Let bleach sit in drain for 5 minutes.
15. Un-clamp drain lines.
16. Select F4: VACUUM ON/OFF to turn system vacuum back on.
17. Go to Pump Prime screen and select F1: PRIME WATER (10x).
18. Close all instrument lids.
19. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
5 - 20
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
Other Maintenance
WARNING: Ensure that the lids are fully open to avoid injury while
performing maintenance.
CAUTION: Using a prying motion may cause damage to the release plate.
CAUTION: When lowering the thermal chamber, ensure the waste tubing on
top of the waste bottle is clear before lowering completely.
Figure 5-19: Release Thermal Chamber
5 - 21
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
2. The tape indicators should align indicating that the chamber is closed
completely (See Figure 5-21).
5 - 22
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
5 - 23
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
5 - 24
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
5 - 25
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
6. Close the reagent lid and ensure that all instrument doors and panels are
closed.
Figure 5-27: Digital Thermometer
5 - 26
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
Calibrating HM Temperature
Note: Complete this procedure using a digital thermometer.
Note: The system must be turned on for more than two hours
before a proper temperature calibration can be performed.
1. Select F4: SYSTEM PREP > F5: TEMP MONITOR.
2. Select F3: CHECK HM INCU and type Y in response to the system warm
up prompt.
3. When the message to place the temperature probe appears, raise the IMT
and reagent lids.
4. Place the digital thermometer probe in the temperature monitoring hole
of the HM thermal ring.
Note: This small hole is in the inside rim of the heater ring at the 6
o'clock position. The HM wheel cover has an arrow indicating
the position.
Figure 5-28: Temperature Probe
5 - 27
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
5. Wait three minutes for the temperature to equilibrate and then read the
temperature on the digital thermometer.
a. If the temperature is between 35° C -41°C, continue with Step 6.
b. If the temperature is outside this range, call the Customer Care
Center - Technical Solution: Inside U.S.: 1-800-441-9250; Outside
U.S. Refer to the local support team.
6. Remove the digital thermometer probe.
7. Select F4: CALIB HM INCU.
8. Enter the temperature from the digital thermometer and press ENTER.
9. Run QC.
Note: If the message “The value entered is outside the allowed
calibration range” displays and the temperature entered is
within the 35°C- 41.0°C temperature range, contact the
Customer Care Center: Inside U.S.: 1-800-441-9250; Outside
U.S. Refer to the local support team.
5 - 28
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
5 - 29
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
5 - 30
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
15. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
Note: To configure custom window check limits, use the “Check Bad
Limit” field. Enter the limit for the software to use in
determining whether a window is acceptable to use. Values
in this field can be between 10 and 35. The default value is
15. (The lower the limit, the more the windows will require
cleaning during this check, but in general windows will need
to be cleaned less often). The system indicates windows that
do not meet the defined limit with red “x”s. These windows
will be presented for cleaning.
5 - 31
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
Needlenose Pliers Grasp the top of the cuvette window with the pliers and pull the window
straight up and out of the cuvette wheel.
Cuvette Extractor Tool 1. With the flat side of the extractor facing the cuvette ring, insert the
extractor straight down until it touches the top of the window.
2. Tilt the handle outward to grip the window and hold it in place. Pull the
extractor up to remove the window.
5 - 32
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
5 - 33
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
1 Check Valve
2 Waste Tubing
5 - 34
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
5. Fill a sample cup with water and place the waste tubing in the cup to
flush any residual bleach from the waste line.
6. Connect the waste tubing to the IMT peristaltic pump tubing.
7. Close the IMT lid.
8. Calibrate the IMT.
5 - 35
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
5 - 36
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
4. Raise the barcode reader and unscrew the alignment bar thumbscrew.
5. Raise the alignment bar out of the way.
5 - 37
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
1 Barcode Reader
2 Thumbscrew
6. Unlock the U-seal solenoid assembly by pulling its curved locking bar to
the right.
7. Move the assembly to the left and out of the way.
5 - 38
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
3 U-Seal Solenoid
8. Pull the cuvette formation assembly arm away from the cuvette ring.
9. Remove the used cuvette diaphragm by lifting an edge and pulling it up
off the heat torch.
Figure 5-39: Cuvette Diaphragm
5 - 39
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
CAUTION: Do not pull the soft tip out of position while replacing the
diaphragm.
10. Install the new cuvette diaphragm.
Note: The circular hole should be toward the bottom of the heat
torch with the U-shaped notches facing outward. Ensure the
diaphragm is pushed down completely.
Figure 5-40: New Cuvette Diaphragm
11. Push the cuvette formation assembly back into position next to the
cuvette ring.
12. Put the U-seal solenoid assembly in its original position and lock it in
place with the curved bar.
13. Position the alignment bar and tighten its screw.
14. Lower the barcode reader.
15. Press any key. When the prompt Do you want to store the NEW system
information? (y/n) displays, type Y to update the System Counters
screen.
16. Close the reagent and IMT lids.
17. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
5 - 40
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
4. Slide the used cartridge off the film holder and discard it.
5. Remove the red plastic key from the center of the new cartridge by
pulling up on its ring.
5 - 41
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
Note: Place the extra diaphragms in the Accessory Spare Parts Kit.
6. Slide the new cartridge onto the film holder.
Note: Ensure the new canister is completely seated onto the film
holder.
7. Pull the left film strip out of the cartridge. Overlap it about 1/2” inch over
the left strip of old film hanging from the system and align the sides of
these two film strips.
8. Use the red tape from the new cartridge to tape these strips together.
Repeat the overlapping, aligning and taping with the right strip of film.
Note: If the red tape is damaged or unusable, use transparent tape.
This tape must be applied to both the front and rear of the
film strips to ensure that they do not separate when pulled
through the instrument.
5 - 42
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
1 Film
4 Film
# Description
5 - 43
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
5 - 44
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
1 Barcode Reader
6. Move the assembly to the left away from the cuvette ring.
5 - 45
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
3 U-Seal Solenoid
7. Unlock the U-seal solenoid assembly by pulling its curved locking bar to
the right.
8. Pull the spring toward the form plunger head and separate it from the
E-ring.
9. Move the E-ring and plunger shaft slightly toward the spring (but still
separate from the spring).
10. Use the needlenose pliers to pull out the E-ring.
5 - 46
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
1 Plunger Shaft
2 Cuvette Diaphragm
3 E-ring
11. Remove the cuvette diaphragm by lifting an edge and pulling it up off the
heat torch.
12. Remove the plunger shaft and tip from the form block.
Figure 5-48: Plunger Shaft
13. If only the spring needs replacing, skip to Step 15. Otherwise, remove the
old soft tip from the plunger head.
Note: If the tip is broken, be sure to remove the remaining part of
the tip. Do not scratch the plunger head.
5 - 47
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
2 Rubber Tip
15. Remove the old spring and place a new spring in the middle of the form
block.
16. Install the plunger shaft.
17. Push the spring toward the plunger head and away from the groove on
the shaft.
18. Push the plunger forward slightly to expose the groove.
19. Install the E-ring, making sure not to crush the spring.
20. Replace the cuvette diaphragm. See “Replacing the Cuvette Diaphragm”
on page 37 of this chapter.
21. Reposition the U-seal solenoid near the cuvette ring and push the curved
locking bar into place.
22. Position the alignment bar and tighten its thumbscrew.
23. Lower the barcode reader.
24. Close the instrument lids.
25. Enter 20 and select ENTER.
26. Select F1: CYCLE to make 20 cuvettes.
27. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
5 - 48
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
6. Grip the sensor firmly and pull it straight down off its spring clip and post
and then disconnect P/J 32B.
7. Install the new cuvette ring sensor.
Note: Push the sensor completely up until it snaps into place.
8. Raise the thermal chamber.
9. Restore power to the system following “Restoring Power After a
Controlled Shutdown” on page 10 of Appendix
5 - 49
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
5 - 50
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
1 Hexagonal Standoff
5 - 51
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
Replacing Fuses
Replacing AC Fuses
Note: Complete this procedure using a screwdriver.
The AC fuse panel is located on the left side of the instrument.
1. Power down the system. See “Performing a Controlled Power Shutdown”
on page 10 of Appendix.
2. Lift the printer shelf.
3. Using a screwdriver, turn the fuse holder to the left (counterclockwise) to
release the fuse from the instrument.
Figure 5-54: Fuse Panel
5 - 52
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
Replacing DC Fuses
1. Power down the system. See “Performing a Controlled Power Shutdown”
on page 10 of Appendix.
2. Open the left front cabinet door. Locate the two vertical rows of fuses on
the card cage.
Figure 5-55: DC Fuses
5 - 53
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
1 Barcode Reader
4. Unlock the U-seal solenoid assembly by pulling its curved locking bar to
the right.
5. Move this assembly to the left and out of the way.
5 - 54
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
3 U-Seal Solenoid
6. Pull the cuvette formation assembly away from the cuvette ring.
5 - 55
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
2 Screw
4 P/J13K
5 - 56
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
1 Screw
2 Assembly Removed
13. Install the new heat torch assembly by reversing the previous steps.
Note: The circular holes should face toward the bottom of the
cuvette diaphragm. The u-shaped notches should face
forward. Be sure the diaphragm is pushed down completely.
14. Restore power to the system following “Restoring Power After a
Controlled Shutdown” on page 10 of Appendix.
15. Close all lids.
16. Perform a System Check.
17. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
5 - 57
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
WARNING: All reaction vessels removed from the wash wheel and
incubate wheel are biohazards. Dispose of them according to
your laboratory’s safe biohazard disposal procedures.
5 - 58
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
6. Unscrew the wash wheel locking knob and then use the knob to lift the
wash wheel out of the instrument.
Figure 5-60: HM Tower
# Description
2 Probe Thumbscrew
4 Wash Wheel
5 Incubate Wheel
5 - 59
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
5 - 60
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
5 - 61
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
CAUTION: The shipping plug (or “T”) must be inserted into the pump head
before installing onto the motor to prevent damage to the pump
rollers as the pump head is pushed onto the motor shaft.
6. Prime the fluid for the pump of the pump head that was replaced.
a. Type the number 3 in the CYCLES field.
b. Select ENTER.
c. Select the appropriate function key for the fluid: F1: HM Wash
Pump.
7. Perform a System Check.
8. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
5 - 62
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
Figure 5-64: Wash Station Raise the IMT lid and reagent lids.
Figure 5-65: Pump Panel Open the front cabinet doors and acess the
pump panel.
5 - 63
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
3. Disconnect the top and bottom tubing from the pump head at its
connectors.
Table 5-7: Pump Head/Motor, Connector and Tubing Identification
5 - 64
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
CAUTION: The shipping plug (“T”) must be inserted into the pump head
before installing onto the motor to prevent damage to the pump
rollers as the pump head is pushed onto the motor shaft.
7. Install the new HM pump and motor assembly.
a. Insert and tighten two screws.
b. Connect the motor electrical connector.
c. Connect the tubing to the pump head.
8. Restore power to the system following “Restoring Power After a
Controlled Shutdown” on page 10 of Appendix.
9. Prime the pump on the assembly that was replaced.
a. Select F4 SYSTEM PREP > F7: PUMP PRIME > F6: HM.
b. Type the number 3 in the CYCLES field.
c. Select ENTER.
d. Prime the pump and motor assembly that was replaced by pressing
on one of the following:
• F1: HM WASH PUMP (chemistry wash)
• F2: REAG CLEANER (reagent cleaner)
• F3: SAMP CLEANER (sample cleaner)
Note: If the wash station pump or motor assembly was replaced,
prime the chemistry wash. If the left pump motor assembly
on the pump panel was replaced, prime the reagent cleaner.
If the right pump motor assembly on the pump panel was
replaced, prime the sample cleaner.
10. Perform a System Check.
11. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
5 - 65
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
2 Probe Thumbscrew
4. Loosen the probe thumbscrew until the probe can be lifted up and out of
the wash station.
5. Discard the probe.
5 - 66
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
6. Position the new wash probe in the wash station so that its dispense
probe is toward the front edge of the wash station.
Note: Gently push the probe all the way down until its lip contacts
the probe holder.
Figure 5-68: Probe
# Description
1 Insert Stylet
2 Lip
3 Sleeve
4 Positioning Groove
5 Aspirate Probe
6 Dispense Probe
5 - 67
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
10. Align the wash probe to the wash wheel by completing the following
steps:
a. Select F7: DIAGNOSTICS > F3: ALIGNMENTS > F6: HM MODULE.
b. Move the cursor to the Wash Probes to Wash Wheel field.
c. Select F7: CHECK ALIGN.
11. Close the instrument lids.
12. Perform a System Check.
13. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
5 - 68
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
4. Using a flat blade screwdriver, remove the two screws that hold the
vacuum sensor tower to the station.
5 - 69
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
5. Set the vacuum sensor tower to the side of the wash station as shown by
allowing it to hang.
CAUTION: Use caution when setting the tower aside so as not to disconnect
any tubing or electrical connections.
Figure 5-71: Vacuum Sensor
5 - 70
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
6. Using a 5/64 Allen wrench, remove the screw that holds the wash probe
home sensor in the system.
7. Disconnect electrical connector P/J 45B.
8. Remove the sensor from the instrument.
9. Install the new wash probe home sensor.
10. Reconnect P/J 45B and reinstall the vacuum sensor tower.
11. Restore power to the system following “Restoring Power After a
Controlled Shutdown” on page 10 of Appendix.
12. Align the wash probes to the wash wheel by completing the steps as
found in the procedure“Aligning the Wash Probes to the Wash Wheel” on
page 9 of Chapter 6 Alignments.
13. Perform a System Check.
14. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
Blue 2 (short) Wash Probe 2 aspirate (bent top) probe and vacuum sensor
Green 4 (short) Wash Probe 1 aspirate (bent top) probe and vacuum sensor
5 - 71
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
5 - 72
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
WARNING: All reaction vessels removed from the wash wheel and
incubate wheel are biohazards. Follow laboratory safe
biohazard disposal procedures to discard reaction vessels.
5 - 73
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
5. Refer to the label on the instrument lid to identify where tubing is located
within the IMT system.
Note: The tubing to be replaced is labeled: R1, D1, D2, F1, F2, X2,
X3, W2, and Waste.
WARNING: The X2, X3, W2, and Waste tubing are biohazards. Follow
laboratory safe biohazard waste disposal procedures to
discard this tubing.
6. After replacing all the tubing, return the QuikLYTE multisensor interface
to its locked, upright position.
7. Select PAUSE.
8. Prime the IMT system with Salt Bridge solution.
9. Run a condition/soak cycle.
10. Calibrate the IMT.
11. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
5 - 74
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
IMT Tubing
Note: Remove and replace tubing one at a time. When routing
tubing, place in plastic tubing guides wherever possible.
Table 5-10: IMT Tubing
Tubing Replacement
• 1: R1 attaches here
• 2: X1 attaches here
5 - 75
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
Tubing Replacement
W 1. Using the 3/32 Allen wrench, remove the two screws that hold the IMT port
to the baseplate.
5. Perform Sample Probe to the IMT Port Alignment. See “IMT Port Alignment”
on page 35 of Alignments.
5 - 76
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
Tubing Replacement
X3 Disconnect and replace this tubing. X3 Connects the IMT Pump to the filter
T-fitting.
Waste Disconnect and replace this tubing. The waste tubing connects to the waste
bottle inside the system cabinet.
5 - 77
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
6. Locate the IMT Rotary Valve (to the right of the IMT interface).
7. Using needlenose pliers, remove the four screws or posts at each corner
in the top of the IMT Rotary Valve.
Figure 5-78: IMT Rotary Valve
8. Lift the rotary valve motor off the valve body. Turn it over and remove
the rotary valve seal from the valve seal holder on the encoder assembly.
9. Press down firmly on the encoder assembly and ensure that it springs
back with a click sound.
Note: If it does not, contact the Customer Care Center - Technical
Solution: Inside U.S.: 1-800-441-9250; Outside U.S. Refer to
the local support team.
5 - 78
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
1 Motor
2 Encoder Assembly
3 Valve Seal
5 Valve Body
10. Clean the inside of the valve body with a gauze pad lightly moistened
with water.
11. Place a pinhead-sized drop of Krytox grease on a gloved forefinger.
Gently rub the gloved forefinger and thumb together, then rub both
sides of the new rotary valve seal between the thumb and forefinger.
12. Install the new rotary valve seal with the flat side against the valve seal
holder on the valve motor assembly.
Note: The X and the slot on the seal should be facing towards the
operator.
13. Reinstall the motor assembly into the valve body by tightening the four
screws using the needlenose pliers.
14. Reconnect the tubing to the Standard A and Standard B bags.
15. Prime Standard A, Standard B, and the Flush solution.
16. Calibrate the IMT.
17. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
5 - 79
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
5 - 80
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
5 - 81
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
11. Select F7: DIAGNOSTICS > F1: ELECTRO/MECH > F2: SAMPLE SYSTEM >
F7: LOCI MODULE > F7: STATISTICS.
12. Select F1: RESET to reset the LOCI statistics. The following message
displays Do you really want to clear the LOCI statistics file? y/n. Press Y
to confirm.
13. In the field next to CYCLE, enter 40 as the number of cycles to perform.
14. Select F8: CYCLE.
15. Exit to the main operating screen.
16. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
5 - 82
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
5 - 83
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
4. Using the screwdriver, loosen the three captive screws that secure the R2
reagent arm to the instrument.
Note: Move the R2 arm counterclockwise to access and loosen
captive screw number 3.
5. Carefully lift the R2 reagent arm and set it aside.
5 - 84
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
5 - 85
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
10. Position the alignment slot in the R2 reagent arm over the alignment pin
in the instrument baseplate. Tighten the captive screws.
11. Restore power to the system following “Restoring Power After a
Controlled Shutdown” on page 10 of Appendix.
12. Perform the R2 Reagent Probe Alignment procedure. See “R2 Reagent
Probe Alignments” on page 22 of Alignments.
13. Perform a System Check and Daily QC.
14. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
5 - 86
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
6. Pull sensor connector P/J 13D up through the hole in the baseplate and
disconnect it.
Figure 5-87: Photometer Home Sensor
# Description
1 R1 Reagent Arm
7. Install the new photometer home sensor and the R1 reagent arm by
reversing the previous steps.
8. Restore power to the system following “Restoring Power After a
Controlled Shutdown” on page 10 of Appendix.
9. Perform Step 1 through Step 4 of the Photometer Alignment procedure.
See “Photometer Alignment ” on page 17 of Alignments.
10. Perform a System Check and Daily QC.
11. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
5 - 87
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
5 - 88
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
3 Thumbscrew
6 HM (2500 µL)
5. Remove the plastic shield on the rear of the pump assembly by loosening
the captive screw at the top of the shield.
6. Locate the sensor to be replaced and disconnect its connector.
7. Use the 5/64 hex wrench to remove the hex screw holding the sensor to
its bracket.
8. Remove and discard the sensor.
5 - 89
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
9. Install the optical sensor on the bracket using the 2-56 x.25 hex screw.
# Description
1 Optical Sensor
2 Locking Nut
3 Adjustment Screw
5 - 90
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
13. Turn the adjusting screw to lower the adjusting screw away from the
optical sensor.
Note: The syringe indicator light (LED) on the front panel of the
appropriate Motor Control Board in the card cage should now
be off.
14. Turn the lead screw on the pump panel to the raise the syringe plunger
and check the syringe gap between the bottom of the syringe and the
top of the metal plunger using a folded piece of paper.
Note: Printer paper should be equal to four thickness to set a gap of
0.005-0.010 inches (0.13-0.25 mm).
15. With the gap held steady, turn the adjustment screw until the syringe
indicator light (LED) on the front panel of the appropriate Motor Control
Board in the card cage just lights.
Table 5-11: Syringe Motor Control Board Slot Chart
5 - 91
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
5 - 92
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
5 - 93
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
1 Sampler Solenoid 1
2 Reagent 2 Solenoid
3 Sampler Solenoid 2
4 HM Solenoid 1
5 Reagent 1 Solenoid
7 HM Solenoid 2
5 - 94
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
8. Display the Pump Prime Menu screen and prime the appropriate pump.
a. Select F4: SYSTEM PREP > F7: PUMP PRIME.
b. Move the cursor to the Cycles field and type 6 for the number of
cycles to be performed.
c. Press the appropriate function key to begin priming the pump for
the solenoid valve that was replaced.
d. While priming, check for any tubing leaks.
e. Continue priming the pump until all the air is out of its tubing.
9. Perform a System Check and the Daily QC.
10. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
5 - 95
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
3 Thumbscrew
6 HM (2500 µL)
5 - 96
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
CAUTION: Use the needlenose pliers on the metal part of the syringe
plunger. Do not use pliers on the glass part of the syringe.
4. Press the pump assembly release button and pull the pump assembly
outward.
5. Remove the plastic shield on the rear of the pump assembly by loosening
the captive screw at the top of the shield.
6. Manually turn the syringe’s pump lead screw to completely lower the
syringe plunger.
Figure 5-95: Pump Lead Screws
9. Fit the new syringe into the upper bracket and loosely tighten the 7/64
Allen screw in the upper bracket.
10. Attach the tubing to the fitting, turning the syringe slightly to obtain
better access.
5 - 97
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
11. Slowly pull the plunger down until it is snug in the bottom syringe
bracket, then attach and completely tighten the thumbscrew.
CAUTION: Use the needlenose pliers only on the metal part of the syringe
plunger. Do not use pliers on the glass part of the syringe.
12. Completely tighten the 7/64 Allen screw at the top of the syringe bracket.
13. Check the syringe gap between the bottom of the syringe and the top of
the metal plunger using folded piece of paper.
Note: The printer paper should be folded to equal four thickness to
set a gap of 0.005-0.010 inches (0.13-0.25 mm).
a. If the gap setting is correct, the printer paper should barely be able
to slide between the bottom of the syringe and the top of the metal
plunger. If adjustment is not required, proceed to step 19.
b. If a gap adjustment is required, continue with step 14.
Figure 5-96: Syringe
# Description
1 Plunger
2 Syringe
3 Foot of Plunger
5 - 98
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
1 Optical Sensor
2 Locking Nut
3 Adjustment Screw
5 - 99
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
15. Turn the adjusting screw to lower the adjusting screw away from the
optical sensor.
Note: The syringe indicator light (LED) on the front of the panel of
the appropriate Motor Control Board in the card cage should
now be off.
16. Turn the lead screw on the pump panel to the right of the syringe
plunger and check the syringe gap between the bottom of the syringe
and the top of the metal plunger using a folded printer paper.
17. With the gap held steady, turn the adjustment screw until the syringe
indicator light (LED) on the front of the panel of the appropriate Motor
Control Board in the card cage just lights.
18. Recheck the gap using step 13.
Table 5-12: Syringe Motor Control Board Slot Chart
19. Display the Pump Prime Menu screen and prime the appropriate pump.
a. Select F4: SYSTEM PREP > F7: PUMP PRIME.
b. Move the cursor to the Cycles field and type 6 for the number of
cycles to be performed.
c. Press the appropriate function key to begin priming the pump for
the solenoid valve that was replaced.
d. While priming, check for any tubing leaks.
e. Continue priming the pump until all the air is out of its tubing.
20. After priming the pump, check the gap setting again.
21. If the gap setting is correct, use the 3/32 allen wrench inserted at the
bottom of the adjusting screw to keep the screw from turning, tighten
the locking nut using the 11/32 allen wrench.
22. After tightening the nut, perform another pump prime. Make a final
check of the gap to ensure that it did not move while the locking nut was
tightened.
23. If further adjustment is required, repeat step 14.
24. Select RESET and recheck the gap.
5 - 100
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
4. Move the probe away from its reagent drain until the sensor is accessed.
5. Disconnect P/J 20C for R1 (or P/J 80D for R2).
6. Using a screwdriver, remove the screw that holds the radial home sensor
to the reagent arm.
7. Remove the sensor.
Figure 5-98: Radial Home Sensor
5 - 101
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
CAUTION: Be careful not to damage any other connectors when cutting tie
wraps.
8. Install the new sensor by reversing the previous steps.
Note: Position the R1 sensor with the yellow and black wires at the
top.
9. Select PAUSE to restart the sampling system.
10. Perform all the reagent arm alignments for the probe on which the
sensor was replaced. See “R1 Reagent Probe Alignments” on page 18 of
Alignments. See “R2 Reagent Probe Alignments” on page 22 of
Alignments. See R3 Probe Alignments in the RMS Operator’s Guide.
11. Close the instrument lids.
12. Perform a System Check.
13. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
4. Move the probe away from its reagent drain (by turning the top shaft of
the R1 reagent arm).
5. Disconnect P/J 20D for R1 or P/J 80B for R2.
5 - 102
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
2 Cable Holder
CAUTION: Be careful not to damage any other connectors when cutting tie
wraps.
6. Using a screwdriver, remove the screw that holds the vertical sensor to
the sensor housing bracket and remove the sensor from the reagent arm.
7. Install the new sensor by reversing the previous steps.
8. Close the instrument lids.
9. Press PAUSE to restart the sampling system.
10. Perform all the reagent arm alignments for the probe on which the
sensor was replaced.
11. Close the system lids.
12. Perform a System Check.
13. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
5 - 103
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
4. Move the probe away from its reagent drain until the sensor can be
accessed.
5. Disconnect P/J 32D.
6. Using a screwdriver, remove the screw that holds the sensor to its
bracket.
7. Remove the sensor from the instrument.
Figure 5-100: Rotational Home Sensor
8. Separate the metal clip from the old sensor and place it on the new
sensor.
5 - 104
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
1 Metal Clip
2 Sensor
3 Bracket
5 - 105
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
1 Ratchet Wrench
2 Offset Wrench
3 Open-End Wrench
5 - 106
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
4. If replacing the R2, move the probe away from its reagent drain until its
reagent probe tip can be accessed.
5. Place a paper towel under the probe to prevent anything from falling into
a reagent cartridge while moving the probe tip.
6. Remove the probe tip and nut by using the 5/16 wrench with either the
offset wrench or the ratchet wrench.
a. Using the 5/16 wrench and the offset wrench:
1) Place the 5/16 wrench on the reagent probe body to keep it
from turning.
2) Slide the 1/4 end of the offset wrench up the reagent probe
and onto the probe tip nut.
3) Turn the offset wrench clockwise to loosen the probe tip nut.
Once loose, unscrew it by hand to remove the probe tip and
nut.
5 - 107
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
1 Open-end Wrench
5 - 108
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
1 Open-End Wrench
7. Insert a new reagent probe tip into a new probe tip nut.
8. Screw the new reagent probe tip onto the reagent probe body
finger-tight.
Figure 5-106: Reagent Probe Tip
# Description
1 Reagent Probe
2 Nut
5 - 109
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
9. Completely tighten the new reagent probe tip using the 5/16 open-end
wrench with offset wrench or ratchet wrench.
a. Using the ratchet wrench to tighten the probe tip:
1) Place the 5/16 wrench on the reagent probe body to keep it
from turning.
2) With the “Off” marking on the ratchet wrench facing up, slide
the 1/4 opening of the ratchet wrench up the reagent probe
and onto the probe tip nut.
3) Turn the ratchet wrench counterclockwise to completely
tighten the probe tip and nut to the probe body.
10. Close the instrument lids and select PAUSE to restart the sampling
system.
11. Select F4: SYSTEM PREP > F7: PUMP PRIME.
12. Move the cursor box to the appropriate reagent probe tip field, and select
ENTER to change the field from NO to YES.
13. Select F1: STORE CHANGES.
14. Prime the reagent pump by:
a. Selecting F4 SYSTEM PREP > F7: PUMP PRIME.
b. Move the cursor to the Cycles field and type 3 for the number of
cycles to be performed.
c. Select the appropriate function key to begin priming the reagent
pump for the tip that was replaced.
15. Perform the Reagent Probe Alignments procedures, see Chapter 6:
Alignments
16. Perform the Daily QC.
17. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
5 - 110
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
2 Bar
3 Tubing Clamps
6. Using the 9/64 Allen wrench, loosen the four screws that secure the drive
motor to its bracket.
7. Slide the drive motor toward the front of the instrument to disengage its
gears.
5 - 111
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
1 Motor Screws
2 Allen Screws
8. Use a screwdriver to remove the three screws that secure the reagent
tray drive assembly to the instrument baseplate.
9. Remove the assembly by carefully lifting it up and out of the instrument.
10. Gently pull the spring bar outward until the reagent tray home sensor
can be pushed out of its hole in the spring bar.
11. Remove one cable clamp and disconnect P/J 50H.
5 - 112
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
1 Springbar
2 Sensor
5 - 113
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
5 - 114
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
1 Knurled Fitting
2 500 µL Syringe
5 - 115
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
8. Disconnect the reagent tubing from the reagent probe by pulling on the
probe knurled fitting.
9. Unscrew the knurled fitting from the tubing and pull the tubing through
the spring guide.
Figure 5-113: Spring Guide
# Description
1 Spring Guide
2 Knurled Fitting
10. Remove the tubing from tubing clamps or guides as necessary and pull it
up through the instrument.
11. Discard the tubing.
12. Start from the reagent probe and thread the new tubing down through
the instrument to the reagent pump.
13. Position the black wrappings (or stamped markings) on the reagent
tubing as shown in Figure 5-114.
5 - 116
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
1 R2 Tubing
2 Tubing Clamps
3 R1 Tubing
4 Tubing Guide
5 Spring Clamp
5 - 117
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
1 R1 Tubing
2 R2 Tubing
14. Attach the knurled fitting removed from the old tubing in step 5 of this
procedure to the pump end of the new tubing by screwing the fitting on
by hand until approximately 1/16 inch (1.5 mm) of the tubing extends
past the fitting.
15. Press the tubing onto the 500 µL reagent syringe fitting.
Figure 5-116: 1/16” of Tubing
16. Install the new reagent tubing through the spring guide and onto the
reagent probe.
5 - 118
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
1 Spring Guide
2 Metal Fitting
3 Knurled Fitting
5 - 119
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
1 Ratchet Wrench
2 Offset Wrench
3 Open-End Wrench
5 - 120
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
4. Remove the probe tip and nut using the 5/16 open-end wrench with
either the offset wrench or the ratchet wrench:
a. Using the 5/16 open-end wrench and the offset wrench:
1) Place the 5/16 wrench on the sample probe body to stabilize it
and prevent it from turning.
2) Slide the 1/4 end of the offset wrench up the sample probe
and onto the probe tip nut.
3) Turn the offset wrench clockwise to loosen the probe tip nut.
Once loose, unscrew it by hand to remove the probe tip and
nut.
5 - 121
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
2 Offset Wrench
5 - 122
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
1 Ratchet Wrench
5. Insert a new sample probe tip into a new probe tip nut and screw it onto
the sample probe body finger-tight.
6. Completely tighten the new sample probe tip using the 5/16 open-end
wrench with offset wrench or ratchet wrench.
a. To use a ratchet wrench to tighten the probe tip:
1) Place the 5/16 wrench on the sample probe body to keep it
from turning.
2) With the “OFF” marking on the ratchet wrench facing upward,
slide the 1/4 opening of the ratchet wrench up the reagent
probe and onto the probe tip nut.
3) Turn the ratchet wrench counterclockwise to completely
tighten the probe tip and nut to the probe body.
7. Close the instrument lids and select PAUSE to restart the sampling
system.
8. Select F4: SYSTEM PREP > F6: SYS COUNTERS.
9. Move the cursor box to the Sample Probe Tip field.
10. Select ENTER to change the field from NO to YES.
11. Select F1: STORE CHANGES.
5 - 123
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
1 Left Port
2 To Sample Probe
3 100 µL Syringe
4. Remove the plastic sleeve from the end of the tubing. Set this fitting
aside for use later in the procedure.
5 - 124
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
5. Open the pump panel to access the tubing on the end of the clot check
board.
6. Slide the plastic sleeve back and remove the tubing from both sides of
the clot check fittings. Remove the plastic sleeves from both pieces of
tubing before removing the tubings from the conduit.
Figure 5-123: Clot Check Tubing
7. Remove the sample tubing from the 100 µL syringe and pull through the
conduit.
8. At the sample probe, gently pull out on the sides of the sample cover and
lift it off the probe assembly.
Figure 5-124: Sample Probe Cover Assembly
9. Remove the sample tubing from the sample probe connection and pull
the tubing up through the system.
10. Press the black sheathed end of the new sample tubing onto the sample
probe connection. Remove the plastic sleeves from the end of the sample
tubing.
5 - 125
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
11. Rout the other end of the sample tubing down through the corrugated
black plastic conduit until it comes out of the opening on the left side of
the clot check board.
12. Slide the plastic sleeve onto the end of the tubing and slide the tubing
onto the clot check board fitting.
13. Attach the sample tubing and press the tubing onto the left port of the
100 µL syringe.
14. Remove the plastic sleeve from the end of the tubing and feed it through
the corrugated black plastic conduit until it comes out of the opening on
the right side of the clot check board.
15. Slide the plastic sleeve on the end of the tubing.
16. Flare the sample tubing if needed. Flare the tubing by inserting the flare
tool from the Spare Parts kit into the end of the tubing that will be
slipped onto the fitting. Slide the tubing onto the clot check board fitting.
17. Close the system lids and press PAUSE to turn on the sampling system.
18. Display the Pump Prime menu.
a. Move the cursor to the Cycles field.
b. Type 3 for the number of prime cycles to be performed.
c. Press ENTER.
d. Select F2: SAMPLER.
19. Check the sample tubing for leaks while the system is priming.
20. Perform a system check and daily QC.
21. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
5 - 126
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
WARNING: Wait at least five minutes after turning off the instrument
power for the source lamp assembly to cool before continuing
this procedure.
2. Lower the cuvette thermal chamber, see “Lowering and Raising the
Thermal Chamber” on page 21 of this chapter.
3. Loosen the two captive thumbscrews that hold the source lamp assembly
in place.
Figure 5-125: Source Lamp Assembly
# Description
1 P/J 72B
2 Screws
4. Unplug the source lamp connector P/J 72 B located inside the cuvette
thermal chamber. Remove the source lamp assembly.
5. Install the new source lamp assembly by reversing Step 2 through Step 4.
5 - 127
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
5 - 128
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
1 Barcode reader
6. Unlock the U-seal solenoid assembly by pulling its curved locking bar to
the right.
7. Move this assembly to the left away from the cuvette ring.
5 - 129
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
3 U-Seal Solenoid
5 - 130
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
1 Tie Wrap
2 Tie Mount
3 P/J 13 L
9. The cable is looped around the tie wrap. With needlenose pliers, break
the tie wrap and remove it.
10. Using the screwdriver gently pry the U-Seal element off its metal frame.
5 - 131
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
1 U-Seal Element
2 Pry Here
11. Snap the new U-seal element onto the metal frame.
12. Curl the end of a new tie wrap to permit easy access.
13. Insert the tie wrap through the tie mount on the cuvette arm.
14. Insert the end of the tie wrap through the head of the tie wrap and form
a loop about 3/4” in diameter.
5 - 132
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
# Description
1 Tie Wrap
2 Tie Mount
3 P/J 13 L
15. Pass the P/J 13L cable through the tie wrap loop and wrap it around once.
Connect P/J13L.
16. Tighten the tie wrap around the cable enough to hold but not pinch it.
17. Cut off the excess tie wrap.
18. Reposition the U-seal solenoid near the cuvette ring and push the curved
locking bar into place.
19. Position the alignment bar and tighten its thumbscrew.
20. Lower the barcode reader.
21. Close the instrument lids.
22. Record completion of this procedure. Refer to the section titled “Using
the Instrument Log Screen” in Chapter 2.
5 - 133
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
1 Cover
2 Filter
5. Remove the old felt filter off of the vacuum pump fitting and replace it.
5 - 134
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
5. Loosen the collar ring on the cap of the water bottle (the collar ring turns
without moving the cap).
6. Remove the dip tube assembly and place it in the spare water bottle.
7. Reconnect P/J 1A.
8. Place the water bottle back in the system.
Note: The water bottle should begin filling with water.
9. Wait until the water bottle is filled, and prime all system pumps:
a. Select F4: SYSTEM PREP > F7: PUMP PRIME.
b. Move the cursor to the Cycles field and type 3 for the number of
cycles to be performed.
c. Select ENTER.
d. Select F1: ALL.
10. Document this replacement on the Instrument Log screen. Refer to the
section titled “Using the Instrument Log Screen” in Chapter 2.
11. Clean the used water bottle. See “Cleaning the Water Bottle” on page 37
of this chapter.
5 - 135
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
# Description
1 Captive Screws
2 Square Handles
4. Use the square handles on the board faceplate to pull the board halfway
out of its slot and then firmly re-seat it.
5. Tighten the two captive screws.
6. Restore power to the system following “Restoring Power After a
Controlled Shutdown” on page 10 of Appendix.
5 - 136
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
Mean SD
5 - 137
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
5 - 138
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
Mean SD
Decontamination Procedure
Note: Perform this procedure only if advised to do so by a Siemens
representative. It takes approximately 50 minutes to
complete these steps.
1. Disable the water system supply:
a. From the Operating Menu, select F6: SYSTEM CONFIG.
b. Move the cursor to “Water In”.
c. Enter a password.
d. Select ENTER to change the selection from PLUMBED to MANUAL.
5 - 139
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
5.25 500 mL
6.00 438 mL
8.00 328 mL
10.00 263 mL
12.00 219 mL
15.00 175 mL
5 - 140
Dimension EXL with LM/EXL 200 System Operator’s Guide Maintenance
5 - 141
Maintenance Dimension EXL with LM/EXL 200 System Operator’s Guide
5 - 142
Dimension EXL with LM/EXL 200 System Operator’s Guide Alignments
Chapter 6: Alignments
6-1
Alignments Dimension EXL with LM/EXL 200 System Operator’s Guide
Alignment Offsets
Each component is set to a specific position during manufacturing. To
compensate for these variations, the system’s components are corrected to
manufacturing specifications using offset values determined when the
component is aligned.
6-2
Dimension EXL with LM/EXL 200 System Operator’s Guide Alignments
6-3
Alignments Dimension EXL with LM/EXL 200 System Operator’s Guide
f. Retighten the screws. Then select F5: BARCODE ALIGN to rerun the
alignment.
Figure 6-1: Barcode Scanner Screws
g. Compare the new alignment values with the alignment limits noted
in Step a and Step b above.
1) If both alignments pass, continue with Step 9.
2) If both alignments fail, contact the Customer Care Center -
Technical Solution: Inside U.S.: 1-800-441-9250; Outside U.S.
Refer to the local support team.
9. Select EXIT and follow the messages to remove the barcode labeled
sample tubes in segment #1, position #5 and position #6.
6-4
Dimension EXL with LM/EXL 200 System Operator’s Guide Alignments
# Description
1 Cuvette Ring
2 Plunger
3 Cuvette Diaphragm
6-5
Alignments Dimension EXL with LM/EXL 200 System Operator’s Guide
9. Select F1: START and visually inspect the position of the cuvette ring.
10. Repeat Step 8 if necessary.
11. Select EXIT.
12. Close the reagent lid.
13. Perform the “Photometer Alignment ” on page 17 of this chapter.
6-6
Dimension EXL with LM/EXL 200 System Operator’s Guide Alignments
HM Module Alignments
This procedure includes alignments of parts on the HM module. To completely
align the HM module, perform the following two alignments as well:
• Sample arm to incubate wheel
• R2 reagent arm to incubate wheel
The HM module component alignments consist of the following:
• Incubate and Wash Wheel Alignments:
a. Incubate wheel to transfer opening
b. Wash wheel to incubate wheel
c. Wash probes to wash wheel
• Vessel Transfer Shuttle Alignments:
d. Vessel transfer shuttle to incubate wheel
e. Vessel transfer shuttle to wash wheel
f. Vessel transfer shuttle to load
All six alignments listed above must be checked/performed to align the HM
module components. Use message prompts as they display to help with each
alignment.
WARNING: Ensure lids are fully open to avoid injury while performing
maintenance.
6-7
Alignments Dimension EXL with LM/EXL 200 System Operator’s Guide
Incubate Wheel to The edges of the transfer station and the incubate wheel slots are
Transfer Opening aligned with each other so that there is an unobstructed path into
the incubate wheel.
Wash Wheel The white alignment mark on the wash wheel is aligned to the white
to Incubate Wheel alignment mark on the incubate wheel.
6-8
Dimension EXL with LM/EXL 200 System Operator’s Guide Alignments
6-9
Alignments Dimension EXL with LM/EXL 200 System Operator’s Guide
Wash Probes The wash probe tips are centered over and above the white
to Wash Wheel alignment dots on the wash wheel.
6 - 10
Dimension EXL with LM/EXL 200 System Operator’s Guide Alignments
5. Select F8: ACCEPT and continue with “Aligning the Vessel Transfer
Shuttle to the Wash Wheel” on page 11 of this chapter.
Table 6-4: Alignment and Correct Positioning
Vessel Transfer Shuttle The edges of the vessel guide on the end of the vessel transfer
to Incubate Wheel shuttle are centered around the top of the vessel in the incubate
wheel.
Vessel Transfer Shuttle The left edge of the shuttle vessel guide on the end of the vessel
to Wash Wheel transfer shuttle is just touching the vessel in the wash wheel.
6 - 11
Alignments Dimension EXL with LM/EXL 200 System Operator’s Guide
6 - 12
Dimension EXL with LM/EXL 200 System Operator’s Guide Alignments
# Description
Vessel Transfer Shuttle The vessel guide on the end of the vessel transfer shuttle is
to Load centered over the top of the vessel in the transfer station.
6 - 13
Alignments Dimension EXL with LM/EXL 200 System Operator’s Guide
5. Select EXIT and follow the message to remove any reaction vessels
placed on the wheel during the alignment procedure.
Note: Remember to replace the HM incubation wheel cover after all
alignments are complete.
6 - 14
Dimension EXL with LM/EXL 200 System Operator’s Guide Alignments
2. Select F7: DIAGNOSTICS > F3: ALIGNMENTS > F6: HM MODULE >
F6: LOCI ALIGNMENTS.
3. Move the cursor first to the Automatic selection above LOCI ARM TO
VESSEL ALIGNMENTS and select F7: CHECK ALIGN.
4. When prompted, place the LOCI alignment gauge on the incubation
wheel just left of the white dot.
6 - 15
Alignments Dimension EXL with LM/EXL 200 System Operator’s Guide
5. Select ENTER.
Note: The HM wheel and the LOCI arm will move through the
auto-align process. When complete, the instrument gives the
opportunity to manually adjust the alignment.
6. Select F8: ACCEPT when complete.
7. Exit to the main operating screen and remember to remove the LOCI
alignment gauge.
6 - 16
Dimension EXL with LM/EXL 200 System Operator’s Guide Alignments
Photometer Alignment
1. Ensure that all instrument doors and lids are closed.
2. Select F7: DIAGNOSTICS > F3: ALIGNMENTS > F4: PHOTOMETER >
F3: ARM ALIGN > F1: START.
Note: Do not touch any other keyboard keys until this alignment
has been completed.
3. When the message New alignment is: (old value: ) Do you want to
store? (y/n) displays, the alignment is complete. The alignment is in
progress when the screen plot at the right slowly displays.
Figure 6-13: Photometer Alignment Screen
6 - 17
Alignments Dimension EXL with LM/EXL 200 System Operator’s Guide
6 - 18
Dimension EXL with LM/EXL 200 System Operator’s Guide Alignments
Note: For a better view of the probe position, use the down arrow
key on the keyboard to lower the probe into the drain.
Figure 6-14: Drain Alignment
3. Visually align the probe using the function keys described on the screen.
4. When the probe is aligned, select F8: ACCEPT.
5. Select F7: CHECK ALIGN and ensure that the component moves and
returns to the new alignment position.
6. Continue with the procedure “R1 Reagent Arm to Target (Cuvette)
Alignment” in this chapter.
6 - 19
Alignments Dimension EXL with LM/EXL 200 System Operator’s Guide
4. If the probe is aligned correctly, the probe will be within the flat circled
tip in the center of the alignment gauge and above the gauge. If an
adjustment is needed, continue with Step 5; if the probe is aligned
correctly, skip to Step 6.
Note: The operator should be able to fit a piece of printer paper
between the probe tip and the gauge and be able to feel
resistance without ripping the paper.
5. Visually align the probe using the function keys described on the screen.
6 - 20
Dimension EXL with LM/EXL 200 System Operator’s Guide Alignments
5. Visually align the probe using the function keys described on the screen.
6. When the probe is aligned, select F8: ACCEPT.
7. Select Exit and follow the messages as they appear to remove the
reagent tray alignment gauge and return the reagent cartridge to its
original slot in the reagent tray.
Note: If a reagent cartridge was removed, place the same cartridge
in the reagent tray slot from which it was removed. In
addition, record the positions of the cartridges prior to
removal.
8. Select F7: CHECK ALIGN and ensure that the component moves and
returns to the new alignment position.
9. Close the instrument lids.
6 - 21
Alignments Dimension EXL with LM/EXL 200 System Operator’s Guide
6 - 22
Dimension EXL with LM/EXL 200 System Operator’s Guide Alignments
# Description
2. Use the 9/64 Allen wrench to turn the adjustment screw on the bracket in
the same direction that the reagent probe needs to be moved.
3. Select F7: CHECK ALIGN after each time the adjustment screw is turned.
When any part of the probe tip or securing nut is above the hole in the
cover of the drain, tighten the two alignment bracket screws.
4. Return to Step 3 of the “R2 Reagent Arm to Drain Alignment” procedure.
6 - 23
Alignments Dimension EXL with LM/EXL 200 System Operator’s Guide
6 - 24
Dimension EXL with LM/EXL 200 System Operator’s Guide Alignments
6 - 25
Alignments Dimension EXL with LM/EXL 200 System Operator’s Guide
6 - 26
Dimension EXL with LM/EXL 200 System Operator’s Guide Alignments
6 - 27
Alignments Dimension EXL with LM/EXL 200 System Operator’s Guide
6. Select F5: INSERT GAUGE and follow the messages on the screen to
insert the reagent tray alignment gauge into the instrument.
Note: If prompted to remove a reagent cartridge, this same
cartridge must be placed into the instrument at the end of
this procedure. Record the positions of the cartridges prior to
removal. The Alignment positions are described and shown
in Table 6-7.
Table 6-7: Alignment Positions
Auto Loader Press F6: CALIB LOADER. There is no alignment position to check. When the
automatic Flex loader stops all movement, the alignment is complete.
R1 Arm Press F1: POS GAUGE. The R1 probe should be above and centered within the target
circle on the gauge. Use the function keys to return to the alignment position.
R2 Arm Press F1: POS GAUGE. The R2 probe should be above and centered within the target
circle on the gauge. Use the function keys to return to the alignment position.
7. The operator should be able to fit a piece of printer paper between the
probe tip and the gauge and be able to feel resistance without ripping
the paper.
8. Select F7: CHECK ALIGN.
9. Use the right arrow key to move to the next mode and repeat Step 6.
10. When the alignment is complete, select F8: ACCEPT.
11. After visually performing an alignment, select F7: CHECK ALIGN and
ensure that the component moves and returns to the new alignment
position.
12. Select Exit and follow the messages that display to remove the reagent
tray alignment gauge and return the reagent cartridge to its original slot
in the reagent tray.
6 - 28
Dimension EXL with LM/EXL 200 System Operator’s Guide Alignments
6 - 29
Alignments Dimension EXL with LM/EXL 200 System Operator’s Guide
6 - 30
Dimension EXL with LM/EXL 200 System Operator’s Guide Alignments
# Description
1 Alignment Gauge
2 Adapter
2. Follow the messages as they display on the screen to place the cup
alignment gauge in the top of an adapter in segment #1, position #1
(marked inside the sample area).
Note: If the probe is aligned correctly, it should be positioned above
the gauge and within the target circle of the gauge.
3. If an adjustment is needed, continue with Step 4; if the probe is aligned
correctly, skip to Step 5.
Note: The operator should be able to fit a piece of printer paper
between the probe tip and the gauge and be able to feel
resistance without ripping the paper.
4. Visually align the probe using the function keys described on the screen.
6 - 31
Alignments Dimension EXL with LM/EXL 200 System Operator’s Guide
7. Select F7: CHECK ALIGN and follow the messages on the screen to move
the cup alignment gauge into the top of an adapter in segment #1,
position #2.
Note: If the probe is aligned correctly, it should be positioned just
above the gauge, within the target circle. If an adjustment is
needed, continue with Step 8. If the probe is aligned correctly,
skip to Step 9.
8. Visually align the probe using the function keys described on the screen.
9. When the probe is aligned, select F8: ACCEPT.
10. Select EXIT and follow the messages on the screen to remove the cup
alignment gauge.
11. Continue with the procedure “Sample Probe to Drain Alignment” in this
Chapter.
6 - 32
Dimension EXL with LM/EXL 200 System Operator’s Guide Alignments
6 - 33
Alignments Dimension EXL with LM/EXL 200 System Operator’s Guide
4. Visually align the probe using the function keys on the screen.
a. F1 and F2 move the sample probe clockwise and counterclockwise.
b. F3 and F4 move the sample probe up and down.
c. F5 and F6 move the incubate wheel clockwise and
counterclockwise.
5. When the probe is aligned, select F8: ACCEPT.
6. Replace the HM incubation wheel cover
7. Continue with the procedure “Sample Probe to the IMT Port Alignment”
in this Chapter.
6 - 34
Dimension EXL with LM/EXL 200 System Operator’s Guide Alignments
3. Visually align the probe using the function keys on the screen.
a. F1 and F2 move the sample probe clockwise and counterclockwise.
b. F3 and F4 move the sample probe up and down.
4. When the probe is aligned, select F8: ACCEPT.
5. After aligning, select F7: CHECK ALIGN.
6 - 35
Alignments Dimension EXL with LM/EXL 200 System Operator’s Guide
CAUTION: Ensure that the sample probe to cup alignment has been
performed before performing this alignment.
1. Select F7: DIAGNOSTICS > F3: ALIGNMENTS > F2: SAMPLER >
F8: PROBE / BAR CODE.
2. Move the box in the Max Depth in: field to the appropriate sample
container.
3. Select F7: CHECK PROBE and follow messages as they display.
4. Set the correct probe depth using the F3 and F4 function keys.
5. Select F7: CHECK ALIGN and ensure that the system moves and returns
to the new alignment position.
Table 6-8: Container and Correct Probe Depth
Primary Tube Align to the primary tube that sits the highest in the segment. The bottom
edge of the nut on the sample probe should be even with the top of the
primary tube.
PED Align to the pediatric (PED) tube that sits highest in the segment. The sample
probe tip should be above the bottom of the pediatric (PED) tube.
SSC Align to the primary tube/SSC combination that sits the highest in the
segments. The sample probe tip should be above the bottom of the SSC. The
size of the sample tube used during this SSC alignment should now be used
with all SSC samples.
• For processing whole blood samples, always place the SSC in the same
size tube used for the SSC sample probe maximum depth alignment.
Refer to the method IFU for sample handling details.
• For processing serum and plasma, primary tube/SSC combinations that
present the SSC at a lower height may be used only if the SSC is filled with
a maximum of 1.0 mL of sample.
6 - 36
Dimension EXL with LM/EXL 200 System Operator’s Guide Alignments
# Description
2 Probe
6 - 37
Alignments Dimension EXL with LM/EXL 200 System Operator’s Guide
6 - 38
Dimension EXL with LM/EXL 200 System Operator’s Guide Troubleshooting
Chapter 7: Troubleshooting
7-1
Troubleshooting Dimension EXL with LM/EXL 200 System Operator’s Guide
Troubleshooting Overview
Help Keys
The Help Keys provide information on:
• A specific screen
• Tasks performed on a screen
• Function keys actions
• A brief procedure for how to perform the tasks or cross-reference to
a procedure in this Operator’s Guide
Table 7-1 shows how to access Help screens.
Table 7-1: Help Screens
ALT+M This key combination displays active error messages. It does not display error
messages that have been viewed and reset.
Help (from the Displays the screens where help for performing a task is available. The Quick
Operating Menu) Index includes the function key sequence to display the screens from the
Operating Menu.
Help (from any Displays information on how to use the screen and its function keys.
other screen)
SHIFT+Help If the cursor is in a field where data can be entered, pressing this key
combination displays acceptable entries for the field.
ALT+Help Displays a screen that explains how Help works and a list of the ALT key
combinations.
CTRL+Help Pressing this key combination displays a list of information for all operating
condition icons. Scroll down on this page to get information on a particular
icon.
7-2
Dimension EXL with LM/EXL 200 System Operator’s Guide Troubleshooting
a. Has the lab started to use any new methods or manufacturing lot
numbers of existing assays?
b. Has anyone replaced or adjusted any system parts?
c. What type of maintenance (daily, monthly, other) was last
performed on the instrument?
4. What common characteristics do the problems share?
5. Is the problem specific to one method or several? If several, do they have
something (such as wavelength) in common?
7-3
Troubleshooting Dimension EXL with LM/EXL 200 System Operator’s Guide
Remote Access
The Customer Care Center - Technical Solution may ask to remotely access an
instrument using the instrument modem. There are two modes of access:
monitor and control. In monitor mode, the specialist can look at files in the
system’s computer memory and/or see the screens. In control mode, the
Customer Care Center specialist can perform the same keyboard functions as
the operator standing at the instrument. The Customer Care Center specialist
will give specific directions over the telephone to establish the remote link
between the system and the Customer Care Center.
Remote Control
WARNING: When taking control of an instrument, ensure that instrument
operators are aware that remote access may move
instrument components thereby presenting a hazard to the
operator.
Remote control is used with various types of servers and/or devices. When an
instrument is engaged in a remote control session where VNC software is
resident on an instrument, a notification will be displayed: Exercise caution
while performing activities involving the physical movement of
instrument modules. Please ensure that users stay clear of the instrument.
7-4
Dimension EXL with LM/EXL 200 System Operator’s Guide Troubleshooting
Chemistry Troubleshooting
Instrument problems are evident when the system displays an error message.
Chemistry problems, however, may not display an error message and may
become evident only when test results are reviewed.
Refer to “Understanding Test Results” on page 2 of Chapter 3: Sample Reports
for details on how to interpret the messages accompanying test results.
The following tips help prevent chemistry problems:
• Perform scheduled maintenance.
• Calibrate or verify a new lot of Flex reagent cartridges before the current
lot runs out.
• Check the expiration date of the calibrator/verifier in use.
• Use fresh quality control materials and check their expiration dates.
• Be sure that QC values are within the assay range identified for each assay.
• Check for interfering substances that may affect test results (see the assay
IFU).
• Assure proper handling and preparation of samples (pretreatment,
preservative, etc.).
7-5
Troubleshooting Dimension EXL with LM/EXL 200 System Operator’s Guide
7-6
Dimension EXL with LM/EXL 200 System Operator’s Guide Troubleshooting
7-7
Troubleshooting Dimension EXL with LM/EXL 200 System Operator’s Guide
Asterisk Displays Next Indicates that foaming Rerun the System Check. If the asterisk
to one or more System occurred in one or more of still displays, follow the appropriate
Check Results the cuvettes during the troubleshooting in this section for the
System Check. system that has the asterisk.
Asterisks Display Indicates that an error Select F5: PROCESS CTRL > F6: ERROR
Instead of Max Diff, condition occurred that LOG, or F7: DIAGNOSTICS >
Mean and SD caused the instrument to F6: ERROR LOG and resolve the error
Calculations abort the system check. condition before rerunning the system
check.
Probe Cleaner Failed Indicates that there was Note: Ensure that all other System
insufficient Sample Probe Check results are acceptable
Cleaner or Reagent Probe before troubleshooting this
Cleaner present in the problem.
appropriate drain. 1. Prime the appropriate probe cleaner
from the Pump Prime screen.
2. Check the tightness of the probe
cleaner tubing connections to the
pump and drain.
3. Check that the probe drain and
fittings are not leaking.
4. Check that the probe cleaner tubing
is completely inserted into and
reaches the bottom of the bottle.
5. Check that the correct cleaner
solution is in place.
No System Check A system condition 1. Check the sample status to see if the
Printout prevented the printout. system check is still processing.
2. Check that F1: START was selected
to start the system check.
3. Check that there is not a system
need for an ABS/CHK Flex reagent
cartridge.
4. Check the short sample load list for a
possible short sample in the cup
that contains the ABS solution.
7-8
Dimension EXL with LM/EXL 200 System Operator’s Guide Troubleshooting
Stray light entering The reagent lid must be closed and all instrument side panels and doors
must be closed when a system check is processing.
Dirty cuvette windows Clean cuvette windows. See “Cleaning Cuvette Windows” on page 28 of
Chapter 5: Maintenance.
Source lamp was This will be the cause only if a new source lamp has been installed. Remove
installed incorrectly and reinstall the source lamp. See “Replacing the Source Lamp” on
page 127 of Chapter 5: Maintenance.
Optical filter is dirty or Remove the specific optical filter that is failing, inspect it, and either clean
has become or replace it.
delaminated
7-9
Troubleshooting Dimension EXL with LM/EXL 200 System Operator’s Guide
Using a new lot of ABS/CHK and Display the Daily Maintenance screen and enter the value from
failing to enter ABS/CHK carton the ABS/CHK carton flap into the Carton Value field for the
flap value into the software. ABS/CHK lot.
Loose tubing on the reagent Check the tightness of all tubing connections on the R1 or R2
syringes. reagent pump panel.
Reagent probe is misaligned. Realign the reagent probe. See “R1 Reagent Probe Alignments”
on page 18 of Chapter 6: Alignments and see “R2 Reagent Probe
Alignments” on page 22 of Chapter 6.
Crimped or pinched reagent Check if the reagent tubing is crimped (bent) or pinched.
tubing. Replace any tubing that appears to be damaged.
Reagent probe tip is not Replace the probe tip. See “Replacing a Reagent Probe Tip” on
functioning properly. page 106 of Chapter 5: Maintenance.
Reagent syringe plunger tips Replace both the 500-µL and the 2500-µL reagent syringes on
are not functioning properly. the reagent pump panel for the appropriate pump. See
“Replacing a Pump Syringe” on page 96 of Chapter 5:
Maintenance.
7 - 10
Dimension EXL with LM/EXL 200 System Operator’s Guide Troubleshooting
Not filling the sample cup with Look at the System Check printout and see what ABS/CHK Flex®
ABS from the same lot that the lot number was used for the System Check by the instrument.
instrument used for the System Ensure that the sample cup was filled with ABS from this same
Check. lot number.
Using an old ABS sample. ABS sample older than one hour should be discarded and
replaced with a fresh ABS sample.
Using a new lot of ABS/CHK and Display the Daily Maintenance screen and enter the value from
failing to enter ABS/CHK carton the ABS/CHK carton flap into the Carton Value field for the
flap value into the software. ABS/CHK lot.
Loose tubing on the syringe. Check the tightness of all tubing connections on the sampler
pump panel.
Sample probe is misaligned. Realign the sample probe. See “Sample Probe Alignments” on
page 29 of Chapter 6: Alignments.
Plugged sample probe or dirty Clean the sample probe and drain. See “Cleaning the Sample
sample drain. Probe and Drain” on page 35 of Chapter 5: Maintenance.
Loss of water to the sample Replace any tubing that appears to be damaged. Tighten (or
drain. replace) any loose sample water tubing. If the sample drain
fitting is broken, contact the Customer Care Center - Technical
Solution: Inside U.S.: 1-800-441-9250; Outside U.S. Refer to the
local support team.
Sample syringe plunger tips are Replace both the 100-µL and the 2500-µL sample syringes on
not functioning properly. the sample pump panel. See “Replacing a Pump Syringe” on
page 96 of Chapter 5: Maintenance.
7 - 11
Troubleshooting Dimension EXL with LM/EXL 200 System Operator’s Guide
• Mean = 10% of the assay value listed on the end flap of the
ABS/CHK carton ±4 mAU, SD ≤ 1.6
Rerun the system check. If the system check continues to fail, follow the
troubleshooting items listed in this section as appropriate and then rerun the
system check.
Table 7-6: System Check Failure and Resolution Table
High Mean for Wash Chemistry Wash Tubing Pinched Ensure that the tubing is not
Probe 1 and 2 pinched or cut.
7 - 12
Dimension EXL with LM/EXL 200 System Operator’s Guide Troubleshooting
Error Messages
The area near the top of the Dimension EXL system screen is used to show only
error messages. The system software has two screens, Active System Errors
and Error Log, to review errors that have occurred.
When the system detects an error condition, it displays an error message in the
error message area, and the alarm sounds.
The system can display only one error message at a time. If more than one
error condition occurs at the same time, the system uses a priority ranking to
determine which one to display.
If system processing stops, reset the instrument by pressing the Reset key to
clear the error message and resume processing.
An error message may have one or more minor error messages associated with
it. These minor error messages provide more specific information about why
the error occurred and are helpful with troubleshooting.
7 - 13
Troubleshooting Dimension EXL with LM/EXL 200 System Operator’s Guide
IMT Troubleshooting
IMT Results Troubleshooting
If the IMT results are questionable and no error message is indicated by the
instrument or test report slip, any of the following issues may have occurred:
• QC is out
• Failed delta check on the LIS
• Results are inconsistent for the patient‘s clinical picture
• Abnormal Anion gap
Follow the steps in this section to troubleshoot these situations.
1. Check for insufficient sample in the sample container.
2. Rerun the sample. If the results are still questionable, perform the basic
troubleshooting steps below for Na, K, and Cl results.
7 - 14
Dimension EXL with LM/EXL 200 System Operator’s Guide Troubleshooting
7 - 15
Troubleshooting Dimension EXL with LM/EXL 200 System Operator’s Guide
Clot Check also includes a bypass feature which permits the operator to bypass
clot check hardware and run the system without clot check functionality if
needed. Clot check must be re-configured within the software, and bypass
tubing must be installed.
7 - 16
Dimension EXL with LM/EXL 200 System Operator’s Guide Troubleshooting
8. Open the instrument front doors to remove the existing clot check
tubing.
Figure 7-4: Syringe Left Port
# Description
1 Left Port
3 100 µL Syringe
9. Remove the plastic sleeve from the end of this tubing and set it aside for
use later in this procedure.
10. Open the pump panel to access the tubing on the end of the clot detector
board.
11. Slide the plastic sleeve back and remove the tubing from the right and
left sides of the clot check metal fittings.
12. Remove the plastic sleeves from both pieces of tubing before removing
the tubings from the conduit.
7 - 17
Troubleshooting Dimension EXL with LM/EXL 200 System Operator’s Guide
13. Remove the sample tubing from the 100 µL syringe by pulling through
the conduit.
14. At the sample probe, gently pull on the sides of the sample cover and lift
it off the probe assembly.
Figure 7-6: Sample Cover Probe Assembly
15. Remove the sample tubing from the sample connection and pull the
tubing up through the conduit.
16. Press the black sheathed end of the new sample tubing onto the sample
probe connection.
17. Remove the plastic sleeves from the end of the sample tubing.
18. Route the other end of the sample tubing down through the corrugated
black plastic conduit until it comes out of the opening on the left side of
the clot check board.
19. Continue routing the tubing through the right side of the conduit all the
way through to the 100 µL syringe.
7 - 18
Dimension EXL with LM/EXL 200 System Operator’s Guide Troubleshooting
20. Slide the plastic sleeve back onto the tubing and connect onto the left
port of the 100 µL syringe.
21. Flare the sample tubing if needed. Flare the tubing by inserting the flare
tool from the Spare Parts kit into the end of the tubing that will be
slipped onto the fitting.
22. Slide the sleeve fitting over the tubing and left port of the 100 µL syringe.
Note: The clot check bypass tubing should look like the image
below.
Figure 7-7: Clot Check (with Bypass)
# Description
23. Close the instrument lids and select PAUSE to turn on the sampling
system.
24. Prime the sample pump by:
a. Selecting F4: SYSTEM PREP > F7: PUMP PRIME.
b. Move the cursor to the Cycles field and type 3 for the number of
cycles to be performed.
c. Select F2: SAMPLER to begin priming the sample pump.
25. Check the sample tubing for leaks while the system is priming.
26. Perform a system check and daily QC.
27. With the bypass tubing installed, perform the following steps:
a. Close the instrument lids and select PAUSE.
b. Display the Pump Prime menu.
c. Move the cursor to the Cycles field.
7 - 19
Troubleshooting Dimension EXL with LM/EXL 200 System Operator’s Guide
7 - 20
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Chapter 8: Configuration
8-1
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Overview
The procedures in this chapter describe how to configure systems and reports
of the Dimension EXL system to meet the specific requirements of the
laboratory. Typically, these configurations are set at the time of installation
and may not need to be changed again. However, operators may want to set
up configurations and do other customizing to meet day-to-day changes in
workload and laboratory operation.
Siemens Healthcare Diagnostics provides validated instructions for using
certain Emit® assays. Contact the Customer Care Center - Technical Solutions
for a copy of the application sheets.
8-2
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Field Meaning
Enter Sample Data Indicates which type of Enter Sample Data screen, NORMAL (single samples) or
Mode BATCH (batch samples), displays for entering patient sample information.
*RMS Option ON indicates that RMS module is used. OFF indicates that RMS module is not
Configured used. Applies to the EXL with LM only.
*Automatic Rerun ON indicates that the system will rerun an assay if there were system errors
during the initial test processing that were resolved by the system. The assay
will be rescheduled without any operator intervention.
Note: This field must be set to ON to use the Automatic Dilution,
Automatic Repeat for Panics and Automatic Reflex Testing.
*Automatic Dilution ON indicates that the system will rerun the assay using a smaller sample
volume when autodilution conditions are met.
*Waste Out PLUMBED indicates that system chemistry wastes are automatically pumped to
an external waste collection system. MANUAL indicates that these wastes are
collected in the instrument waste bottle and must be emptied by the operator.
8-3
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Field Meaning
*Automatic Repeat ON indicates that an automatic rerun of an assay will be run if the test result is
for Panics outside the panic values entered by the operator for that method. See
”Automatic Panic Rerun” in this chapter.
*Automatic Reflex ON indicates that an automatic reflex test will be run if the test result is outside
Testing the reflex limits entered by the operator for that method. See “Automatic
Reflex Testing” on page 10 of this chapter.
*HM Module YES indicates that the HM module is available for use. Both HM fields must be
Configured set to ON.
LOCI Module ON indicates that the LOCI module is available for use.
Configured
Clot Check ON activates clot check for samples and enables the software to take actions if
Configured a clot is detected.
OFF indicates that the instrument will not perform clot check when aspirating
samples. Only the hardware monitoring will be active. The clot check icon will
turn yellow.
8-4
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
*F1: METHOD Enter/view method-specific information for each assay run on Dimension EXL
PARAM system. See “Entering Method Parameters” on page 33 of this chapter.
F2: DEFINE PANELS Create ten panels with up to 20 assays on each panel. All 20 assays can then be
requested with a single keystroke. See “Creating Panel Keys” on page 43 of this
chapter.
*F3: DATE/TIME Change the date and time. To change the date and time
1. Select F5: DATE/TIME.
2. In the Date field, type the day.
3. Select F2: NEXT MONTH to select the month.
4. Type the year.
5. Move the cursor to the Hour and Minute fields and enter the time.
Note: Use the 24-hour time convention for the hour.
6. To display the date/time as MM-DD-YY rather than DD-MM-YY as in steps 2-3,
select ENTER.
Note: The date time display can not be changed if data is already
entered.
7. Select F1: STORE CHANGES before exiting this screen.
*F4: COMPUTER Set up the Dimension EXL system for the laboratory computer operations. See
“Entering Sample ID Information” on page 58 of this chapter.
To establish communications with an LIS:
1. Select F4: COMPUTER > F4: COMMUNICATIONS.
2. Move the cursor to each field on the screen and select ENTER to change its
setting to the appropriate information for the laboratory.
3. Select F2: STORE CHANGES to store changes before leaving this screen.
F5: IMT ON / OFF Configure the IMT system. Press this function key and the IMT status box at the
top of the screen will change appropriately. “IMT Not Config” in the box
indicates that the IMT is OFF.
F6: SELECT PRINTER Configure the printer, creating the header and setting paper specifications for
printed test reports, and if applicable, creating a custom format for an external
printer. See “Configuring the Printer” on page 51 and “Customizing an External
Printer Report” on page 52 of this chapter.
*F7: PASSWORD Change the system password by pressing the function key and follow the
messages as they display.
F8: MORE OPTIONS Displays the System Configuration Secondary Menu screen.
1. Select F1: CLOT CONFIG to go to the Clot Check Configuration menu.
2. Select F5: SAMPLE AUTO REMOVAL to change sample auto removal time.
3. Select F8: TEMPERATURE to override system temperature warnings.
8-5
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Configuration Description
Field Meaning
Translate Clot OFF is the default selection and allows Clot Check Err to be
Check LIS Error transmitted through the LIS.
8-6
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
8-7
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
8-8
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
8-9
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Low and high limits The result is outside this range, either above or below it.
8 - 10
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
WARNING: As with any sample run on the Dimension EXL system, ensure
that there is enough sample in the sample container to run
not only the requested assays but also any subsequent reflex
assays that may be automatically run.
6. To set up a reflex test for another assay, press any arrow key to move the
cursor out of the Run field, then press the test key for the next assay.
8 - 11
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Low and high limits The result is outside this range, either above or below it.
8 - 12
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
8 - 13
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Note: A result with a high panic value on the test report slip has the
letters ”hp“ after it (or ”lp“ for a low panic value). Since two
results are shown because of the automatic panic rerun,
mean, SD, and CV for these results also appear on the test
report. All results are sent to the host computer.
Table 8-7: Automatic Reflex Testing Chart
Why was
Test test run? Test Result
1-CA Original test 13.1 exceeds both the panic and reflex limits for CA.
2-CA Hp rerun This test was run because the 1-CA result exceeded the
from 1-CA panic limits for CA.
3-PHOS Reflex test This test was run because the 1-CA exceeded the reflex
from 1-CA limits for CA.
This test was run because the 3-PHOS 1.2 result exceeds
reflex limits for PHOS. No additional reflex testing will be
4-MG Reflex test done for 4-MG because in this example the MG method did
from 3-PHOS not have a reflex test specified.
5-CA Statistics Printed because more than one CA test was reported.
8 - 14
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
8 - 15
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Note: Only those calculated results that are turned on will have
their changes stored.
Note: If the calculated result always appears with a Hi or Lo error
message, check the Interval field entry for that result. If a
calculated result is turned on but no interval is entered, all
reporting of that calculated result will contain a Hi or Lo error
because the system will use 0.0 - 0.0 as the interval.
Table 8-8: Calculated Results Equations and k Values
8 - 16
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
k2 = 0.056
k2 converts GLUC results reported in mg/dL to millimolar
OSMO Calculated
Osmolality concentration. If GLUC is reported in mmol/L, then k2 = 1.0
k3 = 0.36
k3 converts BUN results reported in mg/dL to millimolar
concentration. If BUN is reported in mmol/L, then k3 =1.0
9
The number 9 is a constant to account for all other osmotically
active solutes present in the solution.
The OSMO calculated result uses the GLUC result in the equation.
8 - 17
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
An unused Reference Interval fluid field, such as CSF/Blood for the BUN assay,
may contain the reference interval for a calculated result.
Table 8-9: Reference Interval Fluid Field
TP CSF/Blood GLOB
TU CSF/Blood FTI
8 - 18
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Calibration Configuration
Define Calibration Product
1. Select the Calib Alert key on the touchscreen.
2. Select F5: DEF CAL PRODCT.
An alternate procedure would be:
1. Select F5: PROCESS CTRL > F1: CALIBRATION.
2. Enter Password.
3. Select F5: DEF CAL PRODCT.
8 - 19
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Key Function
F1: NEW PRODUCT Use to clear the screen and enter additional products.
F2: DELTE METHOD Use to remove the highlighted assay from the list.
F3: CHG PROD NAME Use to enter a different name in the Cal Product field.
F4: CHG PROD LOT Use to enter a different lot number in the Cal Lot field.
F5: CHG PROD EXP Use to enter a different date in the Product Expiration field.
Note: To add tests to the product, use the test keys, then enter the
bottle values in the appropriate Level fields.
7. Select F7: STORE.
8 - 20
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Parameter Meaning
Intercept Close to zero or clinically insignificant. (Intercept 0.000-0.000 is the default value
that turns off intercept as an auto acceptance parameter).
SD Mean Maximum allowable standard deviation of replicates for each calibrated level.
Max Neg Bias Maximum allowable negative bias (residual) for each calibrated level.
Max Pos Bias Maximum allowable positive bias (residual) for each calibrated level.
8 - 21
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
10. When F8: SHOW ACTIVE is selected, these print options are displayed:
Table 8-12: Print Options
Key Function
F6: PRNT ALL PARMS Prints all parameters for all assays listed on the screen in
alphabetical order.
F7: PRNT SEL PARMS Prints parameters for the highlighted assay.
F8: PRINT ACT LIST Prints the list displayed on the screen.
8 - 22
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Correlations
Some situations may require that the Dimension EXL system and an alternate
system or method provide comparable results. If it is necessary to correlate the
Dimension EXL system to the alternate method, the Correlation feature
provides a way to adjust results by applying slope and intercept values derived
from a correlation between the two methods.
The correlation feature uses simple linear regression to calculate the
correlation curve y = mx + b where x is the expected result and y is the
observed result. When you store the correlation curve, the applied slope (mA)
and intercept (bA) are calculated as mA = 1/m and bA = -b/m. The applied slope
and intercept, stored in the Correlation Entry screen, are used to calculate a
correlated result using the follow.ing equation: correlated result =
(mA)(uncorrelated result) + b
When the correlation is performed offline, the applied slope and intercept
values (as calculated above) are entered in the Correlation Entry screen.
Correlation is applied only to patient samples and QC. Correlation is not
applied to calibration; therefore, do not adjust calibrator bottle values when
you calibrate a correlated method.
Correlation Study
A split sample correlation study should be run using at least 20 patient samples
spread throughout the assay range of the assay. Do not use quality control
material or calibrators as the sample. Use the observed results from the study
to calculate slope and intercept values for the assay.
8 - 23
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
8 - 24
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Field Enter
Sample Sample ID
Note: When entering data, only the same number of decimal places
as shown on the Method Parameters screen for an assay are
permitted.
6. After entering all the sample results, select F7: SHOW CORREL’D.
Note: The screen now shows the expected and correlated data and
a linear regression plot for the correlated data.
7. Review the information on this screen.
a. Expected and correlated values should now be nearly identical.
b. Statistics should be close to the following:
slope = 1.000
intercept = 0.000
r = 1.000
8. Decide if the information on this screen is acceptable for the laboratory
operations.
a. If the information is acceptable, skip to Step 10.
b. If the information is not acceptable, continue with Step 9.
Note: To delete all data; place the cursor on the line where the first
data entry mistake occurs and press F4: DELETE TO END.
8 - 25
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
9. Look for possible discrepant result pairs. Delete discrepant result pairs as
determined necessary. To delete a discrepant result pair:
a. Select F7: SHOW OBSERVED.
b. Move the cursor to the result pair line.
c. Select F3: DELETE LINE.
d. Select F7: SHOW CORREL’D.
e. Return to Step 7.
10. Select F6: CORRELATED TO and enter the name of the method or
instrument to which will be correlated the Dimension EXL system.
11. Select F7: SHOW OBSERVED > F8: PRINT to produce a report of the data
entered.
Note: This data will not be able to be viewed after the correlation is
stored.
12. Select F7: SHOW CORREL’D.
13. Select F8: PRINT to produce a report of the correlated data.
14. Select F3: ACCEPT CORR to store the information for the method.
15. Select EXIT to view the Correlation Entry screen.
16. Confirm that the assay range from the Siemens assay IFU is entered in
the Method Parameters screen.
17. Establish and enter new reference interval, panic values, and reflex limits
for the correlated assay.
18. Establish new QC ranges for the correlated method.
8 - 26
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Removing a Correlation
If the correlation data for an assay can no longer be used, simply change the
slope and intercept values to 0 (zero):
1. Select F5: PROCESS CTRL > F2: CORRELATION.
2. Type the password and select ENTER.
3. Move the cursor to the method, or press the test key for the method.
4. Enter 0 for the slope value and select ENTER.
5. Enter 0 for the intercept value and select ENTER.
6. Select F2: STORE.
7. Confirm that the assay range from the Siemens assay IFU is entered in
the Method Parameters screen.
8. Establish and enter new reference interval, panic values, and reflex limits
for the assay.
9. Establish new QC ranges for the assay.
Correlations Date
The F5 function key on the Correlation Entry screen, toggles the display
between “Correlated To” and “Date Correlated”. Selecting
F5: SHOW COR DATE, shows the correlation date where applicable.
8 - 27
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
8 - 28
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Cursor Keys Move the cursor to the assay to hydrate and then enter the
number of tests to hydrate.
Test Keys Press the test key for the assay to hydrate and then enter the
number of tests to hydrate.
Panel Keys Press a predefined panel key, P1–P10, and then enter how
many times this panel will be hydrated.
Pre-programmed Setup Lists Press F1: LOAD SETUP 1 or F2: LOAD SETUP 2 for the
pre-programmed setup list.
3. Select ENTER.
4. Repeat Step 2 through Step 3 until all hydrations have been entered.
5. Select F4: HYDRATE NOW.
Note: If a Timed Hydration is scheduled, and F4: HYDRATE NOW is
pressed, the entire pre-programmed setup will begin to
hydrate.
8 - 29
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
ONLY Hydrations will be performed only on each day through the next
Friday.
Mon to Fri
WEEKLY Hydrations will be performed Monday through Friday of each
week until the hydration period is changed.
ONLY Hydrations will be performed only on each day through the next
Sunday.
Mon to Sun
WEEKLY Hydrations will be performed each day until the hydration period
is changed.
8 - 30
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
8 - 31
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
IMT Configuration
Two items can be customized related to IMT processing for laboratory needs:
• Include ECO2 in Lytes test method key
• Specify the time interval for IMT bleach/condition soak
8 - 32
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
8 - 33
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Standard Vol This field name can be either Standard Vol or Selected Vol.
or Selected Vol When Standard Vol appears, the volume on the screen is the
recommended standard volume of sample for the assay as indicated
in the method IFU.
Selected Vol displays if a volume different from the recommended
standard volume on the method IFU was entered. This volume could
be either:
• A Siemens approved reduced sample volume for the assay, or
• A decreased volume (not approved by Siemens) that is used by
the laboratory. When such volume is used, perform assay
validation in the laboratory for CLIA compliance
Calibration Vol This field cannot be changed. It appears under the Standard Vol field
only for those assays (such as C3, C4, IGA, IGG, IGM) that must use
this volume for their calibrations.
8 - 34
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Auto Dilute Vols Use the keyboard or keypad to enter values for both serum/plasma
and urine samples. Refer to the tables on the pages that follow for
the recommended autodilute AD sample volumes for each
autodilute assay.
Intervals Use the keyboard or keypad to enter the upper and lower limits for
the reference, assay range, and panic intervals for a fluid.
• Reference Intervals:
May be adjusted according to laboratory normal
population.
• Assay Intervals:
Set by Siemens. Only use assay intervals published in
each method’s IFU.
• Panic Intervals:
Can be set using upper and lower limits or can be single
sided. Enter a lower limit to automatically rerun test
results that are lower than this limit. Enter a higher limit
to automatically rerun test results that are higher than
this limit.
Reflex If < or > Use the keyboard or keypad to enter a less than and/or greater than
value which, if exceeded, will trigger the automatic reflex test
entered in the Run field.
Run Use the test keys to enter the method to be automatically reflexed if
the test result for this assay is outside the reflex limits entered in the
Reflex If field. Only one assay can be entered in this Run field.
LOT ID and (C0 - C4) The lot ID and calibration coefficients (C0-C4) are entered
automatically in these fields by the software following the
calibration/verification of the lot. Some coefficients are constant for
certain assays. Refer to the assay IFU for more information. The
system can hold a maximum of two lot numbers per assay in system
memory.
8 - 35
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Automatic Dilutions
The system can perform two types of automatic dilutions:
• Automated Urine Dilutions (AUD) – for BUN, CREA, CRE2, EZCR, PHOS, and
URCA
• Autodilute (AD) – user programmable for urine and serum/plasma samples
Autodilute (AD)
Once the autodilute Method Parameters are entered and the system is
configured to autorerun and autodilute, the Dimension EXL system
automatically performs sample dilutions whenever the calculated result is
above the assay range, or the absorbance of the sample and reagent exceeds
the upper absorbance limit for the assay. To perform the automatic dilution,
the system aspirates a reduced sample volume (similar to the way a manual
dilution would be made). The dilution ratios are based on incremental sample
volume adjustments.
The Dimension EXL system does not perform dilutions based on the Dilution
field on the Enter Sample Data screen – this field is used for convenience when
the instrument calculates the results of the manual dilutions. (If a number is
entered in the Dilution field, the Dimension system will not autodilute the
sample.) For HCG and LHCG and the above AUD only, entering a number in the
Dilution field will not disable automatic dilution.
To configure the Dimension EXL system autodilute feature, first activate both
the autodilute and autorerun features on the System Configuration screen.
1. If either the Automatic Rerun or Automatic Dilution fields is off, move the
cursor to the field and select ENTER to change the setting to ON.
2. When both fields are on, select F1: METHOD PARAMETERS. Type the
password and select ENTER.
8 - 36
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
3. Select the test key of the first method to edit and highlight the first
autodilute volume field.
4. Enter the proper autodilute volume for each sample type. Select
F4: STORE PARAM’S or F5: STORE & PRINT before exiting or choosing
the next method.
Note: To use the Autodilute (AD) feature, the Autorerun and
Autodilute fields on the System Configuration Menu screen
must both be set to ON.
Note: If the system automatically dilutes a sample, the word
“Diluted” appears next to the assay on the printout. If the
sample was diluted and the result is still above the assay
range for the assay, the message “assy rng/dilu” will appear
next to the assay. In this case a manual dilution should be
performed to determine the correct result.
8 - 37
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Standard Recommended
Dilution Other Qualified AD
Method Sample AD Sample
Factor Sample Volumes (µL)
Volume (µL) Volume (µL)
CA 5 2 2.5 3, 4
GLUC 3 2 1.5 —
MALB 17 2 8.5 —
UCFP 10 5 2.0 —
8 - 38
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
WARNING: As with any sample run on the Dimension EXL system, ensure
that there is enough sample in the sample container to run
not only the requested tests but also any subsequent reflex
tests that may be run automatically.
AHDL 3 2 1.5 —
ALB 5 2 2.5 3, 4
ALC 3 2 1.5 —
ALDL 3 2 1.5 —
ALP 7 3 2.3 2, 4
ALPI 7 3 2.3 24
ALTI 35 10 3.5 —
AMM 44 22 2.0
AMY 14 7 2.0 12
AST 40 20 2.0 2, 5, 10
BUN 3 2 1.5 —
CA 5 3 1.7 2, 4
CCRP 12 2 6.0 —
CHOL 3 2 1.5 —
CKI 14 2 7 —
CRBM 3 2 1.5 —
CRE2 20 10 2.0 —
a = The HCG and LHCG methods use an additional 10-fold dilution prior to taking the auto
dilute sample. Therefore a 2 µL autodilute sample for HCG or LHCG is equivalent to a 1:200
dilution. The instrument does calculations automatically.
b = For autodiluted HCG and LHCG results > 200,000 mIU/mL [IU/L], prepare a manual
dilution and enter the appropriate dilution factor on the Enter Sample Data Screen. The
instrument does calculations automatically.
8 - 39
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
CREA 20 10 2.0 5, 8, 15
CRP 3 2 1.5 —
DBI 10 5 2.0 —
ETOH 9 3 3 —
EZCR 6 2 3.0 —
FERR 40 2 20.0 —
FOLA 10 2 5.0 —
GENT 3 2 1.5 —
GLUC 3 2 1.5 —
HCG,a,b 40 2 200.0b —
IBCT 25 12 2.0 —
IGA 10 2 5.0 —
IGG 10 2 5.0 —
IGM 10 2 5.0 —
IRON 40 20 2.0 —
LA 4 2 2.0 3
LDI 8 2 4 —
LHCGa,b 40 2 200.0b —
LIDO 6 2 3.0 —
LIP 4 2 2.0 —
LIPL 3 2 1.5 —
LMMB 60 30 2.0 —
MBI 20 5 4 —
MMB 60 30 2.0 —
MYO 20 2 10.0 —
PHNO 4 2 2.0 3
PHOS 3 2 1.5 —
PTN 4 2 2.0 3
a = The HCG and LHCG methods use an additional 10-fold dilution prior to taking the auto
dilute sample. Therefore a 2 µL autodilute sample for HCG or LHCG is equivalent to a 1:200
dilution. The instrument does calculations automatically.
b = For autodiluted HCG and LHCG results > 200,000 mIU/mL [IU/L], prepare a manual
dilution and enter the appropriate dilution factor on the Enter Sample Data Screen. The
instrument does calculations automatically.
8 - 40
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
SAL 15 5 3.0 8, 10
T4 16 4 4.0 3,5,7,10
TBI 10 5 2.0 —
TGL 4 2 2.0 —
THEO 4 2 2.0 3
TOBR 3 2 1.5 —
TP 15 8 1.9 5, 7, 10
TPSA 40 2 20 —
UCFP 10 5 2.0 —
URCA 17 5 3.4 —
VB12 12 4 3.0 —
VALP 3 2 1.5 —
a = The HCG and LHCG methods use an additional 10-fold dilution prior to taking the auto
dilute sample. Therefore a 2 µL autodilute sample for HCG or LHCG is equivalent to a 1:200
dilution. The instrument does calculations automatically.
b = For autodiluted HCG and LHCG results > 200,000 mIU/mL [IU/L], prepare a manual
dilution and enter the appropriate dilution factor on the Enter Sample Data Screen. The
instrument does calculations automatically.
8 - 41
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
WARNING: As with any sample run on the Dimension EXL system, ensure
that there is enough sample in the sample container to run
not only the requested tests but also any subsequent reflex
tests that may be automatically run.
Reduced Recommended
Dilution
Method Sample Volume AD Sample
Factor
(µL) Volume (µL)
ALTI 20 10 2
AMY 10 5 2
AST 20 10 2
CREA 15 10 1.5
GGT 15 10 1.5
TP 10 5 2
URCA 10 5 2
WARNING: As with any sample run on the Dimension EXL system, ensure
that there is enough sample in the sample container to run
not only the requested tests but also any subsequent
autodiluted tests that may be automatically run.
8 - 42
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
8 - 43
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
HIL Feature
Note: If the Dimension EXL system is connected to an LIS, ensure
that the LIS is able to receive results for tests it did not request
originally. Check with the local LIS consultant.
The HIL feature, which is based on the spectral characteristics of a serum or
plasma sample, provides an index that can alert to potential interference from
hemolysis, icterus, and lipemia in the sample, where:
H = hemoglobin resulting from lysis of red blood cells
I = icterus resulting from endogenous bilirubin
L = lipemia or turbidity caused by insoluble lipids
The HIL feature can be programmed to run automatically in an Operating
Mode that best suits the needs of the laboratory, or by request on individual
samples. Once programmed, the instrument automatically pipets 20 µL of
sample into an empty cuvette along with system water. Spectral absorbance
measurements, taken at 405nm (hemoglobin), 452nm (bilirubin), and 700nm
(turbidity), are used to generate a sample-specific HIL Index. The HIL Index
appears on the report slip as a three-digit value where:
• 1st digit = H index
• 2nd digit = I index
• 3rd digit = L index
Each index value correlates to an approximate concentration range in mg/dL
for each of the potential interferents, as specified in Table 8-20. HIL is not
processed if the maximum of 36 tests for a sample are ordered.
Table 8-20: HIL
H I L
Index
Hemoglobin (mg/dL) Bilirubin (mg/dL) Intralipid ™ (mg/dL)
1 H ≤ 25 I ≤ 2 L ≤ 25
8 - 44
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
The following general conditions are required for HIL to be run automatically:
• Sample must be serum or plasma.
• Sample must be undiluted (dilution does not eliminate spectral
interference).
• Operating mode must be set to ON or AUTO-ON.
• If the operating mode is AUTO-ON, Alert Index values between 2 and 6
must be entered in the HIL Setup screen.
• Sample Mode must be primary tube, bar code tube, sample cup or SSC (as
specified in the selected operating mode).
• HIL will not run automatically in the "limited cup - no level sense" (sample
volume may be limited).
8 - 45
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
HIL Setup
To set up the HIL feature, select an Operating Mode. The selections for
operating modes are listed in Table 8-21. If the mode is "AUTO-ON -Tubes,
Cups, SSCs" or "AUTO-ON -Tubes, Cups", also enter HIL Alert Index values.
WARNING: As with any sample run on the Dimension EXL system, ensure
that there is enough sample in the sample container to run
not only the requested tests but also any subsequent tests
that may be automatically run.
To specify parameters for HIL setup:
1. Select F5: PROCESS CTRL > F8: MORE OPTIONS.
2. Enter Password and select ENTER.
3. Select F6: HIL SETUP.
4. Select F1: NEXT MODE to display the Operating Mode that best suits
laboratory needs.
Table 8-21: Operating Modes
Operating Modes
AUTO-ON -Tubes, Cups, HIL is run automatically when an assay with an alert index between 2 and
SSC's 6 is requested and the sample mode is primary tube, bar code tube,
sample cup or SSC. The test report displays the "HIL Interf" message for an
affected assay if any of the measured HIL index values is greater than or
equal to the corresponding Alert Index for that assay.
AUTO-ON Tubes, Cups HIL is run automatically when a method with an alert index between 2
and 6 is requested and the sample mode is primary tube, bar code tube or
sample cup. The test report displays the "HIL Interf" message for an
affected assay if any of the measured HIL index values is greater than or
equal to the corresponding Alert Index for that assay.
ON HIL is run automatically and reported for all serum and plasma samples
when the sample mode is primary tube, bar code tube, sample cup or
SSC. The test report does NOT display the "HIL Interf" message for an
affected assay.
OFF HIL is not run automatically for any sample. However, the operator can
request HIL by including it in the list of tests requested by LIS or by
pressing the HIL test key when ordering tests manually. The test report
displays the HIL index, but does NOT display the "HIL Interf" message for
an affected assay.
8 - 46
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
8 - 47
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
8 - 48
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
4. Select EXIT.
Table 8-22: Status Configuration Screen Fields and Selection Choices
Field Choices
Sound audio alarm when any STAT OFF alarm will never sound
alert is triggered ON alarm will sound for all configured alerts
Alert when a STAT sample is OFF alert key will not turn yellow when sample processing is
available finished.
ALWAYS alert key will turn yellow when any STAT sample is
finished processing.
SELECTABLE allows the operator to select a specific sample and
select the F8: Complete Alert key:
• If the sample line is white (alert OFF) when F8 is selected, it
turns yellow while processing and blue when finished.
• If the sample line is yellow when F8 is selected, it turns white
(alert OFF).
Alert when a STAT sample is OFF alert key will not turn red when sample processing is finished
available but has an error with no with an error and no result
result ON alert key will turn red when sample processing is finished with
an error and no result
Alert when a STAT has been OFF alert key will not turn red for samples not started
entered but not run in a specified ON alert key turns red if a sample does not start processing within
time the specified time period
Time from STAT entry to alert - in Can be used only if the previous entry is ON. Enter the number of
minutes minutes (1-60) a STAT request can wait before the Not Started
alert is triggered.
Time for STAT alerts to be Enter the number of minutes (1-120) that an alert will be
displayed - in minutes displayed on the STAT Samples screen. After the time has elapsed,
the information is removed from the display.
8 - 49
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Configuring QC Alerts
1. Select QC Alert.
2. Select: F4: CONFIG ALERTS.
3. For each parameter, move the cursor to the required field and select
ENTER to toggle between ON and OFF.
4. Enter a number between 1 and 240 (minutes) to define the time interval
for expiration.
Note: This is done only if the operator turns ON the Alert when QC
expires.
5. Select Exit to store changes
8 - 50
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Field Explanation
Report Title This creates the header that appears on your printed test report.
Enter a title for the report (up to 30 characters).
System Printer Indicates the status of the system printer. Select F1: SYS PRINTER
to select another status.
Spacing between reports Enter the number of centimeters of blank paper that will be
between printed test reports.
Minimum number of lines Enter how many test result lines will appear in all test reports.
External Printer Indicates the status of the external printer. Select F2: EXT PRINTER
to select another status.
8 - 51
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Header Description
Report Title & Banner The date and time the report was printed are placed on the last line of the
label.
Patient Header The label section fields can be chosen from any information on the Enter
Information Sample Data Screen
Test Result Line - Header This section consists of only two lines, both use the same line and column
positioning data. This section sets the headers for the data.
Test Result Line - Data The Data Item column contains results information that can be selected for
the report.
Note: The field “Chemistry” means that the full name of the method
(such as Triglyceride) will be printed on the report; the field
“Test” means that the method abbreviation will appear on
the report.
8 - 52
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
The above settings should produce the output shown in Figure 8-4.
8 - 53
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Note: Follow the dashed lines to view how the line and column
specifications relate from the format to printout.
Note: Use “-1” in the Column field to prevent printing.
Note: To align the title entries use spaces. The “^^^^” in Figure 8-4
indicate spaces added using the space bar to center the titles.
8 - 54
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Limits Side
Includes an A column and a B column, depending on whether the assays use
one or two monitoring checks. The fields in each column define the limit using
either percentage or SD limits. Above Mean Factor numbers are given as a
percentage factor of the mean absorbance result. For the Mean Plus/Minus SD
fields, the number entered is the factor times the SD to define the limit around
the mean.
8 - 55
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Assay A B A B A B
8 - 56
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Assay A B A B A B
8 - 57
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
8 - 58
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Field Explanation
Segmented Wheel Setup The Mode field on blank Enter Sample Data screens defaults to this.
Default Mode Use F8: NEXT SETTING to select other modes.
Default Fluid The Fluid field on blank Enter Sample Data screens defaults to this
entry. Use F8: NEXT SETTING to select other fluids.
Sample Edit Indicates the type of sample container that will be assigned to all
downloaded samples that are assigned segment positions by the
operator from the Sample Status screen. Use F8: NEXT SETTING to
select a sample container type.
If this field is set to No Default, sample positions cannot be assigned
for downloaded samples using the Sample Status screen.
Pediatric Tube Segments Indicates those segments that have been configured for Pediatric
tubes. Move the cursor to this field, select F5: ADD SEGS, enter the
segment letters and select ENTER, then select F7: STORE. The Mode
field on the Enter Sample Data screen defaults to PED Tube when
positions in these segments are entered.
PED Tube I.D. (mm) The inside diameter in millimeters of the PED tube used by the
laboratory.
SSC Indicates segments that have been configured for SSC use. Move the
cursor to this field, select F5: ADD SEGS, and use the keyboard to
enter the segment letters. Select ENTER, then F7: STORE. The Mode
field on the Enter Sample Data screen defaults to SSC when positions
in these segments are entered.
SSC I.D. (mm) The inside diameter in millimeters of the SSC. The inside diameter of
the SSC is 8 mm.
8 - 59
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Field Explanation
Priority Panel (schedule) Allows specific assays within samples to be given a higher scheduling
priority than the sample itself. This enables critical assays to be
processed in a STAT-like manner while the remainder of the assays for
the sample are processed with the priority given on the Enter Sample
Data screen (or downloaded through a host computer). When a
priority panel schedule is requested on a sample, the system raises the
priority of the assays in the designated panel one level above the
other assays requested on that sample. Enter the number of the panel
key defined for priority scheduled tests. Then select “STAT only” or “All
samples.”
Note: The priority scheduling option is not active for QC
samples or samples that include multiple test requests
of the same assay. All tests in the priority panel must
be ordered for scheduling to occur.
• If “STAT only” is selected, the priority panel is performed only on
samples that were entered as STAT.
• If “All samples” is selected, the priority panel is performed on all
samples, regardless of their priority.
Priority Panels:
The Priority Panel (schedule) and Priority Panel (report) fields do not
need to have the same information in them. Define different panels
for use in each of these fields.
8 - 60
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Field Explanation
Priority Panel (report) Enables the system to print a test report for a sample as soon as all the
assays in the panel are completed.
All assays in the priority panel must be ordered for the report to print.
Before enabling priority panel reporting, ensure that the host
computer is capable of receiving two reports for the same sample. If
the operator is not sure of this, check with the local host computer
consultant.
In addition, when the remaining assays of the sample are completed,
another report is prepared that includes all assays (including the
priority panel assays) requested on the sample. The priority of the
priority panel report printout is STAT, so that it can be distinguished
from the final sample report.
Note: The priority reporting option is not active for QC
samples or samples that include multiple test requests
of the same assay. Enter the number of the panel key
defined for priority report tests. Select “STAT only” or
“All samples” as discussed above for Priority Panel
(schedule).
Bar Code Configuration The test panel that is run on a barcoded sample if the system cannot
Default Test Panel find any patient data for the sample. Use the keyboard to enter a
panel number. If a default panel is not selected, enter a zero in this
field.
Processing Whole Blood Samples: Do not configure a default panel if
whole blood samples are processed. If a default panel is not
configured, deconfigure the option to avoid the potential for
inappropriately processing the default panel tests from a whole blood
sample.
Download Pretreats Indicates whether or not samples with pretreatment tests can be
downloaded from the laboratory LIS. Use F2: PRETREATS to select
YES or NO.
Bar Code Label Format The format of the barcode used by the laboratory. Use F8: NEXT
SETTING for other formats recognized by the system.
Label Length The field length of the barcodes used by the laboratory. This field can
be zero or a number from 3 to 12.
Leading Zeros Indicates whether or not the system recognizes zeros that precede a
barcode. If YES is selected, enter all the zeros that precede the
barcode when entering a Sample No. on the Enter Sample Data
screen. Use F3: ZEROS’S YES/NO to make a selection.
Translate TCO2 to ECO2 If turned ON, the system interprets an LIS request for assay TCO2 as a
request for tassayECO2.
Translate Low Volume If turned ON, the system interprets LIS requests for assay HCG, MMB
Test Method Names and TNI as requests for LHCG, LMMB, and LTNI when using the
reduced test Flex reagent cartridges for those assay.
8 - 61
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Fields Explanation
Total Results Includes all patient tests that produced a reportable result. Test results
with an “aborted test,” “no reagent,” or “not calibrated” processing
message are not counted because they did not consume reagent.
Conditional Results Includes all patient tests that produced a reportable result and included
an error code with the result. The Conditional Results count is included in
the Total Results count.
QC/Calib Includes all QC and Calibration tests that produced a printed result and
consumed reagent. The QC/Calib count is not included in the Total
Results count.
Last Saved The count that was in the Total Results column when F8: SAVE COUNTS
was last pressed.
Tests Since Last Save The difference between the Total Results column and the Last Saved
column.
Last Save Date The date and time when F8: SAVE COUNTS was last pressed.
8 - 62
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
5. Select F2: CHANGE DATE RANGE and record the new start date.
6. Select ENTER. Record the new end date and select ENTER.
7. The system will display a storing data and checking data integrity
message.
8. Select F1: STORE DATA to store the date to the USB drive. When finished,
disconnect the device and label properly.
Note: The USB drive is not intended to be used as a storage device,
but only as a transfer device for data.
8 - 63
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
8 - 64
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Field Description
SerumQC1
SerumQC2
QC level
SerumQC3
UrineQC1
UrineQC2
Prod Name The name entered by the operator in the Patient Name field (on Enter Sample Data
screen) or downloaded from the LIS.
QC Lot # The lot number entered by the operator in the Sample No. field (on Enter Sample
Data screen) or downloaded from the LIS.
Flex Lot # Lot number of the Flex® reagent cartridge used in the QC run.
Note: LYTE will appear in this field for Na, K, and Cl.
Inst ID Unique identifier entered in the Instrument ID field in the Communication Set Up
screen.
8 - 65
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Field Description
Patient Name Name of the patient entered in the ENTER DATA screen or downloaded from LIS.
Sample No. Sample number entered on the ENTER DATA screen or downloaded from LIS.
Inst ID Unique identifier entered in the Instrument ID field in the Communication Set Up
screen.
8 - 66
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Field Description
Prime Date Date and time that the pump was primed.
HM Wash If no HM:
SD/IMT Dil SD Dil 10% value ±2 mAU
SD ≤ 1.4
8 - 67
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Field Description
ABS/CHK Lot Contains the lot number of the CHK or ABS lex
Carton Value Contains the user entered bottle values of CHK or ABS
Empty Date Date and time that the cuvette waste was emptied.
Waste Emptied Shows Yes or No to indicate whether the cuvette waste was emptied or not.
8 - 68
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Field Description
Calc Type Selection options include LINEAR, VERIFY, POWER, LOGIT, EXPLOG
C0 Coefficent 0
C1 Coefficent 1
C2 Coefficent 2
C3 Coefficent 3
C4 Coefficent 4
Trip Level Options include: trip point, extended read vs, short read
Scaler B
Scaler C
Scaler D
8 - 69
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Field Description
SD SD of level 2 results.
SD 3 SD of level 3 results.
SD 4 SD of level 4 results.
SD 5 SD of level 5 results.
8 - 70
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Field Description
Ref Range Hi High end of the QC reference range for this fluid level when the calibration was
accepted.
QC Out? Lo, Hi, or blank, depending on the result compared to the Ref Range.
Note: QC info is repeated 4 more times to accommodate 5 QC results.
8 - 71
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Field Description
Initials The initials of the person who stored the log. Will be XXX if stored automatically.
Dil Chk Passed Will either be DIL CHK PASSED in GREEN or DIL CHK FAILED in RED or blank if no
DIL CHK was run
Uncorr Na SD
Uncorr K SD
Cal Passed Will be either CALIBRATION PASSED in GREEN or CALIBRATION FAILED in RED
or blank if no CALIBRATION was run
K Slope 53-65
CI Slope -55-40
8 - 72
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Field Description
Std B Lot ID
Flush Lot ID
Diluent Lot ID
Sensor Lot ID
TC02 Lot ID
8 - 73
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Field Description
8 - 74
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
8 - 75
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
8 - 76
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
# Description
8 - 77
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
3. With the assay highlighted, select F5: ASSIGN KEY. Press the new keys to
assign to the assay. If the new key assignment causes an assay that will
be run in the laboratory to be unassigned, reassign it following steps 2
and 3.
4. Repeat Step 2 and Step 3 to program keys for other assays.
5. After changing the assays, select F8: STORE and enter a password.
Note: To return to the default key assignments in the software,
select F6: DEFAULT KEYS.
8 - 78
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
User-Defined Assays
This section contains all the information needed to understand and operate
the User-Defined Method feature on the Dimension EXL system. With this
feature operators can define their own assays using reagents that are not
purchased from Siemens. Operators can define these assays by choosing the
characteristics that fully describe an analysis. Up to ten assays can be defined
and stored.
Siemens provides validated instructions for using certain Emit assays. Contact
the Customer Care Center - Technical Solutions for the assay-specific
Application Sheet.
Assays defined with the User-Defined Methods feature are requested for
sample processing in the same way that Siemens assays are requested. They
can be processed along with Siemens assays in random, batch, or profile
modes.
Before using User-Defined Methods, operators should be familiar with the
basic operation of the Dimension EXL system because the procedures for
calibrating, performing quality control, and running patient samples are the
same as for Siemens assays.
To program and run a user-defined method, perform the following 12 steps in
sequence:
Table 8-35: Programming a User Defined Assay
1. Enter the assay reaction parameters on the “Identifying the Method” on page 80
User-Defined Method screen. “Defining the Reagent and Sample Deliveries” on
page 81
“Defining the Photometry Reading Times” on
page 84
“Defining the Flex Cartridge Configuration” on
page 85
2. Program the user defined calculations on the “Defining the Calculation” on page 87
template and calculation screens.
5. Store the assay reaction parameters and “Storing a User-Defined Method” on page 89
calculation routine.
7. Enter the assay parameters under the System “Entering User-Defined Method Parameters” on
Configuration screen. page 90
8 - 79
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Customer Responsibility
The Customer assumes all responsibility for the selection of the proper
reagents and entering the proper test parameters, use of the proper test
protocol, correctness of the test results, and any associated errors or
omissions.
Warranty
SIEMENS EXPRESSLY DISCLAIMS ALL WARRANTIES WITH RESPECT TO THIS
USER-DEFINED METHODS PRODUCT WHETHER EXPRESS OR IMPLIED,
INCLUDING WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE.
Since Siemens does not manufacture the reagents that our customers may use
in the user-defined Flex reagent cartridge, the warranty for the Dimension EXL
system does not extend to the performance of user-defined reagents
(including user-defined test results or standard Dimension EXL system test
results that are affected by user-defined testing), their effect on the system
operation and types and frequency of maintenance, or their effect on operator
safety.
WARNING: Never assign the same name to more than one method.
This could result in erroneous or misleading test results.
4. Press F1: NEXT SELECT. to select a mode.
8 - 80
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Field Information
Channel Ten channels (numbers 1–10) are available for user programming.
Name User-defined method names always start with an X. You can add
up to three additional alpha-numeric characters (XABC, XAB1, and
so on). Each user-defined method must have a different name.
8 - 81
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Reagent Delivery R1
1. Enter a delivery time (in seconds) for the reagent to be delivered. The
delivery time for R1 is preset and cannot be changed.
2. Select a component (Component 1) for the reagent delivery by pressing
F1: NEXT SELECT.
3. Enter the volume of Component 1 to be added to the cuvette.
4. If your method requires more than one component, repeat Step 2 and
Step 3 for Components 2 and 3 as necessary.
5. Enter the volume of chase to follow the reagent delivery.
6. Select a mix level for the reagent delivery by pressing F1: NEXT SELECT.
Note: All methods must specify a volume for at least one
component of R1.
Sample Delivery
1. Enter the volume of sample to be added to the cuvette.
Note: Sample volumes can range from 2 µL to 60 µL.
2. Enter the volume of chase to follow the sample delivery.
3. Select a mix level for the sample addition by pressing F1: NEXT SELECT.
8 - 82
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Field Explanation
Component A component is a reagent that is stored in a well of the Flex cartridge. The
cartridge may contain up to five different components, but a component may
be stored in more than one well. it is possible to add as many as three
components in any single reagent delivery. Components are designated by the
letters A–E and correlate to well positions 1–5 in the Flex cartridge. Refer to
“Filling a Flex Cartridge” for details on specifying the location of components
within a cartridge. No order of use or delivery is implied by these letters.
Volume Enter volume in increments of whole µL. The total sample volume delivered
(including chase) must be less than 100 µL. Total volume, V, in the cuvette
(includes reagent, chase, and sample volumes) must be between 350 and
500 µL. Siemens recommends that at least 15 µL of chase per reagent delivery
is used to ensure that the reagents are adequately washed from the probe.
Chase Chase is a diluent (water) supplied by the instrument to flush the reagent lines,
sample probes, or systems after an addition. Enter any amount of chase
(water) that is consistent with the volume limitations described above and the
mix strength described below.
Mix Describes how vigorously to mix the reagent components (or sample) and
chase. GENTLE, MODERATE, STRONG, or NONE are the available choices. The
mix strength involves a combination of how long the solution is mixed, how
much energy is used, and how many pulses are applied. For gentle mixes, the
ultrasonics pulse for longer times at a lower energy level, while strong mixes
occur over a shorter time period with a higher energy level.
When selecting a mix level:
• be careful to prevent foaming.
• consider the chemical properties (such as viscosity) of the reagents and
sample.
The reagent chase volume must be at least 20 µL when a reagent mix is
requested. The sample chase volume must be at least 10 µL when a sample
mix is requested.
8 - 83
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
1. Enter a time (in seconds) for the first photometry reading (P1).
P2 0 (zero) to 675 seconds, but must be later than the first reading (P1).
8 - 84
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
1. Enter the component location for each well in the Flex cartridge by
putting the letter identification of the component under the appropriate
cartridge configuration number (well number 1, 2, 3, 4, 5, or 6) where it
is located in the Flex cartridge.
a. To do this, position the cursor in the parentheses under the well
number and press F1: NEXT SELECT until the proper letter appears.
2. Enter the number of tests for each well. The total number of tests must
be equal for each component (but not necessarily for each well) within a
Flex cartridge.
3. Enter the Well Life, which is the number of hours that each well will
remain stable after the well has been punctured by the reagent probe.
WARNING: The Well Life, On-Board Life, and Calibration fields should be
carefully determined prior to routine use of a user-defined
method.
4. Enter the On-Board Life, which is the number of hours that the Flex
cartridge remains stable after it has been placed in the instrument.
5. Enter how often this method must be calibrated (in hours).
8 - 85
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Field Explanation
Component Reagent component designation and well location. Reagent components can be
located in one or more wells. Nominal Flex cartridge dead volume is 200 µL per well
when used with a user-defined method. The maximum volume of each well is 4.0
mL.
Relates to the number of test volumes in a cartridge well. This is the number of tests
available for all components in a cartridge. For example:
CARTRIDGE CONFIG 1 2 3 4 5 6
COMPONENT A B B C
NUMBER OF TESTS 20 10 10 20
Notice that, as required, components A, B, and C each have a total of 20 tests
Number available.
of Tests
Well Life The number of hours that a reagent (a component in a given well) remains stable
after the well has been punctured by the instrument. This can be up to 760 hours.
The default time is 72 hours.
On-Board The number of hours that reagents in a cartridge remain stable after the cartridge
Life has been placed on the instrument. This can be a maximum of 760 hours. The
default time is 720 hours (or 30 days).
The instrument maintains the cartridge at 2°–8°C as with all other reagents.
Calibration The maximum number of hours allowed between calibration. This can be a
maximum of 8760 hours (or 365 days). The default time is 2160 hours (or 90 days).
Calibration will remain valid within the same lot number of a method, as with
Siemens chemistries.
WARNING: Well Life, On-Board Life, and Calibration should be carefully determined prior
to routine use of a user-defined method. Use caution in determining the
calibration life with respect to variations in reagent preparation. Ensure that
the calibration will hold from cartridge to cartridge.
8 - 86
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
8 - 87
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
8 - 88
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Field Explanation
Measuring Filter 293, 340, 383, 405, 452, 510, 540, 577, 600, 700, or NONE.
Blanking Filter 293, 340, 383, 405, 452, 510, 540, 577, 600, 700, or NONE.
Depletion Factor The system calculates a P1/P2 rate and a P2/P3 rate. If these rates differ from
each other by more than a given percentage (the depletion factor you
entered), an absorbance error message will appear on the printed test
results. A depletion factor must be entered on three-point calculation
templates.
Dilution Indicates the weighting factor or multiplier that compensates for the dilution
of the sample. This applies when sample or reagent is added between
photometry readings.
For example:
• R1 (at –60.0 seconds) involved the addition of 370 µL to the cuvette.
• P1 is at –30.0 seconds (the volume in the cuvette is 370 µL at this time).
• S1 (30 µL) the addition of 30 µL of sample occurred at 0.0 seconds.
• P2 is 65.0 seconds after the sample was added
(Total volume in the cuvette is now 370 + 30 = 400 µL.)
The dilution value is 400 µL divided by 370 µL (1.081).
The dilution value may be positive or negative. An undiluted sample would
have a dilution value of:
• 0 or 1.0 for a rate method
• 1.0 for an end point method
Initial Optical No limitations. Enter an IOD (in mAU) in the numerical field.
Density (IOD) Next to this field are parentheses where the operator must indicate if they
want an absorbance error message to be printed on the test results report.
Select BELOW if the operator wants the message to display when results fall
below the IOD that was entered. Select ABOVE if the operator wants the
message to appear when results fall above the IOD that was entered. Select
INACTIVE if the operator does not want a message to be printed.
Final Optical No limitations. Enter an FOD (in mAU) in the numerical field.
Density (FOD) Next to this field are parentheses where to indicate if the operator wants an
absorbance error message to be printed on the test results report. Select
BELOW if the operator wants the message to display when results fall below
the FOD that was entered. Select ABOVE if the operator wants the message
to display when results fall above the FOD that was entered. Select INACTIVE
if the operator does not want a message to be printed.
8 - 89
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
CAUTION: Do not try to use an eight-well Flex cartridge because this will
damage the reagent probe.
8 - 90
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Field Information
Method Enter the same name that was already assigned to this method on the
Method Parameters screen
Sequence number A unique, five-digit number. This number must be different for all
reagent cartridges for this method.
8 - 91
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
To Request This
User-Defined Test Press this Key Combination
Method
X01 Control/GGT
X02 Control/GLUC
X03 Control/LDI
X04 Control/LYTES
X05 Control/NA/K
X06 Control/PHOS
X07 Control/TBI
X08 Control/TP
X09 Control/TGL
X10 Control/URCA
8 - 92
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
8 - 93
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Field Explanation
Method () The method name displays in this field when the method is selected. The
description “6 well” for the Flex cartridge is always displayed within the
parentheses.
Filters The 3-digit fields to describe the wavelengths to be graphed on the kinetics
screen.
1st field Shows which single wavelength to display on the graph. Press F6: FILTER
DISPLAY to change this field. To display more than one wavelength, choose
“bic” for bichromatic, or “---” to display all ten wavelengths.
2nd and 3rd fields Initially show the Measuring Filter and Blanking Filter wavelengths you
chose when programming the method. These wavelengths appear on the
graph if you chose “bic” for the 1st field. To change the first wavelength,
press F3: 1st BICHROME; to change the second, press F4: 2nd BICHROME.
If a single wavelength or all ten wavelengths are chosen, these last two
fields display the wavelengths programmed into the method. Changing
them does not affect what is displayed on the graph. If turbidimetric mode
is chosen when identifying the method, an asterisk (*) follows the Filters
field until the kinetics reaction has started. After the kinetics reaction has
started, the running time of the assay follows the Filters field regardless of
which mode is selected.
Scale mAU Describes the absorbance scale (in mAU) of the y-axis.
Filename A file name displays in the Filename field when the data is archived. For
user-defined methods, the file name is always “kinetics.ar” and the system
saves the most recent study. Press F5: ARCHIVE DATA to archive the data.
Marker Up to two markers (blue vertical lines, x and o) can be displayed on the
graph.These markers can also be moved during the reaction but only to the
point where the reaction has progressed. Move the cursor to the Marker
field and select Enter to select the marker(s) that you to be viewed or move.
Use the arrow keys to move the marker(s) that were selected. Markers can
continue to be moved throughout the reaction process.
mA, bic dmA mAU with respect to a bichromatic reaction, specifically the change (delta)
in mAU, x and o.
mA/min With respect to bichromatic reactions, this is the change (delta) in mAU per
minute.
8 - 94
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Note: Changing filters using F3, F4, and F6 does not affect the
programmed method. They only change the wavelengths
displayed on the Method Kinetics screen being viewed, but
do not change what has been programmed in the method
itself.
Note: If using an external printer, all the photometrics data from
the kinetics assay that was run will be printed when
F5: ARCHIVE DATA is pressed.
8 - 95
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Cartridge Configuration:
Well #
Component
# tests/well
8 - 96
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Reaction Kinetics:
Reaction Timeline
Event
Time
Volume: Component 1
Component 2
Component 3
(Reminder: Volume in the cuvette must be > 350 µL whenever a photometer reading is programmed.)
8 - 97
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
CAUTION: Do not load the Flex cartridge into the automatic loader until this
prompt displays.
6. Observe the messages as the method reaction progresses.
After the time clock has started, you can stop the reaction at any time by
pressing F1: STOP ASSAY.
You can also change the filter display during the reaction and the graph
automatically updates to the new wavelengths.
7. After the reaction is complete, you can:
a. Choose a print option:
1) Press CRTL+P key combination for a printout of the screen.
2) Press F8: PRINT (with 1 or 2 wavelengths displayed) for a
graph of the active wavelength.
3) Press F8: PRINT (with all wavelengths displayed) for a
complete data dump of all readings for all wavelengths.
4) Make note of any data that you want to edit into the
User-Defined Methods screen for that method.
8. Select EXIT to leave the Method Kinetics screen.
Note: The system displays a message to remove the
investigative-use Flex cartridge.
9. Remove and discard the investigative-use Flex cartridge.
8 - 98
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Programming Terms
The Dimension EXL system user-defined software application provides a
versatile method of specifying how raw photometric data is transformed prior
to conversion to analyte concentration through the method’s standard curve.
Examples of simple transformations would include bichromatic endpoints and
two-point rates. More complex transformations include polychromatic sums,
volume-corrected blanking for endpoint reactions, and checking for substrate
depletion through comparison of early/late reaction rates.
The capability provided by the transformation program is similar to the
functionality provided by a four-function calculator. The four common
arithmetic functions—add, subtract, divide, multiply—are provided, as well as
parentheses. Twenty-six temporary memory registers are provided for the
storage of intermediate results. Functions are available for calculating
monochromatic and bichromatic rates and endpoints from the raw
photometric data arrays.
Programming Structure
All programs must have a left brace ({) as the first character and a right brace
(}) as the last. At least one RETURN statement is required to forward
meaningful results from the calculation.
Enclosed within the braces are program statements. Statements are used to
assign intermediate results to registers, test logical conditions, set error codes,
and return transformed mAU to the system.
Implementation
The MAU Calculation screen provides the 25 lines in which to enter a
transformation program. The input is in the form of a simple programming
language, which provides functions to retrieve raw data from the photometric
data arrays, algebraic statements to transform that data mathematically, and
test statements (if/else) that can be used to set error flags. In addition, there
are 26 registers labeled A–Z that can be used for temporary storage of
intermediate results. The final result is RETURNED for input into the standard
curve transformation.
8 - 99
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Statements
Statements can take any of the following forms:
• Register = numerical expression;
• SET errorcode ERROR;
• RETURN numerical expression;
• { statement1; statement2; ...; statement N; }
• IF (logical expression) statement
• IF (logical expression) statement ELSE statement
Examples of these types of statements appear as follows:
8 - 100
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
8 - 101
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
The first statement calculates the rate of reaction between readings P1 and P2,
and assigns it to register A. The second statement calculates the rate between
readings P2 and P3, and assigns it to register B. The IF statement will flag an
absorbance error (if the later rate is less than 80% of the earlier rate) and
RETURN the value of the early rate as the transformed mAU value. Otherwise,
no error is set, and the later rate in register B is returned.
Note that the TRUE clause of the IF/ELSE statement contains a compound
statement, which is used to group the SET and RETURN statements into a
single clause.
IF (logical expression) statement ELSE statement
This statement is similar to the IF statement. However, the ELSE statement is
executed when the logical expression is FALSE.
Expressions
Two classes of expressions are possible, depending on the kind of value
produced by the expression. The first class, referred to as numerical
expressions, produce numerical results that can be assigned to registers or
returned as values. The other class, called logical expressions, produce logical
results from Boolean operations. Logical expressions are used exclusively
within IF statements. The output of a logical expression cannot be used to
assign values to registers, nor can it be returned.
Numerical Expressions
Numerical expressions allow the user to combine values from function calls,
registers, and constants together using the arithmetic operators *, /, +, and – .
The precedence of these operators corresponds to common usage, where *
and / have the same precedence, and are evaluated as encountered in
left-to-right order; + and – have lower precedence than * and /, and are also
evaluated in left-to-right order. Operator precedence and evaluation order can
be modified through use of parentheses, which can be nested. The following
are valid expressions:
a. (a) A + (B + C) * D
b. (b) (A + B) * 0.5
c. (c) MAU(P1,340NM) – MAU(P1,383NM)
d. (d) A/B * RATE(P1,P2,340NM,510NM)
The resulting value from a numerical expression is usually assigned
to a register or returned:
e. (e) RETURN A + (B + C) * D;
f. (f) C = (A + B) * 0.5;
g. (g) Z = MAU(P1,340NM) – MAU(P1,383NM);
h. (h) RETURN A/B
* RATE(P1,P2,340NM,510NM);
8 - 102
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
Logical Expressions
Logical expressions are used within IF statements, which allow a statement or
group of statements to be executed when the logical expression evaluates to
TRUE.
Logical expressions are composed of comparison operations, combined
together using the AND and OR operators. (No NOT operator is provided, since
the comparison operations can be structured to effect the NOT operator).
Comparison operators include > (greater than) , < (less than), and two equal
signs == (equal to). Numerical expressions serve as operands for these
comparison operators. (Note that testing for equality is strongly discouraged,
since floating point calculations often experience rounding errors, which cause
equality comparisons to fail.)
Several AND and OR operators may appear within a logical expression. The
AND operator has higher precedence than the OR operator. The default
precedence and order of evaluation may be altered using parentheses. For
example:
• A>B
TRUE if register A is greater than register B.
• A > B AND C/D < 0.5
TRUE if register A exceeds register B, and C/D is less than 0.5.
• A > 100.0 OR B < 100.0 AND A/B > 0.50
TRUE if either of the two the conditions listed below tests TRUE:
1) register A exceeds 100.0.
2) register B is less than 100.0 and the ratio of register A to
register B exceeds 0.50.
Note: In example (c) the AND operator has precedence over the OR
operator; the AND operation is tested first, despite the fact
that it is the rightmost operator.
The operators AND and OR are lower in precedence than the arithmetic *, /, –,
and + operators. Consequently, there is no need to use parentheses to
surround numerical expressions in comparison operations.
8 - 103
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
Functions
There are several functions that are useful for extracting data from the raw
photometer data arrays. In all cases, arguments must be supplied to the
function that tells the system which photometric reading(s) and which
wavelength(s) are of interest. Photometric readings are specified as P1, P2, P3,
or P4 and correspond to the readings specified on the User-Defined Method
screen. Wavelengths are specified as 293NM, 340NM, 383NM, 405NM,
452NM, 510NM, 540NM, 577NM, 600NM, or 700NM. In addition, the value
---NM is accepted as a null wavelength specifier, which indicates that no
wavelength is desired.
Table 8-46: Available Functions and Their meanings
RATE(Pw, Px, yyyNM, zzzNM) Returns the rate of reaction (mAU/min) between photometric
readings Pw and Px for the bichromatic difference of
wavelength yyyNM minus wavelength zzzNM.
Note that there is no explicit monochromatic rate function. In order to obtain a monochromatic rate,
the null wavelength specifier should be used:
RATE(Pw, Px, yyyNM, ---NM) Returns the monochromatic rate of reaction (mAU/min)
between photometric readings Pw and Px for the wavelength
yyyNM.
8 - 104
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
All assays must provide some volume during the first reagent delivery. Either
chase or reagent component(s) (or both) need to be specified for the R1
delivery.
A well component or aliquot needs to be provided.
If a well has nonzero aliquots but no identifier (A, B, C, D, E), this message will
be displayed. Similarly, an identifier with zero aliquots will generate this
message.
Every reagent component delivered must appear in at least one well.
If a component identifier (A, B, C, D, E) appears in a reagent delivery (R1, R2,
or R3), it must also appear in the reagent cartridge.
Photometries must be time-ordered, and at least 15 seconds apart.
Photometry must occur within the interval <–30...950> seconds.
Please UNDO CHANGES before attempting to advance channel.
This message displays when the user attempts to move to another channel
without storing changes on the current channel.
Syntax errors have been found on this line.
If a syntax error is detected, the system attempts to move the cursor to the line
in which the error was detected. Note that occasionally the syntax error may
be on a preceding line; the program parser was unable to conclude that there
was an error until it had “looked ahead” a bit to see what came later!
The calculation is either too complex or not correctly terminated.
If a trailing right brace is missing, or a compound statement was not correctly
closed, the system will be unable to detect the end of the program. Check that
the braces balance, i.e., that total number of left braces ({) equals the number
of right braces (}).
This error may also occur if an expression is found that contains deeply nested
pairs of parentheses [e.g., ((((1 + (2/3) / (3 * (A) + (((0.1))))))))]. If such an
expression exists, simplify it using register assignments.
The second reagent delivery must fall within the interval <60...920
seconds>.
In order to easily access the reagent delivery stations, reagent deliveries should
fall within this time interval. Deliveries are excluded between 257.3 and 389.3
seconds, since the cuvette cannot be accessed during this time. However,
there is opportunity for access during the interval 389.3 to 461.3 seconds,
which has an undesirable side effect of requiring the system to index cuvettes
until the cuvette enters an accessible zone. The system will accept a reagent
delivery during this interval, although it may sometimes result in wasted
cuvettes.
The sum of all volumes exceeds the maximum cuvette capacity (500 µl).
The cuvette can only hold 500 µL of volume. When this message displays, try
scaling all volumes to bring the total volume under 500 µL.
8 - 105
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
8 - 106
Dimension EXL with LM/EXL 200 System Operator’s Guide Configuration
This password is not correct. You will not be permitted to STORE any
changes.
The user is prompted for the system password upon entry into User-Defined
Methods. If no password or an incorrect password is entered, this message is
presented to the user. If an invalid password is entered, anþattempt to store
using F4: STORE will again result in a prompt for password, at which time
correct entry of the system password will permit data to be stored.
Undo current changes to this channel?
This message appears when F6: UNDO CHANGES is pressed. If the user enters
N, the undo operation is aborted.
When reagent mix is requested, the reagent chase volume must be at
least 20 µL.
The ultrasonic system requires some chase to be expelled as ultrasonic mix
begins. When reagent mix is requested, there must be at least 20 µL of chase
available.
When sample mix is requested, the sample chase volume must be at least
10 µL.
The ultrasonic system requires some chase to be expelled as the ultrasonic mix
begins. There must be at least 10 µL of chase available when a sample mix is
requested.
8 - 107
Configuration Dimension EXL with LM/EXL 200 System Operator’s Guide
8 - 108
Dimension EXL with LM/EXL 200 System Operator’s Guide Appendix
Appendix
9-1
Appendix Dimension EXL with LM/EXL 200 System Operator’s Guide
Safety
General Safety
To operate the system, the operator must be proficient in its operation and
maintenance procedures.
To ensure safety, follow basic precautions.
• Always wear gloves when operating the Dimension EXL system.
• Observe all warnings and cautions in the manual.
• Remove safety guards only if specifically instructed in the procedures.
Replace all guards after completing the procedures.
• Stow cables and tubing properly to eliminate tripping hazards.
• Use only specified cleaners on the Dimension EXL system. Using other than
the specified cleaner will cause imprecision in some methods.
• Review the Flex reagent cartridge Instructions for Use (IFU) for specific
chemicals and safety information about the reagents in each method
cartridge.
Note: If the equipment is used in a manner not specified by the
manufacturer, the protection provided by the equipment may
be impaired.
Instrument Removal
Please contact a Siemens Healthcare Diagnostics service representative for any
system removal for repair or disposal.
9-2
Dimension EXL with LM/EXL 200 System Operator’s Guide Appendix
Safety Notes
Warnings and cautions are included throughout this guide to emphasize
important and critical instructions.
9-3
Appendix Dimension EXL with LM/EXL 200 System Operator’s Guide
Safety Labels
Safety labels are affixed to the instrument to alert operators to safety
considerations.
Table 9-1: Safety Labels and Their Description
Label Description
Potential Biohazard Indicates an area of the instrument that could have been in
contact with biohazardous materials. Do not handle the
contents or touch the area unless properly protected. Refer to
applicable laboratory procedures and to the guidelines set
forth by the Department of Labor (OSHA) 29CFR Part
1910.1030, Occupational Exposure to Bloodborne Pathogens:
Final Rule.
Pinch Hazard Indicates an area of the instrument where the operator can be
exposed to moving parts. Be careful around these parts when
performing diagnostic and maintenance operations with
safety shield removal.
Puncture Hazard Indicates an area of the instrument where the operator can be
exposed to a sharp tip that could puncture the skin. In
particular do not extend fingers under the guard over the
sample area during operation.
Burn Hazard Indicates a heated area of the instrument that can burn an
operator. Be sure to power off the instrument and let these
areas cool before touching them or performing diagnostic and
maintenance operations around them.
9-4
Dimension EXL with LM/EXL 200 System Operator’s Guide Appendix
Label Description
Protective Earth Terminal Indicates the terminal that connects the power supply line
external earth (ground) to the instrument.
Open Lid Indicates an area of the instrument where the operator can be
exposed to a closing lid. Be careful around these parts when
performing diagnostic and maintenance operations.
Crush Hazard Indicates an area of the instrument where the operator can be
exposed to moving parts. Be careful around these parts when
performing diagnostic and maintenance operations.
Link Symbol Indicates that the instrument is being operated from a remote
location.
Printer Shelf Indicates the weight limit for the printer shelf. The shelf is in
the proper position when two clicks are heard, indicating that
both support brackets are locked into place.
9-5
Appendix Dimension EXL with LM/EXL 200 System Operator’s Guide
Label Description
9-6
Dimension EXL with LM/EXL 200 System Operator’s Guide Appendix
Symbols Key
Figure 9-1: IFU Symbols Key
9-7
Appendix Dimension EXL with LM/EXL 200 System Operator’s Guide
Instrument Labeling
Figure 9-2: Instrument Labeling Diagram - Front Open View
9-8
Dimension EXL with LM/EXL 200 System Operator’s Guide Appendix
Controlled Removes power from all areas of the instrument except the compressor.
Before shutting the instrument down, perform a data backup to ensure that
all data and alignment files are saved. The Controlled Shutdown may be
required to perform maintenance procedures.
Emergency Removes power from the entire instrument. No data backup is performed
before power loss. Although the instrument performs an automatic data
backup every 2 hours, any changes to files after the backup and before the
shutdown are not saved.
To remove power from the RMS only or from both the Dimension EXL system
and the RMS, see the RMS Operator’s Guide.
9-9
Appendix Dimension EXL with LM/EXL 200 System Operator’s Guide
3. Turn OFF the UPS power by turning off the main power switch located at
the rear of the UPS, then press the UPS Off button at the front of the UPS.
The UPS will shutdown in approximately 30 seconds.
9 - 10
Dimension EXL with LM/EXL 200 System Operator’s Guide Appendix
9 - 11
Appendix Dimension EXL with LM/EXL 200 System Operator’s Guide
Safety Compliance
The Dimension EXL with LM system has been designed and tested to comply
with these safety standards:
• EN61010-1 (2001)
• EN61010-2-081 (2002)
• EN61010-2-101 (2002)
• EN61326
• EN61326-1
• EN61326-2-6
• CAN/CSA-C22.2 No. 61010-1 (2004)
• CAN/CSA-C22.2 No. 61010-2-081 (2004)
• CAN/CSA-C22.2 No. 61010-2-101 (2004)
• UL61010-1 (2nd edition)
under the following environmental conditions [subclause 1.4]:
Table 9-3: Safety Compliance
Mains supply 115 ±10% VAC or 230 ±10% VAC, 50/60 Hz, 100+/-10% VAC
Pollution Degree Degree 2, normal indoor laboratory environment. Air contains only
non-conductive pollutants with occasional condensation.
Barcode Scanner
The barcode scanner uses Class I LEDs (light-emitting diodes), and is not
hazardous to your eyes.
WARNING: The Dimension EXL system should not be used next to any
Industrial Scientific and Medical (ISM) equipment that must
functionally produce RF energy (such as Diathermy
Equipment).
9 - 12
Dimension EXL with LM/EXL 200 System Operator’s Guide Appendix
Installation Specifications
The Dimension EXL system is installed by a qualified Siemens representative.
Installation includes a full checkout to ensure that the equipment is fully
operational.
Dimensions
Length: 82 in. (208 cm)
Height: 49 in (124 cm) (without raised lids)
Height: 74 in. (188 cm) (Height includes raised lids)
Depth: 41 in. (104 cm)
These dimensions are without the printer shelf elevated.
Add 18 in. (46 cm) to the length if the printer shelf is used.
Minimum Clearances
Left Side: 24 in. (60 cm)
Right Side: 3 in. (8 cm)
Rear: 5 in. (13 cm) beyond LOCI module
The instrument requires 360-degree access and at least 36 in. (91 cm) of
working space on each side. Access to the rear of the instrument requires
moving it forward.
No leveling is required. However, a reasonably level floor is required for proper
operation. The two front casters of the instrument should be in their locked
position during operation.
9 - 13
Appendix Dimension EXL with LM/EXL 200 System Operator’s Guide
Dimensions
Length: 56 in. (142 cm)
Height: 49 in (124cm) (without raised lids)
Height: 72 in. (183 cm) (Height includes raised lids)
Depth: 40 in. (102 cm)
These dimensions are without the printer shelf elevated.
Add 16 in. (41 cm) to the length if the printer shelf is used.
Minimum Clearances
Left Side: 24 in. (60 cm)
Right Side: 3 in. (8 cm)
Rear: 5 in. (13 cm) beyond LOCI module
The instrument requires 360-degree access and at least 36 in. (91 cm) of
working space on each side. Access to the rear of the instrument requires
moving it forward.
No leveling is required. However, a reasonably level floor is required for proper
operation. The two front casters of the instrument should be in their locked
position during operation.
Power Requirements
Power Specifications
Table 9-4: Power Specifications
Maximum
Nominal Line Line Voltage Nominal Line Maximum
Continuous
Voltage vac Range vac Frequency Hz Power
Current
9 - 14
Dimension EXL with LM/EXL 200 System Operator’s Guide Appendix
Leakage Current
Table 9-7: Leakage Current
Leakage current complies with the requirements of CSA C22.2 No.1010.1 safety standards for
laboratory equipment in non-patient vicinity areas.
Water Requirements
The Dimension EXL system requires and external source of CLSI Clinical
Laboratory Reagent Water or the use of a Purified Water Diluent. The external
system must be connected to the water inlet. Maximum rate of water
consumption and maximum rate of liquid waste output is 0.53 gal. (2.0 L)/hr.
Drain line check valves are installed to prevent back siphoning. Maintenance of
the disposal point is the responsibility of the operator and should be selected
in accordance with local hazardous waste guidelines.
9 - 15
Appendix Dimension EXL with LM/EXL 200 System Operator’s Guide
The purified water supply systems for the Dimension EXL system must produce
water to the specifications shown below which meet the definition of CLSI
Clinical Laboratory Reagent Water (Clinical Laboratory Standards Institute,
C3-A4, Vol. 26, No. 22). Water supply maximum valve pressure is <55 psi
9 - 16
Dimension EXL with LM/EXL 200 System Operator’s Guide Appendix
1Not applicable to CLSI Clinical Laboratory Reagent Water (CLRW), but required for proper
instrument performance.
9 - 17
Appendix Dimension EXL with LM/EXL 200 System Operator’s Guide
Printer Shelf
Note: The printer shelf can support a maximum of 50lbs. (23kg)
The Dimension EXL system includes a shelf folded into the left side which can
hold the printer.
1. To raise the shelf, grasp the lower edge and pull out and up until the
supports click into position. Press slightly on the shelf to ensure the
supports are fully engaged.
2. To lower the shelf, pull up on the release levers on both supports. Push
the shelf down and toward the instrument until it clicks into its storage
position
Additional Requirements
Phone Line
A dedicated, direct phone line will be connected to the Dimension EXL system
for its exclusive use.
Host Interfacing
For host interfacing, a 9-pin female connector is required for hookup to the
male connector on the Dimension EXL system host communications port.
Specifications for this phone line are:
• Full Duplex - capable of two way transmission.
• Standard analog phone jack connection - RJ11C or RJ11W (Digital
not supported).
Figure 9-6: Interface Port
9 - 18
Dimension EXL with LM/EXL 200 System Operator’s Guide Appendix
9 - 19
Appendix Dimension EXL with LM/EXL 200 System Operator’s Guide
Instrument Labels
HM Wash Diagram
Figure 9-7: HM Wash Diagram
9 - 20
Dimension EXL with LM/EXL 200 System Operator’s Guide Appendix
9 - 21
Appendix Dimension EXL with LM/EXL 200 System Operator’s Guide
9 - 22
Dimension EXL with LM/EXL 200 System Operator’s Guide Appendix
9 - 23
Appendix Dimension EXL with LM/EXL 200 System Operator’s Guide
9 - 24
Dimension EXL with LM/EXL 200 System Operator’s Guide
Use the notes column to document all revisions made to a method and/or any method revision levels sent by manufacturer.
Manuals must be reviewed by supervisor/lab manager at a minimum frequency of once per year per CLSI/NCCLS GP2-A5 guidelines.
10 - 1
Dimension EXL with LM/EXL 200 System Operator’s Guide
10 - 2
Dimension EXL with LM/EXL 200 System Operator’s Guide
11- 1
Dimension EXL with LM/EXL 200 System Operator’s Guide
11 - 2
Dimension EXL with LM/EXL 200 System Operator’s Guide
Dimension EXL with LM and Dimension EXL 200 integrated chemistry system
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
11- 3
Dimension EXL with LM/EXL 200 System Operator’s Guide
Dimension EXL with LM and Dimension EXL 200 integrated chemistry system
25
26
27
28
29
30
31
11 - 4
Dimension EXL with LM/EXL 200 System Operator’s Guide
11- 5
Dimension EXL with LM/EXL 200 System Operator’s Guide
Dimension EXL with LM and Dimension EXL 200 integrated chemistry system: Daily System Check (ABS)
10
11
12
13
14
15
16
17
18
19
20
21
22
23
11 - 6
Dimension EXL with LM/EXL 200 System Operator’s Guide
Dimension EXL with LM and Dimension EXL 200 integrated chemistry system: Daily System Check (ABS)
24
25
26
27
28
29
30
31
11- 7
Dimension EXL with LM/EXL 200 System Operator’s Guide
11 - 8
Dimension EXL with LM/EXL 200 System Operator’s Guide
Dimension EXL with LM and Dimension EXL 200 integrated chemistry system: Daily System Check (CHK)
LOCI®
Are all
Reagent 1 Reagent 2 Sampler HM RMS* Highest
wavelengths
Value
10
11
12
13
14
15
16
17
18
19
20
21
22
23
11- 9
Dimension EXL with LM/EXL 200 System Operator’s Guide
Dimension EXL with LM and Dimension EXL 200 integrated chemistry system: Daily System Check (CHK)
LOCI®
Are all
Reagent 1 Reagent 2 Sampler HM RMS* Highest
wavelengths
Value
24
25
26
27
28
29
30
31
11 - 10
Dimension EXL with LM/EXL 200 System Operator’s Guide
11- 11
Dimension EXL with LM/EXL 200 System Operator’s Guide
Dimension EXL with LM and Dimension EXL 200 integrated chemistry system QuikLYTE Results
Na K
53 to 53 to Bottle Bottle
Date -55 to -40 > 0.8 < 0.6 <1.4 <0.04 ≤ 1.0 Initials
65 65 Value Value
140.00 4.00
11 - 12
Dimension EXL with LM/EXL 200 System Operator’s Guide
11- 13
Dimension EXL with LM/EXL 200 System Operator’s Guide
Dimension EXL with LM and Dimension EXL 200 integrated chemistry system
Weekly/Monthly Maintenance
Weekly
Monthly
(HM)
* If more than 100 IMT samples are run daily, clean IMT System every 15 days.
** Instruments with RMS Only
^ Perform System Check after replacing HM pump heads
11 - 14
Dimension EXL with LM/EXL 200 System Operator’s Guide
11- 15
Dimension EXL with LM/EXL 200 System Operator’s Guide
Dimension EXL with LM and Dimension EXL 200 integrated chemistry system
Instrument Log
11 - 16
Dimension EXL with LM/EXL 200 System Operator’s Guide
11- 17
Dimension EXL with LM/EXL 200 System Operator’s Guide
11 - 18
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Index
A
About this Guide 1-1
How this Guide is Organized 1-4
About this guide
Using this Guide 1-2
Air Filters
Replacing 5-12
Alert Keys 8-48
Configuration Calibration Alerts 8-50
Configuring QC Alerts 8-50
Configuring STAT Status 8-48
Configuring the Supplies Alert 8-49
Alignments 6-1
Barcode Scanner 6-3
Cuvette Ring 6-5
General Information 6-2
HM Module 6-7
LOCI® Arm 6-15
Offsets 6-2
Printing and Viewing 6-2
Photometer 6-17
R1 Reagent Probe 6-18
R2 Reagent Probe Alignments 6-22
Reagent Tray Alignment 6-27
Sample Probe 6-29
Appendix
Calibration (Verification) Review Flow Chart 4-22
Calibration (Verification)Setup Flow Chart 4-21
Keystroke Combinations 1-19
Operating Passwords 1-20
Understanding Test Report Messages 3-17
Automated System Check
CHK 8-7
Automatic Panic Rerun 8-12
Setting Up 8-12
12 - 1
Index Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
C
Calculated Results 8-15
Calibration 4-2
Alerts 8-50
Calibration Alert 4-4
Configuration
Definine Calibration Auto Acceptance Parameters 8-20
Definine Calibration Product 8-19
Edit Calibration Product 8-19
Entering Data Manually 8-19
Scanning Barcode Data 8-19
Cuvette System Temperature 5-23
Group 4-5
Alerts 4-5
HA1C 4-3
HB1C 4-3
HM Temperature 5-27
IMT 4-18
Electrolyte Coefficients 4-18
IMT System 4-2
Overview 4-2
Photometric and LOCI® Methods 4-3, 4-6
Cancelling Calibration 4-8
Reviewing and Accepting Results 4-7
Setting Up and Running a Calibration 4-6
Viewing History 4-9
Reagent System Temperature 5-25
Review Flow Chart 4-22
Setup Flow Chart 4-21
Troubleshooting 4-11
Calibration Statistics 4-11
Precision of Calibration Results 4-11
Quality Control 4-12
Urine Drugs of Abuse Methods 4-3
User-Defined Methods 8-92
Using Status Lists 4-19
Create a List of Methods 4-19
Create a Reminder List for Expiring Calibrations 4-20
Removing Calibrations from Instrument Memory 4-20
Verifying Photometric Methods 4-13
Calibration and Quality Control 4-1
Calibrating Photometric and LOCI® Methods 4-6
Calibrating the IMT System 4-18
Calibration Troubleshooting 4-11
Calibration/Verification Overview 4-2
Group Calibration Alerts/Run Calibration Group 4-5
Quality Control 4-23
Using Calibration Status Lists 4-19
Verification Troubleshooting 4-16
Verifying Photometric Methods 4-13
12 - 2
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Index
Card Cage
Control Board 5-136
Chemistry Troubleshooting 7-5
How to Troubleshoot 7-5
Imprecision 7-7
Inaccuracy 7-6
Printing Filter Data for Test Results 7-7
QC Results Out of Range 7-6
When to Troubleshoot 7-5
CHK
Configuring Automated System Check 8-7
Enter New Carton Value 7-9
Flex Reagent Cartridge Information 2-6
Running A System Check 2-3, 5-138
Specifications 2-5
Clot Check 3-4, 3-11
Clot Check Err Message 3-22
Configured 8-3
Drain on the IMT Port 5-8
Replacing Sample Tubing 5-126
Status Icons 1-29
Clot Check Overview 7-15
Code Compliance
Safety Compliance 9-12
Configuration
Flex® Cartridge 8-85
IMT 8-32
Consumable Screen
IMT 2-23
Correlations 8-23
Correlation Study 8-23
Dimension® EXL ™ System with Other Methods 8-23
Entering Observed Results for System Calculation 8-25
Entering Slope And Intercept 8-26
Printing a List of Correlated Methods 8-27
Removing a Correlation 8-27
Result Reporting For Methods 8-24
Using the Correlation Feature 8-24
Customizing 8-1
Automatic Cuvette Removal 8-8
Automatic Panic Rerun 8-12
Automatic Reflex Testing 8-10
Automatic Reflex Testing and Panic Rerun Report Slips 8-13
Calibration Configuration 8-19
Configuring the Printer 8-51
Configuring Touch Screen Alert Keys 8-48
Correlations 8-23
Creating Panel Keys 8-43
Entering Method Parameters 8-33
Entering Sample ID Information 8-58
External Printer Report 8-52
HIL Feature 8-44
IMT Configuration 8-32
Overview 8-2
12 - 3
Index Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
D
Daily Setup 2-2
Checking for Other Maintenance 2-3
Cleaning the Sample Area and Emptying Cuvette Waste 2-2
Managing Offboard Reagent Lots 2-8
Overview 2-2
Recording Daily Maintenance Results 2-3
Running a System Check 2-3
Understanding the System Counters Screen 2-7
Decontamination 5-139
E
Entering and Running Batch Samples 2-40
Entering Sample Data 2-37
Entering Sample Data Screen Fields 2-38
Error Messages 7-13, 8-104
Active System Errors Screen 7-13
Error Log Screen 7-13
Software Error Troubleshooting Help 7-13
User-Defined List 8-104
F
Flex®
Cartridge Configuration 8-85
Cartridge Configuration Fields 8-86
Filling a Cartridge 8-90
Loader Home Sensor 5-50
Loading Cartridge 8-91
Presence Sensor Board 5-51
12 - 4
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Index
Fuses 5-52
Replacing AC 5-52
Replacing DC 5-53
G
General Instrument Care and Cleaning 5-2
Materials for Cleaning 5-2
Opening the LOCI® Access Lid 5-3
Opening the Reagent Lid 5-3
Removing the HM Incubation Wheel Cover 5-4
H
Heat Torch Assembly 5-54
HIL 8-44
Alert Index Values 8-45
Setup 8-46
Histogram Plot 4-34
Printing Information 4-34
HM Consumables 2-18
Replacing HM Fluids 2-18
HM Incubate Wheel 5-58
HM Incubation Wheel Cover 5-4
HM Module
Alignments 6-7
HM Pump 5-62
HM Pump Heads
Replacing 5-17
HM Wash Probe 5-66, 5-68
Tubing 5-71
HM Wash Probes 5-5
Styletting 5-14
HM Wash Wheel 5-72
Alignment 6-8
I
IMT
Cleaning the System 5-11
Configuration 8-32
Bleach/Condition Soak Interval 8-32
ECO2 Test Method 8-32
Consumables Screen 2-23
Miscellaneous Tubing 5-73
Pump Tubing 5-10
Rotary Valve Seal 5-77
Rotary Valve Tubing 5-80
Using the IMT Consumables Screen 2-22
Using the IMT Logs 2-28
Waste Tubing 5-34
IMT Consumables 2-28
Replace 2-22
Replacing the QuikLYTE® Integrated Multisensor 2-24
Resolving a Failed Dilution Check 2-26
Running a Dilution Check 2-25
12 - 5
Index Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
L
Laboratory Data 8-63
Understanding Data Fields 8-65
Levey-Jennings Plot 4-35
Patient Results 4-35
Printing L-J Patient Information 4-35
Printing L-J QC Information 4-36
QC Results Plot 4-35
Load Errors 3-11
Load List Errors 3-12
Loading Samples 2-41
LOCI®
Arm Alignment 6-15
Arm to Detector Alignments 6-16
Arm to Vessel Alignments 6-15
Insert 5-82
Vacuum Cup 5-83
M
Maintenance 5-1
General Instrument Care and Cleaning 5-2
Monthly Maintenance 5-8
Other Maintenance 5-21
Weekly Maintenance 5-5
Maintenance Logs
Using the Daily Maintenance Logs Screens 2-5
Major Components of the Dimension® EXL ™ system
Major Components of the HM Area 1-10
Major Components of the IMT Area 1-9
Major Components of the LOCI® Module 1-14
Major Components of the Printer Area 1-13
Major Components of the Reagent Area 1-11
Major Components of the Thermal Chamber Area 1-12
12 - 6
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Index
O
Offboard Reagent Lots 2-8
Operating Passwords 1-20
Other Maintenance 5-21
Accessory Spare Parts Kit 5-21
Calibrating Cuvette System Temperature 5-23
Calibrating HM Temperature 5-27
Calibrating Reagent System Temperature 5-25
Checking 2-3
Cleaning Cuvette Windows 5-28
Cleaning the IMT Waste Tubing 5-34
Cleaning the Sample Probe and Drain 5-35
Cleaning the Water Bottle 5-37
Decontamination 5-139
Lowering and Raising the Thermal Chamber 5-21
Replace Sample Tubing (Clot Check Configured) 5-126
Replacing a Pump Limit Sensor 5-88
Replacing a Pump Solenoid Valve 5-93
Replacing a Pump Syringe 5-96
Replacing a Reagent Arm Radial Home Sensor 5-101
Replacing a Reagent Arm Vertical Home Sensor 5-102
Replacing a Reagent Probe Tip 5-106
Replacing an HM Pump and Motor Assembly 5-62
12 - 7
Index Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
P
Panel Keys 8-43
Pediatric Tubes
Using 2-34
Performing Power Shutdowns and Startups 9-9
Important Power Safety Information 9-9
Performing a Controlled Power Shutdown 9-10
Performing an Emergency Shutdown 9-10
Restoring Power After a Controlled Shutdown 9-10
Restoring Power After an Emergency Shutdown 9-11
Types of Shutdowns 9-9
Controlled 9-9
Emergency 9-9
Photometer
Alignment 6-17
Filter Wheel Dual Sensor 5-84
Home Sensor 5-86
12 - 8
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Index
Photometric Methods
Verification
Cancelling 4-15
Reviewing and Accepting Results 4-14
Setting Up and Running 4-13
Preparing Samples 2-2
Checking for Adequate Sample Volume 2-35
Placing Barcode Labels on Sample Tubes 2-35
Sample Preparation 2-35
Types of Containers 2-29
Using Pediatric Tubes 2-34
Using Primary Sample Tubes 2-31
Using Sample Cups 2-29
Using SSC (Small Sample Containers) 2-32
Configuring SSC (Small Sample Containers 2-32
Printer
Configuring 8-51
Paper 5-88
Processing Samples 2-43
Adding Samples While the System is Processing 2-43
Processing Samples from Standby Status 2-43
Samples Downloaded from an LIS 2-44
Configuring 2-44
Pump Limit Sensor 5-88
Pump Solenoid Valve 5-93
Q
QC
Alerts 8-50
Reviewing Results 4-42
User-Defined Methods 8-92
QC Patient Results 4-32
Quality Control 4-23
Crossover QC 4-24
Daily QC 4-23
Defining QC Panels 4-29
Defining QC Products 4-28
Editing QC Panels 4-30
Editing QC Products 4-29
Entering Ranges 4-27
Grouping Alerts 4-31
New Reagent and Control Lots 4-24
Previously Defined QC Product 4-30
Processing Samples 4-24
Program 4-23
QC Materials 4-24
Review 4-24
Setting Expiration Period and Checking Status 4-27
Urine Drugs of Abuse Methods 4-25
Using Alert Key 4-26
12 - 9
Index Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
R
R1 Reagent Arm
Delivery 8-82
Drain Alignment 6-18
Reagent Tray Alignment 6-21
Target Cuvette Alignment 6-19
R1 Reagent Probe 6-18
R2 Reagent Arm
Drain Alignment 6-22
HM Incubate Wheel Alignment 6-26
Probe Alignments 6-22
Reagent Arm Rotational Home Sensor 5-104
Reagent Tray Alignment 6-25
Target Cuvette Alignment 6-24
R2 Reagent Probe 5-5
R2, R3 5-19, 8-83
Reaction Vessels 2-20
Adding Reaction Vessels 2-20
Discarding Reaction Vessels 2-21
Reagent Arm Radial Home Sensor 5-101
Reagent Arm Vertical Home Sensor 5-102
Reagent Cartridge Inventory 2-13
Adding Reagent Cartridges 2-14
Manually Entering a Reagent Cartridge Barcode 2-15
Removing Reagent Cartridges 2-16
Using Third Lots of Reagent Cartridges 2-17
Reagent Hydrations 8-28
Cancelling a Reagent Hydration 8-29
Hydrating a Specific Cartridge Lot 8-28
Hydrating Using the Inventory/Hydration Screen 8-29
Setting a Timed Hydration Schedule 8-30
Reagent Probe Tip 5-106
Reagent Temperature 1-25
Reagent Tray
Alignment 6-27
Reagent Tray Home Sensor 5-110
Reference Range Indicators 3-16
Replacing
IMT Consumables 2-22
Reports
External Printer 8-52
Creating 8-52
Customizing 8-52
Format 8-53
Result Monitor 8-55
Accumulated Results Side 8-55
Limits Side 8-55
Methods and Settings 8-56
S
Safety and Specifications
Safety 9-2
Biohazard and Probe Safety 9-2
12 - 10
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Index
12 - 11
Index Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
12 - 12
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Index
T
Test Counters 8-62
Test Keys 8-76
Assignments 8-76
Programming 8-77
Test Report 3-5
Messages 3-17
Abnormal Assay 3-17
Abnormal Reaction 3-17
Aborted Test 3-18
12 - 13
Index Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
12 - 14
Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide Index
U
Unacceptable Photometer Ranges
Photometer 7-9
Unacceptable Ranges 7-9
Understanding Data Fields
Daily Maintenance Fields mtdat.xls File 8-67
IMT Logs Fields imtdat.xls File 8-72
Instrument Log Fields instrdat.xls File 8-74
Photometric Calibration Fields cldat.xls File 8-69
QC Results Fields qcdat.xls File 8-65
Test Results Fields rsdat.xls File 8-66
Understanding the System Counters Screen 2-7
Dimension® EXL ™ System Counters Screen 2-7
Using Instrument Log Screen 2-7
UPS (Uninterruptible Power Source) 1-27
U-Seal Element 5-129
User-Defined Methods 8-79
Calibration and QC 8-92
Customer Responsibility 8-80
Defining Flex® Cartridge Configuration 8-85
Defining Photometry Reading Times 8-84
Allowable Time Ranges 8-84
Defining Reagent and Sample Deliveries 8-81
Reagent and Sample Delivery Fields 8-83
Reagent Deliveries R2 and R3 8-83
Reagent Delivery R1 8-82
Sample Delivery 8-82
Defining the Calculation 8-87
Writing 8-87
Entering Method Parameters 8-90
Error Messages List 8-104
Filling a Flex® Cartridge 8-90
Flex® Cartridge Configuration Fields 8-86
Identifying the Method 8-80
Loading the Flex® Cartridge 8-91
Predefined Calculation Template 8-88
Fields 8-89
Programming Structure 8-99
Expressions 8-102
Functions 8-104
IF ELSE Statement 8-102
Implementation 8-99
Logical Expressions 8-103
Numerical Expressions 8-102
Statements 8-100
Programming Terms 8-99
Reviewing Method Kinetics 8-93
Running 8-92
Storing a Method 8-89
12 - 15
Index Dimension® EXL ™ with LM/EXL ™ 200 System Operator’s Guide
Warranty 8-80
Using this Guide
College of American Pathologists 1-2
For Additional Help 1-3
V
Vacuum Pump Muffler Filter 5-134
Verifcation Troubleshooting 4-16
Precision of Results 4-16
Statistics 4-16
W
Wash Wheel Home Sensor 5-58
Water Bottle 5-37, 5-135
Weekly Maintenance 5-5
Cleaning HM Wash Probes and the R2 Reagent Probe 5-5
12 - 16