Professional Documents
Culture Documents
ICP-OES Plasma Troubleshooting Guide
ICP-OES Plasma Troubleshooting Guide
ICP-OES Plasma Troubleshooting Guide
Plasma
Troublshooting Guide
Agilent Restricted
ICP-OES Plasma Troubleshooting Guide
Contents
Introduction ...............................................................................................................................................................4
First steps ...................................................................................................................................................................4
1. Check the reported error message ................................................................................................................4
RF Error list .................................................................................................................................................................5
RF RAM Failure .......................................................................................................................................................5
RF ROM Checksum Failure .....................................................................................................................................5
HV Supply Failure ...................................................................................................................................................6
Plasma Has Gone Out.............................................................................................................................................6
Plasma compartment door is open or emergency RF-off button is depressed .....................................................6
Insufficient Water Flow ..........................................................................................................................................6
RF Power Tube Air Flow Failure .............................................................................................................................6
RF Power Tube Voltage Failure ..............................................................................................................................7
Power Unattainable ...............................................................................................................................................7
Maximum power timeout failure...........................................................................................................................7
Over current failure ...............................................................................................................................................7
RF failure ................................................................................................................................................................7
Troubleshooting procedures......................................................................................................................................8
A. Plasma does not light .................................................................................................................................8
B. Plasma lights and then goes out ................................................................................................................9
RF Interface board .............................................................................................................................................. 10
Filament Power Supply ....................................................................................................................................... 11
High Voltage Power Supply (HVPS) Checks ......................................................................................................... 12
The HVPS module is activated. ....................................................................................................................... 12
The HVPS module is delivering the requested voltage and current. .............................................................. 14
RF Generator Assembly....................................................................................................................................... 16
1. Fibre Optic Looker ................................................................................................................................... 16
2. RF Power Tube ........................................................................................................................................ 17
3. Current Sensor ........................................................................................................................................ 17
4. Grid Circuit .............................................................................................................................................. 18
5. Anode Circuit........................................................................................................................................... 20
Appendix ................................................................................................................................................................. 21
RF Interface board details ................................................................................................................................... 21
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 2 of 40
ICP-OES Plasma Troubleshooting Guide
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 3 of 40
ICP-OES Plasma Troubleshooting Guide
Introduction
Problems with the 700 Series ICP-OES plasma RF can sometimes be difficult to troubleshoot. The aim of this
troubleshooting guide is to guide service personnel to efficient fault resolution.
Prior to commencement of work on the RF plasma system please read and understand the following warning:
The High Voltage Power Supply contains no serviceable parts, do not attempt
to disassemble.
First steps
1. Check the reported error message
The first step to begin troubleshooting is to identify the error message that is reported at the time of the plasma failure.
This may be the software error number and message reported by the customer or found by viewing the contents of the
Error and Debug log files:
The Error and Debug log files can be found in the following locations:
ICP Expert, ICP Expert II v1.x.x, ICP Expert II v2.x.x -Windows XP:
C:\Program Files\Varian\ICP Expert II 7xx-ES\Run\log
ICP Expert II v2.x.x - Windows 7:
C:\Users\Public documents\Varian\ICP Expert II\Log
For RF issues it is possible to receive multiple error messages at the same time. For example if the plasma goes out
suddenly to a hardware fault it is common to receive “Plasma has gone out” along with other RF error messages. In this
case the “Plasma has gone out” message is because of the hardware fault causing the plasma to fail. The attention should
be placed on other messages aside from the “Plasma has gone out” message.
If the only message received is “Plasma has gone out” then the cause may well be something other than the RF system.
Once the error messages are obtained review the RF error list on page 5 for the suggested action.
Other symptoms can be of significant importance too and need to be considered when determining troubleshooting
actions. The types of other symptoms are often noises such as a pop, loud bang or crackling which indicates an electrical
discharge.
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 4 of 40
ICP-OES Plasma Troubleshooting Guide
RF Error list
Table 1 lists the Software error number the message along with the corresponding firmware error number and message.
Once the error message for the plasma fault has been determined refer to table 1 to determine the next actions to take.
RF RAM Failure
Cause Fault detected with RF Interface board RAM
Actions
1. Check seating of PROM in socket, IC 7 on RF interface board
2. Replace RF Interface board
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 5 of 40
ICP-OES Plasma Troubleshooting Guide
HV Supply Failure
Cause The High Voltage Power Supply has reported an error.
Actions
1. Check that the circuit breaker on rear of the instrument is switched on.
2. Verify all interlocks are enabled to allow the HVPS to energise. These include Air, water,
switch Interlocks and filament voltage
3. Verify if the HVPS is being energised; refer to troubleshooting checks to see if the HVPS is
being energised on page 12
4. Test the HVPS in isolation, if it fails to provide voltage then replace. Refer to page 15 for the
procedure.
5. Possible short on high voltage supply line in RF Generator. Follow troubleshooting for the
Anode circuit on page 20.
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 6 of 40
ICP-OES Plasma Troubleshooting Guide
Power Unattainable
Cause The requested power level by software is not able to be achieved by the RF system.
Actions
1. The RF tube has reached the end of its life and needs to be replaced. If a lower power level
can still be delivered without error this verifies the RF tube has reached the end of its life.
2. Current transformer faulty, follow troubleshooting to check the current sensor circuit on
page 18.
3. Possible short on high voltage supply line in RF Generator. Follow troubleshooting for the
Anode circuit page 20.
RF failure
Cause The RF Generator failed to oscillate (no grid current detected)
Actions
1. The RF tube has failed, test with new RF tube.
2. Fault in grid bias circuit, refer to troubleshooting the grid circuit on page 18
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 7 of 40
ICP-OES Plasma Troubleshooting Guide
Troubleshooting procedures
A. Plasma does not light
If no plasma is formed at all then the following items should be checked.
1. Argon purity.
If the argon is not the required 99.996% purity or is contaminated the plasma may not form. Check the grade of
argon being used. If using cylinders try another cylinder, if possible try a cylinder from another batch or supplier.
Check the argon supply and regulator for leaks.
Long lengths of plastic tubing should not be used for the argon as plastic is air permeable.
Issues with liquid argon supplies are extremely rare and are normally not suspected.
Connections:
Check all connections of the sample introduction system for cracks, missing or damaged items. This includes the
torch, torch connection gas lines, transfer tube, spray chamber, o-rings, nebuliser, drain and sample tubing.
Purge:
The torch may fail to light at the first attempt if the air in the torch and spraychamber has not been fully purged.
Repeat the torch light procedure. You may wish to increase the purge time in the Instrument setup Ignition tab if
this occurs regularly. Also, the nebulizer flow must be allowed to drop to zero before ignition. This is particularly
important for axial instruments. This may require a longer Delay time, especially if you have an Argon Saturator
Accessory fitted.
4. Organics samples.
After running organic samples, the torch area and spraychamber may fill up with organic vapors. Use a
higher power and longer purge time for the ignition. If this is unsuccessful, clean the spraychamber before
attempting to light the plasma again.
A higher power level is required to ignite and maintain the plasma when running low boiling point organic samples
such as methanol. Ignition parameters can be set on the Ignition tab of the Instrument Setup screen.
5. Torch alignment.
Check that the torch is correctly positioned in the work coil. Refer to checking torch installation on page 27 also
check the torch is set to the correct depth in the torch stand as shown below:
The top of the torch intermediate tube should The intermediate tube of the torch should be
sit at least 2 mm below the induction coil. approximately 3 mm to the left of the induction coil
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 8 of 40
ICP-OES Plasma Troubleshooting Guide
8. Incorrect gas flows for Plasma, Auxiliary and Nebuliser flows. Verify correct gas flows for these gasses. Refer to
700 series Service Manual chapter 10 for details on how to check these flows.
9. No seeding spark.
Check that the seeding spark is reaching the torch during the ignite phase. The ignitor may have failed or the
spark is discharging to some other point due to contamination.
3. Check spray chamber is draining correctly and there are no droplets of liquid in the torch.
If the plasma stays lit and then goes out some time after check the following:
1. The spray chamber is not draining.
If the spray chamber is not draining correctly then it may fill up and allow liquid to enter the torch putting out
the plasma.
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 9 of 40
ICP-OES Plasma Troubleshooting Guide
RF Interface board
The RF interface board controls and monitors the plasma generating system. These functions include:
• Cooling fan air flow monitoring
• Cooling water flow
• Safety interlocks
• Filament supply status
• High voltage supply status
• Power oscillator grid current
• Plasma ON
• RF power level control (ALC)
1. The following interlocks must be satisfied before the HVDC power supply is turned on
• RF Door and RF cabinet interlock bi-stable switches closed
• Water On
• Fan On
• Filament On
You can visually check the status of these interlocks by looking at either the Software indicators in the
Diagnostics software or the LED state indicators on the RF interface board.
LED state indicators on the RF interface board
D7 ON = Water Flow OK (green)
D8 ON = Cooling Fan OK (green)
D9 ON = Filament Supply OK (green)
D13 ON = Plasma/HV interlock OK (green)
Refer to the RF Interface board details on page 21 for the location of the LED’s
For details on the Diagnostics software, refer to Section 7 of the 700-ES series Service Manual.
2. During the Plasma Ignition sequence, some of the monitors in the plasma generating system are ignored until the
start of the Run phase. These include;
• Plasma ON
• RF power level control (ALC)
If there is a fault with any of these, the plasma may ignite for a short time and then go out when the Plasma
Ignition sequence starts the Run phase.
You can visually check the status of these interlocks by looking at the LED state indicators on the RF interface board.
LED state indicators on the RF interface board
D6 ON = Plasma ON (yellow)
D14 ON = ALC fail (red)
The “ALC fail” LED state indicator will be on momentarily when the plasma first
NOTE ignites - this is normal.
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 10 of 40
ICP-OES Plasma Troubleshooting Guide
If a fault with the ‘Plasma ON’ monitoring is suspected, then the Fibre Optics checks should be carried out, refer
page 16.
If a fault with the ‘ALC fail’ monitoring is suspected, then the Current Sensor checks should be carried out, refer
page 17.
The filament draws ~22A at 6.61 VRMS +/- 5% measured at the filament transformer during the first 200 hours and
reduces to 6.17VRMS thereafter. All voltages measured at the filament transformer terminals.
There are four main areas to check for correct filament supply operation:
1. The Filament power supply is activated.
This can be tested by simply observing if LED indicator “D9” on the RF interface board is ON.
If LED indicator “D9” is OFF, ensure that the Filament supply is activated in the instrument software.
If the filament supply is activated, but the LED indicator ‘D9’ is off, this may indicate:
• Failed Solid State Relay (SSR)
• Failed Instrument Power Control board.
Figure 6 Measuring the filament voltage Figure 5 Measuring the filament current
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 11 of 40
ICP-OES Plasma Troubleshooting Guide
If the measured voltage and current are not as expected, then possible causes are;
• Failed filament transformer,
• Failed filament Solid State Relay (SSR)
• Failed instrument power board
3. The electrical resistance of the filament wires from the transformer terminals to the RF tube socket.
This can be tested by measurement to ensure good electrical contact is made at all electrical connections and
terminations. There should be little (if any) electrical resistance from the filament transformer terminals to the RF
tube socket. Note: Circuit resistance is less than multimeter lead resistance.
If the LED 2 (Green) is ON, both internal fans are rotating and LED energised is ON, the HVPS is activated.
If the LED 2 (Green) is OFF and the internal fans are not rotating, ensure that the HVPS (HV (Stage 2)) is enabled in
the instrument software.
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 12 of 40
ICP-OES Plasma Troubleshooting Guide
If the HVPS is enabled in the software, but the LED 2 (Green) on the HVPS and/or the Red LED inside the HVPS are
OFF, this indicates that the contactor (internal to the HVPS) has not been energised.
2. Disconnecting and reconnecting 2PL1 on the HVPS. As you disconnect and reconnect 2PL1, you should
hear the contactor clicking as the contacts open and close.
NOTE The power supply to the contactor is routed through all the safety interlocks.
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 13 of 40
ICP-OES Plasma Troubleshooting Guide
Filament (Stage 1) and HV (Stage 2) MUST be enabled first before the RF power will
NOTE be applied.
a) View the contents of the Error and Debug log files and check for any reported RF system faults.
b) If there are faults such as Grid Bias Failure, the RF Interface board will shut down the RF system and you will not
be able to measure the output voltage signals.
You can over-ride this error and force the HVPS to remain on by:
1. Connect a ‘clip lead’ from TP1 (+5v) to the anode of D17 on the RF Interface board. Refer to Figure 11 Clip
lead from TP1 +5V to the anode of D17
2. Move ALC test link to LK1 to the test position. Refer to RF Interface board details on page 21 for the location
of the ALC test link
3. Apply an RF power of 1.0kW
4. Measure the HVPS output voltage signals (TP19) using a voltmeter. If the output voltage signals are as
expected ~3.8Vdc, then the HVPS is OK.
If the HVPS checks out OK then the cause of the Grid Bias error is a failure of the RF Generator to oscillate. The
RF Generator assembly needs to be inspected for a cause.
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 14 of 40
ICP-OES Plasma Troubleshooting Guide
D17 Anode
TP1 +5v
A suitable clip lead would be p/n 8184000200 12IN Test Clip-BLK or Pomona 3781
NOTE Minigrabber® Test Clip Patch Cord.
c) If the output voltage signals are not as expected and there are no error messages displayed.
Then using a voltmeter, check that;
• PL4 pin 14 (Enable) on the HVPS is > +2.4vdc.
• PL4 pin 1 (Cntrl) on the HVPS is ~ +5.7vdc.
d) If the output voltage signals are not as expected and a “Current Overload Failure” or “ALC Failure” is recorded,
then an overcurrent or overvoltage has been reported by the HVPS. At the time the error is reported the
overcurrent orange LED on the HVPS will momentarily light for an overcurrent fault and the red overvoltage
LED for an ALC failure. The RF interface board will then shut down the HVPS and you will not be able to
measure the HVPS output voltage signals. In this case it is necessary to test the HVPS in isolation from the RF
Generator assembly. Please follow the procedure testing the HVPS in isolation on page 15 to determine if the
HVPS or RF Generator is at fault.
Refer to Figure 27 for the location of the overcurrent and overvoltage LED’s.
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 15 of 40
ICP-OES Plasma Troubleshooting Guide
RF Generator Assembly
There are several areas inside the RF box that can be checked, these include;
1. Fibre Optic Looker
2. RF Tube
3. Current sensor
4. Grid circuit
5. Anode circuit
If LED D6 is OFF and the plasma is on, light is not being detected from the fibre optic looker. The RF system will
be shut down the RF, firmware error [6] and the following error will be displayed PC. “Plasma has gone out”.
Typically if the above error is the only error written in the error log file, this indicates that the fibre optical
looker failed to see the light from the plasma.
To confirm a fault with the Fibre optic looker, the monitoring of the fibre optic looker can be over-ridden by
placing LK2 on the RF Interface board into the “Test” position. Refer to RF interface Board details on page 21
for the location of LK2.
With LK2 in the test position, the LED D6 will be ON and the plasma ON button in the
software will not be active. Therefore to ignite the plasma, place LK2 in the ‘Run’
NOTE
position and press the Plasma ON button, as the ignition sequence enters the ignite
phase move LK2 to the ‘Test’ position.
If the plasma ignites and runs normally with LK2 in the ‘Test’ position and goes out when LK2 is returned to the ‘Run’
position. This indicates one of the following conditions
a) The optical fibre is faulty.
If this is the case, check if the tip of the optical fibre is damaged.
If it is damaged, you can cut off the damaged section and strip the outer sheath as detailed below, then
reposition the fibre in the direction of the plasma.
Remove 2.5mm
of outer sheath
Figure 12 Repairing the fibre end Figure 13 Fibre optic position in torch box
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 16 of 40
ICP-OES Plasma Troubleshooting Guide
c) The plasma is below the intensity necessary to be detected. With certain types of samples (particularly
organics) and low plasma power settings the plasma may not be sensed even though it is on. In this case try
increasing the plasma power 0.1 to 0.2kW.
2. RF Power Tube
The RF tube has a finite life and as the RF tube ages, it will require more anode current to deliver the requested
RF power level. Eventually the RF tube will require more anode current than the HVPS can deliver (470mA) and
the HVPS will then go into current limit (LED 1 (Orange) on the HVPS is ON). In this case the RF Interface board
will shut down the RF system, the RF firmware error [15] is written in the error log and the following error will be
displayed on the PC “Over current failure”
Typically the average life of an RF tube in an instrument is around 5500 Filament on hours.
If the ‘Filament On’ hours reported in the DETAILS page of the Diagnostics software is greater than this, the RF
tube should be suspected and replaced.
If the RF Tube is known to be good, then the HVPS checks on page 12 should be carried out.
If the RF Tube and HVPS are known to be good, then the Current Sensor checks on page 17 should be carried out.
3. Current Sensor
The RF current in the induction coil is monitored by a current sense circuit located between the return lead of
the induction coil and ground. The RF current level is then fed back to the Automatic Level Control circuit on
the RF Interface board where it is compared to the required power level as set by the instrument software.
The requested output voltage signal to the HVPS is then adjusted so that the RF current in the induction coil
matches the requested power level in the instrument software.
If the RF Interface board is unable to match the measured RF current in the induction coil to the requested RF
power level in the instrument software, there are several error messages that may be displayed.
“Power unattainable”
This indicates that the requested output voltage signal to the HVPS (TP21) has exceeded 10V.
A failed RF Tube should be suspected.
“ALC Failure”
This indicates that the delivered output voltage signal from the HVPS (TP19) has exceeded 7kV and the RF
Interface board has been unable to match the measured RF current in the induction coil to the requested RF
power level in the instrument software.
The status of the ALC can be observed on the LED indicator “D14” on the RF interface board.
If this LED is ON, a failed ALC circuit on the RF Interface board or a failed current sensor circuit should be
suspected.
If the RF tube and HVPS are known to be good, then a failed HV Filter Component should be suspected.
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 17 of 40
ICP-OES Plasma Troubleshooting Guide
With LK1 in the test position the plasma will appear a little weaker than
NOTE when LK1 is in the 'Run' position.
CAUTION If LK1 is removed from the RF Interface board while the plasma is ON, the plasma will
run at full RF power (2kW). If the AUX and PLASMA gas flow settings are not
appropriate for this power setting, the torch may melt.
Now you can measure the voltages by setting a power level of 1.2kW,
- The delivered output voltage signal (TP19) should be ~ 5Vdc
- The delivered output current signal (TP20) should be ~ 3.3Vdc
To confirm that the Current sensor board is OK and that the fault is with the ALC circuit on the RF Interface
board, measure the feedback voltage from the current sensor circuit (TP14 on the RF Interface board) while
LK1 is in the test position. The voltage at TP14 should read ~ 4.3 Volts.
4. Grid Circuit
Grid current is monitored on the RF interface board to determine if the RF power Generator is oscillating. If the
grid current fails to flow, the ‘RF Failure’ error message will be displayed on the PC. This fault condition will
inhibit plasma ignition and cause the plasma to shut down if detected during plasma operation.
There are several electronic components inside the RF box that may fail. Below are some checks that can be to
done.
at 1.2KW
The anode voltage should be ~5kV, which can be measured as 5V at TP19 on the RF interface board.
The grid current should be 300mA, which can be measured as 3v at TP20
The 'Grid bias' voltage should be > 55mA, which can be measured as > -5.5v at TP16.
A low value for the Grid Bias voltage usually indicates that the tube is near the end of its life
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 18 of 40
ICP-OES Plasma Troubleshooting Guide
If all the inductors in the RF box visibly look ok, not distorted or burnt, including the induction coil, check the
grid circuit capacitors C2, C3, C4, C5 and C19 for a possible short circuit. Check resistors R1, R3 and R4 for
open circuit along with RFC2. D1 inside the Filter compartment may also be short circuit; this can be verified
by measuring the resistance from TP16 on the RF interface board to ground. IF D1 is good the resistance is
~100 ohms. Ensure the RF is not energised when measuring the resistance of D1.
If C5 (25pF 5KV capacitor connected from the grid inductor to ground) is open
circuit the RF efficiency will be poor. If it is open it will look damaged or cracked
NOTE
or may not be soldered properly.
You should be able to check the capacitors for a short circuit by removing the RF tube and measuring the
resistance between the grid pins on the RF Tube socket and chassis with a multimeter. If all is well, you
should measure about 7.0K ohms. You can also do this by measuring the resistance from L2 to ground,
which should also be 7.0K ohms.
This is not a 100% reliable test as the capacitor(s) may break down only when
NOTE voltage is applied.
It is also possible that C19 may have failed, in which case the Generator may partly function since part of the
grid bias resistance R1 (2.2K) will be in circuit. Any of these components failing would show a grid bias failure
error.
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 19 of 40
ICP-OES Plasma Troubleshooting Guide
5. Anode Circuit
The HV filter capacitors C13, C14 & C15 can fail either externally, leaving a visible track mark across the external
surface of the capacitor(s), or internally, which is unfortunately invisible. Typically, this type of failure will force
the HVPS into current limit, which is indicated by the yellow LED on the HVPS and an “over current” error
message. A noise may also be heard when HV is being applied, this may indicate some discharge across some
components in the RF Generator, typically this is the filter capacitors.
There are two methods to check the High Voltage filter capacitors;
· You can test whether the capacitors are breaking down by testing the HVPS in isolation, Refer to page 15. If
you see no yellow “overcurrent” LED activity with the red wire removed, this is a strong indication (but not
proof) that the capacitors are the root cause, though it may still be the RF tube.
· You can remove both filament wires from the filament transformer, and try and apply some RF power in no
plasma mode. With the filament wires disconnected, there is should be no anode current flowing. If there is
anode current measured (any voltage measured at TP20 on the RF interface board), replace the HV filter
capacitors C13, C14 & C15.
The anode circuit also includes capacitors C6, C7 and C8. If one of these have failed it will give the same symptoms
as the filter capacitors, however experience has shown these capacitors to be very reliable.
A remote possibility is either RFC4 or RFC5 and the insulated feedthroughs in the HV filter compartment
breaking down to chassis, but there may be visible signs of such a failure.
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 20 of 40
ICP-OES Plasma Troubleshooting Guide
Appendix
RF Interface board details
D6 Plasma on
D9 Filament Voltage OK
D13 Interlocks OK
TP17
TP16
TP18
TP19
TP20
Plasma Test Jumper (LK2)
D17 Anode
Adjust RV1
Axial -5.715V
TP5 -15V
Radial -5.302V TP2
Sig Earth
ALC Test Jumper (LK1)
TP3 +15V
TP6 +12V
TP1 +5V
Sig Earth
TP10
TP11
TP15
TP13
TP14
TP21
TP4
TP8
TP9
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 21 of 40
ICP-OES Plasma Troubleshooting Guide
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 22 of 40
ICP-OES Plasma Troubleshooting Guide
The numbers given in the tables above are typical values to be used as a guide
NOTE only. They do not form part of any specification.
The plasma may extinguish at the limits of plasma power (0.6kW and 2.0kW).
This is normal for some systems and does not represent a fault. The method
NOTE
allows up to 1.5kW for a radial and 1.7kW for an Axial system.
CAUTION When setting RF power of 1.5kW and above, be sure to increase the AUX gas flow to
2.25l/min and PLASMA gas flow to 18 l/min, otherwise the torch may melt.
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 23 of 40
ICP-OES Plasma Troubleshooting Guide
Plasma off
Power Level Anode Voltage Anode Current Grid current Mod voltage Calibration point
1VDC=1000 10VDC=1amp 1VDC=10mA 0 to 10VDC 5.302VDC Radial
5.715VDC Axial @
1.8kW
TP19 TP20 TP16 TP21 TP13
0.6
0.8
1.0
1.2
1.4
1.5
1.7
1.8
2.0
Plasma on
Power Level Anode Voltage Anode Current Grid current Mod voltage Calibration point
1VDC=1000 10VDC=1amp 1VDC=10mA 0 to 10VDC 5.302VDC Radial
5.715VDC Axial @
1.8kW
TP 18 (Gnd) TP19 TP20 TP16 TP21 TP13 -TP2 (sig gnd)
0.6
0.8
1.0
1.2
1.4
1.5
1.7
1.8
2.0
CAUTION When setting RF power of 1.5kW and above, be sure to increase the AUX gas flow to
2.25l/min and PLASMA gas flow to 18 l/min, otherwise the torch may melt.
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 24 of 40
ICP-OES Plasma Troubleshooting Guide
Radial
The radial work coil can be aligned using the Pre optics torch align tool 7210021100. The work should also be
centered vertically about the mounting posts. Also ensure there is approximately 9.4mm between the top and
bottom turns of the work coil and the middle turn is centered.
9.4mm
Axial
The axial work coil can be aligned using the axial alignment tool Mk III p/n 9910104400. The kit consists of an
interface alignment disk, coil alignment nut and a coil alignment shaft. The following instructions give details on
using the alignment tool:
Figure 19 Remove cone from the interface Figure 20 Fit Interface alignment disk to the
interface. Note: this will only fit in one orientation.
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 25 of 40
ICP-OES Plasma Troubleshooting Guide
Figure 21 Install the coil alignment nut into Figure 22 Insert coil alignment shaft through work
the work coil and center within the work coil. coil and tighten into the interface disk
Also check the work coil for correct orientation as shown below and that there is 7.5mm between work coil
spacing’s.
7.5mm
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 26 of 40
ICP-OES Plasma Troubleshooting Guide
If you find the coil is too close at any point remove the torch and bend the coil gently by hand, refit the torch and
check again.
Figure 25 Using an inspection mirror to check torch is centered in the work coil
and not touching at any point.
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 27 of 40
ICP-OES Plasma Troubleshooting Guide
LED 1 LED 10
Mains connection
Figure 27 HVPS
LED 1 Orange Status LED's and connections
Overload
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 28 of 40
ICP-OES Plasma Troubleshooting Guide
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 29 of 40
ICP-OES Plasma Troubleshooting Guide
Instrument front
2 3 5
4
6
7
9 10 11 12
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 30 of 40
ICP-OES Plasma Troubleshooting Guide
1
2
3 4
5
4
7
6
8
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 31 of 40
ICP-OES Plasma Troubleshooting Guide
Anode compartment
1
4
5
2
8
3 6
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 32 of 40
ICP-OES Plasma Troubleshooting Guide
Grid compartment
5
1
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 33 of 40
ICP-OES Plasma Troubleshooting Guide
Filter Assembly
1
2 4
3
5 7
6
8
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 34 of 40
ICP-OES Plasma Troubleshooting Guide
2 4 5
6 7
9 10 11 12
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 35 of 40
ICP-OES Plasma Troubleshooting Guide
2 3 4
1 5
8 9
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 36 of 40
ICP-OES Plasma Troubleshooting Guide
Igniter compartment
1 2 3 4 5 6 7
9 10
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 37 of 40
ICP-OES Plasma Troubleshooting Guide
1 2 3 4 5 6 7
10
11
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 38 of 40
ICP-OES Plasma Troubleshooting Guide
1 2 3
5
6
7
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 39 of 40
ICP-OES Plasma Troubleshooting Guide
Cables
Assy loom ICP Simul – Filament Resistors 110645900 Resistor mounted on outside of filter compartment. 2 required.
Assy Loom ICP RF Interlock 110651200 Loom including filter and anode compartment interlock switches
Issued: May 2013 – SPSD Product Support Copyright ã 2013 Agilent Technologies Agilent Restricted Page 40 of 40