Professional Documents
Culture Documents
WI-04!05!54 - Service-Manual CT Motion - Rev 03.1
WI-04!05!54 - Service-Manual CT Motion - Rev 03.1
WI-04!05!54 - Service-Manual CT Motion - Rev 03.1
Trademark Notice
TM
CT motion is a trademark of ulrich GmbH & Co. KG.
®
ulrich medical is a registered trademark of ulrich GmbH & Co. KG.
Names and terms used for products from other manufacturers mentioned in this service
manual may be connected to registered trademarks or trade names. ulrich GmbH & Co. KG
acknowledges their legal status without expressly mentioning their underlying legal form, e. g.
TM or ®. Changes to these legal forms are recognized without reservation.
service@ulrichmedical.com
www.ulrichmedical.com
Table of Contents
Table of Contents
1 About This Document.......................................................................................... 1-10
1.1 Purpose ..................................................................................................................... 1-10
1.2 Scope ........................................................................................................................ 1-10
1.3 Currentness .............................................................................................................. 1-10
1.3.1 Change history ............................................................................................ 1-10
1.4 Applicable documents ............................................................................................... 1-12
1.5 Highlighting and Symbols ......................................................................................... 1-13
1.5.1 Warnings ..................................................................................................... 1-13
1.5.2 Text Markings .............................................................................................. 1-13
1.6 Abbreviations ............................................................................................................ 1-14
7.1.6 Reading out data / Calibrating CT motion touch terminal ........................... 7-83
7.2 Service flap ............................................................................................................... 7-86
7.2.1 Removing service flap for service work/testing ........................................... 7-86
7.3 Front module ............................................................................................................. 7-86
7.3.1 Disassembling front module for service work/testing .................................. 7-86
7.4 Device base .............................................................................................................. 7-88
7.4.1 Disassembling/opening device base for service work/testing ..................... 7-88
7.4.2 Replacing LP5 charger electrical components before S/N CTM1510600 ... 7-90
7.4.3 Replacing LP5 charger electrical components of device base with S/N
CTM1510600 or higher ............................................................................... 7-90
7.4.4 Replacing Fuse 5A FF for Charger Electrical Components before S/N
CTM1510600 ............................................................................................... 7-91
7.4.5 Replacing Fuse 5A FF for Charger Electrical Components at device base
with S/N CTM1510600 or higher ................................................................. 7-91
7.4.6 Replacing cable 11 battery charging transistor before S/N CTM1510600 .. 7-91
7.4.7 Replacing cable 11 battery charging transistor of device base with S/N
CTM1510600 or higher ............................................................................... 7-92
7.4.8 Replacing power supply unit before S/N CTM1510600 .............................. 7-92
7.4.9 Replacing power supply of device base with S/N CTM1510600 or higher . 7-93
7.4.10 Replacing batteries SP000445-ET before S/N CTM1510600 ..................... 7-95
7.4.11 Replacing batteries SP000445-ET of device base with S/N CTM1510600 or
higher ........................................................................................................... 7-96
7.4.12 Replacing cable 14 LEDs battery charging before S/N CTM1510600 ........ 7-97
7.4.13 Replacing cable 14 LEDs battery charging of device base with S/N
CTM1510600 or higher ............................................................................... 7-97
7.4.14 Replacing double caster or double caster with brake ................................. 7-98
7.5 Injector head ............................................................................................................. 7-99
7.5.1 Replacing gas spring ................................................................................... 7-99
7.5.2 Replacing rail system ................................................................................ 7-100
7.5.3 Replacing cross plate A ............................................................................. 7-101
7.5.4 Replacing safety screw for injector head .................................................. 7-101
7.5.5 Replacing patient tubing sensor ................................................................ 7-102
7.5.6 Replacing high-pressure filter holder (SP000558-ET) .............................. 7-102
7.5.7 Replacing ECG cable SP000365-ET ........................................................ 7-103
7.5.8 Replacing drip cup holder SP000043-ET and pins SP000042-ET ........... 7-104
7.5.9 Replacing injector head housing ............................................................... 7-105
7.5.10 Replacing On/Off switch SP000392-ET before S/N CTM1721200 ........... 7-107
7.5.11 Replacing On/Off switch SP002223-ET with S/N CTM1721200 or higher7-108
7.6 Feeding unit ............................................................................................................ 7-110
7.6.1 Replacing bottle lift complete (SP000298-ET) .......................................... 7-110
7.6.2 Replacing spring catch .............................................................................. 7-110
7.6.3 Replacing holding arm for bottle guide (SP000304-ET)............................ 7-111
7.6.4 Replacing spike holder .............................................................................. 7-111
1.2 Scope
These work instructions apply to persons authorized by ulrich medical, particularly
technicians, who perform service work on the XD 8000 injector system and perform technical
safety checks according to test report F-04-05-23.
1.3 Currentness
The service manual corresponds to the status of the product at the time the service manual
was created.
Subject to technical changes.
Possession of this service manual does not automatically entail inclusion in the change
management.
Instructions for Use “ulrichINJECT CT motion ulrich medical contrast media injector”
according to your local language.
When needed, request the current version of the instructions for use from ulrich
medical.
1.5.1 Warnings
Information highlighting risks to the patient and/or user and/or device is marked as follows:
Instructions.
Indicates a situation where the operator has to take specific action.
1. Multi-step instructions.
2. Observe the sequence.
Result of an instruction.
Measure to take in order to prevent hazards referred to in a warning.
1.6 Abbreviations
Abbreviation Explanation
B Bonding
CM Contrast medium
ESD Electrostatic discharge
FOC Fiber-optic cable
Hex Hexagon size
IFU Instructions for use
NaCl 0.9% saline solution
pc. Piece
PE Protective earth
POF Plastic optical fiber
PWM Pulse width modulation
RPM Revolutions per minute
Table 4. Abbreviations
2 Safety Information
2.1 Manufacturer's responsibility
ulrich medical is only responsible for effects on safety, reliability and performance of our
products if the following requirements are met:
Training based on the instructions for use, the accompanying safety-related information
and the maintenance notes was performed on the correct handling, use, and operation of
the medical device, and on the correct use together with other medical devices, objects
and accessories.
Assembly, enhancements, readjustments, changes or repairs are to be performed by
persons who have the required specialized knowledge to service medical devices based
on their training and practical experience, and who have been trained an authorized by
ulrich medical.
These persons have required premises (including quality, size, equipment, and
furnishings), as well as the required devices and other working materials, and are able to
perform service works properly and verifiably with respect to their type and scope.
The products are only used in accordance with the instructions for use and the service
manual.
Applicable standards are observed.
3.1 Packaging
3.1.1 Reusable packaging
3.6 Terminal
3.6.1 Terminal with injector S/N lower than CTM1608001
Figure 9. Configuration of the terminal front with injector S/N lower than CTM1608001
1 Display 3 Pedestal
2 Main switch
Figure 10. Configuration of the terminal back with injector S/N lower than CTM1608001
Figure 11. Configuration of the terminal back with injector S/N CTM1608001 or higher
Figure 12. Configuration of the terminal front with injector S/N CTM1721215 or higher
1 Display 3 Pedestal
2 Main switch (on the back of the
terminal)
Figure 13. Configuration of the terminal bottom with injector S/N CTM1721215 or higher
1 Two USB interfaces for Bluetooth 2 Connection point for 12V DC mains adapter
dongle and additional USB sticks
3 Main switch
1 Valve assembly with valve membrane 7 Pump impeller with 3 guide rollers and 3
2 Cover locking mechanism active pump rollers
6 Hose threading/spindle
Figure 15. Configuration of the device base with S/N lower than CTM1510600
Figure 16. Configuration of the interior device base with S/N lower than CTM1510600
Figure 17. Configuration of the device base with S/N CTM1510600 or higher
Figure 18. Configuration of the interior device base with S/N CTM1510600 or higher
4 Starting up
4.1 CT motion pedestal version
4.1.1 Assembling main assemblies
Service flap opened (see section 7.2.1, p. 7-86).
Connecting cables
1. Connect cables (see section 7.3.1, p. 7-86).
2. Tighten four mounting screws (1) and washers
with the hexagon socket, hex 8.
3. Insert front module and close (see section 7.3.1, p. 7-86). In doing so, ensure that the
cables are not pinched.
7. Enter serial number noted on the terminal under Menu ► Setting ► System ►
Bluetooth. Only enter digits.
8. Shut down terminal and restart.
Connection icon on the terminal display
changes from “no connection” (1) to
“connection established” (2).
Software 04.01.02 or newer is required for terminals with injector S/N CTM1721215 and
higher.
WARNING! Ceiling version may fall down or collide with scanner or ceiling.
Patients or operators may be injured or patients may be exposed to radiation a second
time which includes a double dose of contrast media.
Check if the room height is sufficient for the installation.
Check the distance to the scanner.
Check torques for components already installed.
Secure mount from inadvertent handling if assembly is not finished in one day.
Make sure that the Ondal ceiling version is installed in accordance with the
regulations to ensure proper functioning.
Grounding connection
1. Remove grounding bolts (1) present and screw
screw M4 x 70 into drill hole.
3. To crimp the grounding wires: Use ring cable lugs contained in the assembly set.
4. At the grounding point (M4 bolt) of the ceiling tube flange, connect the following grounding
cables:
– Grounding conductor (PE) of cable 44 with preassembled cable lug
– Grounding conductor (PE) securely connected to the in-house network with the cable
lug preassembled by the customer.
– Grounding conductor (PE) of the cable 3b to the power supply unit. Crimp grounding
conductor beforehand with the cable lug present on the flange plate (DIN 46234-4-2.5
for M4 bolt and wire cross-section of 1 mm² to 2.5 mm²).
– Grounding conductor (PE) of the cable 7c to the device supply. Crimp grounding
conductor beforehand with the cable lug (DIN 46234-4-2.5 for M4 bolt and wire cross-
section of 1 mm² to 2.5 mm²).
– Grounding of the fixed-height arm for CT motion
– For a double arm: Grounding of the second fixed-height arm for the third-party device
The C-arm for CT motion does not have any PA connection and no PE grounding
connection since the C-arm cannot come into contact with mains voltage.
WARNING! Wall version may fall down or collide with scanner or ceiling.
Patients or operators may be injured or patients may be exposed to radiation a second
time which includes a double dose of contrast media.
Check if the room height is sufficient for the installation.
Check the distance to the scanner.
Check torques for components already installed.
Secure mount from inadvertent handling if assembly is not finished in one day.
Make sure that the Ondal wall version is installed in accordance with the regulations
to ensure proper functioning.
Grounding connection
All components of the power supply unit box have already been factory-grounded with
appropriate grounding cables.
Figure 60. Feeding unit, back – injector Figure 61. Device base, back – injector
S/N lower than CTM1721441 S/N lower than CTM1510600
Figure 62. Feeding unit, back – injector Figure 63. Device base, back – injector
S/N CTM1721441 or higher, S/N CTM1510600 or higher
pedestal version
5.3.2 Cable connection FOC and power cable/main switch (if available)
Checking condition of the mains connection cable
Ensure that the mains connection cable is clean and free of defects:
– No visible damage such as, for example, cracks, kinks, pinching or faulty insulation.
The function of the air bubble detector on the patient connection is checked during the
final performance test.
Replace hose guide rolls at least every two years during the TSC.
1. Ensure that the pump wheel and the rollers including bearings are clean and free of
defects.
2. If needed: Clean pump wheel, rollers including bearings.
3. Ensure that the hose guide rolls move smoothly and are undamaged.
4. Ensure that the coating of the counter bearing is undamaged.
5.5 Terminal
5.5.1 Inscriptions
Ensure that all inscriptions and stickers are
complete and legible:
– Connections of devices with injector S/N
lower than CTM1721215 (see Figure 66,
p. 5-54)
– Connections (1) of devices with S/N
CTM1721215 or higher (see Figure 68,
p. 5-54)
– On/off button before S/N CTM1721215
(see Figure 65)
– On/off button (2) of devices with S/N Figure 65. On/Off button – injector S/N
CTM1721215 or higher (see Figure 68, lower than CTM1721215
p. 5-54)
– Type label of devices with injector S/N
lower than CTM1721215 (see Figure 67,
p. 5-54)
– Type label of devices with injector S/N
CTM1721215 or higher (see Figure 69,
p. 5-54)
5.5.3 On/off button, sockets, connectors, cables and power supply unit
Ensure that the on/off button, sockets, connectors, cables and power supply unit are clean
and free of defects.
5.5.4 Display
1. Check display with regard to function; switch display on and off.
2. Ensure that the display is clean and free of defects.
3. Ensure that any pixel errors which may occur do not impair the legibility of the display
texts.
5.5.5 Loudspeaker
1. Select Menu ► System settings ► Volume in the terminal.
Volume input field appears.
2. Check the operability of the loudspeaker by pressing the Test button.
A rhythmic, intermittent tone will be heard at maximum volume at the terminal.
5.6 Charger/battery
5.6.1 Battery display on the injector display
Ensure that the charging status of the battery is shown correctly on the injector display.
5.7 CM heater
5.7.1 Heater module
Ensure that the lid mounts for media containers are clean and free of defects.
– No visible damage such as cracks, dents, broken or missing connecting elements.
– Door hinges can move smoothly.
– Door closes easily.
Heating element
Measure resistance with multimeter between pin 1 and pin 2 of the heater.
Target value: 80 Ω ±20 Ω
Sensor
1. Measure resistance of both sensors at room temperature:
− Sensor 1: PIN 3 and PIN 4
− Sensor 2: PIN 5 and PIN 6
2. Estimate room temperature.
3. Actual values compared with target values in diagram.
− Maximum deviation from target values: ±20 Ω
− Maximum deviation between sensors: ±4 Ω (correspond to approx. 1°C deviation)
Extension cords between the connection cable and test device according DIN EN 62353
(VDE 0751-1) make the measured values worse.
Measure grounding conductor resistance at the following measurement points with test
device according to DIN EN 62353 (VDE 0751-1).
Mobile stand:
1st measurement point: Bonding (injector pedestal)
2nd measurement point: Lower cable holder (mains cable)
Ceiling suspension:
1st measurement point: Extension arm
2nd measurement point: C arm
As a test liquid, NaCl, water or, if available, contrast medium can be used.
When conducting functional testing with dummies, the injector may return different results
that in the case of tests with a complete tubing system.
5.9.6 CM sensor
1. Vent both CM accesses.
2. On the terminal, adjust test program with a flow rate of 7 mL/s and 50 mL CM.
3. Start test program.
4. Pivot CM bottles and bottle lift forward.
Injector switches from the current active CM bottle to the inactive bottle. The switch must
take place before air passes through the CM valve.
5. Repeat test for the second CM bottle.
The target value range for pressure may exceed the pressure limit momentarily because
the manometer is set to measure peaks. This is not a malfunction of the injector.
Test Volume Flow rate Cannula Target value range Target value
prog. [mL] [mL/s] [mm] for volume [mL] range for time [s]
1 80 8 No 76 – 84 8.5 – 11.5
2 80 2 1.1 x 33 76 – 84 38.5 – 41.5
9. Fill 1-mL syringe completely with air and inject upwards via 3-way valve. In doing so,
ensure the following:
– Distance between air detector and bubble at least 50 cm so that the bubble passes
through the detector at full speed.
– Bubble must be closed.
10. Start injection with 10 mL/s.
Motor switches off if the bubbles pass through the air trap.
Air bubbles may be transported to the end of the pump tubing at a maximum.
A warning message appears on the injector display.
An acoustic warning sounds and a warning message appears on the terminal display.
The opening pressure of the valves is reached if the water is expelled at the end of the
mobile tube testing unit.
Test reports
Further details can be found in the following documents:
1. Test report CT motion XD 8000 (F-04-05-23)
2. Functional testing protocol wall/ceiling mount (F-04-05-06)
6 Service program
The authorized service technician or distributor is required to ensure that
the logfiles are saved regularly, sustainably and on a long-term basis.
the service program is always kept updated. The most current version is available on the
service SharePoint:
https://sp1178.hostedoffice.ag/Ulrich-Service-Extern/default.aspx.
new updates are required to be installed as soon as possible.
9. Select a folder via Folder recording data (2) to define where to store the transferred data
of the sensors (see Figure 75, p. 6-65).
10. Select a folder via Update folder (3) to define where to store the injector update file.
11. Click OK (4) to confirm your selection.
12. Click Connect device (6) to set up the
connection to the injector. If the connection is
successfully established, the green Heartbeat
LED is blinking.
The firmware of the injector consists of three files packed in a ZIP file. Optionally, parameters,
display texts and character set may be present in the ZIP file.
After each update check if the update was successful by reading the version numbers:
1. Select Information in the side navigation.
2. Check version numbers.
To distinguish the roll pump versions easily from each other, refer to the figures below:
- The counter bearing of the new roll pump is completely coated (1) and the holder is
metal (2).
- The counter bearing of the old roll pump is only coated on the inside (3) and the
holder is covered with a plastic cover (4).
Figure 87. New roll pump Figure 88. Old roll pump
If the tubing system is not filled, click Fill (5) to fill the NaCl part of the pump tubing and
the patient tubing.
Use an appropriate pump tubing to adjust the toll pump correctly. A tolerance of 100 +/-
2.5 ml with correction factor 100 is acceptable for appropriate pump tubing.
One year later to the entered date the user will receive a reminder to have conducted a
TSC again.
7 Service work
7.1 Software
After finish of the service activity, the following points are to be checked and documented
in the test report F-04-05-23: 5 (see section 5.13, p. 5-63).
Flash Drive, 16 GB
The Servicekey.txt file is an empty text file. This can be created on one's own.
Be aware that the file is case-sensitive.
2. Update software:
– Use the following command to switch to the directory with update files:
cd□/media/PKIusb/Rev123
– Start update with the following command:
sh□installTerminalSW.sh□123
– Follow the installation script instructions.
The license file must be updated at the initial installation of the terminal software or during
changes to the available options.
For the update, the most recent version on the USB stick is automatically selected, for
example, Rev231 is more recent than Rev221.
In order to avoid errors with an update to version 04.00.03 or newer, the cables 46 and 22b
need to be disconnected.
Therefore, the following steps must be carried out:
1. Open injector:
Pull out front module (see section 7.3.1,
p. 7-86).
Cable 46 of filter holder (1) and 22b of reed
switch (2) are accessible.
Updating software
1. Copy update folder RevXYZ (i.e. Rev403) to
the servicestick.
2. Insert servicestick at terminal.
3. Open Menu ► Service Menu ► Software
Update.
4. Software update is carried out.
5. Application is restarted.
Assembly
1. Place service flap SP000014-ET on service window (see Figure 120, p. 7-86)
2. Completely screw in both screws (2) into service flap (1) with Phillips screwdriver size 7.
The safety screws can be best unscrewed using a shortened hexagon socket wrench.
Assembly
1. Mount front module in holder (see Figure 122, p. 7-87):
– Insert front module in holder.
– Screw in safety screws (1) on both sides of the front module all the way using the
hexagon socket wrench, size 5.
2. Connect cables:
– Connect power plug.
– Connect charger electrical components cable.
– Attach both grounding conductors.
– If the fiber-optic cable (FOC) is used: Connect FOC cable.
3. Insert front module and close (see Figure 121, p. 7-86).
– Push front module holder onto spherical heads (2) all the way to the back.
– Screw in grub screws (1) right and left all the way using the hexagon socket wrench,
size 5. In doing so, ensure that there is only a minimal gap between the housing and
the front module.
– Loosen locking device (ceiling mount)
To reduce the gap between the housing and the front module:
Press down alternately on both sides of the front module and further tighten the grub
screws.
WARNING!
Exposed electrical components
Electrical hazard
Break circuit before servicing in opened device base.
NOTICE!
Temporary high currents while closing circuit
Blowing fuse of LP5 charger electrical components
Bypass fuse with conductive object while closing circuit.
Disassembly
1. Prize open three dummy plugs over safety
screws (1) at device base using a small
screwdriver and remove (see Figure 123).
2. Loosen three safety screws (1) at device base
using Torx screwdriver size 20 and remove
with washers (see Figure 123).
Assembly
Mount device head on base of device (see section 4.1.1, p. 4-32).
Disassembly
1. Detach cable connections on board.
2. Loosen and remove two mounting screws (1)
on board.
3. Remove board by lifting it out.
Assembly
1. Insert board in holder.
2. Completely screw two mounting screws (1) into the board (see Figure 128, p. 7-90).
3. Reconnect cable and board.
7.4.3 Replacing LP5 charger electrical components of device base with S/N CTM1510600 or
higher
Disassembly
1. Remove cable connections from board.
2. Loosen and remove four mounting screws (1)
on board with Torx screwdriver size 10 (see
Figure 129).
3. Remove board
Assembly
1. Place board properly on boreholes in device base.
2. Completely screw four mounting screws (1) with Torx screwdriver size 10 in board (see
Figure 129).
3. Reconnect cable and board.
7.4.4 Replacing Fuse 5A FF for Charger Electrical Components before S/N CTM1510600
Device base opened (see section 7.4.1, p. 7-88).
7.4.5 Replacing Fuse 5A FF for Charger Electrical Components at device base with S/N
CTM1510600 or higher
Disassembly
Carefully prize out and remove fuse (2) with small screwdriver (see Figure 129, p. 7-90).
Assembly
Insert fuse (2) in holder (see Figure 129, p. 7-90).
The fuse holder may bend outwards when the fuse is removed.
Bend fuse holder back to the original position.
Disassembly
1. Detach connector from LP5 board.
2. Loosen and remove two mounting screws (2)
on white fixing socket.
3. Remove charging transistor together with fixing
socket.
4. Loosen two additional mounting screws to
detach fixing socket from transistor.
5. Remove cable 11 with charging transistor
SP000371-ET.
Figure 131. Cable 11 battery charging
transistor
Assembly
1. Completely screw two mounting screws into fixing socket and transistor.
2. Completely screw fixing socket using two mounting screws (1) into the device base (see
Figure 131, p. 7-91).
3. Insert connector in connection of LP5 board.
7.4.7 Replacing cable 11 battery charging transistor of device base with S/N CTM1510600 or
higher
Disassembly
1. Remove connector from LP5 board.
2. Loosen and remove two mounting screws (1)
on device base with Torx screwdriver size 10
(see Figure 132).
3. Remove charging transistor with fixing socket.
4. Loosen two additional mounting screws (3)
from fixing socket (2) with Torx screwdriver size
15 to remove fixing socket from transistor (1)
(see Figure 133, p. 7-92).
5. Remove cable 11 with charging transistor Figure 132. Cable 11 battery charging
SP000371-ET. transistor
Assembly
1. Screw two mounting screws (3) with Torx
screwdriver size 15 into fixing socket (2) and
transistor (1) (see Figure 133).
2. Screw fixing socket with two mounting screws
(1) and Torx screwdriver size 10 into device
base (see Figure 132, p. 7-92).
3. Insert connector in connection of LP5 board.
Disassembly
1. Remove cable ties and position cable such that
the power supply unit can be reached.
2. Loosen screws on inlet cables (1) and remove
cables.
3. Loosen and remove four mounting screws (2).
4. Remove power supply unit.
Figure 134. Power supply unit
Assembly
1. Insert power supply unit in device base.
2. Completely screw in four mounting screws (2) (see Figure 134, p. 7-92).
3. Completely screw inlet cable (1) into inlets using screws.
4. Attach new cable ties to cables.
5. Measure the following reference values using
Secutest:
6. Grounding conductor resistance
7. Device leakage current
8. Insulation resistance
The measurement should be taken at bonding
bolts (1) and ECG grounding cable
connection (2).
7.4.9 Replacing power supply of device base with S/N CTM1510600 or higher
Disassembly
1. Loosen and remove screws (1) with Torx
screwdriver size 20 (see Figure 137).
2. Lift power supply unit out of device base,
loosen screws on inlet cables (2) with Phillips
screwdriver and remove cables (see Figure
137).
3. Remove power supply unit.
Assembly
1. Screw inlet cables with screws and Phillips
screwdriver into inlets (see Figure 138 and
Figure 139, p. 7-94).
2. Put power supply unit into device base.
3. Screw in four mounting screws (1) with Torx
screwdriver size 20 (see Figure 137, p. 7-93).
Disassembly
Assembly
1. Place battery (3) in device base (see Figure 141, p. 7-95).
2. Rotate battery holder (2) over screw (1) and insert screw through drill hole (see Figure
140, p. 7-95).
3. Completely screw nut onto screw (1).
4. Put connectors and connection cable on batteries.
7.4.11 Replacing batteries SP000445-ET of device base with S/N CTM1510600 or higher
Disassembly at device base from S/N CTM1510600
1. Loosen and remove screw (1) on closed side of
battery holder with Torx screwdriver size 20
(see Figure 142).
2. Loosen screw (4) on opened side of battery
holder (2) with Torx screwdriver size 20 by
three turns (see Figure 142).
3. Push battery holder (2) to the left out of screw
(4) and put it aside (see Figure 142).
4. Loosen connectors on batteries (see Figure
143, p. 7-96). Figure 142. Batteries
5. Remove batteries (3) (see Figure 142).
Ensure that cables which are connected to the battery holder are below the battery
holder during the assembly.
Disassembly
1. Remove cable ties on cable 14.
2. Loosen connector (1) on board.
3. Carefully pull LEDs (2) out of the LED holder.
Assembly
1. Insert LEDs (2) carefully in LED holder (see Figure 144, p. 7-97).
2. Insert connector into (1) connection point on board.
3. Attach new cable ties to cable.
7.4.13 Replacing cable 14 LEDs battery charging of device base with S/N CTM1510600 or
higher
Disassembly
1. Remove cable tie from cable 14 (1).
2. Remove cable tie from cable collection on
separating wall (2) (see Figure 145).
3. Remove connector (1) from board (see Figure
146, p. 7-98).
4. Carefully pull LEDs (3) out of LED holder (see
Figure 144, p. 7-97).
Assembly
1. Carefully insert LEDs (2) in LED holder (see
Figure 144, p. 7-97).
2. Insert connector (1) into connection point on
board (see Figure 145, p. 7-97).
3. Attach new cable tie to cable collection on
separating wall (2) (see Figure 145, p. 7-97).
4. Attach new cable tie to cable (1) (see Figure
146).
Disassembly
1. Loosen locking nut on double caster with
wrench size 22 (see Figure 147).
2. Unscrew and remove double caster with
wrench size 22 from device base.
Assembly
1. Secure the threads of the double caster Loctite 243.
2. Screw double caster with wrench size 22 onto device base.
3. Completely screw locking nut with wrench size 22 on double caster (see Figure 147,
p. 7-98).
During service work on injector head, screws can fall through the clearance hole between
the column and the head.
Seal clearance hole between the column and the head.
Assembly
1. Place gas spring (2) on holder (1) (see Figure 150, p. 7-100).
2. Place locking washer on holder.
3. Completely screw in gas spring clockwise in attachment on housing (see Figure 149,
p. 7-99).
4. Completely screw in both safety screws (2) on holder of gas spring (1) using Torx
screwdriver size 20 (see Figure 148, p. 7-99).
Disassembly
1. Loosen and remove four mounting screws (2)
and (3) on the rail system (1) using the angled
Torx wrench, size 20.
2. Remove rail system (1).
Assembly
1. Fully screw in two mounting screws in drill holes (3) on the rail system (1) in the injector
head housing. To do this use Torx wrench size 20 (see Figure 151, p. 7-100).
2. Screw two mounting screws into drill holes (2) in the same way. In doing so, ensure that
the rail system is parallel to the cross plate and the guide pins.
Disassembly
1. Remove cable ties (3) on cross plate.
2. Loosen and remove four mounting screws (2)
on cross plate (1) using Torx screwdriver size
20.
3. Remove cross plate (1).
Assembly
1. Position/place cross plate (1) on injector head housing (see Figure 152, p. 7-101).
2. Completely screw in four mounting screws (2) on cross plate (1) using Torx screwdriver
size 20.
3. Secure cable on cross plate using cable ties (3).
Disassembly
The safety screw can be best unscrewed using a shortened hexagon socket wrench.
Assembly
Completely screw safety screw into rail system using hexagon socket wrench size 5 (see
Figure 153, p. 7-101).
Disassembly
1. Loosen connector of patient tubing sensor from cable 26.
2. Detach cable ties on fixing socket MB 3A
(SP000260-ET) (3).
3. Unscrew both mounting screws (2) using Torx
screwdriver size 10 and remove together with
washers.
4. Remove media sensor (1).
Figure 154. Media sensor ABD patient
tubing
Assembly
1. Place O-ring in holder for media sensor.
2. Insert media sensor (1) in holder for media sensor (see Figure 154, p. 7-102).
3. Completely screw in both mounting screws (2) using a Torx screwdriver, size 10.
4. Attach cable ties on fixing socket MB 3A (SP000260-ET) (3).
5. Connect connector of the ABD with connection cable of media sensor on cable 26
(SP000385-ET).
Disassembly
1. Loosen cable of high-pressure filter holder from
reed switch.
2. Loosen two screws (1) on filter holder (2) with
Torx screwdriver, size 20, shortened in length.
Assembly
1. Introduce high-pressure filter holder (1) in feeding unit and screw in using screws (2) (see
Figure 155, p. 7-102).
2. Connect cable of high-pressure filter holder with reed switch.
Disassembly
1. Remove cable ties (1).
Assembly
1. Push cable through the opening from outside to inside.
2. Thread in threaded ring (1), washer and plastic ring (see Figure 159, p. 7-103).
3. Fasten threaded ring (1) with a screwdriver (see Figure 157, p. 7-103).
4. Connect cable lug to the cable with crimping tool.
5. Attach ECG cable with cable tie (1) to injector housing (see Figure 156, p. 7-103).
Make sure that the cable tie is loose enough to alow free movement of the cable.
6. Attach ferrite (2) with cable tie to ECG cable and attach the cable tie to the mount (see
Figure 158, p. 7-103).
7. Attach cable lug to grounding point (1) (see Figure 158, p. 7-103).
Assembly
1. Insert two pins for drip cup holder
SP000042-ET (2).
2. Fasten pins with two M5 nuts and washers (1).
3. Attach drip cup holder SP000043-ET to pins
(2).
Disassembly
1. Remove three screws (1) with Torx
screwdriver, size T20.
2. Remove tray.
Assembly
1. Fit injector head housing to injector base and tighten four mounting screws (8) and
washers with hexagon socket, hex 8 (see Figure 167, p. 7-106).
2. Connect cables (see section 7.3.1, p. 7-86).
3. Attach two receptacles of locking bolts (7) to new injector head housing using two screws
(6) with Torx screwdriver, size T8 (see Figure 166, p. 7-106).
4. Push feeding unit into the front.
5. Fasten feeding unit to housing using six screws (5) with Torx screwdriver, size T20 (see
Figure 165, p. 7-106).
6. Attach cables with cable ties to housing (see section 7.5.7, p. 7-103).
7. Attach connector (3) for grounding cables with one screw to housing using Torx
screwdriver, size T20 (see Figure 163, p. 7-105).
8. Connect grounding cables (2) to connector (3) (see Figure 163, p. 7-105).
9. Attach tray with three screws (1) using Torx screwdriver, size T20 (see Figure 162,
p. 7-105).
Disassembly
Assembly
1. Thread in cable.
2. Push On/Off switch from outside to inside.
3. Connect cable of On/Off switch to power board.
After mounting On/Off switch SP002223-ET please update the firmware of the injector
to SW 04.00.03 or newer.
Disassembly
1. Remove cable tie (1).
2. Remove connectors of on/off switch from
boards (2 and 3).
Assembly
1. Put cable of On/Off switch through the opening
inside the front module (see Figure 172,
p. 7-109).
2. Push fastening bushing into the front module
according to the shape of the opening (6).
3. Push On/Off switch into the fastening bushing
(7).
4. Connect the connectors of the On/Off switch to
the boards (2 and 3) (see Figure 170,
p. 7-108).
5. Attach cable tie (1) (see Figure 170, p. 7-108). Figure 173. On/Off switch in front
module
Functioning test
Additionally to the test items mentioned above, switch the injector on and off again to
check the function of the On/Off switch as well as its blue light.
Assembly
Place bottle lifts on cover ABD SP000357-ET.
1. Completely screw in thumb screws 1 into bottle lifts (see Figure 176, p. 7-110).
2. Fold up bottle lifts.
3. Close heater assembly (if present).
Disassembly
1. Loosen and remove two screws (1) on the back
on guide bushing for spring catch using Torx
screwdriver size 10.
2. Press spring catch (2) inwards and remove.
Assembly
1. Push spring catch inwards and place in holder holes of the bottle lift (see Figure 177, p. 7-
110).
2. Completely screw in two screws (1) on the back on guide bushing for spring catch using
Torx screwdriver size 10.
Disassembly
1. Open retaining clip (1).
2. Remove holding arm for bottle guide by lifting it
out.
Assembly
Push holding arm for bottle guide (SP000304-ET) off bottle lift.
Assembly
Push spike holder onto bottle lift.
Assembly
1. Place cover feeding unit on feeding unit.
2. Insert wiring (1) of heater (see Figure 182, p. 7-112).
3. Fold up cover feeding unit (see section Figure 181, p. 7-112).
4. Completely screw in three screws (1) on cover feeding unit with Torx screwdriver size 20
(see Figure 180, p. 7-112).
Disassembly
1. Loosen and remove four screws (1) of cover
media sensor (SP000357-ET) with Torx
screwdriver size 20.
2. Remove cover media sensor (SP000357-ET).
Assembly
1. Insert media sensor ABD with mounted O-Ring for media sensor in spike bearing
SP000054-ET.
2. Completely screw in screws in the holder on the media sensor using Torx screwdriver
size 8 (see Figure 184, p. 7-113).
3. Guide cable with additional O-Ring (1) through
recess (2) in spike bearing.
4. Guide cable over handrail and insert through
recess in feeding unit.
Disassembly
1. Disassemble conductor board holder:
− Loosen six screws (1) with Torx
screwdriver size 20.
− Remove conductor board holder.
Assembly
1. Insert pressure measurement complete (SP000096-ET) in the front module.
2. Screw in grounding cable (1) with screw using Torx screwdriver size all the way in
pressure measurement complete (see Figure 190, p. 7-115).
3. Completely screw in four cylinder head screws Torx 10, M3x8, A2 SP000222 (part of
SP001996-ET) with locking washers shape S for M3 SP000574 (part of SP001996-ET)
and washers 3.2, A2 SP000310 (part of SP001996-ET) (2) using a Torx screwdriver, size
10 into the front module (SP000081-ET).
4. Place conductor board holder on front module.
5. Completely screw in six screws (1) using Torx screwdriver size 20 (see Figure 189,
p. 7-115).
Disassembly
1. Remove whiteTorx screw covers.
2. Loosen and remove screws with Torx screwdriver, size 20.
Because of the press fit, it may be difficult to pull out the hose guide roll.
Use significant force if necessary to pull out the hose guide roll.
Assembly
1. Clean all components and affected sites.
Disassembly
1. Detach fixation:
− Hold nut (1) on fixation with open-end
wrench, hex 8.
− Loosen screw (2) on fixation using Torx
screwdriver, size 20, and remove together
with nut.
2. Loosen and remove nuts (3) on all four guiding
bolts using open-end wrench, hex 10.
The ring segment on the left (viewed from the front) may fall when the pressure ring is
removed.
Assembly
1. Grease counter bearing on the overlying, oblique outer side with Gleitmo 810.
2. Loosen threaded ring using special tool XD 8123.
3. Insert counter bearing (1) with the dowel pin facing upwards in the housing of the roll
pump (see Figure 192, p. 7-117).
4. Insert pressure ring (2) in housing of the roll pump.
5. Completely screw in nuts (3) on all four guiding bolts using open-end wrench, hex 10 (see
Figure 193, p. 7-117).
6. Attach fixation:
− Hold nut (1) on fixation with open-end wrench, hex 8.
− Completely screw in screw (2) on fixation using Torx screwdriver size 20.
7. Adjust rollers (see section 7.8.3, p. 7-118).
1. Detach fixation:
− Hold nut (1) on fixation with open-end
wrench, hex 8.
− Loosen screw (2) on fixation using Torx
screwdriver, size 20, and remove together
with nut.
7.8.4 Replacing cover cap roll pump SP000162-ET before S/N CTM1721200
Injector cover opened.
Disassembly
For devices prior to S/N CTM1300247, the cover caps are adhered.
For devices starting with S/N CTM1300247, the cover caps are simply mounted.
Assembly
1. For devices prior to S/N CTM1300247: Secure the edge of the cover cap (1) with Loctite
454.
2. Place cover cap (1) on roll pump shaft and press down.
7.8.5 Replacing cover base part roll pump before S/N CTM1721200
Service flap opened (see section 7.2.1, p. 7-86).
Front module removed (see section 7.3.1, p. 7-86).
Hose threading disassembled (see section 7.9.1, p. 7-125).
Disassembly
1. Remove connectors:
− Loosen ribbon cable (1) on controller
board.
− Loosen connectors (2) on drive roll pump.
− Remove drive (see section 7.9.2, p. 7-126).
Assembly
1. Secure drive roll pump on cover base part (see Figure 192, p. 7-117):
− Place drive on cover base part.
− Clean screws and secure thread with Loctite 243.
− Completely screw in four locking screws (1) into cover base part using Torx
screwdriver size 25.
2. Secure drive roll pump with cover base part onto pump housing (see Figure 191,
p. 7-116):
− Place drive roll pump (3) together with cover base part and cable on pump housing.
− Clean screws and secure thread with Loctite 243.
− Place grounding cable on drill hole (2) and fully screw in screw with grounding cable
on cover base part using Torx screwdriver size 20.
− Completely screw in three locking screws on cover base part using Torx screwdriver
size 25 (1). In doing so, screw in the two longer screws in the holding plate and the
two shorter screws in the pump housing.
3. Connect connectors (see Figure 189, p. 7-119):
− Lay cable under the holding plate.
− Connect connectors (1) on drive roll pump.
4. Secure offset base (see Figure 190, p. 7-115):
− Clean locking screws and secure thread with Loctite 243.
− Completely screw in locking screw (1) into offset base (2) using Torx screwdriver size
20.
5. Mount drive (see section 7.9.2, p. 7-126).
Disassembly
1. Loosen and remove set screws (2) in
coupling (1) using hexagon socket wrench size
2.
2. Loosen and remove set screws on counter
piece of the coupling on housing upper part
using hexagon socket wrench size 2.
3. Remove coupling (1).
Assembly
1. Place coupling (1) over fitting key on drive shaft on motor (see Figure 202, p. 7-121).
2. Clean set screws and secure thread with Loctite 243.
3. Completely screw in set screws in counter piece of the coupling on housing upper part
using hexagon socket wrench size 2.
4. Completely screw in set screws (2) in coupling (1) using hexagon socket wrench size 2.
7.8.7 Replacing roll pump (upper component) SP000106-ET before S/N CTM1721200
Service flap opened (see section 7.2.1, p. 7-86).
Front module disassembled (see section 7.3.1, p. 7-86).
Hose threading disassembled (see section 7.9.1, p. 7-125).
Cover base part for roll pump disassembled (see section 7.8.5, p. 7-119).
Disassembly
1. Remove compensation coupler SP000710-
ET (1).
Assembly
1. Place roll pump (upper component) SP000106-ET on front module. In doing so, ensure
the following:
− The drive shaft of the hose threading is correctly positioned in the upper component.
− The upper part does not attach to the drive shaft.
2. Completely screw in four screws (1) in the roll pump (upper component) using Torx
screwdriver size 30 (see Figure 205, p. 7-122).
3. Completely screw in two screws (2) into roll pump (upper component) using Torx
screwdriver size 25.
4. Push shim ring SP000740-ET (1) on hose threading (see Figure 204, p. 7-122).
5. Push spring-lock washer (2) on hose threading.
6. Place compensation coupler SP000710-ET (1) on hose threading (see Figure 203, p. 7-
122).
7. Check gap between rollers and counter bearings and adjust if necessary (see section
7.8.3, p. 7-118).
To adjust the roll pump after mounting, please update the firmware of the injector to
SW 04.00.03 or newer. Adjust the volume via service program 2.5 or newer (see section
6.6.7, p. 6-69).
Disassembly
1. Remove connectors (1).
2. Remove cable tie (2) from USB cable.
Assembly
1. Place roll pump on the inside of the front module.
2. Completely screw in four screws (8) into front module using Torx screwdriver, size 30 (see
Figure 210, p. 7-124).
3. Push shim ring (7) on hose threading (see Figure 209, p. 7-124).
4. Push spring-lock washer (6) on hose threading.
5. Place compensation coupler SP000710-ET (5) on hose threading (see Figure 208, p. 7-
124).
6. Mount roll pump drive (see section 7.9.4, p. 7-127).
7. Attach grounding cable (3) incl. locking washer and washer with a Torx screwdriver, size
20 (see Figure 207, p. 7-123).
8. Connect connectors (1 and 2) on power board and controller board (see Figure 206, p. 7-
123).
9. Adjust the roll pump with the service program (see section 6.6.7, p. 6-69).
If the roll pump is not adjusted with the service program, a motor error may be displayed.
Disassembly
1. Remove closing plug (1) using suitable, pointed
tool.
2. Remove screw under closing plug:
− Hold posterior part of drive shaft with open-
end wrench, hex SW 5.5.
− Loosen and remove screw with Torx
screwdriver size 20.
3. Pull spindle (2) out of holder.
When the drive shaft is pulled out, the locking washer, shin ring and compensating
coupling SP000710-ET fall out.
Seal clearance hole between the column and the head.
Assembly
1. Grease sleeve bearing sites (1) on drive shaft
SP000159-ET with Gleitmo 810.
Disassembly
1. Pull out spring-lock washer (1) for hose
threading on the inner side of the front module
using an appropriate tool.
2. Remove hose threading with shim ring (3) and
compensation coupler SP000710-ET (2).
Assembly
1. Push drive shaft of hose threading into front module and hold it.
2. Place shim ring (3) and compensation coupler SP000710-ET (2) on drive shaft of the
hose threading.
3. Push spring-lock washer on the hose threading between shim ring (3) and compensation
coupler SP000710-ET (2).
Disassembly
1. Loosen cable 21 (2).
2. Loosen and remove screws (3) with Torx
screwdriver size 10.
3. Remove drive (4).
Assembly
1. Ensure that compensating coupling SP000710-ET (1) is correctly positioned during
reassembly:
− Axis of drive shaft and motor must be in alignment (see Figure 215, p. 7-127).
2. Place drive (4) on offset base.
3. Completely screw in screws (3) using a Torx screwdriver, size 10.
4. Place cable 21 (2) in power board SP-000216-ET.
5. Lay cable under holding plate of roll pump.
7.9.4 Replacing drive of hose threading SP000378-ET with S/N CTM1721200 or higher
Service flap opened (see section 7.2.1, p. 7-86).
Front module removed (see section 7.3.1, p. 7-86).
Disassembly
1. Loosen cable from power board SP000216-ET
(1).
2. Loosen and remove screws (2) with Torx
screwdriver, size 10.
3. Remove drive (3).
Assembly
Ensure that compensation coupler SP000710-
ET (4) is correctly positioned and axis of drive
shaft and motor are aligned.
1. Place drive (3) on offset base (see Figure 216,
p. 7-127).
2. Completely screw in screws (2) using a Torx
screwdriver, size 10.
3. Put cable (1) in power board SP000216-ET.
Disassembly
1. Remove cables of the valve assembly (1)
(CM1≙V1; NaCl≙V2; CM2≙V3).
2. Loosen four locking screws (2) with Torx
screwdriver size 20 and remove together with
washers.
3. Remove valve assembly.
Ensure that each solenoid block is as parallel as possible to the valve tappet.
Check that the outer edge of the solenoid block and the edge of the holding arm are
parallel.
6. Perform fine adjustment of the valve assembly (see section 5.9.11, p. 5-61).
7. Secure position of the solenoids against shifting:
− Fully tighten both safety screws (grub screws SP000548) using hexagon socket
wrench size 2.
8. Dismantle alignment gauge:
− Open injector cover.
− Remove alignment gauge valve assembly XD 8075 from protective cap for valve
assembly.
− Close injector cover.
Disassembly
1. Remove locking toggle:
− Remove screw cap (1) on locking
toggle (2).
− Loosen screw under screw cap (1) using a
Torx screwdriver size 25 and remove.
− Remove locking toggle (2).
Assembly
1. Secure cover locking mechanism (see Figure 224, p. 7-131):
− Place cover locking mechanism (2) on front module.
− Completely screw in four safety screws (1) into cover locking mechanism (2) using
Torx screwdriver size 25.
2. Secure light barrier (see Figure 223, p. 7-131):
− Place light barrier (2) on cover locking mechanism.
− Completely screw in both screws (1) into light barrier with Torx screwdriver size 8.
3. Secure cable 35 SP000391-ET with cable ties to cover locking mechanism (see
Figure 224, p. 7-131).
4. Place drive shaft head (1) on square socket (2) of the drive shaft (see Figure 221,
p. 7-130).
5. Secure locking toggle (see Figure 220, p. 7-130).
− Place locking toggle (2) on drive shaft head. In doing so, ensure that the locking
toggle is horizontally aligned.
− Secure screw using Loctite 245.
− Fully screw the screw into the locking toggle using a Torx screwdriver, size 25.
− Place screw cap (1) over screw in the locking toggle.
7.11.2 Replacing cover locking mechanism SP002026-ET with S/N CTM1721200 or higher
Service flap opened (see section 7.2.1, p. 7-86).
Front module pulled out (see section 7.3.1, p. 7-86).
After mounting the cover locking mechansim please update the firmware of the
injector to SW 04.00.03 or newer.
Disassembly
1. Remove white cover of Torx screw (1) (part of
SP001996-ET).
Assembly
1. Insert slide bearing (8) in opening in the front module (see Figure 228, p. 7-133).
2. Place cover locking mechanism (7) on slide bearing aligned with screw holes (6) (see
Figure 227, p. 7-133).
3. Completely screw in three safety screws (6) into cover locking mechanism (7) using Torx
screwdriver, size 25.
The locking mechanism is automatically adjusted by the firmware after turning the injector
back on.
Assembly
1. Hold device cover on pressure plate and insert ribbon connector (1) in LP4 user interface
board (SP000276-ET) (see Figure 231, p. 7-135).
2. Mount device cover on pressure plate: Completely screw in five screws (1) using Torx
screwdriver size 10 (see Figure 230, p. 7-134).
3. Place dummy plugs over screws.
Make sure the dummy plugs line up precisely with the device cover.
Disassembly
1. Open detents of the connectors (1).
2. Loosen ribbon cable from connections.
Assembly
1. Place distance rolls on drill holes in cover.
2. Place board on display holder (SP000275-ET) and distance rolls.
3. Completely screw in four screws (2) of the display holder (SP000275-ET) using Torx
screwdriver size 10 (see Figure 233, p. 7-135).
4. Place board with display holder on device cover.
5. Completely screw in four screws (1) into board with Torx screwdriver size 8.
6. Place ribbon cable on board connections and then secure.
7.12.3 Replacing display foil (SP000280-ET)
Service flap opened (see section 7.2.1, p. 7-86).
Front module pulled out (see section 7.3.1, p. 7-86).
Device cover disassembled (see section 7.12.1, p. 7-134).
When working on conductor boards: Ensure ESD safeguard.
Disassembly
1. Disconnect ribbon cable of the display foil from LP4 user interface board.
2. Remove self-adhesive display foil from cover.
3. Clean surface of the display foil to remove adhesive remains.
Assembly
1. Feed cable (1) of the new display foil through
the cutout (2) in the cover.
Disassembly
1. Loosen spring pressure part SP000281-ET (1)
with pliers and remove it.
2. Remove four screws (2) with Torx screwdriver,
size 10.
Assembly
1. Install cable 31 SP000388-ET (9) (see section 7.13.4, p. 7-142).
2. Put the pressure plate SP000279-ET on the hinges (see Figure 240, p. 7-138).
3. Attach pressure plate to hinges with six nuts and washers (5) using a hexagon wrench,
size 8 (see Figure 237, p. 7-137).
4. Attach two screws (8) with Torx screwdriver, size 10 (see Figure 239, p. 7-137).
5. Attach blank holder SP000406-ET (7) with two screws (6) using Torx screwdriver, size 10
(see Figure 238, p. 7-137).
6. Place dummy plugs (4) over screws (see Figure 237, p. 7-137).
7. Put counter bearing SP000284-ET (3) into
appropriate opening (10).
8. Attach cover SP000285-ET (3) with four screws
(2) using Torx screwdriver, size 10 to pressure
plate (see Figure 236, p. 7-137).
Disassembly
1. Disconnect cable 23 SP000382-ET (connection
cable valve assembly) from power board LP1
(1).
2. Disconnect ribbon cable (connection cable
encoder) from controller board LP2 (2).
Assembly
During assembly there is no risk of mixing up the connectors since these are precisely
coordinated.
Assembly
1. Connect three ribbon cables (2) (see Figure 248, p. 7-141).
2. Attach power board LP1 to spacer bolts using two screws (1) and Torx screwdriver, size
25 (see Figure 247, p. 7-140).
7.13.3 Replacing controller board LP2 (SP000217-ET)
Service flap opened (see section 7.2.1, p. 7-86).
Front module removed (see section 7.3.1, p. 7-86).
Conductor board block removed (see section 7.13.1, p. 7-139)
Disassembly
1. Hold and loosen two spacer bolts (1) with
wrench, size 8.
2. Remove two spacer bolts (1).
Assembly
1. Connect three ribbon cables (2) (see Figure 250, p. 7-141).
2. Attach controller board LP2 finger-tight to spacer bolts (1) using a wrench, size 8 (see
Figure 249, p. 7-141).
After assembling:
1. Update firmware.
2. In the service program click New Pump (6) or Old Pump (7) to select the appropriate
correction mode (see Figure 89, p. 6-70) according to the roll pump in use (see
section 6.6.7, p. 6-69).
3. Perform functional tests 38 to 49 (see F-04-05-23).
Disassembly
1. Loosen connector (1) from ribbon cable on
controller board (SP000217-ET).
2. Remove connector from ribbon cable using
side cutter.
Assembly
1. Adjust cable 31 (1) such that the red
connector (2) is in the recessed area.
2. Feed cable through cover (3), guide sleeve (4)
and cover (5).
3. Screw in lower cover (3). Use short screws for
this purpose.
4. Push ribbon cable through front module and
screw in upper cover (5).
5. Lay cable 31 up to connection point (1) on the
controller board and mark length up to
connection point on cable (see Figure 251, Figure 254. Cover on pressure plate
p. 7-142).
6. Push red connector of SP000388-ET to mark
on ribbon cable and crimp using parallel pliers.
In doing so, ensure that the nose (1) is on the
red side of the ribbon cable.
7. Cut off any material projecting from the ribbon
cable.
8. Connect ribbon cable with controller board.
Injectors with S/N lower than CTM1721526 require Bluetooth module SP000225-ET
and injectors with S/N CTM1721526 or higher require Bluetooth module SP002513-
ET.
If you need to replace Bluetooth module SP000225-ET with Bluetooth module
SP002513-ET, the injector system has to be upgraded with software 04.03.00 or
newer and a black terminal (see section 3.6.3, p. 3-27).
Software 04.03.00 is also compatible with Bluetooth module SP000225-ET.
Disassembly
1. Remove cable of controller board (1).
2. Remove screws (2) and plastic nuts SP000226
(part of SP001996-ET) (3) (see Figure 257, p.
7-144) with Torx screwdriver, size 10.
Assembly
1. To attach Bluetooth module to the holder, use screws (2) (see Figure 256, p. 7-143) and
plastics nuts SP000226 (part of SP001996-ET) (3) (see Figure 257, p. 7-144) with Torx
screwdriver, size 10.
2. Plug in cable of controller board (1) (see Figure 256, p. 7-143).
7.14 Heater
After finish of the service activity, the following points are to be checked and documented
in the test protocol F-04-05-23: 16 – 18 and 37 (see section 5.13, p. 5-63).
Disassembly
1. Loosen four screws (1) on heater assembly
using Torx screwdriver size 20.
2. Remove cover feeding unit.
Assembly
1. Feed in cable through opening in cover (see Figure 258, p. 7-144).
2. Place heater assembly left and right on cover.
3. Push seal (2) upwards.
4. Completely screw in screws (1) using a Torx screwdriver, size 20.
Disassembly
1. Release connectors (1) on reed switch.
Assembly
1. Place reed switch sleeve (1) in housing (see Figure 261, p. 7-145).
2. Screw PA nut onto reed switch sleeve (see Figure 260, p. 7-145). In doing so, do not
tighten PA nut all the way.
3. Screw in reed contact (1) using socket wrench
hex 19 until the hexagon socket is flush with
the front module.
Assembly
1. Pull the O-Ring SP000094 (part of SP001996-
ET) over the media sensor.
2. Insert the media sensor with the O-Ring
SP000094 (part of SP001996-ET) into the front
module with handle SP000081-ET.
3. Attach air bubble detector to front module with
cylinder head screws SP000222 (part of
SP001996-ET), locking washers SP000574
(part of SP001996-ET) and washers SP000310
(part of SP001996-ET) using a Torx
screwdriver, size 10. Figure 265. Assembly order
8 Troubleshooting
8.1 Common problems
Problem Cause/detection Solution Section
Gap size of the roll pump Adjustment ring for gap size Clean upper part of roll pump 7.8.7
cannot be adjusted. adjustment is clogged with using warm water.
contrast medium. If necessary, replace roll pump
upper component SP000106-
ET.
Roll pump gap has The contact surface between Replace roll pump upper 7.8.7
narrowed on its own. the pump housing and the component SP000106-ET.
counter bearing is corroded and
effloresced.
Injector pauses for no Play-Pause button is defective. Replace display foil SP000280- 7.12.3
identifiable reason. ET.
Switch contact in high-pressure Change high-pressure filter 7.5.6
filter holder briefly opens during holder.
the injection and immediately
closes once again as soon as
the injection is paused.
Battery run time is too Batteries have lost capacity. Change both batteries 7.4.10 or
low. Charging LED on SP000445-ET. 7.4.11
device base turns green.
The battery run time is Battery charge is defective/ Change fuse F1. 7.4.4
too low. Charging LED on charge current is too low
device base does not turn (≤100 mA): Fuse F1 on LP5
green. charger electrical components is
defective.
Battery charge is defective/ Change charging transistor 7.4.6
charge current is too low SP000320-ET.
(≤100 mA): Changing transistor
is defective
Battery charge is defective/ Change power supply unit 7.4.8
charge current is too low SP000446-ET/ SP000617.
(≤100 mA): Power supply unit
provides too little power.
Hose threading does not Spiral conveyor traps pump Change spiral conveyor 7.9.1
function reliably. tubing between spiral conveyor SP000160-ET.
and front module.
Spiral conveyor does not deliver Change motor SP000378-ET. 7.9.2
the tubing to the bottom of the Change power board 7.13.2
front module. SP000216-ET.
Spiral conveyor does not deliver Change motor SP000378-ET. 7.9.2
the tubing fully upwards. Change power board
SP000216-ET.
In the case of information, the next steps are shown in the display.
258 E ERROR_DO Incorrect setting of the reed Readjust the reed contacts. 7.15
– OR_SENSO contacts.
267 RS 1. Check light barrier and blank sheet
Door got wedged with knob
and is opened manually. for damage.
Note: Errors 2. Completely replace cover locking 7.11
267 and 264 mechanism SP000230-ET.
indicate the Firmware version is out of Software/firmware update. 7.1
right reed date.
contact. Error
Also occurs occasionally Turn injector off and on. 4.1.4
261 indicates
directly after switching on the
the light
injector.
barrier/blank
sheet.
357 E ERROR_SAF A-D converter receiving 1. Check sensor with CT motion 5.9.2
ETY_ADC_V undefined values from service program.
ALUES_TEM temperature monitoring of 2. Replace heater CM2. 5.4
PERATURE_ heater 2. 3. Check cable. 7.13.5
CA_B Heater on the terminal is Connect heater CM2 on injector or See IFU
activated but no heater disable heater in terminal.
installed at the injector.
368- I INFORMATI Information about battery Connect injector to power supply! See IFU
378 ON_BATTER voltage
Y_VOLTAGE
384- I INFORMATI Information about the battery Connect injector to power supply. See IFU
390 ON_BATTER level
Y_LEVER
529 E ERROR_UI_ Main controller could not find 1. Check UI with service program. 5.4
NOCONNEC UI. 2. Check ribbon cable and connectors. 7.12.2
TION 3. Change UI.
4. Replace LP2 controller board. 7.12.4
530 E ERROR_SAF Safety controller could not find 1. Reinstall software. 7.1
ETY_NOCON main controller. 2. Replace LP2 controller board.
NECTION No firmware on safety, LP2
defective.
531 E ERROR_RE Disconnection with terminal. 1. Wait a few minutes. –
MOTE_NOC 2. Restart injector.
ONNECTION 3. Restart terminal.
533 E ERROR_INV Program was not correctly 1. Check Bluetooth connection. 4.1.4
ALID_SEQU transferred from the terminal: 2. Replace Bluetooth dongle.
ENCE_CHEC Poor Bluetooth connection or 3. Replace Bluetooth board injector.
K software bug.
534 E ERROR_DIF One of the communication Install compatible software versions 7.1
FERENT_PR protocols between the main, on terminal and injector.
OTOCOL_VE safety, UI and terminal is not
RSION compatible.
535 E ERROR_MIS No device serial number in the 1. Update software/firmware. 7.1
SING_DEVIC terminal. 2. Enter terminal serial number in
E_VERSION terminal.
Error resolution
1. Press power button for 5 seconds to reboot terminal.
2. Wait until application is completely started.
3. Plug in the keyboard and the USB stick that has the service key and the current terminal
software ("Rev403" folder).
4. Close application by pressing “Ctrl”, “Alt”, “Shift” and “u”.
ulrich medical boot screen is displayed.
5. Switch screens by pressing "Alt" and until linux-wpp1:~#_ appears (see Figure 267,
p. 8-178).
6. Enter mc.
Midnight Commander appears (with a blue background).
7. Use the arrow keys to navigate to the first line ( /…) (see Figure 268, p. 8-178).
8. Continue selecting the top line (/…) and confirming your selection by pressing "Enter" until
the following screen appears:
9. There must not be any files listed after the /var folder except for sdb1.
If there are still any files here named Terminal*.tar.gz or Rev*.tar.gz (see
Figure 269, p. 8-179), they can be selected with the cursor and deleted by pressing
"F8".
11. Once sdb1 is the only file left in the /var folder, navigate to the /media folder.
12. Select the /PKIusb folder.
13. Select the Rev403 folder there (see Figure 270, p. 8-179).
Error resolution
1. Plug USB keyboard into the terminal, while the USB stick for the update remains plugged
in.
Please contact the ulrich medical Technical Service Devices team by phone
+49 (0) 731 9654-111 to request the login data.
SP000376-ET Cable 18 1
CM heater
SP000384-ET Cable 25 1
Media sensors CM1, NaCl, CM2
SP000365-ET Cable 4b 1
Connection ECG grounding set
SP000378-ET Cable 21 1
Motor hose threading
SP000274-ET Display 1
SP000382-ET Cable 23 - 1
connection cable valve assembly
Power board LP1
9.15 Heater
Art. no. Designation Pc. Figure
SP000070-ET Heater left 1
9.16 Misc.
Art. no. Designation Pc. Figure
SP001205-ET Pictogram sheet CT motion 1
9.17 Tools
Art. no. Designation Pc. Figure
XD4R00A00- 1 not illustrated
Silicone spray 500 mL
001
SP001996-ET Hardware Assortment 1 not illustrated
SP000512-ET GLEITMO 810 white grease paste 1 not illustrated
XD2000E- 1 not illustrated
Technical safety check TSC kit
STK-CT
XD 8074 Alignment gauge valve assembly 1
10 Wiring diagrams
10.1 Pedestal version