Professional Documents
Culture Documents
M2P11 Anal Chem
M2P11 Anal Chem
M2P11 Anal Chem
pubs.acs.org/ac
r XXXX American Chemical Society A dx.doi.org/10.1021/ac201540t | Anal. Chem. XXXX, XXX, 000–000
Analytical Chemistry ARTICLE
Figure 2. Mass spectra of human and chicken lysozymes: (A) full scan spectrum of human and chicken lysozymes in multiple charge states, (B) full scan
spectrum of human lysozyme and PTM carrying eight charges in plasma, (C) enlarged spectrum of human lysozyme carrying eight charges, and (D)
enlarged spectrum of human dehydrated lysozyme carrying eight charges.
chicken lysozyme (IS), the most intensive isotopic ions were If the resolving power increases enough to resolve these peaks, only
m/z 1788.6280, 1788.7540, 1788.8810, 1789.0060, 1789.1300, the MS response of the resolved peaks will be used as the signal
1789.2550, 1789.3800, 1789.5060, 1789.6300, and 1789.7580; for quantitation while the MS response from background inter-
for human lysozyme (analyte), m/z 1836.9140, 1837.0400, ference between these peaks will be excluded (Figure 3D). As a
1837.1590, 1837.2890, 1837.4160, 1837.5560, 1837.6670, result, the assay sensitivity (signal-to-noise ratio) for quantitation
1837.7920, 1837.9150, and 1838.0410; for dehydrated human will be enhanced. This is especially important for most biological
lyzozyme, m/z 1834.9910, 1835.0320, 1835.1610, 1835.2850, samples, such as plasma or serum samples, which contain large
1835.4040, 1835.5350, and 1835.6580. The EIC window for amounts of interfering components. As shown in Figure 4, the
quantitation was set at (10 ppm. Peak areas were calculated using interference of background ions in plasma spreads over the entire
QualBrowser software (Thermo Fisher Scientific, Waltham, MA). mass range of the analyte protein’s isotopic peaks of interest.
The standard curve was fitted to a linear regression model with a Utilization of HR-MS in the analysis of biological samples
1/x2 weighting factor using in-house software. All statistic para- provides an opportunity for minimizing such interference.
meters were calculated by Microsoft Excel. The exact molecular In order to obtain effective improvement in selectivity and
weight of the most abundant isotope was determined by Protein sensitivity using this approach, three critical interrelated factors
Prospector (University of California, San Fransisco). need to be considered: mass spectrometer resolving power,
molecular weight of the protein analyte, and the EIC window
’ RESULTS AND DISCUSSION for quantitation. We propose here that the value of the resolving
power needs to be at least 4 times the molecular weight of the
High-Resolution Mass Spectrometry Detection. When a protein analyte, and the EIC window should be set around the
strategy of MS detection is used for protein bioanalysis, HR-MS peak width of each isotopic ion. The following is the rationale of
has unique advantages over low-resolution approaches. In a com- this consideration using the 15 kDa human lysozyme as an
monly used ESI source, proteins can be protonated into multiple example. The full-scan mass spectrum of an ionized lysozyme
charge states (Figure 2A). The ion at each charge state consists should present every two adjacent isotopic peaks one Dalton
of a cluster of isotopic peaks (Figure 2B,C), which cannot be apart, which is ∼60 ppm (1 Da/15 kDa) apart regardless of
resolved in low-resolution mass spectrometry (Figure 3A,B). the charge states. When the resolving power of 10k (full width at
C dx.doi.org/10.1021/ac201540t |Anal. Chem. XXXX, XXX, 000–000
Analytical Chemistry ARTICLE
half-maximum (FWHM): 1/10k = 100 ppm, baseline peak width together, suggesting no advantage compared to using one EIC
= 200 ppm) was employed on the low-resolving power linear ion window to cover the whole cluster of isotopic ions in the low-
trap MS (LTQ) using the zoom/ultrazoom scan function, the resolution approach. When the resolving power further increased
mass spectrometer was unable to resolve the isotopic peaks from to 4 times of the protein’s molecular weight (60k), a gap of one
the baseline (Figure 3B). When the resolving power was increased peak width in the background emerged between adjacent iso-
on an HR-MS to 30k (twice the protein’s molecular weight), topic peaks (FWHM, 1/60k ≈ 15 ppm, baseline peak width ≈
baseline separation of the isotopic peaks was just barely achieved peak gap width ≈ 30 ppm (Figure 3D)) and enabled clear
(FWHM, 1/30k ≈ 30 ppm, baseline peak width ≈ 60 ppm separation of the background from the signal of lysozyme. If the
(Figure 3C)). However, if all isotopic peak signals were used EIC window was narrowed down to one peak width accordingly,
for quantitation, EIC windows of these peaks would all be joined interference signal present in the gaps could be removed without
sacrificing the analyte’s peak signal.
Taken together, it is recommended to use the following
equation to obtain resolving power that allows adequate separa-
tion of isotopic peaks for HR-MS quantitation of an intact
protein:
R ¼ kM
Figure 4. LCMS spectra comparison of lysozyme and matrix interference in human plasma: (A and B) full scan MS of lysozyme and (C and D) full
scan MS and enlarged full scan MS of interference peaks eluted after lysozyme.
Figure 5. Effect of EIC window on assay selectivity and sensitivity: Figure 6. Effect of HPLC column temperatures (T) on the chromato-
(A) EIC window, (0.5 m/z; (B) EIC window, (30 ppm; (C) EIC graphy: (A) T = 30 °C, (B) T = 40 °C, (C) T = 50 °C, (D) T = 60 °C, and
window, (10 ppm; (D) EIC window, (1 ppm. (E) T = 70 °C.
lysozyme was narrowed down from (0.5 m/z to (10 ppm, the selectivity, and the capability to simultaneously differentiate
peak intensity ratio of the analyte over the interference increased proteins from their products of potential PTMs and BTX as well
5 times. Further reduction of the extraction window from as other homologues.
(10 ppm to (1 ppm caused reductions in the peak intensity Assay Development of Sample Preparation and Chroma-
of both analyte and interference and thus did not significantly tographic Method. An LC method with good chromatographic
improve the analyte/interference ratio (Figure 5C,D). Com- peak shape is often difficult to obtain for protein samples;
bined with consideration of minimizing the matrix interference in however, it is essential for achieving quality bioanalysis results.
plasma samples, the EIC window was set to a little less than the Sample preparation is another important part of assay develop-
baseline peak width (at (10 ppm). ment. Although the objective of this research was not to emphasize
The example of lysozyme quantitation demonstrates improve- the development of sample preparation and chromatographic
ment of the signal-to-noise ratio using the isotopic cluster peaks conditions for the model analyte, assay development was con-
from one charge state (z = 8). Full scan mass spectrometry ducted to have a workable assay to demonstrate the concept of
detection enables the capability in summing signals of isotopic HR LCMS bioanalysis of proteins. For the sample preparation,
cluster ions from multiple charged states. The ion selection for protein precipitation and SPE were tested. The results suggested
quantitation depends on the distribution of signal intensity in that SPE provided better sample cleanup (results not shown).
different charged states. In this case, the protonated lysozyme Different combinations of organic and aqueous phases were used
was concentrated at the z = 8 charge state in the optimized to obtain the optimal conditions for wash and elution steps of
ionization conditions. The isotopic peaks from the other charged SPE sample preparation. The result indicated that using 4080%
states were <30% of the intensity of the highest z = 8 peak and of ACN as elution solvent had the highest recovery of the analyte,
thus excluded for quantitation. For analyte proteins evenly ionized while there was minimal elution of the analyte when less than 20%
into different charge states, all ions in the charge states with ACN was used. Therefore, 15% of ACN in washing solution and 70%
significant signal intensity should be included in the combined of ACN in elution solvent were used to remove polar interference
EIC to achieve the best sensitivity. components and collect cleaned up samples, respectively.
For quantitation of proteins with higher molecular weights, a For the HPLC chromatography development, one chromato-
mass spectrometer with higher resolving power is required. The graphic factor we tested was column temperature. As shown in
higher the resolving power of a mass spectrometer, the better its Figure 6A, room temperature chromatography exhibited broad
selectivity; and therefore, better sensitivity as a narrower EIC peak width and poor peak shape. Even at elevated column
window is utilized to remove the background interference. For temperature up to 50 °C (Figure 6C), the peak shape was still
example, to fully utilize this approach to quantify an intact undesirable, with multiple overlapping wide peaks. When the
antibody with 150 kDa molecular weight, it requires a 600 000 column temperature increased to 70 °C (Figure 6D,E), the peak
practical resolving power based on the above model equation. It narrowed down to ∼20 s wide and became symmetric. This
is still a big challenge to achieve such high resolving power with phenomenon suggested full equilibrium of the different isoforms
high scan speed. When ultra high-resolution mass spectrometers of the lysozyme at this temperature. In addition, both isocratic
become available and are compatible with HPLC chromato- and gradient chromatography were tested. The results showed
graphy, they will provide a powerful tool to quantitate a range that isocratic chromatography caused significant peak tailing for
of intact proteins in biological matrixes with good sensitivity, lysozyme, while a slow gradient chromatography could achieve
E dx.doi.org/10.1021/ac201540t |Anal. Chem. XXXX, XXX, 000–000
Analytical Chemistry ARTICLE
1.92 ( 0.60
6.0 ( 1.3
average
human plasma human plasma human plasma human plasma human plasma human plasma human plasma human plasma
lot 8
1.1
7.2
lot 7
1.4
6.7
lot 6
3.0
5.1
lot 5
1.9
7.4
Figure 7. EIC of human plasma sample no. 1 from Table 1 with (10 ppm
lot 4
1.9
7.3
window: (A) EIC of IS (chicken lysozyme); (B) EIC of the analyte
(human lysozyme), inserted panel, EIC of the analyte with the EIC
window of (0.5 m/z; and (C) EIC of the dehydrated lysozyme.
lot 3
2.0
4.2
good peak shape and acceptable separation from the major
interference peaks. Thus, by combination of high-temperature
chromatography and mild gradient conditions, satisfactory LC
conditions for the quantitation were achieved (Figure 7B).
Assay Evaluation. Because of the existence of endogenous
lot 2
2.2
human lysozyme in human plasma, the assay evaluation was 4.7
performed using standards and QCs prepared by spiking human
Table 1. Quantitation of Native Lysozyme and Dehydrated Lysozyme
2.7
5.3
results were mainly from the low QC level. Excluding the low QC
level, the assay performance was from 4% to 28% for the accuracy
and from 7% to 13% for % CV.
Endogenous lysozyme concentrations in eight individual lots
ratio of dehydrated lysozyme
lysozyme concn (μg/mL)
to lysozyme (%)
concentration obtained was 1.92 ( 0.60 μg/mL, while the that it had the same charge state and isotopic distribution as the
literature value obtained using a radioimmunoassay was 4.6 ( human lysozyme (Figure 2C,D). These observations suggested
0.8 μg/mL.16 The difference could be explained by the fact that that this component was related to the human lysozyme. The
the radioimmunoassay might include the responses of nonspe- mass shift of the most abundant isotopic isomer was calculated as
cific binding or of other protein homologues with different 17.9648 Da ((1837.5404 1835.2948) 8) less than the native
sequence contexts (cross-reactivity). In order to ensure monkey human lysozyme, indicating loss of a molecule of H2O (dehydration,
plasma as a proper surrogate for human plasma for the prepara- loss of 18.0105 Da). This dehydrated product was also present in
tion of standard curves and QC samples, a human plasma sample the purified human lysozyme used for standard curve and QC
was diluted 4-fold by monkey plasma and measured again for preparation. The EIC peak area ratio of the dehydration product
lysozyme concentration. Calculated concentrations from the versus the native lysozyme in human plasma was 6.0% ( 1.3%,
diluted and undiluted samples differed by 11% (2.44 μg/mL vs close to the ratio for the low QC samples (5.2 ( 1.1%). The
2.73 μg/mL, respectively). Although further improvement of dehydration could result from PTM on serine, threonine, or
assay performance may be needed for drug development applica- tyrosine residues. Gas phase fragmentation in the ion source or
tions, the accuracy and precision obtained here can be regarded the ion trap may also cause dehydration of a protein or a peptide,
as sufficient for biomarker discovery and pharmacokinetic anal- although it most commonly happens during MS/MS analysis
ysis of therapeutic proteins in the drug discovery stage. when the collision energy is applied.18
Investigation of the Protein PTMs and Its Implications. Taken together, along with the quantitation of the target intact
The proposed approach has significant advantages in obtaining protein, investigation of the extra information that could be lost
information on PTMs as well. One advantage is in detection of the in the “bottom-up” approach could enable a new way to identify
formation of disulfide bridges within the analytes. Compared with biomarkers related to disease, populations, achievement of
the common “bottom-up” approach by breaking disulfide bonds efficacious dose, or other factors that could be critical for clinical
before protein digestion, the quantitation of intact proteins can study design and data interpretation.
maintain their disulfide bridges during analysis. In this study of
human lysozyme, no reducing agent was used in the sample pre-
paration. The spectrum of octuply charged human lysozyme ’ CONCLUSIONS
showed m/z 1837.5404 as the most abundant isotopic ion We report here the first attempt for protein bioanalysis using
(Figure 2B,C). The molecular weight of the most abundant iso- HR LCMS detection of the intact protein. The major advan-
tope was, therefore, calculated as 14 692.2648 Da ([1837.5404 tages of this strategy are specific detection of the intact protein,
1.0073] 8), which is 7.9086 Da less than the theoretical independent of fragmentation efficiency, fully utilizing a high
molecular weight of the unfolded lysozyme (14 700.1734 Da). resolving power to achieve good selectivity, and quantifying
This observation demonstrated that the eight cysteine residues in proteins with PTM characterization. In this study, only one
the natural human lysozyme lost eight hydrogens and formed charge state of the ionized protein was used for the example. As
four disulfide bonds during post-translational folding. As forma- the molecular weight of the protein increases and multiply
tion of a disulfide bond (2.0156 Da) only causes a marginal charged ions are more evenly distributed in intensities, feasibility
change to the large molecular weight of a protein, high resolution of utilizing the most abundant ion peaks by HR-MS detection will
and high accuracy are crucial to identify the most abundant result in further improvement of the sensitivity of the assay. Along
isotope and to assign its exact mass for determination of the with more advanced instrument development for higher resolu-
accurate number of disulfide bridges. In this regard, the HR-MS tion, faster scan speed, higher charging efficiency, and the ad-
quantitation method is also capable of detecting changes of the vancement of computer algorithms for signal summation, EIC
number disulfide bridges caused by mutations, e.g., one less window optimization, and calibration of standard regression, this
disulfide bond formed in a mutant human lysozyme with the analytical approach could provide further enhanced selectivity
Cys65 replaced by an Ala.17 and sensitivity and therefore become a more valuable approach
Another advantage of this approach is the capability of with wider applications in protein bioanalysis for drug
detecting and semiquantifying the other components chemically development.
derived from the analytes by PTM, e.g., chemical modifications of
the backbone amino acids and potential BTX products. When a
protein containing hundreds of amino acids is denatured during ’ ASSOCIATED CONTENT
sample processing, PTMs on one or two residues normally cause
minimal changes to the overall properties of the whole molecule bS Supporting Information. Standard curves and quality
in LCMS. It would be expected that any of these components control samples. This material is available free of charge via the
extracted during sample preparation would elute at a very similar Internet at http://pubs.acs.org.
retention time as the analyte. HR LCMS facilitates the identi-
fication of PTM products without prior knowledge by comparing ’ AUTHOR INFORMATION
multiple characteristics, such as chromatographic peak shape, Corresponding Author
distribution of isotopic isomers, and charge states with those of *E-mail: Qin.Ji@BMS.com.
the analyte protein. This approach can be readily used for
semiquantitative analysis of specific PTMs. In the study of human
lysozyme, a component with similar LCMS properties of
lysozyme was detected in human plasma samples but not in ’ ACKNOWLEDGMENT
the blank monkey plasma (Figure 7C). This component exhib- The authors would like to thank Mr. Huidong Gu for writing
ited the same chromatographic retention time and similar peak an Excel Macro for the regression model and weighting selection
shape to the human lysozyme. The HR-MS spectrum indicated and Heidi Snapp for the assistance in manuscript preparation.
G dx.doi.org/10.1021/ac201540t |Anal. Chem. XXXX, XXX, 000–000
Analytical Chemistry ARTICLE
’ REFERENCES
(1) Ezan, E.; Bitsch, F. Bioanalysis 2009, 1 (8), 1375–1388.
(2) Ji, Q. C.; Rodila, R.; Gage, E. M.; El-Shourbagy, T. A. Am. Pharm.
Rev. 2004, 7, 84–89.
(3) Carr, S. A.; Anderson, L. Clin. Chem. 2008, 54, 1749–1752.
(4) Heudi, O.; Barteau, S.; Zimmer, D.; Schmidt, J.; Bill, K.;
Lehmann, N.; Bauer, C.; Kretz, O. Anal. Chem. 2008, 80, 4200–4207.
(5) Ji, C.; Sadagopan, N.; Zhang, Y.; Lepsy, C. Anal. Chem. 2009,
81, 9321–9328.
(6) Wu, S. T.; Ouyang, Z.; Olah, T. V.; Jemal, M. Rapid Commun.
Mass Spectrom. 2011, 30 (25), 281–290.
(7) Xu, Y.; Mehl, J. T.; Bakhtiar, R.; Woolf, E. J. Anal. Chem. 2010,
82, 6877–6886.
(8) Ji, Q. C.; Rodila, R.; Gage, E. M.; El-Shourbagy, T. A. Anal. Chem.
2003, 75, 7008–7014.
(9) Ji, Q. C.; Rodila, R.; Morgan, S. J.; Humerickhouse, R. A.;
El-Shourbagy, T. A. Anal. Chem. 2005, 77, 5529–5533.
(10) Ji, Q. C.; Gage, E. M.; Rodila, R.; El-Shourbagy, T. A. Rapid
Commun. Mass Spectrom. 2003, 17, 794–799.
(11) Kellie, J. F.; Tran, J. C.; Lee, J. E.; Ahlf, D. R.; Thomas, H. M.;
Ntai, I.; Catherman, A. D.; Durbin, K. R.; Zamdborg, L.; Vellaichamy, A.;
Thomas, P. M.; Kelleher, N. L. Mol. Biosyst. 2010, 6, 1532–9.
(12) Arnaud, C. H. Chem. Eng. News 2010, 88 (25), 10–15.
(13) Yoshimura, K.; Toibana, A.; Nakahama, K. Biochem. Biophys.
Res. Commun. 1988, 150, 794–801.
(14) Peters, C. W.; Kruse, U.; Pollwein, R.; Grzeschik, K. H.; Sippel,
A. E. Eur. J. Biochem. 1989, 182, 507–516.
(15) Taniyama, Y.; Yamamoto, Y.; Nakao, M.; Kikuchi, M.; Ikehara,
M. Res. Commun. 1988, 152, 962–967.
(16) Peeters, T. L.; Depraetere, Y. R.; Vantrappen, G. R. Clin. Chem.
1978, 24, 2155–2157.
(17) Taniyama, Y.; Yamamoto, Y.; Kuroki, R.; Kikuchi, M. J. Biol.
Chem. 1990, 265, 7570–7575.
(18) Wysocki, V. H.; Resing, K. A.; Zhang, Q.; Cheng, G. Methods
2005, 35, 211–222.