Professional Documents
Culture Documents
ICU Ventilator - Group 4
ICU Ventilator - Group 4
ICU Ventilator - Group 4
2
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
The COVID-19 pandemic has brought attention to ventilators, but few people are familiar with
what they do or how they work. When a patient is unable to breathe adequately on their own, a
ventilator pumps air—usually with extra oxygen—into their airways. Patients who have had their
lung function severely impaired, such as by an injury or an illness like COVID-19, may require a
ventilator. It's also used to help people breathe while they're having surgery.
CHAPTER 1: INTRODUCTION
3
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
Invasive ventilation
Tracheal intubation is invasive ventilation that involves physically inserting a
tube into the patient's throat to take over respiratory function. This obstructs the
airway, so patients are sedated for this procedure.
The most common reason for admission to the ICU is invasive mechanical
ventilation, which requires access to the trachea, most commonly via an endotracheal
tube. The following are the most common indications for invasive ventilation,
according to large multinational surveys:
+ Coma 16%
+ COPD 13%
+ ARDS 11%
+ Heart failure 11%
+ Pneumonia 11%
+ Sepsis 11%
+ Trauma 11%
+ Postoperative Complications 11%
+ Neuromuscular Disorders 5%.
Many researchers are still unsure about COVID-19, but we do know that those
infected with the novel coronavirus experience symptoms such as fever, cough, and
sore throat, among other things.
If the body's immune system fails to fight the infection, it can spread to the lungs
and cause acute respiratory distress syndrome, which can be fatal (ARDS). The
alveoli (tiny air sacs that allow oxygen to reach the bloodstream and remove carbon
dioxide) fill with fluid in ARDS, reducing the ability of the lungs to provide enough
oxygen to vital organs.
"ARDS causes severe pulmonary inflammation, but the main issue is that it
renders portions of the lungs useless," Dr. Ferrante explains. "It can be life-
threatening, and many of these patients will require a ventilator."
1.2. History and Revolution
The history of mechanical ventilators, positive and negative pressure, invasive
and noninvasive ventilation will all be explored in this discussion. The goal will be to
figure out what developments in clinical medicine caused the mechanical ventilator to
be refined. The capabilities of today's ICU ventilators, as well as the evolution of
positive-pressure ICU ventilators, will be discussed. All of this will serve as a
backdrop for speculation about the future ICU ventilator.
Negative-pressure ventilators
During the nineteenth century and the first half of the twentieth
century.
The negative-pressure ventilator was the most common device
for providing ventilatory assistance.
A full-body type ventilator was the first description of a
negative-pressure ventilator.
In 1838
John Dalziel, a Scottish physician, was the first to describe this "tank ventilator" . It was
made up of an airtight box in which the patient was kept in a sitting position. Manually
4
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
pumping air into and out of the box created negative pressure. Similar types of manually
operated negative-pressure ventilators were developed by a number of other groups.
In 1904
Sauerbrach also invented a negative-pressure operating
chamber.
With the development of the iron lung, originally designed and
built by Drinker and Shaw but manufactured and sold by Emerson,
negative-pressure ventilation became a much more clinical reality.
During the worldwide poliomyelitis epidemics from 1930 to
1960, this approach to ventilatory support reached its pinnacle.
The first intensive care units (ICUs) were established to care
for dozens of patients of all ages who required negative-pressure ventilation due to
poliomyelitis. The Boston Children's Hospital created a large negative-pressure chamber that
could hold four children at once and allow a nurse to care for them from within the chamber.
Other negative-pressure chambers, such as the "raincoat" and the "chest cuirass,"
were developed and used with varying degrees of success over time.
In the 1960s
However, due to a number of factors, there was a move away from negative-
pressure ventilation. The first volume-controlled ICU/anesthesia ventilators appeared.
At the end of WWII
The development of jet aviation led to the development
of small, portable intermittent positive-pressure breathing
(IPPB) devices, such as the Bennett and Bird IPPB machines.
Complications with the use of negative-pressure ventilation
made it impossible to continue using it in newly developed
ICUs. These devices were large, heavy, and cumbersome as a
group, and it was difficult to avoid excessive leaking (which
usually resulted in the patient's body cooling); they had
trouble maintaining effective ventilation, were unable to
maintain high airway pressure or establish PEEP, and access
to the patient was restricted.
5
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
6
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
7
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
Gas entered this system from the left and preferredentially entered a bellows
from which the patient could spontaneously breathe.
Gas was directed to a second bellows if the first bellows filled completely.
The gas from the bellows was delivered to the patient as a positive-pressure breath
once it was filled.
All breaths were either spontaneous or mandatory (if the patient became apneic)
or a mix of both depending on the flow of gas into the system, the setting of the
bellows capacity, and the patient's spontaneous minute volume.
The primary problem with this initial system was that the patient could breathe
the entire minute volume with a very rapid and shallow breathing pattern, but it did
provide the first form of closed-loop control.
8
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
9
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
10
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
2.2. Components
2.2.1. Source of power
Gas supply
- 1 bar = 100 kPa = 14.5 psi.
- 100% Oxygen or medical air ( 21% Oxygen and 79% Nitrogen).
- Either Wall or Bottled Gas.
- Air can be from a compressor.
- Colour Coded.
a. Wall Gas
- Wall gas = pressure of 350 - 400 kPa.
- Colour Coded.
- Indexed.
b. Bottled Gas
- Gas bottles 15MPa but has a
regulator that reduces it to
350 - 400kPa.
- Pin indexed.
Power Supplies
- Main power is regulated and reduced into set voltages.
+ ±5 Volts
+ ±12 Volts
+ ±24 Volts
- Loss of power Alarm for most ventilators.
- Battery power that is charging when on mains power and in use when power
failure.
Pressure Regulators
- From 350-400 kPa to 110-170 kPa (depending on the ventilator).
- Need to be calibrated if they are prone to drifting.
- Ventilator controls the trans-respiratory system pressure
- This trans-respiratory system gradient determines the depth or volume of respiration.
- Based on this a ventilator can be positive or negative pressure ventilator.
2.2.2. Control of gas delivery
Gas Blender
A gas blender is required to control the mixture of air, oxygen, anaesthetic gas or
whatever else you might be using the ventilate your patient. One may not need any
such gas blender if one is discussing some sort of stripped-down domiciliary model
which runs on room air alone, and which does not accept an exogenous oxygen source.
Gas Accumulator
A gas accumulator might be a component of a ventilator which requires a
precise control of gas mixtures and which cannot rely on proportioning valves to
produce this level of precision, eg. where the gas flows are very low. An example
of this is the accordionlike “bellows” of an anaesthetic machine; it is used to
maintain a reservoir of a stable gas mixture.
11
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
12
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
13
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
- Flow is obstructed. When air passed through tubing opens the flap.
- Changed the pressure on the strain gage.
- Increase in flow is a reduction in resistance.
- Within the patient circuit so must be cleaned after each use.
- Can become brittle and break.
- Can be changed while the vent is in use ( but alarm will be triggered).
- Solenoids control the flow not the sensors so only an alarm will sound.
Gas concentration
Voltaic cells or spectrophotometers are commonly used to monitor gas
concentration. The oxygen supply sensor, for example, is typically an oxygen cell that
generates an output voltage proportional to the partial pressure of oxygen in the
inspiratory gas pipe.
Pressure
Historically, pressure in the circuit was measured using aneroid manometers,
which are pressure sensors that detect air pressure by the action of the air on the
elastic lid of an evacuated box. Integrated silicon wafer pressure transducers have
replaced these in current ventilators, at a tenth of the cost and with far greater
accuracy.
Volume
In current ventilators, volume is computed using flow data rather than being
measured directly. The major variable over which the intensivist had any control in
previous ventilator designs (e.g., the bellows and piston types) was a directly
measured volume.
Gas Mixing and Flow Control
- Solenoids control flow and mix of gases.
- Proportional: Changing current = changing flow.
- Solenoids work in unison to control flow.
- Control:
+ Tidal volumes.
+ Patient flow rates.
+ Respiratory rates.
+ I/E ratios and oxygen flow.
- Gases are either mixed in tubing or a separate chamber.
- Most ventilators have a separate chamber for the gas to be
inspired.
+ Used for mixing gas as well as reserve chamber for gas as well as reserve
chamber for gas to be inspired.
2.2.4. Safety Features
Filters
- Gas Supplies not always clean.
- Filters to protect the equipment as well as the patient.
- Water can also be present and so a water trap is also useful.
Water traps must be emptied.
- Filters and water traps on the inlet.
- Filters in the patient circuit.
- Bacterial filters to prevent infection to patients.
14
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
The safety features are some devices and measures which ensure that the
patient does not come to any additional harm from being ventilated (beyond the
already brutal effects which are integral to the process). These consist of filters and
alarms.
Inspiratory filters of the ventilator promote purity of inspired gas (eg. by
removing airborne particles and bacteria from the inspired gas mixture).
Expiratory filters keep the ICU workers safe. Expired gas is filtered to
keep the ventilator from constantly spitting out huge clouds of aerosolized
germs created in the patient's airways, which smell like a sewer. Expiratory
filters are also commonly used to protect ventilator components from the hot
and humid gases that must be exhaled, which would damage sensor
measurements and shorten the device's lifespan.
Alarms are frequently built into the program to prevent unintended
modifications to the ventilator settings and strange ventilation
misapplications. In general, they are devices that alert you when the patient's
state or ventilator performance has exceeded the parameters that you have
determined to be safe. Non-software alarm-like features, such as mechanical
blow-off valves to relieve excess pressure when the patient coughs, are also
included into ventilators.
2.3. Ventilator Settings
2.3.1. Input
CPU
15
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
16
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
Baseline variable
- The variable controlled during expiration phase.
- Mostly its pressure.
Limit variable
- A variable(pressure,volume,flow) is 'limited' if it increases to to a preset before
inspiration cruds.
- Inspiration is NOT terminated when a variable has reached its limit.
Pressure controller
- Ventilator controls the trans-respiratory system pressure.
- This trans-respiratory system gradient determines the depth or volume of respiration.
- Based on this a ventilator can be positive or negative pressure ventilator.
Volume controller
• Volume cycled ventilation delivers a:
- Set volume;
- With a variable Pressure - determined by resistance, compliance and inspiratory
effort.
Time controller
- Measures and controls inspiratory and expiratory time.
- These ventilators are used in newborns and infants
Modes of Ventilation
MODE is a preprogramed settings that tells the ventilator how to oxygenate the
patient.
• Time cycled ventilators
• Volume cycled ventilators
• Pressure cycled ventilators
• High Frequency Ventilators
Time cycled ventilator
- Time delivery of gas flow; tidal volume = flow rate x inspiration time.
- Delivers relatively constant Tidal Volume.
- Allows precise control of gas delivery.
- Types: IMV-Bird, Foregger 210, Emerson.
Volume Cycled Ventilator
- Inspiratory gas flow terminated after preselected volume delivery.
- Pressure in circuit determined by tidal volume and Compliance.
- Delivered Tidal volume changes with changes in Compliance.
- Types: Bennett, Ohio 560, Bourns Bear 1, Monaghan, Siemons Servo.
Pressure cycled ventilators
- Gas flow continues until preset pressure develops.
- Tidal volume = flow rate x time until pressure is reached.
- Variable volume if circuit pressure varies change in compliance).
- Types: bird mark, bennett PR.
17
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
18
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
Flow Waveforms
• Inspiratory flow is controlled by setting the peak flow and flow waveform.
• The peak flow rate is the maximum amount of flow delivered to the patient during
inspiration, whereas the flow waveform determines the how quickly gas will
be delivered to the patient throughout various stages of the inspiratory cycle.
• There are four different types of flow waveforms available.
These include the square,
decelerating (ramp),
accelerating
sine/sinusoidal waveform
Pressure waveforms
- Rectangular
- Exponential rise
- Sine
• Can be used to monitor
+ Air trapping (auto-PEEP)
+ Airway Obstruction
+ Bronchodilator Response
+ Respiratory Mechanics (C/Raw)
+ Active Exhalation
+ Breath Type (Pressure vs Volume)
+ PIP, Pplat
+ CPAP. PEEP
+ Asynchrony
+ Triggering Effort
2.3.4. Alarms
Inspiratory time
- Inspiratory time is a combination of the inspiratory flow period and time taken for
inspiratory pause. The following diagram depicts how the addition of an inspiratory
pause extends total inspiratory time.
Expired Air
- Once the air has been expired by the patient:
+ No longer heated.
+ Water trap to remove moisture.
+ Enters expiratory block heated to stop rain out.
+ Filtered to stop infection from getting into the unit as well as contaminating the
environment.
+ Either enters room air or is scavenged.
2.3.5. Exhalation System
- Controls:
+ Timing of breaths.
+ Volume of the air delivered to the patient.
- During inhalation the exhalation valve closes to stop the flow of gas out of the.
system and into the patient this inflates the patient's lungs.
19
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
- During exhalation the valve is opened to allow gas from the patient out through the
expired limb.
- Usually removable to allow for cleaning.
- Usually heated to stop condensation.
2.3.6. PEEP System
- Positive End Expired Pressure.
- Keeps lungs slightly inflated.
- Reduces damage to lungs.
- Maintained by closing the expiratory valve.
AutoPEEP
During expiration alveolar pressure is greater than circuit pressure until
expiratory flow ceases. If expiratory flow does not cease prior to the initiation of the
next breath gas trapping may occur. Gas trapping increases the pressure in the alveoli
at the end of expiration and has been termed: dynamic hyperinflation; autoPEEP;
inadvertent PEEP; intrinsic PEEP; and occult PEEP.
Dangers of PEEP
- High intrathoracic pressures can cause decreased venous return and decreased
cardiac output.
- May produce pulmonary barotrauma.
- May worsen air-trapping in obstructive pulmonary disease.
- Increases intracranial pressure.
- Alterations of renal functions and water metabolism.
Physiology of PEEP
- Reinflates collapsed alveoli and maintains alveolarinflation during exhalation
PEEP
↓
Decreases alveolar distending pressure
↓
Increases FRC by a veolar recruitment
↓
Improves ventilation
↓
Increases V/Q, improves oxygenation, decreases work of breathing
2.3.7. Clinical Terms
- PEEP = Positive End Expiratory Pressure: Pressure remaining in the circuit at the
end of expiration.
- PIP = Peak Inspiratory Pressure: Maximum pressure of inspiration.
- FiO2 = Fractional Inspired Oxygen: Displayed as a percentage.
- TE = Expiratory Time.
- I/E Ratio = ratio of time for inspiration and expiration.
- MAP = Mean Airway Pressure: The average pressure that the circuit has.
20
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
21
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
22
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
23
INTERNATIONAL UNIVERSITY
BIOMEDICAL ENGINEERING DEPARTMENT
CHAPTER 4: DEVELOPMENT
24
other hand, is a crisis. This could signify the development of a tension pneumothorax,
which necessitates rapid action. Clearly, the alert conditions in each of these three
circumstances should differ significantly. The ventilator of the future will be able to
comprehend these patterns, resulting in distinct alert circumstances.
The ventilators of the future will not display endless amounts of irrelevant data that
are useless to the practitioner. Only a limited number of separate pieces of data may be
processed by a single person. Information, not just lines of data, will be presented by the
next generation of ventilators. The clinician will be able to quickly determine if the
patient's status has altered using simple graphs or figures. At least one ventilator already
uses a figure to reflect a change in the patient's state. Important connected variables will
be presented so that the doctor can identify whether or not change has occurred quickly.
Tidal volume and airway pressure, for example, will be displayed in such a way that the
patterns in these variables may be easily discerned. In addition, information that has not
previously been available will be made available.
The number of breaths with missed triggers, double-triggers, or an unusually short
or long inspiratory or expiratory time will be calculated and shown, as well as the
asynchrony index. The presence of conditions linked to the emergence of auto-PEEP
will be recognized and shown.
The most significant function of this new generation of ventilators will be to assist
in decision-making. Following each alarm condition, a list of possible causes and
solutions will be provided. Changes in ventilator variables will be identified, and the
clinician will be advised of the change, as well as the possible causes and solutions.
From the ventilator's operation manual to the evidence that supports a recommended
action, the ventilator screen will provide access to a library of information.
On all ventilation modes, closed-loop ventilation control will be provided. These
new ventilators will be able to modify gas flow to increase synchronization between the
patient and the ventilator. They'll be able to understand the airway pressure and flow
waveform during both volume and pressure ventilation, and alter the flow waveform,
peak inspiratory flow, rising time, and termination criteria automatically to guarantee
gas supply is in sync with the patient's wishes. This is becoming a more relevant
element in ventilator operation as we learn that asynchrony can have a significant
impact on patient outcomes. On at least one ventilator, automatic termination criteria
change is already available.
All of these anticipated changes suggest that future users of mechanical ventilators
will need to be much more prepared than current users. They'll need to know everything
there is to know about the new features' operational difficulties. They'll have to be able
to tell when one characteristic is more important than the other. They'll have to ensure
that the ventilator is doing what it's supposed to be doing and that the patient is reacting
to the intervention as predicted. Clinicians in charge of these patient-ventilator systems
in the future will need to be far more capable than those in charge now.
25
REFERENCES
6. Nieminen MS, Bohm M, Cowie MR, et al. Executive summary of the guidelines on
the diagnosis and treatment of acute heart failure: the task force on acute heart failure
of the European Society of Cardiology. Eur Heart J 2005; 26: 384e416.
7. Murase K, Tomii K, Chin K, et al. The use of non-invasive ventilation for life
threatening asthma attacks: changes in the need for intubation. Respirology 2010; 15:
714e20.
9. Kerby GR, Mayer LS, Pingleton SK. Nocturnal positive pressure ventilation via
nasal mask. Am Rev Respir Dis 1987;135(3):738–740.
10. McPherson SP, Spearman CB. Respiratory therapy equipment, 2nd edition. St.
Louis: Mosby; 1981:333–492
11. https://www.slideserve.com/brody/ventilator
12. https://www.slideshare.net/AnanyaNanda/principles-of-icu-ventilators
13. https://www.linkedin.com/pulse/icu-design-future-planning-designing-tarun-
katiyar/?fbclid=IwAR0VTrlmXgPNGrQvMCG4IxojxjNIVVgbqJU5wkEIcT1aJhvzm
d-_QKZ4rKk
14. https://resident360.nejm.org/content_items/277
15. https://www.slideserve.com/toan/patients-safety-in-intensive-care-unit
26