Professional Documents
Culture Documents
An Xray Gas Position Sensitive Detector: Construction and Characterization
An Xray Gas Position Sensitive Detector: Construction and Characterization
An Xray Gas Position Sensitive Detector: Construction and Characterization
This article is copyrighted as indicated in the article. Reuse of AIP content is subject to the terms at: http://scitationnew.aip.org/termsconditions. Downloaded to IP:
216.165.95.79 On: Wed, 10 Dec 2014 17:28:31
An x .. ray gas position sensitive detector: Construction and characterization
A. F. Barbosa
Centro Brasileiro de Pesquisas F/sicas, RUG Xavier Sigaud, 150, 22290 Rio de Janeiro, Brazil
A. Gabriel
EMBL, B. P.156X F-3804J Grenoble, France
A Craievich
Laboratorio Nacional de Luz Sfncrotron, C. P. 6192, 13081 Campinas, SP and Instituto de Ffsica, USP,
Silo Paulo, Brazil
(Presented on 31 August 1988)
A linear x-ray gas position sensitive detector with delay line readout has been constructed. The
detector is described, characterized, and used for detecting x-ray diffraction patterns from
polycrystals.
I. FEATURES
n photon
beryllium
The linear position sensitive detectors consist of a gas window
chamber, a metallic anode wire, and a cathode connected to
a delay line. As in proportional detectors, the incoming pho-
tons produce electrical charges in the gas, which induce a cathode
pulse on the cathode. The measured values of the traveling
times of the induced charges from the point of collection to
the ends of the delay line, are used to determine the position
of the photons. 3 FIG. I. Detection cell (schematic).
2315 Rev. Sci. !nstrum. 60 (7), July 1989 0034-6748/89/012315-03$01.30 @ 1989 Amer!can Institute of Physics 2315
This article is copyrighted as indicated in the article. Reuse of AIP content is subject to the terms at: http://scitationnew.aip.org/termsconditions. Downloaded to IP:
216.165.95.79 On: Wed,
•••• -. , •••• -.-••••••••••• 10
,' •••>;•••••••••••••••••• Dec 2014
:.:.:.:.:.:.:.;.;.:.:.:.; 17:28:31•••; •••••••: •.•••••••: ••
••••••: ••••••••••::.:.:.:.:.:.:.:.:-:.:.:.:.:.;.:
>;>: •••••••• -.~••••••••- ••••••• ,".-•••-, •••• ~,.,. ,._ ••••
FIG. 2. Assembling of the detector, (a) detection cell, (b) beryllium win- "H'" 2175 2250 "".0 2",,00
dow. VOLTAGE (V)
II. CHARACTERIZATION
The characterization of the detector has been carried
out by using a standard x-ray generator producing filtered
CuKa radiation, which mainly contains photons of approxi-
eoo
f
mately 8 keY. 520
f
The counting rate of the detector as a function of the
high voltage between anode and cathode, is represented in "0
Fig. 4. The counting rate exhibits a "plateau" above
V = 2250 V. Consequently, the operating high voltage of the
detector was chosen to be 2300 V.
The resolution parameter R is defined as the full width
I
at half maximum intensity (FWHM) of the counting distri- _L_j
bution for an incoming beam of negligible cross section. This S .•
parameter has been determined by using a slit of 17 11m BEAM POSITION X (em)
between the x-ray source and the detector. The measured
FIG. 5. Linearity test (channel number as a function ofthe beam position).
values of R vary slightly as a function of the position of the
beam around the average value of R = 430.um (FWHM).
With the total length of the detector window L equal to 80
mm, the relative resolution becomes R /L = 0.5%.
The linearity of the detector has been evaluated by mea-
suring the intensity profiles recorded for a narrow incoming
beam (width = 90.urn) for different beam-incidence posi-
tions. The center of gravity of the distributions in the multi-
channel analyzer as a function of the beam position is plotted
in Fig. 5. This figure shows a qualitatively good linearity; the
c d e
FIG. 3. Electronic system for data readout, (a) detector, (b) preamplifiers,
(c) amplifier-discriminators, (d) time-delay generator. (e) time-to-ampli- DiFFRACTION ANGl E (degrees)
2316 Rev. SCi.lnstrum., Vol. 60, No.7, July 1989 Data acquisition and image detectors 2316
This article is copyrighted as indicated in the article. Reuse of AIP content is subject to the terms at: http://scitationnew.aip.org/termsconditions. Downloaded to IP:
216.165.95.79 On: Wed, 10 Dec 2014 17:28:31
III. APPL.ICATIONS
~
s.o
g (oj
For testing purposes the linear detector has been used to
0::
III
III
::E 4.5
record the x-ray diffraction patterns of polycrystalIine sili-
:>
z con and C 23 H.,.s paraffin sampleso For comparison, a similar
I-
Z experiment has been carried out for the same paraffin sample
.
:> 3.0
0
u ." using a scintillation counter and the standard step scanning
" procedure.
1.5
:
.. The diffraction diagram for the silicon sample is repre-
0.0 ~ "OI\'P~_ ........ sented in Fig. 7 for diffraction angles ranging from 84° up to
2.0 4.0 s.o 6.0 98°. The CuKa doublet for the 333 and 224 Bragg reflections
DIFFRACTION ANGLE (dogr•• s) is partially resolved in the obtained experimental profile.
The small angle diffraction pattern corresponding to the
paraffin sample is plotted in Fig. 8(a). This diagram shows
the 002, 004, and 006 Bragg reflections from the crystalline
modification of the studied paraffin. The same portion of the
pattern has been recorded using the step scanning procedure
with a resolution slit of 90 pm. The width at half maximum
intensity of the 002 Bragg peak profiles in Fig. 8(a) (posi-
tion sensitive detector) and Fig. 8(b) (step scanning) are
0.18° and 0.13", respectively. This difference is expected be-
cause of the different resolution parameter associated with
each procedure.
IV. REMARKS
2.0 4.0 6.0 8.0 10 The characteristics of the first x-ray gas position sensi-
tive detector built at CBPF, are comparable with those of
FIG. 8. Diffraction patterns from a paraffin sample using: (a) the position similar detectors reported in previous works. 4 The detector
sensitive detector (counting time: 300 s) and (b) the step scanning proce- is now used for x-ray diffraction and small-angle scattering
dure (total scanning time: 2400 s). measurements. These experiments will allow us to verify the
long-term reliability and, eventually, to improve the perfor-
mance of the detector.
integral nonlinearity has been determined to be lower than
1%. The data of Fig. 5 provide the factor which should be
ACKNOWLEDGMENT
used to obtain the coordinate x of the incoming photon from
the analyzer channel number N: (I.l.x/I:J.N) = 0.154 One of the authors (AoF.B.) is grateful to PAPERJ
mm/channel. (Rio de Janeiro Institution) for its financial support.
The plot in Fig. 6 represents the x-ray intensity as a
function of the relative position of the incoming beam. We
can see that the variation in photon count number is approxi- Ie. E. T. Gon,(alves da Silva, A. R. Rodrigues, and A. F. Craievich, Synch.
mately 1.5%. The statistical relative error for the number of Rad. News 1, 28 (l988).
2A. R. B. de Castro, J. E. A. Verdugo, and P. T. Fonseca (to be published).
recorded photons Nis 0/$) = 0.5%. From this we esti- 'A. Gabriel, Rev. Sci. lnstrum. 48, 1303 (1977).
mate the relative variation in the intrinsic detector efficiency 4R. A. Boie, O. Fischer, Y. Inagaki, F. e. Merrit, Vo Radeka, L. C. Rogers,
to be about 1%. and D. M. Xi, Nucl. lnstrum. Methods 201,93 (1982),
2317 Rev. Sci. Instrum., Vol. 60, No.7, July 1989 Data acquisition and image detectors 2317
This article is copyrighted as indicated in the article. Reuse of AIP content is subject to the terms at: http://scitationnew.aip.org/termsconditions. Downloaded to IP:
...'.•.•.'.•.•.••••..-•.•••.. •••;••••••••••'";••'•••.•.•:<.:.:.:.:.:.:.:.;.:.:.:.:.;.:.;••• ..••.•:•.216.165.95.79
:.:.~.:.:.:.:.:.:.:.:.:.;·7.·.·.·.·.·.·.'.:.:.:.:.:.:.:.:.;.:.:.:.:.:.:.;.:.:.?~ -:•••:••• On:
..•..•;'••..•..'.'.'Wed,
~ •.••• 10 Dec 2014
•••.••••.••.•.•••••• 17:28:31
:.:.:.:.:.:.;.:.:.:.;.:.;.:.:.;.;
:.:.:.:.:.:.:':.:.:.:.:o:.:.:.;.;';O~'.'. ••••.•.•••.•'••
~.:.:.:.:.:.:.;.:o;.;.;.;.~.~ y.·A·.·•••• " I ....... _.. • .... _. .. "" . .- .••.••-.-•••. __
;';>•••'.'.' •••' •••:.~ •••••••-.-•••••••••••.•••• -.....................-...... •• ~.'.-.- "0' •••••