Product Data

No. MPDCT0162EAD
Multislice HELICAL CT SCANNER

MULTI

APPLICATION

This system, the Aquilion MULTI, is a


multislice Helical CT scanner capable of
scanning the whole-body.
The system generates four slices per
revolution, using the Selectable Slice-
thickness Multi-row Detector (SSMD).
Since it can also perform high speed
reconstruction at a maximum 12 frames
per second, every examination can be
processed at ultra-high speed.
In combination with the Aspire CI
technology, it is possible to quickly
perform more accurate scanning and
examinations.

FEATURES

• High-speed scanning
This system permits simultaneous resolution of 2.5 mm at 0.25%, and The image of any slice thickness can
4-slice acquisition per 0.5 second high-contrast resolution of 0.35 mm. be generated later by reconstruction
rotation using the SSMD and pro- Multislice helical CT can be per- retry. For example, after the 2-mm
viding surpossed volumetric data formed using thin slices in routine slice thickness scanning is performed,
acquisition. The waiting time for examinations, so that high-precision, the images of several slice thicknesses
cooling of the X-ray tube is reduced, 3D and MPR images can be generat- can be generated from raw data such
because the exposure time is short. ed from the high-resolution voxel as a 10-mm thick image for routine
Therefore, patient waiting time and data. In CT angiography of the examinations, a 5-mm thick image
throughput is improved. brain, for example, a 40-mm for detail examination, and 2-mm
distance can be scanned in 5.5 thick image for generation of a 3D
• Wide range of slice thick- seconds at a slice thickness of 1 mm. volume.
nesses The use of “stack scan” with thin
The SSMD, which provides both slices minimizes partial volume • Multislice CT fluoroscopy
high-speed and high-resolution artifact. (option)
scanning, allows slice thicknesses Conventional CT fluoroscopic shows
of 0.5 mm, 1 mm, 2 mm, 3 mm, 4 • High patient throughput only a single slice, but multislice CT
mm, 5 mm, 8 mm, or 10 mm, to be In combination with fast scanning, permits realtime image reconstruction
selected, based on the objectives of the high-power 60-kW X-ray genera- to display 3 images obtained by com-
the examinations. tor and 7.5-MHU X-ray tube, bining 4 slices. Multislice CT
enable the data for routine examina- fluoroscopy significantly improves
• High quality image tions and the biopsy and 3D genera- operability in biopsy and interven-
The system achieves low-contrast tion to be acquired in 1 scan. tional procedures.
Product Data
No. MPDCT0162EAD

• Easy operations • Optimal space utilization


The system consists of two stand-alone consoles (one The Aquilion MULTI consists of only three compo-
for monitoring and the other for image processing). nents - gantry, couch and console. Even though it has
Both consoles are provided with a hybrid keyboard, a large gantry aperture of 72 cm in diameter and a
monitor and mouse, and are capable of simultaneous couchtop width of 47 cm for enhanced patient
parallel processing, significantly improving efficiency in comfort and safety, the whole system fits into an area
performing image processing and diagnosis. The use of as small as 25 meter square.
a large-size 21-inch color monitor, a sophisticated
GUI, and high-performance CPU has improved ease
of operation and reduced operation time.

COMPOSITION

• High speed 3D processing system, VolumeView


MULTI (CTD-11A)
standard composition • CBP-Study (CSCP-001A)
(Model: TSX-101A/4) • CT Cardiac Functional Analysis software***
(1) Aquilion system (TSX-101A/2) (CCFA-001A)
• Gantry ..............................................................1 • Stereotactics (CAU-11A)**
• Patient couch....................................................1 • Diagnostic image processing system, Independent type
• Console ............................................................1 Anet50 (TAN-003A)
• Accessories • System transformer (CETF005C)
– Set of phantoms ..........................................1 • Quantitative bone mineral study system**
– Head rest......................................................1 (CBM-13C)
– Set of operation manuals ............................1 • Display system for dental application (CDP-06C)
Note: Console desk is not included as standard • Multislice CT fluoroscopy (TSXF-003C)
(2) Multislice CT system upgrade kit (CGS-22A/2) • Diagnostic image display unit (Flat type)
• SSMD ..............................................................1 (CMM-003B)
• Data Acquisition System (DAS) • Pediatric scanning system (CHKS-001A)
for multislice CT..............................................1 • Patient observation system (CPM-05A)*
• MUDAT for multislice CT ............................1 • Video interface (CVF-02B)
• Gantry modification kit ..................................1 • DICOM Storage SCP (COT-30A)
• Console for multislice CT................................1 • DICOM MWM (COT-32A)
• Phantom set......................................................1 • DICOM MPPS (COT-33C)
• System software ................................................1 • DICOM Q/R SCP (COT-34C)
• Set of operation manuals..................................1 • DICOM Q/R SCU (COT-35C)
• PC online interface (COT-37B)
This kit is assembled into the main Aquilion system at • Color printer interface (CCP-02A)
the manufacturing site, prior to delivery. • ECG-gating scan system (CHEG-002A/1)
• ECG-gating reconstruction system (CHEG-002A/2)
Optional items • Extend raw data disk (EISC-006A)
• Diagnostic image display unit (color 21 inch HQ) • Extend image data disk (EISC-005A)
(CMCM-002B) • Extension couch-top kit (CACT-005A, CACT-007A)
• Diagnostic image display unit (LCD 18 inch)
(CMFM-001A) * Not available for sale in European countries.
• Image processing software, ALATOVIEW ** Not available for sale in U.S.A.
(TIWS-001B) *** Work-in-progress
• Fly-through software (CFT-02A)
2
Multislice HELICAL CT SCANNER
MULTI

PERFORMANCE SPECIFICATIONS • Tube position for


scanoscope: 0°, 90°, 180° and 270°
Scan parameters Any arbitrary angle can be
• Scan regions: Whole body, including head specified (in 5° increments)
• Scan system: 360° continuous rotate/rotate • Gantry aperture: 720 mm diameter
• Scan time
– CT scan: 0.32 s (Partial), 0.5, 0.75, 1, Patient couch
1.5, 2 and 3 s (360°) • Vertical movement
– Scanoscope: Arbitrary up to 14 seconds System: Hydraulically driven
depending on the length of – Speed of vertical
the scan field along the movement: 16 to 30 mm/s
patient axis. – Stroke: Approx. 650 mm
• Scan cycle time (for 0.5 s scan) – Minimum couchtop
– SCAN & VIEW height: Approx. 300 mm
mode: 1.5 s (one image display) Remote control from the
– SCAN & SCAN console is possible.
mode: 1.5 s (rapid sequence, couch Note: This function allows the user to check the
movement 10 mm) image on the console and adjust the couchtop
Note: Scan cycle time refers to the time between one without leaving the console. Adjustment can
scan initiation and the next. be made in 10 mm increments at the console.
SCAN & VIEW mode permits immediate • Couchtop movement
viewing of images upon completion of each System: Motor driven or manual
individual slice. – Speed of movement: 100 mm/s (fast)
• Scan field 10 mm/s (slow)
– CT scan: ø180 mm (SS) – Stroke: 1,820 mm
ø240 mm (S) – Scannable range: 1,430 mm (with head rest)
ø320 mm (M) – Step feed pitch: 0.5 to 600 mm in 0.5 mm
ø400 mm (L) unit
ø500 mm (LL) – Reproducibility: ±0.25 mm
– Scanoscope: Repeatable to within
Lateral direction Axial direction ±0.25 mm after 600 mm of
movement
240 mm (S)
Arbitrary from – Remote control from the console is possible.
400 mm (L) • Couchtop width: 470 mm
200 mm to 1,390 mm
500 mm (LL) • Load limitation
• Slice thickness/ – Max. allowable load: 205 kg (450 lb)
Number of slices: 0.5 mm × 4
1 mm × 4 Voice-recorded instruction and scan
2 mm × 4 system (VoiceLink)
3 mm × 4 Vocal instructions to the patient can be recorded electron-
4 mm × 4 ically by the operator and replayed during scan sequences
5 mm × 4 automatically as a part of the eXam Plan.
8 mm × 4 • Available number
10 mm × 2 of messages: Max. of 32
• Number of slices: Max 4 slices/scan • Available number
• Gantry tilting angle: From forward 30° to of seconds: Max. of 128 s for a total of
backward 30° 32 messages.
(in increments of 0.5°)
Remote control from the
console is possible.
3
Product Data
No. MPDCT0162EAD

• Delay time setting: Time between the end of • Reconstruction method: MUSCOT Reconstruction
reproduction and scan start (MUltiSlice COne-beam
can be set up to 10 s in Tomography)
increments of 1 s. • The relation between the slice thickness and the
imaging area scanned in 5 s (0.5 s, 10 rotations) in the
Helical scan longitudinal direction are shown for helical pitches 3.0,
• X-ray tube rotation 4.5, and 6.0.
speed: 0.5, 0.75, 1, 1.5 s/360° Image slice Setting slice Helical pitch (Note)
• Maximum time of one thickness thickness 3.0 4.5 6.0
continuous scan: 100 s
• Scan start time delay: Min. 1 s 0.5 mm 0.5 mm 15.0 mm 22.5 mm 30.0mm
Setting is possible in steps of 1 mm 1 mm 30.0 mm 45.0 mm 60.0 mm
0.1 s. 2 mm 2 mm 60.0 mm 90.0 mm120.0 mm
• Scan field in axial 5 mm 5 mm 150.0 mm225.0 mm300.0 mm
direction: Max. 1390 mm/scan
Up to 10 scan plans are 10 mm 8 mm 240.0 mm360.0 mm480.0 mm
programmable in one eXam Note: The couchtop traveling distance per rotation is
Plan. (Multiple and/or shown in proportion to the slice thickness.
Multi-directional Helical)
• Couch speed: This can be specified Dynamic scan
arbitrarily within the range of • Scan time: 0.5, 0.75, 1, 1.5 s/360°
2.5 mm to 96 mm/s. (in • Programmable time: Max. 1 hour
steps of 0.1 mm/s) • Number of
• Helical scan pitch 2.5/3/3.5/4.5/5/5.5/6 programmable scans: Max. 100 scans
• Image reconstruction • Maximum time of one
time: Min. 1.2 s continuous scan: 100 s
• Real time Helical • Scan plan
reconstruction time: 0.13 s/image – Scan interval: Min. interval is 1 s.
(1 slice/512 × 512 matrix) Setting is possible in steps of
0.08 s/image 0.1 s in scan interval more
(1 slice/256 × 256 matrix) than 1 s.
• SureStart Note: When a scanning mode with patient couch
– Max continuous movement is used, the minimum scan interval
scan time: 100 s is limited by the time for movement.
– Region of interest – Scan plan
(ROI): Max. 3 ROIs programming: Up to 360 different
– CT number Sequences can be pre-
measurement interval: 0.125 s programmed.
– Scan start time delay: Min. 3 s • Scan start time delay: Min. 0.5 s
– Display function: CT number lapse time Setting is possible in steps of
• Specification of 0.1 s.
reconstruction position: By entering the couchtop • Maximum scan rate: Max. 200 scans/100 s
position, or on the (0.5 s scan, 200 rotations)
scanogram (in minimum • Image reconstruction
units of 0.1 mm) – Full image: 1 image/360°
• Reconstruction Mode: Full image – Half image: 1 image/229°
Half image • Partial image: Reconstruction is possible in
Detail image steps of 0.1 s.

4
Multislice HELICAL CT SCANNER
MULTI

• Reconstruction time: 0.5 s Data processing


• Real time reconstruction • Reconstruction matrix: 512 × 512
time: 0.13 s /image • Picture element (pixel) size
(1 slice/512 × 512 matrix) – CT image:
0.08 s /image Unit: mm
(1 slice/256 × 256 matrix) Scan field SS S M L LL
• Rapid sequence scan: Minimum scan cycle time * to * to * to * to * to
Approx. 1.5 s at 10 mm Pixel size
0.35 0.47 0.63 0.78 0.98
pitch with 0.5 s scan
• Multi-position (*: depending on Vari-Area or Zoom factors)
dynamic scan: Maximum number of 8
positions – Scanogram: Unit: mm
Enlargement
X-ray generation Standard 2× 4× 8× 16×
ratio (area)
• X-ray beam shape: Fan shaped, fan angle 49.2°
• X-ray exposure: Continuous 2.00 (LL) 1.41 1.0 0.71 0.50
• X-ray tube voltage: 80, 100, 120 and 135 kV Pixel size 1.00 (L) 0.71 0.50 – –
• X-ray tube current: 10 mA to 500 mA 0.50 (S) – – – –
(in increments of 10 mA)
• X-ray tube heat capacity: 7.5 MHU • Reconstruction
• X-ray tube cooling rate: 1,386 kHU/min (max.) filter functions: Total of 99 available filter
1,008 kHU/min (actual) spaces
• Focal spot size Region
– IEC standard,
From FC01 Abdomen with BHC
nominal: 0.9 mm × 0.8 mm (small)
1.6 mm × 1.4 mm (large) From FC10 Abdomen without BHC
From FC20 Brain with BHC
X-ray detection From FC30 Inner ear and bone
• Detection system: Solid state detectors From FC40 Brain without BHC
• Main detector: 896 channels × 34 elements From FC50 Lung
• Data acquisition: 896 channels × 4 rows FC70 Maintenance
• Reference detector: 1 set From FC80 Super resolution mode for
• Viewing rate: 1,800 views/s (max.) inner ear, bone and lung
FC90 High resolution mode

• Image reconstruction time


– CT scan: 0.5 s (for 0.5 s scan)
– Scanoscope: Reconstructed and displayed
simultaneously with scanning
(real time reconstruction)
• Data processor
– Central processing
unit: 64-bit microprocessor
– Memory size: 1 Gbyte
– Magnetic disk unit: Raw data, 36 Gbytes
Image data, 9 Gbytes
– Magneto-optical disk
unit: 5-inch type, 2.6 Gbytes
(double sided)

5
Product Data
No. MPDCT0162EAD

Data storage – Enlargement (2×, 4× for L and LL size; 8×, 16× for
• Magnetic disk LL size)
– Raw data: max. 4,000 rotations • CT image processing
(0.5 s scan) – ROI setting and processing
– Image data: max. 16,000 · ROI shape: Point, rectangular, polygonal,
• Magnetooptical disk: Approximately 9,600 images elliptical, irregular
or 600 rotations raw data per · ROI processing: Mean value, standard devia-
cassette (2.6 Gbyte) tion, area, number of pixels
· ROI display: 3 ROIs can be displayed on
Image display an image
• Display monitor: 21-inch color × 2 units · ROI control: Size, position, rotation
• Monitor matrix: 1280 × 1024 – Measurement of distance and angle between two
• Image matrix: 1024 × 1024 (max.) points
• CT number – Profile (oblique profile also available)
display range: From −1,024 to +8,191 – Histogram
• Window width: Continuously variable – CT number display
(adjustable at variable speed) – Mark display (grid display, scale display)
• Window level: Continuously variable – Volume calculation
(adjustable at variable speed) – Enlargement (2×, 4× and arbitrary size)
• Preset window: 3 kinds of window settings – Addition/subtraction between images
may be preset for each image. – Band display (non-linear windowing)
• Window types: Linear, non-linear (6 user- – Comment and arrow insertion
programmable), and double – Top/bottom, right/left, black/white reversal of
windows. image
• Image retrieval – Image filtering
– Method: On-screen menus and – Screen save
keyboard – High speed axial interpolation
– Mode: Image, series and patient – MultiView (Auto MPR)
• Autoview function: Software control, • Raw data processing
function key – Multi-zooming (processing of raw data for 4 ROIs
• Multi-frame display: Reduction/cut-off display, and multiple images)
ROI processing – Stack reconstruction
• Inset scanogram display – Protect/Unprotect
• Selective related information display – Play/Reverse reconstruction
• Cine display • System management
– Number of frames: Up to 128 – Warm-up
– Image display speed: The display speed can be – Calibration data
varied up to a maximum of – Picture retouching
20 frames/s. – Scanogram/CT image switching

Image processing Clinical application


• Scanogram processing • Dynamic study
– Slice position display (display of planned slice, – A mathematical function is fitted to the time curve
preset slice, and last scanned slice) generated for the designated ROI within an image,
– Anatomical scale (display of relative position, taking and the functional parameters are displayed.
any point selected as standard as 0) – ROI shape: Rectangular, polygonal, ellip-
– Slice position setting tical, irregular

6
Multislice HELICAL CT SCANNER
MULTI

– Fitting function: Gamma function (Perform • MPR


least squares approximation.) 3 orthogonal planes/oblique image
Note: The above parameters may be modified or Curved MPR
additional parameters may be provided to
upgrade the system. Image transfer
• Functional image • Ethernet network: Toshiba protocol
– In addition to processing images, Aquilion can (100 Base-T) DICOM storage SCU
create functional images or image maps as well.
– Parameters which can be used to generate Auto filming
a functional image: Peak height • Ethernet network: Toshiba protocol
Peak time DICOM PRINT
Appearance time
Area under curve IMAGE QUALITY
Mean transit time
2nd moment • Noise
Transit time – Standard deviation: Less than Less than
Note: These parameters may be modified or addition- 0.25% 0.5%
al parameters may be provided to upgrade the – Scan parameters
system. · Tube voltage: 120 kV 120 kV
– Size of the calculation · Tube current: 500 mA 400 mA
area: 1 × 1 to 10 × 10 pixels, · Scan time: 0.5 s 1s
variable · Function type: FC41 FC70
– Area to be calculated: By specifying a rectangular · Slice thickness: 10 mm 10 mm
ROI, the ROI is divided into · Scan field: S S
elements of the size above – Phantom: ø16 cm ø24 cm Water
and parameters for each CATPHAN
element are calculated. • Spatial resolution: 13 lp/cm at MTF10%
– Resultant file: Filed as a 512 × 512 matrix 14.5 lp/cm at MTF2%
image. 18 lp/cm at cut off
– Scan parameter
· Tube voltage: 120 kV
3D color image processing · Tube current: 200 mA
High quality 3D images can be obtained very quickly · Scan time: 1s
with easy operation. · Slice thickness: 2 mm
• 3D surface rendering · Scan field: S
– Clipping, texture or non-texture · Function type: FC90
• 3D volume rendering – Phantom: ø16 cm CATPHAN
– Maximum intensity projection (Max-IP) • High contrast resolution
– Minimum intensity projection (Min-IP) – High resolution
– X-ray volume rendering mode (FC90): 0.35 mm ±0.05 mm
– Intensity volume rendering – Standard mode
– Shaded volume rendering (opacity curve can be set (FC30): 0.55 mm ±0.05 mm
arbitrarily) – Scan parameters
• Display/processing function · Tube voltage: 120 kV
Zooming, panning, measurement (distance, angle), · Tube current: 300 mA
annotation, cutting, drilling · Scan time: 0.5 s
• Cine display · Slice thickness: 2 mm
• Segmentation · Scan field: S (30 mm: use of zooming)
Partial extraction can be performed
7
Product Data
No. MPDCT0162EAD

– Phantom: Toshiba standard patient. X-ray high-voltage generator is built-in the


• Low contrast resolution: gantry. The system employs a high-frequency inverter for
3 mm 5 mm generating and stabilizing the high voltage supplied to the
at 0.3% at 0.3% X-ray tube. Included in the generator electronics are
– Scan parameters circuits for controlling the speed of the rotating anode in
· Tube voltage: 120 kV 120 kV the X-ray tube. Use of a high-frequency inverter system
· Tube current: 500 mA 300 mA results in high power output coupled with excellent sta-
· Scan time: 0.5 s 0.5 s bility. In addition, the system is compact and light
· Function type: FC41 FC41 weight.
· Slice thickness: 10 mm 10 mm
· Scan field: S S Patient couch
– Phantom: ø20 cm ø20 cm The patient couch is located in front of the gantry and
CATPHAN CATPHAN supports the patient. The entire unit moves vertically and
– Dose the top moves longitudinally. In case of an emergency,
· Surface: 35 mGy 21 mGy the couchtop may be pulled out manually with very little
• CTDI (CT Dose Index, unit: /100 mAs) effort. The couchtop can be lowered to a minimum
– Head center: 18.0 mGy height of 30 cm from the floor. This facilitates transfer-
surface: 19.4 mGy ring the patient from a low bed or stretcher.
– Body center: 7.4 mGy
surface: 13.6 mGy Console
– Phantom: Dosimetry Phantom The two stand-alone consoles (one for scanning, and the
Head ø160 mm, acrylic other for image processing) are both provided with a
Body ø320 mm, acrylic hybrid keyboard, monitor and mouse. These consoles can
be operated independently, and simultaneously parallel
processing can significantly improve the efficiency of
SYSTEM COMPONENTS AND THEIR image processing and diagnosis.
FUNCTION • Functions of the console for scanning
– Selection of scan parameters
Gantry – Scanoscope control
The gantry and the patient couch constitute the scanner. – Scan control
The scanner uses a fan-shaped continuous X-ray beam to – Remote control of couchtop movement
scan the region to be examined. Transmitted X-rays are – Remote control of gantry tilt
detected and converted into electric signals by SSMD. • Functions of the console for scanning and image
The gantry consists of the main body and its support processing
mechanism. The X-ray tube and the SSMD are – Window level and window width adjustment
mounted facing each other on either side of the gantry – Other various mouse-operated image processing
aperture. The X-ray tube and detectors rotate continu- functions
ously about the aperture of the gantry. A slip-ring is
employed to transmit power between the gantry and the Magneto-optical disk unit
rotating X-ray high-voltage generator assembly. The magneto-optical disk unit permits repeated recording
The gantry may be tilted forward and backward in order and retrieval of data, like a floppy disk drive. It is a
to perform tilt scanning. Three-dimensional alignment compact, externally mounted, large-capacity digital
lights are provided for setting slice positions. Gantry and storage device that can be used to save a large amount of
patient couch operating controls are located on both sides data.
of the front of the gantry housing. The patient guide Images can be automatically archived to the magneto-
display indicates the scan status to the operator and the optical disk unit via the eXam Plan.

8
Multislice HELICAL CT SCANNER
MULTI

OPERATING FEATURES technician to check the scan field on the image,


contrast study timing, presence of patient's body
Patient handling and positioning motion, etc. Therefore the patient can be released
• The couchtop can be lowered to within 300 mm of immediately after scanning.
the floor thus simplifying patient transfer to and from • SureStart function helps the technician to initiate the
a bed or stretcher. helical scanning at peak enhancement timing of
• Fast and accurate patient positioning is assisted by contrast study. SureStart starts monitoring the scan
alignment lights mounted on the gantry aperture. from the start of contrast study at a certain slice
• High-precision couchtop positioning is possible from position, to monitor the change in CT number on the
the integrated console or by manual operation from image being displayed in real time. When the contrast
the control panel and clear digital readouts are located reaches the set value, the helical scan automatically
on the gantry. starts. This technique provides contrast study at peak
• Patient observation system (optional) in the operator's enhancement timing regardless of each patient's blood
room ensures patient safety during scan procedure. flow speed and also saves contrast medium.
• The couchtop can be pulled out manually in case of
an emergency. Data processing
• Rapid reconstruction reduces reconstruction time to as
Scanning short as 0.5 second for 0.5 second scans. (Scan &
• Toshiba Scanoscope provides a radiographic-like full- View mode)
view image of the patient for high-precision advance • A variety of reconstruction algorithms are available
planning of the slice positions. according to the anatomy being studied and the
• The axial length of the scanning field for Scanoscope clinical objectives. These include algorithms for the
can be adjusted up to 1,390 mm at any time even abdomen, head, bone, lung, high-resolution, soft
while monitoring the image being reconstructed in real tissue, etc.
time. This minimizes patient radiation dosage.
• The auto index functions allows automatic incre- Image display and processing
mental couchtop movement based on the slice posi- • Reconstructed images are automatically displayed
tions determined while viewing the Scanogram. according to window settings preset in the eXam Plan.
• The eXam Plan function allows simple selection of • The window save function allows the user to store an
pre-programmed scanning parameters for routine image with window settings different from the ones set
examinations and for efficient patient throughput. by the eXam Plan.
• The Vari-area function allows the user to pre-select a • Filter parameters can be customized through simple
region of interest for zooming using raw data and on-screen menu selections. These parameters include
immediate post-scan analysis. Zooming using raw the number of filtering passes, matrix size and filter
data yields higher resolution than enlarging an already coefficients.
reconstructed image. • Images can be rotated and reversed either right/left,
• Dynamic and rapid sequence scan modes. top/bottom, or black/white.
• Multislice Helical Scan acquires raw data by rotating • The Multi-frame feature allows up to 36 images to be
the X-ray tube continuously while moving the patient retrieved and simultaneously displayed on the screen.
continuously through the scanner. A volume of data is • The three-dimensional image display allows the gener-
acquired from which slices can be reconstructed at any ation of color three-dimensional and real time MPR
point along the patient axis. This scan mode is best images from the volumetric scan data acquired using
used for rapid patient scanning within one breath hold Helical Scan. This results in higher definition and
and for high definition three-dimensional and MPR image quality than images reconstructed from conven-
imaging. tional single slice scanning. This is because helical
• Realtime helical reconstruction mode makes it possible scanning achieves better data continuity along the
to observe the images being scanned in real time at patient axis than conventional scanning.
maximum 12 frames per second. This mode shows
shift in the slice position in real time and helps the
9
Product Data
No. MPDCT0162EAD

Image storage and archiving DIMENSIONS AND MASS


• The system comes with a 45 Gbyte magnetic hard
disk as standard equipment allowing for the on-line
storage of approximately 16,000 images with max. Dimensions Mass
Unit
4,000 rotations raw data. L × W × H mm (in) kg (lb)
• A 2.6 Gbyte magneto-optical disk is provided as 960 × 2,330 × 1,950 1,750
Gantry
standard equipment. The image storage capacity of (37.8 91.7 76.8) (3,857)
the magneto-optical disk is approx. 9,600 images per 2,390 × 620 × 90
3 330
Patient couch
disk. (94.1 24.4 15.4) (727)
Console (when a
Image filming recommended 800 × 1,600 × 1,265 250
• Filming of images can be carried out manually or dedicated desk* is (31.5 63.0 49.8) (551)
automatically from the console. used)
• Automatic filming sends an entire study to the laser 420 × 1010 × 700 135
Navibox
camera. Filming is carried out in background mode so (16.5 39.8 27.6) (298)
that other scanner and image processing functions can *Toshiba recommends a width of over 1,600 mm
be performed without interruption or delay.
SITING REQUIREMENTS
Patient throughput
Patient throughput and cost effectiveness have been given Power requirements
great consideration in designing and building the • Phase: Three-phase
Aquilion MULTI CT scanner. • Voltage: 200 V*
• The system incorporates a 7.5 MHU X-ray tube with • Frequency: 50 Hz or 60 Hz ±0.5 Hz
a very fast cooling rate of 1,008 kHU/min, in actual • Capacity: 100 kVA
use. • Voltage fluctuation
• High speed scans can be conducted in as little time as due to load change: Less than 5%
0.32 second per scan. Routine scans can be performed • Power voltage
as quickly as 0.5 second per scan. fluctuation: Less than 10%**
• In multislice helical scanning, the scanning can be per- * Please consult Toshiba for solutions to other voltages
formed at maximum 12 slices per second. and excessive power fluctuations.
• Real time reconstruction for scanoscope. ** This represents the total voltage fluctuation of load
• CT image reconstruction in 0.5 second for 0.5 second variation and power.
routine scans.
• Routine scan cycle times as short as 1.5 seconds. (Scan Grounding
& View mode) In accordance with local regulations for medically used
• Ease of operation with an optimized use of hybrid electrical equipment.
keyboard, mouse-driven menus and large color
monitor. Ambient conditions
• Convenient patient handling with a couchtop that can • Scanner room: Gantry
be lowered very near the floor. Patient couch
– Room temperature: 18°C to 28°C
– Relative humidity: 40% to 80%
(no condensation)
• Operator's room: Console
– Room temperature: 16°C to 28°C
– Relative humidity: 40% to 80%
(no condensation)
Refer to the site planning guide MOICT0047EA or the
site planning manual No. 2A201-377E for details on the
10
air-conditioning work.
Multislice HELICAL CT SCANNER
MULTI

OUTLINE DRAWING

960
(37.8)
380 580
(15) (22.8)
30° 30°
(TILT ANGLE)

(76.8)
1,950

2,057
(81)
1,016
(40)

2,330
(91.7)

Gantry
(18.5)

(24.4)
470

620

1,820 2,000
(71.7) (78.7)
Couchtop stroke

1,490 2,390
(58.7) (94.1)
(25.7)
655
Stroke

(41.1)
300 (37.5)

1045
955

(15.4)
390
(11.8)

Patient Couch
Unit: mm (in)

11
Product Data Multislice HELICAL CT SCANNER
No. MPDCT0162EAD
MULTI

OUTLINE DRAWING

1,600 800
(31.5)
(63.0)
(49.8)
1,265
(31.5)
800
(27.6)
700

Navibox

1,010 420
(39.8) (16.5)

Unit: mm (in)

Console

Note: Console desk is not included as standard. The photograph of the system on the cover differs from the drawing above.

TOSHIBA CORPORATION
MEDICAL SYSTEMS COMPANY
http://www3.toshiba.co.jp/medical
©Toshiba Corporation 1999 all rights reserved. Toshiba Corporation Medical Systems Company
meets internationally recognized standards
Design and specifications subject to change without notice. for Quality Management System Toshiba Nasu Operations meets the Environmental
ISO 9001, ISO 13485, EN 46001. Management System standard, ISO 14001.
2001-10 TME/SB
Printed in Japan