Professional Documents
Culture Documents
6-Gas Chromatography-Mass Spectrometry (GC-MS) - Working Principle and Applications.
6-Gas Chromatography-Mass Spectrometry (GC-MS) - Working Principle and Applications.
1. Introduction
Gas chromatography-mass spectrometry (GC-MS) is a hyphenated technique
consisting of two analytical procedures coupled in series that is a Gas Chromatography
(GC) for separation of volatile compounds and Mass Spectrometry (MS) for
identification and quantification of molecular weight of each separated component. The
GC helps in separation of multiple components mixture so that they can reach the mass
spectrometry one at a time. Thus, the molecular mass and other fragments mass values
describe only for a single component at a time.
87
Fig. 1. GC-chromat
G togram
a. Inj
njector : Sammples are vapporized
b. Caarrier gas: Nitrogen
N (N2),
) Helium (H He) or Hydroogen (H2)
c. Ov ven: Isothermmal or gradiient temperatture
d. Co olumns: Cappillary Colum mn and Packked column
e. Deetectors: Diffferent detecctors are useed dependingg upon selecctivity and/oor
seensitivity of detection
d forr specific com
mpound. Foor example,
Electron--capture deetector (ECD D): ECD is sensitive likke FID, but has a limiteed
mic range. Mostly
dynam M usedd in the analyysis of electtronegative atoms
a containing organiic
molecules such as
a halogens, phosphorous
p s and nitro groups.
g
Amobile Astationnary
The equillibrium connstant (K) iss termed as partition coefficient orr distributioon
consttant is defin
ned as the degree of thet solute molar
m concenntration (CM) distributees
themselves betweeen the mobiile and statioonary phasess.
88
K = [CM] / [CM]
S M
Where, [CM] is molar concentration of the analyte in the SP and [CM] is molar
S M
concentration in the MP. If K is constant over a wide range of analyte concentrations,
then [CM] is directly proportional to [CM] and chromatographic peak is symmetrical,
S M
Gaussian distribution and retention time are independent of the amount of analyte
injected.
Retention Times of an analyte is defined as the time it takes after sample injection for
the analyte to reach the detector and elute. The time for non-interacted species with SP to
reach the detector is defined as the dead time (tD), if taken in volume called dead volume
(VD). Thus.the rate of migration of a non-interacted species is the same as the rate of
motion of the mobile phase expressed as,
Rate = L / t ………………….(i)
M D
Where, L is the column length and t is the retention time of the mobile phase/
D
non-interacted species.
Similarly, the linear rate of a solute molecule is calculated by dividing the column
length by the retention time t ,
R
The linear rate of a solute molecule can be expressed as a function of the rate of
migration of the unretained species,
Where, S (t) is the fraction of time the solute spends in the mobile phase.
M
HETP = L / N …….………….(iv)
89
Fig. 2. GC-chromatogram
Determining the number of theoretical plates: From Fig. 2, the number of theoretical
plates can be calculated,
2 2
N = 16 (t / W) = 5.55(t / W ) ………..(v)
R R 1/2
Where, W is the peak width measured in the same units as t and W is the peak
R 1/2
width measured at half the peak height.
The resolution of a column is defined as its ability to separate a mixture of
compounds. For example, compounds Ψ and Θ were separated by the GC column as
shown in Fig. 3.
Fig. 3. GC-chromatogram
Determining the column resolution: After some mathematical manipulation, the
required parameters to compute the column resolution is as follow:
90
R = (Θt - Ψt ) / (W + W ) = Δt / (W + W ) …………(vi)
S R R Θ Ψ R Θ Ψ
The chromatogram generally plotted in detector signal vs. time, gave Gaussian-
shaped (or bell shaped) peaks. In reality, non-Gaussian peak shapes will often occur. The
front side of a peak might be drawn out while the tail on the right is steep called fronting,
might be due to large amount of a sample is introduced into the column or wrong
combination of packing material. The right side of the peak is drawn out while the tail on
the left is steep called tailing and usually occurs when the solute has a concentration
dependent distribution coefficient. Fronting and tailing will result in less accurate
quantitative analysis and peaks broadening.
Equipment calibration: The GC-MS instrument needed to be tuned before each
analysis. The calibration consists of injecting a known volume of a standard and
measuring the time between injection and elution. The retention time of a compound
remained the same for a given set of variables (temperature, flow rate and column
length). If the quantitative analysis is required, it might need to compare peak heights or
areas of the peak with that of one of the standards. The most ancient method for
quantitative analysis is the internal standard calibration method (ISTD) and involves
the preparation of a series of standards that approximate the composition of unknown. In
this procedure, a measured quantity of an internal standard is introduced into each
standard and sample. Chromatograms for the standards are generated and peak areas are
plotted as a function of analyte amount or concentration. Such plot should yield a straight
line passing through the origin and gives qualitative and quantitative analysis of different
components present in the mixture (Fig. 4).
Fig. 4. An ISTD calibration curve (2.5 area counts = 1.0 units of compound amount)
91
ionization and frragmentationn in ion souurce region, mass separaation in ion analyzer annd
detecction processses, respecttively. For example, a mixture of three com mponents goot
separrated in GC followed byy MS gave threet mass spectra
s for each
e componnent (Fig. 5).
GC and
a MS used d together too give better understandiing of substaance in idenntification annd
compposition indeependently.
Figg. 5. GC and
d MS chrom
matograms
Instrrumentation n: All masss spectromeeters have three major componennts (Fig. 6):
(i) Ionizer (ion so
ource), (ii) Ioon/mass Anaalyzer and (iiii) Detector
Fig. 6. MS
M instrum
ment
92
(i) Ionizer: A sample vaporized at >400 0C and 10-3 Pa pressure is introduced in the mass
spectrometer where molecule get ionized using different ionization methods that include
electron impact (EI), In-beam (desorption) electron impact (DEI), Chemical ionization
(CI), Alternate CI-EI (ACE), Rapid heating (desorption) Chemical ionization (DCI),
Atmospheric pressure ionization (API), Atmospheric pressure chemical ionization
(APCI), Electrospray ionization (ESI), Thermospray ionization (TSP), Laser ionization
(Laser desorption), Matrix assisted Laser ionization (MALDI), Field ionization (FI) and
Field desorption (FD), Fission fragment ionization or Plasma desorption mass
spectrometry (PDMS), Secondary ion mass spectrometry (SIMS), Liquid SIMS, Fast
atom bombardment (FAB), Photon and multiphoton ionization (MPI) and Resonance
Ionization, etc. Mostly, MS ionizers work with positive ions (negative is rare). For
example, in EI-MS technique, 70 eV electron energy is used for molecule ionization
where atom knocked out one or more electrons off to give a positive ion (molecular ion)
and lower the electron energy. The high energy radical cation (mol. ion) undergoes
further fragmentation to minimize its energy (Scheme 1).
General reaction:
70 eV
A-B-C (vapor) + e- ------------Æ A-B-C.+ + 2e-
(radical cation)
In the ionization chamber, the vaporized molecules are bombarded with a stream
of electrons emitted from electrically heated metal coil. Some of the collisions are
energetic enough to knock out one or more electrons to make positive ions. Mostly, the
positive ions formed carry single charge because it becomes much more difficult to
remove further electrons from a positive ion. The free electron(s) are attracted towards
the electron trap (+vely charged plate called electron collector). The ions produced in the
ionization chamber travel forward direction only due to repeller ((+vely charged plate)
93
placeed from other side. In ion source and mass analyzer a secctions, a higgh vacuum is
i
requiired to avoid
d high energyy electrons too strike withh air moleculles (Fig. 7).
Fig. 7. Ion
nization Chaamber
The positiive ions are repelled awaay from the very positivve ionization chamber annd
pass through threee slits, haviing potentiall difference ini decreasinng order from
m + 10,000 to t
0 volts. Thereforee, all the ionns are accelerrated and haave the samee kinetic enerrgy at the ennd
of 0 volts
v (Fig. 8)).
(ii) Mass/Ion
M Analyzer:
A A
Advancemen nt in GC-M
MS machine was obtainned by usinng
differrent mass an
nalyzers for separating the
t ion beam
m into beamss of differennt masses. Foor
exammple,
a.. magnetic sector analyyzer (MSA): gave high resolution, exxact mass
b. quadrupole analyzer (Q): gave loow (1 amu) resolution,
r faast, cheap
c.. Time-of-fflight analyzzer (TOF): haas no upper m/z limit annd high throuughput
d. Ion trap mass
m analyzeer (QSTAR):: gave good resolution
e.. Fourier Transformat
T tion Ion Cyclotron
C R
Resonance (
(FT-ICR): g
gave highesst
resolution
n, exact masss but costly.
94
Based on n use of different
d tyype of ion source andd mass analyzer, masss
specttrometry hass been develloped such as ESI-QTO OF (electrospray ionizattion source +
quadrrupole masss filter + tim
me-of-flight mass
m analyzzer), MALD
DI-QTOF (m matrix assisteed
desorrption ionizaation + quaddrupole + tiime-of-flightt mass analyyzer), GC-M
MS (separatees
volatile compoun nds in gas coolumn and ID’s
I by masss), LC-MS (separates compounds
c i
in
HPLC C column and a ID’s byy mass), MS-MS
M (Tanndem mass spectromettry) (separatte
comppound fragm ments by maggnetic field and
a ID’s by mass), LC/L LC-MS/MS (Tandem LC C
and Tandem
T MS)) (separates by
b HPLC, ID D’s by mass and AA seqquence), etc.
In the maass analyzer,, the accelerrated ions arre then enterring into maass separatioon
unit, depending upon
u their mass/charge
m (
(m/z or m/e)) ratio. For example,
e maagnetic sectoor
analyyzer (MSA) deflects ionss by an electromagnetic field accordding to theirr masses (Figg.
9). Lighter ions area deflectedd more com mpared to heaavier. The amount
a of deeflection alsso
depennds on the number
n of poositive chargges on ion. The
T more thee ion is chargged, the morre
it getts deflected.
In Fig. 9, ion stream A is most deeflected becaause ions havve minimum m mass/chargge
ratio. Ion stream C is the leaast deflectedd because ionns have maxximum mass/charge ratioo,
assumming all are +1 ions. Thhus, stream A has the ligghtest ions, stream
s B thee next lightesst
and stream
s C thee heaviest. Lighter
L ions are going too be more deeflected thann heavy ones.
Only stream B ions i will reaach to the detector
d in this
t case. For
F stream A ions to be b
recorrded, we neeed less electrromagnetic field
f to be appplied. Convversely, for stream
s C ionns
to be recorded, wew need stroonger electroomagnetic field to be appplied whichh can be donne
by chhanging the magnetic
m fieeld.
(iii) Detector:
D Only
O stream B ions reachh to the ion detector. Thhe other ionss collide witth
the walls
w where they
t will picck up electroons and be neutralized
n (uuncharged). Finally, theey
get reemoved fromm the mass spectrometer
s r by the vacuuum pump. Ions
I detectioon is based on
o
eitherr charge or momentum
m v
value. In moodern instrum
ments, for larrge signals a Faraday cuup
is useed to collectt ions and measure
m the current. In old instrumeents, photoggraphic platees
are used
u to meassure the ionn abundance at each maass/charge raatio. Most detectors
d useed
ampliifying the ion
i signal using
u a colllector similaar to a phootomultiplierr tube. Thesse
95
ampliifying detecctors includee: electron multipliers,
m channeltrons and multichhannel plates.
The gain
g is contrrolled by chaanging the high
h voltage applied to thhe detector. A detector is
i
selectted for its speed, dynam mic range, gain
g and geoometry. Somme detectors are sensitivve
enouggh to detect single ions.
When an n ion hits thhe metal suurface, its chharge is neuutralized byy an electroon
jumping from thee metal on thhe ion (rightt hand diagraam). That leaves a spacee amongst thhe
electrrons in the metal and the
t electronss in the wirre shuffle allong to fill it.
i A flow of o
electrrons in the wire is dettected as ann electric cuurrent whicch can be amplified
a annd
recorrded throughh computer software. Laarger the ionns knockingg, the greateer will be thhe
curreent which ind
dicate the inttensity of thee peak.
The outpu ut from the chart recordder is usuallly simplifiedd into a "stiick diagram".
This shows the reelative current produced by ions of varying
v masss/charge ratio.
96
(a) (b)
1.3 Applications
assessment for the total organic carbons (TOCs) that include pesticides, endocrine
disruptors, chlorofluoro-compounds, protein degradation and radioactive products for
drinking water. Very low levels of total organic carbon are essential with a typical
resistivity of 18.2 MΩ. cm, a very low TOC value of less than 2 ppb and bacteria levels
below 0.1 CFU/ml – is highly recommended for ultra pure water which can be achieved
with GC-MS analysis.
GC-MS is used as the tool of choice for tracking organic pollutants in the
environment. There are some compounds for which GC-MS is not sufficiently sensitive,
including certain pesticides and herbicides, but for most organic analysis of
environmental samples, including many major classes of pesticides, it is very sensitive
and effective.
GC-MS can analyze the particles from a human body in order to help link a
criminal to a crime. The analysis of fire debris using GC-MS is well established. There is
even an established American Society for Testing Materials (ASTM) standard for fire
debris analysis. GCMS/MS is especially useful here as samples often contain very
complex matrices and results, used in court, need to be highly accurate.
GC-MS is used for detection of illegal narcotics and supplant drug-sniffing dogs.
It is also commonly used in forensic toxicology to find drugs and/or poisons in biological
specimens of suspects, victims, or the deceased.
GC-MS is the main tool used in sports anti-doping laboratories to test athletes'
urine samples for prohibited performance-enhancing drugs, for example anabolic
steroids.
1.3.7 Astrochemistry
98
Several GC-MS have left earth. Two were brought to Mars by the Viking
program. Venera 11 and 12 and Pioneer Venus analyzed the atmosphere of Venus with
GC-MS. The Huygens probe of the Cassini-Huygens mission landed one GC-MS on
Saturn's largest moon, Titan. The material in the comet 67P/Churyumov-Gerasimenko
will be analyzed by the Rosetta mission with a chiral GC-MS in 2014.
1.3.8 Medicine
References
-----
99