Professional Documents
Culture Documents
A5 Fluidics SV in Chamber Position v2.0
A5 Fluidics SV in Chamber Position v2.0
aperture
4Diff
V34 Sample
RBC Waste
BASO - Dilu
aperture
cleaning line
Aperture
Dilu 1
V6
V7
V8
Pressure
Dilu 3
(Yellow) Lyse
Open to Air 4m Tube
(Impedance)
Reservoir 1.
Reservoir 2.
V16/1
Reservoir
Reservoir
reservoir
Vacuum
Diff-5p
Diluent
Diluent
Sheath Laser
Lyse
1 2 3 4
Pressure
Dilu 4
BASO
V5
V3 SV SV
inner outer
Diatro Dil-5p
Lyse 2
Diff-5p IN
ring ring
Diatro Diff-5p Diatro Lyse-5p V39/1
4Diff IN
V38/1 V4 reagent Dilu 7
reagent reagent 1 2
container
V23
container container 3
Dilu 6
V36/1 V40/1 V42/1 V37/1 V41/1 V43/1 (Green) RBC
(Orange) (Yellow) 3
Vacuum Glass
reservoir
reservoir
Sheath
(Laser)
V1/1
V20 Blood
V1/1
4m Tube Sensor
1
V12
V1
1 2 Resistance
Pump 2
2
Waste 3 (4m tube)
2
4Diff
V13
container V2
3
V32
Sample Needle
(Red) Needle
1
1
Opt.
2
(Red) Waste
V15
3
2
V24 Dilu 10 Needle
1 2 Laser
Dilu 2 Dilu 7
3
BASO - Dilu
Dilu 5 cleaning line
V33 V18 V25
1 2 1 2 1 2
Dilu 4 Dilu 10 Dilu 6 WBC
3 3 3
Empty
Chamber
Pressure Lyse 1
MIX RBC
Lyse 1 Lyse 2 Dilu 3
V44 V17 V26 V28 V29 V30 WBC pre-
1 2 1 2 1 2 1 2 1 2 1 2
Dilu 1 Dilu 8 Air 3 Air 2 Dilu 9 heater
3 3 3 3 3 3
Diff-5p IN
WBC/HGB
chamber
chamber
chamber
Chamber
Vacuum
MIX
RBC
MIX
V36 V37 V38 V39 V40 V41 V42 sample V34 V35
V43
V43/1 1 2
V36/1 1 2
V37/1 1 2
V38/1 1 2
V39/1 1 2
V40/1 1 2
V41/1 1 2
V42/1 1 2
Air 1
3 3 3 3 3 3 3 3
Dilu 8 Dilu 9
Air 3
Diff-5p
Air 1 Air 2
Lyse
Dilu
Dilu
Dilu
Dilu
Dilu
Dilu
BASO
V10
V11
V9
Dilutor Dilutor Dilutor Dilutor
Size Cat. no. Drawing number Change
Nr.1 Nr.2 Nr.3 Nr.4
Drawn by: Rajkai Márton / Tóth Eszter A2 Abacus 5 Fluidic diagram_Shear Valve in Chamber Position V2.0
(Diluent) (Diff-5p) (Lyse) (Air)
Released: 21.06.2016 Scale No scale Page 1/1