Instrument reprocessing
in the dental practice
dentsplysirona.com
02 I 03

Infection control systems


for a high level of safety DAC Universal,
DAC Universal S
Infection control in dental practices is becoming even more important, and with The combination machine cleans, lubricates
(if necessary) and disinfects1 / sterilizes2 up to six
such significance comes increased monitoring. Ensure all-round protection for your- straight and contra-angle handpieces, turbines,
self, your practice team and your patients by using instrument reprocessing with a ultrasonic / sonic handpieces and tips, nozzles of
multifunctional syringes and powder jet devices as
high level of hygienic safety and comprehensive documentation options. Infection well as powder jet handpieces in approx. 15 minutes1 /
control solutions from Dentsply Sirona are suitable for the cleaning, care, disinfection 21 minutes2 – including cooling.

and sterilization of dental instruments. Regardless of the design of your infection


control workflows, we have the appropriate solution. 1
DAC Universal
2
DAC Universal S
04 I 05

Instrument reprocessing
Straight and contra-angle handpieces, turbines, ultrasonic / sonic handpieces and
tips, nozzles of multifunctional syringes and powder jet devices as well as powder jet
handpieces place increased requirements on diligent reprocessing as a result of
the narrow media channels and the angled interior spaces. Difficulty is increased by
technological contaminations such as abrasion and oil residues in addition to typical
contaminations from treatments such as blood, saliva, secretions and tissue.

In principle, straight and contra-­ cleaning, disinfection or sterilization ­ edical devices must be defined in
m
angle handpieces and t­ urbines must (unwrapped) and wrapped the operating procedures. The
be reprocessed after each patient sterilization. Machine reprocessing reprocessing guidelines from the
treatment and require special care increases process r­ eliability, ­relevant manufacturers must be
due to design cavities. Rotating whereby the occupational safety for taken into account. All reprocessing
instruments can be classified as the practice staff is also increased. steps as well as cleaning and
semi-critical (non invasive use) or Mechanical reprocessing is disinfection1 / steriliziation2
critical instruments (invasive use). preferable to manual reprocessing measures should s­ ubsequently be
Depending on the country, the for these reasons. All workflows compiled in the hygiene plan of the
procedure of reprocessing contains: relating to the reprocessing of operating ­practice.

1
DAC Universal
2
DAC Universal S
06 I 07

The infection control cycle in Instrument reprocessing


machine instrument reprocessing methods
Cleaning and disinfection Sterilizer classification

Machine cleaning and disinfection – The standard for small steam


the safe approach to reprocessing ­sterilizers EN 13060 ­differentiates
It begins with correct hand hygiene sterilization2, packaging – if neces- DAC Premium, the instruments are between three classes of ­sterilization
With machine reprocessing, all
and thorough disinfection of all sary – and wrapped sterilization. rendered sterile. The packaging programs:
­process steps involved in cleaning
patient-related surfaces. Emphasis is The effectiveness and repeatability ­protects against recontamination
and disinfection are performed
placed on the correct reprocessing of the reprocessing processes with during storage and transportation. Class B – the universal
by an infection control system.
of medical devices – an ever-present DAC Universal / DAC Universal S are All important parameters and sterilization type
Cleaning is performed using water
process. It affects all instruments checked during validation. The the success of reprocessing are
and, where necessary, cleaning This autoclave sterilizes wrapped
that are brought into the reprocess- authorities recognize this validation ­documented after the cycle has
agents are added. Disinfection is and unwrapped solid products and
ing room after treatment. Such process. Medical devices designated been completed and then archived
mostly thermal without the addition hollow items in accordance with
instruments are then put through for sterile use are wrapped and on the practice computer.
of chemicals. According to the RKI the manufacturer’s specifications.
comprehensive reprocessing con- sealed with SiroSeal Premium.
guidelines, machine reprocessing Devices with such programs are
sisting of cleaning, disinfection1 or During subsequent sterilization in
methods are preferable to manual referred to as class B sterilizers
methods, and thermal disinfection is (e.g. DAC Premium /
favored over chemical disinfection. DAC Professional).
Machine cleaning and disinfection
devices comply with the require- Class S – for sterilization
01 ments of the international standard of medical devices
Flushing (approx. 30 sec.) of the transmission EN ISO 15883-1 / -2. This autoclave sterilizes wrapped
instruments at the treatment center and unwrapped solid products and
Manual cleaning and disinfection hollow items in accordance with
In the case of manual reprocessing, the manufacturer’s specifications
10 02
cleaning and disinfection chemicals (see manufacturer’s declaration).
Usage Removal of the drill
from spray bottles are used to clean Devices with such programs are
the inside and outside of instruments. referred to as class S sterilizers (e.g.
To ensure effectiveness, standard- DAC Universal S). The sterilization
ized work instructions must be result satisfies the same quality
09 03 ­followed very closely. The manual requirements as for Class B
Documented Wipe disinfection of method is very time intensive. ­sterilizers.
­release and the instruments at the For medical devices of invasive use,
­storage treatment center the machine reprocessing is gener- Class N – for thermal disinfection
ally recommended. This autoclave is used for
unwrapped solid products. Class N
Semi-manual cleaning cannot be used with hollow items.
08 04 and disinfection
DAC Premium Disposal of
Numerous care and infection control
Sterilization single-use instruments
devices offer automation of a part of
and proper preparation
the reprocessing process. The miss-
ing process steps have to be carried
out manually or by a different
machine system (see page 17
05 “Market overview of care and
07 infection control devices”).
Collet chuck care (weekly)
SiroSeal Premium
Packaging
06
DAC Universal /
DAC Universal S
cleaning, lubrication
and disinfection1 or
sterilization2

1
DAC Universal
2
DAC Universal S
08 I 09

DAC Universal and


DAC Universal S
Advantages

Ease of use Cost-effective and environmental Switch off infection control risks:
• New design friendly reprocessing
• Touch display with intuitive user interface • Low operating and consumption costs – Switch on DAC Universal / DAC Universal S
• Guided maintenance workflow Check & Clean no use of cleaning and disinfection chemicals
and only up to 800 ml1  /  900 ml2 water Comply with infection control standards at the touch of a button and avoid cross
consumption per cycle
• Low investment costs in instruments thanks to
contamination: Completely safe with DAC Universal / DAC Universal S. Your patients
cooling at the end of the process and therefore and employees can rely on this all-round protection and put their complete trust in
quick return to service
the treatment with the reprocessed instruments.
AG A I N
IVE ST
CT V

A
Conformity with standards

IR
Fully automatic reprocessing Legal certainty

LY

US
FUL
The cleaning and disinfection process of DAC Universal is Fully active against viruses:

ES
• Six instruments in approx. 15 minutes1 / • Cleaning and disinfection1 / sterilization2 carried out in compliance with the international standard Reprocessing with

ES

FUL
21 minutes2 process which can be validated EN ISO 15883-1 / -2 for cleaning and disinfection devices.

US
DAC Universal and

LY
IR
• Internal and external cleaning, lubrication • Cleaning and disinfection process in

A
V CT
The cleaning process of DAC Universal S is carried out DAC Universal S
ST
(if needed) and disinfection1 / sterilization2
IVE
accordance with EN ISO 15883-1 / -2 AG A I N

of straight and contra-angle handpieces, in compliance with the international standard EN ISO
• Sterilization process in accordance with The thermal disinfection of DAC Universal as
turbines, ultrasonic / sonic handpieces and 15883-1 / -2, the sterilization process in accordance with
EN ISO 13060 and ISO 17665-1 well as the sterilization of DAC Universal S,
tips, nozzles of multifunctional syringes and EN ISO 13060 and ISO 17665-1.
powder jet devices as well as powder jet • Routine Control with chemical indicator class 5 are not only bactericidal and fungicide but also
handpieces and PCD (Process Challenge Device) fully virucidal. Proven full virucidal1 activity.
Efficacy spectrum with relevant examples:
• Process safety through automatic program
selection bactericidal TBC, S. aureus

• LAN interface for electronical documentation fungicide C. albicans


virucidal HPV, HBV, HCV, HIV, COVID-19,
influenza, adenoviruses, noroviruses

1
DAC Universal
2
DAC Universal S 1
Tested with temperature resistant parvoviruses.
12 I 13

DAC Universal S – with sterilization


Reprocessing of rotating instruments
Fully automated reprocessing process
in a single cycle with the Program Pink Lid
DAC Universal S cleans, lubricates (if necessary) and sterilizes up to six straight and
contra-angle handpieces and turbines in a fully automated process. Furthermore,
ultrasonic / sonic h
­ andpieces and tips, nozzles of multifunctional syringes and powder
jet devices as well as powder jet handpieces can be reprocessed at a very high level
of hygienic safety in DAC Universal S.

Lid variations
Internal cleaning with cold water Fully automated lubrication
1. Preliminary cleaning 4. Lubrication:
 The drive channels
2. Leak test are lubricated (sufficient for the
next treatment)
3. Internal
 cleaning: The internal
spray and drive channels are
rinsed with water

Pink Lid White Lid


For the reprocessing of straight and For the reprocessing of ultrasonic /
contra-angle handpieces, turbines sonic handpieces and tips, nozzles
and contra-angle handpiece heads. of multifunctional syringes and
powder jet devices as well as
powder jet handpieces.

External cleaning with cold water Sterilization and cooling


5. External cleaning: Pulse wash 6. Heating up to 134 °C
procedure (multi-cyclical 7. Back-flush:
 Saturated steam is
cleaning method) directed through the instruments
8. Sterilization: 3 min. at 134 °C
9. Cooling
10. The lid opens slightly
11. The lid can now be opened fully

Reprocessing with the White Lid


(identical process to the Program Pink Lid, but without lubrication)

Internal cleaning with cold water External cleaning with cold water Sterilization and cooling
14 I 15

Information on the validation


of DAC Universal and
DAC Universal S
The following information is provided in line
with the statutory requirements

Validation is a process that tests process parameters or after a change


the effectiveness and reproducibility in loading. The inspection qualifica-
of the reprocessing procedure. It is tion and operation qualification are
composed of installation qualification omitted in the renewed performance
(IQ), operational qualification (OQ) qualification.
and performance qualification (PQ).
Batches must be documented; this
If the relevant authorities demand can be carried out with a printer,
complete and comprehensive initial using the practice software (also via
validation on-site in the practice, a network) or via a USB data-logger
there are various dealers and service system.
providers that offer such on-site
validation services. Complete initial Routine control tests must be done
validation locally in the practice due to the recommendations of
includes a comprehensive perfor- the manufacturer. E.g. DAC Universal S
mance qualification in addition to requires chemical indicators with
the installation qualification and every cycle and a steam penetration
operational qualification. test with a PCD regarding ISO
17665-1 once a week.
The renewed performance qualifica-
tion (revalidation) must be carried Maintenance as recommended
out after 12 months. With lasting by the manufacturer must be per-
stability of the processes as well as formed at the latest after two years
existing risk assessment by the oper- or 3,000 cycles. A maintenance kit is
ator, the interval can be raised on up available (REF. 67 15 689). The dealer
to 24 months / 4000 cycles over the technician must be allowed approx.
longterm. Revalidation is also three working hours.
required after changes have been
made to the device that influence the
18 I 19

Requirements and information Market overview of


on installation care and infection control devices
Requirements on location: Place DAC Universal / DAC Universal S in a well-ventilated For the reprocessing of turbines and straight and contra-angle handpieces
location on a flat, heat-resistant table top near to a power source. A compressed air DAC DAC Assistina QUATTRO iCare+ Melatherm PWD PG 8581 STAT- X-Cid 2 Lubrina 2
connection of 5 to 8 bar flow pressure is required. The recommended minimum Universal
S
Universal TWIN care Plus 10 Evolution 8531 MATIC
smart
distance to the wall is 10 cm. Further, there must be enough space to enable Manufacturer Dentsply Dentsply W&H KaVo NSK MELAG MIELE MIELE SciCan Micro- Morita
Sirona Sirona Mega
DAC Universal / DAC Universal S to be opened upward. The total height of the open Cycle time approx. approx. approx. approx. approx. approx. approx. approx. approx. approx. approx.
DAC Universal is 59 cm. The minimum height should be 70 cm in order to prevent 21 min. 15 min. 10 sec. 1 min. 15 min. 60 min. 60 min. 42 min 10 min. 30 min. 20 s /
w/o turbine
possible injuries when opening the lid (risk of crushing). drying approx.
40 s / c / a
  

Capacity 6 6 1 4 4 23 6 44 3 3 4
(instr.)
Weight [kg] 26 26 7.5 10 14 80 55 65 7.3 8 8
Installation of DAC Universal / DAC Universal S
Process documentation: Water n n – – – n n n – – –
A RS232 interface: recommended for printer
connection

LAN interface: recommended for the connection Waste water n n – – – n n n – – –


connection
with PC, laptop
A Compressed air n n n n n – – – n n n
connection 0.5 – 0.5 – 0.5 – 0.4 – 0.5 – 0.45 – 0.5 - 0.3 -
AC connection:
B 220 - 240 VAC, 50–60 Hz, 1,300 W
0.8 MPa 0.8 MPa 1 MPa 0.6 MPa 0.6 MPa 0.6 MPa 0.8 MPa 0.5 Mpa
External n n – – n n n n – n –
cleaning (water) (water) (cleaner) (cleaner) (cleaner) (cleaner) (cleaner)
B C
Water supply: Water from a water treatment system can be Internal n n – – n n n n n n –
connected to the water connection via a 6 mm hose. We cleaning (water) (water) (cleaner) (cleaner) (cleaner) (cleaner) (cleaner) (cleaner)
recommend NitraDem Direct ­Connect 2 (REF. 66 93 233) as (spray channel)

A the direct connection. Water can also be filled manually into Internal n n – – – n n n – – –
the water tank. Note: Water quality must be < 3 µS / cm. cleaning (water) (water) (cleaner) (cleaner) (cleaner)
(drive channel)
Note: Maximum water pressure is 6 bar.
Oil n n n n n – – – n n n
maintenance
Compressed air connection:
D Connect clean and dry air (6 mm hose).
Disinfection – n – – n n n n – n –
(thermal) (chemical) (thermal) (thermal) (thermal) (chemical)
The air ­pressure must be between 5 and 8 bar flow pressure
Sterilization n – – – – – – – – – –
(short-term air consumption: approx. 60 Nl / min. at 5 bar).
Electronic o o –1 –1 n o o o n – –
C Waste water: The waste water hose must be manufactured from
E heat-resistant material and has a diameter of 6 mm. The
documentation
Instruments n n – – n n n n – n –
maximum length is 3 m. Please use the original waste water can be directly (additional (additional (additional (additional
container (REF. 60 78 526) or an original siphon (REF. 61 26 341) used for manual or manual or manual manual or
automated automated external automated
for direct connection to the waste water system. semi-critical B
D internal, internal, cleaning, internal,
external external thermal external
cleaning cleaning disinfection / cleaning
and thermal and thermal unwrapped and thermal
disinfection / disinfection / sterilization) disinfection /
unwrapped unwrapped unwrapped

E Ultrasonic /  n n
sterilization)

sterilization)
– – n n n – –
sterilization)

sonic tips
Ultrasonic /  n n – – – n n n – – –
sonic
handpieces
Attachments n n – – – n n n – – –
for multifuncti-
onal syringes
Nozzles of n n – – – n n n – – –
powder jet
devices

n available

optional
o
– not available

1
Device is not a washer-disinfector / sterilizer
Information based on manufacturers’ details (February 2020)
26 I 27

Water supply
High-quality treated water (< 3µS / cm) is required for DAC Universal / DAC Universal S.
The majority of standard water reprocessing systems do not guarantee a ­constant flow
NitraDem Direct Connect 2 – connections
or required return flow protection that would meet this requirement. In order to avoid NitraDem Direct Connect 2 features 3 water outlets.
cycle interruptions due to poor water quality, we recommend the NitraDem Direct Depending on the type, up to three devices can be connected.
Connect 2 water treatment system.

Water treatment

EN 1717
Category 5
Water pressure: max. 10 bar, min. 2 bar
EN 1717 for direct
highest water temperature: 40 °C
Category 5 connection
for direct
connection

NitraDem filter

NitraDem Direct Connect 2

Category 5 conform
for direct connection A B C
i
Manual filling
EN 1717
Category 5
for direct
connection
or

Demineralized water
WARNING!
Could have differing
­water quality!

NitraDem filter

DAC Premium DAC Universal / Water pistol or


DAC Universal S DAC Universal / DAC Universal S

Direct connection: A water filter (20 μm) must be installed upstream of DAC Universal / REF. 66 93 233 NitraDem Direct Connect 2
i DAC Universal S to ensure that the water is clean. This serves to protect your DAC Universal / NitraDem Direct Connect 2 has three water outlets, from which demineralized water is available at a pressure
DAC Universal S. REF. 63 12 214 of 5-6 bar.

At outlet A a infection control system can be optionally connected with or without a water suction
A pump. The system is not recommended for connection to cleaning and disinfection devices.

Water outlets B and C are used to connect infection control systems without a water suction pump,
B C e.g. water pistol, DAC Universal or DAC Universal S.
30 I 31

Instrument reprocessing in the hygiene area


The infection control area should consist of separate areas which must be designated for
the reprocessing of instruments for semi-critical and critical ­applications. These reprocessing
areas must be differentiated into the areas “Dirty”, “Clean” and “Storage.” It is recommended that
these three areas are marked accordingly. DAC Universal / DAC Universal S must be
positioned in the unclean area, directly on the border to the clean area.
32 I 33

Technical data
DAC Universal S

Installation prerequisites

Electrical power supply ~ 100-127 VAC / 200-240 VAC 50 / 60 Hz


Power consumption 1.3 kW
Compressed air connection Input pressure: 5.0 - 8.0 bar; max.
Short-term air consumption: approx. 60 NI / min. at 5 bar
Oil Can capacity 0.2 l
Water tank capacity 2.3 l
Water consumption DAC Universal: approx. 800 ml per cycle (Program Blue Lid)
DAC Universal S: approx. 900 ml. per cycle (Program Pink Lid)
Water quality < 3 μS / cm
Height open / closed (with Blue / Pink Lid) 59 cm / 40 cm x 40 cm x 42 cm
x width x depth
Capacity up to 6 instruments
Weight 26 kg
Minimum distance from wall / ceiling 10 cm  /  70 cm

Installation kit

Designation Description / models REF.


“Siphon” installation kit (Touch) Including: 67 09 880
• Siphon with direct connection
• Manometer
• Hose
• Waste water filter (6 pcs.)
• NitraClean cleaning tablets (50 pcs.)
“Waste water tank” installation kit Including: 66 98 299
(Touch) • Waste water tank
• Manometer
• Hose
• Waste water filter (6 pcs.)
• NitraClean cleaning tablets (50 pcs.)

Documentation

RS 232 connections e. g. printer, documentation software, data logger

LAN e. g. documentation software

Programs

Program Pink Lid Cleaning, sterilization and lubrication of straight and contra-angle
handpieces, turbines and contra-angle heads 134 °C, 3 min.
sterilization, entire cycle time: approx. 21 min1 including cooling
Program White Lid Cleaning and sterilization of ultrasonic / sonic handpieces and tips,
nozzles of multifunctional syringes and powder jet devices as well
as powder jet handpieces 134 °C, 3 min. sterilization, entire cycle
time: approx. 21 min1 including cooling

1
The cycle times can vary.
38 I 39

Ordering information and accessories


DAC Universal S Pink Lid White Lid1

Adapter for straight and Adapter for ultrasonic / sonic


Product REF. REF. handpieces, tips and
contra-angle handpieces REF.
multi-functional syringe
Pink Lid: Check & Clean kit: 67 42 899 Dentsply Sirona TE/ 66 86 682 attachments
• Pink Lid • Check & Clean Cap (EU) Classic Adapter Touch
• 1 bottle of NitramOil • Check & Clean Lid Adapter for ultrasonic / sonic handpieces:
67 47 534
lubrication concentrate •N itraClean cleaning tablets ISO/INTRAmatic® adapter 60 51 648 Sirona SiroSonic TL / 65 36 135
(GB)
• Water filter • Syringe PerioSonic adapter
• Hose • Cotton rolls
• Combination filter •S crewdriver for waste KaVo and Bien-Air contra- 60 51 663 Sirona SiroSonic / 65 36 143
• Power cable water filter angle handpiece adapter L adapter
• Screwdriver for adapter •W aste water filter
• USB stick DAC Universal S
• Indicator holder EMS Piezon straight 66 13 538
• 1 paper of chemical indicators (30 pcs.) Adapter for turbines REF. ­handpiece adapter

“Siphon” installation kit (Touch): 67 09 880 Satelec Slim 66 23 438


• Siphon with direct connection Sirona quick coupling 60 51 697
• Manometer R/F/B adapter
• Hose
KaVo MULTIflex adapter 60 51 655 Satelec Newtron LED 66 23 446
• Waste water filter (6 pcs.)
• NitraClean cleaning tablets (50 pcs.)
“Waste water tank” installation kit (Touch): 66 98 299 Satelec Newtron 66 23 420
• Waste water tank W&H Roto quick adapter 60 51 671
• Manometer
• Hose
BienAir UNIFIX adapter 60 51 713 KaVo SONICflex handpiece 67 32 056
• Waste water filter (6 pcs.)
adapter for KaVo SONICflex
• NitraClean cleaning tablets (50 pcs.)
2003 and KaVo SONICflex
2008
NSK PTL adapter 60 51 804
Adapter for multi-functional syringe nozzles:
Accessories REF. Accessories REF.
Sirona Sprayvit nozzle 65 36 150
NSK QDJ adapter 60 51 812 adapter
NitramOil #2 lubrication 62 59 118 Pink Lid, incl. indicator holder, 67 42 907
concentrate (blue, 6 bottles) without adapters
for DAC Universal Borden adapter, 60 51 861 Adapter for ultrasonic / sonic tips:
2 - 3 holes Sirona 65 36 127
White Lid, incl. indicator holder, 67 42 931
without adapters ultrasonic tip adapter:
Castellini CERAMIC 60 51 762 For the instrument tips
NitraClean tablets 66 35 499 FREEDOM adapter SiroSon S/C8/L;
Lid holder 67 09 856 SiroSonic/L; SiroSonic TL;
(pack of 50)
Midwest/ISO 4/5-hole 60 51 853 PerioSonic1
turbine with fixed connection EMS 66 10 708
PCD DAC Universal S 67 42 956 Waste water tank with hose 60 78 526 adapter ultrasonic tip adapter
Siphon with direct connection 61 26 341 Morita Alpha adapter 60 51 911
Satelec 66 10 716
Waste water filter (6 pcs.) 66 98 166 DAC Universal thermal printer 60 51 770 ultrasonic tip adapter
printer paper 65 99 018 Morita CP4 adapter 60 51 929
KaVo 67 35 646
SONICflex 2003  tip adapter
Indicator holder 67 43 624
Osada OFJ adapter 60 85 745
Adapter for nozzles of powder jet devices:

Chemical indicators 67 42 857 Yoshida QUICK JOINT 63 23 831 EMS AIR-FLOW®2 66 23 461
(510 pcs., for DAC Universal S) adapter

EMS AIR-FLOW® Handy 66 23 453


Check & Clean Lid 67 09 997

1
 he White Lid is not intended for the reprocessing of implant,
T
Check & Clean Cap 67 10 003 endo and CEM tips.
2
Currently not available
Subject to technical changes and errors in the text, Order No. A91100-M45-C197-01-7600, Printed in Germany, Dispo No. 4605, OEW17 WS 0319 V0