INTRODUCTION AND TECHNICAL DATA 1

Philips Healthcare
INSTALLATION 2

BuckyDiagnost VS SETTING-TO-WORK 3

9890 010 8364.


PERFORMANCE CHECKS 4

Level 1 Documentation
HANDOVER PREPARATION 5

FAULT FINDING 6

REPLACEMENT 7

MECHANICAL DRAWINGS 8

DRAWINGS/DIAGRAMS VS BASIC 9

DRAWINGS/DIAGRAMS VS ADVANCED 10
This document and the information contained in it is strictly re‐
served for current Philips Healthcare ("Philips") personnel, Philips
licensed representatives and Philips customers who have pur‐
chased a valid service agreement for use by the customer's desig‐
nated in-house service employee on equipment located at the cus‐
tomer's designated site. Use of this document by unauthorized per‐
sons is strictly prohibited. This document must be returned to
Philips when the user is no longer licensed and in any event upon
Philips' first written request. DRAWINGS/DIAGRAMS VS DIGITAL 11
Printed in Hamburg
Federal Republic of Germany
©2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED

Manual Order No. 4512 984 28037 REV A DRAWINGS/DIAGRAMS VS REPLACE‐


MENT FOR VE/VT 12
BuckyDiagnost VS

Proprietary Notice:

This document and the information contained in it is proprietary and confidential information
of Philips Healthcare (”Philips”) and may not be reproduced, copied in whole or in part, adapt‐
ed, modified, disclosed to others, or disseminated without the prior written permission of the
Philips Legal Department. Use of this document and the information contained in it is strictly
reserved for current Philips personnel and Philips customers who have a current and valid
license from Philips for use by the customer’s designated in-house service employee on equip‐
ment located at the customer’s designated site. Use of this document by unauthorized per‐
sons is strictly prohibited. Report violation of these requirements to the Philips Legal Depart‐
ment. This document must be returned to Philips when the user is no longer licensed and in
any event upon Philips’ first written request.

Warranty Disclaimer Language:

Philips provides this DOCUMENT without warranty of any kind, implied or expressed, includ‐
ing, but not limited to, the implied warranties of merchantability and fitness for a particular pur‐
pose.

Limitations of Liability Language:

Philips has taken care to ensure the accuracy of this document. However, Philips assumes
no liability for errors or omissions and reserves the right to make changes without further no‐
tice to any products herein to improve reliability, function, or design. Philips may make im‐
provements or changes in the product(s) or program(s) described in this document at any time.

2 CSIP Level 1 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
LOPAD BuckyDiagnost VS

SERVICE MANUAL - Subsystem

BuckyDiagnost VS

Type no: 9890 010 8364.

Manual Order No. 4512 984


28036 released: 06/2019
List of Pages and Drawings

1 ...4

Z-1.5 04.1 A4 4512-982-0180.


1 - 1 ... 1 - 9 (09.0)
9 - 1 ... 9 - 2 (09.0)

2 - 1 ... 2 - 70 (09.0) Z1-1 09.0 A3 BuckyDiagnost VS


2Z-10 b/08.0 A3 4512 983 11021
Z2-9 06.0 A4 4512 983 11361
3 - 1 ... 3 - 9 (09.0)
Z1-2 07.0 A3 BuckyDiagnost VS
Z1-3 06.0 A3 BuckyDiagnost VS
4 - 1 ... 4 - 2 (09.0) Z1-4 06.0 A4 BuckyDiagnost VS
Z1-5 07.0 A3 BuckyDiagnost VS
Z2-1 06.0 A3 BuckyDiagnost VS
5 - 1 ... 5 - 2 (09.0) Z2-2 06.0 A4 BuckyDiagnost VS
Z2-5 06.0 A4 BuckyDiagnost VS
Z2-6 06.0 A4 BuckyDiagnost VS
6 - 1 ... 6 - 10 (09.0)
Z3-1.1 06.0 A4 BuckyDiagnost VS
Z3-1.2 06.0 A4 BuckyDiagnost VS
7 - 1 ... 7 - 93 (09.0)
10 - 1 ... 10 - 2 (09.0)

8 - 1 ... 8 - 2 (09.0) Z1-1 09.0 A3 BuckyDiagnost VS


2Z-10 b/08.0 A3 4512 983 11021
Z-1.1.1 05.1 A4 4512-982-0180. Z2-9 06.0 A4 4512 983 11361
Z-1.2.1 05.1 A4 4512-982-0180. Z1-10 07.0 A3 BuckyDiagnost VS
Z-1.3.1 05.1 A3 4512-982-0180. Z1-11 07.0 A3 BuckyDiagnost VS
Z-1.4.1 05.1 A4 4512-982-0180. Z1-12 06.0 A3 BuckyDiagnost VS
Z-1.4.3 05.1 A3 4512-982-0180. Z1-13 06.0 A4 BuckyDiagnost VS
Z-1.4.5 05.1 A4 4512-982-0180. Z1-4 06.0 A4 BuckyDiagnost VS
Z-1.4.7 05.1 A3 4512-982-0180. Z1-14 06.0 A4 BuckyDiagnost VS
Z-4.1 05.1 A4 4512-982-0180. Z2-1 06.0 A3 BuckyDiagnost VS
Z-4.3.1 05.0 A3 4512-982-0180. Z2-3 06.0 A4 BuckyDiagnost VS
Z-4.3.2 05.1 A3 4512-982-0180. Z2-4 06.0 A4 BuckyDiagnost VS
Z-4.4 05.0 A4 4512 982 0180. Z2-5.1 06.0 A4 BuckyDiagnost VS
Z-4.6.1 05.0 A1 4512 982 0180. Z2-6.1 06.0 A3 BuckyDiagnost VS
Z-4.6.2 05.1 A0 4512 982 0180. Z2-8 06.0 A3 BuckyDiagnost VS

4512 984 28037 REV A CSIP Level 1 3


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS LOPAD

Z3-2.1 06.1 A4 BuckyDiagnost VS Z2-7 06.0 A3 BuckyDiagnost VS


Z3-2.2 06.1 A3 BuckyDiagnost VS Z3-6 06.0 A4 BuckyDiagnost VS
Z3-3.1 08.0 A4 BuckyDiagnost VS Z3-3.1 08.0 A4 BuckyDiagnost VS
Z3-3.2 08.0 A3 BuckyDiagnost VS Z3-3.2 08.0 A3 BuckyDiagnost VS
Z3-4.1 a/09.2 A3 BuckyDiagnost VS Z3-5.1 a/09.2 A3 BuckyDiagnost VS
Z3-4.2 a/09.0 A3 BuckyDiagnost VS Z3-5.2 a/09.0 A3 BuckyDiagnost VS
Z3-4.3 a/06.0 A3 BuckyDiagnost VS Z3-5.3 a/06.0 A3 BuckyDiagnost VS
Z3-6.1 08.0 A4 BuckyDiagnost VS Z3-6.1 08.0 A4 BuckyDiagnost VS
Z3-6.2 06.2 A3 BuckyDiagnost VS Z3-6.2 06.2 A3 BuckyDiagnost VS
Z3-8.1 06.1 A4 BuckyDiagnost VS Z3-7.1 06.0 A4 BuckyDiagnost VS
Z3-8.2 06.1 A3 BuckyDiagnost VS Z3-7.2 06.0 A3 BuckyDiagnost VS
Z3-8.1 06.1 A4 BuckyDiagnost VS
11 - 1 ... 11 - 2 (09.0 ) Z3-8.2 06.1 A3 BuckyDiagnost VS

Z1-1 09.0 A3 BuckyDiagnost VS 12 - 1 ... 12 - 2 (09.0)


2Z-10 b/08.0 A3 4512 983 11021
Z2-9 06.0 A4 4512 983 11361 Z1-1 09.0 A3 BuckyDiagnost VS
Z1-20 07.0 A3 BuckyDiagnost VS 2Z-10 b/08.0 A3 4512 983 11021
Z1-21 07.1 A3 BuckyDiagnost VS Z2-9 06.0 A4 4512 983 11361
Z1-22 06.0 A3 BuckyDiagnost VS Z1-30 06.0 A3 BuckyDiagnost VS
Z1-23 06.0 A3 BuckyDiagnost VS Z1-31 06.0 A3 BuckyDiagnost VS
Z1-24 08.0 A3 BuckyDiagnost VS Z1-32 06.0 A3 BuckyDiagnost VS
Z1-25 06.0 A4 BuckyDiagnost VS Z2-20 06.0 A4 BuckyDiagnost VS
Z1-26 06.0 A3 BuckyDiagnost VS Z2-21 06.0 A4 BuckyDiagnost VS
Z1-27 07.0 A3 BuckyDiagnost VS Z3-2.1 06.1 A4 BuckyDiagnost VS
Z2-10 06.0 A3 BuckyDiagnost VS Z3-2.3 06.0 A3 BuckyDiagnost VS
Z2-11 09.0 A3 BuckyDiagnost VS Z3-2.4 06.0 A3 BuckyDiagnost VS
Z2-12 06.0 A4 BuckyDiagnost VS Z3-2.5 06.0 A3 BuckyDiagnost VS
Z2-6.1 06.0 A3 BuckyDiagnost VS

4 CSIP Level 1 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
INTRODUCTION AND TECHNICAL DATA BuckyDiagnost VS

INTRODUCTION AND TECHNICAL DATA


Contents
1 Purpose of manual ........................................................................................................... 1 - 2
2 Compatibility ..................................................................................................................... 1 - 3
3 Technical data .................................................................................................................. 1 - 3
3.1 Mechanical data .............................................................................................................................. 1 - 3
3.2 Environmental data ......................................................................................................................... 1 - 4
3.3 Electrical data .................................................................................................................................. 1 - 5
3.4 Tools/Material required .................................................................................................................... 1 - 5
3.5 Remote service data ....................................................................................................................... 1 - 6
3.6 Transport data ................................................................................................................................. 1 - 6
3.7 Pre–installation requirements .......................................................................................................... 1 - 6
3.8 Pre-installation data ......................................................................................................................... 1 - 6
3.8.1 Wall holder fixation .......................................................................................................................... 1 - 6
3.8.2 Column fixation ................................................................................................................................ 1 - 6
3.8.3 Cable outlets/cable length ............................................................................................................... 1 - 7
4 Safety information ............................................................................................................ 1 - 7
5 Compliance information .................................................................................................... 1 - 7
5.1 Remote control ................................................................................................................................ 1 - 7
6 Equipment identification and labeling ............................................................................... 1 - 8
7 Abbreviations and definitions ........................................................................................... 1 - 8
8 Manual history .................................................................................................................. 1 - 9

4512 984 28037 REV A CSIP Level 1 (19.0) 1-1


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INTRODUCTION AND TECHNICAL DATA

1 PURPOSE OF MANUAL
This documentation is valid for the wall stand BuckyDiagnost VS.

VS basic VS advanced VS advanced tilting VS digital (detector) VS digital tilting (detector)


VS digital prepared VS digital prepared tilting
(cassette) (cassette)

(Fig. 1-1)

NOTE The wall stand BuckyDiagnost can be used in the systems BuckyDiagnost,
DigitalDiagnost, Eleva systems and as replacement for existing BuckyDiagnost VE/
VT wall stands.

The wall stand types VS basic, VS advanced (tilting) and VS digital prepared (tilting) are equipped with the
Bucky unit 2 cassette sensing/non-sensing and used in BuckyDiagnost systems.

The wall stand types VS digital (tilting) are equipped with the Bucky unit 2 digital (detector) and used in
DigitalDiagnost systems.

The wall stand types VS basic and VS advanced are equipped with the Bucky unit 2 cassette sensing/non-sens‐
ing and used in Eleva systems or for replacement of existing wall stands BuckyDiagnost VE/VT.

The wall stand VS is characterized by:

– height adjustable stand with manual cassette or digital detector unit


– manual or motorized movement of stand in vertical direction
– vertical tracking (follow-up control)
– optional motorized tilting
– ”move to position” function: motorized movements of detector to predefined positions
– user interfaces on both side of the detector housing for motorized movements, collimation and AMPLIMAT
field override
– wireless remote control unit
– manual or digital bucky unit with oscillating grid mechanism, changeable grid and parking slit for two grids
– automatic exposure control with 5 field measuring chamber

1-2 4512 984 28037 REV A


CSIP Level 1 (19.0)
© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INTRODUCTION AND TECHNICAL DATA BuckyDiagnost VS

2 COMPATIBILITY
The items listed are compatible with the BuckyDiagnost VS. For further information refer to the PMS/Siebel cata‐
logue or instructions for use.

System compatibility:

BuckyDiagnost Ceiling System

BuckyDiagnost Floor System

BuckyDiagnost Vertical Radiography

BuckyDiagnost Casualty Floor/Ceiling System

DigitalDiagnost TH

DigitalDiagnost TH with application SW release ≥ 1.4

DigitalDiagnost VR with application SW release ≥ 1.4

Options
Spacer 300mm 9890 010 8418x
Floor fixing device 9890 010 8421x
Remote control 9890 010 8438x 9890 010 8438x
Bucky unit 2 sensing (CAN) 4512 201 0256x
Bucky unit 2 sensing (non-CAN) 4512 201 0258x
Bucky unit 2 non-sensing (CAN) 4512 201 0257x
Bucky unit 2 non-sensing (non-CAN) 4512 201 0259x
Bucky unit 2 digital 4512 201 0255x
X-Ray grid lines/cm=36 r=12 fo=110 9896 010 6013x
X-Ray grid lines/cm=36 r=12 fo=140 9896 010 6015x
X-Ray grid lines/cm=36 r=8 fo=180 9896 010 6014x
X-Ray grid lines/cm=36 r=8 fo=110 9896 010 6016x
X-Ray grid lines/cm=36 r=8 fo=140 9896 010 6017x

Accessories
Cassette holder standard 9890 010 8425x
Cassette holder UK 9890 010 8426x
Display unit/Display unit attachment parts 4512 201 0208x/4512 201 0284x
Lead apron 9890 010 8420x
Babix holder 9890 010 8422x
Armrest height adjustment, tiltable 9890 010 8424x
VS basic patient stretch grip 9890 010 8423x

3 TECHNICAL DATA

3.1 MECHANICAL DATA


Crate / Box Dimension (mm) Weight (Newton)
BD VS basic column Length Width Height (10 N = 1 kg)
Packed 2220 760 550 1350
Installed see drawing Z-1.1.1, Z-1.1.2 100

4512 984 28037 REV A CSIP Level 1 (19.0) 1-3


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INTRODUCTION AND TECHNICAL DATA

Crate / Box Dimension (mm) Weight (Newton)


BD VS advanced column Length Width Height (10 N = 1 kg)
BD VS digital column
Packed 2220 760 550 1550*/1700**
Installed see drawing Z-1.2.1 ... Z-1.4.8 1200*/1350**

Crate / Box Dimension (mm) Weight (Newton)


BD VS column Length Width Height (10 N = 1 kg)
(Bucky / detector carriage)
Packed 1030 970 600 1000*/900**
Installed see drawing Z-1.1.1; Z-1.2.1; Z-1.4.1, Z-1.4.5 750*/650**

Crate / Box Dimension (mm) Weight (Newton)


BD VS tiltable unit Length Width Height (10 N = 1 kg)
(Bucky / detector carriage)
Packed 1030 970 900 1600*/1500**
Installed see drawing Z-1.3.1; Z-1.4.3, Z-1.4.7 1300*/1200**

Crate / Box Dimension (mm) Weight (Newton)


BD VS digital detector Length Width Height (10 N = 1 kg)
Packed 920 880 600 440
Installed 480 533 45 200

* for BuckyDiagnost VS with Bucky unit 2 sensing/non-sensing

** for DigitalDiagnost with digital detector

NOTE The specified measures are maximum values - they can slightly deviate depending
on the packaging class!
The packaging department reserves the right to change the contents of the packing
units respectively.

3.2 ENVIRONMENTAL DATA


Operation Operation Stock/Transport Stock/
BuckyD. VS BuckyD. VS BuckyD. VS transport
basic/standard digital digital detector
Temperature in °Celsius 0 ...40 18 ... 30 -25 ... 70 -10 ... 55
Temperature in °Celsius / Hour N/A N/A N/A N/A
Humidity in % (non-condensing) 5 ... 95 20 ... 75 10 ... 100 10 ... 95
Gradient in % / hour N/A N/A N/A N/A
Vibrations / Shock range in Hz 5-500 10 - 60 - 5-9
Vibrations / Shock amplitude in mm 0.035 0.035 - 3.5
Vibrations / Shock acceleration in g 0.5 0.5 - 1
Shock acceleration in g < 15 <5 < 25 < 25
Shock pulse duration in msec 6 6 - 6
Air pressure in Hecto-pascal 700 ... 1060 700 ... 1060 500 ... 1060 700 ... 1060

1-4 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INTRODUCTION AND TECHNICAL DATA BuckyDiagnost VS

Acoustic noise level: Standby and operation 55dB or less


Air cooling: see PRD system room requirements
EMC: IEC 601-1-1

3.3 ELECTRICAL DATA


Equipment related:
Nominal voltages: 220/230 VAC
Nominal current: 1.7 A/0.6 A (max./standby)
Nominal frequency: 50 Hz/60 Hz

Mains supply related:


Supply configurations: 230 VAC
Mains resistance: less or equal 1 Ω
Measured current 1.7 A/0.6 A (max / standby)
(inrush, standby, maximum):
Voltage variation: +10%, -15%
Voltage impulse: 1500V peak
Voltage surge: according to IEC60601-1-2, EN 61000-4-(2...8), EN 61000-4-11
Voltage sag: according to IEC60601-1-2, EN 61000-4-(2...8), EN 61000-4-11
Static frequency variation: 50 Hz or 60 Hz+/-1Hz
Dynamic frequency variation: according to IEC60601-1-2, EN 61000-4-(2...8), EN 61000-4-11
Harmonic voltage distortion according to IEC60601-1-2, EN 61000-4-(2...8), EN 61000-4-11
(single, total):
Neutral to ground voltage variation: according to IEC60601-1-2, EN 61000-4-(2...8), EN 61000-4-11
Neutral to ground voltage impulse: according to IEC60601-1-2, EN 61000-4-(2...8), EN 61000-4-11

3.4 TOOLS/MATERIAL REQUIRED


Tools:
Pre–installation
Delivered: N/A
To be ordered: N/A
To be arranged locally: N/A

Installation Delivered:
To be ordered: N/A
To be arranged locally: N/A
- TC 129 Standard tool kit
- Hammer drill
- Spirit level
- Service PC
- Cable pulling tool for conduits
- Fork-lift
Material:
Pre–installation
Delivered :
To be ordered: N/A
To be arranged locally: N/A
Foundation frame or plater for the column, if the floor does not have the
required thickness and strength. See the static calculation in the room
planning.

4512 984 28037 REV A


CSIP Level 1 (19.0) 1-5
© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INTRODUCTION AND TECHNICAL DATA

Installation
Delivered: Wall holder for the fixation of the column at the wall
To be ordered: N/A
To be arranged locally: Mounting bolts for the fixation of the wall holder. See the static calcula‐
tion in the room planning.
Anchor bolts for the fixaton of the column on the floor. See the static cal‐
culation in the room planning.

3.5 REMOTE SERVICE DATA


N/A

3.6 TRANSPORT DATA


Refer to drawing Z-5 “Transport data“.

3.7 PRE–INSTALLATION REQUIREMENTS


It is supposed that the pre-installation includes all work to allow the mechanical and electrical installation of the
shipped subsystem and components at once.

– The cable ducts and cable conduits are installed already and have the required dimensions.
– The doorways must be sufficient to pass through with the BuckyDiagnost VS column and
vertical/tiltable bucky unit.
– The room needs to be clean.

3.8 PRE-INSTALLATION DATA

3.8.1 Wall holder fixation


The wall holder must be fixed with two screws. The wall mounting points have to be suitable for tensile strength
of Fs ≥ 1500 N each screw.

3.8.2 Column fixation


– Fixation of the column with wall holder

The column must be fixed to the floor with three screws. The floor mounting points have to be suitable for ten‐
sile strength of Fs ≥ 500 N each screw.

– Fixation of the column without wall holder (Stand-alone version with floor fixing device)

The column can be mounted to the floor without using a wall holder. In this case a floor fixing device (9890 010
8421x) has to be attached to the column. The two floor fixing devices must be fixed with two screws each, the
floor mounting points have to be suitable for tensile strength of Fs ≥ 3000 N each screw.
The column must be fixed to the floor with three screws additionally. The floor mounting points have to be suita‐
ble for tensile strength of Fs ≥ 3000 N each screw.

NOTE For the fixation of the column at the floor the max. available height between the col‐
umn base plate and the bottom side of the counterweight carriage is 35 mm.

1-6 CSIP Level 1 (19.0)


© 2020 Koninklijke Philips Electronics N.V. 4512 984 28037 REV A
ALL RIGHTS RESERVED
INTRODUCTION AND TECHNICAL DATA BuckyDiagnost VS

counterweight carriage

max. 35mm

column
base plate

(Fig. 1-2)

3.8.3 Cable outlets/cable length

– Wall connection box

The interface cables between the wall stand and other system components (e.g. generator, table) are connect‐
ed to the wall connection box. The cable outlets for flush-mounted cabling are either located in the upper area
or in the lower area of the wall junction box (see mechanical drawings Z-4.2; Z-4.5; Z-7.1).

For surface-mounted cabling the cable duct should end at either the top side or at the bottom side of the wall
connection box. In this case an adequate area of the cover has to be cut off (see mechanical drawing Z-7.1)

The length of the cables measured from the wall connection box is 20 m except two 30 m cables for
BuckyDiagnost VS digital (fiber–optic and communication).

– Stand–alone version of BuckyDiagnost VS

For this version the wall connection box is fixed on the rear side of the column. The cable outlet has to be car‐
ried out in the floor on the rear side of the column.

4 SAFETY INFORMATION
The general legal and factory safety recommendations for this X–ray equipment and the following recommenda‐
tions must be strictly observed!
Start of installation, operation and maintenance work and especially electrical work must only be executed by
trained and authorized persons. This equipment must only be serviced by properly educated service specialists
who have received general and system–specific training as performed by Philips Medical Systems.

WARNING Ionizing radiation!


Every X-ray unit produces ionizing radiation that may be harmful if not controlled prop‐
erly. Therefore, it is recommended to operate the equipment in accordance with the
guidelines set down by the National Council on Radiation Protection.
The system/component must be switched OFF during work.

5 COMPLIANCE INFORMATION

5.1 REMOTE CONTROL


The device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:

4512 984 28037 REV A CSIP Level 1 (19.0) 1-7


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INTRODUCTION AND TECHNICAL DATA

1. The device may not cause harmful interferences.

2. The device needs to accept any interferences received, including interferences that may cause undesired
operation.

For class B digital device or peripheral, the instructions provided to the user shall include the following or similar
statement, placed in a prominent location in the text of the manual.

NOTE The equipment has been tested and found to comply with the limits for a class B digi‐
tal device, pursuant to Part 15 of the FCC Rules. These limits are designed to pro‐
vide reasonable protection against harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instruction, may cause harmful intererence
to radio communications. However, there is no guarantee that interference will not oc‐
cur in a particular installation. If this equipment does cause harmful interference to
radio or televisions reception, which can be determined by turning the equipment off
and on, the user is encouraged to try to correct the interference by one or more of
the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the re‐
ceiver is connected.
Consult the dealer or an experienced radio/TV technician for help.

6 EQUIPMENT IDENTIFICATION AND LABELING


Refer to 2Z-10.

7 ABBREVIATIONS AND DEFINITIONS


Abbreviation Explanation
AGenT Automatic Generator Tool
APR Anatomically Programmed Radiography
BIST Built–In Self Test
BUF Bucky Unit Family
MSB Mini Signal Bus
MTP Move To Position
PCB Pixium Control Board
PMS Philips Medical Systems
POST Power–On Self Test
VS Vertical Stand

1-8 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INTRODUCTION AND TECHNICAL DATA BuckyDiagnost VS

8 MANUAL HISTORY
Date Version Name Reason of changes
01-10-04 V1 REV AA M.Hasche
03-01-05 V1 REV AB
04-12-06 V1 Vers 00 M. Hasche ▪ Level 1 implementation
▪ Replacement implementation
▪ Introduction VS for BuckyDiagnost Systems
12-18-06 V1 Vers 01 ▪ Text update section 1 chapter 1 & 3
▪ Removed X-Scope link in section 4 chapter 2.11
▪ Drawings update
12-01-09 4512 984 28036 Complete manual revision

▪ Update VP1
▪ New front cover support + front cover
▪ Securing the vertical unit/tilting unit at the col‐
umn with Loctite

4512 984 28037 REV A CSIP Level 1 (19.0) 1-9


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

INSTALLATION
Contents
1 Introduction, tools, room check ........................................................................................ 2 - 3
1.1 Overview over BuckyDiagnost VS types ......................................................................................... 2 - 3
2 Items supplied .................................................................................................................. 2 - 5
3 Unpack, transport, positioning .......................................................................................... 2 - 6
3.1 Box dimensions ............................................................................................................................... 2 - 6
3.2 Roadway ......................................................................................................................................... 2 - 6
3.3 Unpacking ....................................................................................................................................... 2 - 6
3.3.1 Unpacking of the column ................................................................................................................. 2 - 6
3.3.2 Unpacking of the vertical unit .......................................................................................................... 2 - 8
3.3.3 Unpacking of the tilting unit ............................................................................................................. 2 - 8
4 Installation of BuckyDiagnost VS ..................................................................................... 2 - 9
4.1 Installation of the column ............................................................................................................... 2 - 10
4.1.1 Installation of the wall–mounted version ........................................................................................ 2 - 10
4.1.2 Installation of the stand–alone version .......................................................................................... 2 - 12
4.2 Installation of the Counterweights ................................................................................................. 2 - 16
4.3 Installation of the vertical unit/tilting unit at the column ................................................................. 2 - 17
4.3.1 Installation of the 300 mm spacer (optional) ................................................................................. 2 - 21
4.4 Change from left-hand operation to right-hand operation for BuckyDiagnost VS Basic ................ 2 - 23
4.5 Change from left-hand operation to right-hand operation for BuckyDiagnost VS advanced ......... 2 - 32
4.6 Change from left-hand operation to right-hand operation for BuckyDiagnost VS digital ............... 2 - 41
4.7 Installation of the wall connection box VP ..................................................................................... 2 - 52
4.7.1 Wall–mounted version ................................................................................................................... 2 - 52
4.7.2 Stand–alone version ...................................................................................................................... 2 - 52
4.7.3 Corrugated hose fixing .................................................................................................................. 2 - 53
4.8 Cable connections for BuckyDiagnost VS .................................................................................... 2 - 54
4.8.1 Cable connections for BuckyDiagnost VS basic ........................................................................... 2 - 55
4.8.2 Cable connections for BuckyDiagnost VS advanced ................................................................... 2 - 56
4.8.3 Cable connections for BuckyDiagnost VS advanced tilt ................................................................ 2 - 57
4.8.4 Cable connections for BuckyDiagnost VS digital/VS digital tilting ................................................. 2 - 58
4.8.5 Cable connections for BuckyDiagnost VS digital prepared/VS digital prepared tilting .................. 2 - 59
4.9 Cable connections for replacement of BuckyD. VE/VT with BuckyD. VS ..................................... 2 - 60
4.9.1 Cable connections for BuckyDiagnost VE sensing replacement with VS basic ............................ 2 - 60
4.9.2 Cable connections for BuckyDiagnost VE non-sensing replacement with VS basic ..................... 2 - 61
4.9.3 Cable connections for BuckyDiagnost VE sensing/tracking replacement .................................... 2 - 62
4.9.4 Cable connections for BuckyDiagnost VT replacement ............................................................... 2 - 63
5 Installation of the digital detector .................................................................................... 2 - 63

4512 984 28037 REV A CSIP Level 1 (19.0) 2-1


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

6 Installation of accessories .............................................................................................. 2 - 69


6.1 Installation of the radiation protection lead apron for BuckyDiagnost VS ...................................... 2 - 69
6.2 Installation of babix holder for BuckyDiagnost VS ........................................................................ 2 - 69
6.3 Installation of the display unit (only for BuckyDiagnost VS digital/Tilting) .................................... 2 - 69
6.4 Installation of the patient stretch grip (only for BuckyDiagnost VS Basic) ..................................... 2 - 69
6.5 Installation of the floor fixing device for BuckyDiagnost VS .......................................................... 2 - 69
6.6 Installation of the cassette holder / cassette holder UK for BuckyDiagnost VS ............................ 2 - 69
7 Final work ....................................................................................................................... 2 - 69

2-2 CSIP Level 1 (19.0)


© 2020 Koninklijke Philips Electronics N.V. 4512 984 28037 REV A
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

1 INTRODUCTION, TOOLS, ROOM CHECK


Before starting the installation check that all tools and materials that are listed in section 1 are available and the
pre-installation room requirements are fulfilled. Refer to the planned room layout.

1.1 OVERVIEW OVER BUCKYDIAGNOST VS TYPES

VS basic VS advanced VS advanced tilting VS digital (detector) VS digital tilting (detector)


VS digital prepared VS digital prepared tilting
(cassette) (cassette)

(Fig. 2-1)

Short technical description VS types :

BuckyDiagnost VS basic
– Manual height movement of vertical frame with bucky carriage
– Bucky unit sensing / non-sensing

BuckyDiagnost VS advanced
– Manual height movement or optional motorized height movement of vertical frame with bucky carriage
– Bucky unit sensing / non-sensing
– User interface keypads
– Tracking as option (height potentiometer)

BuckyDiagnost VS advanced tilting


– Manual height movement of vertical frame with bucky carriage
– Additional motorized tilt movement of vertical frame with bucky carriage
– Bucky unit sensing / non-sensing
– User interface keypad
– Tracking as option (height potentiometer)

4512 984 28037 REV A CSIP Level 1 (19.0) 2-3


© 2020 Koninklijke Philips Electronics N.V. ALL
RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

BuckyDiagnost VS digital prepared


– Motorized height movement of vertical frame with bucky carriage
– Bucky unit sensing / non-sensing
– User interface keypad
– Tracking (height potentiometer)
– Prepared for later upgrade to digital detector

BuckyDiagnost VS digital prepared tilting


– Motorized height movement of vertical frame with bucky carriage
– Additional motorized tilt movement of vertical frame with bucky carriage
– Bucky unit sensing / non-sensing
– User interface keypad
– Tracking (height potentiometer)
– Prepared for later upgrade to digital detector

BuckyDiagnost VS digital
– Motorized height movement of vertical frame with bucky carriage
– Bucky unit 2 digital including digital detector
– User interface keypad
– Tracking (height potentiometer)

BuckyDiagnost VS digital tilting


– Motorized height movement of vertical frame with bucky carriage
– Additional motorized tilt movement of vertical frame with bucky carriage
– Bucky unit 2 digital including digital detector
– User interface keypad
– Tracking (height potentiometer)

The labels with 12NC are positioned as follows:

Wall connection box


(label inside wall connection box)

Bucky unit 2 digital


(only for VS digital with detector)

Tilting unit
(only with option tilt)

Vertical frame
Bucky unit 2 (label includes type of wall stand)
(sensing / non-sensing)

Column

(Fig. 2-2)

2-4 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

VS type Wall connection Column VB Vertical unit VA Bucky unit


box VP Tilting unit VT
BD VS basic Wall box basic Basic Basic Bucky unit 2 non-sens‐
45121332903x 45122010270x 45122010275x ing
(including wall box) 4512 201 0257x
45121333061x Bucky unit 2 sensing
(excluding wall box) (CAN)
4512 201 02561
Bucky unit 2 sensing
(non CAN)
4512 201 0258x
BD VS advanced Wall box tracking Manual Vertical tracking Bucky unit 2 non-sens‐
45121332897x 45122010271x 45122010276x ing
Manual with track‐ (including wall box) 4512 201 0257x
ing 45121333010x Bucky unit 2 sensing
45122010272x (excluding wall box) CAN
4512 201 02561
Bucky unit 2 sensing
(non CAN)
4512 201 0258x
BD VS advanced tilt‐ Wall box tilting Manual Tilt tracking Bucky unit 2 non-sens‐
ing 4512133898x 45122010271x 4512 201 0280x ing
Manual with track‐ (including wall box) 4512 201 0257x
ing 45121333012x Bucky unit 2 sensing
45122010272x (excluding wall box) (CAN)
4512 201 0256x
Bucky unit 2 sensing
(non CAN)
4512 201 0258x
BD VS digital pre‐ Wall box digital Motorized Vertical motorized Bucky unit 2 non-sens‐
pared 45122010273x 4512 201 0278x ing
45121332899x
(cassette) (including wall box) 4512 201 0257x
45121333011x Bucky unit 2 digital
BD VS digital (excluding wall box) 4512 201 0366x
(detector) Bucky unit 2 sensing
CAN
4512 201 0256x
BDVS digital pre‐ Wall box digital Motorized Tilting motorized Bucky unit 2 non-sens‐
pared tilting (cas‐ 45121332899x 45122010273x 4512 201 0282x ing
sette) (including wall box) 4512201 02557x
45121333013x Bucky unit 2 digital
BD VS digital tilting (excluding wall box) 4512 201 0366x
(detector) Bucky unit 2 sensing
CAN
4512 201 0256x

2 ITEMS SUPPLIED
– Column with:
– 16 x counterweights small
– 16 x counterweights large
– 4 x linings 1 mm
– 4 x linings 2 mm

4512 984 28037 REV A CSIP Level 1 (19.0) 2-5


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

– 2 x Allen screws M8x40 for fixation of the vertical/tilting unit


– 4 x caps for transport locking
– Vertical/tilting unit with:
– Bucky unit 2
– 6 x counterweights small (only for BD VS with digital detector)
– Wall connection box VP
– Installation cables
– Options

3 UNPACK, TRANSPORT, POSITIONING

3.1 BOX DIMENSIONS


Refer to section 1 Introduction, chapter 4.1 “Mechanical data“.

3.2 ROADWAY
Refer to drawing Z-5 “Transport data”.

3.3 UNPACKING
The following procedure should be carried out by two engineers.

3.3.1 Unpacking of the column

Unpacking
0.5 h 0.5 h

▪ Remove the screws (1) on both sides of the trans‐


port box.
▪ Dismantle the box.

(Fig. 2-3)

2-6 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Remove the counterweights (1).

BuckyDiagnost Counterweights
2
VS
basic 18 x small (1 x with clamp)
advanced 12 x small (1 x with clamp)
12 x large (1x with clamp) 1 1
digital 16 x small (1 x with clamp)
16 x large (1 x with clamp)
4
▪ Remove the top cover (2).
3
▪ Remove the front cover which is located below
the column (3).
▪ Remove accessories bag (4).
▪ Remove the two screws of the transport lock (5).

5
(Fig. 2-4)

NOTE The wall fixation angle (1) shall remain at the column. Do not throw it away!

▪ Remove nut and washer (2).

(Fig. 2-5)

4512 984 28037 REV A CSIP Level 1 (19.0) 2-7


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

Contents of the accessories bag:

– 4 x linings 1 mm
– 4 x linings 2 mm
– 4 x blind caps
– 2 x screws M8 for vertical unit/tilting unit fixation.

(Fig. 2-6)

3.3.2 Unpacking of the vertical unit


▪ Dismantle the box.
▪ Remove the accessories bag containing the fixing
screws for the wall connection box (1).
▪ Remove the wall connection box (2).
▪ Remove the cables from the pallet (3).
▪ Remove the transport fixation (4) (only for
BuckyDiagnost VS advanced).

(Fig. 2-7)

3.3.3 Unpacking of the tilting unit


▪ Dismantle the box.
▪ Remove the cover of the wall connection box (1).
2
▪ Remove the screws (3) and take out the transport
lock (2) on both sides.
▪ Remove the transport fixation screws (4) from the
wall connection box.

3
1

(Fig. 2-8) 2

2-8 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Remove the counterweights 2 x 6 small/large (1).

(Fig. 2-9) 2

▪ Remove the plank of the pallet (1).

(Fig. 2-10)

4 INSTALLATION OF BUCKYDIAGNOST VS 1

Installation of VS
2h 2h

4512 984 28037 REV A CSIP Level 1 (19.0) 2-9


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

4.1 INSTALLATION OF THE COLUMN


WARNING The counterweight carriage
needs to stay secured with the
four safety screws with red wash‐
ers (1) until the vertical/tilting unit
and the counterweights are instal‐
led!

1
1

(Fig. 2-11)

4.1.1 Installation of the wall–mounted version


▪ Mark the positions of the fixing points according
to the template Z-4.6.1. wall
For the correct installation place refer to the re‐
spective planned room layout.
▪ Drill three holes depending on the fixing material
in the middle of the marked area.

(Fig. 2-12)

2 - 10 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Bring the column into its final position.


▪ Mark the holes on the wall. 120
▪ Take away the column.
▪ Drill the holes for the wall fixing depending on the 2x
fixing material in the middle of the marked area of
the two fixing points.

1990
(Fig. 2-13)

NOTE For the fixing of the column to the floor the max. available height between the col‐
umn base plate and the bottom side of the counterweight carriage is 35 mm.

counterweight carriage

max. 35mm

column
base plate

(Fig. 2-14)

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 11


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

▪ Align the column horizontally and vertically with a


spirit level. Level with linings, if necessary.

WARNING Do not remove the transport safe‐


ty device (1).

▪ Fix the column to the floor and fasten the bolts.

(Fig. 2-15)

▪ Fix the column to the wall with two screws and


washers.

(Fig. 2-16)

4.1.2 Installation of the stand–alone version


CAUTION During installation of the floor fixing device the second engineer has to secure the
upright column against tipping over!

2 - 12 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

Preconditions:

▪ Unpack the floor fixing device.


▪ Mark the positions of the fixing points according
to the template Z-4.6.2.
– For the correct installation place refer to the re‐
spective planned room layout.
▪ Drill holes depending on the fixing material in the
middle of the marked area.
▪ Lift the column and position it at the installation
place.

(Fig. 2-17)

▪ Unscrew (1) (3 x) and remove the cover of the


floor fixing device (2 x).

(Fig. 2-18)

▪ Loosen the torx screws TX40 (1) (3 x) of the guid‐


ing rail.

(Fig. 2-19)

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 13


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

▪ Insert the guiding rail from the top into the slot.
– The opening of the floor fixing device needs
to be outside.
▪ Repeat the procedure with the second floor fixing
device.

(Fig. 2-20)

▪ Fix the torx screws TX40 of the guiding rail (1) (3


x).
▪ Repeat the procedure with the second floor fixing
device.

(Fig. 2-21)

NOTE Before proceeding with further installation steps make sure that the counterweight car‐
riage is secured with the four safety screws!

▪ Bring the column into its final position.


▪ Align the column horizontally and vertically with a
spirit level. Level with linings, if necessary.
▪ Fix the column (1) and the floor fixing device (2)
(2 x) to the floor and fasten the bolts.

2
1
(Fig. 2-22)

2 - 14 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Install the cover of the floor fixing device with


three screws (1) (2 x).

(Fig. 2-23)

▪ Replace the wall holder (1) by the holder plate for


the wall box (2) at the column. 1 2

(Fig. 2-24)

▪ Install the cable duct (1) with four screws on the


column.
1

(Fig. 2-25)

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 15


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

4.2 INSTALLATION OF THE COUNTERWEIGHTS


▪ Insert the large counterweight plates in the rear
carriage (1). Use the counterweight with clamp as
the top plate in the stack of plates.
▪ Insert the small counterweight plates in the front 1
carriage (2). Use the counterweight with clamp as
2
the top plate in the stack of plates.

Bucky Diagnost VS Counterweights


basic 17 x small plates
advanced 11 x small plates
11 x large plates
advanced tilting 17 x small plates
17 x large plates
digital (with detector) 15 x small plates
15 x large plates
(Fig. 2-26)
digital tilting (with detector) 21 x small plates
21 x large plates
digital prepared 11 x small plates
(with Bucky unit 2 cassette) 11 x large plates
digital tilting prepared 17 x small plates
(with Bucky unit 2 cassette) 17 x large plates

NOTE One counterweight plate remains and can be used for customization of the weight
balance.

▪ Fix the counterweights with the clamps and se‐


cure both clamps with screws (1).
1 1

(Fig. 2-27)

2 - 16 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

4.3 INSTALLATION OF THE VERTICAL UNIT/TILTING UNIT AT THE COLUMN


NOTE If a spacer is part of the delivery, continue with chapter “Installation of the 300 mm
spacer (optional)” before installing the vertical unit/tilting unit.

NOTE The screws fastening the vertical unit/tilting unit to the column have to be secured
with Loctite!

▪ Hang up the vertical unit (1).

(Fig. 2-28)

CAUTION Make sure both fixing screws (2) are properly inserted in the associated lock holes.

▪ Attach the vertical unit with two screws (2) from inside the column.
▪ Secure the screws with Loctite.

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 17


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

Installation of the tilting unit:


▪ Move the pallet and place it in front of the column.
Use a fork lift.
▪ Remove the wall connection box from the pallet
and place it on the floor, next to the column.

(Fig. 2-29)

▪ Secure the tilting unit against falling from the pal‐


let (1).
▪ Pull the bar to the front and lift up the tilting unit.
Push the bar to the rear and let down the tilting
unit carefully on top of the transport lock (2).

(Fig. 2-30)

2 - 18 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Move the pallet towards the column and hang up


the tilting unit at the two fixing points (1). 1
▪ Lower the pallet.

(Fig. 2-31)

CAUTION Make sure both fixing bolts (1) are properly inserted in the associated lock holes.

▪ Attach the tilting/vertical unit with two Allen


screws (1) from inside of the column.
▪ Secure the screws with Loctite.

(Fig. 2-32)

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 19


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

▪ Lift the tilting module and remove the pallet.

(Fig. 2-33)

NOTE Keep back safety screws for later service!

▪ Remove the four counterweight safety screws (1)


from both sides of the column.
▪ Press the four caps delivered with the accesso‐
ries bag into the tapped holes.

1
1

(Fig. 2-34)

2 - 20 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Remove the transport safety device (1).

(Fig. 2-35)

NOTE After installation of the tilting unit it might be necessary to adjust the tilting unit buffer.
Refer to chapter 'Adjustment of the tilting unit buffer' in Setting-to-work.

4.3.1 Installation of the 300 mm spacer (optional)


NOTE For systems with option tomography the spacer is part of the standard delivery.

In addition to tomography the spacer is required in


the following applications:

1. Ceiling suspension operated from the front side Tomo alpha drive
over tilted wall stand.
Operating
2. Patient trolley over tilted wall stand. side

Spacer 300mm

(Fig. 2-36)

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 21


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

NOTE Secure the screws of the spacer with Loctite!

For vertical unit:

▪ Turn the vertical unit carefully on its pallet to have


access to the rear side.
2
▪ Insert the fixing points of the spacer into the ob‐
long holes (1) of the vertical unit rear side.
1

icn-09-v-712-16079-090000-10.iso

(Fig. 2-37)

NOTE Use the screws delivered with


the spacer!

▪ Install the screws (1; 2 in figure above) from in‐


side the spacer and secure them with Loctite.

(Fig. 2-38)

For tilting unit:

▪ Install the spacer at the tilting unit inserting the fix‐


ing points of the spacer into the oblong holes of
the tilting unit (1).
1

icn-09-v-712-16079-090000-11.iso

(Fig. 2-39)

2 - 22 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Insert the screws delivered with the spacer from


inside the spacer.
▪ Secure the screws with Loctite.
1

(Fig. 2-40)

After installing the spacer at the vertical unit/tilting unit, the vertical unit/tilting unit can be installed at the column
as described in chapter 'Installing the vertical unit/tilting unit at the column'.

▪ Insert the fixing points of the column (1) into the


oblong holes of the spacer. 1
▪ Fasten the vertical unit/tilting unit with the two
screws (2) from inside the column.
▪ Secure the screws with Loctite.

2
icn-07-v-712-17007-4668-03.iso

(Fig. 2-41)

4.4 CHANGE FROM LEFT-HAND OPERATION TO RIGHT-HAND OPERATION FOR BUCKYDIAGNOST VS BASIC

Change of the operation side


3,0 h

NOTE Reuse the 5-field Amplimat chamber, the grid and the Bucky unit 2 when changing
the operation side.

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 23


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

WARNING DANGER OF INJURY!


Make sure to insert the four coun‐
terweight safety screws with red
washers to prevent the system
from moving up automatically.

▪ Move the counterweight carriage to its parking po‐


1
sition and fix it with the four safety screws (1), two
on each side of the column. 1
▪ Switch OFF the system.

(Fig. 2-42)

▪ Remove the two screws (1) on the lower side of


the BuckyDiagnost VS.

1
(Fig. 2-43)

▪ Remove the front cover starting from the lower


side of the Bucky unit. The upper side of the cov‐
er is snapped in on top of the Bucky unit.

(Fig. 2-44)

▪ Remove the three screws (1) and remove the


rear cover.

(Fig. 2-45)

2 - 24 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Disconnect the protective earth conductor.

VA X13 (1).

(Fig. 2-46)

▪ Remove the grid.


▪ Remove the two screws (2) and remove the angle 1
plate (1) .

(Fig. 2-47)

▪ Remove the screw (1).


1
▪ Remove the screw (2) and turn the metal bracket
(3) to the left side so that the Amplimat chamber
can be removed.

3 2

(Fig. 2-48)

▪ Take out the Amplimat chamber until there is ac‐


cess to the plug VAVA PH X1.
▪ Remove the cable tie.
▪ Remove the metal clamp (1).
▪ Remove the plastic clamp (2).
▪ Disconnect VAVA PH X1 (3) carefully.
▪ Remove the Amplimat chamber carefully.

(Fig. 2-49)

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 25


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

▪ For right-hand operation:


The position of the dip switch has to be changed
from the right-hand to the left-hand side.
▪ For the left-hand operation:
The position of the dip switch has tobe changed
from the left-hand to the right-hand side.

left-hand operation
right-hand operation

(Fig. 2-50)

▪ Disconnect VAVA X11 (1) and VAVA X1 (2).


1 2
▪ Remove the four screws that connect the Bucky
unit 2 to the vertical frame.
▪ Remove the Bucky unit 2 and turn it upside down
onto a suitable work space.

(Fig. 2-51)

The cassette tray must be modified before it can be used in a right-hand operated system. The cam unit has to
be adapted according to the operation side.

NOTE For Bucky unit 2 non sensing a second cam unit is necessary as the cam units for
right and left-hand operation differ in their design. Order a cam unit with 12 NC
4512-201-04221 for right-hand operation or 12 NC 4512-201-04211 for left-hand op‐
eration.
For Bucky unit 2 sensing no change of the cam units is necessary.

▪ Pull out the cassette tray.


▪ Remove the six screws (1) and remove the metal
plate. 1
▪ Disconnect the protective earth conductor (2).

(Fig. 2-52)

2 - 26 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

NOTE Take care that the Bowden cable


(1) is placed above the cams. left-hand version

▪ Remove the old cam unit and place the new one
into the so far unused opposite seating.
▪ Check the operability of the cam unit. The cams
should be moveable.
▪ Connect the protective earth conductor.
▪ Install the metal plate. Lock the screws with Loc‐ right-hand version
tite blue.
▪ Check the cassette tray for smooth operation.

(Fig. 2-53)

Jumper configuration for Bucky unit 2 sensing CAN4512 201 0256x

▪ Check that jumper W21 is set (BUF controller),


icn-09-v-712-7966-090000-59.iso X15
the other pins remain unused. Set surplus jump‐ W16
rd H41 1
2
ers on only one pin. W17
W18 S1 Service
rd H42

W19 rd H40 gn H44


W20
W21
W22
U20 yw H38 gn H43
W23 S2 Reset gn H39
W24
2 1 X9 1 X12
1 1 1 X13
X11 X8 X7

(Fig. 2-54) Setting of jumper W21

Jumper configuration for Bucky unit 2 non sensing/sensing no CAN4512 201 0257x / 4512 201 0258x

▪ Set the jumper (BUF controller) according to the icn-09-v-712-7966-090000-58.iso


W16 W16
operation side. Set surplus jumpers on only one W17
W18 S1 Service
W17
W18 S1 Service
W19 W19
pin. W20
W21 U20
W20
W21 U20
W22 W22
W23 S2 Reset W23 S2 Reset
W24 W24
2 1 2 1
1 1
X11 X11

(Fig. 2-55) Setting of jumper W19 and W20

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 27


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

For the reconstruction from left-hand to right-hand op‐


eration the following steps have to be carried out at left-hand operation
the vertical module:

▪ Remove the grip (1). 1 2


3
▪ Disconnect the plug (2).
▪ Remove the cover (3).

(Fig. 2-56)

Install the components in the following order:


right-hand operation
▪ Install the grip (1) as shown in the figure.
▪ Connect the plug (2). 4
3 2 1
▪ Install the cover (3).
▪ Put the cables to the left side. They do not have
to be fixed (4).

(Fig. 2-57)

For the reconstruction from right-hand operation to left-hand operation the same steps have to be carried out,
but mirrored (see figures).

2 - 28 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Insert the Bucky unit 2 from the operational side


into the vertical module and push it until there are
ca.10cm
only 10 cm left (see figures).
Make sure not to damage or squeeze the cables.

left-hand operation

ca.10cm

right-hand operation

(Fig. 2-58)

▪ Route the cables through the opening into the


Bucky unit 2.
▪ Push the Bucky unit 2 completely into position
and fasten it with the four screws to the vertical
module.
▪ Check the smooth operation of the cassette tray.
▪ Connect VAVA X11 and VAVA X1.

(Fig. 2-59)

▪ Connect VAVA PH X1 (3).


▪ Install the metal clamp (1).
▪ Install the plastic clamp (2).

NOTE Take care that all wires are


placed under the black housing
of the Amplimat chamber!

(Fig. 2-60)

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 29


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

▪ Push the Amplimat chamber completely into the


3
Bucky unit.
▪ Reinstall the metal bracket (1) and fasten the
screws (2) and (3).

1 2

(Fig. 2-61)

▪ Reinstall the angle plate (1) and fasten the two


screws (2). 1

(Fig. 2-62)

▪ Connect the protective earth conductor VA


X13 (1).

(Fig. 2-63)

▪ Reinstall the rear cover and fasten the three


screws (1).

(Fig. 2-64)

2 - 30 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Position the front cover on the frame (1).


1

icn-09-v-712-16079-090000-07.iso

(Fig. 2-65)

▪ Reinstall the front cover starting from the upper


side of the Bucky unit. The upper side of the cov‐
er is snapped in on top of the Bucky unit.

(Fig. 2-66)

▪ Fasten the two screws (1) on the lower side of the


BuckyDiagnost VS.

1
(Fig. 2-67)

▪ Check whether the cassette tray works properly. When Bucky unit 2 sensing is used, check the format recog‐
nition of the cassette tray.
▪ Remove and insert the grid and check whether the grid moves freely.

NOTE Do not forget to program the new right-hand version in X-Scope on the bucky control‐
ler.
In systems with AWS (Aquisition Work Spot) via FSF (Field Service Framework).

▪ Check the operation of the Bucky unit:


– Run a test exposure.

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 31


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

4.5 CHANGE FROM LEFT-HAND OPERATION TO RIGHT-HAND OPERATION FOR BUCKYDIAGNOST VS AD‐
VANCED

Change of the operation side


3,0 h

NOTE Reuse the 5-field Amplimat chamber, the grid and the Bucky unit 2 when changing
the operation side.

WARNING DANGER OF INJURY!


Make sure to insert the four coun‐
terweight safety screws with red
washers to prevent the system
from moving up automatically.

▪ Move the counterweight carriage to its parking po‐


1
sition and fix it with the four safety screws (1), two
on each side of the column. 1
▪ Switch OFF the system.

(Fig. 2-68)

▪ Remove the two screws (1) on the lower side of


the BuckyDiagnost VS.

(Fig. 2-69)

2 - 32 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Remove the front cover starting from the lower


side of the Bucky unit. The upper side of the cov‐
er is snapped in on top of the Bucky unit.

(Fig. 2-70)

▪ Remove the three screws (1) and remove the 1


3
rear cover (2) .
▪ Disconnect the protective earth conductor
VA X12 (3).

(Fig. 2-71)

▪ Remove the grid.


▪ Remove the two screws (2) and remove the angle 1
plate (1).

(Fig. 2-72)

▪ Remove the screw (1).


1
▪ Remove the screw (2) and turn the metal bracket
(3) to the left side so that the Amplimat chamber
can be removed.

3 2

(Fig. 2-73)

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 33


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

▪ Take out the Amplimat chamber until there is ac‐


cess to the plug VAVA PH X1.
▪ Remove the cable ties.
▪ Remove the metal clamp (1).
▪ Remove the plastic clamp (2).
▪ Disconnect VAVA PH X1 (3) carefully.
▪ Push the Amplimat chamber into its original posi‐
tion and turn the metal bracket in a way that the
Amplimat chamber cannot move.
(Fig. 2-74)

▪ Check that the dip switch of the Amplimat cham‐


ber is in the right-hand position (left-hand opera‐
tion) for 5-field operation.

left-hand operation

(Fig. 2-75)

▪ Disconnect VAVA X11 (1) and VAVA X1 (2).


1 2

(Fig. 2-76)

▪ Remove the four screws that connect the Bucky


unit 2 to the vertical frame.
▪ Remove the Bucky unit 2.

(Fig. 2-77)

2 - 34 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

For Bucky unit 2 non-sensing the cassette tray has to be modified when it is used in a right-hand operated sys‐
tem. The cam unit has to be adapted according to the operation side.

NOTE For Bucky unit 2 non-sensing a second cam unit is necessary as the cam units for
right and left-hand operation differ in their design. Order a cam unit with 12 NC
4512-201-04221 for right-hand operation or 12 NC 4512-201-04211 for left-hand op‐
eration.
For Bucky unit 2 sensing no change of the cam units is necessary.

▪ Pull out the cassette tray.


▪ Remove the six screws (1) and remove the metal
plate. 1
▪ Disconnect the protective earth conductor (2).

(Fig. 2-78)

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 35


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

NOTE Take care that the Bowden cable


(1) is not placed above the cams. left-hand version

▪ Remove the old cam unit and place the new one
into the opposite seating that has not yet been
used.
▪ Check the operability of the cam unit. The cams
should be moveable.
▪ Connect the protective earth conductor.
right-hand version
▪ Install the metal plate. Lock the screws with Loc‐
tite blue.
▪ Check the cassette tray for smooth operation.

(Fig. 2-79)

Jumper configuration for Bucky unit 2 sensing CAN4512 201 0256x

▪ Check that jumper W21 is set (BUF controller),


icn-09-v-712-7966-090000-59.iso X15
the other pins remain unused. Set surplus jump‐ W16
rd H41 1
2
ers on only one pin. W17
W18 S1 Service
rd H42

W19 rd H40 gn H44


W20
W21
W22
U20 yw H38 gn H43
W23 S2 Reset gn H39
W24
2 1 X9 1 X12
1 1 1 X13
X11 X8 X7

(Fig. 2-80) Setting of jumper W21

Jumper configuration for Bucky unit 2 non sensing/sensing no CAN4512 201 0257x / 4512 201 0258x

▪ Set the jumper (BUF controller) according to the icn-09-v-712-7966-090000-58.iso


W16 W16
operation side. Set surplus jumpers on only one W17
W18 S1 Service
W17
W18 S1 Service
W19 W19
pin. W20
W21 U20
W20
W21 U20
W22 W22
W23 S2 Reset W23 S2 Reset
W24 W24
2 1 2 1
1 1
X11 X11

(Fig. 2-81) Setting of jumper W19 and W20

2 - 36 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Loosen but do not remove the screws of the angu‐


lar plate (1) which carries the snap-in latch and re‐
2
move it through the oblong holes.
▪ Remove the screw and remove the metal plate
(2) that covers the cables. 4
▪ Remove the two screws and remove the flap (3).
▪ Remove the grid stop (4).
3

(Fig. 2-82)

For the reconstruction from left-hand to right-hand operation the following steps must be carried out at the verti‐
cal module:

▪ Remove the cable ties (1).


left-hand operation

(Fig. 2-83)

▪ The cables have to be arranged as shown in the


figure: right-hand operation
– Protective earth conductor VAVA X11 (1),
– VAVA PH X1 (2), 1
– VAVA X1 (3). 2
4
3
▪ Put the cables to the left side.
Fix them with cable ties to the support (4).

(Fig. 2-84)

For the reconstruction from right-hand operation to left-hand operation the same steps have to be carried out,
but mirrored (see figures).

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 37


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

▪ Install the flap (1) without fastening the screws.


Shut the flap and align it to the vertical module.
Then fasten the screws.
▪ Install the angular plate (2) through the oblong 2
holes and fasten it that way that the flap stays
aligned to the vertical module when being shut.
▪ Install the metal cable cover (3) and fasten the
screw. 4 3
▪ Install the grid stop (4). 1
▪ Check the flap assembly for smooth operation.

(Fig. 2-85)

▪ Insert the Bucky unit 2 from the operation side in‐


to the vertical module and push it until there are
only 10 cm left before the final position is
reached. Make sure not to damage or squeeze
the cables.

left-hand operation

right-hand operation

(Fig. 2-86)

2 - 38 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Route the cables through the opening into the


Bucky unit 2.
▪ Push the Bucky unit 2 completely into position
and fasten it with the four screws to the vertical
module.
▪ Check for smooth operation of the cassette tray.
▪ Connect VAVA X11 and VAVA X1.

(Fig. 2-87)

▪ Connect VAPH X1 (3).


▪ Install the metal clamp (1).
▪ Install the plastic clamp (2).

NOTE Take care that all wires are


placed under the black housing
of the Amplimat chamber!

(Fig. 2-88)

▪ Push the Amplimat chamber completely into the


3
Bucky unit.
▪ Reinstall the metal bracket (1) and fasten the
screws (2) and (3).

1 2

(Fig. 2-89)

▪ Reinstall the angle plate (1) and fasten the two


screws (2). 1

(Fig. 2-90)

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 39


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

▪ Connect the protective earth conductor VA 1


3
X12 (3).
▪ Reinstall the rear cover (2) and fasten the three
screws (1).
2

(Fig. 2-91)

▪ Position the front cover on the frame (1).


1

icn-09-v-712-16079-090000-07.iso

(Fig. 2-92)

▪ Reinstall the front cover starting from the upper


side of the Bucky unit. The upper side of the cov‐
er is snapped in on top of the Bucky unit.

(Fig. 2-93)

▪ Fasten the two screws (1) on the lower side of the


BuckyDiagnost VS.

(Fig. 2-94)

2 - 40 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Check for the correct operation of the cassette tray. When Bucky unit 2 sensing is used, check the format
recognition of the cassette tray.
▪ Remove and insert the grid and check whether the grid moves freely.

▪ Set jumper X2 in the wall connection box according to the operation side.

Set jumper X2

1-2 left-hand operation 123

2-3 right-hand operation 1 2 3


123

X2
5 Field control
VP3
4512 133 02102

X35 X34
X31 1 10 1
X5
1
X32
9 GND

9
1 X33
1
6

15
8 16

X7 15 8 9
5

0V GND

(Fig. 2-95)

NOTE Do not forget to program the new right-hand version in X-Scope on the bucky control‐
ler.
In systems with AWS (Aquisition Work Spot) via FSF (Field Service Framework).

– Run a test exposure.

4.6 CHANGE FROM LEFT-HAND OPERATION TO RIGHT-HAND OPERATION FOR BUCKYDIAGNOST VS DIGITAL

Change of the operation side


3,0 h

NOTE Reuse the 5-field Amplimat chamber, the digital detector and the Bucky unit 2 digital
when changing the operation side.

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 41


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

WARNING DANGER OF INJURY!


Make sure to insert the four coun‐
terweight safety screws with red
washers to prevent the system
from moving up automatically.

▪ Move the counterweight carriage to its parking po‐


1
sition and fix it with the four safety screws (1), two
on each side of the column. 1
▪ Switch OFF the system.

(Fig. 2-96)

▪ Remove the two screws (1) on the lower side of


the BuckyDiagnost VS.

(Fig. 2-97)

▪ Remove the front cover starting from the lower


side of the Bucky unit.

(Fig. 2-98)

▪ Remove the three screws (1) and remove the 1


3
rear cover (2).
▪ Disconnect the protective earth conductor
VA X12 (3).

(Fig. 2-99)

2 - 42 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Remove the grid.


▪ Remove the nut (2) and remove the end stop (1)
at the left-hand side of the Bucky unit 2 digital. 3
▪ Disconnect the plug VAVA X1 (3).

(Fig. 2-100)

▪ Pull out the 5-field measuring chamber to get ac‐


cess to the connector VAVA PH X1 (3).
▪ Remove the metal clamp (1) and the plastic
clamp (2).
▪ Disconnect VAVA PH X1 (3).

(Fig. 2-101)

▪ Remove the 5–field measuring chamber (1).

(Fig. 2-102)

▪ Remove the nut (2) and remove the end stop (1)
at the right-hand side of the Bucky unit 2 digital.
3
▪ Unplug the connector of the electromagnet VA X2
(3).

2
1

(Fig. 2-103)

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 43


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

▪ Disconnect the cables at the detector:


1 3
– Earth wire VAVA P X11 (lower bolt) (1).
– Fiber optic link VAVA P B1 (2).
– Communication cable VAVA P X2 (3) with 2
clamp (4).
– Power supply cable VAVA P X1 (5).
5
4

(Fig. 2-104)

▪ Disconnect the cable of the fan X6 (1).


VA PCB X6

1 C6

C5

X1

X2

K1

X3

X5

C8

C9

X4

X66
icn-09-v-712-7966-090000-60.iso

(Fig. 2-105)

CAUTION All cables have to be fixed with tape to protect them from damage when the digital
detector is removed from the Bucky unit 2 digital.

▪ Fix the left-hand ribbon cable and earth wires with 1


some tape (1) to the left-hand side of the Bucky 2
unit 2 digital. Position them above the earth bolt
(2) to protect them from damage when the digital
detector is replaced.

(Fig. 2-106)

2 - 44 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Fix the right-hand ribbon cable with some tape (1)


to the right -hand side of the Bucky unit 2 digital.

(Fig. 2-107)

▪ Pull the detector (1) carefully out of the Bucky unit


2 digital.

1
(Fig. 2-108)

▪ Remove the four screws that connect the Bucky


unit 2 digital to the vertical frame.
▪ Remove the Bucky unit 2 digital.

(Fig. 2-109)

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 45


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

▪ Loosen but do not remove the screws of the angu‐


lar plate (1) which carries the snap-in latch and re‐
2
move it through the oblong holes.
▪ Remove the screw and remove the metal plate
(2) that covers the cables. 4
▪ Remove the two screws and remove the flap (3).
▪ Remove the grid stop (4).
3

(Fig. 2-110)

For the reconstruction from left-hand to right-hand operation the following steps have to be carried out at the
vertical module:

▪ Remove the cable ties (1).


left-hand operation

(Fig. 2-111)

▪ The cables have to be arranged as shown in the


figure: right-hand operation
– Protective earth conductor VAVA X11 (1),
– VAVA PH X1 (2), 1
– VAVA X1 (3). 2
4
3
▪ Put the cables to the left side.
Fix them with cable ties to the support (4).

(Fig. 2-112)

For the reconstruction from right-hand operation to left-hand operation the same steps have to be carried out,
but mirrored (see figures).

2 - 46 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Install the flap (1) without fastening the screws.


Shut the flap and align it to the vertical module.
Then fasten the screws.
▪ Install the angular plate (2) through the oblong 2
holes and fasten it in such a way that the flap
stays aligned to the vertical module when being
shut.
▪ Install the metal cable cover (3) and fasten the 4 3
screw. 1
▪ Install the grid stop (4).
▪ Check the flap assembly for smooth operation.

(Fig. 2-113)

▪ Insert the Bucky unit 2 digital from the operation


side in the vertical module and fasten it with the
four screws. Make sure not to damage or
squeeze the cables.

left-hand operation

right-hand operation

(Fig. 2-114)

CAUTION Make sure not to damage the wires and cables in the Bucky unit 2 digital.

▪ Insert the detector carefully in the Bucky unit 2 digital.

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 47


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

▪ Reconnect the cables at the detector:


1 3
– Earth wire VAVA P X11 (lower bolt) (1).
– Fiber optic link VAVA P B1 (2).
– Communication cable VAVA P X2 (3) with 2
clamp (4).
– Power cable VAVA P X1 (5).
5
4

(Fig. 2-115)

▪ Reconnect the cable of the fan X6 (1).


VA PCB X6

1 C6

C5

X1

X2

K1

X3

X5

C8

C9

X4

X66
icn-09-v-712-7966-090000-60.iso

(Fig. 2-116)

▪ Install the end stop (1) on the right-hand side of


the Bucky unit 2 digital and tighten the nut (2).
3
▪ Plug in the connector of the electromagnet VA X2
(3).

2
1

(Fig. 2-117)

CAUTION Ensure that the wires are located under the housing of the 5–field measuring chamber.

2 - 48 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Insert the 5-field measuring chamber carefully.


▪ Connect VAVA PH X1 (3).
▪ Install the metal clamp (1) and the plastic clamp
(2).
▪ Push the 5-field measuring chamber fully into the
Bucky unit 2 digital.

(Fig. 2-118)

▪ Install the end stop (1) at the left-hand side of the


Bucky unit 2 digital and tighten the nut (2).
▪ Reconnect the plug VAVA X1 (3). 3

(Fig. 2-119)

CAUTION Ensure that the cables behind the rear cover do not obstruct the movement of the
grid frame.

▪ Connect the earth wire VA X12 (3). 1


3
▪ Reinstall the rear cover (2) of the Bucky unit 2 dig‐
ital and fasten the three screws (1).

(Fig. 2-120)

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 49


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

▪ Position the front cover on the frame (1).


1

icn-09-v-712-16079-090000-07.iso

(Fig. 2-121)

▪ Reinstall the front cover (1).


▪ Reinstall the two Allen screws (2).

(Fig. 2-122)

Amplimat configuration

▪ Set jumper X2 in the wall connection box according to the operation side.

2 - 50 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

Set jumper X2

1-2 left-hand operation 123

2-3 right-hand operation 123 1 2 3

X2
5 Field control
VP3
4512 133 02102

X35 X34
X31 1 10 1
X5
1
X32
9 GND

9
1 X33
1
6

15
8 16

X7 15 8 9
5

0V GND

(Fig. 2-123)

NOTE Keep back safety screws for lat‐


er service!

▪ Remove the four safety screws (1) from the col‐ 1


umn.
1
▪ Press the four caps into the tapped holes.
▪ Switch ON the system.
▪ Insert the grid.

(Fig. 2-124)

Test:

▪ For detector calibration refer to the respective system manual.


▪ After the change of the operation side you have to do a re-calibration of the detector orientation.

Name Value Description


DDFFrontWall left Definition of the side where the wall stand control panel is
mounted and how the detector is mounted.
This influences the way the image is shown. Image could
be up-side down.
HeadSideTable left Definition of the side where the patient head is placed.
This influences the way the image is shown. Image could
be up-side down.

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 51


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

NOTE If one of the items is changed the Detector Calibration is invalid. ➔ Re-Calibration

NOTE Do not forget to program the new right-hand version in X-Scope on the bucky control‐
ler.
In systems with AWS (Aquisition Work Spot) via FSF (Field Service Framework).

▪ Check the operation of the Bucky unit:


– Check the grid function.
– Run a test exposure.

4.7 INSTALLATION OF THE WALL CONNECTION BOX VP

4.7.1 Wall–mounted version


▪ Mark the position of the fixing points for the wall
connection box according to the template Z-4.2
resp. Z-4.5.
1 1
▪ Drill four holes for the upper and lower screws.
▪ Fix the upper screws (1).
▪ Hang up the wall connection box.
▪ Fix the lower screws (2).

2 VP
Wall box basic

2
VP
Wall box digital

(Fig. 2-125)

4.7.2 Stand–alone version


NOTE The minimal distance between the wall and the installed column must be 500 mm!

2 - 52 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Hang up the wall connection box (1) at the column.


▪ Move the safety plate (2) downwards in order to 2
secure the wall connection box. Fix the two scews.
1

(Fig. 2-126)

4.7.3 Corrugated hose fixing


Column cable connection:

CAUTION The work on the electronic system is allowed to be carried out by authorized person‐
nel only! Before starting the electrical connection check that the unit is voltage free!

▪ Fix the corrugated hose (1) of the column with the


clamp (2).
▪ Use cable-ties to fix the cables (3).

2 3

(Fig. 2-127)

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 53


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

System cable connections:

▪ Fix the system cables with cable-ties (1).


▪ Fix the cable clamp (2) to the wall connection box.

(Fig. 2-128)

4.8 CABLE CONNECTIONS FOR BUCKYDIAGNOST VS


NOTE Identify the wall connection box with the 12NC code No. (1).

– For BuckyDiagnost VS digital the label with the


12NC code No. (1) is placed variably in the wall
connection box.

(Fig. 2-129)

2 - 54 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

4.8.1 Cable connections for BuckyDiagnost VS basic


Wall box basic code No. 4512 133 2903x (not available for order)
▪ Connect VPI X7 (mains 230 V AC).
– Use the cable clamp (1) for screening. VP Wall box basic
for BuckyDiagnost VS basic
▪ Connect VPI X10 (MSB) (only for sensing).
▪ Connect VPI X1 wall stand release (only for non-
sensing).
▪ Connect VP X32 Amplimat chamber.
▪ Connect VPI X3 vertical brake. N2

Earthing:

▪ Connect the earth wire


VP X11 to MEX.
X1

X10

X3

X7
X11

VPI Interface Basic


(Fig. 2-130)

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 55


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

4.8.2 Cable connections for BuckyDiagnost VS advanced


Wall box tracking code No. 4512 133 2897x (not available for order)
▪ Connect VPI X7 (mains 230 V AC).
– Use the cable clamp (1) for screening. VP Wall box tracking
for BuckyDiagnost VS advanced
▪ Connect VPI X1 wall stand release (only for non-
sensing).
▪ Connect VPI X10 (MSB) (only for sensing).
▪ Connect VPI X5 tracking (optional).
▪ Connect VPI X3 vertical brake. N2
▪ Connect VPI X4 height potentiometer (optional with
tracking). X4
▪ Connect VP3 X31 Amplimat chamber.
▪ Connect VP3 X33 Amplimat display. X3
X5

Earthing: X10

▪ Connect the earth wire


VP X11 to MEX. X1

X7
VP3 X11

VPI
Interface Basic / Tracking

X31

X33

VP3
5-field measuring chamber

(Fig. 2-131)

2 - 56 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

4.8.3 Cable connections for BuckyDiagnost VS advanced tilt


Wall box tilting code No. 4512 133 2898x (not available for order)
▪ Connect VPI X7 (mains 230 V AC).
– Use the cable clamp (1) for screening. VP Wall box tilting
▪ Connect VPI X1 wall stand release (only for non- BuckyDiagnost VS advanced tilt
sensing).
▪ Connect VPI X10 (MSB) (only for sensing).
▪ Connect VPI X3 vertical brake.
▪ Connect VPI X4 height potentiometer (optional for
tracking). T1
X3
N5
▪ Connect VP3 X31 5-field measuring chamber.
▪ Connect VP3 X33 5-field measuring display. X4

Earthing:

▪ Connect the earth wire L1


X10
VP X11 to MEX.
X12
X1

VP3

VPI
X11
1
X7
Interface wall

X31

X33

VP3
5-field measuring chamber

(Fig. 2-132)

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 57


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

4.8.4 Cable connections for BuckyDiagnost VS digital/VS digital tilting


Wall box digital code No. 4512 133 2899x (not available for order)
▪ Connect VPI X7 mains 230 V AC.
– Use the cable clamp (1) for screening.
▪ Connect VPI X55 communication.
▪ Connect VPI X10 MSB.
▪ Connect VP1 X2 column motor.
▪ Connect VP1 X6 / VP1 X19 safety switch, column
brake, coupling.
▪ Connect VP1 X8 tracking potentiometer.
▪ Connect VP3 X31 (5–field measuring chamber dis‐
play)
▪ Connect VP3 X33 (5 field measuring chamber).

Earthing:

▪ Connect the earth wire VP X11.

(Fig. 2-133)
2 - 58 CSIP Level 1 (19.0) 4512 984 28037 REV A
© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

4.8.5 Cable connections for BuckyDiagnost VS digital prepared/VS digital prepared tilting
Wall box digital code No. 4512 133 2899x (not available for order)
▪ Connect VPI (VP) X7 mains 230 V AC.
– Use the cable clamp (1) for screening.
▪ Connect VPI VP X10 (MSB / CAN).
▪ Connect VP1 X2 column motor.
▪ Connect VP1 X6 / VP1 X19 safety switch, column
brake, coupling.
▪ Connect VP1 X8 tracking potentiometer.
▪ Connect VP3 X31 (5–field measuring chamber dis‐
play).
▪ Connect VP3 X33 (5 field measuring chamber).

Earthing:

▪ Connect the earth wire VP X11.

(Fig. 2-134)
4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 59
© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

4.9 CABLE CONNECTIONS FOR REPLACEMENT OF BUCKYD. VE/VT WITH BUCKYD. VS

4.9.1 Cable connections for BuckyDiagnost VE sensing replacement with VS basic


Code No. N/A
▪ Connect VPI (VP) X7 (mains 230 V AC).
– Use the cable clamp (1) for screening.
VP Wall box
▪ Connect VPI (VP) X2 sensing. for BuckyDiagnost VE sensing replacement
▪ Connect VPI (VP) X1 release wall stand.
▪ Connect VP3 X32 Amplimat chamber.

Earthing:

▪ Connect the earth wire N2


VP X11 to MEX.

(VP)X2

(VP)X2
(VP)X1

(VP)X7

X11

VPI Interface Basic / Tracking

(Fig. 2-135)

2 - 60 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

4.9.2 Cable connections for BuckyDiagnost VE non-sensing replacement with VS basic


Wall box basic code No. 4512 133 2903x (not available for order)
▪ Connect VPI (VP) X7 (mains 230 V AC).
– Use the cable clamp (1) for screening. VP Wall box basic
for BuckyDiagnost VS basic
▪ Connect VPI (VP) X1 wall stand release.
▪ Connect VP X32 Amplimat chamber.
▪ Connect VPI X3 vertical brake.

Earthing:
N2
▪ Connect the earth wire
VP X11 to MEX.

(VP)X1
X1

X3

(VP)X7

X11

VPI Interface Basic


(Fig. 2-136)

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 61


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

4.9.3 Cable connections for BuckyDiagnost VE sensing/tracking replacement


Wall box tilting code No. 4512 133 2897x (not available for order)
▪ Connect VPI (VP) X7 (mains 230 V AC).
– Use the cable clamp (1) for screening. VP Wall box tilting
for BuckyDiagnost VS advanced replacement
▪ Connect VPI (VP) X1 wall stand release.
▪ Connect VPI (VP) X2 sensing.
▪ Connect VPI (VP) X5 tracking.
▪ Connect VPI X3 vertical brake.
▪ Connect VPI X4 height potentiometer.
▪ Connect VP3 X31 Amplimat chamber.
▪ Connect VP3 X33 Amplimat release. N2
Earthing:
X4
▪ Connect the earth wire VP X11 to MEX.
(VP)X5

(VP)X2

(VP)X1
X3

(VP)X7
VP3 X11

VPI
Interface Basic / Tracking

X31

X33

VP3
5-field measuring chamber

(Fig. 2-137)

2 - 62 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

4.9.4 Cable connections for BuckyDiagnost VT replacement


Code No. N/A
▪ Connect VPI (VP) X7 (mains 230 V AC).
– Use the cable clamp (1) for screening. VP Wall box
for BuckyDiagnost VS advanced tilt
▪ Connect VPI (VP) X1 wall stand release. replacement
▪ Connect VPI (VP) X2 sensing.
▪ Connect VPI (VP) X3 tilt position.
▪ Connect VPI (VP) X5 tracking .
▪ Connect VPI X3 vertical brake.
▪ Connect VPI X4 height potentiometer.
▪ Connect VP3 X31 Amplimat chamber. N2
▪ Connect VP3 X33 Amplimat release.

Earthing: X4

▪ Connect the earth wire VP X11 to MEX. (VP)X5

(VP)X2
(VP)X3

(VP)X1
X3

(VP)X7
VP3 X11

VPI
Interface Basic Tilting

X31

X33

VP3
5-field measuring chamber

(Fig. 2-138)

5 INSTALLATION OF THE DIGITAL DETECTOR

Installation of the digital


detector
0,5 h 0,5 h

NOTE Note
Do not open the inner plastic bag before the detector temperature has reached room
temperature . The acclimatization time may be up to 24 hours. It may be shortened if
the still sealed inner plastic bag containing the detector is removed from the two card‐
board boxes.

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 63


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

NOTE For the BuckyDiagnost VS with tilting option the installation of the detector can be
carried out with the Bucky unit in horizontal position.
For the BuckyDiagnost VS without tilting option the installation has to be carried out
in vertical position.

▪ Tilt the bucky module to 90° position by pushing the tilt button on the user interface (only possible if a tilting
module is installed).

CAUTION Make sure that the system is se‐


cured with the four counter‐
weight safety screws (1)!

1
1

(Fig. 2-139)

▪ Switch OFF the system at ENF1. 1


3
▪ Remove the rear cover (2) of the Bucky unit 2 dig‐
ital (three screws (1) .
▪ Disconnect the earth wire
VA X12 (3). 2

(Fig. 2-140)

▪ Remove the end stop (1) from the right side of the
Bucky unit 2 digital by loosening the nut (2).
3
▪ Unplug the connector of the electromagnet
VA X2 (3).

2
1

(Fig. 2-141)

2 - 64 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Remove the end stop (1) from the left side of the
Bucky unit 2 digital by removing the nut (2).
▪ Disconnect the plug VAVA X1 (3). 3

(Fig. 2-142)

NOTE Remove the nut of the end stop completely!


If not, it is possible that the nut could be inscrewed and pushed in while inserting the
detector.

▪ Pull out the 5–field measuring chamber to get ac‐


cess to the connector VAVA PH X1 (3).
▪ Remove the metal clamp (1) and the plastic
clamp (2).
▪ Disconnect VAVA PH X1 (3).
▪ Remove the 5-field measuring chamber.

(Fig. 2-143)

NOTE All cables have to be fixed with tape to protect them against damage during insertion
of the digital detector!

▪ Fix the ribbon cable with tape (1) on the right side
of the Bucky unit 2 digital.

(Fig. 2-144)

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 65


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

▪ Fix the cables and earth wires with tape (1) on 1


the left side of the Bucky unit 2 digital. Position 2
them above the earth bolt (2) to protect them
against damage.

(Fig. 2-145)

▪ Install both rubber guiding pieces (1) at the bot‐


tom side of the detector.
▪ Apply a uniform layer of the talcum powder to the
rubber guiding pieces and to the rubber hold-
downs inside the Bucky unit 2 digital.
1

(Fig. 2-146)

CAUTION Make sure not to damage the wires and cables inside the Bucky unit 2 digital.

▪ Push the detector (1) carefully into the Bucky


unit 2 digital.

1
(Fig. 2-147)

CAUTION Ensure that the wires are located under the housing of the 5–field measuring chamber.

2 - 66 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ Insert the 5-field measuring chamber carefully.


▪ Connect VAVA PH X1 (3).
▪ Install the metal clamp (1) and the plastic clamp
(2).
▪ Push the 5-field measuring chamber fully into the
Bucky unit 2 digital.

(Fig. 2-148)

▪ Set the sliding switch SAPH S1 to the correct po‐


sition.

Position 1 → user side left and right S1 Settings Chamber Layout


all 5 fields
(all 5 fields are active) and
field 1 - 2 - 3
only field
4-2-5
(head left) (head right)
default
settings

icn-08-v-712-17007-0000-01.iso

(Fig. 2-149) Position 1 - all 5 fields are active

▪ Connect the cables at the detector:


1 3
– Earth wire VAVA P X11 (lower bolt) (1).
– Fiber optic link VAVA P B1 (2).
– Communication cable VAVA P X2 (3) with 2
clamp (4).
– Power cable VAVA P X1 (5).
5
4

(Fig. 2-150)

▪ Install the end stop (1) on the left side of the


Bucky unit 2 digital and tighten the nut (2).
▪ Reconnect the plug VAVA X1 (3). 3

(Fig. 2-151)

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 67


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

▪ Install the end stop (1) on the right side of the


Bucky unit 2 digital.
▪ Reconnect VA X2 (2). 1
2

(Fig. 2-152)

CAUTION Ensure that the cables behind the rear cover do not obstruct the movement of the
grid frame.

▪ Connect the earth wire VA X12 (3). 1


3
▪ Install the rear cover (2) of the Bucky unit 2 digital
(three screws (1).

(Fig. 2-153)

NOTE Keep back safety screws for later service!

▪ Remove the four counterweight safety screws (1).


▪ Press the four caps delivered with the accesso‐
ries bag into the tapped holes.
▪ Switch ON the system at ENF1.

1
1

(Fig. 2-154)

2 - 68 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
INSTALLATION BuckyDiagnost VS

▪ For detector calibration refer to the respective system manual.

6 INSTALLATION OF ACCESSORIES

6.1 INSTALLATION OF THE RADIATION PROTECTION LEAD APRON FOR BUCKYDIAGNOST VS


Refer to the unit manual, Manual Code No. 4512 988 00491

6.2 INSTALLATION OF BABIX HOLDER FOR BUCKYDIAGNOST VS


Refer to the unit manual, Manual Code No. 4512 988 00741

6.3 INSTALLATION OF THE DISPLAY UNIT (ONLY FOR BUCKYDIAGNOST VS DIGITAL/TILTING)


Refer to the unit manual, Manual Code No. 4512 988 00501

6.4 INSTALLATION OF THE PATIENT STRETCH GRIP (ONLY FOR BUCKYDIAGNOST VS BASIC)
Refer to the unit manual, Manual Code No. 4512 988 02451

6.5 INSTALLATION OF THE FLOOR FIXING DEVICE FOR BUCKYDIAGNOST VS


Refer to the unit manual, Manual Code No. 4512 984 29751

6.6 INSTALLATION OF THE CASSETTE HOLDER / CASSETTE HOLDER UK FOR BUCKYDIAGNOST VS


Refer to the unit manual, Manual Code No. 4512 984 30011

7 FINAL WORK
Insertion of the column front cover:

▪ Switch ON the power and move the vertical/tilting


unit to its lowest position.
▪ Insert the front cover of the column from the top.
▪ Fix the front cover with the screws (1).
▪ Switch OFF the power.

(Fig. 2-155)

4512 984 28037 REV A CSIP Level 1 (19.0) 2 - 69


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS INSTALLATION

Insertion of the top cover of the column:

▪ Insert the top cover of the column.


▪ Turn out the grub screw (1) to secure the top cover.

(Fig. 2-156)

Insertion of the wall connection box cover:

▪ Hold the cover sloped and insert it at the top (1) 1 1


of the wall connection box.
▪ Attach the cover at the bottom of the wall connec‐
tion box. 2
▪ Fix the cover with screws (2).
Wall box basic
2

Wall box digital

(Fig. 2-157)

2 - 70 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
SETTING-TO-WORK BuckyDiagnost VS

SETTING-TO-WORK
Contents
1 Functions of the control panel .......................................................................................... 3 - 2
2 Adjustment of the remote control ..................................................................................... 3 - 4
3 Adjustment of the wall stand height .................................................................................. 3 - 4
3.1 Height Adjustment with system controller (for sensing systems with CAN bus) ............................. 3 - 4
3.2 Height Adjustment at Controller VP1 (for non-sensing systems without CAN bus) ........................ 3 - 4
4 Adjustment of the wall stand tilting angle ......................................................................... 3 - 6
4.1 Tilt Adjustment with system controller (for sensing systems with CAN bus) ................................... 3 - 6
4.2 Tilt Adjustment at Controller VT1 (for non-sensing systems without CAN bus) .............................. 3 - 6
5 Adjustment of the tilting unit buffer ................................................................................... 3 - 7
6 HARDWARE ..................................................................................................................... 3 - 9
7 SOFTWARE ..................................................................................................................... 3 - 9

4512 984 28037 REV A CSIP Level 1 (19.0) 3-1


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS SETTING-TO-WORK

1 FUNCTIONS OF THE CONTROL PANEL


The control panel is part of BuckyDiagnost VS advanced and BuckyDiagnost VS digital and is available in three
versions, equipped as follows:

– BuckyDiagnost VS advanced

Functions No. 2 + 3 + 7
– BuckyDiagnost VS advanced tilting

Functions No. 1 + 2 + 3 + 6 +7
– BuckyDiagnost VS digital tilting / digital prepared tilting

Functions No. 1 + 2 + 3 + 4 + 5 + 6 +7

No. Symbol Description

Auxiliary (AUX) selected T racking

1 (not part of wireless remote control)


1

LED illuminated: tracking switched on


(not part of wireless remote control)
Tracking

Select auxiliary
2

2
Select measuring field group:
Out of 5 measuring fields a group of 3 is selected.
Measuring fields are selected with an APR
program resp.
. measuring field selection on the
generator. The LEDs indicate the position of the
active measuring fields.

3 3
Switch on light beam lamp

Set collimator
4
Memory

4 Collimator
Memory
- select last manual value 5 6
- set maximum size

Toggle between portrait and landscape


(only with digital detector)

(Fig. 3-1)

3-2 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
SETTING-TO-WORK BuckyDiagnost VS

(Fig. 3-2)

4512 984 28037 REV A CSIP Level 1 (19.0) 3-3


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS SETTING-TO-WORK

2 ADJUSTMENT OF THE REMOTE CONTROL


For systems with Bucky Controller functioning as system controller refer to unit manual X-Scope for detailed
information about the adjustment of the remote control.

For systems with a PC based system controller (e.g. Eleva systems) refer to the accordant Field Service Frame‐
work (FSF) for detailed information about the adjustment of the remote control.

NOTE The batteries of the remote control can be empty and must be charged in the charg‐
ing unit for approx. 1 hour. The remote control can be used during charging time.

3 ADJUSTMENT OF THE WALL STAND HEIGHT


NOTE The height adjustment is necessary in case of an incorrectly adjusted height potenti‐
ometer VBR or after replacement of the height potentiometer VBR.

3.1 HEIGHT ADJUSTMENT WITH SYSTEM CONTROLLER (FOR SENSING SYSTEMS WITH CAN BUS)
For systems with Bucky Controller functioning as system controller refer to unit manual X-Scope for detailed
information about the height adjustment of the wall stand BuckyDiagnost VS.

For systems with a PC based system controller (e.g. Eleva systems) refer to the accordant Field Service Frame‐
work (FSF) wall stand height adjustment procedure.

3.2 HEIGHT ADJUSTMENT AT CONTROLLER VP1 (FOR NON-SENSING SYSTEMS WITHOUT CAN BUS)

3-4 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
SETTING-TO-WORK BuckyDiagnost VS

▪ Remove the cover from the wall connection box


to get access to the controller PCB VP1 (1).

(Fig. 3-3)

▪ Switch S2 (Service Mode) to ON.


1 1

▪ Move the vertical unit to the lowest working posi‐


7 9

X22 X19
W10
tion and press service button S3 to store the value.
5 8
1 3 1
1 1 4
1
X21
X6 X2
▪ Move the vertical unit to the highest working posi‐ X23
X16
V51 yellow

tion and press service button S3 to store the value. 1 V49 yellow

▪ Switch S2 (Service Mode) to OFF.


▪ Check the vertical movement function. W13 1

X1
▪ Re-Install the cover. X14
1

1
W6
1

1
W9 W5
X20
1 W4 V53 yellow
1

W12 W3
VP 1 V54 yellow

1 1

W8 W7
1
9

1 X3

W11
15
8

D3 S2 1
1

HIGH
D1 ON OFF

D2 S3
5

X17
9

LOW X13
6

S1
1

ON
V11
V10

V12
V9

X9 X8
OFF
W1 S4
BOOT
CAN0
CAN1
CAN2
CFG0
CFG1
CFG2
CFG3

1
yellow
red

yellow
gree
green

1
1
RESET
n

(Fig. 3-4)

4512 984 28037 REV A CSIP Level 1 (19.0) 3-5


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS SETTING-TO-WORK

4 ADJUSTMENT OF THE WALL STAND TILTING ANGLE


NOTE The height adjustment is necessary in case of an incorrectly adjusted tilt potentiome‐
ter VTR or after replacement of the tilt potentiometer VTR.

4.1 TILT ADJUSTMENT WITH SYSTEM CONTROLLER (FOR SENSING SYSTEMS WITH CAN BUS)
For systems with Bucky Controller functioning as system controller refer to unit manual X-Scope for detailed
information about the tilt adjustment of the wall stand BuckyDiagnost VS

For systems with a PC based system controller (e.g. Eleva systems) refer to the accordant Field Service Frame‐
work (FSF) wall stand tilt adjustment procedure.

4.2 TILT ADJUSTMENT AT CONTROLLER VT1 (FOR NON-SENSING SYSTEMS WITHOUT CAN BUS)

▪ Remove the bottom cover (1) with 2.5 mm Allen


key to get access to the tilt controller PCB VT1 (2).

1 1
1
7 1
9 3
1
1

on
1
1 1 X19
X1
1
1

1
5 1

on 8 4
Reset 15
6 9 1 9
1 5 8 1

(Fig. 3-5)

3-6 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
SETTING-TO-WORK BuckyDiagnost VS

▪ Switch S2 (Service Mode) to ON.


1 1

▪ Move the tilting unit to the 0° position using a spi‐


7 9

X22 X19
W10
rit level for verification.
5 8
1 3 1
1 1 4
1
X21
X6 X2
▪ Press service button S3 to store the value. X23
X16
V51 yellow

V49
▪ Move the tilting unit to the 90° position using a spi‐
yellow
1

rit level for verification.


▪ Press service button S3 to store the value. W13 1

X1
▪ Switch S2 (Service Mode) to OFF. X14
1

▪ Check the tilting function. 1


W6

1
1

▪ Re-install the cover. W9 W5


X20
1 W4 V53 yellow

1
W12 W3
VT 1 V54 yellow

1 1

W8 W7

1
9
1 X3

W11

15
8
D3 S2 1
1
HIGH
D1 ON OFF

D2 S3

5
X17

9
LOW X13

6
S1

1
ON
V11
V10

V12
V9

X9 X8
OFF
W1 S4
BOOT
CAN0
CAN1
CAN2
CFG0
CFG1
CFG2
CFG3

1
yellow
red

yellow
gree
green

1
1
RESET
n

(Fig. 3-6)

5 ADJUSTMENT OF THE TILTING UNIT BUFFER


Items supplied during installation:

– 3 x washer (2522 600 86032)

Adjustment procedure:

▪ Move the Bucky module into the 90° position.


▪ Check if the Bucky module buffer rests solidly on the arm so that the Bucky module does not move.
▪ Check if the Bucky module is horizontal using a spirit level.

If the Bucky module is moving or is not horizontal, the buffer at the tilting unit rear side can be adjusted. If an
adjustment is necessary carry out the following steps:

4512 984 28037 REV A CSIP Level 1 (19.0) 3-7


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS SETTING-TO-WORK

▪ Remove the upper rear side cover (2) of the


Bucky module by removing the four screws (1).

0911-24-01

(Fig. 3-7)

NOTE The buffer has been secured with Loctite 222 by the supplier. Set the pliers as close
as possible to the threaded bolt when removing the buffer. This avoids damage to
the buffer.

▪ Remove the buffer (1) at the Bucky module rear


side.
1

0911-24-01

(Fig. 3-8)

▪ Insert the necessary number of washers behind the buffer.


▪ Install the buffer.
▪ Move the Bucky module into the 90° position.
▪ Check if the Bucky module is horizontal using a spirit level.

▪ Repeat this procedure until the Bucky module is leveled out in the 90° position.
▪ Clean the buffer and apply Loctite to the threaded bolt.
▪ Install the buffer.

3-8 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
SETTING-TO-WORK BuckyDiagnost VS

▪ Reinstall the upper rear side cover (2) fastening


the four screws (1).

0911-24-01

(Fig. 3-9)

6 HARDWARE
For jumper settings and DIP switch settings see schematic diagrams.

7 SOFTWARE
For detailed information about programming of the wall stand BuckyDiagnost VS refer to manual X–Scope for
systems with Bucky controller.

NOTE The Bucky controller must be reset to activate the settings after the modification.

4512 984 28037 REV A CSIP Level 1 (19.0) 3-9


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
PERFORMANCE CHECKS BuckyDiagnost VS

PERFORMANCE CHECKS
Contents
1 Acceptance ....................................................................................................................... 4 - 2
2 Performance check .......................................................................................................... 4 - 2

4512 984 28037 REV A CSIP Level 1 (19.0) 4-1


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS PERFORMANCE CHECKS

1 ACCEPTANCE
N/A

2 PERFORMANCE CHECK
▪ Switch ON power after finishing programming and adjustment.
▪ Check all movements and functions of the wall stand.

4-2 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
HANDOVER PREPARATION BuckyDiagnost VS

HANDOVER PREPARATION
Contents
1 Handover preparation ....................................................................................................... 5 - 2

4512 984 28037 REV A CSIP Level 1 (19.0) 5-1


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS HANDOVER PREPARATION

1 HANDOVER PREPARATION
▪ Remove all dirt and grease residues with alcohol.

5-2 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
FAULT FINDING BuckyDiagnost VS

FAULT FINDING
N/A

4512 984 28037 REV A CSIP Level 1 (19.0) 6-1


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

REPLACEMENT
Contents
1 Replacement procedures at the column ........................................................................... 7 - 2
1.1 Replacement of the vertical drive assembly ................................................................................... 7 - 2
1.2 Replacement of the vertical brake .................................................................................................. 7 - 5
1.3 Replacement of the microswitch with roller follower ........................................................................ 7 - 9
1.4 Replacement of the potentiometer 10k for vertical movement ...................................................... 7 - 11
1.5 Replacement of the chain assembly for VS advanced, VS digital ................................................ 7 - 13
1.6 Replacement of the set of springs for torque-switch .................................................................... 7 - 17
1.7 Replacement of the set of castors for the vertical carriage .......................................................... 7 - 21
2 Replacement procedures at the vertical/tilting unit ......................................................... 7 - 26
2.1 Replacement of the tilting unit tracking or tilting unit motorized ................................................... 7 - 26
2.2 Replacement of the vertical unit tracking or vertical unit motorized ............................................. 7 - 33
2.3 Replacement of the front cover ..................................................................................................... 7 - 37
2.4 Replacement of the complete Bucky unit 2 .................................................................................. 7 - 38
2.5 Replacement of the complete Bucky unit 2 digital ......................................................................... 7 - 44
2.6 Replacement of the digital detector ............................................................................................... 7 - 54
2.7 Replacement of the 5-field measuring chamber ............................................................................ 7 - 59
2.8 Replacement of the vertical unit safety cover assembly ............................................................... 7 - 62
2.9 Replacement of the set of actuators for safety switches for vertical unit ....................................... 7 - 63
2.10 Replacement of the microswitch with roller follower ...................................................................... 7 - 67
2.11 Replacement of the front cover support ........................................................................................ 7 - 70
2.12 Replacement of the set of actuators for safety switches front cover ............................................. 7 - 71
2.13 Replacement of the potentiometer 10k for tilt movement .............................................................. 7 - 74
2.14 Replacement of the PCB VT1 ....................................................................................................... 7 - 77
2.15 Replacement of the flap assembly for grid parking ....................................................................... 7 - 78
2.16 Replacement of the tilt axis drive motor ........................................................................................ 7 - 79
2.17 Replacement of the coupling tilt drive ........................................................................................... 7 - 84
2.18 Replacement of the user interface keypad and cable ................................................................... 7 - 88
2.19 Replacement of the patient hand grip right or left ......................................................................... 7 - 92
3 Replacement procedures at the wall connection box ..................................................... 7 - 93
3.1 Replacement of the PCB VP1 ....................................................................................................... 7 - 93

4512 984 28037 REV A CSIP Level 1 (19.0) 7-1


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

1 REPLACEMENT PROCEDURES AT THE COLUMN

1.1 REPLACEMENT OF THE VERTICAL DRIVE ASSEMBLY


Code no. 4512 133 2808x

Replacement of the
vertical drive assembly
0.5 h

Tools/Material required:

– TC129 Standard tools


– Ladder

Scope of delivery: 5
4
– Vertical drive assembly incl. fixing material:
6
– 1 x Vertical drive assembly (1)
– 1 x Clutch disk (2)
– 1 x Split pin (3)
– 2 x Washer (4)
– 2 x Plastic washer (5)
– 2 x Screws (6)
– 4 x Washer (0.2 mm) for adjusting the clutch (7) 7
1

3 2

(Fig. 7-1)

Replacement procedure:

▪ Remove the top cover.

▪ Position the vertical / tilting module so that there


is free access to the fixing screws of the clutch (1).
▪ Switch OFF the system . 1

(Fig. 7-2)

7-2 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Remove cable ties (1).


▪ Remove connectors from the drive (2) and the
clutch (3). 1 3
▪ Remove the lower connector with pink wire and
the upper connector with blue wire of cable VBS
from the micro switch (4).
2

4
(Fig. 7-3)

▪ Loosen the two allen screws 2.5 mm (1) of the


clutch from the chain axis.
1

(Fig. 7-4)

▪ Remove the drive assembly from the column by


loosening screws M10 (2 x).

(Fig. 7-5)

4512 984 28037 REV A CSIP Level 1 (19.0) 7-3


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Replace the split pin (1).


▪ Replace the clutch disk at the chain axis (2). 2

(Fig. 7-6)

▪ Put the plastic washers on the drive assembly.

(Fig. 7-7)

▪ Attach the drive assembly at the clutch disk.


▪ Fix screws M10 (2 x) handtight.

(Fig. 7-8)

7-4 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

NOTE Ensure that the drive assembly and the clutch are straight aligned and use the feeler
gauge for determing the distance between drive assembly and clutch.

▪ Insert washers for adjusting the distance between


clutch and drive assembly (1), pull the clutch disk 2
to the drive assembly and fix the grub screws (2).
1
▪ Fix screws M10 (2 x) (3).

(Fig. 7-9)

▪ Connect all switches and fix cable ties vice versa.


▪ Check the function of the drive assembly.
▪ Reinstall the top cover.

1.2 REPLACEMENT OF THE VERTICAL BRAKE


Code no. 4512 133 3015x

Replacement of the
vertical brake
1.5 h

Tools/Material required:

– TC129 Standard tools


– Ladder

Scope of delivery:

▪ 1 x vertical brake
▪ 6 x screw
▪ 4 x adjusting washer

Replacement procedure:

4512 984 28037 REV A CSIP Level 1 (19.0) 7-5


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Move the counterweight carriage into its parking


position and fix it with the four safety screws (1),
two on each side of the column.
▪ Remove the top cover of the column.

1
1

(Fig. 7-10)

▪ Remove the two grub screws at the axis using an


1.5 mm Allen key .
1

(Fig. 7-11)

▪ Loosen the screws (1) using a Torx screwdriver


T15.
2
▪ Remove the holder plate (2) with potentiometer.

(Fig. 7-12)

7-6 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Remove the nut at the axis using a 19 mm double


open end wrench.

icn-07-v-712-16079-4424-08.iso

(Fig. 7-13)

▪ Remove the metal plate at the new vertical brake


by loosening the two screws.

icn-07-v-712-16079-4424-09.iso

(Fig. 7-14)

NOTE The brake must be released in order to remove the brake disk. Release the brake
either by pressing the brake release button at the user interface or supplying a volt‐
age of 26 volts at connector X1.

4512 984 28037 REV A CSIP Level 1 (19.0) 7-7


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Fasten the metal plate to the old brake disk.


▪ Mount the screw (1) in the middle of the axis clock‐
wise.
▪ Remove the brake disk using the metal plate as a
handle.

(Fig. 7-15)

▪ Remove the five screws (1) at the rear side of the


magnetic unit of the vertical brake.
▪ Install the magnetic unit of the new vertical brake.
Fasten the five screws.

WARNING Danger of injury!


Take care not to squeeze your
fingers between the magnetic
unit and the brake disk.

(Fig. 7-16)

▪ Install the new brake disk. Press the brake re‐


lease button and add the brake disk to the magnet.
▪ Reinstall the nut at the axis.
▪ When the brake is released check the air gap be‐
tween the magnet and the new brake disk. One of
the adjusting washers has to be insertable, two ad‐
justing washers have to be too thick for the air gap.
▪ Remove the splint pin to determine the number of
adjusting washers. Start with the number already
at the axis.
▪ Reinstall the potentiometer in exactly the same po‐
sition as before. After installing the potentiometer
fasten the grub screws at the axis.
(Fig. 7-17)

NOTE Check the function of the potentiometer by testing the function of tracking. If tracking
does not work properly, adjust the tracking function according to the Field Service
Framework (FSF).

7-8 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Reinstall the top cover of the column.

NOTE Keep back safety screws for later service!

▪ Remove the four saftey screws.

1.3 REPLACEMENT OF THE MICROSWITCH WITH ROLLER FOLLOWER


Code no. 4512 133 2949x

Replacement of the
microswitch with roller follower
0.5 h

Tools/Material required:

– TC129 Standard tools


– Ladder

Scope of delivery:

– 1 x Microswitch with roller follower

(Fig. 7-18)

Replacement procedure:

▪ Remove the top cover of the column.

▪ Remove connectors from the microswitch (1).


▪ Remove the screws (2). 2

(Fig. 7-19)

4512 984 28037 REV A CSIP Level 1 (19.0) 7-9


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Remove the holder plate with the microswitch.

(Fig. 7-20)

NOTE The new microswitch must be released!

▪ Replace the microswitch on the holder plate.

(Fig. 7-21)

▪ Reinstall the holder plate at the vertical drive as‐


sembly.

(Fig. 7-22)

▪ Connect all connectors.


▪ Check the function of the microswitch.
▪ Reinstall the cover.

7 - 10 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

1.4 REPLACEMENT OF THE POTENTIOMETER 10K FOR VERTICAL MOVEMENT


Code no. 4512 133 3000x

Replacement of the potentiometer


10K for vertical movement
1h
(Fig. 7-23)

Tools/Material required:

– TC129 Standard tools


– Ladder

Scope of delivery:

– 1 x Potentiometer 10 kΩ for vertical movement


with cable VBR X4

(Fig. 7-24)

Replacement procedure:

▪ Move the vertical/tilting module to its highest position.


▪ Remove the top cover from the column.

▪ Disconnect the connector (1).


▪ Remove both grub screws (2) with 1.5 mm allen 1
2
key.

(Fig. 7-25)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 11


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Remove both torx screws TX20 (1).


▪ Remove the holder plate (2) with potentiometer.
2

(Fig. 7-26)

▪ Remove the potentiometer (1) from the holder


plate (2). 1

(Fig. 7-27)

▪ Turn the potentiometer counter-clockwise to stop


position (1), then approx. 90° back (2) to avoid
1
damage at the potentiometer during setting-to-
work procedure.

2
90
°

(Fig. 7-28)

7 - 12 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Insert the potentiometer into the brake.


▪ Fix the holder plate with torx screws (1). 2 3
▪ Fix the grub screws (2).
▪ Connect VBRX4 (3).

1
(Fig. 7-29)

Setting-to-work procedure:

▪ Ensure that the resistance between pin 1 and 3 is


9.5 kΩ.
▪ Ensure that the voltage is approx. 4.85 V.
▪ Continue with chapter 3. Setting-to-work.
▪ After the adjustment of the wall stand height
check the vertical movement function of the wall
stand.
▪ Reinstall the cover.

(Fig. 7-30)

1.5 REPLACEMENT OF THE CHAIN ASSEMBLY FOR VS ADVANCED, VS DIGITAL


Code no. 4512 133 3002x

Replacement of the
chain assembly for VS adv., digital
1.5 h

Tools/Material required:

– TC129 Standard tools


– Ladder
– Transport locking (four safety screws)
– Parking support for vertical/tilting module

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 13


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

Scope of delivery:

– 1 x Chain incl. fixing material

(Fig. 7-31)

NOTE The system has to be switched on during replacement procedure so that the brake
works!

▪ Remove the top cover and the front cover from the column.

▪ Remove the clamps from the upper counter‐


weights (1).
▪ Remove counterweights (2) to get access to the
chain fixation screws. 2 1

(Fig. 7-32)

7 - 14 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Move the counterweight carriage to its parking po‐


sition and fix it with the four safety screws (1), two
on each side of the column.

1
1

(Fig. 7-33)

▪ Position a parking support below (1) the vertical/


tilting module.
– The chain has to be released while the coun‐
terweight carriage is secured in its parking po‐
sition.

(Fig. 7-34)

▪ Remove the chain fixation screws (1) from the


counterweight carriage.

(Fig. 7-35)

NOTE The old bushing has to be re-used!

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 15


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Remove the chain fixation screw (1) from the car‐


riage of the vertical/tilting module.
– Re-use the old bushing.

(Fig. 7-36)

▪ Remove the old chain.


▪ Fit the new chain in the sprocket.

(Fig. 7-37)

▪ Fix the chain at the carriage of the vertical/tilting


module (1).
– Use the old bushing.

(Fig. 7-38)

7 - 16 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Fix the chain at the counterweight carriage (1).

(Fig. 7-39)

NOTE Keep back safety screws for later service!

▪ Remove the four safety screws.


▪ Press the four caps delivered with the accessories bag into the tapped holes.
▪ Move the counterweight carriage to working position and insert the counterweights.

▪ Fix the counterweights with the clamps and se‐


cure both clamps with screws (1).
1
1

(Fig. 7-40)

▪ Check the function.


▪ Install the column front cover and the top cover.

1.6 REPLACEMENT OF THE SET OF SPRINGS FOR TORQUE-SWITCH


Code no. 4512 133 3004x

Replacement of the set of springs


for torque-switch
0.5 h

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 17


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

Tools/Material required:

– TC129 Standard tools


– Ladder

Scope of delivery:

– 2 x Spring for torque-switch incl. fixing material

(Fig. 7-41)

Replacement procedure:

▪ Remove the top cover from the column.


▪ Disconnect the connectors from the microswitch.

▪ Remove both torx screws TX20.

(Fig. 7-42)

▪ Unscrew the distance block (1).


▪ Loosen the holder (2).

2
1

(Fig. 7-43)

7 - 18 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Remove the holder from the gear.

(Fig. 7-44)

▪ Remove the four torx screws (1) and replace the


springs.

(Fig. 7-45)

▪ Attach the holder at the gear.

(Fig. 7-46)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 19


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Fix the screw (1) from the rear side.

(Fig. 7-47)

▪ Attach the distance block (1).

(Fig. 7-48)

▪ Mount both torx screws TX20.

(Fig. 7-49)

▪ Connect all cables and connectors.


▪ Check the function.
▪ Reinstall the top cover.

7 - 20 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

1.7 REPLACEMENT OF THE SET OF CASTORS FOR THE VERTICAL CARRIAGE


Code no. 4512 133 3006x

Replacement of the set of castors


for the vertical carriage
1.5 h

Tools/Material required:

– TC129 Standard tools


– Ladder

Scope of delivery:

– 1 x Set of castors incl. fixing material

(Fig. 7-50)

Replacement procedure:

▪ Remove the top cover and the front cover from the column.

▪ Move the counterweight carriage to its parking po‐


sition and fix it with the four safety screws with
red washers (1), two of them on each side of the
column.

1
1

(Fig. 7-51)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 21


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Remove the vertical/tilting module from its car‐


riage.
– Remove the screws (1) (2x).
– Detach the module and park it on a support (2).

1 2
(Fig. 7-52)

▪ Remove the chain from the carriage (1).

(Fig. 7-53)

▪ Remove the screws (2x) from the head of the col‐


umn.

(Fig. 7-54)

7 - 22 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Lift the head of the column and position it.

(Fig. 7-55)

▪ Remove the carriage (1).

(Fig. 7-56)

▪ Replace all castors at the carriage.

(Fig. 7-57)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 23


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

NOTE Clean the track before re-inserting the carriage.

▪ Re-insert the carriage.


▪ Adjust the castors.

(Fig. 7-58)

▪ Reinstall the head of the column.

(Fig. 7-59)

7 - 24 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Mount the chain at the carriage (1).

(Fig. 7-60)

▪ Tighten the castors.


▪ Adjust the castors.

(Fig. 7-61)

▪ Hang up the tilting module on the fixation points


(1).
▪ Fasten the tilting module with the original screws
(2) from inside of the vertical carriage. 1
▪ Secure the screws with Loctite.

2
(Fig. 7-62)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 25


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

NOTE Keep back safety screws for lat‐


er service!

▪ Remove the four safety screws (1). 1


▪ Press the four caps delivered with the accesso‐ 1
ries bag into the tapped holes.

(Fig. 7-63)

▪ Check the function.


▪ Reinstall the top cover and the front cover of the column.

2 REPLACEMENT PROCEDURES AT THE VERTICAL/TILTING UNIT

2.1 REPLACEMENT OF THE TILTING UNIT TRACKING OR TILTING UNIT MOTORIZED


Code no. 4512 133 3012x or 4512 133 3013x

Replacement of the tilting unit


tracking or tilting unit motorized
1.0 h

NOTE Before disconnecting any cable, check the labeling, so the cable can be reconnected
in the same position.

CAUTION Danger of injury!


If the rear covers are removed never put your hands or fingers inside until the move‐
ment of the counterweight carriage is blocked.
The four safety screws have to be inserted!

Tools/Material required:

– TC129 Standard tools


– Transport locking (four safety screws with red washers)

Scope of delivery:

– 1 x tilting unit tracking or tilting unit motorized

7 - 26 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

NOTE When replacing the tilting unit tracking or tilting unit motorized, reuse the front cover,
the 5-field Amplimat chamber and the Bucky unit 2.

WARNING Danger of injury!


Make sure to insert the four coun‐
terweight safety screws to pre‐
vent the system from moving up
automatically.

▪ Move the counterweight carriage into its parking


1
position and fix it with the four safety screws (1),
two on each side of the column. 1
▪ Switch off the system at ENF1.

(Fig. 7-64)

Replacement procedure: 1
3
Remove the front cover. Refer to chapter Replace‐
ment of the front cover.

▪ Remove the rear cover (2) loosening the three


screws (1). 2
▪ Disconnect the protective earth conductor VA
X12 (3).
▪ Remove the grid.

(Fig. 7-65)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 27


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Remove the angle plate (1) loosening the two


screws (2).

icn-09-v-712-16079-090000-09.iso

(Fig. 7-66)

Removal of the 5-field Amplimat chamber

Remove the 5-field Amplimat chamber. Refer to chapter Replacement of the 5-field Amplimat chamber.

Removal of the Bucky unit 2

Remove the Bucky unit 2. Refer to chapter Replacement of the Bucky unit 2.

▪ Remove the cable connections at the wall connec‐


tion box VP.
▪ Loosen the two Allen screws (1) from inside of
the column.

(Fig. 7-67)

7 - 28 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V. ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Remove the vertical module by uncoupling the


unit at the two fixation points (1). 1

(Fig. 7-68)

▪ Move the pallet in front of the column. Use a fork-


lift.

(Fig. 7-69)

▪ Secure tilting unit from falling from the pallet (1).


2
▪ Pull the bar to the front and lift up the tilting unit. 1
Push the bar to the front and let down the tilting
unit carefully on top of the transport lock (2).

(Fig. 7-70)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 29


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Move the pallet towards the column and hang up


the tilting unit on the two fixation points (1). 1
▪ Lower the pallet.

(Fig. 7-71)

CAUTION Make sure both fixation bolts (1) are properly inserted in the associated lock holes.

▪ Fix the tilting unit with two Allen screws (1) from
inside of the column.
▪ Secure the screws with Loctite.

(Fig. 7-72)

7 - 30 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Insert the Bucky unit 2 from the operation side


(grid insertion side) into the vertical module. Make
sure not to damage or squeeze the cables.

left-hand operation

right-hand operation

(Fig. 7-73)

▪ Put the cables through the opening into the Bucky


unit 2.
▪ Push the Bucky unit 2 completely into position
and fasten it with the four screws to the vertical
module.
▪ Check for smooth operation of the cassette tray.
▪ Connect VAVA X11 and VAVA X1.

(Fig. 7-74)

▪ Connect VAVA PH X1(3).


▪ Install the metal clamp (1) and the plastic clamp
(2).

NOTE Take care that all wires are


placed under the black housing
of the Amplimat chamber!

(Fig. 7-75)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 31


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Push the Amplimat chamber completely into the


3
Bucky unit.
▪ Reinstall the metal bracket (1) by fastening the
screws (2) and (3).

1 2

(Fig. 7-76)

▪ Install the angle plate (1) by fastening the two


screws (2).

icn-09-v-712-16079-090000-09.iso

(Fig. 7-77)

▪ Connect the protective earth conductor VA 1


3
X12 (3).
▪ Reinstall the rear cover (2) by fastening the three
screws (1).
2

(Fig. 7-78)

For the installation of the front cover see chapter Replacement of the front cover and follow the instructions in
reverse order.

7 - 32 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

NOTE After replacement of the tilting unit it might be necessary to adjust the tilting unit buf‐
fer. Refer to chapter 'Adjustment of the tilting unit buffer' in Setting-to-work.

2.2 REPLACEMENT OF THE VERTICAL UNIT TRACKING OR VERTICAL UNIT MOTORIZED


Code no. 4512 133 3010x or 4512 133 3011x

Replacement of the vertical unit


tracking or vertical unit motorized
1h

Tools/Material required:

– TC129 Standard tools


– Transport locking (four safety screws with red washers)

Scope of delivery:

– 1 x vertical unit tracking or vertical unit motorized

NOTE When replacing the tilting unit tracking or tilting unit motorized, reuse the front cover,
the 5-field Amplimat chamber and the Bucky unit 2.

WARNING Danger of injury!


Make sure to insert the four coun‐
terweight safety screws to pre‐
vent the system from moving up
automatically.

▪ Move the counterweight carriage into its parking


1
position and fix it with the four safety screws (1),
two on each side of the column. 1
▪ Switch OFF the system at ENF1.

(Fig. 7-79)

Replacement procedure:

Removal of the front cover

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 33


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Remove the front cover (see chapter Replace‐ 1


3
ment of the front cover, page 7-37).

▪ Remove the rear cover (2) loosening the three


screws (1).
▪ Disconnect the protective earth conductor VA 2
X12 (3).

(Fig. 7-80)

▪ Remove the grid.


▪ Remove the angle plate (1) loosening the two
screws (2).
2

icn-09-v-712-16079-090000-09.iso

(Fig. 7-81)

Removal of the 5-field Amplimat chamber

▪ Remove the 5-field Amplimat chamber (see chapter Replacement of the 5-field measuring chamber, page 7-
59).

Removal of the Bucky unit 2

▪ Remove the Bucky unit 2 (see chapter Replacement of the complete Bucky unit 2 , page 7-38).

7 - 34 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Remove the cable connections at the wall connec‐


tion box VP.
▪ Loosen the two Allen screws (2) from inside of
the column.
▪ Remove the vertical unit by uncoupling the unit at
the two fixation points (1).
1
Installation of the vertical unit:

▪ Hang up the new vertical unit by coupling the unit


at the two fixation points (1).
▪ Fix vertical unit with two screws (2) from inside of
the column.
2
▪ Secure the screws with Loctite.

(Fig. 7-82)

CAUTION Make sure both fixation screws (2) are properly inserted in the associated lock holes.

▪ Connect the cables at the wall connection box.

▪ Insert the Bucky unit 2 from the operation side


(grid insertion side) into the vertical module. Make
sure not to damage or squeeze the cables.

left-hand operation

right-hand operation

(Fig. 7-83)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 35


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Put the cables through the opening into the Bucky


unit 2.
▪ Push the Bucky unit 2 completely into position
and fasten it with the four screws to the vertical
module.
▪ Check for smooth operation of the cassette tray.
▪ Connect VAVA X11 and VAVA X1.

(Fig. 7-84)

▪ Connect VAVA PH X1 (3).


▪ Install the metal clamp (1) and the plastic clamp
(2).

NOTE Take care that all wires are


placed under the black housing
of the Amplimat chamber!

(Fig. 7-85)

▪ Push the Amplimat chamber completely into the


3
Bucky unit 2.
▪ Reinstall the metal bracket (1) by fastening the
screws (2) and (3).

1 2

(Fig. 7-86)

7 - 36 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Install the angle plate (1) by fastening the two


screws (2).

icn-09-v-712-16079-090000-09.iso

(Fig. 7-87)

▪ Connect the protective earth conductor 1


3
VAVA PH X11 (3).
▪ Reinstall the rear cover (2) by fastening the three
screws (1).
2

(Fig. 7-88)

For the installation of the front cover see chapter Replacement of the front cover and follow the instructions in
reverse order.

2.3 REPLACEMENT OF THE FRONT COVER


Code no. 4512 201 0279x

Replacement of the front cover

0.25 h

Tools/Material required:

– TC129 Standard tools

Scope of delivery:

– 1 x Front cover

Replacement procedure:

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 37


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Loosen both Allen screws (2).


▪ Remove the old front cover (1).

(Fig. 7-89)

▪ Position the new front cover on the frame (1).


1

icn-09-v-712-16079-090000-07.iso

(Fig. 7-90)

▪ Fasten both Allen screws (1).

(Fig. 7-91)

2.4 REPLACEMENT OF THE COMPLETE BUCKY UNIT 2


Code no. 4512 201 0256x
Code no. 4512 201 0258x
Code no. 4512 201 0257x

Replacement of the complete


Bucky unit 2
2.0 h

Tools/Material required:

7 - 38 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

– TC129 Standard tools


– Transport locking (four safety screws with red washers)

Scope of delivery:

– 1 x Bucky unit 2

NOTE When replacing the Bucky unit 2, reuse the replaceable grid and the 5-field measur‐
ing chamber of the old Bucky unit.

Replacement procedure:

WARNING Danger of injury!


After removing the Bucky unit 2
the system is not weight bal‐
anced. Make sure to insert the
four counterweight safety screws
(1) to prevent the system from
moving up automatically.
1
▪ Move the counterweight carriage into its parking 1
position and fix it with the four safety screws (1),
two on each side of the column.

(Fig. 7-92)

▪ Switch OFF the system at ENF1.

▪ Loosen the two Allen screws (1).


▪ Remove the front cover (2).

(Fig. 7-93)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 39


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Remove the three screws (1). 1


3
▪ Remove the rear cover (2).
▪ Disconnect the earth wire VA X12 (3).
▪ Remove the grid.
2

(Fig. 7-94)

▪ Remove the screws (2) and (3).


3
▪ Remove the metal bracket (1).

1 2

(Fig. 7-95)

▪ Pull out the 5-field measuring chamber to get ac‐


cess to the connector VAVA PH X1 (3).
▪ Remove the metal clamp (1) and the plastic
clamp (2).
▪ Disconnect VAVA PH X1 (3).
▪ Remove the 5–field measuring chamber.

(Fig. 7-96)

▪ Disconnect VAVA X11 (1) and VAVA X1 (2).


1 2

(Fig. 7-97)

7 - 40 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Remove the four screws that connect the Bucky


unit 2 to the vertical frame.
▪ Remove the Bucky unit 2.

(Fig. 7-98)

▪ Insert the new Bucky unit 2 from the operation


side into the vertical module. Make sure not to
damage or squeeze the cables.

left-hand operation

right-hand operation

(Fig. 7-99)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 41


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Put the cables through the opening into the Bucky


unit 2.
▪ Push the Bucky unit 2 completely into position
and fasten it with the four screws to the vertical
module.
▪ Check for smooth operation of the cassette tray.
▪ Connect VAVA X11 and VAVA X1.

(Fig. 7-100)

CAUTION Make sure that the wires are lo‐


cated under the housing of the 5- 1
field measuring chamber (1).

2
▪ Reinsert the 5-field measuring chamber. 4
▪ Reconnect VAVA PH X1 (2).
▪ Remount the metal clamp (3) and the plastic
clamp (4).
3
(Fig. 7-101)

▪ Reinstall the metal bracket (1) by fastening the


3
two screws (2) and (3).
▪ Fix the cable with cable ties.

1 2

(Fig. 7-102)

CAUTION When reinstalling the rear cover, ensure that the cables behind the rear cover do not
obstruct the movement of the grid frame.

7 - 42 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Insert the grid. 1


3
▪ Reconnect the earth wire VA X12 (3).
▪ Reinstall the rear cover (2).
▪ Remount the three screws (1).
2

(Fig. 7-103)

▪ Position the front cover on the frame (1).


1

icn-09-v-712-16079-090000-07.iso

(Fig. 7-104)

▪ Reinstall the front cover (1).


▪ Remount the two Allen screws (2).

(Fig. 7-105)

NOTE Keep back safety screws for later service!

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 43


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Remove the four safety screws from the column


(1).
▪ Press the four caps delivered with the accesso‐
ries bag into the tapped holes.

1
1

(Fig. 7-106)

▪ Switch ON the system at ENF1.

Test:

▪ Check the functioning of the Bucky unit 2:


– Check the grid function.
– Run a test exposure.

2.5 REPLACEMENT OF THE COMPLETE BUCKY UNIT 2 DIGITAL


Code no. 4512 201 0255x
Code no. 4512 201 0366x

Replacement of the complete


Bucky unit 2 digital
2.0 h

Tools/Material required:

– TC129 Standard tools


– Transport locking (four safety screws with red washers)

7 - 44 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

Scope of delivery:

– 1 x Bucky unit 2 digital

Front

(Fig. 7-107)

NOTE When replacing the Bucky unit 2 digital, reuse the replaceable grid, the 5-field meas‐
uring chamber and the digital detector of the old Bucky unit 2.

Replacement procedure:

WARNING Danger of injury!


After removing the Bucky unit 2
digital the system is not weight
balanced. Make sure to insert
the four counterweight safety
screws (1) to prevent the system
from moving up automatically.
1
▪ Move the counterweight carriage into its parking 1
position and fix it with the four safety screws (1),
two on each side of the column.

(Fig. 7-108)

▪ Switch OFF the system at ENF1.

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 45


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Loosen the two Allen screws (2).


▪ Remove the front cover (1).

(Fig. 7-109)

▪ Remove the three screws (1). 1


3
▪ Remove the rear cover (2).
▪ Disconnect the earth wire VA X12 (3).
▪ Remove the grid.
2

(Fig. 7-110)

▪ Remove the end stop (1) at the left side of the


Bucky unit 2 digital by loosening the nut (2).
▪ Disconnect the plug VAVA X1 (3). 3

(Fig. 7-111)

7 - 46 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Remove the end stop (1) at the right side of the


Bucky unit 2 digital by loosening the nut (2).
3
▪ Unplug the connector of the electromagnet VA X2
(3).

2
1

(Fig. 7-112)

▪ Disconnect the cables at the detector:


1 3
– Earth wire VAVA P X11 (lower bolt) (1).
– Fiber optic link VAVA P B1 (2).
– Communication cable VAVA P X2 (3) with 2
clamp (4).
– Power supply cable VAVA P X1 (5).
5
4

(Fig. 7-113)

▪ Pull out the 5-field measuring chamber to get ac‐


cess to the connector VAVA PH X1 (3).
▪ Remove the metal clamp (1) and the plastic
clamp (2).
▪ Disconnect VAVA PH X1 (3).
▪ Remove the 5–field measuring chamber.

(Fig. 7-114)

▪ Disconnect the cable of the fan X6 (1).


VA PCB X6

1 C6

C5

X1

X2

K1

X3

X5

C8

C9

X4

X66
icn-09-v-712-7966-090000-60.iso

(Fig. 7-115)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 47


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

CAUTION All cables have to be fixed with tape to protect them against damage when removing
the digital detector from the Bucky unit 2 digital.

▪ Fix the left hand cables and earth wires with 1


some tape (1) on the left side of the Bucky unit 2 2
digital. Position them above the earth bolt (2) to
protect them against damage when removing the
digital detector.

(Fig. 7-116)

▪ Fix the right hand ribbon cable with some tape (1)
on the right side of the Bucky unit 2 digital.

(Fig. 7-117)

▪ Pull the detector (1) carefully out of the Bucky unit


2 digital.

1
(Fig. 7-118)

7 - 48 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Remove the angle plate (1) by loosening the two


screws (2).
▪ Remove the two screws (3) on both sides of the
Bucky unit 2 digital.
2
▪ Exchange the Bucky unit 2 digital.
▪ Remount the two screws (3).
▪ Reinstall the angle plate (1) and remount the two
screws (2).
1
3
icn-09-v-712-16079-090000-08.iso

(Fig. 7-119)

▪ Remove the end stop (1) at the left side of the


new Bucky unit 2 digital by loosening the nut (2).

(Fig. 7-120)

▪ Remove the end stop (1) at the right side of the


new Bucky unit 2 digital by loosening the nut (2).
3
▪ Unplug the connector of the electromagnet VA X2
(3).

2
1

(Fig. 7-121)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 49


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Fix the left hand cables and earth wires with 1


some tape (1) on the left side of the Bucky unit 2 2
digital. Position them above the earth bolt (2) to
protect them against damage when inserting the
digital detector.

(Fig. 7-122)

▪ Fix the right hand ribbon cable with some tape (1)
on the right side of the Bucky unit 2 digital.

(Fig. 7-123)

▪ Apply a uniform layer of the talcum powder to the


rubber guiding pieces (1) of the digital detector
and to the four rubber hold-downs inside the
Bucky unit 2 digital.

(Fig. 7-124)

CAUTION Make sure not to damage the wires and cables of the Bucky unit 2 digital.

7 - 50 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Carefully install the detector (1).

1
(Fig. 7-125)

CAUTION Make sure that the wires are lo‐


cated under the housing of the 5- 1
field measuring chamber (1).

2
▪ Reinsert the 5-field measuring chamber. 4
▪ Reconnect VAVA PH X1 (2).
▪ Remount the metal clamp (3) and the plastic
clamp (4).
3
(Fig. 7-126)

▪ Reconnect the cables at the detector:


1 3
– Earth wire VAVA P X11 (lower bolt) (1).
– Fiber optic link VAVA P B1 (2).
– Communication cable VAVA P X2 (3) with 2
clamp (4).
– Power supply cable VAVA P X1 (5).
5
4

(Fig. 7-127)

▪ Reconnect the cable of the fan X6 (1).


VA PCB X6

1 C6

C5

X1

X2

K1

X3

X5

C8

C9

X4

X66
icn-09-v-712-7966-090000-60.iso

(Fig. 7-128)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 51


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Install the end stop (1) at the right side of the


Bucky unit 2 digital and tighten the nut (2).
3
▪ Reconnect the connector of the electromagnet
VA X2 (3).

2
1

(Fig. 7-129)

▪ Install the end stop (1) at the left side of the


Bucky unit 2 digital and tighten the nut (2).
▪ Connect the plug VAVA X1 (3). 3

(Fig. 7-130)

CAUTION When reinstalling the rear cover, ensure that the cables behind the rear cover do not
obstruct the movement of the grid frame.

▪ Insert the grid. 1


3
▪ Reconnect the earth wire VA X12 (1).
▪ Reinstall the rear cover (2).
▪ Remount the three screws (3).
2

(Fig. 7-131)

7 - 52 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Position the front cover on the frame (1).


1

icn-09-v-712-16079-090000-07.iso

(Fig. 7-132)

▪ Reinstall the front cover (1).


▪ Reinstall the two Allen screws (2).

(Fig. 7-133)

NOTE Keep back safety screws for later service!

▪ Remove the four safety screws from the column


(1).
▪ Press the four caps delivered with the accesso‐
ries bag into the tapped holes.

1
1

(Fig. 7-134)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 53


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Switch ON the system at ENF1.

Test:

▪ For detector calibration refer to the respective system manual.


▪ Check the functioning of the Bucky unit 2 digital.
▪ Check the grid function.
▪ Run a test exposure.

2.6 REPLACEMENT OF THE DIGITAL DETECTOR


Code no. 4512 131 3516x

Replacement of the digital detector

1.5 h

Tools/Material required:

– TC129 Standard tools


– Transport locking (four safety screws with red washers)

Scope of delivery:

– 1 x Digital detector

(Fig. 7-135)

Replacement procedure:

WARNING Danger of injury!


Make sure to insert the four coun‐
terweight safety screws(1) to pre‐
vent the system from moving up
automatically.

1
▪ Move the counterweight carriage into its parking
1
position and fix it with the four safety screws (1),
two on each side of the column .

(Fig. 7-136)

▪ Switch OFF the system at ENF1.

7 - 54 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Remove the rear cover (1) of the Bucky unit 2 dig‐ 1


3
ital, three screws ( 2).
▪ Disconnect the earth wire VA X12 (3).

(Fig. 7-137)

▪ Remove the end stop (1) at the left side of the


Bucky unit 2 digital by loosening the nut (2).
▪ Disconnect the plug VAVA X1 (3). 3

(Fig. 7-138)

▪ Remove the end stop (1) at the right side of the


Bucky unit 2 digital by loosening the nut (2).
3
▪ Unplug the connector of the electromagnet VA X2
(3).

2
1

(Fig. 7-139)

▪ Disconnect the cables at the detector:


1 3
– Earth wire VAVA P X11 (lower bolt) (1).
– Fiber optic link VAVA P B1 (2).
– Communication cable VAVA P X2 (3) with 2
clamp (4).
– Power supply cable VAVA P X1 (5).
5
4

(Fig. 7-140)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 55


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Pull out the 5-field measuring chamber to get ac‐


cess to the connector VAVA PH X1 (3).
▪ Remove the metal clamp (1) and the plastic
clamp (2).
▪ Disconnect VAVA PH X1 (3).
▪ Remove the 5–field measuring chamber.

(Fig. 7-141)

NOTE All cables have to be fixed with tape to protect them against damage when removing
the digital detector from the Bucky unit 2 digital.

▪ Fix the left hand cables and earth wires with 1


some tape (1) on the left side of the Bucky unit 2 2
digital. Position them above the earth bolt (2) to
protect them against damage when replacing the
digital detector.

(Fig. 7-142)

▪ Fix the right hand ribbon cable with some tape (1)
on the right side of the Bucky unit 2 digital.

(Fig. 7-143)

7 - 56 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Pull the detector (1) carefully out of the Bucky unit


2 digital.

1
(Fig. 7-144)

▪ Remove both rubber guiding pieces (1) from the


old detector and install them on the bottom side
of the new detector.
▪ Apply a uniform layer of the talcum powder to the
rubber guiding pieces and to the rubber hold-
downs inside the Bucky unit 2 digital. 1

(Fig. 7-145)

CAUTION Make sure not to damage the wires and cables of the Bucky unit 2 digital.

▪ Insert the new detector (1) carefully into the


Bucky unit 2 digital.

1
(Fig. 7-146)

CAUTION Ensure that the wires are located under the housing of the 5–field measuring chamber.

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 57


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Reinsert the 5-field measuring chamber carefully.


▪ Reconnect VAVA PH X1 (3).
▪ Reinstall the metal clamp (1) and the plastic
clamp (2).
▪ Push the 5-field measuring chamber fully into the
Bucky unit 2 digital.

(Fig. 7-147)

▪ Connect the cables at the detector:


1 3
▪ Earth wire VAVA P X11 (lower bolt) (1).
▪ Fiber optic link VAVA P B1 (2).
▪ Communication cable VAVA P X2 (3) with 2
clamp (4).
▪ Power cable VAVA P X1 (5).
5
4

(Fig. 7-148)

▪ Reinstall the end stop (1) on the right side of the


Bucky unit 2 digital and tighten the nut (2).
3
▪ Plug in the connector of the electromagnet VA X2
(3).

2
1

(Fig. 7-149)

▪ Reinstall the end stop (1) at the left side of the


Bucky unit 2 digital and tighten the nut (2).
▪ Reconnect the plug VAVA X1 (3). 3

(Fig. 7-150)

7 - 58 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

CAUTION
Ensure that the cables behind the rear cover do not obstruct the movement of the
grid frame.

▪ Reconnect the earth wire VA X12 (3). 1


3
▪ Reinstall the rear cover (2) of the Bucky unit 2 dig‐
ital (three screws; 1).

(Fig. 7-151)

NOTE Keep back safety screws for lat‐


er service!

▪ Remove the four safety screws (1) from the col‐ 1


umn.
1
▪ Press the four caps delivered with the accesso‐
ries bag into the tapped holes.

(Fig. 7-152)

▪ Switch ON the system at ENF1.

Test:

▪ Check the function.


▪ For detector calibration refer to the accordant system manual.

2.7 REPLACEMENT OF THE 5-FIELD MEASURING CHAMBER


Code no. 9890 000 7000x

Replacement of the
5-field measuring chamber
0.5 h

Replacement procedure:

▪ Switch OFF the system at ENF1.

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 59


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Remove the three screws (1). 1


3
▪ Remove the rear cover (2).
▪ Disconnect the earth wire VA X12 (3).

(Fig. 7-153)

For Bucky unit 2 digital:

▪ Remove the end stop (1) at the left side of the


Bucky unit 2 digital by loosening the nut (2).
▪ Disconnect the plug VA X1 (3). 3

(Fig. 7-154)

For Bucky unit 2:

▪ Remove the screws (2) and (3).


3
▪ Remove the metal bracket (1).

1 2

(Fig. 7-155)

▪ Pull out the 5-field measuring chamber to get ac‐


cess to the connector VAVA PH X1 (3).
▪ Remove the metal clamp (1) and the plastic
clamp (2).
▪ Disconnect VAVA PH X1 (3).
▪ Remove the old 5–field measuring chamber.

(Fig. 7-156)

7 - 60 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Set the sliding switch VAVA PH S1 of the new 5-


field measuring chamber according to its position
in the old 5-field measuring chamber (position 1 -
S1 Settings Chamber Layout
all 5 fields are active; suitable for left-hand and all 5 fields
and only field
right-hand operation). field 1 - 2 - 3 4-2-5
(head left) (head right)
default
settings

icn-08-v-712-17007-0000-01.iso

(Fig. 7-157)

CAUTION Ensure that the wires are located under the housing of the 5–field measuring chamber.

▪ Insert the new 5-field measuring chamber carefully.


▪ Reconnect VAVA PH X1 (3).
▪ Reinstall the metal clamp (1) and the plastic
clamp (2).
▪ Push the 5-field measuring chamber fully into the
Bucky unit 2 digital.

(Fig. 7-158)

For Bucky unit 2 digital:

▪ Reinstall the end stop (1) at the left side of the


Bucky unit 2 digital and tighten the nut (2).
▪ Reconnect the plug VAVA X1 (3). 3

(Fig. 7-159)

For Bucky unit 2:

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 61


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Reinstall the metal bracket (1) by fastening the


3
two screws (2) and (3).
▪ Fix the cable with cable ties.

1 2

(Fig. 7-160)

CAUTION Ensure that the cables behind the rear cover do not obstruct the movement of the
grid frame.

▪ Reconnect the earth wire VA X12 (3). 1


3
▪ Reinstall the rear cover (2) of the Bucky unit 2 dig‐
ital (three screws; 1).

(Fig. 7-161)

▪ Switch ON the system at ENF1.


▪ Check the function.

2.8 REPLACEMENT OF THE VERTICAL UNIT SAFETY COVER ASSEMBLY


Code no. 4512 133 3007x

Replacement of the vertical


unit safety cover assembly
0.25 h

Tools/Material required:

– TC129 Standard tools

Scope of delivery:

– 1 x Vertical frame safety cover assembly incl. fixing material

Replacement procedure:

7 - 62 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Tilt the module 90° (1) and move the tilting mod‐
ule to its highest position (2).

2
1

(Fig. 7-162)

▪ Remove the eight screws (1) and the old cover as‐
sembly (2).
▪ Install the new vertical frame cover assembly.

(Fig. 7-163)

NOTE The collision protection switches of the tilting unit are located behind the vertical
frame cover assembly!

▪ Check the function of the vertical frame cover assembly. When pressing on the edges of the cover, a click‐
ing sound of the collision protection switches has to be heard.
▪ Check that the cover does not collide with any part of the tilting unit.

2.9 REPLACEMENT OF THE SET OF ACTUATORS FOR SAFETY SWITCHES FOR VERTICAL UNIT
Code no. 4512 133 3008x

Replacement of the set of safety


switches for vertical unit
1.0 h

Tools/Material required:

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 63


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

– TC129 Standard tools

Scope of delivery:

– 1 x Set of actuators for safety switches for vertical


frame incl. fixing material

(Fig. 7-164)

Replacement procedure:

▪ Move the counterweight carriage into its parking


position and fix it with the four safety screws with
red washers (1), two on each side of the column.

1
1

(Fig. 7-165)

▪ Loosen the two Allen screws (2).


▪ Remove the front cover (1).

(Fig. 7-166)

▪ Remove the 5-field measuring chamber and the Bucky unit 2/Bucky unit 2 digital (see chapter Replacement
of the 5-field measuring chamber, page 7-59) and (see chapter Replacement of the complete Bucky unit 2 ,
page 7-38)/(see chapter Replacement of the complete Bucky unit 2 digital, page 7-44).

7 - 64 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Remove the screws (1) (2 x) and pull out the guid‐


ing rails for grid parking (2).

(Fig. 7-167)

▪ Remove the spring (1).


▪ Replace the rocker lever (2).
– Pay attention to the roller follower of the micro‐ 2 1
switch (3).
▪ Reinstall the spring (1).

(Fig. 7-168)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 65


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Repeat the procedure with all safety switches (1).


1 1

1
1
(Fig. 7-169)

▪ Insert the guiding rails for grid parking (2) and fix
the screws (1) (2 x).

(Fig. 7-170)

▪ Reinstall the Bucky unit 2/Bucky unit 2 digital, the 5-field measuring chamber and the digital detector (when
using Bucky unit 2 digital) (see chapter Replacement of the complete Bucky unit 2 , page 7-38)/(see chap‐
ter Replacement of the complete Bucky unit 2 digital, page 7-44), (see chapter Replacement of the 5-field
measuring chamber, page 7-59) and (see chapter Replacement of the digital detector, page 7-54).

NOTE Keep back safety screws for later service!

7 - 66 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Remove the four safety screws from the column.


▪ Check the function.
▪ Reinstall the front cover.

2.10 REPLACEMENT OF THE MICROSWITCH WITH ROLLER FOLLOWER


Code no. 4512 133 2949x

Replacement of the microswitch


with roller follower
1.0 h

Tools/Material required:

– TC129 Standard tools


– Transport locking (four safety screws with red washers)

Scope of delivery:

– 1 x Microswitch with roller follower.

(Fig. 7-171)

Replacement procedure:

▪ Move the counterweight carriage into its parking


position and fix it with the four safety screws (1),
two on each side of the column.

1
1

(Fig. 7-172)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 67


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Remove the front cover (1) of the tilting/vertical


unit by loosening the screws (2).

(Fig. 7-173)

▪ Remove the digital detector (when using Bucky unit 2 digital), the 5-field measuring chamber and the Bucky
unit 2/Bucky unit 2 digital. (see chapter Replacement of the digital detector, page 7-54) (see chapter Replace‐
ment of the 5-field measuring chamber, page 7-59)(see chapter Replacement of the complete Bucky unit 2 ,
page 7-38)(see chapter Replacement of the complete Bucky unit 2 digital, page 7-44)

▪ Remove the screws (1) (2 x) and pull out the guid‐


ing rails for grid parking (2).

(Fig. 7-174)

7 - 68 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Replace the microswitch with roller follower (1).

(Fig. 7-175)

▪ Insert the guiding rails for grid parking (2) and fix
the screws (1) (2 x).

(Fig. 7-176)

▪ Reinstall the Bucky unit 2 / Bucky unit 2 digital, the 5-field measuring chamber and the digital detector (when
using Bucky unit 2 digital) according to chapter 2.4, 2.5 and 2.6.

NOTE Keep back safety screws for later service!

▪ Remove the four safety screws from the column.


▪ Check the function.
▪ Reinstall the front cover.

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 69


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

2.11 REPLACEMENT OF THE FRONT COVER SUPPORT


Code no. 4512 133 01101

Tools/Material required:

– TC129 Standard tools

Scope of delivery:

– 2 x Front cover support

icn-09-v-712-16079-090000-05.iso

(Fig. 7-177)

Replacement procedure:

▪ Remove the front cover (1) of the tilting/vertical


module by loosening the screws (2).

(Fig. 7-178)

▪ Replace the front cover supports (1).

icn-09-v-712-16079-090000-06.iso

(Fig. 7-179)

7 - 70 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Reinstall the front cover of the vertical/tilting module.

2.12 REPLACEMENT OF THE SET OF ACTUATORS FOR SAFETY SWITCHES FRONT COVER
Code no. 4512 133 3009x

Replacement of the set of safety


switches front cover
1.5 h

Tools/Material required:

– TC129 Standard tools


– Transport locking (four safety screws with red washers)

Scope of delivery:

– 1 x Set of safety switches front cover

(Fig. 7-180)

Replacement procedure:

▪ Move the counterweight carriage to its parking po‐


sition and fix it with the four safety screws (1), two
on each side of the column.

1
1

(Fig. 7-181)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 71


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Remove the front cover (1) from the tilting/vertical


module by loosening the screws (2).

(Fig. 7-182)

▪ Remove the front cover support.

icn-09-v-712-16079-090000-06.iso

(Fig. 7-183)

▪ Remove the circlip and washer (2 x).

(Fig. 7-184)

▪ Remove the complete Bucky unit 2 / Bucky unit 2 digital with digital detector, and the 5-field measuring cham‐
ber according to chapter 2.4. / 2.5. and 2.6.

7 - 72 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Remove the old actuator.

(Fig. 7-185)

▪ Fit washer and circlip (2 x).

(Fig. 7-186)

▪ Position the new actuator.

(Fig. 7-187)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 73


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Fix the actuator from the inner side.

(Fig. 7-188)

▪ Repeat the procedure with the second actuator.


▪ Reinstall the Bucky unit 2 / Bucky unit 2 digital with digital detector and the 5-field measuring chamber.
▪ Reinstall the front cover support.

NOTE Keep back safety screws for later service!

▪ Remove the safety screws from the column.


▪ Press the four caps delivered with the accessories bag into the tapped holes.
▪ Check the function.
▪ Reinstall the front cover.

2.13 REPLACEMENT OF THE POTENTIOMETER 10K FOR TILT MOVEMENT


Code no. 4512 133 3001x

Replacement of the potentiometer


10K for tilt movement
1.0 h

Tools/Material required:

– TC129 Standard tools


– Spirit level

7 - 74 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

Scope of delivery:

– 1 x Potentiometer10 kΩ for tilt movement incl. ca‐


ble and holder plate

(Fig. 7-189)

NOTE Ensure that the grub screw at the potentiometer is fixed!

Replacement procedure:

▪ Move the tilting module to its highest position (1)


and tilt the module 90° (2).
▪ Remove the bottom cover (3) with 2.5 mm Allen
key. 1

(Fig. 7-190)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 75


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Disconnect connector VT1 X8 from the controller


board (1).
1

X8
1
X8

6 9 15 9
1 5 8 1

1
(Fig. 7-191)

▪ Remove the cable tie (1).


▪ Loosen the torx screws TX20 (2) from holder plate.
▪ Remove the holder plate incl. potentiometer (3).

1 2 3
(Fig. 7-192)

▪ Pre-adjust the new potentiometer to 9.5 kΩ.

(Fig. 7-193)

▪ Insert the new potentiometer and push it into the clutch.


▪ Fix the holder plate.
▪ Connect the cable to VT1 X8 at the control board.
▪ Ensure that the resistance between pin 1 and 3 is 9.5 kΩ.
▪ Ensure that the voltage is approx. 4.85 V.

7 - 76 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Continue with chapter 3. Setting-to-work.


▪ After the adjustment of the wall stand tilt check the tilting function of the wall stand.
▪ Reinstall the cover.

NOTE For software adjustment of the wall stand tilt angle refer to manual X-Scope for sys‐
tems with Bucky Controller or Field Service Framework for systems with PC based
controller (Eleva systems).

▪ Fix the cable with cable tie.


▪ Reinstall the bottom cover.

2.14 REPLACEMENT OF THE PCB VT1


Code no. 4512 133 0200x

Replacement of the
PCB VT1 controller
1.5 h

Tools/Material required:

– TC129 Standard tools

Scope of delivery:

– 1 x PCB VT1

Replacement procedure:

▪ Tilt the module 90° (1) and move the tilting mod‐
ule to its highest position (2).

2
1

(Fig. 7-194)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 77


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Remove the bottom cover with 2.5 mm Allen key.


▪ Disconnect all connectors from the controller 1
board VT1 (1).
1 1
1
7 1
9 3
1
1

on
1
1 1
X19 1
X1
1

1
1 5

4 8 on
9 15
1 9 6 Reset
1 8 5 1

(Fig. 7-195)

▪ Replace the controller board VT1.


▪ Connect all connectors to the controller board VT1 (1).
▪ Check the function of the system.
▪ Reinstall the bottom cover with 2.5 mm Allen key.

2.15 REPLACEMENT OF THE FLAP ASSEMBLY FOR GRID PARKING


Code no. 4512 133 2861x

Replacement of the flap assembly


for grid parking
0.5 h

Tools/Material required:

– TC129 Standard tools

Scope of delivery:

– Flap assembly Set with close latch

Replacement procedure :
2
▪ Loosen but do not remove the screws of the angu‐
lar plate (1) which carries the close latch and re‐
move it using the oblong holes.
▪ Remove the flap (2) by loosening the two screws.

(Fig. 7-196)

7 - 78 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Install the flap (2) without fastening the screws. Shut the flap and align it to the vertical module. Then fasten
the screws.
▪ Install the angular plate (1) using the oblong holes and fasten it that way that the flap stays aligned to the
vertical module when being shut.
▪ Check the flap assembly for smooth operation.

2.16 REPLACEMENT OF THE TILT AXIS DRIVE MOTOR


Code no. 4512 133 2928x

Replacement of the
tilt axis drive motor
2.0h

Tools/Material required:

– TC129 Standard tools


– Loctite

Scope of delivery:

– 1 x Tilt axis drive motor incl. cable

(Fig. 7-197)

Replacement procedure:

▪ Move the tilting module to its highest position (1).


▪ Remove the bottom cover (2) with 2.5mm Allen
key.
1

(Fig. 7-198)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 79


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Disconnect connector VT1 X3 from the controller


board (1).

1
X8

X3
6 9 15 9
1 5 8 1

(Fig. 7-199)

▪ Remove screws TX30 (1) (4x).


▪ Pull out the holder plate with motor and sprocket.

(Fig. 7-200)

▪ Loosen grub screw (1).

(Fig. 7-201)

7 - 80 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Pull out the sprocket (1).

1
(Fig. 7-202)

▪ Disconnect the connector (1).

(Fig. 7-203)

▪ Remove screws TX20 (1) (4x).


▪ Replace the motor and fix it with screws TX20 (1)
(4x).
– Secure the screws with Loctite.

(Fig. 7-204)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 81


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Insert the sprocket (1) and fix grub screw (2).

1 2
(Fig. 7-205)

▪ Connect the connector (1).

(Fig. 7-206)

NOTE Shift the motor unit in direction (3) until the plastic clutch (1) has minimum clearance
in both directions. Too much clearance (approx. < 2mm) can lead to motor noise dur‐
ing movement.

7 - 82 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics
N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Join the clutch together (1).


▪ Fix the holder plate with screws TX30 (2) (4x).

1
(Fig. 7-207)

▪ Connect connector VT1 X3 at the controller board


(1).

1
X8

X3
6 9 15 9
1 5 8 1

(Fig. 7-208)

▪ Check the function.

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 83


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Install the bottom cover (1) with 2.5mm Allen key.

(Fig. 7-209)

2.17 REPLACEMENT OF THE COUPLING TILT DRIVE


Code no. 4512 133 2947x

Replacement of the
coupling tilt drive
2.0h

Tools/Material required:

– TC129 Standard tools

Scope of delivery:

– 1 x Coupling tilt drive comprising


– 2 x Sprocket
– 1 x Plastic clutch

(Fig. 7-210)

Replacement procedure:

7 - 84 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Move the tilting module to its highest position (1).


▪ Remove the bottom cover (2) with 2.5mm Allen
key.
1

(Fig. 7-211)

▪ Disconnect connector VT1 X3 from the controller


board (1).

1
X8

X3
6 9 15 9
1 5 8 1

(Fig. 7-212)

▪ Remove screws TX30 (1) (4x).


▪ Pull out the holder plate with motor (2) and plastic
clutch (3).

3
(Fig. 7-213)

4512 984 28037 REV A 7 - 85


CSIP Level 1 (19.0)
© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Loosen grub screw (1) and replace the sprocket


(2).

(Fig. 7-214)

▪ Loosen grub screw (1) and replace the sprocket


at the motor side.

1
(Fig. 7-215)

NOTE Shift the motor unit in direction (3) until the plastic clutch (1) has minimum clearance
in both directions. Too much clearance (approx. < 2mm) can lead to motor noise dur‐
ing movement.

7 - 86 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Join the clutch together (1).


▪ Fix the holder plate with screws TX30 (2) (4x).

3
1
(Fig. 7-216)

▪ Connect connector VT1 X3 at the controller board


(1).

1
X8

X3
6 9 15 9
1 5 8 1

(Fig. 7-217)

▪ Check the function.

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 87


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Install the bottom cover (1) with 2.5mm Allen key.

(Fig. 7-218)

2.18 REPLACEMENT OF THE USER INTERFACE KEYPAD AND CABLE


Code no. 4512 133 318xx

Replacement of the
user interface keypad and cable
1.5 h

Tools/Material required:

– TC129 Standard tools

Scope of delivery:

– 1 set of keypad and cable

Replacement procedure:

▪ Loosen the two screws (1) on the lower side of


the BuckyDiagnost VS.

(Fig. 7-219)

7 - 88 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Remove the front cover starting from the lower


side of the Bucky unit. The upper side of the cov‐
er is snapped in on top of the Bucky unit.

(Fig. 7-220)

▪ Remove the rear cover (2) loosening the three 1


3
screws (1).
▪ Disconnect the protective earth conductor VA
X12 (3).
2

(Fig. 7-221)

▪ Take out the Amplimat chamber until there is ac‐


cess to the plug VAVA PH X1 (3).
▪ Remove the cable tie.
▪ Remove the metal clamp (1) and the plastic
clamp (2).
▪ Disconnect VAVAP H X1 (3) carefully.
▪ Remove the Amplimat chamber carefully.

(Fig. 7-222)

▪ Disconnect VAVA X11 (1) and VAVA X1 (2).


1 2

(Fig. 7-223)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 89


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Remove the four screws that connect the Bucky


unit 2 to the vertical frame.
▪ Remove the Bucky unit 2.

(Fig. 7-224)

▪ Remove the metal plate (1) that covers the cables


by loosening the screw.
▪ Disconnect the cable connectors at the PCB. 1
2
▪ Remove the flap of the grid parking by loosening
the two screws.
▪ Loosen the screws on both sides of the user key‐
pad (2). 3

▪ Remove the keypad and the cable.

(Fig. 7-225)

▪ Insert the cable parts into the slots (3) in the verti‐
cal module. 2
▪ Install both parts of the user keypad by fastening
the screws (2).
▪ Connect the cables to the PCB in the vertical mod‐
ule.
▪ Reinstall the flap and align it to the vertical frame.
▪ Reinstall the cable cover and fasten the screw.

(Fig. 7-226)

7 - 90 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

▪ Insert the Bucky unit 2 from the operation side


(grid insertion side) into the vertical module. Make
sure not to damage or squeeze the cables.

left-hand operation

right-hand operation

(Fig. 7-227)

▪ Put the cables through the opening into the Bucky


unit 2.
▪ Push the Bucky unit 2 completely into position
and fasten it with the four screws to the vertical
module.
▪ Check the smooth operation of the cassette tray.
▪ Connect VAVA X11 and VAVA X1.

(Fig. 7-228)

▪ Connect VAVA PH X1 (3).


▪ Install the metal clamp (1) and the plastic clamp
(2).

NOTE Take care that all wires are


placed under the black housing
of the Amplimat chamber!

(Fig. 7-229)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 91


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS REPLACEMENT

▪ Push the Amplimat chamber completely into the


3
Bucky unit.
▪ Turn the metal bracket (1) to the right side so that
the Amplimat chamber cannot be moved and fas‐
ten the screws (2) and (3).

1 2

(Fig. 7-230)

▪ Install the angle plate (1) by fastening the two


screws (2).

icn-09-v-712-16079-090000-09.iso

(Fig. 7-231)

▪ Connect the protective earth conductor VA 1


3
X12 (3).
▪ Reinstall the rear cover (2) by fastening the three
screws (1).
2

(Fig. 7-232)

For the installation of the front cover see chapter Replacement of the front cover and follow the instructions in
reverse order.

2.19 REPLACEMENT OF THE PATIENT HAND GRIP RIGHT OR LEFT


Code no. 4512 133 3254x or 4512 133 3255x

Replacement of the
patient hand grip
1.5 h

7 - 92 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
REPLACEMENT BuckyDiagnost VS

Tools/Material required:

– TC129 Standard tools

Scope of delivery:

– 1 x Patient hand grip left or right

Replacement procedure:

Not yet available

3 REPLACEMENT PROCEDURES AT THE WALL CONNECTION BOX

3.1 REPLACEMENT OF THE PCB VP1


Code no. 4512 133 0200x

Replacement of the
PCB VP1 controller
1.5 h

Tools/Material required:

– TC129 Standard tools

Scope of delivery:

– 1 x PCB controller board VP1

Replacement procedure:

▪ Remove the cover from the wall connection box


(1).
▪ Remove all cables and connectors from board
1
VP1 (2).
▪ Replace the controller board.
▪ Connect all cables and connectors vice versa.
▪ Ensure the correct jumper settings according
Z3-3.1 (VP1).
▪ Check the function.
▪ Reinstall the cover.

(Fig. 7-233)

4512 984 28037 REV A CSIP Level 1 (19.0) 7 - 93


© 2020 Koninklijke Philips Electronics N.V. ALL
RIGHTS RESERVED
MECHANICAL DRAWINGS BuckyDiagnost VS

MECHANICAL DRAWINGS
Contents
BuckyDiagnost VS basic / Wall mounted version / Mechanical dimensions ................................................Z-1.1.1
BuckyDiagnost VS advanced / Wall mounted version / Mechanical dimensions .........................................Z-1.2.1
BuckyDiagnost VS advanced tilt / Wall mounted version / Mechanical dimensions ...................................Z-1.3.1
BuckyDiagnost VS digital with BU 2 (cass) / Wall mounted version / Mech. dimensions ............................Z-1.4.1
BuckyDiagnost VS digital tilt w. BU 2 (cass) / Wall mounted version / Mech. dimensions...........................Z-1.4.3
BuckyDiagnost VS digital with BU 2 dig (detector) / Wall mounted version / Mech. dimensions .................Z-1.4.5
BuckyDiagnost VS digital tilt with BU 2 dig (detector) / Wall mounted version / Mech. dimensions ............Z-1.4.7
BuckyDiagnost VS basic/advanced / Wall mounted version / Template .........................................................Z-4.1
BuckyDiagnost VS basic / Wall mounted version / Floor fixing Template ....................................................Z-4.3.1
BuckyDiagnost VS basic / Stand alone version / Floor fixing / Template .....................................................Z-4.3.2
BuckyDiagnost VS digital Wall mounted version Template.............................................................................Z-4.4
BuckyDiagnost VS advanced/digital / Template for floor fixing (wall mounted version) ...............................Z-4.6.1
BuckyDiagnost VS advanced/digital / Template for floor fixing (stand alone Version) .................................Z-4.6.2
BuckyDiagnost VS Wall holder / Mechanical dimensions ...............................................................................Z-1.5

4512 984 28037 REV A CSIP Level 1 (19.0) 8-1


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS MECHANICAL DRAWINGS

8-2 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
min. 000 1 )
I-------, 1

--;_ ___ .[_-_]Bt------t--�

L[)
n
L[)

0 L[)
0 0
L[)
0
CD O N
o m X X
N m 0 0
E E
en
0

+I
n
0 L[)

+1
n
Cs.

,-T 10
1
I I 0
� -I I I 0
n
0 I I I Ill: C
-+-' � E
0
0 \_ _J
/
0,
C
•-
406
C
C 550±50
0
7 1 0±50
0:::
-+-' 1 39
u
::J
u
0 50
L
Q_

Ref. cross point

(D
m
L[)

� :;::;
-
0
I

N
N �
I I
4 1 7±7.5
:;
DO

I :; 1) space for cassette and grid exchange

BuckyDiognost VS bosic
U) N
DO
0 m

Woll mounted version


N
n :;::;
'T

Mechonicol dimensions
<C
0 500 1000

..........mm

982 0180. CSIP Level 1 (05.1) Z-1.1.1


Philips Electronics NY
reserved
Drawings/Diagrams VS basic BuckyDiagnost VS

DRAWINGS/DIAGRAMS VS BASIC
Contents
Survey of units BuckyDiagnost VS ...................................................................................................................Z1-1
BuckyDiagnost VS / Labeling .........................................................................................................................2Z-10
Conversion of measuring chamber ..................................................................................................................Z2-9
BuckyDiagnost VS Basic ..................................................................................................................................Z1-2
Power supply BD VS basic...............................................................................................................................Z1-3
MSB connection and Pin assignment BD VS basic_advanced ........................................................................Z1-4
BuckyDiagnost VS Basic / Advanced Release circuit with BU 2 non sensing .................................................Z1-5
Earthing diagram BD VS basic_advanced ......................................................................................................Z2-1
Wall box basic BD VS basic .............................................................................................................................Z2-2
Vertical unit BD VS basic .................................................................................................................................Z2-5
Vertical frame BD VS basic ..............................................................................................................................Z2-6
PCB Interface basic BD VS basic .................................................................................................................Z3-1.1
PCB Interface basic BD VS basic .................................................................................................................Z3-1.2

4512 984 28037 REV A CSIP Level 1 (19.0) 9-1


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS Drawings/Diagrams VS basic

9-2 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
VP
VB
VP
VB

VT

VA

VAG2
VA
VAPH
VAG1
VA = Vertical frame
VAG1 = User interface left
VD VAG2 = User interface right
VAPH = 5-Field measuring chamber
VAVA = Bucky unit 2 digital
(BuckyDiagnost VS digital)
VAVA
VAVA Bucky unit 2 sensing/ non sensing
(BuckyDiagnost VS basic/ advanced)
VAPH VB = Vertical unit
VD = Display
VP = Wall box
VT = Tilting unit

BuckyDiagnost VS basic BuckyDiagnost VS advanced


BuckyDiagnost VS digital
A3 09-10-21 Schr.

BuckyDiagnost VS
Survey of units

BuckyDiagnost VS CSIP Level 1 (19.0 Z1-1


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
3) "Wall Box"
High End
Wallstand
Segment Control Unit M 3) I
(Optional, if no table is present!
Wall Box
BuckyUnit 2 sensing / non sensing
Only for DigitalDiagnost 1.x, not
necessary for BuckyDiagnost VS digital)

3) B Logo Address
BuckyUnit 2 digital Typen No. Electr. data
IEC60601-1 IEC60601-2-32

I FDA
UL / CSA IEC
Logo Address
WEEE
CE Environmental
Typen No. Electr. data

Typen No.
IEC60601-1 IEC60601-2-32

Logo
H UL / CSA
FDA
IEC

Barcode

UL / CSA
WEEE
CE Environmental

Typen No.
Logo
M

UL / CSA
Barcode
E
Outside
F G Inside

3) E Vertical unit G
with BuckyDiagnost VS Basic G
with BuckyDiagnost VS advanced /
advanced tilt
3) H

Barcode
Address
Family PEI Address

Familie Pei

Typen No.
Typen No. Barcode
Package 1) Address
Typen No. Barcode

Logo Address
Typen No. UL / CSA
A2/A3 08-06-04 Schr. / Kö

3) F 1) Package VS basic
Package VS advanced
2) Package VS advanced tilt
Package VS advanced tilt replacement BuckyDiagnost VS
Package VS basic replacement Package VS digital tilt
Tilting unit Package 2) Address Package VS advanced replacement Labeling
3) Necessary for BuckyDiagnost VS digital
Typen No. Barcode Package VS digital

4512 983 11021 CSIP Level 0 (b/08.0) 2Z-10


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS BuckyDiagnost VS

Left hand Right hand


operation operation
(default)

Hardware Programmings

BuckyDiagnost TH2/TF (MSB)


BuckyDiagnost VS basic
1 2 1 2
S1 in position 1 5 Field measuring chamber S1 in position 2 S1
S1 (default)

BuckyDiagnost VS advanced / digital


5-Field control
Position 1-2 Jumper VP3X2 Position 2-3
(default for VS)

Software Programmings
Bucky Controller (X-Scope)

Program Manual / Bucky RAD / Wallstand / Basic Setup /

Left (default) Bucky Device Orientation Right


A3/A4 06-06-13 Schr.

BuckyDiagnost TH2/TF (MSB)


Conversion of measuring chamber

4512 983 11361 CSIP Level 0 (06.0) Z2-9


C 2004 Philips Medical Systems
P B
Wall box Basic Vertical unit Basic
4512 133 2903x 4512 201 0270x

0V N2 20,4V AC
230V AC
A Vertical frame basic
4512 201 0275x (including wall box basic)

I Interface
Basic
SIX54 / VSIX54 X6 X8
230VAC
(NPX2 ED Eleva) X7 X9
Mains connection X5 X1
230VAC
X3 X3 Y1 Brake
X2
X21 X1
X25 X4
X20 X5
X40
X16
26VDC X17
SIX4 / VSIX4
(TRAX6/4/14 ED Eleva)
MSB CAN MSB_CAN
VA
(only for sensing) Bucky unit 2
X10 X14
and for non sensing sensing 4512 201 0256x
Failsafe X13 UI non sensing 4512 201 0257x
X12 Brake release switch
WAX3 X1 D-Sub
X11 X1
Release wall stand X2
X32
(only for non sensing) S1
not for non sensing
Failsafe

X1 MSB_CAN

X32 PH
EZ X31 X1 5Camber
Field measuring

(EZX12 ED Eleva)
Amplimat

V
Wall stand
A2/A3 07-11-28 Schr.

BuckyDiagnost VS Basic

BuckyDiagnost VS CSIP Level 0 (07.0)


C Koninklijke Philips Electronics N.V.
Z1-2
ALL RIGHTS RESERVED
SIX54 / VSIX54 I
Mains Interface basic
connection X8
X16 Multifuse 26V DC X11 VAX1
X7 F4
F3 L + 1 2,5A 15 15
230V~
L2 1 1
N2 ~
26V DC
A
switched
1A in 20,4V AC out 5,0A H1 Bucky unit 2 sensing / non sensing
N 3 3 2 8 GND 8
switched
N - =
GND

I
Interface
P basic / tracking
V
Wall box Basic
Wall stand
BuckyDiagnost VS Basic
A3 06-01-16 Schr.

BuckyDiagnost VS Basic
Power supply
BuckyDiagnost VS CSIP Level 1 (06.0) Z1-3
C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Pin assignment of Mini System Bus (MSB)
Bucky unit type
PIN Bucky unit 2 sensing (CAN) Bucky unit 2 non sensing
1 CAN GND not used
2 CAN L in not used
1 3 CAN L out not used
4 not used not used
5 not used not used
6 sync grid GND sync grid GND
7 sync grid sync grid
8 26V DC GND 26V DC GND
9 CAN 12V not used
10 CAN H in not used
11 CAN H out not used
12 not used not used
15 13 start grid start grid
14 start grid GND start grid GND
15 26V DC 26V DC

VP Wall box Basic VA Vertical frame


X10 X11 VAX1
8 8 8
15 15 15

MSB out
MSB in
9 9 9
1 1 1

male female female

VPI Bucky unit 2


Interface Basic sensing / non-sensing
A3 06-01-16 Schr.

BuckyDiagnost VS Basic
BuckyDiagnost VS Advanced
MSB connection and Pin assignment

BuckyDiagnost VS CSIP Level 1 (06.0) Z1-4


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
N2
230V
20,4VAC

X8 X16
230V VPX7

Optimus generator Multifuse


see 3Z-21 2,5A X11 X1 X16
15 +26V 15 14 F1 26V Power +5V
~ supply
EWAX3 X1 = 1,5A GND

4 constant +24V(+26V) br 1 +26VXG +5V


PO 26VRF H1
(not used)
+15V bl 6 13 Start + 13 10 Grid
U9 control

2 yw 4 14 Start - 14 12
RQSNX/ 3 7 Sync + 7 13
or CPU
+26V
CTRLX K1

1 rd 2 6 Sync - 6 11
7 vt 5 8 8 15
GND EX
ground
I VA
Interface Basic Bucky unit 2 non sensing
Interface Tracking BuF Controller

P Wall box Basic A Vertical frame basic

V Wall stand
A3 07-09-25 Schr.

BuckyDiagnost VS Basic / Advanced


Release circuit with BU 2 non sensing

BuckyDiagnost VS CSIP Level 1 (07.0) Z1-5


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
MEX M6 M6
screw screw

X11 X11
VBX11
M6 M6
screw screw
VPX11
X11 X11
Bucky unit 2
VBX11 sensing / non sensing
screw

B
M4
VAVAX13
VAVAX14 X12
VPX11 Vertical Rear
front
VAX12 unit X12
Tilting screw
unit VAX13 VAVAX12 20,4V AC
X11
0
230V
N2
X13 screw
VTX11 rear M4
VAVAX11 cover

connected
VAVAX15 VAVAX16 through
VTX12 cable base
cover frame
internal metal
sheet
screw
X16 X15

VAX11 X12
Grid frame

bottom X14 X13

P Basic Wall box cover


Cassette tray

T VA Bucky unit 2
tilting unit sensing / non sensing
(only with

BuckyDiagnost VS advanced BuckyDiagnost VS basic V


Wallstand
advanced tilt)
A Vertical frame
A3 06-01-16 Schr.

BuckyDiagnost VS
Basic / Advanced
Earthing diagram

BuckyDiagnost VS CSIP Level 1 (06.0) Z2-1


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
VPI X16:2
20,4V VPI X16:1
AC
0V

F4 5A

N2 20,4V AC
N2 20,4V AC

F3 1A
241V AC

219V AC
230V AC

0V
VPI X8:1
VPI X8:3

14 15
VA X1
(Brake release switch VAS1)

VA VA X1
MSB CAN
1

SIX4 / VSIX4
X1
MSB CAN
(only for

X11
sensing)
VPI

X10
1

15
Interface Basic
X1 D-Sub
1
9

15
VB X1

1
(VBY1 vertical brake) WA X3
Release
(only for non
(not used) sensing)
X6
H1
1
X3

X16

230V
X8

X7
Multifuse

N2 F4 0V
2,5A
F1

N2 F4 20,4V
Mains
1

230Vac
1
1

X11 VAX11
N2 F3 230V

VBX11
N2 F3 0V

MEX
A4 06-01-16 Schr.

VP
BuckyDiagnost VS Basic
Wall box Basic
4512 133 2903x

BuckyDiagnost VS CSIP Level 1 (06.0) Z2-2


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
VP
VP X11
IX3
X1.2
X1 X11
X1.1

X1 X11

X1

X1.1 X1.2
rd blk

bl blk
A4 06-01-16 Schr.

VB
BuckyDiagnost VS Basic
Vertical unit

BuckyDiagnost VS CSIP Level 1 (06.0) Z2-5


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
S1 X1
Brake release VPIX1
switch

X1 VPIX1

S1 X1.1 X1.2
1 blk 1

2 rd 2

14
A4 06-01-16 Schr.

15

VA
BuckyDiagnost VS Basic
Vertical frame

BuckyDiagnost VS CSIP Level 1 (06.0) Z2-6


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
F1
Multifuse 2,5A
X1 1

X3
1

X16
1
H1 9

1
X1 D-Sub
1
15
X8 X11
1

1
X6
1
X7 not used
15
X10

230V
A4 06-01-16 Schr.

VPI
BuckyDiagnost VS
Interface basic
4512 133 0204x

BuckyDiagnost VS CSIP Level 1 (06.0) Z3-1.1


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
X7
Mains 3
connection 2 X3 VBY1 Vertical
230Vac 1 2 Brake
1
X8 3
VPN2 3
2
1 0V GND

PO 26V X1
R7 15
VAS1
X16 R2 V9 User Interface
1 F1 VDR
V10 Brake release
R8 14
VPN2 V1 V2 switch
Multi fuse 2,5A
20,4Vac 2 100n
H1 R1
C8 C9

0V GND

X6
2 1 CAN termination
Jumper
(not used)

R9 X10
X11
1 CAN GND 1
2 CAN LIN 2
3 CAN LOUT 3
4 4
5 5 SIX4 / VSIX4
6 Sync - Sync- 6
Sync + Sync+
(TRAX6/4/14 ED El)
7 7
8 0V GND (24V GND) 8 MSB
9 9
(only for sensing)
CAN +12V
10 CAN HIN 10
11 CAN HOUT 11
VAVA X1
MSB 12 12
13 Start + Start + 13
Bucky unit 2
14 Start - Start- 14
PO 26V +24V dcP 15
15
X1
D-Sub
PO 26VXG (not used) 1
Sync - 2 WAX3
Sync + 3 Release
Start - 4 Wallstand
24V GND 5 (only for
Start + 6 non sensing)
A4 06-01-16 Schr.

not used 7
not used 9

VPI
BuckyDiagnost VS
Interface basic
4512 133 0204x

BuckyDiagnost VS CSIP Level 1 (06.0) Z3-1.2


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Drawings/Diagrams VS advanced BuckyDiagnost VS

DRAWINGS/DIAGRAMS VS ADVANCED
Contents
Survey of units BuckyDiagnost VS ...................................................................................................................Z1-1
BuckyDiagnost VS / Labeling .........................................................................................................................2Z-10
Conversion of measuring chamber ..................................................................................................................Z2-9
BuckyDiagnost VS Advanced.........................................................................................................................Z1-10
BuckyDiagnost VS Advanced tilt ....................................................................................................................Z1-11
Power supply BD VS advanced .....................................................................................................................Z1-12
Safety circuit BD VS advanced tilt ..................................................................................................................Z1-13
MSB connection and Pin assignment BD VS basic_advanced ........................................................................Z1-4
MSB connection and Pin assignment for BDVS advanced tilt .......................................................................Z1-14
Earthing diagram BD VS basic_advanced ......................................................................................................Z2-1
VP BuckyDiagnost VS advanced / Wall box tracking .......................................................................................Z2-3
PCB Interface tracking for replace VE sensing_tracking ..................................................................................Z2-4
Vertical unit BD VS advanced .......................................................................................................................Z2-5.1
VA BuckyDiagnost VS Advanced tilt digital tilt Vertical frame .......................................................................Z2-6.1
Tilting unit BD VS advanced tilt ........................................................................................................................Z2-8
PCB Interface tracking BD VS.......................................................................................................................Z3-2.1
PCB Interface tracking BD VS.......................................................................................................................Z3-2.2
VPI BuckyDiagnost VS Interface Wall ...........................................................................................................Z3-3.1
VPI BuckyDiagnost VS Interface Wall ...........................................................................................................Z3-3.2
VP1 BuckyDiagnost VS Wall stand controller vertical ...................................................................................Z3-4.1
VP1 BuckyDiagnost VS Wall stand controller vertical ...................................................................................Z3-4.2
VP1 BuckyDiagnost VS Wall stand controller vertical ...................................................................................Z3-4.3
VP3 DigitalDiagnost VM, VM2, BuckyDiagnost VS, 5 Field control ..............................................................Z3-6.1
VP3 DigitalDiagnost VM, VM2, BuckyDiagnost VS digital, 5 Field control ....................................................Z3-6.2
VAG DigitalDiagnost VM, BuckyDiagnost VS, User interface keyboard .......................................................Z3-8.1
VAG DigitalDiagnost VM, BuckyDiagnost VS, User interface keyboard .......................................................Z3-8.2

4512 984 28037 REV A CSIP Level 1 (19.0) 10 - 1


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS Drawings/Diagrams VS advanced

10 - 2 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
VP
VB
VP
VB

VT

VA

VAG2
VA
VAPH
VAG1
VA = Vertical frame
VAG1 = User interface left
VD VAG2 = User interface right
VAPH = 5-Field measuring chamber
VAVA = Bucky unit 2 digital
(BuckyDiagnost VS digital)
VAVA
VAVA Bucky unit 2 sensing/ non sensing
(BuckyDiagnost VS basic/ advanced)
VAPH VB = Vertical unit
VD = Display
VP = Wall box
VT = Tilting unit

BuckyDiagnost VS basic BuckyDiagnost VS advanced


BuckyDiagnost VS digital
A3 09-10-21 Schr.

BuckyDiagnost VS
Survey of units

BuckyDiagnost VS CSIP Level 1 (19.0 Z1-1


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
3) "Wall Box"
High End
Wallstand
Segment Control Unit M 3) I
(Optional, if no table is present!
Wall Box
BuckyUnit 2 sensing / non sensing
Only for DigitalDiagnost 1.x, not
necessary for BuckyDiagnost VS digital)

3) B Logo Address
BuckyUnit 2 digital Typen No. Electr. data
IEC60601-1 IEC60601-2-32

I FDA
UL / CSA IEC
Logo Address
WEEE
CE Environmental
Typen No. Electr. data

Typen No.
IEC60601-1 IEC60601-2-32

Logo
H UL / CSA
FDA
IEC

Barcode

UL / CSA
WEEE
CE Environmental

Typen No.
Logo
M

UL / CSA
Barcode
E
Outside
F G Inside

3) E Vertical unit G
with BuckyDiagnost VS Basic G
with BuckyDiagnost VS advanced /
advanced tilt
3) H

Barcode
Address
Family PEI Address

Familie Pei

Typen No.
Typen No. Barcode
Package 1) Address
Typen No. Barcode

Logo Address
Typen No. UL / CSA
A2/A3 08-06-04 Schr. / Kö

3) F 1) Package VS basic
Package VS advanced
2) Package VS advanced tilt
Package VS advanced tilt replacement BuckyDiagnost VS
Package VS basic replacement Package VS digital tilt
Tilting unit Package 2) Address Package VS advanced replacement Labeling
3) Necessary for BuckyDiagnost VS digital
Typen No. Barcode Package VS digital

4512 983 11021 CSIP Level 0 (b/08.0) 2Z-10


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS BuckyDiagnost VS

Left hand Right hand


operation operation
(default)

Hardware Programmings

BuckyDiagnost TH2/TF (MSB)


BuckyDiagnost VS basic
1 2 1 2
S1 in position 1 5 Field measuring chamber S1 in position 2 S1
S1 (default)

BuckyDiagnost VS advanced / digital


5-Field control
Position 1-2 Jumper VP3X2 Position 2-3
(default for VS)

Software Programmings
Bucky Controller (X-Scope)

Program Manual / Bucky RAD / Wallstand / Basic Setup /

Left (default) Bucky Device Orientation Right


A3/A4 06-06-13 Schr.

BuckyDiagnost TH2/TF (MSB)


Conversion of measuring chamber

4512 983 11361 CSIP Level 0 (06.0) Z2-9


C 2004 Philips Medical Systems
P B
Wall box tracking Vertical unit
4512 133 2897x Column manual 4512 201 0271x
Column manual with tracking 4512 201 0272x

A Vertical frame tracking


0V N2 20,4V AC 4512 201 0276x (including wall box tracking)
230V AC

I Interface
Tracking

SIX54 / VSIX54
X8
Mains connection X7 230VAC
230VAC
X1 User-
SZ1 X5 Interface
BuCo X5 X3 Y1 Brake
G1
(only with left
option tracking) X1
X4

Vertical potentiometer
X4 R (only for option tracking)
SIX4 / VSIX4 X16 26VDC
MSB CAN
(only for sensing)
and for non sensing
MSB_CAN VA
X10 Bucky unit 2
Failsafe X1 X3
G sensing 4512 201 0256x
User-Interface non sensing 4512 201 0257x
X2 keyboard
WAX3 X1 D-Sub
Release wall stand X4
X11
(only for non sensing)
X32
not for non sensing
Failsafe

User-
Interface

G2
right

X1 MSB_CAN

EX X1
X33
Amplimat Display X35 PH
X32 X1 5Camber
Field measuring

EZ X31
Amplimat X31 VP3 X34

5 - Field control

V
Wall stand
A2/A3 07-11-28 Schr.

BuckyDiagnost VS Advanced

BuckyDiagnost VS CSIP Level 0 (07.0)


C Koninklijke Philips Electronics N.V.
Z1-10
ALL RIGHTS RESERVED
P B Vertical unit T Tilting unit A Vertical frame tilting / tracking
4512 201 0280x (including wall box tilting)
Column manual 4512 201 0271x
Wall box tilting Column manual with tracking 4512 201 0272x 4512 201 0280x
4512 133 2898x Stretch grip

S1 S2

L L1 L 230V AC
(optional)
Mains
N
filter
N
0V
N5 26V DC

X20

Safety switches contour


X30 S3 S4 S5 S6
X14
I Interface
Wall
SIX54 / VSIX54
(NPX2 ED Eleva) X7
230VAC
X8
VT1
Mains connection
230VAC
X1 Tilt User-
Interface
X3 Y1 Brake controller G1
left
X2
X4
X5 Tilting potentiometer
Vertical potentiometer
X4 R (option tracking) R
X16 26VDC X9
X17
SIX4 / VSIX4
(TRAX6/4/14 ED Eleva)
MSB CAN MSB_CAN
VA
Bucky unit 2
(only for sensing) X10 M X2 X1 X3
G sensing 4512 201 0256x
and for non sensing X17
non sensing 4512 201 0257x
Failsafe Tilting motor User-Interface
X2 keyboard
X12
X4
WAX3 X1 D-Sub X11
Release wall stand X6 Brake
(only for non sensing)
not for non sensing
Failsafe X1Power

X3 MSB_CAN User-
Interface

G2
X21
right
X8

X1 MSB_CAN

EX X1
X33
Amplimat Display X35 PH
X32 X1 5Camber
Field measuring

EZ X31
(EZX12 ED Eleva)
X31 VP3 X34

Amplimat 5 - Field control

V
Wall stand
A2/A3 07-11-28 Schr.

BuckyDiagnost VS Advanced tilt

BuckyDiagnost VS CSIP Level 0 (07.0)


C Koninklijke Philips Electronics N.V.
Z1-11
ALL RIGHTS RESERVED
I
Mains Interface wall
X16 X17 X1
connection X7 X8 L1 (VT)1
F1 L + F2 26V DC 1 1 26V DC 1 Controller
230V~
L2 1 1 L L/ 230V AC N5 tilt movement T
in out 5A in 26V DC out 12,5A (optional)
switched N 3 3 0V~ switched GND 2 2 GND 2 Tilting unit
- (optional)
N N/ N
M3 M3 Mains
filter

X11 VAX1
15 26V DC 15
A
P 8 GND 8
Bucky unit 2 sensing / non sensing
Wall box tracking
I
Interface wall
Interface Tracking V
Wall stand
BuckyDiagnost VS Advanced
A3 06-01-17 Schr.

BuckyDiagnost VS Advanced
BuckyDiagnost VS Advanced tilting
Power supply
BuckyDiagnost VS CSIP Level 1 (06.0) Z1-12
C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
VPN5 26VDC
2) for more details see VPI Interface Tracking Z3-2.2
VPIX16:1
VPIX1:1

X1:1

X2:1 X17:14 X6:1

VPIX2
2)
VPIX3
W10 VTM VAG1,2 VTY1
1
2
M
Tilting Handswitch Brake
3 motor brake release
VPIX2 2)
1) Safety switches contour VPIX3
can be activated from
front side and rear side X2:2 X17:15 X6:2
of the tilting unit.

2 K2
X30:1 X14:1 3
K3 5
1
2
S3 7
1
W9
1
S4
Safety 2 2
switches
contour 1) 1 3 W13
S5 3
2
2
1
S6X30:2 X14:4
1
2
X20:1
W12 VT1
1 10
1
Stretch grip
S1 2 Tilting K2
2 11
switches
1
3 controller
S2
2 X20:2
Safety
relay
K4 tilting
5

VA 3 D1
D1 Brake
Safety
deactivation signal
relais
out signal
X1:2

VPIX1:2

VPN5 GND
A4 06-01-17 Schr.

BuckyDiagnost VS Advanced tilting


Safety circuit

BuckyDiagnost VS CSIP Level 1 (06.0) Z1-13


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Pin assignment of Mini System Bus (MSB)
Bucky unit type
PIN Bucky unit 2 sensing (CAN) Bucky unit 2 non sensing
1 CAN GND not used
2 CAN L in not used
1 3 CAN L out not used
4 not used not used
5 not used not used
6 sync grid GND sync grid GND
7 sync grid sync grid
8 26V DC GND 26V DC GND
9 CAN 12V not used
10 CAN H in not used
11 CAN H out not used
12 not used not used
15 13 start grid start grid
14 start grid GND start grid GND
15 26V DC 26V DC

VP Wall box Basic VA Vertical frame


X10 X11 VAX1
8 8 8
15 15 15

MSB out
MSB in
9 9 9
1 1 1

male female female

VPI Bucky unit 2


Interface Basic sensing / non-sensing
A3 06-01-16 Schr.

BuckyDiagnost VS Basic
BuckyDiagnost VS Advanced
MSB connection and Pin assignment

BuckyDiagnost VS CSIP Level 1 (06.0) Z1-4


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Pin assignment of Mini System Bus (MSB)
Bucky unit type
PIN Bucky unit 2 sensing (CAN) Bucky unit 2 non sensing
1 CAN GND not used
2 CAN L in not used
1 CAN L out not used
3
4 not used not used
5 not used not used
6 sync grid GND sync grid GND
7 sync grid sync grid
8 26V DC GND 26V DC GND
9 CAN 12V not used
10 CAN H in not used
11 CAN H out not used
12 not used not used
15 start grid start grid
13
14 start grid GND start grid GND
15 26V DC 26V DC

VPI VP Wall box Basic


Interface Wall
CAN
terminator X12 X3
8 8
15 15

W15
default
(without
9 9
CAN terminator) 1 1

female male

X10 X11
8 8
15 15

MSB out
MSB in
9 1
9 Tilt
VT1 controller
1

male female

VA 8
15
VAX1
Vertical frame
Bucky Unit 2
A3 06-01-17 Schr.

1
9
sensing / non-sensing

female

BuckyDiagnost VS Advanced tilting


MSB connection and Pin assignment

BuckyDiagnost VS CSIP Level 1 (06.0) Z1-14


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
MEX M6 M6
screw screw

X11 X11
VBX11
M6 M6
screw screw
VPX11
X11 X11
Bucky unit 2
VBX11 sensing / non sensing
screw

B
M4
VAVAX13
VAVAX14 X12
VPX11 Vertical Rear
front
VAX12 unit X12
Tilting screw
unit VAX13 VAVAX12 20,4V AC
X11
0
230V
N2
X13 screw
VTX11 rear M4
VAVAX11 cover

connected
VAVAX15 VAVAX16 through
VTX12 cable base
cover frame
internal metal
sheet
screw
X16 X15

VAX11 X12
Grid frame

bottom X14 X13

P Basic Wall box cover


Cassette tray

T VA Bucky unit 2
tilting unit sensing / non sensing
(only with

BuckyDiagnost VS advanced BuckyDiagnost VS basic V


Wallstand
advanced tilt)
A Vertical frame
A3 06-01-16 Schr.

BuckyDiagnost VS
Basic / Advanced
Earthing diagram

BuckyDiagnost VS CSIP Level 1 (06.0) Z2-1


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
20,4V VPI X16:1
VPI X16:2
AC
0V
F4 5A

N2 20,4V AC
N2 20,4V AC

F3 1A
241V AC

219V AC
230V AC

0V
VPI X8:3

VPI X8:1

Interface
VPI Basic / Tracking

14 15

1
VA GX1 SZ1 X11
User interface BuCo

X5
(only with
option tracking)

X1

9
VB X4
Height potentiometer

1
X4
1

9
X2
EX X1

1
VP3 VA VA X1
Amplimat EZ X31
X35 MSB CAN

1
Amplimat
1

Display

X11
X32
Bucky unit 2

1
X2

15

X10
SIX4 / VSIX4

1
9

X31

X1 D-Sub
MSB CAN
1
0V GND

(only for sensing)


X7

X3

VAVA PHX1
1

15
1

1
5 Field chamber
9

WA X3
X32

1)

X40
1
X2

VB X1 Release
U4

(not used)
(VBY1 vertical brake)
GND

(only for
X33
1

X6
X5

H1

non sensing)
9

X17 1
X3
X35

VP3
X16

230V
VPI X32
X8

X7
5-Field chamber
Multifuse
10

2,5A
X34
16

F1

VAG X2
1

Mains
230Vac
1) Jumper X2 (without function on PCB version 1)
1

1 - 2 5-field measuring chamber for


left hand operation (default for VS). X11
N2 F4 20,4V

N2 F3 230V

VAX11
VBX11
N2 F4 0V

N2 F3 0V

2 - 3 5-field measuring chamber for


right hand operation MEX
A4 06-01-17 Schr.

VP
BuckyDiagnost VS Advanced
Wall box tracking
4512 133 2897x

BuckyDiagnost VS CSIP Level 1 (06.0) Z2-3


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
VPI X16:1+26V DC
VPI X16:2 0V
gr

LED
T1
N5
VB X1
N5 Y1 vertical brake
F2 VT1 X6
12,5A
Tilt controller brake signal

F1 VT1 X8
Tilt controller height poti signal
5A
br VB X4
Height potentiometer
bl

X55 X56
1
1
1

10
1
X3
4

10
L N

1
5
5
X2
1
X32

X15 X4
X5

Multifuse 1,35A
230V AC

C1
2
1
3

F1

1
5

8
C2

2
VT1 X3

4
MSB CAN
Tilt controller
L1
EX X1
SIX4 / VSIX4
X14

X11
X13

X12
Amplimat

X10
EZ X31
Display Amplimat MSB CAN
(only for sensing)
1

9
1
L/ N/
1

1
9
9

9
9

X31

VA VA X1
1
0V GND

MSB CAN
1
X7

X21

X23
1

X1 D-Sub X65

1
2

Bucky unit 2
2

VAVA PHX1 9
3
9 1

5 Field chamber
Interface wall

WAX3 Release
X32

1
1)

X20

2 1

1
1
X2

X22
2
6

(only for
VPI
U4

9
3
7

bl br
GND

non sensing)
3
4
X33

8
1

X5

R1
2

1
9

X24

X25
2
1

2
3

1
1

R3
X35

VP3
X1
2
1

VT1 X21
2

5-Field chamber 230V AC


10

X17

VT1 X1 power
X34
16

1
1

1
1

VAG X2
X7
X8

X6
2

2
2

X16

1
2

1) Jumper X2 (without function on PCB version 1)


3
3

1 - 2 5-field measuring chamber for


X11
N5 26VDC

left hand operation (default for VS).


VAX11
VBX11 Mains connection
N5 0V

2 - 3 5-field measuring chamber for


right hand operation MEX 230 Vac
SIX54 / VSIX54
A4 06-01-17 Schr.

VP
BuckyDiagnost VS Advanced tilting
Wall box tilting
4512 133 2898x

BuckyDiagnost VS CSIP Level 1 (06.0) Z2-4


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
VP
X1
1E

VP
art
h

IX
wir

VP

4
e

Po
IX

te
3B

nt
X11

io
ra

m
ek

et
e r
X1.2
X1 X4
X1.1

X1
X11

X4

Height potentiometer
bl rd blk (optional with tracking)
Upper position
2
4,5V +- 0,1V
3 1 between
pin 3-2
R 5,0V power supply
Poti between 3-1

X1

X1.1 X1.2
rd blk

bl blk
A4 06-01-17 Schr.

VB
BuckyDiagnost VS Advanced
Vertical unit

BuckyDiagnost VS CSIP Level 1 (06.0) Z2-5.1


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
1 S3:2 S4:2 1
4 4
2 S5:1 S6:1 2

VAX12 VAX13
S4 S5

VT1X20:2
4 2 S2:1 2 4
S1:2
G1 VT1X20:1 1
G2
User interface S11 User interface
left VP3:X34 S2 right
5 field cont. S1...S2 = Switches for stretch
grip detection
X2
(only for option tilting)
1 10

X3 X4
G
1 2

39 40

User
G1 interface
G2 S3...S6 = Safety switches contour
keyboard collision detection
X1 (only for option tilting)
39 40

1 2

15 1

1) Safety switches contour can be activated from


VT1:X17
front side and rear side of the tilting unit.
Wall stand
controller

VAX11

S3:1
S6:2

1 2 X30

S3 S6
VT1 X14

1 X30:1 S5:2 1
A3 06-01-17 Schr.

4 2 S4:1 X30:2 2 4

VA
BuckyDiagnost VS Advanced tilt digital tilt
Vertical frame

BuckyDiagnost VS CSIP Level 1 (06.0) Z2-6.1


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
View B

VTM Tilt motor

gr
blk
cover yw
rd
rd
blk

View A

View B

VAGX1 VTM VAS1..S2 VAX30


(User interface) Tilt motor (Stretch grip switches) (Safety contour switches VAS3...S6)

View A
1 1
1
X14
X21
X17 X20 1 1 1 7

W9 W12 3 9

X22
1
VP3 X35
on (VT)1 1
X23
5-field control VTR Tilt potentiometer
Tilt controller X16
1
(control fields)
W11
1 W13
1 1 Potentiometer adjustment :
1
2
X1
X19
BR1 1 W10 X6 In 90° position of tilting unit
D1 W8
1 1 3 4,5V+-0,1V between pin 3-2
W7 1
5,0V Power supply between pin 3-1
1
X9
W6 W5 W4 W3
1

X8 S3 1 5

S2
yw rd br
on
X3
S4 4 8
A3 06-01-23 Schr.

Bu St
1 green green green green X2
Reset 6 9 15 9 C38
1
X13
5
off W2
8 1
V54 V50 GND 26V V51 V49 (VT)1 X8

VTR VPI X5 VPIX12 VPIX17


(Tilt potentiometer) (height potentiometer (MSB CAN) (26V DC) VT
via Interface wall, for BuckyDiagnost VS advanced tilt
tracking option only)
Tilting unit

BuckyDiagnost VS CSIP Level 1 (06.0) Z2-8


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
X40 1
F1
Multifuse 2,5A
X1
X2 1 1

X3 D-Sub
X3 9 X4 1

1
X17 1

1
X32 1

X16
1 X1 D-Sub
1
H1 9

1
1 X2 D-Sub
15
X8 X11 9
1

1
X6
1
X7 (not used) 15 X10 X5
1
230V
9
A4 06-11-07 Schr.

VPI
BuckyDiagnost VS
Interface Tracking
4512 133 0203x

BuckyDiagnost VS CSIP Level 1 (06.1) Z3-2.1


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
X7 X3 VBX1 Vertical
Mains 3 Brake Y1 X6
2 1 Jumper
connection 2 2
230Vac 1 (not used)
1
3
X8 X11 R9 X10
VPN2 3
2
GND 26V DC X32 1 CAN GND 1
1 2 CAN L in 2
1 26V DC 26V DC 3 CAN L out 3
R7 4 4
2 5 5
F1 VP3 X35
6 Sync - Sync- 6
X16 Multi fuse 3 5 Field control 7 Sync + Sync+ SIX4 / VSIX4
R2 V9 7
2,5A VDR Power 26V DC (24V GND) 8 MSB
1 V10 8 GND 26V DC
VPN2 R8 GND 26V DC 9 CAN +12V 9 (only for sensing)
V1 V2 VAVA X1
20,4Vac 2 100n MSB
10 CAN -H in 10
11 CAN -H out 11
H1 R1 Bucky unit 2
C8 C9 12 12
13 Start + Start + 13
14 Start - Start- 14
GND 26V DC 26V DC +24V dcP 15
15
X1
D-Sub
PO 26VXG 1
(not used) 2 WAX3
3 Release
4 (only for
5 non sensing)
6
7
9
V8
VAG1 VAG
VAG2 X1 X1
Brake release 14 14
switch 15VDC
A3
15 15
X5
1 1 GND BuCo 26V BuCo 1
9 9 Collimator light trigger (24V DC) 9
6 6 Collimator shutter transversal close (24V DC) 8
5 5 Collimator shutter transversal open (24V DC) 7
8 8 Collimator shutter lateral close (24V DC) 5
7 7 Collimator shutter lateral open (24V DC) 4
VAG1 / VAG2 15VDC SZ1 X5
Light and shutter
control User GND BuCo 2 BuCo
X4 V5 R4
Interface
1 16 Tracking current (4...20mA) 6 (only for tracking)
15
12 14
VBR
Vertical unit 2
R3 A2 3
A1 1
+
potentiometer 4 13
(optional for 3 - 2,5,9
tracking)
A3 06-11-07 Schr.

GND BuCo GND BuCo

VPI
BuckyDiagnost VS
Interface Tracking
4512 133 0203x
BuckyDiagnost VS CSIP Level 1 (06.1) Z3-2.2
C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
C1 X10 X65 X1D-Sub
2 1 1
10 10 X1 X17 X16
1 2 F1 9 1 R1
Multifuse 1,35A 9 1
1 1 1

2
X11 2 2 2

9
9
1 1
X15 R3
1 1 8 4 2 X9
X12 X25
X56 X55 1 2 3
9
5 1
X22 1 1

X4 1 5 1 2
X23
1 2 3 3 X7
X13 X20
X5 1 5 1 2 1 2 3
9
5 6 7 8 X24
1 1 2 3 4 1
X2 1 4 2
X14 1 2 3 X8
9 1 2 3
X3 1 4 1 3

1 X21
X32
C2 230V AC

Jumper X15
8 4

Without CAN termination (for BuckyDiagnost, DigitalDiagnost 1.x)


5 1

8 4

With CAN termination (for DigitalDiagnost 2.x)


5 1
A4 08-04-17 Schr.

VPI
BuckyDiagnost VS
Interface wall
4512 133 0205x

BuckyDiagnost VS CSIP Level 1 (08.0) Z3-3.1


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
without CAN termination (default)
Jumper X15
8 4
X15 8 7 6 5 4 3 2 1 Without CAN termination (for BuckyDiagnost, DigitalDiagnost 1.x)
5 1

1 8 4

100R
With CAN termination (for DigitalDiagnost 2.x)
R2 5 1
2
X65
1
CAN termination
2 CAN L in X10 X11 X12 X13 X14
3 CAN GND CAN GND 1 1 1 1 1
4 CAN L in 2 2 2 2 2
CAN 5
CAN L out
3 3 3 3 3
(not used)
6
not used
4 4 4 4 4 X1 Vertical controller
7 CAN H in 5 5 5 5 5 2 VP1:X1
not used X16 1 26V DC
8 sync grid GND 6 6 6 6 6
VPN5 2 GND (only VS digital,
9 CAN 12V sync grid 7 7 7 7 7 26V DC 1 26V VS digital tilting)
not used 8 8 GND 26VDC 8 8 8
CAN 12V 9 9 9 9 9
2 C2 X17 Tilting controller
CAN H in 10 10 10 10 10 2 VT1:X1
X1 (D-Sub) 11 11 11 11 11 1 100n 26V DC
CAN H out 1
1
12 12 12 12 12
not used
2 sync grid GND 1
start grid 13 13 13 13 13 F1 1,35A Multifuse
3 sync grid
Release 4
start grid GND 14 14 14 14 14
X32
start grid GND 2
Grid 15 15 26VDC 15 15 15 1
5 not used
Exposure 2
6 not used
(only for start grid
3
non sensing) 7
MSB CAN
MSB CAN MSB CAN MSB CAN MSB CAN
8 BD TH SIX4 VAVA:X1 VT1:X3 VP1:X3 VP4:X1
9 DD or SCU VSIX4 Bucky unit 2 digital BD VS Wall stand BD VS Wall stand BD VS User interface
Bucky unit 2 sensing controller tilting controller vertical Remote control
Bucky unit 2 non sensing (only VS digital, (only VS digital,
VS digital tilting) VS digital tilting)

X21
1 20,4V ACP
Connection height potentiometer RS 485 Connection
Mains supply VPN2
2
X7 X8 X4 X5 X55 X56 20,4 VAC 3
** 1 Poti +10V 1 1 1
20,4V ACN
X20
1 L1 1 GND 1
Mains 2 2 VPL1 2 Poti slider 2 2 2 X24
VBR 2
connection Mains 3 Poti GND 3 VT1:X8 3 3
1 9,26V ACP
3 N 3 PCB X4 3
230V AC filter 4 4 VPN4 2 9,26V ACN
4 RXD 485 B 4 Pixium Contr. 9,26 VAC 4
5 screen 5 RS485 5 5 Board X23 5 PCB X2
6 6 1 9,26V ACP1
RXD 485 A
VPN3 2 9,26V ACN1
6 Pixium Contr.
7 TXD 485 B 7 Board
X9 Connection brake
9,26 VAC 7
3 * 8 TXD 485 A 8 X22
A3 08-04-17 Schr.

X3 X2 9 9 1
8
2 VPL2 1 1
+12Volt 24V DC
Brake +
1 Mains 10 10 2 GND 24VDC
filter VBY1:X1 2 Brake - 2 VT1:X6 VPN1
24 VDC 3
3 3
R3 VPI
4 4 X25
* Via mains filter VPL2 to VPN1, VPN3, VPN4 Charging 1
Multifuse BuckyDiagnost VS
2 1
** Via mains filter VPL1 to VPN2, VPN5 station 2 Interface wall
Remote
only for BuckyDiagnost VS Advanced tilting only for BuckyDiagnost VS digital, VS digital tilt
control
3 4512 133 0205x

BuckyDiagnost VS CSIP Level 1 (08.0) Z3-3.2


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Jumper Function Default
W1 EPROM type 29F040 2-3, 4-5
W2 1-2 -> RAD system 1-2
W3 Motor current limitation (default value 13A) 1-2
W4 Motor current limitation (default value 13A) 1-2
X3 X13 W5 Motor current limitation (default value 13A) 1-2
V51 V49 26V GND V53 V54 1
off
1 W6 Motor current limitation (default value 13A) 1-2
C38 Reset
X2 yellow yellow yellow yellow W7 Only available on PCB version 02004 open
S4 W8 Only available on PCB version 02004 open
4 on
8 W2 S2 W9 used for safety circuit at optional tilting c ontroller VT1 (contour switches VAS3…6),
1-2
1 must be set to 1-2 at vertical c ontroller VP1 to activate safety circuit.
Service
W10 1-2 -> Torque safety switch VBS by-pass ed (test), 2-3 Torque s afety s witch ac tive 2-3
switch 1.)
S3 W11 EPROM type 29F040 1-2
5 1 X8 used for safety circuit at optional tilting c ontroller VT1 (stretch grip s witches
Service W12 VAS1,2) must be set to 1-2 at vertic al c ontroller VP1 to activate safety circuit. 1-2
button 2.) 1
W55 X55 W13 only used with DigitalDiagnos t VM , VS default 1-2 ! 1-2
W6 W5 W4 W3 W55 only used with DigitalDiagnos t VM , VS default 1-2 ! 1-2
1 1 X9
K4 1
1 W7 1
5 3 7 1 X1 Current limiting
8 6 4 2 1 VS digital tilting VS digital
1 W8 D1 (vertical and tilting motorized) (only vertical motorized)
W10 BR1 BOOT off off
2 1 CAN0 off off
X19 11 12 K2
CAN1 off off
1 W1
W11 CAN2 off off
1 1 1 W13 CFG0 off off
X6 X16 X50 1 1
1 D3 D2 CFG1 on off
7 5 13 1 K3
VP1 S1 1 CFG2 off off
X23 Vertical controller HIGH LOW on CFG3
CFG2
CFG3 off off
CFG1
X22 CFG0
CAN2
1 CAN1 off on flashing
CAN0
1 BOOT V9 (yellow) Service Mode inactive Service Mode active -
1 Selftest failed or safety circuit
X5 V10 (red) Selftest ok During selftest
9 3
W9 W12 1
V12 green open
X60 9
V9 yellow V11 (yellow) - CAN bus not connected CAN bus active
7 1 1 1 V10 red V12 (green) Controller not ready Controller ready -

X21 X20 X17 V11 yellow


V49 (yellow) Power PO26V Mot not ok 26V power supply motor VBM ok -
X14 1
26V power supply vertical brake
1 1 V51 (yellow) Power PO26V BRK not ok
VBY1 and clutch VBY2 ok
-
26V power supply safety circuit /
V53 (yellow) Power PO26V EL not ok
wall stand controller VP1 D1 ok
-
26V power supply wall stand
V54 (yellow) Power PO26V DIS not ok
display VD ok
-

1.) Service switch S2


Off: during normal operation (default)
On: vertical motorized movement possible in case of wrong or missing potentiometer value VBR (fault finding) or
mode to set lower and upper potentiometer value for wall stands without CAN bus in old world systems (adjustment) see 2.)
A3 09-12-07 Schr./Schu

2.) Service button S3


Storage of potentiometer value VBR for wall stands without CAN bus in old world systems (precondition: S2 On)
1. move to lowest working position of vertical unit and press S3
2. move to uppermost working position of vertical unit and press S3

VP1
BuckyDiagnost VS
Wall stand controller vertical
4512 133 02004/5

BuckyDiagnost VS CSIP Level 1 (a/09.2) Z3-4.1


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
X23 PO 26V MOT Motor final stage X2
PO26V
1 EL Mot + 1 rd
EN VBM
X10 M Vertical
2 P2:12 DIR C38 4
U8 3 Mot - 2 blk motor
U13 PO 0V
3 PWN
P2:10 250 4400 1000 500 200 5

PO 26V
Fault drive K4
PO 0V
X22 BRK
current +5V
PO 26V 1 2 3 1 2 3 1 2 3 1 2 3
1 limiting
BRK 3

2 Brake
V36 W6 W5 W4 W3
P7:7 PO 0V
3 Encoder IN 4 rd
PO 0V PO 15V HI
1 2 3 W7 1) 5 blk

P2:14
A Kanal PO 0V 6 yw Hall sensor
vertical drive

U15
1 2 3 W8 1)
X17 User interface IN P2:15
B Kanal 7 gr

5 Column A screen 8
P2:1
6 Column B P2:2
D1
7 Column C U16 Analog IN X8 VBX4
P2:3 1 br 1
VREF
VAG X1 8 Column D P2:4 2 blk 2 10K VBR
User interface AIN 0 Potentiometer
(only VS digital) 9 Column E P2:5 PO 26V MOT 3 rd 3
Temp Amp. screen 5
P5:2 PO 0V
PO 0V
P5:0 X9
PO 26V EL 1
VREF 1) Only available on
26V 14 vertical PCB version 02004
Torque Pot tilt 2
VAG1, VAG2
Handswitch
1 2 3 W10 P5:3
3
user interface 15
Hand switch
screen 5
User interface OUT PO 0V

X50 X17
2 Row 1 1
P8:0
4
Row 2 2
P8:1
Row 3 3
X55 W55 P8:2 U19
1 1
only VM Row 4 4
P8:3
(safety circuit) 2 2 default
Row 5 13 VAGX1
3 P8:4
User interface
K3 (only VS digital)
X6 PO 26V EL
1

2 1 3 13 UI LED A 10
Torque (VBS) P8:5
PO 0V
safety switch 4 U20 UI LED B 11
VBX1 2 5 7 P8:6
V20
Brake 2 PO 26V BRK Brake vertical UI LED C 12
VBY1 1 1
W13 P8:7
PO 0V
screen 16
3
+5V
X16 2
Safety circuit br 2
connection 1
VT1X16 rd 4

9 2 10 1 K2 V9 V10 V11 V12


X14 PO 26V
8 3 11 12 EL
1 P4:7
PO 0V PO 0V
2
1
P6:5
3 D12 Safety relais out
2 P6:6 +5V
4 2 K4
3 PO 0V P6:7
X20 W9 1 2 3 W12
1
Reset
RST IN +5V
2
S4
A1/A3 09-07-06 Schr.

Stretch grip
1
Contour PO 0V
3
P5:11 +5V
S2
VBX2
X19 PO 0V

1
PO 0V
Clutch V21
VBY2 3 P5:12
Clutch
S3
PO 0V PO 0V

VP1
BuckyDiagnost VS
Wall stand controller vertical
4512 133 02004/5
BuckyDiagnost VS CSIP Level 1 (a/09.0) Z3-4.2
C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
5V 21V 15,3V
1
N3 1) N1 N2
X2 U in
V17 U4
V18
X35 X34
1 10 1
X31 4 4 4
1 U10 1 U8 1 U9
X5
1 X32 GND
9 4
U12
1
9
1
X33
4
U11 1
1 6
4
15
8 U13
16
1
X7 8
15 4 9
U14 5
1
0V GND
1) Jumper X2 (without function on PCB version 1)
1 - 2 5-field measuring chamber for
left hand operation (default for VS).
2 - 3 5-field measuring chamber for
right hand operation (default for VM)
A4 08-12-09 Schr.

VP3
DigitalDiagnost VM, VM 2
BuckyDiagnost VS
5 Field control
4512 133 0210.
BuckyDiagnost VS CSIP Level 1 (08.0) Z3-6.1
C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
PO 26V PO 26V
BRK DIS X13
X21 Serial I/O 1
PO 26V
1 BRK D13
V34 3
2 Lock P3:10

7
P7:6 P6:4
PO 0V 2 VD
P3:11 Display
3 8
P6:3
P3:15
VP3 X35 PO 0V 5
4
5 Field P3:0
measuring chamber PO 0V 10
5
in / out P3:1
screen 11
6
U1 P3:2
U9 +5V
7 X60
P3:3
1
8 P3:13
P3:4 2
U26 3
9
P3:5 P3:8 4
+5V
Interface encoder
PO 0V V64 V63 V30 V62 (VM only)

PO 0V 5
8
CAN / SAN +5V D1 9
PO 5V CAN uP + PROM + RAM + A/D PO 0V
X5
1

B2 2
X3 RXD CAN 3
+5V W11
10 29F040 27C4001
P3:12 PO 0V
1 1 1
11 CAN H +5V
2 2 2
D11 P4:3 3 3
2 3
CAN L B3
TXD CAN
3 W1
1
29F040 27C4001
1 1
9 CAN V+ IN OUT PO 5V WRL
U25 +5V 2 2 2
MSB CAN CAN
+5V 3 3 3
VPI X12 1 CAN GND
4 4 4
U26 5 5 5
12 B1 STOP MV +5V
not used
6 RD
1 RAD/URF
A / D Bus
A1...18
W2 D0...7 LOW
2 D2
3

PO 0V RD
+5V PO 0V
A1...18
D0...15 RD
A1...18
D8...15

D3 HIGH

Power supply
MP2 +5V PO 0V
PO 26V
X1
br 1 1 2
VPIX17 26Vdc L3 U17 +5V

bl 2
GND S1
2 1
PO 0V PO 26V EL POL4
1
4 3
BR1 P5:5
6 5
2 P5:6
PO 26V 8 7
PO 26 V PO 26V P5:7
U18 PO 15V UI 9
MOT BRK DIS P5:8 10
12 11
P5:9
14 13
P5:10
V49 V51 V54 V53 16 15
P5:4

PO 0V PO 0V PO 0V PO 0V Dipswitch

PO 0V

VP1

A1/A3 06-04-20 Schr.


BuckyDiagnost VS
Wall stand controller vertical
4512 133 02004
BuckyDiagnost VS CSIP Level 1 (a/06.0) Z3-4.3
C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
X31 X32
1 +400V +400V 1
3 Field 3 (right) +15V Field 3 3
4 Release chamber Release chamber 4
5 +15V +15V 5
6 -15V -15V 6 VA PH X1 / VAVA PHX1
EZ X31
(EZ X12 ED Eleva) 7 Signal chamber Signal chamber 7 5 Field
from Generator 8 8 Measuring chamber
RF 0V chamber RF 0V chamber
11 Field 1 (left) Field 1 11
12 Field 2 (center) Field 2 12
13 0V XG 0V XG 13

X7 R43
VCC
R37 R40
14
VCC VCC Field 4
U10 5 U8 5 U9 5 +15V VCC Field 5 15
0V GND 5 U12
4 +15V VCC
4
X35 0V GND 4 0V GND 0V GND 4 U11
5 Control field center +15V VCC
FD2 R45 4
U13
6 Control field left +15V VCC
0V GND R44 4
4 Control field right U14

VAG User interface keyboard FD L FD R 0V GND R46 4


VAG User interface keyboard
+15V FD3
FD1 0V GND R47
VAG1 X3 VAG2 X4 X2 X34 VAG2 VAG1
User interface left User interface right FD4 0V GND User interface User interface
15 15 10 10
FD5 X34 X2 X4 right X4 X3 left X3
left select R50 7 9 6 LED Field 3 5 6 5
14 14 1 1 R19
chamber left
R51 5 5
R6 2 LED Field 1 1 2 1

right select R49


12 12 2 2 R21 8 8 10 LED Field 4 9 10 9 1 3
chamber right
2
R10 R48 5 4
Control 9 7 4 LED Field 5 3 4 3
upper select
11 11 3 3 R20 R52/53
chamber upper 6 6 8 LED Field 2 7 8 7
R8
10 10 15V DC
lower select
13 13 4 4 R22 chamber lower
R12 VP3 5 Field control X33 EX X1
1 1 FD1
2 2 FD2
3 3 1 3
FD3
X35 4 4 2
R16 FD4
VS Advanced 2 Move chamber 90°
5 5 FD5 5 4
VPI X35
R17 7 7
Interface Tracking X5
VS Advanced tilting 3
VCC X2 CH Amplimat display
VT1 X21 GND 3
Tilting controller 7 2 - 3 right hand operation of wall stand
Chamber position 0 2
R64
A3 06-09-29 Schr.

8 1 - 2 left hand operation of wall stand


Chamber position 1 1
VS digital / digital tilt
VP1 X21 9
Vertical controller GND
+15V VCC (5V)
1
R62 V25 +21V VP3
U in
N1 N2 DigitalDiagnost VM, VM2
Multi fuse R71 R75 R77/78
N3 R67 BuckyDiagnost VS digital
R65/66
0,5A
R73 R76 V18 R68 V17 5 Field control
4512 133 0210.
GND GND GND GND GND
BuckyDiagnost VS CSIP Level 1 (06.2) Z3-6.2
C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
VP3:X34
5-Field control

1 10
39 40 X2 1 2

User interface left User interface right


VAG1 X3 X4 VAG2

X1
1 2 39 40
15 1

VT1:X17 Tilting controller (VS advanced tilt / VS digital tilting)


VP1:X17 Vertical controller (VS digital)
VP2:X17 Tilt arm controller (VM)
VPI :X1 Interface Tracking (VS advanced)
A4 06-03-01 Schr.

VAG
DigitalDiagnost VM
BuckyDiagnost VS
User interface keyboard
4512 133 0206x

BuckyDiagnost VS CSIP Level 1 (06.1) Z3-8.1


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Row2 P
ColumnD..P
ColumnC..P
ColumnB..P
ColumnA..P
Row3 LED8
Row3 LED7
Row2 LED6 X2
Row2 LED5 1
2
Row2 LED4
3
4
X3 X4 5 VP3 X34
1 2 LEDA L4 1 2 LEDA L4 6 5-Field control
3 4 LEDB L5 3 4 LEDB L5 7
5 6 LEDC L6 5 6 LEDC L6 8
7 8 LEDA L7 7 8 LEDA L7 9
9 10 LEDB L8 9 10 LEDB L8 10
11 12 11 12
13 14 13 14
VAG1 15 16 15 16
VAG2 User interface right
User interface 17 18 17 18
left 19 20 19 20
21 22 21 22
23 24 23 24
25 26 ROW1 LED1 25 26
LEDA L1 27 28 ROW1 LED2 LEDA L1 27 28
LEDB L2 29 30 ROW1 LED3 LEDB L2 29 30
LEDC L3 31 32 ROW3 LED9 LEDC L3 31 32
LEDD L9 33 34 ROW4 LED10 LEDC L9 33 34
35 36 LEDA L10 35 36 LEDA L10
37 38 37 38
Brake A 39 40 Brake B 39 40

X1
Row 1 1
Row 2 2
Row 3 3
Row 4 4
ColumnA 5
ColumnB 6
Wall stand controller:
ColumnC VT1:X17 VS advanced tilt
7
VS digital tilting
ColumnD 8
VP1:X17 VS digital
ColumnE 9
VP2:X17 VM
10
11 Interface Tracking:
12 VPI :X1 VS advanced
A3 06-03-01 Schr.

13
14
15
16
VAG
DigitalDiagnost VM
BuckyDiagnost VS
User interface keyboard
4512 133 0206x

BuckyDiagnost VS CSIP Level 1 (06.1) Z3-8.2


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Drawings/Diagrams VS digital BuckyDiagnost VS

DRAWINGS/DIAGRAMS VS DIGITAL
Contents
Survey of units BuckyDiagnost VS ...................................................................................................................Z1-1
BuckyDiagnost VS / Labeling .........................................................................................................................2Z-10
Conversion of measuring chamber ..................................................................................................................Z2-9
BuckyDiagnost VS digital (with detector) .......................................................................................................Z1-20
BuckyDiagnost VS digital tilting (with detector) ..............................................................................................Z1-21
Connection Diagram BD VS digital (with BU 2 sensing) ................................................................................Z1-22
Connection Diagram BD VS digital tilt (with BU 2 sensing)............................................................................Z1-23
BuckyDiagnost VS digital Power supply.........................................................................................................Z1-24
BuckyDiagnost VS digital Safety Circuit vert. motorized ................................................................................Z1-25
BuckyDiagnost VS digital tilt Safety circuit vert. and tilting motorized ............................................................Z1-26
BuckyDiagnost VS digital MSB connection and pin assignment ....................................................................Z1-27
BuckyDiagnost VS digital Earthing diagram ...................................................................................................Z2-10
BuckyDiagnost VS Wall box digital ................................................................................................................Z2-11
BuckyDiagnost VS digital Vertical unit ...........................................................................................................Z2-12
VA BuckyDiagnost VS Advanced tilt digital tilt Vertical frame .......................................................................Z2-6.1
VT BuckyDiagnost VS digital tilt Tilting unit......................................................................................................Z2-7
Pixium control board (PCB) wiring diagramm ..................................................................................................Z3-6
VPI BuckyDiagnost VS Interface Wall ...........................................................................................................Z3-3.1
VPI BuckyDiagnost VS Interface Wall ...........................................................................................................Z3-3.2
VT1 BuckyDiagnost VS Wall stand controller tilting ......................................................................................Z3-5.1
VT1 BuckyDiagnost VS Wall stand controller tilting ......................................................................................Z3-5.2
VT1 BuckyDiagnost VS Wall stand controller tilting ......................................................................................Z3-5.3
VP3 DigitalDiagnost VM, VM2, BuckyDiagnost VS, 5 Field control ..............................................................Z3-6.1
VP3 DigitalDiagnost VM, VM2, BuckyDiagnost VS digital, 5 Field control ....................................................Z3-6.2
VP4 BuckyDiagnost VS Remote control .......................................................................................................Z3-7.1
VP4 BuckyDiagnost VS Remote control .......................................................................................................Z3-7.2
VAG DigitalDiagnost VM, BuckyDiagnost VS, User interface keyboard .......................................................Z3-8.1
VAG DigitalDiagnost VM, BuckyDiagnost VS, User interface keyboard .......................................................Z3-8.2

4512 984 28037 REV A CSIP Level 1 (19.0) 11 - 1


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS Drawings/Diagrams VS digital

11 - 2 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
VP
VB
VP
VB

VT

VA

VAG2
VA
VAPH
VAG1
VA = Vertical frame
VAG1 = User interface left
VD VAG2 = User interface right
VAPH = 5-Field measuring chamber
VAVA = Bucky unit 2 digital
(BuckyDiagnost VS digital)
VAVA
VAVA Bucky unit 2 sensing/ non sensing
(BuckyDiagnost VS basic/ advanced)
VAPH VB = Vertical unit
VD = Display
VP = Wall box
VT = Tilting unit

BuckyDiagnost VS basic BuckyDiagnost VS advanced


BuckyDiagnost VS digital
A3 09-10-21 Schr.

BuckyDiagnost VS
Survey of units

BuckyDiagnost VS CSIP Level 1 (19.0 Z1-1


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
3) "Wall Box"
High End
Wallstand
Segment Control Unit M 3) I
(Optional, if no table is present!
Wall Box
BuckyUnit 2 sensing / non sensing
Only for DigitalDiagnost 1.x, not
necessary for BuckyDiagnost VS digital)

3) B Logo Address
BuckyUnit 2 digital Typen No. Electr. data
IEC60601-1 IEC60601-2-32

I FDA
UL / CSA IEC
Logo Address
WEEE
CE Environmental
Typen No. Electr. data

Typen No.
IEC60601-1 IEC60601-2-32

Logo
H UL / CSA
FDA
IEC

Barcode

UL / CSA
WEEE
CE Environmental

Typen No.
Logo
M

UL / CSA
Barcode
E
Outside
F G Inside

3) E Vertical unit G
with BuckyDiagnost VS Basic G
with BuckyDiagnost VS advanced /
advanced tilt
3) H

Barcode
Address
Family PEI Address

Familie Pei

Typen No.
Typen No. Barcode
Package 1) Address
Typen No. Barcode

Logo Address
Typen No. UL / CSA
A2/A3 08-06-04 Schr. / Kö

3) F 1) Package VS basic
Package VS advanced
2) Package VS advanced tilt
Package VS advanced tilt replacement BuckyDiagnost VS
Package VS basic replacement Package VS digital tilt
Tilting unit Package 2) Address Package VS advanced replacement Labeling
3) Necessary for BuckyDiagnost VS digital
Typen No. Barcode Package VS digital

4512 983 11021 CSIP Level 0 (b/08.0) 2Z-10


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS BuckyDiagnost VS

Left hand Right hand


operation operation
(default)

Hardware Programmings

BuckyDiagnost TH2/TF (MSB)


BuckyDiagnost VS basic
1 2 1 2
S1 in position 1 5 Field measuring chamber S1 in position 2 S1
S1 (default)

BuckyDiagnost VS advanced / digital


5-Field control
Position 1-2 Jumper VP3X2 Position 2-3
(default for VS)

Software Programmings
Bucky Controller (X-Scope)

Program Manual / Bucky RAD / Wallstand / Basic Setup /

Left (default) Bucky Device Orientation Right


A3/A4 06-06-13 Schr.

BuckyDiagnost TH2/TF (MSB)


Conversion of measuring chamber

4512 983 11361 CSIP Level 0 (06.0) Z2-9


C 2004 Philips Medical Systems
P Wall box
4512 133 2899x

L L1 L 230V AC N5
Mains
0V 26V DC
N filter N

N2
230V AC 20,4V AC
0V
(X21) A Vertical frame
4512 201 0278x (including wall box)

User-

I Interface
Wall X8
L2
Interface

G1
ENP X3 (for DiDi 2.x)
SIX54 / VSIX54 (for DiDi 1.x) X9
L
Mains
L
230V AC
N1 left
230V AC 24V DC (X22)
Mains N filter N 0V
connection X7
230V AC X24
230V AC
N3
X23 (X23)

B Vertical unit motorized


9,26V AC
0V
X22
X21
230V AC
N4 4512 201 0273x
X25
0V
9,26V AC (X24) X1 X3
G
Filter X20 Remote control User-Interface
X2 keyboard
X8 modul X7 charging station
X1 X4
Receiver 26V DC
VP4 X4 X5
X16
X1
X4 R Vertical potentiometer
Remote

ETRB X5 (for DiDi 2.x)


control X3 X4 X1
X14 MSB_CAN
X8
X1
VA
SIX4 / VSIX4 (for DiDi 1.x)
CAN modul
Y1 Brake User- Bucky unit 2 digital
MSB CAN X10 Interface 4512 201 0366x
X13
Connector
X3
G2
X1 D-Sub X12 X6
Pins 2-3, right
Communication X55
X11
Pins 10-11 X2 P
Acquisition Work Spot (AWS) /
SUN Workstation
X56 shortened
VP1 X19 M Y2 Clutch Pixium
RSX2 / RS1X2
Vertical Detector
controller S Safety switch torque
PCB
Pixium X3 X1
D X2
X3
M M Vertical motor
control board X5
X2 RS232
Display X200 X13
X21 X17 B1 Image

X2 X4 X1
X6 Fan Power supply

Fan1 X2

Fan2 X3
C1
BUF
Fan3 X4 Controller
X1 X16
Acquisition Work Spot (AWS) / X50 MSB CAN
SUN Workstation
FPC B2 Image data
FOL
A2/A3 07-08-10 Schr.

EX X1 X35 PH
VP3
5-Field measuring
Amplimat Display X33 X32 X1 Chamber
EZ X31 Amplimat X31 X34
5 - Field control

V Wall stand

BuckyDiagnost VS digital (with detector)

BuckyDiagnost VS CSIP Level 1 (07.0) Z1-20


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
P Wall box
4512 133 2899x

T Tilting unit A Vertical frame


4512 201 0282x (including wall box)
L L1 L 230V AC
N5 4512 201 0282x
Mains 26V DC
0V
N filter N Stretch grip

S1 S2
230V AC
N2
20,4V AC (X21)
0V
(optional)

ENP X3 (for DiDi 2.x)


SIX54 / VSIX54 (for DiDi 1.x)
I Interface
Wall
X8
X9
L L2
Mains
L N1
230V AC 24V DC
(X22)
X20

Safety switches contour


0V

V
Mains connection 230V AC 230V AC N filter N
(TR X12 ED Eleva
X7 X30 S3 S4 S5 S6
X14
SAN Generator interface) X24 N3
230V AC 9,26V AC X30
X23 (X23)
0V
Wall stand
X22
X21 N4
VT1 User-
Interface

X25
230V AC 9,26V AC
(X24) Tilt controller G1
0V left
X20 X17 X1
Filter Power

B Vertical unit motorized


X8 modul X7
X1
Receiver 26V DC
VP4 X4 X5
X16
4512 201 0273x Tilting potentiometer
Remote

ETRB X5 (for DiDi 2.x)


control X3 X4 X1
X14 MSB_CAN
X1
Remote control
charging station
R
X8
VA
SIX4 / VSIX4 (for DiDi 1.x) CAN modul Bucky unit 2 digital
MSB CAN X10
X17
X1 X3
G 4512 201 0366x
(TP CPC X11 ED Eleva CAN) X13 X3 X4 R Vertical potentiometer User-Interface
X8
X1 D-Sub X12 X1
M X2 X2 keyboard
X4
Y1 Brake
Communication
Acquisition Work Spot (AWS) /
X55
X11
Tilting motor
P
SUN Workstation
X56 Pixium
RSX2 / RS1X2 X6 Detector
(TP FPC X2 ED Eleva RS485)
VP1 X2 PCB
X19 M Y2 Clutch User- Pixium X3 X1
Interface
Vertical control board X5
X2 RS232
controller S Safety switch torque G2
B1 Image
right
X3
D X2 M M Vertical motor
Display X200 X2 X4 X1
X16 X16 X6 Fan Power supply
X13 X21 safety

Fan1 X2
X3
MSB_CAN
Fan2 X3
C1
BUF
Fan3 X4 Controller
X1 X16
X50 MSB CAN
Acquisition Work Spot (AWS) /
SUN Workstation
FPC B2 Image data
(TP FPC B2 ED Eleva) FOL
A2/A3 07-08-10 Schr.

EX X1 X35 PH
Amplimat Display X33 X32 X1 5-Field measuring

VP3
Chamber
EZX31 Amplimat X31 X34
(EZ X12 Amplimat
ED Eleva) 5 - Field control

BuckyDiagnost VS digital tilting (with detector)

BuckyDiagnost VS CSIP Level 1 (07.1) Z1-21


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
P Wall box
4512 133 2899x

L L1 L 230V AC N5
Mains
0V 26V DC
N filter N

N2
230V AC 20,4V AC
0V
(X21)
A Vertical frame
4512 201 0278x (including wall box)

I Interface
Wall X8
L2
SIX54 / VSIX54 X9
L L
230V AC
N1 User-
Interface
230V AC Mains 24V DC (X22)
0V
Mains
connection X7
N filter N G1
left
230V AC X24
230V AC
N3
X23 9,26V AC (X23)

X22
X21
0V

N4
B Vertical unit motorized
4512 201 0273x
230V AC
X25 9,26V AC (X24)
0V
Filter
Remote control
X8 modul X7 charging station
X1 X3
Receiver 26V DC X1 G
X16
VP4
Remote
X4 X5 X17
X1
X4 R Vertical potentiometer User-Interface
X2 keyboard
control X1 X8
X3 X4 X4
X14 MSB_CAN X1
SIX4 / VSIX4 CAN modul
Y1 Brake
MSB CAN X10
X13 X3
Connector X6
X1 D-Sub X12
Pins 2-3,
X11
Pins 10-11 X2 User-
shortened
VP1 X19 M Y2 Clutch Interface

G2
Vertical
controller S Safety switch torque right

X3
X2 M M Vertical motor
X21 X17

VA
Bucky unit 2 sensing
4512 201 0256x

X1 MSB CAN
A2/A3 06-01-18 Schr.

EX X1 X35 PH
5-Field measuring
Amplimat Display
EZ X31 Amplimat
X33
X31
VP3 X32
X34
X1 Chamber

5 - Field control

V Wall stand

BuckyDiagnost VS digital
with Bucky unit 2 sensing

BuckyDiagnost VS CSIP Level 1 (06.0) Z1-22


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
P Wall box
4512 133 2899x

T Tilting unit A Vertical frame


4512 201 0282x (including wall box)
L L1 L 230V AC
N5 4512 201 0282x
Mains 26V DC
0V
N filter N Stretch grip

S1 S2
230V AC
N2
20,4V AC (X21)
0V
(optional)

SIX54 / VSIX54 I Interface


Wall
X8
X9
L L2
Mains
L N1
230V AC 24V DC
(X22)
X20

Safety switches contour


Mains 0V
connection X7 230V AC N filter N

V S3 S4 S5 S6
X30
230V AC
X14
X24 N3
230V AC 9,26V AC X30
X23 (X23)
0V
Wall stand User-
X22
X21 N4
VT1 Interface

X25
230V AC 9,26V AC
(X24) Tilt controller G1
0V left
X17 X1
Filter Power
X8 modul X7

VP4 X4
Receiver
X5
X16 26V DC
X1
B Vertical unit motorized
4512 201 0273x Tilting potentiometer
Remote
control X3 X1 R
X4 Remote control
X14 MSB_CAN charging station X8
SIX4 / VSIX4 CAN modul X1
MSB CAN X10 X3
X17
X1 G
X13 X3 X4 R Vertical potentiometer User-Interface
X8 X2
X1 D-Sub X12 X1
M X2 keyboard
X4
Y1 Brake Tilting motor
X11

X6

VP1 X19
X2
M Y2 Clutch User-
Interface
Vertical
controller S Safety switch torque G2
X3
right VA
X2 M M Vertical motor
Bucky unit 2 sensing
4512 201 0256x
X16 X16
X21 safety

X3
MSB_CAN

X1 MSB CAN
A2/A3 06-01-19 Schr.

EX X1 X35 PH
Amplimat Display X33 5-Field measuring
X32 X1 Chamber
EZX31 Amplimat X31 VP3 X34

5 - Field control

BuckyDiagnost VS digital tilting


with Bucky unit 2 sensing

BuckyDiagnost VS CSIP Level 1 (06.0) Z1-23


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
P X1 X1
Wall box digital 1 1
(VP)1
+26V DC Controller
vertical
X16 2 GND +26 V DC 2
movement
230V AC
F1 L
N5 + F2 +26V DC
1

5,0A in 26V DC out 12,5A


22...32V 2
0V~ GND +26V DC X17 X1
- (VT)1
N 1 +26V DC 1 Controller T
tilt movement
2 GND +26 V DC 2 (optional) Tilting unit

A Bucky unit 2 digital


X6 X1 C1
1 1 BUF Controller
I GND +15V DC
2
Interface wall 3 M Fan
4 2
+15V DC
I
X6 X1 C1
BUF Controller
Interface wall GND +24V DC
1 1
X8 L1 X21 2
1 F3 L + F4 1 M Fan
L L/ 230V AC
N2 20,4V ACP 3
Mains
4 +15V DC 2
connection X7 in out 1,0A in 20,4V AC out 5,0A +24V DC
1) 2) 3 0V~ 20,4V ACN 3
230V AC L3 /L2 1
switched N N/ N - X20 X2 X3 X1
230V AC L3'1)/L2'2) Mains 1 20,4V ACP 1 1 3
2 GND +24V DC
continious filter
power 2 20,4V ACN 2 2 8
N 3 +24V DC
X9 L2 X24
2 L L/
F5 L + F6 1 3 9,26V ACP 3
230V AC
N4 9,26V ACP 3 Sense +24V DC 9

in out 0,5A 9,26V AC out 3,2A 2 4 9,26V ACN 4


in 9,26V ACN 4 Trigger 4
3 0V~ continuous
N N/ power -
N Power
X23 in
F7 L + F8 5
230V AC N3 9,26V ACP1 1 9,26V ACP1 5 Power 5 Trigger 5 Power
in
0,5A in 9,26V AC
continuous out 3,2A 9,26V ACN1 2 6 6
out
6 2
0V~ 9,26V ACN1 GND +5V DC
power -
N
X22
L N1 + 1 7 24V DC 7 7
A3 08-06-19 Schr.

230V AC +24V DC 7 +5V DC


24V DC
in
0V~ continuous out GND +24V DC 2 8 GND +24V DC 8 8 GND +5V DC 1
N power - PCB P
9
V Pixium control
board
9 -5V DC 6
Pixium
detector
Wall stand

1) only for DigitalDiagnost 2.x


2) only for DigitalDiagnost rel. 1.x
BuckyDiagnost VS digital
BuckyDiagnost VS CSIP Level 1 (08.0) Power supply
C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Z1-24
VPN5 26VDC

VPIX16:1

VPIX1:1

X1:1

9
K2
8
X6:3 X2:1 X17:14 X6:1 X19:1
W10
1
2
VBS VBM VAG1,2 VBY1 VBY2
3 Torque M Vertical
switch motor Brake
Handswitch Clutch
brake release

X6:4 X2:2 X17:15 X6:2 X19:3

2 K2
3
K3 5

W9 7 W13
1
3
2
2
3
1

W12
1 VP1 10
2 Vertical controller K2
11
3

Safety
relay
K4 vertical
5

D1
3 D1 D1
Brake Clutch
A4 06-01-18 Schr.

Safety relais deactivation signal activation signal


out signal
X1:2

VPIX1:2 BuckyDiagnost VS digital


Safety circuit
VPN5 GND Vertical motorized (without tilting)

BuckyDiagnost VS CSIP Level 1 (06.0) Z1-25


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
VPN5 26VDC VPN5 26VDC

VPIX16:1 VPIX16:1
VPIX1:1 VPIX1:1

X1:1
X1:1

9
K2
8
X6:3 X2:1 X6:1 X19:1 X2:1 X17:14 X6:1
W10 W10
1 1-2 only tilting 1 VTM VAG1,2
2
VBS VBM VBY1 VBY2 motorized
2
VTY1
3 Torque M Vertical 2-3 vertical+
3
M
tilting motori. Handswitch
switch motor Brake Tilting Brake
Clutch brake release
motor

X6:4 X2:2 X6:2 X19:3 X2:2 X17:15 X6:2


X16:4 X16:4

X16:2 Safety cable X16:2

2 K2
X30:1 X14:1 3
2 K2 K3 5
1
3 2
S3 7
5 W9
K3 1 1
S4
W9 7 W13
Safety
switches
2 2
W13
1 contour 1) 1 3
3 S5 3
2 2
2 2
3 1
1 S6 X30:2 X14:4
1
2

W12 X20:1
W12
1 VP1 10 1 1 VT1 10
2 Vertical controller K2 S1 2 Tilting K2
11 Stretch grip 2 11
3 switches
1
3 controller
S2
2 X20:2
Safety Safety
relay relay
K4 vertical
5 K4 tilting
5

3 D1 D1 VA 3 D1
D1 Brake Clutch D1 Brake
Safety relais deactivation signal activation signal Safety
deactivation signal
out signal relais
A3 06-01-18 Schr.

X1:2 out signal


X1:2

VPIX1:2 VPIX1:2
1) Safety switches contour can be activated from BuckyDiagnost VS digital tilt
front side and rear side of the tilting unit.
VPN5 GND VPN5 GND Safety circuit
Vertical and tilting motorized

BuckyDiagnost VS CSIP Level 1 (06.0) Z1-26


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
VP
VPI VP Wall box
VP1 Interface Wall Remote
VP4
VPI X13 X3 VP4 control
8 8
15 15

VT X1
8
15
9

VA
9 1
1

female male

X14 Vertical
9
1

8
15 VP1 controller
female

9
1

female

CAN
terminator X12 X3
8 8
15 15

W15
default
(without
CAN terminator) 1
9 9
1

female male

X10 X11
8 8
15 15

MSB out Tilt


MSB in VT1 controller
Pin assignment of Mini System Bus (MSB) 9

(optional)
9 1
1

male female

Bucky unit type


PIN Bucky unit 2 digital Bucky unit 2 sensing (CAN)
1 CAN GND CAN GND
2 CAN L in CAN L in
1 3 CAN L out CAN L out
4
5
not used
not used
not used
not used
VA 8
15

6 sync grid GND sync grid GND Vertical frame


7 sync grid sync grid VAX1
8 not used 26V DC GND 9
9 CAN 12V CAN 12V 1

10 CAN H in CAN H in female

11 CAN H out CAN H out


12 not used not used Bucky Unit 2 Digital
A3 07-02-12 Schr.

15 13 start grid start grid


14 start grid GND start grid GND
15 not used 26V DC
BuckyDiagnost VS digital
BuckyDiagnost VS digital tilt
MSB connection and Pin assignment

BuckyDiagnost VS CSIP Level 1 (07.0) Z1-27


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Vertical unit PE
screen
Wall box mains cable
without option tilting
Standard X11 X11
screw
L1
Mains
filter

screw
VBX11 VPX12 L2 X11 X11
Mains
filter

VPX11 screw
X12
VAX12 0
N5 26V DC Front
cover
Tilting 230V
X14
unit VAX13 screw
Bucky unit 2 X11
0
N2 20,4V AC
digital
Vertical 230V
X13 screw
VTX11 frame rear M4

N1 24V DC
B cover
P
0
230V
Vertical Pixium connected
through
VAVAP screw unit detector
base
VTX12 X12 cable cover X11
X11 N3 9,26V AC internal
frame
0 metal
230V sheet
VAVA screw screw X16 X15
X12 0
N4 9,26V AC X12
VAX11 230V VA Grid frame

bottom Bucky unit 2 digital X13


VAVA
X11 P StandardWall box cover

Rear

VAVA T X12
cover

X14 Tilting unit


A Vertical frame
V
Wallstand
(option)
A3 06-01-18 Schr.

BuckyDiagnost VS digital

BuckyDiagnost VS digital
Earthing diagram

BuckyDiagnost VS CSIP Level 1 (06.0) Z2-10


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
X1 0V GND

Filter modul X7
16
X7
1 15
8
9
2 1
1
3
Filter
X4 X33 9
1

VP4 X8
1
X2
open-no
X1
X32 1
9
CAN termination 9
X5
Remote control
1
2 (default) GND
CAN Modul 1 10 1 X31
U1 U2 X34 X35

S1 U4
D1 D2 D3 D4
1 RP1
X2 rd yw gr rd

1
VP3 X2

Receiver 5-Field chamber 1)

FOL X50
1

Communication
Detector Detector
VAVAPB1 (VS digital only)
X3 X13
W2
241V AC F4 20,4V F5 0V 26V
C38
GND 1
off
1
N5:230V AC VPI 0,5A yellow yellow Reset
230V AC 5A 1

219V AC 0V X21 219V AC 8 4


S2
on S4
X11
Service
230V AC
T1 N5:0V 0V F3 VPI 0V X2 switch 1.)
S3 X8
F6
5 1
1A X24 9,26V 241V AC Service
N2 AC 3,2A W6 W5 W4 W3
button 2.) 1

20,4V AC
N4 K4
X1
1
1
W7 2)
X9
1
9,26V AC 5
8
3
6
7
4
1
2
Current limiting
1
1 W8 2) D1
W10 X6 X16 BR1
2 1
X19 11 12 K2
1 W1
W11
+ VPI
F7 0V
L2:N / 1 1
1
W13
1
D3 D2
VP1
0,5A 1
L1:N / N1 - X22
7 5 13 1 K3 S1 1
N 219V AC
+ X23 on CFG3
HIGH LOW
L2:L / Wall stand controller
N5 L2 :L / CFG2

L 24V DC CFG1
1)
- 0V 230V AC X22 CFG0

26V DC
VPI F8 241V AC
1
CAN2
CAN1

X23 9,26V 1 CAN0


BOOT

AC
3,2A
9 3
W9 W12 V12 green

N3 7 1 1 1
MP2 V9
V10
yellow
red
9,26V AC V11
X21 X14 1 X20 X17
+5V yellow
screening of mains cable
1

X12
1

N1:L
N3:230V AC
N4:230V AC
L/ 230V AC L 230V AC VPI X14 C2
Interface wall X21
9 X32 X3

L2
3 1 4 1

1) CFG1 on = Vertical / Tilting motorized


3 2 1 1
VPI X8 X8
X24
X20
3 2 1
2 X13 1 X2
N3:0V
4
1 4 3 2 1 9 CFG1 off = only vertical motorized
N4:0V N/ N X25
8 7 6 5 X23
1
1
5 1 X5
3 2 1 2

N1:N X22
X7 3

2
3 2 1
9
X12
X15 5 1 X4
2) Only available on PCB version 02004
+26V AC R3
1 1 5

VPI X16 L/
LED
1 1
230V AC L
-
3 2 1
X1 D-Sub X65 X55 X56
X6 2 4 8
1 1
X11
gr N2:0V
L1
9
A3 09-07-06 Schr.

VPI X9 X16 X17 X1 9 9


1
2 2 2
R1

F2
2

N/ N X10 F1

N5
1 1 1
1 Multifuse 1,35A
1 1 1 2 1

12,5A N2:230V AC 10 10

F1 9 1 2

C1

5A
VP
BuckyDiagnost VS
Wall box digital

BuckyDiagnost VS CSIP Level 1 (09.0) Z2-11


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
X3.1
X3
X3.2 2 VP
1X X1
VP VP 1

/ X19
1X
8
X11

VP1X6
gr blk yw rd

rd blk

M
Motor

X1.2
X2.2 X1 X4.2
X2 X1.1 X4
X2.1 X4.1

bl pk

bl rd blk Upper position


4,5V +- 0,1V
2
Y1 between
Y2 pin 3-2
3 1
Clutch Brake
5,0V power supply
R between 3-1
Poti

S Top view Y1 = brake activ during power off


1 bl

24

pk

Safety switch
torque

X2 X3 X1

X2.1 X2.2 X3.1 X3.2 X1.1 X1.2


blk br
rd rd rd rd rd blk
thick

rd or
bl blk
bl br
gr bl

blk yw pk yw
thick
A4 06-01-18 Schr.

X4
X4.1 X4.2
rd rd
blk
bl
wt
bl VB
BuckyDiagnost VS digital
Vertical unit

BuckyDiagnost VS (06.0) Z2-12


C Philips Medical Systems
1 S3:2 S4:2 1
4 4
2 S5:1 S6:1 2

VAX12 VAX13
S4 S5

VT1X20:2
4 2 S2:1 2 4
S1:2
G1 VT1X20:1 1
G2
User interface S11 User interface
left VP3:X34 S2 right
5 field cont. S1...S2 = Switches for stretch
grip detection
X2
(only for option tilting)
1 10

X3 X4
G
1 2

39 40

User
G1 interface
G2 S3...S6 = Safety switches contour
keyboard collision detection
X1 (only for option tilting)
39 40

1 2

15 1

1) Safety switches contour can be activated from


VT1:X17
front side and rear side of the tilting unit.
Wall stand
controller

VAX11

S3:1
S6:2

1 2 X30

S3 S6
VT1 X14

1 X30:1 S5:2 1
A3 06-01-17 Schr.

4 2 S4:1 X30:2 2 4

VA
BuckyDiagnost VS Advanced tilt digital tilt
Vertical frame

BuckyDiagnost VS CSIP Level 1 (06.0) Z2-6.1


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
View B

VTM Tilt motor

gr
blk
cover yw
rd
rd
blk

View A

View B

VAGX1 VTM VAS1..S2 VAX30


(User interface) Tilt motor (Stretch grip switches) (Safety contour switches VAS3...S6)

View A
1 1

X20 X14
1 X21
X17 1 1 1 7

W9 W12 3 9

1
1

(VT)1 VTR Tilt potentiometer


X22

on
X23

Tilt controller
1
1

X16
W11
1 W13
1 1 Potentiometer adjustment :
1
2
X1
X19
BR1 W10 X6 In 90° position of tilting unit
4,5V +-0,1V between pin 3-2
1

1 3
D1 W8
5,0V Power supply between pin 3-1
1
W7 1
1
X9
W6 W5 W4 W3
1

X8 S3 1 5

S2
yw rd br
X3
S4 on 4 8
A3 06-01-23 Schr.

Bu St
1 green green green green X2
Reset 6 9 15 9

(VT)1 X8
C38
1 5 8 1
V54 V50 GND V51 V49 26V
off W2
X13

VTR VPIX12 VPIX17 VP1X16 VT


(Tilt potentiometer) (MSB CAN) (26V DC) (Safety)
BuckyDiagnost VS digital tilt
Tilting unit

BuckyDiagnost VS CSIP Level 1 (06.0) Z2-7


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
X1 (Power BUD board): Only systems with digital
1 bucky unit DDF
X66 1 24V AC
2 24V AC
X2 (Power in):
X4 1 9,26V AC
1 2 9,26V AC

3 9,26V AC
6 9,26V AC
4 20,4V AC
7 20,4V AC

H4 gn 8 24V DC
+5V 9 GND 24V DC

X3 (Power detector):

1 GND +24V
2 +24V
3 SENSE +24V
4 TRIGGER
X5
5 TRIGGER
1 6 GND +5V
7 +5V
1 8 GND -5V
J1 SW1 9 -5V

X4 (Communication in):

X3 1 GND 12V from SUN


4 RXD-485-B
1 6 RXD-485-A
7 TXD-485-B
H6 yw 8 TXD-485-A
9 +12V from SUN

H5 yw X5 (Communication detector):
1 RXD-232
H2 gn 2 TXD-232
+24V
3 GND 5V from FSXD
4 +5V from FSXD
1
X6 (Fan):
X2 1 GND +24V
9 2 GND +24V
3 +24V
J1: 4 +24V
1 X1
only for X66
fault finding F1 not used
1A slow
1
Caution:
H3 gn LEDs:
NEVER connect the -5V
Detector with jumper H1 24V AC Trigger, gn
plugged in. H2 +24V, gn
This can damage the H3 - 5V, gn
Detector! H4 +5V, gn
H5
on, while detector power OK
H6
J2:
S1:
Schr.

Hardware
autoreset via 1 H1 gn HW-reset detector
generator
A4 06-01-18

control desk
disabled
if bridged!
Pixium control board (PCB)
J2 wiring diagram
1 (4512 130 91836)
X6
1

BuckyDiagnost VS CSIP Level 1 (06.0) Z3-6


C Koninklijke
c Philips Electronics N.V.
ALL RIGHTS RESERVED
C1 X10 X65 X1D-Sub
2 1 1
10 10 X1 X17 X16
1 2 F1 9 1 R1
Multifuse 1,35A 9 1
1 1 1

2
X11 2 2 2

9
9
1 1
X15 R3
1 1 8 4 2 X9
X12 X25
X56 X55 1 2 3
9
5 1
X22 1 1

X4 1 5 1 2
X23
1 2 3 3 X7
X13 X20
X5 1 5 1 2 1 2 3
9
5 6 7 8 X24
1 1 2 3 4 1
X2 1 4 2
X14 1 2 3 X8
9 1 2 3
X3 1 4 1 3

1 X21
X32
C2 230V AC

Jumper X15
8 4

Without CAN termination (for BuckyDiagnost, DigitalDiagnost 1.x)


5 1

8 4

With CAN termination (for DigitalDiagnost 2.x)


5 1
A4 08-04-17 Schr.

VPI
BuckyDiagnost VS
Interface wall
4512 133 0205x

BuckyDiagnost VS CSIP Level 1 (08.0) Z3-3.1


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
without CAN termination (default)
Jumper X15
8 4
X15 8 7 6 5 4 3 2 1 Without CAN termination (for BuckyDiagnost, DigitalDiagnost 1.x)
5 1

1 8 4

100R
With CAN termination (for DigitalDiagnost 2.x)
R2 5 1
2
X65
1
CAN termination
2 CAN L in X10 X11 X12 X13 X14
3 CAN GND CAN GND 1 1 1 1 1
4 CAN L in 2 2 2 2 2
CAN 5
CAN L out
3 3 3 3 3
(not used)
6
not used
4 4 4 4 4 X1 Vertical controller
7 CAN H in 5 5 5 5 5 2 VP1:X1
not used X16 1 26V DC
8 sync grid GND 6 6 6 6 6
VPN5 2 GND (only VS digital,
9 CAN 12V sync grid 7 7 7 7 7 26V DC 1 26V VS digital tilting)
not used 8 8 GND 26VDC 8 8 8
CAN 12V 9 9 9 9 9
2 C2 X17 Tilting controller
CAN H in 10 10 10 10 10 2 VT1:X1
X1 (D-Sub) 11 11 11 11 11 1 100n 26V DC
CAN H out 1
1
12 12 12 12 12
not used
2 sync grid GND 1
start grid 13 13 13 13 13 F1 1,35A Multifuse
3 sync grid
Release 4
start grid GND 14 14 14 14 14
X32
start grid GND 2
Grid 15 15 26VDC 15 15 15 1
5 not used
Exposure 2
6 not used
(only for start grid
3
non sensing) 7
MSB CAN
MSB CAN MSB CAN MSB CAN MSB CAN
8 BD TH SIX4 VAVA:X1 VT1:X3 VP1:X3 VP4:X1
9 DD or SCU VSIX4 Bucky unit 2 digital BD VS Wall stand BD VS Wall stand BD VS User interface
Bucky unit 2 sensing controller tilting controller vertical Remote control
Bucky unit 2 non sensing (only VS digital, (only VS digital,
VS digital tilting) VS digital tilting)

X21
1 20,4V ACP
Connection height potentiometer RS 485 Connection
Mains supply VPN2
2
X7 X8 X4 X5 X55 X56 20,4 VAC 3
** 1 Poti +10V 1 1 1
20,4V ACN
X20
1 L1 1 GND 1
Mains 2 2 VPL1 2 Poti slider 2 2 2 X24
VBR 2
connection Mains 3 Poti GND 3 VT1:X8 3 3
1 9,26V ACP
3 N 3 PCB X4 3
230V AC filter 4 4 VPN4 2 9,26V ACN
4 RXD 485 B 4 Pixium Contr. 9,26 VAC 4
5 screen 5 RS485 5 5 Board X23 5 PCB X2
6 6 1 9,26V ACP1
RXD 485 A
VPN3 2 9,26V ACN1
6 Pixium Contr.
7 TXD 485 B 7 Board
X9 Connection brake
9,26 VAC 7
3 * 8 TXD 485 A 8 X22
A3 08-04-17 Schr.

X3 X2 9 9 1
8
2 VPL2 1 1
+12Volt 24V DC
Brake +
1 Mains 10 10 2 GND 24VDC
filter VBY1:X1 2 Brake - 2 VT1:X6 VPN1
24 VDC 3
3 3
R3 VPI
4 4 X25
* Via mains filter VPL2 to VPN1, VPN3, VPN4 Charging 1
Multifuse BuckyDiagnost VS
2 1
** Via mains filter VPL1 to VPN2, VPN5 station 2 Interface wall
Remote
only for BuckyDiagnost VS Advanced tilting only for BuckyDiagnost VS digital, VS digital tilt
control
3 4512 133 0205x

BuckyDiagnost VS CSIP Level 1 (08.0) Z3-3.2


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Jumper Function Default
W1 EPROM type 29F040 2-3, 4-5
W2 1-2 -> RAD system, 2-3 URF system 1-2
W3 Motor current limitation (default value 13A) 1-2
W4 Motor current limitation (default value 13A) 1-2
X3 X13 W5 Motor current limitation (default value 13A) 1-2
V51 V49 26V GND V53 V54 1
off
1 W6 Motor current limitation (default value 13A) 1-2
C38 Reset W7 open
X2 yellow yellow yellow yellow Only available on PCB version 02004
W8 Only available on PCB version 02004 open
4 on S4
8 W2 S2 W9 1-2 -> Contour switches VAS3…6 in safety circuit by-passed (for test only).
2-3 -> Contour safety switches active 2-3
1 Service W10 1-2 -> for VS Advanc ed tilting, 2-3- > for VS digital tiltin g depends on configration
switch 1.)
S3 W11 EPROM type 29F040 1-2
5 1 X8 W12
1-2 -> Stretch grip switches VAS1…2 in safety circuit by-passed (for test only).
2-3
1 - 2 Tilting only (VS Advanced tilt) Service 2-3 -> Stretch grip switches VAS1…2 active
button 2.) 1
W55 X55 2 - 3 Tilting and vertical motorized W13 only used with DigitalDiagnos t VM , VS default 1-2 ! 1-2
(VS digital tilting) W6 W5 W4 W3 W55 only used with DigitalDiagnos t VM , VS default 1-2 ! 1-2
1 1 X9
K4 1
1 W7 1
5 3 7 1 X1 Current limiting
8 6 4 2 1
VS digital tilting VS Advanced tilting Replacement for BD VT
1 W8 D1 (vertical and tilting motorized) (only tilting motorized)
W10 BR1 BOOT off off off
2 1
X19 11 12 K2 CAN0 off off off
1 W1 CAN1 off off off
1 1 1 W13 W11 CAN2 off off off
1 CFG0 on on on
1
X6 X16 X50 1
D3 D2 CFG1 on off off
7 5 13 1 K3
VT1 S1 1 CFG2 off off off
CFG3 off off off
X23 Tilting controller HIGH LOW on CFG3
CFG2
CFG1
X22 CFG0
CAN2
1 CAN1
CAN0
1 BOOT
1
X5
9 3
W9 W12 1
V12 green
X60 9
V9 yellow off on flashing
7 1 1 1 V10 red V9 (yellow) Service Mode inactiv Service Mode active -

X21 X20 X17 V11 yellow


V10 (red) Selftest ok Selftest failed or safety circuit open During selftest
X14 1
1 1 V11 (yellow) - CAN bus not connected CAN bus active
V12 (green) Controller not ready Controller ready -
V49 (yellow) Power PO26V Mot not ok 26V power supply motor VTM ok -
V51 (yellow) Power PO26V BRK not ok 26V power supply ok -
26V power supply safety circuit /
Service switch S2
1.) V53 (yellow) Power PO26V EL not ok
wall stand controller VP1 D1 ok
-
Off: during normal operation (default) V54 (yellow) Power PO26V DIS not ok 26V power supply ok -
On: tilting movement possible in case of wrong or missing potentiometer value VTR (fault finding) or
mode to set 0° and 90° potentiometer value for wall stands without CAN bus in old world systems (adjustment) see 2.)
A3 09-12-04 Schr./Schu

2.)Service button S3
Storage of potentiometer value VTR for wall stands without CAN bus in old world systems (precondition: S2 On)
1. move tilting unit to 0° position and press S3
2. move tilting unit to 90 ° position and press S3

VT1
BuckyDiagnost VS
Wall stand controller tilting
4512 133 02004/5

BuckyDiagnost VS CSIP Level 1 (a/09.2) Z3-5.1


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
VPI X40 VP3 X35 PO 26V Motor final stage
Interface tracking 5-field control X21 BRK

1
PO 26V
PO 26V MOT
X2
BRK
PO 26V PO 26V Mot + 1 rd
V34 EN
2 Lock VTM
Not Tilted Move Chamber 90° M
DIR C38 4 Tilt motor
P7:6 U8 3 Mot - 2 blk
PO 0V
PO 0V PWN
250 2600 1300 1300 800 5
Fault drive K4
3 PO 0V
GND GND P3:15 Current +5V
1 2 3 1 2 3 1 2 3 1 2 3
PO 0V limiting
4 3
Tilted Field right P3:0
5 W6 W5 W4 W3
Shutter trans open Field central in/out P3:1 Hall sensor
PO 0V
tilt drive
Shutter trans close Field left
6
U1 P3:2
Encoder IN 4 rd
U9 PO 15V HI

Shutter long open Field select 0


7
P3:3
1 2 3 W7 3) 5 blk

8 A Kanal PO 0V 6 yw
Shutter long close Field select 1 P3:4 P2:14
9
Light Trigger Default (left/right) P3:5 U15
only for replacement
1 2 3 W8 3)
BuckyDiagnost VT
sensing X17 User interface IN P2:15
B Kanal 7 gr
(old world interface)
5 Column A screen 8
P2:1
6 Column B P2:2
D1 1) 2)
7 Column C U16 Analog IN X8 VPI X5 VPI X4
P2:3 1 br VTR 1 1 VBR
VREF
VAG X1 8 Column D P2:4 2 yw 10K 2 2 10K
User interface AIN 0
9 Column E P2:5 PO 26V MOT 3 rd 3 3
Temp Amp. screen 5 Tilting
P5:2 PO 0V Height potentiometer
potentiometer
PO 26V EL only (for tracking option)
PO 0V
14 vertical P5:0 X9
Torque 1 VTR
1 2 3 W10 VREF
15 Pot tilt 2 10K
Hand switch P5:3
3

X50 screen 5
Tilting
potentiometer
2 User interface OUT PO 0V

4
X17 1) Potentiometer connection for
Row 1 1 BuckyD. digital, BuckyD. digital tilt
P8:0
Row 2 2 2) Potentiometer connection for
X55 W55 P8:1 BuckyD. advanced tilt
1 1
Row 3 3
P8:2 U19
2 2 default
Row 4 4
3 P8:3
K3 Row 5 13 VAGX1
X6 PO 26V EL
1 P8:4
User interface
3 13

4 PO 0V
VBY1 VBX1 VPIX3 VPIX2 UI LED A 10
Vertical brake 2 2 2 2 5 7 P8:5
V20 U20 3) Only available on
PO 26V BRK Brake vertical UI LED B 11
PCB version 02004
1 1 1 1
W13 P8:6
PO 0V UI LED C 12
3 P8:7
X16 2
screen 16
Contour switches br 2 +5V
VAS3...S6 can be Safety 1
be activated through VP1X16 rd 4
a mechanical
interface from front 9 2 10 1 K2
and rear side of the X14 PO 26V
V9
tilting unit ! 1 8 3 11 12 EL V10 V11 V12
rd
PO 0V PO 0V
Safety 2
1
P4:7
Contur switches 3 D12 Safety relais out
2 P6:5
Tilting
br 4 2 K4
3 PO 0V P6:6 +5V

X20 W9 1 2 3 W12
1
VAS2 VAS1 P6:7
rd 2
Stretch Stretch grip Reset
A1/A3 09-07-06 Schr.

1 RST IN +5V
grip switches br
Contour
3 S4
PO 0V
X19 PO 0V
P5:11 +5V
1 S2
3
V21
PO 0V
PO 0V
Clutch P5:12
S3
PO 0V
VT1
BuckyDiagnost VS
Wall stand controller tilting
4512 133 02004/5
BuckyDiagnost VS CSIP Level 1 (a/09.0) Z3-5.2
C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
PO 26V

Serial I/O
DIS X13
CAN / SAN +5V 1

PO 5V CAN
D13
3
P3:10

7
P6:4
B2
X3 +5V
RXD CAN
P3:11
2
10
8
not used
11 CAN H P6:3

5
D11
2 CAN L B3 PO 0V 10
TXD CAN
3
screen 11

9 CAN V+ IN OUT PO 5V
U25
MSB CAN CAN
+5V
VPI X12 1 CAN GND
U26
12 B1 STOP MV
not used
6
1 RAD/URF
W2 2

PO 0V

W11 29F040 27C4001


P3:12
1 1 1
+5V
2
Power supply 2 2
MP2 P4:3
3
3 3

PO 26V
X1 W1
VPIX17 26Vdc br 1 1 2
U17 +5V 29F040 27C4001
L3 1
1 1
WRL
GND bl 2
+5V 2 2 2
3 3 3
PO 0V PO 26V EL
1
4 4 4
BR1
2 D1 5 5 5

PO 26V PO 26 V PO 26V uP + PROM + RAM + A/D +5V


MOT U18 PO 15V UI
BRK DIS
A / D Bus RD
A1...18
D0...7
D2 LOW
V49 V51 V54 V53

PO 0V PO 0V PO 0V PO 0V
RD PO 0V
A1...18 +5V
D0...15 RD
A1...18
D8...15

D3 HIGH

X23 PO26V +5V PO 0V


1 EL
X10
2 P2:12
U13
3
P2:10 S1
PO 26V 2 1
X22 BRK POL4
4 3
PO 26V P5:5
1
BRK 6 5
P5:6
V36 8 7
2 Brake P5:7
P7:7 P5:8 10 9
3
12 11
PO 0V
A1/A3 01-11-09 Schr.

P5:9
14 13
P5:10
16 15
P5:4

Dipswitch

PO 0V

VT1
BuckyDiagnost VS
Wall stand controller tilting
4512 133 02004
BuckyDiagnost VS CSIP Level 1 (a/06.0) Z3-5.3
C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
5V 21V 15,3V

N3 1) N1 N2
X2 U in

V17 U4
V18

X35 X34
1 10 1
X31 4 4 4

1 U10 1 U8 1 U9
X5
1 X32 GND
9 4

U12
1
9
1
X33
4

U11 1
1 6

4
15
8 U13
16
1

X7 8
15 4 9
U14 5

1
0V GND

1) Jumper X2 (without function on PCB version 1)


1 - 2 5-field measuring chamber for
left hand operation (default for VS).

2 - 3 5-field measuring chamber for


right hand operation (default for VM)
A4 08-12-09 Schr.

VP3
DigitalDiagnost VM, VM 2
BuckyDiagnost VS
5 Field control
4512 133 0210.

BuckyDiagnost VS CSIP Level 1 (08.0) Z3-6.1


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
X31 X32
1 +400V +400V 1
3 Field 3 (right) +15V Field 3 3
4 Release chamber Release chamber 4
5 +15V +15V 5
6 -15V -15V 6 VA PH X1 / VAVA PHX1
EZ X31
(EZ X12 ED Eleva) 7 Signal chamber Signal chamber 7 5 Field
from Generator 8 8 Measuring chamber
RF 0V chamber RF 0V chamber
11 Field 1 (left) Field 1 11
12 Field 2 (center) Field 2 12
13 0V XG 0V XG 13

X7 R43
VCC
R37 R40
14
VCC VCC Field 4
U10 5 U8 5 U9 5 +15V VCC Field 5 15
0V GND 5 U12
4 +15V VCC
4
X35 0V GND 4 0V GND 0V GND 4 U11
5 Control field center +15V VCC
FD2 R45 4
U13
6 Control field left +15V VCC
0V GND R44 4
4 Control field right U14

VAG User interface keyboard FD L FD R 0V GND R46 4


VAG User interface keyboard
+15V FD3
FD1 0V GND R47
VAG1 X3 VAG2 X4 X2 X34 VAG2 VAG1
User interface left User interface right FD4 0V GND User interface User interface
15 15 10 10
FD5 X34 X2 X4 right X4 X3 left X3
left select R50 7 9 6 LED Field 3 5 6 5
14 14 1 1 R19
chamber left
R51 5 5
R6 2 LED Field 1 1 2 1

right select R49


12 12 2 2 R21 8 8 10 LED Field 4 9 10 9 1 3
chamber right
2
R10 R48 5 4
Control 9 7 4 LED Field 5 3 4 3
upper select
11 11 3 3 R20 R52/53
chamber upper 6 6 8 LED Field 2 7 8 7
R8
10 10 15V DC
lower select
13 13 4 4 R22 chamber lower
R12 VP3 5 Field control X33 EX X1
1 1 FD1
2 2 FD2
3 3 1 3
FD3
X35 4 4 2
R16 FD4
VS Advanced 2 Move chamber 90°
5 5 FD5 5 4
VPI X35
R17 7 7
Interface Tracking X5
VS Advanced tilting 3
VCC X2 CH Amplimat display
VT1 X21 GND 3
Tilting controller 7 2 - 3 right hand operation of wall stand
Chamber position 0 2
R64
A3 06-09-29 Schr.

8 1 - 2 left hand operation of wall stand


Chamber position 1 1
VS digital / digital tilt
VP1 X21 9
Vertical controller GND
+15V VCC (5V)
1
R62 V25 +21V VP3
U in
N1 N2 DigitalDiagnost VM, VM2
Multi fuse R71 R75 R77/78
N3 R67 BuckyDiagnost VS digital
R65/66
0,5A
R73 R76 V18 R68 V17 5 Field control
4512 133 0210.
GND GND GND GND GND
BuckyDiagnost VS CSIP Level 1 (06.2) Z3-6.2
C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Antenna
Filter modul

CAN Modul

Receiver

X1

X4 X5
8 8

1 1
S2 S1
J2 J1 RP2
1 1 1 S2 not used
J1
3 3
J2

U2
U1

S1
D1 D2 D3 D4
1 RP1
X2 rd yw gr rd

Receiver

8 8 X1

8
15
CAN Modul
1 1

X3 X4
X2
9
open-no 1
CAN termination
(default)

Filter modul

X7
1
2
3
A4 06-01-18 Schr.

Filter
X8

VP4
1
2

BuckyDiagnost VS
Remote control
Receiver with CAN Modul 4512 133 0290x
Filter modul with cable X7 4512 133 0194x

BuckyDiagnost VS CSIP Level 1 (06.0) Z3-7.1


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
CAN Modul X3 X4
male female
Receiver X1 Antenna
1
L1 L4
CAN CAN VCC 1 CAN
CAN VCC 2
VCC
CAN VCC 2
L3
Termination L2
X2 1 2
3 CAN GND 3 CAN U4
male CAN GND
GND 4 CAN GND 4
X1 5 5
CAN VCC CAN VCC
female
1 CAN GND 6 6
2 CAN L in N1 U6
7 RxD CAN 7
3 CAN Lout U1 16
8 TxD CAN 8
4 MHz
N2
X4 X5 S1
male female 8 1
MSB CAN 5 EN MV EN MV 1 EN MV 1 7 2
7 EN XR EN XR 2 EN XR 2 6 3
Wall stand
5 4
controller 8
6 3 Rel 2 3
1 2 3 4
VP1 X14 U1
9 CAN 15V 4 Rel 1 4
RP1
10 CAN H in 5 Txd PIC 5
8 7 6 5
J1
11 CAN Hout 6 Txd K2 6 3
12
7 Rxd K2 7 2
13 X2
14
8 RxD PIC 8 1 male
1
15 24V DC
U2
VCC D1 D2 D3 D4 U3 2
(+5V)
TxD EXT 3
4
CAN RxD EXT 5
GND 6
7
J2 8
3
9
2
Filter modul 1

X7 X8 5 6 7 8
male male
1 24V DC U2 RP2
1
Charging S2
VPIX25 2 GND 24V DC FI 1 2 station 4 3 2 1 8 1
3 7 2
6 3
5 4
A3 06-01-18 Schr.

VP4
BuckyDiagnost VS
Remote control
Receiver with CAN Modul 4512 133 0290x
Filter modul with cable X7 4512 133 0194x

BuckyDiagnost VS CSIP Level 1 (06.0) Z3-7.2


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
VP3:X34
5-Field control

1 10
39 40 X2 1 2

User interface left User interface right


VAG1 X3 X4 VAG2

X1
1 2 39 40
15 1

VT1:X17 Tilting controller (VS advanced tilt / VS digital tilting)


VP1:X17 Vertical controller (VS digital)
VP2:X17 Tilt arm controller (VM)
VPI :X1 Interface Tracking (VS advanced)
A4 06-03-01 Schr.

VAG
DigitalDiagnost VM
BuckyDiagnost VS
User interface keyboard
4512 133 0206x

BuckyDiagnost VS CSIP Level 1 (06.1) Z3-8.1


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Row2 P
ColumnD..P
ColumnC..P
ColumnB..P
ColumnA..P
Row3 LED8
Row3 LED7
Row2 LED6 X2
Row2 LED5 1
2
Row2 LED4
3
4
X3 X4 5 VP3 X34
1 2 LEDA L4 1 2 LEDA L4 6 5-Field control
3 4 LEDB L5 3 4 LEDB L5 7
5 6 LEDC L6 5 6 LEDC L6 8
7 8 LEDA L7 7 8 LEDA L7 9
9 10 LEDB L8 9 10 LEDB L8 10
11 12 11 12
13 14 13 14
VAG1 15 16 15 16
VAG2 User interface right
User interface 17 18 17 18
left 19 20 19 20
21 22 21 22
23 24 23 24
25 26 ROW1 LED1 25 26
LEDA L1 27 28 ROW1 LED2 LEDA L1 27 28
LEDB L2 29 30 ROW1 LED3 LEDB L2 29 30
LEDC L3 31 32 ROW3 LED9 LEDC L3 31 32
LEDD L9 33 34 ROW4 LED10 LEDC L9 33 34
35 36 LEDA L10 35 36 LEDA L10
37 38 37 38
Brake A 39 40 Brake B 39 40

X1
Row 1 1
Row 2 2
Row 3 3
Row 4 4
ColumnA 5
ColumnB 6
Wall stand controller:
ColumnC VT1:X17 VS advanced tilt
7
VS digital tilting
ColumnD 8
VP1:X17 VS digital
ColumnE 9
VP2:X17 VM
10
11 Interface Tracking:
12 VPI :X1 VS advanced
A3 06-03-01 Schr.

13
14
15
16
VAG
DigitalDiagnost VM
BuckyDiagnost VS
User interface keyboard
4512 133 0206x

BuckyDiagnost VS CSIP Level 1 (06.1) Z3-8.2


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Drawings/Diagrams VS replacement for VE/VT BuckyDiagnost VS

DRAWINGS/DIAGRAMS VS REPLACEMENT FOR VE/


Contents
VT
Survey of units BuckyDiagnost VS ...................................................................................................................Z1-1
BuckyDiagnost VS / Labeling .........................................................................................................................2Z-10
Conversion of measuring chamber ..................................................................................................................Z2-9
Replacement for BuckyDiagnost VE sensing with BD VS basic ....................................................................Z1-30
Replacement for BuckyDiagnost VE sensing_tracking with BD VS advanced...............................................Z1-31
Replacement for BD VT sensing_tracking with BD VS advanced tilting ........................................................Z1-32
Wall box tracking - Replacement for BD VE sensing BD VS basic ................................................................Z2-20
Wall box tracking - Replacement for BD VT sensing BD VS basic ................................................................Z2-21
PCB Interface tracking BD VS.......................................................................................................................Z3-2.1
PCB Interface tracking for replace VE sensing .............................................................................................Z3-2.3
PCB Interface tracking for replace VE sensing_tracking ...............................................................................Z3-2.4
VPI BuckyDiagnost VS / Replacement for BD VT sensing / Interface Tracking............................................Z3-2.5

4512 984 28037 REV A CSIP Level 1 (19.0) 12 - 1


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS Drawings/Diagrams VS replacement for VE/VT

12 - 2 CSIP Level 1 (19.0) 4512 984 28037 REV A


© 2020 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
VP
VB
VP
VB

VT

VA

VAG2
VA
VAPH
VAG1
VA = Vertical frame
VAG1 = User interface left
VD VAG2 = User interface right
VAPH = 5-Field measuring chamber
VAVA = Bucky unit 2 digital
(BuckyDiagnost VS digital)
VAVA
VAVA Bucky unit 2 sensing/ non sensing
(BuckyDiagnost VS basic/ advanced)
VAPH VB = Vertical unit
VD = Display
VP = Wall box
VT = Tilting unit

BuckyDiagnost VS basic BuckyDiagnost VS advanced


BuckyDiagnost VS digital
A3 09-10-21 Schr.

BuckyDiagnost VS
Survey of units

BuckyDiagnost VS CSIP Level 1 (19.0 Z1-1


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
3) "Wall Box"
High End
Wallstand
Segment Control Unit M 3) I
(Optional, if no table is present!
Wall Box
BuckyUnit 2 sensing / non sensing
Only for DigitalDiagnost 1.x, not
necessary for BuckyDiagnost VS digital)

3) B Logo Address
BuckyUnit 2 digital Typen No. Electr. data
IEC60601-1 IEC60601-2-32

I FDA
UL / CSA IEC
Logo Address
WEEE
CE Environmental
Typen No. Electr. data

Typen No.
IEC60601-1 IEC60601-2-32

Logo
H UL / CSA
FDA
IEC

Barcode

UL / CSA
WEEE
CE Environmental

Typen No.
Logo
M

UL / CSA
Barcode
E
Outside
F G Inside

3) E Vertical unit G
with BuckyDiagnost VS Basic G
with BuckyDiagnost VS advanced /
advanced tilt
3) H

Barcode
Address
Family PEI Address

Familie Pei

Typen No.
Typen No. Barcode
Package 1) Address
Typen No. Barcode

Logo Address
Typen No. UL / CSA
A2/A3 08-06-04 Schr. / Kö

3) F 1) Package VS basic
Package VS advanced
2) Package VS advanced tilt
Package VS advanced tilt replacement BuckyDiagnost VS
Package VS basic replacement Package VS digital tilt
Tilting unit Package 2) Address Package VS advanced replacement Labeling
3) Necessary for BuckyDiagnost VS digital
Typen No. Barcode Package VS digital

4512 983 11021 CSIP Level 0 (b/08.0) 2Z-10


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost VS BuckyDiagnost VS

Left hand Right hand


operation operation
(default)

Hardware Programmings

BuckyDiagnost TH2/TF (MSB)


BuckyDiagnost VS basic
1 2 1 2
S1 in position 1 5 Field measuring chamber S1 in position 2 S1
S1 (default)

BuckyDiagnost VS advanced / digital


5-Field control
Position 1-2 Jumper VP3X2 Position 2-3
(default for VS)

Software Programmings
Bucky Controller (X-Scope)

Program Manual / Bucky RAD / Wallstand / Basic Setup /

Left (default) Bucky Device Orientation Right


A3/A4 06-06-13 Schr.

BuckyDiagnost TH2/TF (MSB)


Conversion of measuring chamber

4512 983 11361 CSIP Level 0 (06.0) Z2-9


C 2004 Philips Medical Systems
P B
Wall box Basic Vertical unit Basic
4512 133 2903x 4512 201 0270x

A
0
Vertical frame basic
230V
N2 20,4V AC 4512 201 0275x (including wall box basic)

I Tracking
Interface

SX54 / VSX54
Mains connection X8
230VAC X7 230VAC

X1
X3 Y1 Brake
X2 VA
X1
Bucky unit 2
sensing 4512 201 0256x
non sensing 4512 201 0257x

X16
26VDC

MSB_CAN
VA
Bucky unit 2
SZ1 X12 sensing (non CAN)
UI
BuCo X1 D-Sub 4512 201 0258x
Brake release switch
Release wallstand
SZ1 X10 X11 X1
X2
BuCo S1
Sensing wallstand

X1 MSB Analog

X32 PH
EZ X31 X1 5Camber
Field measuring

Amplimat

V
Wall stand
A2/A3 06-01-18 Schr.

BuckyDiagnost VS Basic
Replacement for Bucky D. VE sensing

BuckyDiagnost VS CSIP Level 1 (06.0)


C Koninklijke Philips Electronics N.V.
Z1-30
ALL RIGHTS RESERVED
P B
Wall box Basic with tracking Vertical unit
4512 133 2897x Column manual (tracking)

A
0
Vertical frame tracking
230V
N2 20,4V AC 4512 201 0276x (including wall box basic)

I Interface
Tracking
SX54 / VSX54
Mains connection X7
X8
230VAC
230VAC
X1 User-
SZ1 X5 Interface
X5 X3 Y1 Brake
BuCo
Tracking
G1
left
X1
X4

Vertical
X4 R potentiometer
X16 26VDC

MSB_CAN
VA
SZ1 X12 X3
Bucky unit 2
BuCo X1 D-Sub
X1 G sensing (non CAN)
Release wallstand
User-Interface 4512 201 0258x
X2 keyboard
X4
SZ1 X10 X2 X11
BuCo X32
Sensing wallstand

User-
Interface

G2
right

X1 MSB Analog

EX X1
Amplimat Display X33
X35 PH
X32 X1 5Camber
Field measuring

EZ X31
Amplimat X31 VP3 X34

5 - Field control

V
Wall stand
A2/A3 06-01-18 Schr.

BuckyDiagnost VS Advanced
Replacement for Bucky D. VE sensing / tracking

BuckyDiagnost VS CSIP Level 1 (06.0)


C Koninklijke Philips Electronics N.V.
Z1-31
ALL RIGHTS RESERVED
P B T A Vertical frame tilting / tracking
4512 201 0280x (including wall box basic)
Wall box Basic tilting / tracking Vertical unit Tilting unit
4512 133 2898x Column manual 4512 201 0271x 4512 201 0280x Stretch grip
Column manual with tracking 4512 201 0272x
S1 S2

L L1 L 230V AC N5
(optional)
Mains
filter 0V 26V DC
N N

X20

Safety switches contour


X30 S3 S4 S5 S6
I Interface
Tracking
X14

SX54 / VSX54
Mains connection
230VAC
X7
230VAC
X8
VT1
SZ1 X5
X5
X1 Tilt controller User-
Interface
BuCo Tracking X3 Y1 Brake
SZ1 X11
G1
X3 left
BuCo Tilt pos. X2
X4
X5 Tilting potentiometer
not connected Vertical potentiometer
X4 R (option tracking) R
X16 26VDC X8
X17

MSB_CAN
VA
SZ1 X12 Bucky unit 2
BuCo Release X1 D-Sub M X2 X1 X3
G sensing (non CAN)
X17
X40 Tilting motor User-Interface 4512 201 0258x
X2 keyboard
X4
SZ1 X10 X11
X2 X32 X6 Brake
BuCo Sensing

X1Power

X3 MSB_CAN User-
Interface

G2
X21
right
X8

X1 MSB Analog

EX X1 X33
Amplimat Display X35 PH
X32 X1 5Camber
Field measuring

EZ X31
Amplimat X31 VP3 X34

5 - Field control

V
Wall stand
A2/A3 06-01-18 Schr.

BuckyDiagnost VS Advanced tilt


Replacement for Bucky D. VT sensing / tracking

BuckyDiagnost VS CSIP Level 1 (06.0)


C Koninklijke Philips Electronics N.V.
Z1-32
ALL RIGHTS RESERVED
20,4V VPI X16:1
VPI X16:2
AC
0V
F4 5A

N2 20,4V AC
N2 20,4V AC

F3 1A
241V AC

219V AC
230V AC

0V
VPI X8:3

VPI X8:1

Interface
VPI Tracking

1
VA X1
(Brake release switch VAS1) SZ1 X5

X5
BuCo

X1

9
VB X4
Height potentiometer

1
SZ1 X10

X4
1

9
X2
BuCo

1
VA VA X1

X11
MSB Analog
1

X32

1
X2

15

X10
1
9

X1 D-Sub
X3

15
1

1
1
X40

SZ1 X12
VB X1 BuCo

(not used)
(VBY1 vertical brake)
X6
H1
X17 1
X3

X16

230V
X8

X7
Multifuse
2,5A
F1

Mains
230Vac
1

1
N2 F4 20,4V

X11
N2 F4 0V

N2 F3 230V

VAX11
VBX11
N2 F3 0V

MEX
A4 06-01-19 Schr.

VP
BuckyDiagnost VS Basic
Wall box tracking
(replacement for Bucky D. VE with sensing)

BuckyDiagnost VS CSIP Level 1 (06.0) Z2-20


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
VPI X16:1 +26Vac
VPI X16:2 0V
gr

LED

N5
T1
N5
F2
12,5A

F1
5A
br
bl
Interface
VPI Tracking

1
L N SZ1 X5

X5
230V AC

BuCo
0V

Tracking

X1

9
VB X4
Height potentiometer
SZ1 X10

1
L1 BuCo

X4
1

9
X2
EX X1 Sensing

1
VT1X6 VP3
Amplimat EZ X31 VA VA X1
Brake X35

X11
Amplimat
1

Display MSB CAN

X1 D-Sub
X32

1
X2

15

X10
L/ N/

1
9

X31

9
1
0V GND

VT1 SZ1 X11


X7

BuCo
VPX3

VAVA PHX1 X21


VPI X16:1 bl

VPI X16:3 blk

Tilt pos.
1

15
5 Field chamber
1

1
9

1
X32

SZ1 X12
1)

X40
1
X2

VB X1 BuCo
U4

(not used)
(VBY1 vertical brake)
GND

Release
X33
1

X6
X5

H1
9

X17 1
X3
X35

VP3
X16

230V
VPI X32
X8

X7

5-Field chamber
Multifuse
10

2,5A
X34
16

F1

VAG X2
1

Mains
1) Jumper X2 (without function on PCB version 1)
230Vac
1

1 - 2 5-field measuring chamber for


left hand operation (default for VS). X11
L1 230V blk N/

VAX11
2 - 3 5-field measuring chamber for VBX11
L1 0V bl L/
N5 26Vac

right hand operation (default for VM) MEX


VT1 X1

N5 0V
power
A4 06-01-19 Schr.

VP
BuckyDiagnost VS Advanced
Wall box tilting
(replacement for Bucky D. VT with sensing)

BuckyDiagnost VS CSIP Level 1 (06.0) Z2-21


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
X40 1
F1
Multifuse 2,5A
X1
X2 1 1

X3 D-Sub
X3 9 X4 1

1
X17 1

1
X32 1

X16
1 X1 D-Sub
1
H1 9

1
1 X2 D-Sub
15
X8 X11 9
1

1
X6
1
X7 (not used) 15 X10 X5
1
230V
9
A4 06-11-07 Schr.

VPI
BuckyDiagnost VS
Interface Tracking
4512 133 0203x

BuckyDiagnost VS CSIP Level 1 (06.1) Z3-2.1


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
X7 X3 VBX1 Vertical X2
Mains 3
2 Brake Y1 X11 D-Sub
connection 2 1 GND Uref 3
230Vac 1 1 2 Format y 5
3 Cass Press 2
3 4 Tray Close 7
X8 5 SZ1X10
3 6 BuCo
VPN2 GND 26V DC X32 VAVA X1 7
2
MSB sensing
26V DC 1 8
1 26V DC Bucky unit 2
R7 9 PO Uref 4
2 10 Format x
F1 VP3 X35 6
5 Field control 11 Cass Exc 9
X16 Multi fuse 3
2,5A VDR R2 V9 Power 26V DC 12 +26V 1
1 V10 13 +26V 8
R8 GND 26V DC
VPN2 V1 V2 14
20,4Vac 2 100n 26V DC
H1 R1 15 X1
C8 C9 D-Sub
+26V 1
Free exp1 7
GND 26V DC SZ1 X12
Sync- 2
Sync+
BuCo
3
Release
GND 26V DC 5
wallstand
Free exp2 9
Start+ 6
Start- 4
GND 26V DC

VAX1 X1
VAS1 14 14
Brake release
button user
interface 15 15
A3 06-01-17 Schr.

VPI
BuckyDiagnost VS Basic
Replacement for BD. VE sensing (without tracking)
Interface Tracking
4512 133 0203x
BuckyDiagnost VS CSIP Level 1 (06.0) Z3-2.3
C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
X7 X3 VBX1 Vertical X2
Mains 3 Brake Y1 X11 D-Sub
2
connection 2 1 GND Uref 3
230Vac 1 1 2 Format y 5
3 Cass Press 2
3 4 Tray Close 7
X8 5 SZ1X10
VPN2 3 6 BuCo
2
GND 26V DC X32 VAVA X1 7
1 MSB sensing
1 26V DC 26V DC 8
R7 Bucky unit 2 9 PO Uref 4
2
F1 VP3 X35 10 Format x 6
Multi fuse 3 5 Field control 11 Cass Exc 9
X16 2,5A R2 V9 1
1
VDR Power 26V DC 12 +26V
V10
R8 GND 26V DC 13 +26V 8
VPN2 V1 V2 14
20,4Vac 2 100n 26V DC
H1 R1 15 X1
C8 C9
D-Sub
+26V 1
Free exp1 7
GND 26V DC SZ1 X12
Sync- 2
Sync+
BuCo
3
Release
GND 26V DC 5
wallstand
Free exp2 9
Start+ 6
Start- 4
GND 26V DC

V8
VAG1 VAG
VAG2 X1 X1
Brake release 14 14
switch 15VDC
A3
15 15
X5
1 1 GND BuCo 26V BuCo 1
9 9 Collimator light trigger (24V DC) 9
6 6 Collimator shutter transversal close (24V DC) 8
5 5 Collimator shutter transversal open (24V DC) 7
SZ1 X11
8 8 Collimator shutter lateral close (24V DC) 5
BuCo
7 7 Collimator shutter lateral open (24V DC) 4 (only for tracking)
VAG1 / VAG2 15VDC
Light and shutter
control User GND BuCo 2
Interface
X4 V5 R4
1 16 Tracking current (4...20mA) 6
15
12 14
VBR
Vertical unit 2
R3 A2 3
A1 1
+
potentiometer 4 13
(optional for 3 - 2,5,9
tracking)
A3 06-01-17 Schr.

GND BuCo GND BuCo

VPI
BuckyDiagnost VS Advanced
Replacement for BD. VE sensing / tracking
Interface Tracking
4512 133 0203x
BuckyDiagnost VS CSIP Level 1 (06.0) Z3-2.4
C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
X7 X3 VBX1 Vertical
3 X2
Mains 2 Brake Y1 X11 D-Sub
2
connection 1 GND Uref 3
1 1 2
230Vac Format y 5
3 3 Cass Press 2
4 Tray Close 7
X8 5 SZ1X10
3
0V GND 6 BuCo
VPL1 2 VAVA X1 7
MSB sensing
1 PO 26V X2 8
R7 R17 1
Bucky unit 2 9 PO Uref 4
F1 VT1X6 10 Format x 6
U8
X16 Multi fuse Tilt controller 11 Cass Exc 9
2,5A VDR R2 V9 Brake control
1 V10 12 +26V 1
VPN5 (for tilt) R8 13 +26V 8
V1 V2 2
26V DC 2 100n 14
26V DC
H1 R1
C8 C9 15 X1
X17 D-Sub
1 +26V 1
VTX1
2 0V GND Free exp1 7
26V DC SZ1 X12
Sync- 2
BuCo
X3 D-Sub Sync+ 3
Release
2 24V DC BuCo GND 26V DC 5
wallstand
SZ1X11 3 Free exp2 9
24V DC BuCo
BuCo Start+ 6
7 BWS not tilted
Tilt position Start- 4
8 BWS tilted
V8 GND 26V DC

X40
4 R10 U1
3 15V DC
1 R11 U2
A3
VT1X21 2 X5
Tilt controller 5 R12 U3 GND BuCo
1
26V BuCo
Collimator shutter
Collimator shutter transversal open (24V DC) 7
control,
Light control, 6 R13 U4
Tilt angle Collimator shutter transversal close (24V DC) 8

7 R14 U5 SZ1X5
BuCo
Collimator shutter lateral open (24V DC) 4 (only for
8 R15 U6 tracking)
Collimator shutter lateral close (24V DC) 5
9 R16 U7

Collimator light trigger (24V DC) 9

15VDC
GND BuCo 2
X4 V5 R4
1 16 Tracking current (4...20mA) 6
15
12 14
VBR 2
R3 A2 3
A1 1
+
A3 06-01-19 Schr.

Vertical unit 4 13
potentiometer 3 - 2,5,9
(optional for
tracking) GND BuCo GND BuCo

VPI
BuckyDiagnost VS
Replacement for BD. VT sensing
Interface Tracking
4512 133 0203x
BuckyDiagnost VS CSIP Level 1 (06.0) Z3-2.5
C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED