VIDAS®Ferritin (FER)
VIDAS Ferritin is an automated quantitative test for use on the VIDAS family instruments for the determination of human
Ferritin in human serum or plasma (lithium heparin or EDTA) using the ELFA technique (Enzyme Linked Fluorescent
Assay).
bioMérieux SA English - 1
®
VIDAS Ferritin (FER) 06036 K - en - 2015/01
Hazard statement
H318 : Causes serious eye damage.
H373 : May cause damage to organs through prolonged or repeated exposure.
H315 : Causes skin irritation.
H302 : Harmful if swallowed.
Precautionary statement
P280 :Wear protective gloves/protective clothing/eye protection/face protection.
P305 + P351 + P338: IF IN EYES: Rinse cautiously with water for several minutes. Remove contact lenses, if present
and easy to do. Continue rinsing.
P309 + P311 : IF exposed or if you feel unwell: Call a POISON CENTER or doctor/physician.
** Signal Word: DANGER
Hazard statement
H318 : Causes serious eye damage.
Precautionary statement
P280 :Wear protective gloves/protective clothing/eye protection/face protection.
P305 + P351 + P338: IF IN EYES: Rinse cautiously with water for several minutes. Remove contact lenses, if present
and easy to do. Continue rinsing.
For further information, refer to the Material Safety Data Sheet.
bioMérieux SA English - 2
®
VIDAS Ferritin (FER) 06036 K- en - 2015/01
bioMérieux SA English - 3
®
VIDAS Ferritin (FER) 06036 K - en - 2015/01
bioMérieux SA English - 4
®
VIDAS Ferritin (FER) 06036 K- en - 2015/01
Menopausal women:
Range of values 0 - 24.4 ng/ml 24.4 - 118 ng/ml 118 - 278 ng/ml Mean
Prevalence 5% 45% 45% 151 ng/ml
If concentrations are lower than 20 ng/ml in women and 30 ng/ml in men, tests for martial deficiency should be
performed. If concentrations are greater than 250 ng/ml in women or 350 ng/ml in men, tests should be performed for
inflammatory, infectious, hepatic or tumoral pathologies, or abnormal iron storage (idiopathic or secondary
hemochromatosis). “It is recommended that each laboratory establishes its own reference values from a rigorously
selected population.”
PERFORMANCE Specificity
Studies performed using VIDAS Ferritin gave the following
Tested compound Cross-reactivity (%)
results:
Spleen ferritin 106
Measurement range
Liver ferritin 121
The measurement range of the VIDAS Ferritin reagent
nd
extends from 1.5 to 1200 ng/ml (2 .IS NIBSC 80/578). Heart ferritin 28
Analytical detection limit Placental ferritin 137
Defined as the smallest concentration of ferritin which is
significantly different from the zero concentration with a
probability of 95%: < 1.5 ng/ml.
Hook effect
No hook effect was found up to ferritin concentrations of
100,000 ng/ml.
Precision
Within-run reproducibility:
5 samples were tested 30 times in a same run.
Sample 1 2 3 4 5
Dose 15.3 102 239 466 924
(ng/ml)
CV% 6.2 4.6 5.0 4.0 4.0
Between-run reproducibility
5 samples were tested singly in 24 different runs on the
same VIDAS.
Sample 1 2 3 4 5
Dose 16.5 128 234 537 1121
(ng/ml)
CV% 4.4 5.9 7.0 4.9 4.6
bioMérieux SA English - 5
®
VIDAS Ferritin (FER) 06036 K - en - 2015/01
Accuracy
Dilution test
Three samples were diluted in the (R1) diluent and tested singly in 3 series. The ratio of the mean concentration
measured over the expected concentration is expressed as a mean recovery percentage.
Comparison with other test methods 6. VERNET M., GUILLEMIN C., RYMER JC. et al., Etude
comparative de cinq méthodes d'immunodosage de la
Correlation was established between the VIDAS Ferritin Ferritine sérique chez des polytransfusés., L'Eurobiologiste.,
kit and another commercially available reagent. 1994,Tome XXVIII, N° 213, 43-311 / 49-317.
VIDAS Ferritin = 1.16 X + 10.3 r = 0.99 (n = 95) 7. KIMBER R.J., RUSAKI Z., BLUNDEN R.W., Iron deficiency
and iron overload: serum ferritin and serum iron in clinical
WASTE DISPOSAL medecine Pathology., 1983, 15, 497-503.
Dispose of used or unused reagents as well as any other 8. MONGIN M., Contexte pathologique des variations de la
contaminated disposable materials following procedures sidérémie. Feuillets de Biologie., 1988, vol. XXIX, n° 161, 49-
for infectious or potentially infectious products. 53.
It is the responsibility of each laboratory to handle waste 9. PARIS M., VERNET-NYSSEN M., DEZIER J.F., Variations
and effluents produced according to their nature and pathologiques du fer, de la transferrine et de la ferritine
degree of hazardousness and to treat and dispose of sérique.Le Pharmacien Biologiste., 1986, tome XX, n° 161,
them (or have them treated and disposed of) in 31-34.
accordance with any applicable regulations. 10. RYMER J.C., VERNET M., Dosage de la ferritine sérique.
Qualités et défauts. Immunoanal. Biol. spéc., 1990, 19, 51-
LITERATURE REFERENCES 55.
11. VERNET M.,Commission "Fer et Protéines de transport"
1. AISEN P., Iron transport and storage proteins. Ann. Rev. Evaluation de l'intérêt diagnostique de la ferritinémie en
Biochem., 1980, 49, 357-393. pathologie humaine mesurée à l'aide de divers systèmes
2. CHALLAND G.S., MICKAELDOUDIS A., WATFA R.R., actuels de réactifs prêts à l'emploi. Journée SFBC du
COLES S.J., MACKLIN J.L., Distribution of haemoglobin in 14.01.88. Information scientifique du biologiste., 1989, 15 (2),
patients presenting to their general practitioner, and its 93.
correlation with serum ferritin. Ann. Clin. Biochem., 1980, 27,
15-20.
3. IMBERT M., PRIOLET G., RYMER J.C., SULTAN Co.,
Révaluation des stratégies pour le diagnostic des carences
martiales. Ann. Biol. Clin., 1987, 45, 541-545.
4. SULTAN C., HENNY J., IMBERT M., INTRATOR L.,
JOUAULT H., Le dépistage précoce des carences martiales.
Le concours médical., 1985, 107-42, 3971-3973.
5. REVENANT M.C., VERNET M., RYMER JC. et al., Etude
comparative de six systèmes d'immunodosage de la Ferritine
sérique au cours de maladies rhumatismales.,
L'Eurobiologiste., 1994, Tome XXVIII, N°213, 35-303 / 41-
309.
bioMérieux SA English - 6
®
VIDAS Ferritin (FER) 06036 K - en - 2015/01
INDEX OF SYMBOLS
Symbol Meaning
Catalog number
Manufacturer
Temperature limit
Use by date
Batch code
Date of manufacture
WARRANTY
bioMérieux disclaims all warranties, express or implied, including any implied warranties of MERCHANTABILITY AND
FITNESS FOR A PARTICULAR USE. bioMérieux shall not be liable for any incidental or consequential damages. IN NO
EVENT SHALL BIOMERIEUX’S LIABLITY TO CUSTOMER UNDER ANY CLAIM EXCEED A REFUND OF THE
AMOUNT PAID TO BIOMERIEUX FOR THE PRODUCT OR SERVICE WHICH IS THE SUBJECT OF THE CLAIM.
REVISION HISTORY
Change type categories :
N/A Not applicable (First publication)
Correction Correction of documentation anomalies
Technical change Addition, revision and/or removal of information related to the product
Administrative Implementation of non-technical changes noticeable to the user
Note: Minor typographical, grammar, and formatting changes are not included in the
revision history.
Release
Part Number Change Type Change Summary
date
INDEX OF SYMBOLS
Administrative REVISION HISTORY
2015/01 06036K
CONTENT OF THE KIT (60 TESTS)
Technical WARNINGS AND PRECAUTIONS
BIOMERIEUX, the blue logo, SPR and VIDAS are used, pending and/or registered trademarks belonging to bioMérieux or one of its
subsidiaries, or one of its companies.
Any other name or trademark is the property of its respective owner.