Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

IN0277022105V1

POC βhCG infinosis™


Beta-Human Chorionic Gonadotropin
REF: IN027702 25 tests • For in vitro diagnostic use only. For professional use only. Add 20 μL of plasma/serum to the buffer tube.
• Carefully follow the instructions and procedures described in this instructions Step 3: Mixing
Please note before testing. Mix the specimen well with buffer for 5-10 seconds by tapping or inverting the
The measured hCG value of a patient’s sample can vary depending on the • The test cassette should remain in its original sealed pouch until ready to use. tube.
testing procedure used. The laboratory finding must therefore always contain a Do not use it if the pouch is damaged or the seal is broken.
Step 4: Loading
statement on the hCG assay method used. hCG values determined on patient • Do not use reagents beyond the labeled expiry date. Add 80 μL of sample mixture and load it onto the sample well of the test cassette.
samples by different testing procedures cannot be directly compared with one • Do not mix or use components from kits with different Lots.
another and could be the cause of erroneous medical interpretations. • Don’t use Test Cassette if its Lot does not match with ID Chip that is inserted Step 5: Testing
onto the instrument. Put the Test Cassette on the operation platform. 15 minutes later, insert the Test
If there is a change in the hCG assay procedure used while monitoring therapy,
• The Infinosis™ βhCG assay should be used with Infinosis™ 2020 FIA analyzer. Cassette onto the Cassette Holder and click “Test”. The result will show in the
then the hCG values obtained upon changing over to the new procedure must be
• The tests should be applied by professionally trained staff working in certified display and print out when click “Print”.
confirmed by parallel measurements with both methods.
laboratories at some remove from the patient and clinic at which the sample is Please refer to the Infinosis™ 2020 FIA analyzer Operation Manual for details.
taken by qualified medical personnel.
Intended use Infinosis™ βhCG assay is single use only. Do not re-use it.
The Infinosis™ βhCG is an immunofluorescence-based lateral flow assay for the • Limitations - interference
• The Test Cassette and Analyzer should be used away from vibration and
in vitro quantitative determination of the human chorionic gonadotropin (hCG) plus • The assay is unaffected by icterus (bilirubin < 410 μmol/L or < 24 mg/dL),
magnetic field. During normal usage, the Test Cassette may introduce minute
the hCG β‐subunit in human whole blood, serum or plasma. hemolysis (Hb < 0.621 mmol/L or < 1.0 g/dL), lipemia (Intralipid < 1400 mg/dL)
vibration, which should be regarded as normal.
and biotin (< 327 nmol/L or < 80 ng/mL).
• Use separate clean pipette tips and detector buffer vials for different specimens.
Summary1-5 The pipette tips and detector buffer vials should be used for one specimen only. • Criterion: Recovery within ± 10 % of initial value.
Similarly to LH (Luteinizing hormone), FSH (Follicle‐stimulating hormone) and • No interference was observed from rheumatoid factors up to a concentration of
• Do not smoke, eat, or drink in areas in which specimens or kit reagents are
TSH (Thyroid‐stimulating hormone), human chorionic gonadotropin (hCG) is a 3400 IU/mL and samples from dialysis patients.
handled.
member of the glycoprotein family and consists of 2 subunits (α‐1 and β‐chains) • There is no high-dose hook effect at hCG concentrations up to 1000000 mIU/
• Blood specimens, used test cassettes, pipette tips and detector buffer vials are
which are associated to form the intact hormone. The α‐chains in all four of these mL.
potentially infectious. Proper laboratory safety techniques, handing and
glycoprotein hormones are virtually identical, whereas the β‐chains have greatly • For diagnostic purposes, the results should always be assessed in conjunction
disposal methods should be followed in accordance with standard procedures
differing structures and are responsible for the respective specific hormonal with the patient’s medical history, clinical examination and other findings.
and relevant regulations observed by microbiological hazard materials.
functions. The results should be interpreted by the physician along with clinical findings
• Limits and ranges
hCG is produced in the placenta during pregnancy. In non‐pregnant women, it can and other laboratory test results.
also be produced by tumors of the trophoblast, germ cell tumors with trophoblastic The test will be applied on a routine basis and not in emergency situations. Measuring range

components and some non‐trophoblastic tumors. 0.1‐10000 mIU/mL (defined by the lower detection limit and the maximum of the
hCG assays detecting the intact hCG plus the free β‐subunit are well established Storage and stability master curve). Values below the lower detection limit are reported as < 0.1 mIU/
markers as an aid in the management of patients with trophoblastic tumors and mL. Values above the measuring range are reported as
together with AFP in patients with testicular and other germ cell tumors. • Store the extraction buffer at 4-30°C. The buffer is stable up to 24 months. > 10000 mIU/mL.
• Store Infinosis™ test cassette at 4-30°C, shelf life is up to 24 months.
Test principle
• Test cassette should be used within 1 hour after opening the pack. Lower limits of measurement
Sandwich principle. Total duration of assay: 15 minutes Lower detection limit
Sample is added to the sample well of the test, then the fluorescence-labeled
Specimen collection and preparation Lower detection limit: 0.1 mIU/mL

The test can be performed with either serum or plasma.
detector anti-βhCG antibody binds to βhCG antigen in blood specimen. As the The detection limit represents the lowest analyte level that can be distinguished
sample mixture migrates on the nitrocellulose matrix of test strip by capillary • Collect serum samples in accordance with correct medical practices. from zero. It is calculated as the value lying two standard deviations above that of
action, the complexes of detector antibody and βhCG are captured to anti-βhCG • Using standard phlebotomy procedure, collect a venipuncture whole blood the lowest standard (master calibrator, standard 1 + 2 SD, repeatability study, n =
specimen using a blood collection tube. If collecting plasma use a blood
antibody that has been immobilized on test strip. 21).
collection tube containing suitable anticoagulant (EDTA recommended).
The more βhCG antigen is in blood specimen, the more complexes are
accumulated on test strip. Signal intensity of fluorescence of detector antibody • Separate the serum/plasma from blood as soon as possible to avoid hemolysis.
• Test should be performed immediately after the specimens have been collected.
Dilution
reflects amount of βhCG captured and Infinosis™ FIA analyzer shows βhCG Samples with hCG concentrations above the measuring range can be diluted with
Do not leave the specimens at room temperature for prolonged periods.
concentrations in blood specimen. The default results unit of Infinosis™ βhCG Sample Dilute. The recommended dilution is 1:100. The concentration of the
Specimens may be stored at 2-8°C for up to 3 days. For long-term storage,
test is displayed as x mIU/mL from Infinosis™ FIA analyzer. diluted sample must be > 200 mIU/mL.
specimens should be kept below -20°C.
Patient Sample Dilution(1:100): Dispense 0.010 mL(10 uL) of each patient
Reagents specimen into 1 mL of sample diluent. Cover and vortex or min thoroughly by
Materials provided Quality control inversion. Testing must be completed within 8 hours.
Each Infinosis™ βhCG test cassette contains internal control that satisfies routing
• Test cassette, 25 pcs, individually packaged
quality control requirements. This internal control is performed each time a patient
• ID chip, 1 pcs Expected values
sample is tested. This control indicates that the test cartridge was inserted and
• Sample buffer, 25 vials <5.0 mIU/mL
read properly by Infinosis™ 2020 FIA analyzer. An invalid result from the internal
• IFU, 1 copy Results from a multicenter study in clinical centers are listed below.
control causes an error message on Infinosis™ 2020 FIA analyzer indicating that
the test should be repeated. • ≤ 3 mIU/mL hCG for 97.5 % of the values obtained from 286 healthy, non‐
Materials required (but not provided) pregnant premenopausal women. The corresponding upper 95 % confidence
• Infinosis™ 2020 FIA analyzer limit ranges up to 5.5 mIU/mL.
• Transfer pipette set (100 μL size) Test procedure
• ≤ 7 mIU/mL hCG for 97.5 % of the values obtained from 133 healthy,
• Alcohol pads Refer to Infinosis™ 2020 FIA analyzer Operation Manual for the complete postmenopausal women. The corresponding upper 95 % confidence limit
• Centrifuge (for plasma and serum only) instructions on use of the test. The test should be operated in room temperature. ranges up to 8.5 mIU/mL.
• Timer Step 1: Preparation • < 2.5 mIU/mL hCG for 97.5 % of the values obtained from 369 men. The
Check/insert ID Chip into the analyzer. corresponding upper 95 % confidence limit ranges up to 3.0 mIU/mL.
Precautions and warnings Step 2: Sampling

202105 V1 English 1/2 Infinosis™ βhCG/IFU


IN0277022105V1

POC βhCG infinosis™


Beta-Human Chorionic Gonadotropin

• During pregnancy (weeks of pregnancy - defined as completed weeks of Method comparison


pregnancy beginning with the start of the last menstruation phase), the following A comparison of the Infinosis™ βhCG assay (y) with the Roche Elecsys HCG+β
values have been determined. (x) using clinical samples gave the following correlations:
Data are given only for the weeks of gestation for which the case numbers (n)
were greater than 10.
Number of samples measured: 98
Linear regression
diasino
y = 1.036x - 1.095 DiaSino Laboratories Co., Ltd
Weeks of gestation N hCG mIU/mL r = 0.9882 No.68, Jingnansi Road, National Eco & Tech Development Area
Zhengzhou, China. 450000
Median 5-95th percentile The sample concentrations were between approx. 1 and 18550 mIU/mL. Technical Support: ts@diasino.com
3 17 16.8 6.6-58.3 www.diasino.com
4 57 158 11.4-692 Analytical specificity
5 41 1477 198-8009 For the monoclonal antibodies used, the following cross-reactivities were found:
TSH: not detectable, LH 0.12 %, FSH < 0.1 %.
6 39 3899 265->20000
7 25 >20000 3787->20000 Functional sensitivity
8-12 17 >20000 >20000 < 1.8 mIU/mL

The maternal age and weight and the gestational age in days was given for each The functional sensitivity is the lowest analyte concentration that can be
sample. reproducibly measured with an intermediate precision CV of 20 %.
Median values and the 5th and 95th percentile were calculated for the completed
gestational weeks - see within the table above.
References
1. Schwarz S, Berger P, Wick G. The Antigenic Surface of Human Chorionic
Distribution of Infinosis™ βhCG results from healthy subjects and patients with Gonadotropin as Mapped by Murine Monoclonal Antibodies. Endocrinology
benign and malignant diseases: 1986;118(1):189-197.
The results from patients with benign and malignant diseases are summarized 2. Sturgeon CM, McAllister EJ. Analysis of hCG: clinical applications and assay
data from measurements with the Infinosis™ βhCG+β assay requirements. Ann Clin Biochem 1998;35:460-491.
Cencentration 3. Hoermann R, Berger P, Spoettl G, et al. Immunological Recognition and Clinical
N Percent (%)
mIU/mL Significance of Nicked Human Chorionic Gonadotropin in Testicular Cancer.
Healthy subjects 788 ≤4 >4 - ≤ 8 >8 - ≤ 100 >100 >1000 Clin Chem 1994;40(12):2306-2312.
Males 369 98.92 1.08 4. Mann K, Hörmann R. hCG (human chorionic gonadotropin). In: Thomas L (ed.).
Females 286 98.60 1.40 Clinical Laboratory Diagnosis, TH-Books, Frankfurt, 1st English edition
Premenopause 1998:971-976, 8th German edition 2012:1668-1669.
Females 133 66.17 33.08 0.75 5. Sturgeon C. Practice Guidelines for Tumor Marker Use in the Clinic. Clin Chem
Postmenopause 2002;48(8):1151-1159.

Malignant 937
diseases In vitro diagnostic Refer to instruction
Chorioncarcinoma 71 9.86 11.27 21.13 9.86 47.89 use for use
Seminoma 64 85.94 6.25 7.81 0 0
Germ cell tumor 82 79.27 2.44 1.22 6.10 10.98
Yolk sac tumor 32 21.88 6.25 21.88 9.38 40.63 Expiry date Manufacturing date
Ovarian cancer 37 70.27 21.62 8.11 0 0
Gestational 196 20.92 10.71 31.63 20.92 15.82
trophoblastic Batch number Test per kit
diseases
Mole 91 2.20 13.19 23.08 21.98 39.56
Others 364 51.65 9.90 11.54 12.36 14.55
Catalog number Do not re-use
Each laboratory should investigate the transferability of the expected values to its
own patient population and if necessary determine its own reference ranges.
Store between
Specific performance data Manufacturer
4-30℃
Representative performance data are given below. Results obtained in individual
laboratories may differ.

Precision
Intra-assay
Determined by using 10 tests in the same batch to test with TSH control, CV ≤
15%
Inter-assay
Determined by using 3 tests in 3 random and continuous batches to test with TSH
control, CV ≤ 20%

202105 V1 English 2/2 Infinosis™ βhCG/IFU

You might also like