CM-D-001 Calibration Method For Caliper
CM-D-001 Calibration Method For Caliper
CM-D-001
) ( Copy No
اﻟﺘﻮﻗﯿﻊ: اﻟﻮظﯿﻔﺔ :اﻟﻤﺪﯾﺮ اﻟﻌﺎم اﻻﺳﻢ :راﻣﻲ ﻣﺤﻤﻮد زﯾﺎدة اﻋﺘﻤﺎد اﻟﻤﺪﯾﺮ اﻟﻌﺎم
1 Introduction
1-1 The objective of the equipment calibration is for reliable and accurate results.
2 Purpose
2-1 This procedure describes the calibration method of all types calipers as
specified in DIN 862 standard
3 Scope
This method applies to all types of caliper (vernier –dial-digital) also the depth caliper up to range 600
mm.
4 Definitions/Abbreviations
4-1: Vernier caliper:
A measuring device used for external and internal measurements that consists of a main scale
with a fixed jaw and a sliding jaw with an attached vernier scale.
5 Calibration scope
7 Procedure
7-1 preparing:
a) The measuring surfaces of the gauge blocks and the gauge measuring surfaces
for the caliper are cleaned using petroleum ether. Any grit or other particles are also cleaned from the
scale. The surfaces are wiped afterwards using the lint free tissue (or special synthetic cloth).
b) Measuring surfaces are checked to see if there are any scratches which may impede calibration.
c) The main and the vernier Scales are checked for any marking points or numbers on the scale which may
be missing or worn.
d) Dial scale is checked for the straightness of the pointer and its distance from the scale, The pointer must
be of the same width as the lines on the scale.
Main office: 41 El Esraa St. with 26 July St. – El Mohandseen – Giza – Egypt Page 2 of 5
Lab: 37a Sayeda Aisha St. - 7th district - 6th October city - Egypt
Tel./Fax: (002) 0233034440 Mob: (002) 01023310801/2
www.tagegypt.com info@tagegypt.com
Calibration Method
Issue No.: 01.0
Issue Date: 03 Jan 2021 For All type of Caliper
Doc. No: CM-D-001
e) Digital scale is checked for numerical legibility and for any scratches on the display, If the numbers are
of poor resolution (checked at display value 88.88) the battery is changed and resolution is checked again,
the numbers must be clearly visible in each measuring position.
f) The fit between slider and beam shall permit the slider to be moved smoothly over the entire length of the
beam.
Main office: 41 El Esraa St. with 26 July St. – El Mohandseen – Giza – Egypt Page 3 of 5
Lab: 37a Sayeda Aisha St. - 7th district - 6th October city - Egypt
Tel./Fax: (002) 0233034440 Mob: (002) 01023310801/2
www.tagegypt.com info@tagegypt.com
Calibration Method
Issue No.: 01.0
Issue Date: 03 Jan 2021 For All type of Caliper
Doc. No: CM-D-001
a) By the end of calibration process re-clean the caliper by cotton immersed in benzene then by dry cloth.
b) Cover the caliper jaws only by thin layer of oil or Vaseline.
c) Put the caliper in its case or wooden box.
d) Put the sticker of calibration.
8 Estimation of The Measurement Uncertainty
8-1 The uncertainty budget for thermometers calibration as in table blew
Uncertainty Budget for calibration 0-150 mm Digital Caliper
Uncertainty source estimate degree of freedom Type distribution divisor standard uncertainty Variance
master/part
temperature 0.86 ∞ B rectangular 1.732051 0.496521232 0.246533
difference
Summation 8.738267
Combined standard
uncertainty
u 2.956056
Main office: 41 El Esraa St. with 26 July St. – El Mohandseen – Giza – Egypt Page 4 of 5
Lab: 37a Sayeda Aisha St. - 7th district - 6th October city - Egypt
Tel./Fax: (002) 0233034440 Mob: (002) 01023310801/2
www.tagegypt.com info@tagegypt.com
Calibration Method
Issue No.: 01.0
Issue Date: 03 Jan 2021 For All type of Caliper
Doc. No: CM-D-001
9 References
9-1 DIN 862 standard
9-2 Refer to manufacturer’s manual for additional information.
9-3 ISO 17025 International Standard.
9-4 M3003.