You are on page 1of 3

Water inlet

H01
Flush Drain
H02
Flow diagram 5008S (all options)
Drain
F01
H19
VF2
A04

V02 Compressor
EBM
V19 V21

S03 F06

A05
V30
V11 V12 V13 V14
PT7 PT8
P H15
F04 Temp V23

CD7 V24 H42


H16
V38 H04
H14
F05 H36
A02
H09

PT4 V22 V26 V31


VF1 V29s
V03 Temp S10
CD4 S27 Shunt
V28 H32

Dialyzer
V15 V16 V17 V18 S28 S04
S05
H03 S09
Temp

H07 S06
PT2 P05 P06
S01
S16
S15 P S36
H18
V32

P04 CD9
S29
P M
S38
S07 S30
H33
H37

Temp
H08
CD5 A03 P03 S11 PT9 P V33
F08 V25 F07
H43
V09 H17
S08
Temp
P01 H42
PT5

H06

V50 V51

V04 S50 S51

H50 H51
H05
S35
V35s

V41 V40 V36


S53

VB1 VS1
F12 F11
V20 V34 H23
S20 S18
VB2 VS2
F10 F16
S19 S17

H20 H35 bibag H22 H21 Safety System


D1 D2 S22
S21
Operating System
S47

VS3s

VERSION 1.3
VS4s

VS5s

VS6s

F14

H29
CDS
Chapter 11: Functional description

O Legend

Explanation of symbols

Axx Elements for adjustment PTx Temperature sensors


CDx Conductivity cells Sxx Other sensors
Fxx Filters Vxx Valves
Hxx Hydraulic components Vxxs Small valves (s)
Pxx Pumps

Legend

A02 Loading pressure valve, balancing chamber P01 Degassing pump


A03 Relief valve P03 Flow pump
A04 Pressure reducing valve, water inlet P04 UF pump
A05 Check valve, water inlet P05 Bicarbonate pump
P06 Concentrate pump
CD4 Conductivity cell, bicarbonate
CD5 Conductivity cell, bicarbonate PT2 Sensor, temperature control
CD7 Conductivity cell, overall conductivity PT4 Temperature sensor, CD bicarbonate
CD9 Conductivity cell, OCM PT5 Temperature sensor, CD bicarbonate
PT7 Temperature sensor, overall CD and temperature display
F01 Filter, water inlet PT8 Sensor, temperature re-adjustment
F04 Filter stage 1 PT9 Temperature sensor, CD OCM
F05 Filter stage 2
F06 Hydrophobic filter, compressor EBM S01 Float switch, water inlet chamber
F07 Particle filter, dialysate S03 Pressure transducer, dialyzer inlet
F08 Filter, UF pump S04 Sensor, shunt interlock inlet
F10 Filter, disinfectant S05 Sensor, shunt interlock open
F11 Suction tube, concentrate with filter S06 Sensor, shunt interlock outlet
F12 Suction tube, bicarbonate with filter S07 Pressure transducer, dialyzer outlet
F14 Filter, CDS S08 Blood leak detector
F16 Filter, disinfectant 2 S09 Level sensor, secondary air separator
S10 Sensor, air separation valve
H01 Male connector, water inlet S11 UF pump, CD inlet
H02 Male connector, Flush S15 Pressure transducer, balancing chamber switching
H03 Water inlet chamber S16 Pressure transducer, fill dry concentrate bag
H04 Heat exchanger S17 Level sensor, concentrate rinse chamber
H05 Degassing orifice S18 Sensor, concentrate rinse chamber lock
H06 Degassing chamber S19 Level sensor, rinse chamber, bicarbonate
H07 Heater rod chamber S20 Sensor, bicarbonate rinse chamber lock
H08 Heater rod S21 bibag® flap, bag operation position
H09 Primary air separator S22 bibag® connected
H14 Balancing chamber S27 Sensor 1, substituate port
H15 Sampling valve S28 Sensor 2, substituate port
H16 Dialyzer coupling, to dialyzer S29 Sensor 1, rinse port
H17 Dialyzer coupling, from dialyzer S30 Sensor 2, rinse port
H18 Secondary air separator S35 Leakage sensor, hydraulics
H19 Male connector, drain S36 UF pump, CD outlet
H20 Male connector, disinfectant S38 UF pump, monitoring
H21 Rinse chamber, concentrate S47 bibag® flap, cleaning position
H22 Rinse chamber, bicarbonate S50 Bicarbonate air separator sensor
H23 Connector, bibag® S51 Concentrate air separator sensor
H29 Male connector, CDS S53 sensor for suction tube flap
H32 Substituate port
H33 Rinse port
H34 Male connector, compressor EBM
H35 Male connector, disinfectant 2
H36 Male connector, dialyzer coupling, to dialyzer
H37 Male connector, dialyzer coupling, from dialyzer
H42 Connector, potential equalization
H43 Male connector, vent water inlet chamber
H50 Air separation chamber, bicarbonate (continued next page)
H51 Air separation chamber, concentrate

Fresenius Medical Care 5008S SM 9/06.09 11-15


Chapter 10: Repair

V02 Water inlet valve


V03 Recirculation valve, cleaning
V04 Degassing orifice bypass valve
V09 Fill valve, dry concentrate bag
V11 Balancing chamber valve
V12 Balancing chamber valve
V13 Balancing chamber valve
V14 Balancing chamber valve
V15 Balancing chamber valve
V16 Balancing chamber valve
V17 Balancing chamber valve
V18 Balancing chamber valve
V19 Drain valve
V20 Disinfection valve
V21 Vent valve, filter stage 1
V22 Retentate valve
V23 Test valve
V24 Dialyzer inlet valve
V25 Dialyzer outlet valve
V26 Bypass valve
V28 Fill valve, secondary air separator
V29s Air separation valve
V30 Outlet valve
V31 Substituate valve
V32 Rinse valve 1, H(D)F
V33 Rinse valve 2, H(D)F
V34 Disinfection valve
V35s Negative pressure valve 2
V36 Rinse valve
V38 Bypass valve, heat exchanger
V40 Shutoff valve, disinfection
V41 Rinse valve, disinfection
V50 Sensor, bicarbonate air separation valve
S51 Sensor, concentrate air separation valve
V53 Sensor for suction tube flap

VB1 Fill valve, bibag®


VB2 Deaeration valve, bicarbonate rinse chamber

VF1 Flush valve


VF2 Flush valve, drain

VS1 Rinse valve, concentrate rinse chamber


VS2 Deaeration valve, concentrate rinse chamber
VS3s CDS, shutoff valve OS
VS4s CDS, shutoff valve SS
VS5s CDS valve, test
VS6s CDS valve, inlet

Fresenius Medical Care 5008S SM 9/06.09 10-9