How to Simulate Shelf Life


for Ageing Trials
Table I: The Arrhenius equation.
Arrhenius’rate equation is an equation
giving the rate R of a thermally activated,
physical process:R = Ro exp(Ea/kT)
where Ro is a constant,Ea is the activation
energy,k is Boltzmann’s constant,and T is
the absolute temperature.
A series of guides to tackling manufacturing problem areas.
Source:Hawley’s Condensed Chemical Dictionary,
13th Edition, John Wiley and Sons Ltd,Chichester,UK

A ccording to ISO 11607, Packag-


ing for Terminally Sterilised
Medical Devices, and United States
of the rate of chemical reaction. It is
then assumed that doubling the rate
of chemical reaction doubles the rate
been sterilised:
■ Chemical characterisation signature
of system components
Food and Drug Administration regula- of decay of packaging and product. ■ Seal strength of pack welds
tions, medical devices must be given The accelerated ageing factor is ■ Puncture and abrasion resistance
an expiry date, which indicates how known as Q10. A Q10 figure of 2 is ■ Porosity of the protective barrier
long they may be stored prior to use. considered conservative. Higher ■ Does the product and packaging
It is necessary to substantiate any shelf- figures may be used if they are shown still look good?
life claims by verifying product and to be appropriate by testing or if
package performance following ageing. literature information justifies it. Product tests
Real-time ageing is the best way to do Clearly, there is a limit to the The manufacturer should demon-
this, but no organisation can afford to temperature that can be applied to a strate that the product performs to
delay its product launch by five years medical product or package that is specification following ageing. Nor-
whilst waiting for the results of real- made of plastic. Accelerated ageing- mally, this requirement is contained
time ageing tests. A method of accel- must be carried out below the glass within quality assurance procedures,
erated ageing is required that realistically transition temperature of any of the which would have been written
recreates what a product may experi- components of the system. It is during product development.
ence during storage. generally accepted that 60 °C is the
To limit the range of conditions a maximum temperature that is suitable Summary
product may be subject to, it is common for most products (ASTM F1980). Medical device manufacturers are
to detail storage recommendations on compelled to validate the shelf-life
the packaging and/or in the user instruc- The process claims of their products. If the materi-
tions. However, it must be assumed A typical cycling programme would als used in the product and packaging
that products will be transported by be one year of real-time ageing where can withstand elevated temperatures,
air and road and that perfect conditions Q10 is 2. This is equivalent to 23 days accelerated ageing regimes may be
cannot be maintained at all times. at 55 °C and 70% relative humidity, 3 used to simulate real-time ageing. If
days at –20 °C, and 23 days at 55 °C elevated temperatures are detrimental
The principle and below 20% relative humidity. to the components involved, then
Accelerated ageing is achieved by As with any scientific testing it is real-time ageing is the only option. All
storing the product at an elevated necessary to begin by defining what testing should be carried out on
temperature. Better still, temperature is being tested and the protocol for sterile product when appropriate. mdt
cycling is used because this more testing. There are two things that
closely simulates real life and low simulated ageing should address: Further reading
temperatures can cause damage. The ■ Does the packaging maintain the ■ ASTM F1980 Standard Guide For
flexing caused by temperature cycling sterility of the product? Accelerated Ageing of Sterile Medical
is much closer to the real stresses ■ Does the product still function to Device Packages.
experienced by the product and specification?
packaging over time than storage at There is likely also to be a market-
constant elevated temperature. In ing requirement that the presentation Mark Turner
addition, the effects of humidity can of the product remains unchanged. is Sales Manager for Medical Engineering
be important in some instances. Technologies Ltd, a provider of engineering
The Arrhenius equation is often Packaging tests and scientific services, Hydra House,
used to calculate the factor for The following characteristics should 26 North Street, Ashford TN24 8JR, UK,
increased ageing (Table I). This states be verified before and after acceler- tel. +44 1233 649 204, fax +44 870 056 2153,
that a rise in temperature of 10 °C ated ageing, this ageing and verifica- e-mail: m.turner@met.uk.com
will cause approximately a doubling tion taking place after the product has www.met.uk.com