Download as pdf or txt
Download as pdf or txt
You are on page 1of 10

This is an additional procedure of the

SPECTRALIS
Service Manual
Version 6
Art. No. 96 049-006

Addition 1
Spectralis touch panel: new components, compatibility, and replacement

Heidelberg Engineering GmbH


Tiergartenstraße 15
D-69121 Heidelberg, Germany

Tel.: +49 / 6221 / 6463-0


Fax: +49 / 6221 / 6463-62
support@heidelbergengineering.com
Additional procedure of Spectralis Service Manual
Art. No.: 96 083-001 of 26 August 2011 Page 3 / 10 96049-006_SPECTRALIS_Service
Manual_V006_Addition1_EN.doc

Table of Contents
1.1.1 Purpose ................................................................................................................... 5
1.1.2 Versions .................................................................................................................. 5
1.1.3 How to identify the touch panel version................................................................. 5
1.1.4 HRA2 / Spectralis TSP repair................................................................................. 6
Additional procedure of Spectralis Service Manual
Art. No.: 96 083-001 of 26 August 2011 Page 4 / 10 96049-006_SPECTRALIS_Service
Manual_V006_Addition1_EN.doc

CAUTION!
ESD (electrostatic discharge)
Only work in Electrostatic Protected Area
Use in the field an ESD service kit

Electrostatic discharge (ESD) is the sudden and momentary electric current that flows
between two objects at different electrical potentials. Prevention of ESD bases on Electrostatic
Protective Area (EPA). EPA can be a small working station or a large manufacturing area. The
main principle of an EPA is that there are no highly charging materials in the vicinity of ESD
sensitive electronics, all conductive materials are grounded, workers are grounded, and charge
build-up on ESD sensitive electronics is prevented. For service jobs in non-electrostatic
protected areas we highly recommend the use of an ESD service kit. With this kit the worker is
grounded, and the electronic boards and components should be placed on the grounded ESD
mat.
Additional procedure of Spectralis Service Manual
Art. No.: 96 083-001 of 26 August 2011 Page 5 / 10 96049-006_SPECTRALIS_Service
Manual_V006_Addition1_EN.doc

1.1.1 Purpose

From now the Spectralis touch panel will be delivered with new internal components. This
procedure describes the changes, how to identify the versions, the new part numbers and how
to replace those components.

1.1.2 Versions
In the past we delivered version 2.1 and lower. Now we deliver version 3.1.

Version 2.1 and lower:


 LCD module (p/n: 30487) is not available anymore
 Encoder (p/n 40945) is not available anymore

Version 3.1:
 LCD module, new p/n 30487-003
 Encoder, new p/n 40975-003

1.1.3 How to identify the touch panel version


Go to service screen, see chapter 6.4 Touch Panel Service Screens, select “Touchp.”

If the touch panel does not work at all, you have to open it and identify the version.
Additional procedure of Spectralis Service Manual
Art. No.: 96 083-001 of 26 August 2011 Page 6 / 10 96049-006_SPECTRALIS_Service
Manual_V006_Addition1_EN.doc

1.1.4 HRA2 / Spectralis TSP repair


Encoder replacement

1. Check TSP PCB-Board Revision (see chapter 6.4 Touch Panel Service Screens)
select “Touchp.”

for Board Rev 3.1 you need:


Spare part: 40945-003 TSP Encoder

for Board Rev 2.1 (or lower) you need:


Spare part: 40945-003 TSP Encoder
32072 TSP Adapter Kit

Encoder Adapter Kit (Board with Cable)

These tools are required:


Screwdrivers, Hexagon socket key (size 12)
Additional procedure of Spectralis Service Manual
Art. No.: 96 083-001 of 26 August 2011 Page 7 / 10 96049-006_SPECTRALIS_Service
Manual_V006_Addition1_EN.doc

2. Remove all seven screws on the backside of the TSP

3. Remove the black brightness control knob by pushing a screwdriver carefully through
the two holes next to the TSP encoder
Additional procedure of Spectralis Service Manual
Art. No.: 96 083-001 of 26 August 2011 Page 8 / 10 96049-006_SPECTRALIS_Service
Manual_V006_Addition1_EN.doc

4. Remove the screw nut with the socket key

5. Disconnect encoder cable

 
Additional procedure of Spectralis Service Manual
Art. No.: 96 083-001 of 26 August 2011 Page 9 / 10 96049-006_SPECTRALIS_Service
Manual_V006_Addition1_EN.doc

6. Insert encoder with washer and mount it with lock washer and screw nut
Turn encoder into the correct position and tighten nut

7. Board Revision 3.1:


Reconnect encoder cable to TSP Board (check correct orientation of the plug)
Additional procedure of Spectralis Service Manual
Art. No.: 96 083-001 of 26 August 2011 Page 10 / 10 96049-006_SPECTRALIS_Service
Manual_V006_Addition1_EN.doc

8. Board Revision 2.1 (or lower):


Connect encoder cable and TSP Adapter Cable to TSP Adapter Board
(check correct orientation of the plug)
Use screw M3x6 with lock washer and mount board to thread standoff

9. Connect TSP Adapter Cable to TSP Board (check correct orientation of the plug)