Professional Documents
Culture Documents
Immulite 2000 - PM Manual
Immulite 2000 - PM Manual
Immulite 2000 - PM Manual
Maintenance Instructions
System
Preventive Maintenance Instructions
Document Version
Siemens Healthcare GmbH reserves the right to change its products and services at any
time.
In addition, manuals are subject to change without notice. The hardcopy documents cor‐
respond to the version at the time of system delivery and/or printout. Versions to hard‐
copy documentation are not automatically distributed.
Please contact your local Siemens Healthineers office to order a current version or refer to
our website http://www.siemens-healthineers.com.
Disclaimer
Siemens Healthcare GmbH provides this documentation “as is” without the assumption of
any liability under any theory of law.
The content described herein shall be used by qualified personnel who are employed by
Siemens Healthcare GmbH or one of its affiliates or who are otherwise authorized by Sie‐
mens Healthcare GmbH or its affiliates to use such documents.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens Healthcare GmbH or one of its affiliates are not entitled to
use this documentation without prior written authority.
Copyright
“© Siemens Healthineers, 2020” refers to the copyright of a Siemens Healthineers entity
such as:
Siemens Healthcare GmbH - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Healthcare Private Ltd. - India
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA
Siemens Healthcare Diagnostics Products GmbH - Germany
Corindus Inc. - USA
1 Overview 4
2 PM Checklist 6
2.1 Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2 Power Supply Voltages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.2.1 Checking the Voltages at the DC Distribution PCB . . . . . . . . . . . . 8
2.2.2 Checking the Voltage at the Sample Carousel PCB . . . . . . . . . . . . 8
2.2.3 Checking the Voltages at the PMT Power Supply . . . . . . . . . . . . . 8
2.3 Dual Resolution Diluter (DRD) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3.1 Replacing the DRD Seals and Springs . . . . . . . . . . . . . . . . . . . . . . 9
2.3.2 Lubricating the DRD. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.4 Dual Pipettor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.4.1 Checking the Sample and Reagent Pipettor Probes . . . . . . . . . . . 11
2.4.2 Verifying Tube Bottom Position of Sample Arm Z . . . . . . . . . . . . . 11
2.5 Bead Carousel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.6 Tube Hopper Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.7 Reagent Carousel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.8 Tube Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.9 Sample Carousel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.10 Solenoid and Linear Actuator Pumps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2.10.1 Performing a Volume Check for Solenoid and Linear Actuator 19
Pumps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.10.2 Adjusting the Drawback of the Substrate and Water Pumps . . . . 19
2.11 Fluidics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
2.11.1 Replacing Filters and Tubing . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
2.11.2 Replacing the Vacuum Pump Trap . . . . . . . . . . . . . . . . . . . . . . . . 20
2.11.3 Verifying the Vacuum Pump Function . . . . . . . . . . . . . . . . . . . . . 21
2.11.4 Turning on Clot Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2.11.5 Checking the Transducer Waveform . . . . . . . . . . . . . . . . . . . . . . . 22
2.12 Water Test PM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
2.13 Computer Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
2.13.1 General Software Check and Cleanup . . . . . . . . . . . . . . . . . . . . . . 27
2.13.2 Verifying Mirror Mode Functionality . . . . . . . . . . . . . . . . . . . . . . . 27
2.13.3 Scanning the Control-Side Hard Drive for Problems . . . . . . . . . . . 28
2.13.4 Defragmenting the User-Side Hard Drive . . . . . . . . . . . . . . . . . . . 29
2.14 System Temperatures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
2.15 Final Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.1 Version 08 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
2.1 Preparation
WARNING
The instrument could be contaminated with human body fluids.
Wear personal protective equipment and use universal precautions.
1. Enter the Detailed View of the Event Log and look for any unusual activity. If there are
any recurring errors, closely inspect the affected module.
2. Discuss the instrument’s past performance with the customer.
3. Verify that the UPS sustains power when it is unplugged:
a) Remove the UPS power plug from the wall outlet.
b) Confirm that the backup battery light comes on and the IMMULITE 2000/2000 XPi
keeps running.
c) Plug the UPS back into the wall outlet.
CAUTION
Performing preventive maintenance while the user computer is logged in to Win‐
dows may damage the operating system and the IMMULITE software.
To prevent any software damage, make sure that Windows XP or Windows NT dis‐
plays the login screen.
Restart Windows, but do not log in.
4. With the system power off, make sure power connections are secure and free of de‐
fects and that there are no objects on or around the uninterruptible power supply
(UPS).
5. Visually inspect all areas of the system for any signs of potential problems.
- Inspect the pipetting area for any signs of fluidic leaks.
- Inspect the tube feeder area for any signs of dropped beads or dropped tubes
around the hopper.
6. Remove and clean the main fan filter.
- The main fan filter is located on the right side of the system next to the electronic
chassis.
- Clean the filter by holding it under running water and rubbing gently to remove
any dust.
- Make sure the filter is dry before reinstalling it.
7. Remove the dilution well insert and install the replacement included in the PM kit.
Make sure the insert is securely seated in the brass fitting by pressing down on it.
The dilution well sample pipettor positions must be verified using the procedures in the
Alignments section ( Alignments / DCIM-B02.842.01).
8. Clean the trackball with rubbing alcohol using cotton swabs or a lint-free cloth, and
make sure the area where the trackball sits is clean and free from debris.
Refer to the figure below for the areas on the Kensington and Logitech trackballs that
need to be cleaned.
Fig. 1: Trackball Cleaning Points
PM Check the 24 VAC voltage at connector J12. The acceptable range is ~28.
PM Check the 12 VDC voltage at test points 5 and 6. The acceptable range is 11.88
to 12.12.
PM Check the –12 VDC voltage at test points 7and 6. The acceptable range is –11.88
to –12.12.
PM Board SMN 10484630: Check the voltage at connectors J1, J3, J5A/B and pin 2
(Ground). The acceptable range is 4.85 to 5.15.
PM Board SMN 10484630: Check the voltage at connectors J1, J3, J5A/B and pin 3
(+5 VDC). The acceptable range is 4.85 to 5.15.
PM Board SMN 10389935: Check the voltage at connectors J2, J3, and J4 and pin 1
(-5 VDC). The acceptable range is 5.00 to 5.20.
PM Board SMN 10389935: Check the voltage at connectors J2, J3, and J4 and pin 2
(Ground). The acceptable range is 5.00 to 5.20.
PM Board SMN 10389935: Check the voltage at connectors J2, J3, and J4 and pin 3
(+5 VDC). The acceptable range is 5.00 to 5.20.
CAUTION
Contamination may occur when the DRD seals are being replaced.
Wear gloves to avoid contaminating the diluter assembly.
CAUTION
Improper seal installation may cause leaking and poor priming.
When you insert a seal into its cavity in the DRD acrylic block, it should make a
snapping sound to indicate that it is seated correctly.
PM IMMULITE 2000 systems, serial numbers below E0920 only: Clean and lubri‐
cate the sample and reagent pipettor lead screw assemblies with a Teflon-based
lubricant.
Lubrication is not required for IMMULITE 2000 systems, serial number E0920 and
above and IMMULITE 2000 XPi systems.
(1) Inner
(2) Middle
(3) Outer
Make sure that the gasket is seated correctly. If it is not, luminometer chain or high dark
count errors may occur.
Verify the positions using the procedures in the Alignments section ( Align‐
ments / DCIM-B02.842.01).
13. Remove and replace the solid waste exit chute provided in the kit.
a) Disconnect the waste chute sensor from J22 on the dual pipette PCB.
b) Remove all hardware from the waste chute assembly, and discard the original
waste chute.
c) Install the new waste chute.
d) Once the assembly is in place, adjust the chute so that the reaction cups fall
smoothly into the solid waste container.
If the assembly is not adjusted correctly, the used reaction cups may bounce off the
metal bracket and not fall into the solid waste container.
e) Reconnect the sensor to J22 on the dual pipette PCB and verify that the tube drop
sensor is operating correctly by running Tube Chute Test 2000 in the Diagnostics
program.
PM Clean tube drop section of tube processor and replace waste manifold insert.
If the customer uses its own printed labels, make sure they fall within the appropriate
ranges.
PM Check the scan rate for the Microscan scanner. The expected scan rate is 300 to
500 decodes per second.
PM Check the scan rate for the JADAK scanner. The expected scan rate is 20 to 200
milliseconds for the TTR.
PM IMMULITE 2000 XPi only: Check rack loader sensors and operation.
2.10.1 Performing a Volume Check for Solenoid and Linear Actuator Pumps
To verify the dispense volume of the substrate or water pump, you’ll need a beaker and
several reaction cups. They should be set up so that they are in easy reach.
n The substrate pump should dispense a consistent 200 μL (±10 μL).
n The water pump should dispense a consistent 400 μL (±20 μL).
If you stop pressing the green prime button before this step is completed, the dispensed
volumes may be inconsistent.
4. Stop pressing the green button, and return the probe to its holder.
5. Measure the volume of each reaction cup to make sure that the dispensed volumes
are consistent. If they are not, replace the pump.
PM Check volume dispense of substrate (200 μL) and water (400 μL) pumps.
2.11 Fluidics
PM Check all tubing within component deck for discoloration and possible occlusion.
Replace if necessary.
6. To finish the test, in the area where the word Load is displayed, click Blue from the
three color options, and then click No Clot. The system displays a saved blue wave‐
form, as shown in ( Fig. 8 Page 23).
Fig. 8: No Clot Waveform (Blue)
7. Compare your acquired waveform with the blue waveform, which is represents a per‐
fect “no clot.” If the waveform is comparable, the transducer is performing as expec‐
ted.
For contrast, see ( Fig. 9 Page 24), which shows a waveform that represents a clot.
Fig. 9: Clot Waveform (Green)
8. If the transducer has been well primed and the waveform is not comparable, run
Transducer Decon in the Diagnostics program, and then repeat these steps.
If the waveform is still not comparable, the sample probe or transducer may need to
be replaced.
PM Check the transducer waveform.
The WatertestPM diagnostic tests for alkaline phosphatase contamination in the water.
If the water probe touches the water inside the sample tube, wipe the water probe with
a lint-free cloth before returning it to the wash station.
13. When prompted, remove the Water Probe from the tube and return it to the Wash Sta‐
tion.
14. Select Replace Water Probe.
The instrument primes the sample and reagent probes, pipettes water, transfers the
reaction tubes to the luminometer, and dispenses substrate. The photomultiplier tube
(PMT) then reads the reaction tubes. After the instrument reads the tubes, the mes‐
sage “Program Complete” is displayed and the instrument generates a report of the re‐
sults. The figure below shows an example blank result report for a water test PM on
an IMMULITE 2000 or IMMULITE 2000 XPi system.
Tab. 1 Blank Watertest PM Result Report for the IMMULITE 2000 / 2000 XPi Systems
WATER TEST PM
mm-dd-yy hh:mm:ss
Operator: _______________________
15. Evaluate the results. If the Open CPS measurement of any reaction tube is > 1100 CPS,
the instrument may be contaminated.
In this case, consider performing a decontamination procedure, followed by a repeti‐
tion of the WatertestPM 2000 or WatertestPM 2000XPi procedure.
16. To stop running diagnostics, select Exit, then select Quit.
PM Confirm that the Watertest PM – 2000 or the Watertest PM – 2000XPi returns an
Open CPS measurement for each reaction tube of less than 1100 CPS.
PM IMMULITE 2000: Confirm both user-side and control-side CPU fans are spinning.
2.13.2.1 DupliDisk
Make sure that the three LEDs on the front of the DupliDisk are green. See the figure be‐
low, which shows the LEDs not illuminated. When the DupliDisk is saving data the status
LED is amber.
1. Restart the system and watch the messages displayed during the boot sequence.
Make sure that the following message appears: “Array is successful.”
2. Alternatively, you can enter the promise card setup by typing Ctrl+F when you are
prompted during the boot sequence.
1. At the Windows desktop, right-click Start and then click Explore. Navigate to the
C:\Program\intel folder.
2. Double-click shell.exe.
The Intel Matrix Storage Console opens.
3. Verify that both hard drives (shown as Port 0 and Port 1) are green. If not, replace the
defective drive.
Refer to ( Computer and Accessories Replacement / DCIM-B02.841.03).
This procedure does not apply to systems that use an IDE flash drive.
1. Disconnect the keyboard and monitor cable from the user-side computer and recon‐
nect them to the control-side computer.
The monitor should display DOS software.
2. Press Esc.
3. At the prompt type C:\Bin, and press Enter.
4. Type scandisk and press Enter.
5. Click Yes to run a complete scan.
The scan could take 30 minutes or more. If scandisk stops running, or if the results of
the scan indicate that there are several bad sectors, the hard drive should be replaced.
6. When the scan is complete, exit scandisk.
7. Press Ctrl+Alt+Del to restart the control-side computer.
8. Disconnect the keyboard and monitor cable from the control-side computer and re‐
connect them to the user-side computer.
9. Close the Run IMMULITE window, and then double-click the Run IMMULITE icon on
the desktop to reestablish communication between the control-side and user-side
computers.
PM Perform scan disk on control-side hard drive.
6. In the Initials field, type your initials, and in the Password field type PMTracker.
7. Click OK.
The PM Message dialog box opens.
8. Click OK.
The current date is displayed in the Date of Last PM field and the current test count
resets to 0.
9. Click Exit to close the PM Tracker and return to the desktop.
PM Run one set of customer controls on one assay to verify performance.
3.1 Version 08
Section Change Description Change Num‐
ber
( Sample Car‐ Added steps to replace the foam CHG0035595
ousel / Page 17) roller. and
LFT0014886
( Replacing the Added check point to “check LFT0015216
Vacuum Pump wash sump drain for leaks”.
Trap / Page 20)
( Final Revised step 2 and associated LFT0013579
Check / Page 31) check point to “Run one set of
customer controls on one assay
to verify performance”.
siemens-healthineers.com