Professional Documents
Culture Documents
L4 - Heijmen - Athens 2016 - Reference Dosimetry
L4 - Heijmen - Athens 2016 - Reference Dosimetry
L4 - Heijmen - Athens 2016 - Reference Dosimetry
Ben Heijmen
MU-chamber
is a transmission ionization chamber
electronics generates ‘counts’ (MUs)
after a pre-set charge QMU
measured
Calibration of MU-chamber:
electronics is tuned such that
1MU/QMU corresponds with
a dose of 1 cGy in H20
in reference conditions
dose measured
with calibrated
ionization chamber
What accuracy in delivered dose is required?
Delivered dose is main treatment related
predictor for radiotherapy outcome:
local control and toxicity
Prostate cancer
D
TCP
D
1% change in dose results in
% change in TCP
=3
5% change in dose results in
15 % change in TCP
(50%35%)
D
1% ?
(Levegrun, IJROBP 2002) D
What accuracy in delivered dose is required?
A. Brahme
Acta Radiologica Oncology
1984
What is needed to perform reference dosimetry in a hospital?
Needed by hospital:
1. calibrated ionization chamber (i.c.):
ND,w,Co (dose in H2O in 60Co / charge)
2. protocol to use calibrated i.c.
in a therapeutic beam
Outline
1. Global workflow in a hospital
2. Introduction to primary dosimetry standards and standard
laboratories (needed for calibration of i.c.)
3. Detailed example of reference dosimetry in the hospital
4. Details and physics of primary standards of absorbed dose
to water
Global workflow for reference dosimetry in a hospital
1. Hospital buys i.c. recommended by TRS 398
2. Hospital’s i.c. + electrometer to a standard laboratory
for calibration ND,w,Co : dose in water in 60Co /reading
3. Absolute water dose measurements in the hospital
for MU-chamber calibration, using:
- the calibrated i.c.
- TRS398 protocol
Reference Dosimetry
for high energy, external beam photon beam therapy
Outline
1. Global workflow the hospital
2. Introduction to primary dosimetry standards and standard
laboratories (needed for calibration of i.c.)
3. Detailed example of reference dosimetry in the hospital
4. Details and physics of primary standards of absorbed dose
to water
What are dosimetry standards and standard laboratories:
your
calibration = dose/reading in 60Co beam
factor
International Measurement System (IMS) for radiation metrology
Outline
1. Global workflow the hospital
2. Introduction to primary dosimetry standards and standard
laboratories (needed for calibration of i.c.)
3. Detailed example of reference dosimetry in the hospital
4. Details and physics of primary standards of absorbed dose
to water
How to measure absorbed dose in water:
TPR20,10(Q)=D(d=20)/D(d=10)
To determine kQ,Co: use lookup Table in TRS938
Dw,Q = MQ∙ND,w,Co∙kQ,Co∙ki
Outline
1. Global workflow the hospital
2. Introduction to primary dosimetry standards and standard
laboratories (needed for calibration of i.c.)
3. Detailed example of reference dosimetry in the hospital
4. Details and physics of primary standards of absorbed dose
to water
Only 3 methods for are suited for absolute dose measurement
in a primary standard in a standard laboratory:
- calorimetry
- ionization dosimetry
- chemical dosimetry (Fricke)
Calorimetric standard
30 cm
Fe2+ in H2O I0
before irradiation: uv
after irradiation: uv
I
Fe2+ and Fe3+ in H2O
log(I0/I)
M= = 2187 liter mole-1 cm-1
·L
Intermezzo to introduce ionometric standards (I)
w P
Dose in a point P in water:
water
Dw= w(S/)c,w
(P)
Intermezzo to introduce ionometric standards (III)
dEc
(P) O d P
D(P) dx
V V
d (P)
O
dEc
(P)
D(P) dx (P) (S / )
c
Ionometric standard
w P
Dose in a point P in water:
water
Dw= w(S/)c,w
Dw= Dairsw,air
Ionometric standard
standard dosimetry: dose measurement with i.c. in 60Co:
Dw,Co= Dair,Co(sw,air)Co
Dair,Co= Eair,Co/m, Eair,Co is energy deposited in air by electrons
Dair,Co= (QWair,Co)/m
- Q = charge formed in air cavity (measured with electrometer)
- Wair,Co = average energy expended in air per coulomb
- m = mass of air in cavity air
P
Dw,Co= ((QWair,Co)/m)(sw,air)Co i.c
water
TRS398:
- Wair,Co = 33.97 J/C Refinements needed e.g. to correct
- (sw,air)Co = 1.134 for deviations Bragg-Gray (in TRS398)