Professional Documents
Culture Documents
Dosicard-R User's Manual (1) - Fixed
Dosicard-R User's Manual (1) - Fixed
User’s Manual
9236766A
Copyright 2003, Canberra Eurisys SA. All rights reserved.
The material in this document, including all information, pictures, graphics
and text, is the property of Canberra Eurisys SA and is protected by French
copyright laws and international copyright conventions.
Canberra Eurisys expressly grants the purchaser of this product the right to
copy any material in this document for the purchaser’s own use, including as
part of a submission to regulatory or legal authorities pursuant to the
purchaser’s legitimate business needs.
No material in this document may be copied by any third party, or used for
any commercial purpose or for any use other than that granted to the
purchaser without the written permission of Canberra Eurisys.
Canberra Eurisys SA
ZA de l'Observatoire - 4 avenue des Frênes
78067 St Quentin Yvelines Cedex - France
Tel: 33 (0)1 39 48 57 70 FAX: 33 (0)1 39 48 57 80
http://www.canberra.com
The information in this document describes the product as accurately as
possible, but is subject to change without notice.
Choosing a Battery
1. INTRODUCTION......................................................................................................................................1
2. GETTING STARTED ...............................................................................................................................4
3. CONTROLS .............................................................................................................................................6
3.1. FUNCTIONAL DESCRIPTION ................................................................................................................6
3.2. MAIN SCREENS .................................................................................................................................8
4. SETUP ...................................................................................................................................................18
4.1. SET UP DIRECTLY ON INSTRUMENT ..................................................................................................18
4.2. SETUP VIA DATA BASE SOFTWARE ...................................................................................................24
5. OPERATION ..........................................................................................................................................25
5.1. OPERATING MODES.........................................................................................................................26
5.2. “PERMANANENT” MODE DETAILED INSTRUCTIONS ............................................................................27
5.2.1. Manual dose function......................................................................................................28
5.2.2. Dose rate function ...........................................................................................................30
5.2.3. Dose in current day function..........................................................................................31
5.2.4. Test function ....................................................................................................................32
5.2.5. Password protected functions .......................................................................................35
5.3. ALARMS AND INDICATORS FUNCTIONS................................................................................40
5.3.1. Manual dose alarm ..........................................................................................................41
5.3.2. Dose rate alarm................................................................................................................42
5.3.3. Daily dose alarm ..............................................................................................................43
5.3.4. Time alarm........................................................................................................................44
5.3.5. Monthly dose alarm .........................................................................................................45
5.3.6. Three-month dose alarm.................................................................................................46
5.3.7. Indicators..........................................................................................................................47
6. MAINTENANCE.....................................................................................................................................48
6.1. NORMAL MAINTENANCE ..................................................................................................................48
6.2. TROUBLE SHOOTING ......................................................................................................................48
7. APPENDICES ........................................................................................................................................49
7.1. TECHNICAL CHARACTERISTICS ........................................................................................................49
7.1.1. Nuclear characteristics ...................................................................................................49
7.1.2. Electrical characteristics ................................................................................................51
7.1.3. Mechanical & environmental characteristics ...............................................................51
7.1.4. Electromagnetic compatibility .......................................................................................51
7.2. WHEN AND HOW TO CHANGE DOSICARD BATTERY ............................................................................53
7.2.1. When to change the Dosicard battery?.........................................................................53
7.2.2. How to change the Dosicard battery? ...........................................................................55
7.3. TEST PROCEDURE ..........................................................................................................................58
7.3.1. Dose integration test .......................................................................................................58
7.3.2. Calibration and checks ...................................................................................................58
7.4. GLOSSARY.................................................................................................................................59
DosicardTM/R
1. INTRODUCTION
Thank you for purchasing the DosicardTM Personal Dosimeter, developed to help you make a
permanent record of your accumulated dose while continously monitoring your current dose rate.
The Dosicard badge is designed to be used in its vinyl protective pocket which protects it from dirt,
humidity, shocks, contamination, etc…, even during bi-directional infra-red dialog with a multiplexer or
when using the three touch buttons.
With its credit-card size, Dosicard features the smallest stand-alone electronic dosimeter available on
the market. The quantity measured is the personal dose equivalent Hp(10) in γ-rays, which is referred to
throughout this text by its abbreviated form of "dose" and is often expressed in mrem (millirem).
Dosicard provides real time monitoring of the personal dose and dose rate. The knowledge of individual
dosimetry in real time is very important, as it allows the doses received during operations to be reduced,
in accordance with the ALARA (As Low As Reasonably Achievable) strategy. That’s why the date and
time must be set.
The Dosicard Personal Dosimeter carries the " CE " label for electromagnetic compatibility
Dosicard features a silicon detector, complete analog and digital circuitry, including a microcontroller
with large non-volatile memory, an LCD display, and audio and visual alarms. Three touch buttons allow
programming and display setup of the manual dose, dose rate and cumulative doses per
day/month/quarter/year/five years. The recording of the doses received according to a programmable
schedule, the management of the alarms, the display and the touch buttons are powered by a standard
button battery. Non-volatile EEPROM memory safely stores the detailed history of the daily doses,
times when alarm levels have been exceeded, and relevant data concerning the user and the badge
itself.
Thanks to the power of its microprocessor and the capacity of its internal memory, the
dosimeterDosicard contains a very large range of functions which are fully exploited in the
DOSEMANAGER and CARD systems (see the relevant manuals). Depending on their configuration,
these computerised systems can provide the total management of operational dose measurement of for
personnel, from a small medical laboratory of ten people, up to a nuclear center employing a thousand.
1
Different dosimetry configurations:
DOSIMAN is a Dosicard with a special ON/OFF switch for users who do not want to monitor dosimetry
via a database. Once switched off, DOSIMAN clears all data and is ready for next use.
Stand-alone DOSIMAN
BADGE
Mobile PC READER
2
The CARD System enables the complete dose history of a group of people and monitors their presence
in a controlled zone. The zone entrance and exit readers are connected by an Ethernet TCP/IP network
to a central computer which manages the installation’s data base and its event history.
The CARD system is well adapted to large nuclear installations employing several thousands of people.
CONTROLLED AREA
RS 422 BAC
Ethernet
MULTIPLEXER
LCB
BADGE RS 232
READER
Printer
3
2. GETTING STARTED
• Ensure that the badge is equipped with a battery.
• The badge must be placed in its protective pocket which must be closed.
Fixing clip
Protective pocket
• Wake-up Dosicard: To activate the liquid crystal display, press anytouch button.
If the Dosicard displays nothing, consult § 6.2.
• Ensure that the LCD is activated. Ensure that the following indicators are not displayed.
• If the low battery pictogram appears and particularly before opening the back cover, consult
the appendix, paragraph 7.2. Never open the back cover without a reason. The date and time
setting may be lost.
4
• Ensure that your Dosicard has been identified in data base. Check with your Health Physics
manager officer.
Dosicard must be assigned to one worker: one badge number is associated with only one worker
number. If not, Dosicard offers only a small number of functionalities (see § 4).
• Go on test screen. For that, press the button twice when the Dosicard is awake.
MTH YR
Ensure 3, MTH and YR are not blinking. If so, the date and time have not been set. A badge
reader and the attribution software are needed to copy the system time into the Dosicard (see the
user manual of the PC based software such as DOSEMANAGER).
5
3. CONTROLS
Dosicard offers two levels of control:
− direct access on Dosicard front panel,
− software driven setup.
This level of control needs a PC, software and a badge reader (LCB).
Liquid Crystal
Buzzer air vent
Display (LCD)
audible alarm
Three touch
Position of the buttons
silicon diode
Code and
serial No
Rear cover
locking screw
(battery access)
Concise
instructions
for use
6
Touch buttons:
Screen selection: button used to switch from one screen to the next.
Scroll: button used to increase the digits or to pass to the next option inside each screen.
7
3.2. MAIN SCREENS
SLEEP MODE:
c Manual Dose
Hp
* ))
mrem
)
d Doserate
)) )
Hp mrem/h
ª Alarm setup /
Dose History
SET ) ))
Sleep
e Dose in Current Day
) ))
Hp . mrem
• The Sleep Mode switches off the display but NOT the detector.
• All the dose acquisition in process remains running when to in the Sleep Mode.
• This display mode saves energy, and increases the battery’s life.
• This is the default display mode: The display automatically switches to this mode 4 minutes after
the last button press.
• Dosicard quits this display mode as soon as ANY button has been pressed.
• When it leaves the Sleep Mode, Dosicard goes toManual Mode (see further).
• The Sleep Mode can be disabled by using a badge reader (LCB) and the software
DOSEMANAGER or CARD. The display is then always switched on.
8
)
Hp * ))
mrem
1) MANUAL DOSE MODE:
c Manual Dose
Hp
*mrem
)) )
d Doserate
)))
Hp mrem/h
ª Alarm setup /
Dose History
SET ) ))
Sleep
e Dose in Current Day
) ))
Hp . mrem
f Test / Alarm display
)
))
• The Manual Dose Mode is an active mode, identified by the presence of the star in the top right
part of the display.
• Acquisition starts when the subject presses the star button . The displayed star starts blinking
and the previous value is reset.
• The acquisition stops when the subject presses the star button again . The displayed star
stops blinking.
• Press the straight arrow button (during or at the end of the acquisition) to see the elapsed
time.
• If the alarm threshold is exceeded during acquisition, the corresponding alarms are triggered (see
alarm management § 5.3).
9
)) )
Hp mrem/h
2) DOSE RATE MODE:
c Manual Dose
Hp
*
) ))
mrem
dDoserate
)))
Hp mrem/h
ª Alarm setup /
Dose History
SET
) ))
Sleep
e Dose in Current Day
) ))
Hp . mrem
f Test / Alarm display
)
))
• Dose Rate acquisition starts automatically when the detector is switched on (by the badge reader
(LCB or BIO readers)).
The top left corner state triangle starts blinking. In “permanent” mode, Dosicard always acquires
dose.
• The dose rate acquisition stops automatically when the detector is switched off (by the LCB or
BIO readers),in stand-by mode, or out of the time range in daily timer mode. When dose rate
acquisition turns off, the triangle in the upper left corner of the display stops blinking.
• If the Dose Rate alarm threshold is exceeded when the Dose Rate is active, the corresponding
alarms are triggered (see alarm management).
• Some Dosicard dosimeter versions can be set in the factory without this doserate display mode
(Dosicard /H, Dosicard /Z).
10
)) )
Hp . mrem
3) DOSE IN CURRENT DAY MODE:
c Manual Dose
Hp mrem
*
) ))
d Doserate
)) )
Hp mrem/h
ª Alarm setup /
Dose History
SET
) ))
Sleep
e Dose in Current Day
)) )
Hp . mrem
• It’s the accumulated dose since entering a controlled area in zone mode (detector switched on
by a badge entrance reader).
• The daily dose is saved in memory each day at midnight, becoming the most recent of the 90
daily doses.
• One Dosicard holds 90 daily doses permanently, allowing dose summation: « month dose »=sum
of the last 30 daily doses; « 3 months dose » = sum of the last 90 day doses.
• In order to keep a daily history of the Dosicard wearer, it is required to update the database at
least once every 3 months using the LCB reader and the software’s “Database update” feature.
11
4) TEST / ALARM DISPLAY MODE:
c Manual Dose
Hp
* ) ))
mrem
d Doserate
)) )
Hp mrem/h
ª Alarm setup /
Dose History
) ))
SE
T
Sleep
e Dose in Current Day
)
))
Hp . mrem
12
Alarm levels setup displays:
Hp mrem Hp mrem/h
SET
SET
mrem
Hp mrem Hp mrem
5 YR YR
Hp mrem Hp mrem
Pressing the button returns the Dosicard to the normal Dose main screen.
15
Alarm management: type of alarms:
Visual alarm
Alarm pictogram
flashes when the
))) alarm is activated
Hp . mrem
16
Alarm management: alarm level setup (consult § 4.1)
• Available from the display « Waiting for Code » after validation of the 4 hyphens « - - - - », using
the star button .
• Setup available on the Manual dose, the Dose rate, the Time and the Dose Increment (Step).
• Alarm levels memorized: not reset when detector stopped nor when batteries out of service.
• Presence of one Alarm Pictogram as soon as at least one alarm level is set up.
• Reading of the alarm levels available from the Test display mode, after starting the tests (see
Test display mode).
17
4. SETUP
Dosicard stores doses permanently and continuously unless the battery is missing or out of order.
Dosicard parameters can be set
− directly on instrument or
− via a PC based software (an LCB reader, a PC and software are needed).
In order to be operational, the Dosicard has to be assigned to one subject. This can only be done in
PC based software (i.e. DOSEMANAGER; Please consult the relevant user manual).
The Dosicard’s parameters can be set only if the Dosicard has been assigned to one subject via
NOTE
the PC-based software (i.e. DOSEMANAGER or CARD).
The user does not have access to all the alarm levels.
− Incremental beep (1)
− Timer alarm (2)
− Dose rate alarm (3)
− Dose in current day alarm (4)
(1)
and (2) are always available on the Dosicard.
(3)
and (4) are only available if enabled in PC software (see the relevant user manual).
18
HOW DO I SET AN ALARM LEVEL ON Dosicard ?
From the Alarm Setup display screen (see 5) of § 3.2) of Dosicard:
Manual Dose
c * )) )
Hp mrem
Doserate
d )
))
Hp mrem
Test/Alarm display
Alarm level setting f
Manual Dose
Hp
* mrem
)) )
mrem
The following alarm levels are manually accessible by the subject, using the three touch buttons:
− the manual dose level,
− the dose rate level,
− the time level,
− incremental dose level.
The default setting of the badge allows manual setting of the levels.
Nevertheless, it is possible at any time to prevent the subject from changing the levels manually. To do
this, all that is needed is to calibrate the Dosicard differently using the attribution software and a badge
reader.
Access to the alarm level changes can be authorised or prevented depending on the configuration of
the Dosicard.
A null value for a level corresponds to the deactivation of that level.
19
− Access to the manual modification of the alarm levels is available from the Alarm Setup display
screen (see 5) of § 3.2), corresponding to the entry into the « coded access functions » mode.
In this mode the display shows screen as follows :
Figure 3
− From this screen, you must press the “ VAL" button four times to see the screen for setting the
level for the current day’s dose.
Each successive press of the " VAL" button makes the next digit flash, as is shown in the
figures below :
• First touch on the " VAL" button:
SET
The triangle and the
second hyphen will flash.
SET
The triangle and the
third hyphen will flash.
• From this situation, the fourth and last touch on the " VAL" button will then allow access to
the adjustment screens for the following levels :
Dose in Manual alarm level screen:
Hp mrem
Figure 4
Successive touches on the " SCROLL" button will scroll through the four adjustment screen
levels.
20
Dose rate alarm level screen:
Hp mrem/h
Figure 5
SET
The word "SET"
will flash.
Figure 6
SET
The word "SET"
will flash.
mrem
Figure 7
Successive touches on the “ SCROLL" button will scroll through Figure 4 through Figure 7
again.
− To select a level, press the " VAL" button in the desired screen. The word "SET" will then
become continuous and the first hyphen will flash.
− The " VAL" button therefore enables the selection of the level that the wearer wishes to adjust.
Hp
*
mrem
Figure 8
• to leave the adjustment procedure for the level in progress if there has been a touch on the "
VAL" button, and pass on to the adjustment of the next level.
21
1) How to adjust the alarm level for the manual dose
After having selected this level by the " VAL" button from Figure 4.
• Each press of the “ SCROLL" button increases the first digit from 0 to 9.
• The " VAL" button validates the value of the digit displayed and selects the next digit.
• Validation of the fourth digit allows the unit, ‘m’ for milli or ‘µ’ for micro, to be chosen with the "
SCROLL" button.
• Once the unit is correctly displayed, a touch on the " VAL" button accepts the level and unit
displayed and and stores them in the badge’s memory. This final acceptance is accompanied by
a special sound signal indicating that the value has been stored.
• the " VAL" button will enable the same level to be adjusted again.
The dose in current day level may be adjusted between 0 mrem and 999.9 rem.
After having selected this levelwith the " VAL" button(Figure 5).
The adjustment procedure is identical to that for the manual dose level with a range from 0 mrem/h
to 999.9 rem/h.
22
3) How to adjust the alarm level for the time
The alarm level for the time cannot be set from software; it is a local setting only.
After having selected this level by the " VAL" button from Figure 6.
The value of the time level is implicitly expressed in minutes. The range is from 0 to 99 minutes.
The adjustment procedure is identical to that for the other levels, with the exception that there are
only two digits to set and there is no adjustment of the unit.
Incremental dose is the amount of dose acquired by Dosicard before it beeps once.
After having selected this level by the " VAL" button from Figure 7.
The value of the incremental dose can be adjusted between 0 and 99.9 mrem.
The adjustment procedure is identical to that for the other levels, with the exception that there is no
adjustment of the unit (mrem is compulsory).
23
4.2. SETUP VIA DATA BASE SOFTWARE
The data base software allows to set the five following alarm levels:
− manual dose,
− dose rate,
− dose in current day,
− month dose,
− 3 month dose.
Consult the relevant user manual.
24
5. OPERATION
Do not take the dosimeter out of its pocket. Do not put other objects in the pocket.
Do not put the Dosicard in direct sunlight, for example on the back shelf of a car.
It is inadvisable to bring a Dosicard near a functioning portable telephone or a walkie-
talkie, or, generally near any electromagnetic source emitting a magnetic field higherthan
the range detailed paragraph 7.1.4. For the same reason, do not put the dosimeter on a
television set, a microwave oven or a high frequency welding kit
25
5.1. OPERATING MODES
The various modes of configuration and hence of use of the Dosicard enable all the dose measurement
needs of a user to be covered, whether in his normal life or in specific work conditions.
The configuration modes of the Dosicard are as follows:
− "PERMANENT" mode (operating default mode):
In this mode, Dosicard collects the dose data 24 hours a day.
. rem
The top left corner triangle is always blinking.
This mode is suitable for personnel who move from site to site yet report to a center which
possesses an LCB badge reader and the DOSEMANAGER attribution software. All of the Dosicard’s
functions are available in this mode. By means of the computer system to which it is attached, the
wearer of such a badge can keep track of his dosimetry and carry out the relevant processing, both
on the badge and on the information that has been gathered.
− "STAND-BY" mode:
In this mode, the Dosicard will only integrate doses within a controlled area.
Dosicard collects dose data from the time the wearer checks in at an entrance entrance reader to
the time when he checks out at an exit reader.
The top left corner triangle is not blinking: Dosicard does not collect any dose.
The dose recorded is the dose taken inside a controlled area.
Top left corner triangle is blinking inside the controlled area and fixed outside. The detector part
is powered only inside the controlled area.
This mode can be used in order to save the battery power when Dosicard is on the shelf.
The use of a Dosicard in "Stand-by" mode requires a hardware and software environment which are
both adapted to the situation, and can enable management of dose measurement and site access
for several hundred people. This mode is suitable for the dosimetry control of people who work
inside buildings whose areas are defined by entrance and exit readers.
When the badge’s time setting is not correct, for example after a late battery change without a time
NOTE reset by an LCB badge reader, the manual dose and dose rate functions remain valid, but the dose
history will cease to have any meaning.
Dosicard badges are configured in "Permanent" mode at the factory with a user code of 1000 and a
time delay of 4 minutes for the screen.
In "Permanent" mode the badge integrates doses 24 hours a day, even when the screen is not lit up.
The three buttons, MODE, VALidation and SCROLL provide access to the main functions.
An audible beep is emitted each time an active button is pressed.
26
5.2. “PERMANANENT” MODE DETAILED INSTRUCTIONS
This mode for the Dosicard is the default mode.
In this mode, the configuration of the Dosicard is factory set as follows:
− Manual modification of the levels is authorised.
− The manual dose function is activated.
− In case of alarm, the red LED and the buzzer are activated.
− The duration of the audio (buzzer) and visual (red LED) signals corresponding to the setting off of an
alarm is limited to 32 seconds.
− The display switches off automatically after 4 minutes, without depressing any button.
− The unit by default is Rem.
− The user code to access to the dose history is 1000.
The options described above may be modified at any time by the attribution software via a badge
reader.
NOTE: If the display is off, it can be switched on again by simply pressing one of the three buttons.
Access to the various functions available in "Permanent" mode is obtained by pressing the "
MODE" button successively. The functions will be presented in the following order :
− display displayof the manual dose,
− display displayof the dose rate,
− display displayof the daily dose,
− the test function,
− the coded access file function => display displayof the dose history,
− the alarm function => programming the alarm levels.
To change from the "Alarm setup / Dose history" function to the “manual dose” function , press the
button a second time.
c Manual Dose
Hp
*
mrem
)) )
d Doserate
)))
Hp mrem/h
ª Alarm setup /
Dose History
SET ) ))
Sleep
e Dose in Current Day
) ))
Hp . mrem
27
5.2.1. Manual dose function
The Manual dose function enables the wearer of the badge to view a dose that has been integrated
between two instants. Both instants are determined by the wearer by pressing the button.
Figure 9
The mrem or rem units are automatically changed when the measurement range changes:
− 0 mrem to 999.9 mrem,
− 1 rem to 100 rem.
The wearer activates the "Manual dose" function using the " VAL" button:
− A first touch on " VAL" determines the "START" time and begins to integrate the manual dose..
− A second touch on “ VAL" determines the "STOP" time and stops integrating the manual dose.
− A third touch on " VAL" determines a new "START" time, and the star will start to flash again.
By extension, successive pressing of the " VAL" button enables the manual dose function to be
started or stopped.
Since the star flashes only when the manual dose function is active, a quick glance at the display
shows the function’s status.
28
To see the elapsed time since the "START", press the " SCROLL" button.
The display will show Figure 10, as follows :
Figure 10
Minutes and seconds from 00:00 (00 min. 00 sec.) to 59:59 (59 min. 59 sec.), then hours and minutes
from 01:00 (01 hr. 00 min.) to 99:59 (99 hr. 59 min.)
This screen allows the elapsed time since the activation by a "START" of the manual dose function to
be viewed (in the present case, either 1 minute and 22 seconds or 1 hour and 22 minutes).The elapsed
time increments in 2 second steps when the displayed time is in minutes and seconds.
Afterwards, if the increment in the elapsed time (the scrolling of the last digit) no longer occurs every
two seconds, this means that the time that has passed since the "START" is more than 1 hour. If this is
the case, the time displayed is expressed in hours and minutes.
When the elapsed time exceeds 99 hr. 59 min., the display begins to flash and the time no longer
increments.
Each press of the " SCROLL" button toggles the display between manual dose and elapsed time.
The manual dose function may be started or stopped by pressing the " VAL" button when the
display is in the elapsed time mode (Figure 10).
29
5.2.2. Dose rate function
Access to the dose rate function is obtained by pressing the " MODE" button from either Figure 9 or
Figure 10 in the Dose in Current Day function.
The dose rate function enables the value of the instantaneous dose rate measured in real time to be
viewed.
In this mode the display shows Figure 11, as follows :
Figure 11
•
The flashing dot indicates that the dose rate measurement function is activated. The H p logo stands
for the equivalent dose rate per person.
The mrem/h or rem/h units are automatically changed when the measurement range changes :
− 0 mrem/h to 999.9 mrem/h
− 1 rem/h to 99.9 rem/h
− > 100 rem/h flashing
The operator’s actions in this mode are limited, as he can only read the value of the instantaneous
dose rate measured by the Dosicard.
Pressing the " SCROLL" or " VAL" buttons will have no effect.
30
5.2.3. Dose in current day function
Access to the function for viewing the daily dose is obtained by pressing the " MODE" button again
when Figure 11 (display of the dose rate) is being displayed.
In this mode, the display will show Figure 12, as follows :
Figure 12
Unlike the screen for the manual dose, the star does not appear in this display mode.
The wearer can see the value of the integrated dose for the day.
The value of the dose is reset to zero at 00 hr 00 every day.
NOTE : The daily dose will not be displayed if the Dosicard’s date and time have not been set.
31
5.2.4. Test function
The test function lets you quickly check the Dosicard’s operation, particularly the display and
alarms, and verify the current settings for the alarm levels.
Access to the test function of the Dosicard is obtained by pressing the " MODE" button from Figure
12 (display of the daily dose).
In this mode, the display will show Figure 13. This is the default screen that should be seen as soon as
test mode is entered :
Figure 13
If the display shows Figure 14instead, the Dosicard is no longer set to the right date and time :
Figure 14
The test function of the Dosicard is launched by pressing the “ VAL" button when Figure 13 is being
displayed. The test function allows the operation of the buzzer, the LED and all the elements of the
display (segments, letters, words, elements and pictograms) to be checked.
Pressing on the " VAL" button activates the buzzer and the red LED for a short time,
thensequentially displays the following screens :
− all of the elements of the display: Figure 15,
− the level value in force for the manual dose: Figure 16,
− the level value in force for the dose rate: Figure 17,
− the level value in force for the elapsed time (the time passed since the "START" signal): Figure 18,
− the incremental dose value in force: Figure 19.
The display of each of the screens (Figure 15 to Figure 19) lasts for a few seconds in order to allow the
presence of all the elements of the display to be verified and to give time to read the various level
values.
32
)
* ))
The triangle indicating the SET 3 MTH 5 YR
Dosicard’s state, the battery
indicator and the dot " z " above
Hp : . mrem/h
"Hp" will be flashing.
Hp mrem
Figure 16: Level value for the manual dose (1 mrem in this example)
Hp mrem/h
Figure 17: Level value for the dose rate (3 mrem/h in this example)
33
When the sequence of screens Figure 15 to Figure 19 ends, the Dosicard displays the initial screen
Figure 13, corresponding to the entry into test mode, again, as follows :
A new test sequence can then be launched by pressing on the " VAL" button.
During the test sequence, the following points must be verified :
− The buzzer should sound.
− The LED should give a short flash.
− All of the elements of the display should be present.
If any of these points cannot be verified, the Dosicard has an operational fault.
34
5.2.5. Password protected functions
Access to this function is obtained by pressing again on the “ MODE" button from Figure 13 (test
function).
In this mode the display will show Figure 20, as follows :
From this screen, access may be had to two independent functions, as follows:
− display of the dose history
− manual setting of the alarm levels
NOTE : The dose history will not be displayed if the Dosicard’s date and time have not been set.
To access the dose history from screen in Figure 20, the user code must be entered. This code can be
changed at any time by the attribution software via a badge reader.
35
How to enter the user code
− When the "SET" mode is entered (screen Figure 20), the first hyphen will flash. Press the "
SCROLL" button to change the " − " to " 0 " :
SET
Figure 21
Press the " SCROLL" button again to change the first digit from the 0 to the 1 :
SET
Figure 22
Successive touches on the “ SCROLL" button increments the displayed value by one, up to the
value 9.
SET
Figure 23
A further touch on the " SCROLL" button when 9 is displayed (screen Figure 23) will return to
the value 0 (screen No. 13).
− When the desired value is displayed, press the " VAL" button to accept it. and select the next
hyphen.
− As the default user code programmed at the factory is 1000, you should press the " VAL" button
when the display corresponds to Figure 22.
The display will then show Figure 24 :
Figure 24
From this screen, the value of the second digit of the user code, whose default value is " 0 ", must
be set by successive touches on the " SCROLL" button.
Figure 25
36
Once the value of the second digit is set to the desired value, accept it by pressing the " VAL"
button.
The display will then show Figure 26 :
Figure 26
In the same way, the value of the remaining digits are selected and the correct value is validated.
When all four digits have been entered, the screen will appear as follows :
Figure 27
When the displayed four digit number corresponds to the user code programmed into the Dosicard,
accepting the fourth digit will give access to the dose history.
If the four digit number does not correspond to the programmed user code, the dose history display will
be disabled. The Dosicard will then leave the current user code entry mode and return to the manual
dose function, showing Figure 9 once again :
Hp
*
mrem
If another try is required to access the dose history, it will be necessary to press the " MODE"
button four times in order to enter the user code.
At any time, the sequence for programming the user code may be exited by pressing the " MODE"
button. In this case, the Dosicard will return to the manual dose function and the user code procedure
NOTE
will be cancelled. During the user code programming function, only the " SCROLL" and " VAL"
buttons are used. If the time taken to validate a digit is greater than about four minutes, the Dosicard
will return automatically to the manual dose function and the user code procedure will be cancelled.
From Figure 27, when the four digit number corresponds to the user code programmed into the
Dosicard, a touch on the " VAL" button allows access to the dose history.
37
Dose history
This file allows the wearer to view the following information sequentially :
• the monthly dose, corresponding to the sliding total of the last 30 complete days.
For example: On July 15th 1996, the wearer views the monthly dose. The value displayed will
correspond to the dose integrated by the badge from 00:00 on Saturday June 15th 1996 to
23:59:59 on Sunday July 14th. The monthly dose is updated every day at 00:00 (midnight).
30 complete days
Day (D-1)
July 14th 1996
MTH
Hp mrem
Figure 28
From Figure 28, press the " SCROLL" button to view the "3 mth" dose.
• the 3 months dose corresponding to the sliding total of the dose for the last 90 complete days.
For example: On July 15th 1996 the wearer views the « 3 monthly » dose. The value displayed
will correspond to the dose integrated by the badge from 00:00 on Tuesday April 16th 1996 to
23:59:59 on Sunday July 14th. The « 3 monthly » dose is updated every day at 00:00 (midnight).
Day D
July 15th 1996
90 complete days
3 MTH
Hp mrem
Figure 29
From Figure 29, press the " SCROLL" button to view the "YR" dose.
38
• the "YR" dose corresponding to the sliding total of the dose for the last 12 complete calendar months.
For example: On July 15th 1996 the wearer views the "YR" dose. The value displayed will
correspond to the dose integrated by the badge from 00:00 on July 1st 1995 to 23:59:59 on
Sunday June 30th 1996. The "YR" dose is updated on the first day of every calendar month at
00:00 (midnight).
Day D
July 15th 1996
12 complete months
YR
Hp mrem
Figure 30
From Figure 30, press the " SCROLL" button to view the "5 YR" dose.
• the "5 YR" dose corresponding to the sliding total of the dose for the last 60 complete calendar months. The "5
YR" dose is updated on the first day of every calendar month at 00:00 (midnight).
5 YR
Hp mrem
Figure 31
Successive touches on the " SCROLL" button from Figure 31 will scroll through the
dose history again. If no action is carried out for a few minutes, the Dosicard will leave
the dose history and return automatically to the manual dose. Pressing the " MODE"
NOTE button will quit the dose history at any time and return to the manual dose function :
Hp
*
mrem
39
5.3. ALARMS AND INDICATORS FUNCTIONS
In general, a triggered alarm is manifested by three elements: the buzzer, the LED and the liquid crystal
display :
− The buzzer generates audio signals, which vary according to the type of alarm.
− The red LED gives a short flash once every second, whatever the type of alarm.
1 second
< >
− The liquid crystal display indicates the type of alarm which has been or is being set off.
Other alarm configurations can be defined with the attribution software and a badge reader.
The values of the levels are programmed into the badge by the attribution software using a badge
reader :
− if the value of the level is 0, the alarm is not activated,
− an alarm is activated by programming a level other than 0.
40
5.3.1. Manual dose alarm
The manual dose alarm is triggered when the integrated manual dose is greater than the manual dose’s
alarm level.The manual dose alarm can be triggered onlyalarm if the badge’s manual dose function is
activated.
When a manual dose is triggered, the display will show Figure 32, :
The star is
*
continuous.
The Hp symbol and the
triangle indicating the
))) The alarm pictogram
badge’s state will flash.
Hp : .))) will flash.
. mrem
Figure 32
During the complete duration of the alarm, the buzzer will emit the following audio signal:
1 second
< >
This audio signal may be unlimited or limited to 32 seconds, depending on the configuration that has
been programmed.
The audio signal is accompanied by the flashing of the red alarm LED.
When the audio signal stops because it has been limited in time, the visual display (Figure 32) will
remain until it is cancelled by the wearer.
A dose rate alarm can be cancelled by pressing the " VAL" key.
This cancellation will stop the buzzer if it is still audible, and allows display of the other badge functions.
When the audio signal stops because it has been limited in time, Figure 33 will remain on the display
until it is cancelled by the wearer pressing the " VAL" key.
41
5.3.2. Dose rate alarm
The dose rate alarm is triggered when the measured dose rate is greater than the dose rate’s alarm
level.
The dose rate alarm can be triggered only if the badge is activated (the indicator triangle is flashing).
The delay in setting off an alarm on the dose rate may be relatively long (more than a minute for low
rates) so as to integrate a sufficient number of measurements, creating a good statistical accuracy. This
delay assures that a real alarm has been triggered, avoiding false alarms.
When an alarm on the dose rate is triggered, the display will show Figure 33,:
Figure 33
The buzzer will emit the following audio signal while the alarm is active:
1 second
< >
This audio signal may be unlimited or limited to 32 seconds, depending on the configuration that has
been programmed.
The audio signal is accompanied by the flashing of the red alarm LED.
When the audio signal stops because it has been limited in time, Figure 33 will remain on the display
until it is cancelled by the wearer.
An alarm on the dose rate is cancelled by pressing the " VAL" key.
This cancellation will stop the buzzer if it is still audible, and allows display of the other badge functions.
42
5.3.3. Daily dose alarm
The daily dose alarm is triggered when the integrated dose is greater than the daily dose’s alarm
level.When a daily dose alarm is set off, the display will show Figure 34, :
the buzzer will emit the following audio signal while the alarm is active:
1 second
< >
This audio signal may be unlimited or limited to 32 seconds, depending on the configuration that has
been programmed.
The audio signal is accompanied by the flashing of the red alarm LED.
When the audio signal stops because it has been limited in time, Figure 34 will remain on the display
until it is cancelled by the wearer.
The " MODE" and " SCROLL" keys will have no effect.
43
5.3.4. Time alarm
The time alarm is triggered when the elapsed time is greater than the time alarm level.
When a time alarm is set off, the display will show Figure 35 :
Figure 35
During the complete duration of the alarm, the buzzer will emit the following audio signal :
1 second
< >
This audio signal may be unlimited or limited to 32 seconds, depending on the configuration that has
been programmed.
The audio signal is accompanied by the flashing of the red alarm LED.
When the sound signal stops because it has been limited in time, Figure 35 will remain on the display
until it is cancelled by the wearer.
The " MODE" and " SCROLL" keys will have no effect.
44
5.3.5. Monthly dose alarm
The monthly dose alarm is triggered when the integrated monthly dose is greater than the monthly
dose’s alarm level.
When a monthly dose alarm is set off, the display will show Figure 36:
During the 12 second duration of the alarm, the buzzer will emit the following audio signal :
1 second
< >
The audio signal is accompanied by the flashing of the red alarm LED.
When the audio signal stops after 12 seconds, Figure 36 will remain on the display until it is cancelled
by the wearer.
The " MODE" and " SCROLL" keys will have no effect.
45
5.3.6. Three-month dose alarm
three-monththree-monthThe three-month dose alarm is triggered when the integrated dose for the last
ninety days is greater than the three-month dose’s alarm level.
When a three-month dose alarm is set off, the display will show Figure 37, as follows:
3 MTH
))) The alarm pictogram
.))) will be flashing.
The Hp symbol
Hp :
will be flashing. . mrem
Figure 37
During the 12 second duration of the alarm, the buzzer will emit the following audio signal:
1 second
< >
The audio signal is accompanied by the flashing of the red alarm LED.
When the audio signal stops after 12 seconds, Figure 37 will remain on the display until it is cancelled
by the wearer.
The " MODE" and " SCROLL" keys will have no effect.
46
5.3.7. Indicators
The following indicators are given mainly by the buzzer, and it is not necessary to press the " VAL"
key to redisplay of the badge’s other functions.
1 second
< >
5.3.7.2. Indicator that a level value has been taken into account
The taking into account of a new value for a level (see paragraph 4.1 for the procedure of the
adjustment of the levels) by pressing on the " VAL" key is accompanied by an audio signal which
lasts for 2 seconds, as follows:
< >
2 seconds
47
6. MAINTENANCE
SYMPTOMS ACTIONS
LCD screen off The activation of the LCD screen can be timed in order to save the
battery. In this case, a simple pressure on one of the Dosicard’s three
buttons will activate the screen for the programmed duration (typically 4
minutes).
Low battery pictogram The (low battery) pictogram appears when the battery’s charge is
activated low. The buzzer will be activated every hour (7 long beeps). Refer to §
7.2.
Low battery pictogram The “HS” symbol appears 48 hours after the activation of the (low
and “HS” symbol battery) pictogram.
activated From then on, the Dosicard no longer measures doses. Refer quickly to §
7.2.
“StoP” displayed The “StoP” message appears five days after the “HS” symbol, that is, a
week after the (low battery) pictogram is activated.
The battery is then completely down The Dosicard no longer measures
doses and the clock has stopped. Nevertheless, the history of the
integrated doses is preserved in EEPROM memory. In the absence of
battery, the content of the EEPROM memory will be saved for 10 years.
No access to the dose Access to the dose history (1 month, 3 months, 1 year and 5 years
history doses) is protected by a user code. The code defined at the factory
before delivery is: 1000.
If access is denied to the dose history, check whether the original
access code has been modified later by the user (by means of the LCB
badge reader).
48
7. APPENDICES
49
Irradiation on PMMA phantom
range of caesium 137
R[Hp(10)] = 1.04 “ mrem ” / mrem
1,2
1
0,8
R' Hp (10) 0,6
0,4
0,2
0
10 100 1000
E (keV)
2
1,8
1,6
1,4
1,2
R'/R'th 1
0,8
Relative response (alpha)
0,6
0,4 Relative response (théta)
0,2
0
-90 -75 -60 -45 -30 -15 0 15 30 45 60 75 90
Angle (degrees)
Figure 39: Relative response (60Co) as a function of the incident angle
α=0
θ=0
Phantom plan
Phantom plan
50
Irradiation on PMMA phantom
with gamma of 65 keV
2
1,8
1,6
1,4
1,2
R'/R'th
1
0,8
0,6 Relative response (alpha)
0,4
Relative response (théta)
0,2
0
-90 -75 -60 -45 -30 -15 0 15 30 45 60 75 90
Angle (degrees)
51
Table of the CEM compliance limits:
Comply Comply
with the with the
20 V/m ENV 50- ENV 50-
204 204
standard standard
Comply Comply
with the with the
50 V/m ENV 50- ENV 50-
204 204
standard standard
52
7.2. WHEN AND HOW TO CHANGE DOSICARD BATTERY
The Dosicard is supplied with a battery which is a standard lithium 3 V button battery, CR 2450 type
(article code 41514), available in watchmakers’ and photographers’ shops. The life of the battery is 5
months in "permanent" mode and up to about one year in "Stand-by" and "Daily timer" modes.
The Dosicard itself carries out an hourly check of the battery voltage.
− At the first negative check, the (low battery) pictogram will start to flash.
− After this, at each negative hourly check, the buzzer emits 7 long beeps and the date and time are
memorised.
− 48 hours after the first negative check, the dose measurement function of the badge will be de-
activated and the "HS" (Hors Service = Out of service) symbol will appear continuously on the
screen.
− One week after the first negative check, all the functions of the badge will be stopped, and the
indication "StoP" will appear on the screen. From this moment, the badge no longer tells the right
time.
• It is strongly recommended to
change the battery as soon
the battery symbol
appears on the LCD display.
• Two days after the battery
symbol was displayed, the text
"HS" (Out of order) appears an
the dosimetry capabilities are
not longer valid because the
analog part of the Dosicard is
not powered to save the
remaining electric power.
Then, there is a potential risk
that Dosicard loses its time
information and in a few
cases, its complete
configuration (display of "5S").
• Dosicard
Seven days can after
be blocked (no
the battery
change was
symbol available, alarm
displayed, not
the text
valid) in(clock
"StoP" casestopped)
of use of bad
means
quality
that thebattery.
time will certainly be
corrupted and the
configuration has very high
• pItrobability
is totherefore
be modified.highly
recommended to change the
battery AS SOON AS
POSSIBLE after the symbol
is displayed and before
"HS" or "StoP" appear. This
has to be done assuming a
good feedback of each
• Dosicard which is notbatteries
The recommended in use
(stock,
are: holidays, etc...).
DURACELL and VARTA.
53
• If Dosicard configuration has
changed, we recommend to
configure it with one LCB
badge reader. This can be
done by the customer,
representative or by Canberra
if suspected Dosicard is sent
• toWhatever
local service center.historical
happens,
data are available for 10 years
in non-volatile memory
(EEPROM) even without
battery.
54
7.2.2. How to change the Dosicard battery?
Screwdriver
55
• Fit the cover’s side clip in the
Dosicard’s housing hole on the
edge with 30° angle. Be sure
that the screw is in its position
while you install the metallic
cover.
56
SYMPTOMS ACTIO NS
Neither beep nor flash are emitted when you Reopen the cover and wait for at least 1 minute,
close the cover. then close it again firmlyt. One beep and one
light flash are emitted to confirm that the cover is
The red LED remains on all the time.
properly seated. If this is not the case, repeat the
No display on the LCD screen. operation.
TIME LOST: If after the battery change, the daily Set the Dosicard’s time in the with one LCB and
dose and historical data are not readable and the DOSE MANAGER software. The LCB reader is
"month"/"year" symbol are blinking on the LCD mandatory for this setting.
display, Dosicard lost its time parameter.
This usually occurs when the battery is changed
or when the rear metallic cover is opened
WITHOUT the low battery symbol or "HS"
symbol displayed. It may also happen when the
battery is dead ("StoP" displayed) or the battery
is good but the rear metallic cover has not been
closed correctly.
57
7.3. TEST PROCEDURE
Every Dosicard badge has an internal test function which is automatically launched once per day at
midnight.
A test can also be performed manually.
To start the test procedure, see paragraph 4.
58
7.4. GLOSSARY
Buzzer
− The buzzer is a sound warning.
The maximum intensity of the buzzer is 70 dB, next to the badge, without its pocket.
59
Buzz, Buzz dose, Incremental dose
Buzz describes a specific audible signal emitted by the buzzer, consisting of 4 times 2 short rapid
beeps.
Buzz occurs each time the integrated dose increases by a given amount referred to as the incremental
dose. For example, if the incremental dose is 0.3 mrem, the buzz signal will be heard when the
integrated dose reaches 0.3 mrem, 0.6 mrem, 0.9 mrem, 1.2 mrem, etc.
The frequency with which this signal is repeated gives an idea of the level of the dose rate without
having to look at the display of the Dosicard.
60
Dose, Manual dose, Integrated dose, Cumulative dose, Maximum allowable doses
Throughout the text, the " dose-equivalent " quantity is referred to as "dose". The symbol used is Hp: H
for the dose-equivalent and p for personal.
The Dosicard measures the personal dose-equivalent which penetrates to a depth of 10 mm of tissue
equivalent in accordance with ICRU 39 recommendations.
The manual dose is the dose integrated between a " START " and a " STOP ", given either manually
(two button presses by the wearer) or by the system (passage through a badge entrance reader and
then through a badge exit reader)
The integrated dose is the dose received in the current day.
The "MTH" cumulative dose is the result of the addition of the integrated doses for the 30 previous
days.
The "5 YR" cumulative dose is the result of the addition of the cumulative doses of the 60 previous
calendar months.
− A dose of 5 rem per year corresponds to a rate of 2.5 mrem/h during 2000 hours of work a year.
− A dose of 1.5 rem per year corresponds to a rate of 0.75 mrem/h during 2000 hours of work a year.
− A dose of 0.5 rem per year can be reached by aircrew on long-distance flights (0.25 mrem/h, or
2.8 mrem on a Paris-Los Angeles flight).
− A dose corresponding to a natural radioactivity level of 0.01 mrem/h is about 0.1 rem per year
(0.25 mrem per 24 hours, 7 mrem per month).
− For comparison purposes, the dose received during a chest X-ray is of the order of 30 mrem.
Dose equivalent: See Dose
Dose rate, Instantaneous dose rate, Rate
Throughout the text, the "dose-equivalent rate " is referred to as the "rate" or the "dose rate". The
•
symbol used is H p: H for the dose-equivalent, p for personal and • for the rate.
The Dosicard measures the rate of deep dose-equivalent Hp (10) which penetrates to a depth of 10 mm
of tissue equivalent in accordance with the ICRU 39 recommendations.
The dose rate is expressed in mrem/h (millirem per hour), rem/h (rem per hour).
The level of the mean natural radioactivity at sea level is slightly less than 0.01 mrem/h. Natural
radioactivity is higher in some regions or in altitude (0.25 mrem/h in a subsonic plane, 1.1 mrem/h in
Concord and 3.5 mrem/h in low terrestrial orbit).
Light-emitting diode
The red light-emitting diode is a luminous warning which is visible to the wearer when the Dosicard
badge is worn in the pocket in the correct position, that is, with the diode at the top of the protective
pocket.
When an alarm is tripped, short flashes are emitted by the diode at a frequency of one flash per second,
whatever the type of alarm.
The visual alarm is normally given at the same time as the audible alarm and the flashing of the display
(particularly the alarm pictogram •))) ).
61
Liquid crystal display, lcd: See Display
Rem, millirem
The dose-equivalent describes the risk of irradiation from the biological point of view.
For individual supervision, the personal dose-equivalent is often used. Its symbol is Hp, H for the dose-
equivalent and p for personal.
The Dosicard measures Hp(10), which is the personal dose-equivalent which penetrates to a depth of
10 mm of tissue equivalent, in accordance with ICRU 39 recommendations.
62
____________________________________________________________
Warranty
Canberra (we, us, our) warrants to the customer (you, your) that for a period of ninety (90) days from the date
of shipment, software provided by us in connection with equipment manufactured by us shall operate in
accordance with applicable specifications when used with equipment manufactured by us and that the media
on which the software is provided shall be free from defects. We also warrant that (A) equipment manufactured
by us shall be free from defects in materials and workmanship for a period of one (1) year from the date of
shipment of such equipment, and (B) services performed by us in connection with such equipment, such as
site supervision and installation services relating to the equipment, shall be free from defects for a period of
one (1) year from the date of performance of such services.
If defects in materials or workmanship are discovered within the applicable warranty period as set forth above,
we shall, at our option and cost, (A) in the case of defective software or equipment, either repair or replace the
software or equipment, or (B) in the case of defective services, reperform such services.
LIMITATIONS
EXCEPT AS SET FORTH HEREIN, NO OTHER WARRANTIES OR REMEDIES, WHETHER STATUTORY,
WRITTEN, ORAL, EXPRESSED, IMPLIED (INCLUDING WITHOUT LIMITATION, THE WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE) OR OTHERWISE, SHALL APPLY. IN
NO EVENT SHALL CANBERRA HAVE ANY LIABILITY FOR ANY SPECIAL, EXEMPLARY, PUNITIVE,
INDIRECT OR CONSEQUENTIAL LOSSES OR DAMAGES OF ANY NATURE WHATSOEVER, WHETHER
AS A RESULT OF BREACH OF CONTRACT, TORT LIABILITY (INCLUDING NEGLIGENCE), STRICT
LIABILITY OR OTHERWISE. REPAIR OR REPLACEMENT OF THE SOFTWARE OR EQUIPMENT DURING
THE APPLICABLE WARRANTY PERIOD AT CANBERRA'S COST, OR, IN THE CASE OF DEFECTIVE
SERVICES, REPERFORMANCE AT CANBERRA'S COST, IS YOUR SOLE AND EXCLUSIVE REMEDY
UNDER THIS WARRANTY.
EXCLUSIONS
Our warranty does not cover damage to equipment which has been altered or modified without our written
permission or damage which has been caused by abuse, misuse, accident, neglect or unusual physical or
electrical stress, as determined by our Service Personnel.
We are under no obligation to provide warranty service if adjustment or repair is required because of damage
caused by other than ordinary use or if the equipment is serviced or repaired, or if an attempt is made to
service or repair the equipment, by other than our Service Personnel without our prior approval.
Our warranty does not cover detector damage due to neutrons or heavy charged particles. Failure of beryllium,
carbon composite, or polymer windows, or of windowless detectors caused by physical or chemical damage
from the environment is not covered by warranty.
We are not responsible for damage sustained in transit. You should examine shipments upon receipt for
evidence of damage caused in transit. If damage is found, notify us and the carrier immediately. Keep all
packages, materials and documents, including the freight bill, invoice and packing list.
Software License
When purchasing our software, you have purchased a license to use the software, not the software itself.
Because title to the software remains with us, you may not sell, distribute or otherwise transfer the software.
This license allows you to use the software on only one computer at a time. You must get our written
permission for any exception to this limited license.
BACKUP COPIES
Our software is protected by United States Copyright Law and by International Copyright Treaties. You have
our express permission to make one archival copy of the software for backup protection. You may not copy our
software or any part of it for any other purpose.