Professional Documents
Culture Documents
Epic - SSR Board Replacement
Epic - SSR Board Replacement
WARNING: the SSR and the Epic main boards are sensitive to electrostatic discharges.
Handle them with the required caution.
If you remove an SSR board from an EPIC Plus device that will be operated again without the
SSR option, remember to close the socket in the image below with a jumper between the pins
number 1 and 3.
The MASKING feature will be not working if the socket is not properly closed.
Figura 1. The image shows how to close the socket for the MASKING feature after removing an SSR board.