How to replace the SSR board on the EPIC Plus device

WARNING: the SSR and the Epic main boards are sensitive to electrostatic discharges.
Handle them with the required caution.

1. Open the main unit case,


2. release the fixing screw from the board to be replaced,
3. mount the new board paying attention to the correct pin insertion. Handle the board with
care.
4. Fix again the board through the screw.
5. The SSR board is tested and already set; do not change any setting in the hardware.

If you remove an SSR board from an EPIC Plus device that will be operated again without the
SSR option, remember to close the socket in the image below with a jumper between the pins
number 1 and 3.

The MASKING feature will be not working if the socket is not properly closed.
Figura 1. The image shows how to close the socket for the MASKING feature after removing an SSR board.