Deuterium Lamps (D Lamps)
(D2 LAMPS)
Hamamatsu deuterium lamps
LAMPS
FOR HIGH PERFORMANCE
£Problems with prior lamps using quartz glass
Premature deterioration in transmittance
cause by UV light
£We solved these problems
by using UV-transmitting
glass superior to quartz
glass!!
Quality variations due to production
DEVICES process and material inclusions
High resistance to UV light
Troublesome ozone generation
1 0
0 10 100
TIME (h)
1000 10000
High stability
Light output stability Maintains high stability until product lifetime end
3 2
Stability at initial operation Stability after 2000 hours of operation
20 20
15 15
10 10
ABS (10-5 A.U.)
2 Ceramic electrode
£Problems with prior lamps using £We solved these problems by
(Metal electrode) using ceramic electrodes with
excellent temperature
Low stability due to being susceptible to characteristics!!
ambient conditions
Excellent temperature
Large variation in light output because characteristics
electrode spacing is not uniform ensure high stability
Uniform electrode spacing
minimizes variations
in characteristics
P3 P5 P11 P14
*As of Feb. 2016 according to our research *As of Feb. 2016 according to our research
Window material
Applications UV glass Synthetic silica MgF2
HPLC (High Performance Liquid Chromatography)
UV-VIS Spectrophotometer
CE (Capillary Electrophoresis)
Atomic Absorption Spectrophotometer
Thin Layer Chromatography
Water Quality, Air Pollution and Other Environmental Analyzer
Film Thickness Gauge
Semiconductor Testing Equipment
UV Resistance Evaluation of Materials
Photoionization Light Source
Static Electricity Removal by Vacuum UV Light
2
®
X2D2
LAMPS
The X2D2® lamps (high brightness & long
lifetime deuterium lamps) deliver unparalleled
brightness twice that of L2D2 lamps (see page
5) while still maintaining the high stability and
long lifetime offered by the conventional
deuterium lamps. These characteristics will
enhance sensitivity and throughput in various
photometric instruments.
Features Characteristics
●Long life: 2000 h ●Spectral distribution
TLSOB0024EF TLSOB0079EC
●High stability: 0.005 %(p-p) typ. 1 100
80
UV GLASS
(L9518, L10804)
40
0.01
20
0.001 0
160 200 240 280 320 360 400 100 120 140 160 180 200 220 240 400
0° 0° 0°
●L2D2 lamp Aperture size: 0.5 mm -10° -10° -10°
0° 0° 0°
3
Dimensional outline (Unit: mm)
L9518, L10804 L9519, L10904
50 50
-0.05 Tolerance of emitting point (center) -0.05 Tolerance of emitting point (center)
35 - 0.1 35 -0.1
A
A
With respect to axial line A: ±0.1 With respect to axial line A: ±0.1
30 ± 1 With respect to plane B: ±0.1 30 ± 1 With respect to plane B: ±0.1
ARC POINT ARC POINT
+0.15 +0.15
3 + 0.05 3 +0.05
59 MAX.
59 MAX.
ARC POINT ARC POINT
15
50
50
22 - 0.1
22 - 0.1
10
10
+0
+0
32
32
B B
2 × 3.3
5
5
2 × 3.3
14 ± 1
9 MAX.
9 MAX.
AUXILIARY AUXILIARY
IGNITION 22±0.05 22±0.05 IGNITION
DEVICE DEVICE 22±0.05 22±0.05
160 ± 10
160 ± 10
LIGHT OUTPUT
LIGHT OUTPUT
CONNECTOR (MOLEX) CONNECTOR (MOLEX)
CONNECTION HOUSING: 39-01-2040 CONNECTION
HOUSING: 39-01-2040
(70)
(70)
#1: AUXILIARY ELECTRODE CONTACT: 39-00-0428 #1: AUXILIARY ELECTRODE
CONTACT: 39-00-0428
#2: ANODE #2: ANODE
#3: FILAMENT #3: FILAMENT
#4: FILAMENT (GND) TLSOA0105EE #4: FILAMENT (GND) TLSOA0106EE
L9841 50
Tolerance of emitting point (center)
-0.05
35 -0.1 With respect to axial line A: ±0.1
Example of optical system layout for visible light
A
22 - 0.1
+0
50
10
32
25 ± 1
ANGLE OF INCIDENCE
19°
9 MAX.
AUXILIARY
IGNITION LIGHT
DEVICE 22±0.05 22±0.05
OUTPUT
160 ± 10
HALOGEN
LENS
LIGHT OUTPUT LAMP
(VISIBLE LIGHT)
CONNECTOR (MOLEX)
HOUSING: 39-01-2040 CONNECTION L10804, L10904
(70)
Specifications
Type No. L9518 L10804 L9519 L10904 L9841 Unit
Type Standard See-through Standard See-through Standard —
Window material UVglass Synthetic silica MgF2 —
Spectral distribution 185 to 400 160 to 400 115 to 400 E nm
Aperture diameter 0.5 mm
Output stability Drift (Max.) ±0.3 %/h
at 230 nm Fluctuation (p-p) Typ. 0.005 %
Guaranteed life at 230 nm A 2000 h
Discharge starting voltage (Max.) B 400 V dc
Anode current 300 ± 30 mA dc
Tube voltage (Typ.) 90 85 V dc
Voltage 2.5 ± 0.25 V dc
Warm-up
Current (Typ.) 4 A dc
Filament ratings
Voltage 1.7 ± 0.2 V dc
Operating
Current (Typ.) 3.3 A dc
Filament warm-up time (Min.) 20 s
Power supply C C9559, M9521 —
Lamp house E9522-50, E9558-50 —
Bulb wall temperature D 245 to 290 °C
ALamp life end is defined as the point when light output at 230 nm falls to 50 % of its initial value or when output fluctuations exceed 0.05 % (p-p).
BA trigger voltage must be applied to the anode and auxiliary electrode. CThe power supply for the L2D2 cannot be used to operate X2D2 lamps.
DRecommended temperature for operating a lamp in the lamp housing. Consult us on how to measure the temperature.
EDoes not support vacuum evacuation and so should be used in nitrogen atmosphere.
* Custom lamps not listed above will be available on request. Please feel free to contact us.
4
®
L2D2
LAMPS
The L2D2® lamps are UV light sources
with a long service lifetime and high
stability. These L2D2 lamps have charac-
teristics essential for light sources used
in chemical analysis instruments and
provide high measurement accuracy.
Specifications
A B Required
Output stability at 230 nm
Dimen- Guaranteed discharge Anode Tube
Window Spectral Aperture Fluctuation life starting
Type No. Type sional Drift (p-p) current voltage
C
outline
material distribution diameter at 230 nm voltage
Max. Typ. Max. Typ.
(nm) (mm) (%/ h) (%) (h) (V dc) (mA dc) (V dc)
L6565 1.0 4000 350
q
L6301
L6301-50 i
L6303 q
UV glass 185 to 400
L12313 e
L12313-50 u
0.5 ±0.3 0.005 2000 400
L6307 Standard
w
L6309
300 ± 30 80
L7296 r Synthetic
160 to 400
L7296-50 y silica
L12307 w UV glass 185 to 400
L7293 t
MgF2 115 to 400 1.0 — — 2000D 350
L7293-50 o
L6999 q
UV glass 185 to 400
L6999-50 i
See-through 0.5 ±0.3 0.005 2000 400
L9030 r Synthetic
160 to 400
L9030-50 y silica
ASee pages 7 and 8.
BLamp life end is defined as the point when light output at 230 nm falls to 50 % of its initial value or when output fluctuations exceed 0.05 % (p-p).
CA pulse voltage higher than this value must be supplied to start reliable lamp discharge.
DOperating life may vary depending on operating environmental conditions (vacuum atmosphere).
5
Characteristics
●Spectral distribution
Deuterium lamps emit high intensity light in the UV range at wavelengths shorter than 400 nm. Light intensity on the short wavelength side is deter-
mined by the window material used.
■UV glass, Synthetic silica ■MgF2
TLSOB0024EF TLSOB0079EC
1.0 100
IRRADIANCE (µW·cm-2·nm-1 at 50 cm)
SYNTHETIC SILICA
0.1
60
UV GLASS
40
0.01
20
0.001 0
160 200 240 280 320 360 400 100 120 140 160 180 200 220 240 400
0° 0° 0°
0° 0° 0°
See-through type
q L6301, L6565, L6303, L6999 w L6307, L6309 , L12307 e L12313
7 MAX.
7 MAX.
28 ± 1 30±1
7 MAX.
30 ± 1
68 ± 2
68 ± 2
60 ± 2
42 ± 2
42 ± 2
42 ± 2
160 ± 10
160 ± 10
CONNECTION
CONNECTION CONNECTION
L6303
20
20
FILAMENT : BLUE
20
FILAMENT : BLUE
FILAMENT : BLUE FILAMENT . GND : BLUE
FILAMENT.GND : BLACK
7
7
7
50 ± 1 - 0.05
35.0 - 0.1
30 ± 1 50 ± 1
7 MAX.
A
14 ± 1 30 ± 1
8 7 30 ± 1 22.0 ± 0.1 22.0 ± 0.1 ARC
POINT
2 × 3.3
15
15
68 ± 2
68 ± 2
22.0 - 0.1
68 ± 2
+0
ARC
42 ± 2
42 ± 2
POINT
42
ARC B
14 ± 1
15
15
POINT +0.15
FIXED 3 +0.05
RANGE
15 ± 0.5
160 ± 10
160 ± 10
160 ± 10
LIGHT OUTPUT
20
CONNECTION CONNECTION
7
CONNECTION
L7296 6
L7293 L7296-50
20
20
FILAMENT.GND : BLACK FILAMENT : BLUE FILAMENT·GND : BLACK Tolerance of emitting point (center)
ANODE : RED 6 6 ANODE : RED With respect to axial line A: ±0.1
ANODE : RED With respect to plane B: ±0.1
L9030 L9030-50
FILAMENT : BLUE FILAMENT : BLUE
FILAMENT : BLUE FILAMENT : BLUE
ANODE : RED ANODE : RED
TLSOA0017EG TLSOA0011EF TLSOA0075EE
7
See-through type
u L12313 -50 i L6301-50, L6999-50 o L7293-50
- 0.05
35.0 - 0.1 - 0.05
A
-0.05 35.0 - 0.1
7 MAX.
35 -0.1
A
7 MAX.
50 ± 1
7 MAX.
50 ± 1 28 ± 1
A
30 ± 1
ARC 30 ± 1 22.0 ± 0.1 22.0 ± 0.1 ARC
22 ± 0.1 22 ± 0.1 ARC
POINT POINT
POINT
50 ± 1
2 × 3.3 22.0 ± 0.1 22.0 ± 0.1 2 × 3.3
60 ± 2
68 ± 2
ARC
22.0 - 0.1
68 ± 2
37
+0
ARC
22 -0.1
POINT
+0
7
37
B POINT
42
2 × 3.3
15
B
5
O-RING
15
+0.15
3 +0.05 B
50 ± 1
15
22.0 - 0.1
8
FIXED
+0
5
RANGE +0.15
3 +0.05
ARC 15 ± 0.5
POINT
160 ± 10
160 ± 10
LIGHT OUTPUT +0.15
3 +0.05 15
160 ± 10
LIGHT OUTPUT LIGHT OUTPUT
20
FILAMENT : BLUE FILAMENT : BLUE FILAMENT : BLUE
7
7
ANODE : RED ANODE : RED ANODE : RED
6 6 6
Tolerance of emitting point (center) Tolerance of emitting point (center) Tolerance of emitting point (center)
With respect to axial line A: ±0.1 With respect to axial line A: ±0.1 With respect to axial line A: ±0.1
With respect to plane B: ±0.1 With respect to plane B: ±0.1
With respect to plane B: ±0.1
4
An example for optics of See-through type
TOP VIEW
40°
A ARC
POINT
LIGHT
B SCREW PORTION OUTPUT
HALOGEN
C LENS LAMP
1VACUUM FLANGE
2TIGHTENING SCREW SEE-THROUGH
3O-RING JIS B2401 L2D2 LAMP
CALL No. V15 TLSOC0011EF
4SPACER
AMgF2 WINDOW
BGRADED SEAL
CO-RING FIXED RANGE
▼Type No.
TLSO 0053EA L6999, L6999-50, L9030, L9030-50
8
X2D2® L2D2® Power supply for X2D2® lamps / Lamp housing
LAMPS LAMPS
POWER SUPPLY
FOR D2 LAMPS
LAMP HOUSE
Applications using deuterium lamps require very high stability ▲Power supply Left: C9559, Right: M9521 ▲Lamp housing Left: E9522-50, Right: E9558-50
of light output, so using a Hamamatsu dedicated power supply
and lamp house is recommended to operate these lamps. Power supply for L2D2® lamps / Lamp housing
When users are designing their own power supply and lamp
housing, we provide technical support and follow-up to ensure
an optimal optical design so please consult us when needed.
+0.1
2.8 - 0 WEDGE PRESSURE PIN
60
19.2
ARC POINT
3 × M4 DEPTH 5
LAMP FLANGE MOUNTING SURFACE
TOP VIEW (P.C.D.* 40)
5.8
79
3-1 32 22 2 × M3 DEPTH 5
20
C9559 °
22 ± 0.05 22 ± 0.05
3-120°
35.2
140
50
R2
35 18 3 × M3 (P.C.D.* 24) 34 ARC POINT
ARC POINT
5
110 280 +0.2
32.2 -0 OPTICAL SYSTEM MOUNTING SURFACE
FRONT VIEW SIDE VIEW REAR VIEW
TLSOA0114EA
●Supplied lamp output cable
●E9558-50 +0.1
3000 ± 50 2.8 - 0 WEDGE PRESSURE PIN
TLSOA0112EA 60
●M9521
OPTICAL SYSTEM
MOUNTING SURFACE 32
4.2
LAMP FLANGE MOUNTING SURFACE
3 × M4 DEPTH 5 79
(P.C.D.* 40) 29 ± 1 22 2 × M3 DEPTH 5
100
110
3-1
22 ± 0.05 22 ± 0.05
35.2
50
60
+0.1
2.8 - 0 WEDGE PRESSURE PIN
19.2
LAMP FLANGE
3 × M4 DEPTH 5
MOUNTING SURFACE
TOP VIEW (P.C.D.* 40)
5.8
84
3-1 37 22 ± 0.05 2 × M3 DEPTH 5
20
°
22 ± 0.05 22 ± 0.05
3-120°
90
35.2
50
117 200
R2
35 18 3 × M3 (P.C.D.* 24) 32 ARC POINT
ARC POINT
5
FRONT VIEW SIDE VIEW REAR VIEW +0.2
32.2 -0 OPTICAL SYSTEM MOUNTING SURFACE
●E9558
1800 ± 50
+0.1
60
TLSOA0116EA 2.8 - 0 WEDGE PRESSURE PIN
LAMP FLANGE
●M9596 OPTICAL SYSTEM
MOUNTING SURFACE 32 MOUNTING SURFACE
4.2
3 × M4 DEPTH 5 84 29 ± 1
22 2 × M3 DEPTH 5
(P.C.D.* 40)
5
ARC POINT 20°
R2
86.4
3-1
100
22 ± 0.05 22 ± 0.05
35.2
50
118 30 MAX. 5 MAX.
°
102 20
3-1 3 × M3 (P.C.D.* 24) 37 5 34 ARC POINT
FRONT VIEW SIDE VIEW
TLSOA0117EA TLSOA0119EA
Filament ratings
Warm-up Operation
Type No. Voltage Current Voltage Current Applicable lamp
(V dc) (A dc)(Typ.) (V dc) (A dc)(Typ.)
L6565, L7293, L6999, L6999-50, L7293-50
C9598/M9596-2510 2.5 ± 0.2 4 1.0 ± 0.1 1.8
L6301, L6301-50, L9030, L9030-50
C9598/M9596-2517 2.5 ± 0.2 4 1.7 ± 0.2 3.3 L6303
C9598/M9596-3000 3 ± 0.2 5 0 0 L12313
C9598/M9596-1035 10 ± 0.5 0.8 3.5 ± 0.2 0.3 L6307
C9598/M9596-1070 10 ± 0.5 1.2 7 ± 0.4 1 L7296, L6309, L7296-50
C9598/M9596-1555 13.5 ± 0.7 0.5 5.25 ± 0.25 0.3 L12307
10
®
S2D2
LAMPS
The S2D2® lamps are compact deuterium
lamps with a drastically reduced size
compared to ordinary deuterium lamps.
Despite their compact body, the S2D2 lamps
have the same high stability as conventional
deuterium lamps and a unique electrode
structure that delivers high brightness.
▲Left: L13301
Right: L10671D
Features Characteristics
●Long life: 1500 h (L10671D) ●Spectral distribution ●Directivity (Light distribution)
TLSOB0109EB
●Compact
0.010
L10671D L13301
IRRADIANCE (µW·cm-2·nm-1) at 50 cm
0.009
20°
●High output UV continuous 0.007
10° 10°
0.006
spectrum 0.005
L10671D 0° 0°
80
MgF2 (L13301)
-20° -20°
-30° -30°
60
TLS B0005EA TLSOB0114EA
DETECTOR
40
20
TLS C0005EA
0
100 120 140 160 180 200 220 240 400
WAVELENGTH (nm)
Specifications
Parameter Description / Value Unit
Type No. L10671D L13301 —
Window material UV glass MgF2 —
Spectral distribution 185 to 400 115 to 400 nm
Aperture diameter 1.0 mm
Output stability Drift (Max.) ±0.25 %/h
at 230 nm Fluctuation (p-p) (Typ.) 0.005 %
Guaranteed life at 230 nm A 1500 1000 D h
Output current 30 50 mA
Output voltage (Typ.) Approx. 135 V
Voltage 4.2 V
Warm-up
Current (Typ.) 0.55 A dc
Filament ratings
Voltage 3.5 V
Operating
Current (Typ.) 0.5 A dc
Filament warm-up time (Min.) Approx. 25 s
Bulb wall temperature B (Max.) 185 240 °C
Storage temperature -10 to +60 °C
Storage humidity Below 85 % (No condensation) —
ALamp life end is defined as the point when light output at 230 nm falls to 50 % of its initial value or when output fluctuations exceed 0.05 % (p-p).
11 B Position “a” in dimensional outline COperating life may vary depending on operating environmental conditions (vacuum atmosphere).
Dimensional outline (Unit: mm)
* PCD (Pitch Circle Diameter)
●L10671D ●L13301 O-RING FIXED RANGE TEMPERATURE
MEASUREMENT P.C.D.* 10.72 ± 0.2
TEMPERATURE P.C.D.* 10.72 ± 0.2 15.0 ± 0.5
TEMPERATURE TEMPERATURE POINT a
MEASUREMENT POINT a A 58 MAX. A
MEASUREMENT MEASUREMENT
POINT a 20.5 MAX. 8 MAX. 18° 10° POINT a 20 5 15.5 ± 2.0 8 MAX. 18° 10°
9 × 0.7 9 × 0.7
36°
36°
7 8 7 8
36°
36 °
22 MAX.
9 B 9 B
6 6
20.5
15
5 5 ●CONNECTION
●CONNECTION
36 °
36°
3 6°
36 °
4 1 4 1 1: FILAMENT
3 2 1: FILAMENT 3 2
3 (7): ANODE
3 (7): ANODE ARC POINT 5: AUXILIARY
ARC POINT 7.5 6 9 × 0.7 18° 10° 5: AUXILIARY 7.5 6 9 × 0.7 18° 10° ELECTRODE
ELECTRODE EMISSION POINT POSITION TOLERANCE (CENTER) 9: GND
EMISSION POINT POSITION TOLERANCE (CENTER) 9: GND · With Respect To Axis Line A: ±0.3 mm 2, 4, 6, 8: N.C.
· With Respect To Axis Line A: ±0.3 mm 2, 4, 6, 8: N.C. · With Respect To Axis Line B: ±0.3 mm
· With Respect To Axis Line B: ±0.3 mm
TLSOA0109EA TLSOA0016EA
RELATED PRODUCTS
Power supply
●L10671P (for L10671)
3× 3
CN9: CONNECTOR FOR 12 V CN1: JACK FOR DC INPUT VOLTAGE (MOUNTING HOLE) CN4: CONNECTOR
1: + 3 3 FOR 5 V OUTPUT
Parameter Description / Value Unit 2: – (Usage example: Tungsten Lamp)
1: +
2: –
Input voltage (DC) 12 * V CN2: EXTERNAL CONTROL
3
27
CONNECTOR
CN5: CONNECTOR
Power consumption Max. 10 VA FOR 5 V OUTPUT
65
(Usage example: Shutter Solenoid)
S2D2 lamp ON /OFF — 1: +
2: N.C.
20
CN4 output ON/OFF — 3: –
3
8 17 30
55 CN3: CONNECTOR
External control Status signal
18 MAX.
TOP VIEW FOR S2D2 LAMP TOP VIEW
OUTPUT (Bottom Board Only)
S2D2 lamp —
( )
1: GND
2: FILAMENT
CN4
32.5 MAX.
3: N.C.
4: AUXILIARY BOARD = 1.2 mm
Main power ELECTRODE
9.2
5: ANODE
AEach time a signal for external control is input across
2.5 MAX.
SIDE VIEW
terminals 3 and 10, voltages +5 V and -5 V are alternately TLS A0005EA
output at 200 ms intervals.
* Input voltage range is from 8.5 V dc to 13.2 V dc.
1000 ± 100 112.5
BAt position “b” in the L10671H dimensional outline. (When this lamp
housing is installed in equipment, thermal design specs must be consid-
ered to ensure the operating temperature will be within this range.) ▼CONNECTION
SOCKET
23 6 1: ANODE
130
TEMPERATURE 2: AUXILIARY ELECTRODE
MEASUREMENT POINT b 2 × M2.6 8 MAX.
3: FILAMENT
4: GND
28 5 ARC POINT
ARC
POINT FRONT VIEW 6 SIDE VIEW M11 × 0.5 (DEPTH 3.75) *
2 × 2.7 * 10 mm dia. lens can be placed.
FOR M2.6 FLATHEAD SCREW 13
13
2 × M2.6 DEPTH 2
BOTTOM VIEW
(Screw holes for mounting) TLS A0006EA * Please consult us on the housing and vacuum flange for the L13301.
12
RELATED PRODUCTS
Fiber light source
■APPLICATIONS
●Spectrophotometry ●Microtiter plate reader ●Biological measurement
●Environmental measurement ●Pharmaceutical testing ●Semiconductor inspection
●Absorption spectrum measurement ●High-performance liquid chromatography ●Optical component inspection
15
10
●200 nm to 220 nm ●200 nm to 1100 nm
5
0
(enlarged view of enhanced output region) TLSZB0125EA
1.4 10
-5
-10 L12515 L12515
1.2 L10671 (conventional type) L10671 (conventional type)
-15
-20
RELATIVE IRRADIANCE
RELATIVE IRRADIANCE
1
●at 600 nm 1
20
ABSORBANCE (10-5 A.U.)
0.8
15
10
5
0.6 3
0
0.4
Times 0.1
-5
-10 LIGHT GUIDE LIGHT GUIDE
-15 0.2 A7969-08AS A7969-08AS
Core diameter= 0.8 mm CORE DIAMETER= 0.8 mm
-20
TIME (100 s/div.) 0 0.01
TLSOB0107EA 200 202 204 206 208 210 212 214 216 218 220 200 300 400 500 600 700 800 900 1000 1100
Characteristics Features
▲L10290
●Spectral distribution (Typical data) ●High output: Twice intensity
TLSOB0099EB
10 (Compared to conventional model)
L10290
WHEN TUNGSTEN-HALOGEN
LAMP WITH MAX. VOLTAGE
●High stability: Fluctuation 0.004 % p-p (Typ.)
Twice intensity L10290
(equivalent to 2 × 10-5 A.U.)
RELATIVE IRRADIANCE
WHEN TUNGSTEN-HALOGEN
(compared to conventional model) LAMP WITH MIN. VOLTAGE
CONVENTIONAL MODEL ●Long life lamp: 2000 hours
●External control
1
●Shutter function
●Filter holder (Sold separately)
LIGHT GUIDE
A7969-08AS
CORE DIA.= 0.8 mm
0.1
200 300 400 500 600 700 800 900 1000 1100
WAVELENGTH (nm)
13
VUV light source
■APPLICATIONS
●Electrostatic remover ●Semiconductor inspection ●Dechucking of electrostatic chucks
●Film thickness measurement ●Material resistance evalution ●LCD manufacturing equipment
●Spectrophotometry ●Photoionization source
Features Characteristics
●High brightness: 6 times ●Spectral distribution ●Arc distribution
TLSOB0079EB TLS B0007EB
650
(Compared to L2D2 lamp) 100
600
●High stability: Fluctuation 0.05 % p-p (Max.) 550 High
RELATIVE IRRADIANCE (%)
efficiently neutralizes large areas in a vacuum. Left: Light source, Right: Power supply -30° TLSZB0105EA
■WARRANTY
Lamps are warranted for a period of one year from the date of shipment. If a lamp is found to be defective within this warranty period, Hamamatsu will
replace the defective lamp without charge. (This warranty is limited to replacement of the defective lamp.) Even if within the warranty period (one
year), the warranty shall not apply to cases where the lamp operation time has exceeded the guaranteed life, or the trouble was caused by incorrect
operation or natural or man-made disasters.
■DISPOSAL OF LAMPS
When disposing of the used lamp, take appropriate measures in compliance with applicable regulations regarding waste disposal and correctly
dispose of it yourself, or entrust disposal to a licensed industrial waste disposal company.
In any case, be sure to comply with the regulations in your country, state, region or province to ensure the used lamp is disposed of legally and
correctly.
Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office.
Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are
subject to change without notice. No patent rights are granted to any of the circuits described herein. ©2016 Hamamatsu Photonics K.K.