Professional Documents
Culture Documents
6 7 8 Biopotential Amplifiers
6 7 8 Biopotential Amplifiers
2
Basic Features of Biopotential Amplifier
Purpose:
To provide voltage and/or current gain to increase the amplitude of weak
electric signals of biological origin
Features:
High input impedance (minimize the loading effects of the amplifier inputs)
Protection circuitry (limit the possibility of introducing dangerous micro-shocks
or macro-shocks at the input terminals of the amplifier)
Low output impedance (low with respect to the load being driven)
Adequate output current (to supply the power needed to drive the load)
Basic Features of Biopotential Amplifier… cont.
Features
Bandlimited frequency response (match the frequency response of the signal
being measured to eliminate out-of-band noise)
Quick calibration (include a signal source and a number of selectable fixed
gains settings)
High common-mode rejection for differential amplifiers (common mode
signals are frequently larger than the biopotentials being measured)
Additional specific requirements for each application
5
Biopotential amplifier Schematic
6
Differential Amplifier
v3
v4
Solves Low Zi problem --- Solves low Zi problem and add gain
but common mode signa; is also amplified
Rough sketch of the dipole field of the heart when the R wave is
maximal. The dipole consists of the points of equal positive and
negative charge separated from one another and denoted by the
dipole moment vector M.
Frontal-Plane Vector ECG
+ VF –
Augmented Leads
Precordial chest leads are used to record the voltage difference between
these electrodes and Wilson’s Central Terminal.
Standard ECG
V4
II
Right leg
electrode
Electrocardiograph
Driven
Sensing Lead-fail right leg ADC Memory
electrodes detect
Sensing electrodes circuit
Auto calibration
Preamplifier Isolated
Auto Baseline
power
Baseline restoration calibration restoration
supply
Microcomputer Control
program
Control software Keyboard
Frequency distortion
High-frequency loss rounds the sharp edges of the QRS complex.
Low-frequency loss can distort the baseline (no longer horizontal) or cause monophasic waveforms
to appear biphasic.
Saturation/cutoff distortion
Combination of input amplitude & offset voltage drives amplifier into saturation
Positive case: clips off the top of the R wave
Negative case: clips off the Q, S, P and T waves
Ground loops
Patients are connected to multiple pieces of equipment; each has a ground (power line or common
room ground wire)
If more that one instrument has a ground electrode connected to the patient, a ground loop exists.
Power line ground can be different for each item of equipment, sending current through the patient
and introducing common-mode noise.
Open lead wires
Can be detected by impedance monitoring.
Artifacts
An electrocardiograph has a broad frequency response so that its amplifier has a first-
order time constant of 16 s. The electrocardiograph amplifier has a broad dynamic
range of input voltages, but any input voltage greater than 2mV will be out the range
of its display and cut off.
While recording the ECG of a patient, a transient occurs that has an amplitude of 10
mV, and this causes the ECG to fall out of the range of the instrument’s display. If the
ECG R wave has an amplitude of 1 mV, how long will it take for the entire signal to be
visible on the display?
Power-Line Coupling
Sources
Power lines
Electromagnetic radiation
Patient leads become antennas, especially if detached.
Sources
Radio and Television
Radar
Research equipment
Electrosurgical devices
Arching fluorescent lights (needing replacement)
Remedy
Employ capacitors shunting the inputs to ground (eg., 200 pF).
Do not lower the input impedance of the amplifier.
Amplifier Protection
Electrostatic discharge
High voltages due to electrosurgical equipment
Leads shorted to high voltage by hospital personnel
Remedy
Voltage limiting devices on each input lead are used to
protect the equipment
Noise: 40 mV p-p
Frequency response
.05 to 150 Hz
100 - 200 mA
through the patient
Amplifiers for other Biopotentials
EMG
EEG
Glass micropipette Amp
Amplifiers for other Biopotentials
Skeletal
Function: Generate Force
Excitable
Contractible
Skeletal Muscles
Fibers
(10 mm to 80 m m Dia.)
Myofibrils
(100-1000th)
EMG signals
Frequency: 25 Hz to several kHz.
Amplitudes 100 µV to 90 mV, depending on the type of signal and
electrodes used.
Thus EMG amplifiers must have a wider frequency response than ECG
amplifiers,
They do not have to cover so low a frequency range as the ECGs.
EMG Amplifier: Electrodes
Skin-surface electrodes:
signals peak amplitudes are of the order of 100 µV to 1 mV.
Electrode impedance is relatively low, ranging from about 200 to 5000 ohm,
The amplifier must have somewhat higher gain than the ECG amplifier, and
Its input characteristics should be almost the same as those of the ECG amplifier.
The 3rd electrode can be stuck to the bone in elbow of the same arm
and that is connected to Reference in the EMG circuit.
Basics and Design concepts: Amp and Filters
Frequencies: 25 Hz (or lower) up to 1 kHz (or higher).
The usable energy of EMG signal is dominant between 50-150 Hz.
DC offset or Bias problem
Bias Adjustment
HPF
The dominant common mode voltage signals on our bodies is usually a
50/60‐ Hz sine wave that is capacitively coupled to us
Reject this signal by using differential amplifier with high common‐mode
rejection
Basics and Design concepts : Envelope Detection
Add a circuit that will give us a “running average” of the amplitude of the
EMG signal:
Calculate the amplitude of this signal by rectifying it
However, diodes need approximately 700 mV of forward bias before they begin
conducting, and our signals (after amplification) are less than this.
We must use a “precision rectifier” circuit,
Smooth the rectified signal to generate an “envelope” of the signal
Basics and Design concepts: Block Diagram
Basics and Design concepts: Precision Rectifier
70 Prof Mohamed El-Brawany 07-Dec-21
Basics and Design concepts: Envelope Detection