Download as pdf or txt
Download as pdf or txt
You are on page 1of 20

Data

MAMMOMAT 1000 / 3000 Nova


Universal mammography system for screening
and diagnosis

s
medical
MAMMOMAT 1000 / 3000 Nova

Universal mammography system


for screening and diagnosis

With MAMMOMAT 1000 and MAMMOMAT 3000 Our commitment to increase patient
Nova we are offering two analog mammography comfort is demonstrated by innovations such as:
systems that meet the demands of reliable mammo- • Flex2 – the compression plate that automatically
graphy examinations as well as the European adapts to the patient‘s anatomy
guidelines for screening examinations.
The systems are ready for the connection of CR • SoftSpeed – reduces the speed of the com-
readers* from different manufacturers, providing pression plate as soon as it touches the breast
the first introduction into the digital world. • Opcomp – optimized compression with auto-
matic selection of the required pressure
Our mid-range mammography system
MAMMOMAT 1000 is designed for high patient
User convenience for an efficient workflow
throughput and is mainly used for screening.
With our high-end mammography system Isocentric rotation
MAMMOMAT 3000 Nova you have your clinical
Due to the motorized, isocentric rotation of the
applications in one unit.
swivel arm, the patient can be positioned quickly
The system supports both screening examinations
and accurately.
and diagnostic applications featuring special views,
The object table remains at the correct height
magnifications, specimen radiography and stereo-
between projections throughout the examination,
tactic biopsy, making the system a highly efficient
even if the angle is changed.
multi-use platform.
Format changes*, 1)
Both systems offer:
The unique dual Bucky design enables quick and
Excellent image quality easy switching between 18 cm x 24 cm and
24 cm x 30 cm formats.
Equipped with special mammography X-ray tubes,
If the object table is changed, the system auto-
the MAMMOMAT 1000 and MAMMOMAT 3000
matically collimates to the film format used.
Nova allow optimal imaging of breasts of different
In addition, each object table can be manually
thickness and density.
replaced by a stereotactic table or magnification
• Opdose – Preprogrammed auto-selection of X-ray table.
parameters for optimized dose to the patient
Quickswitch
High patient comfort and patient access
Cassettes can be changed easily within seconds
The narrow X-ray tube unit and slim design enable using only one hand.
easy and accurate positioning even of patients
sitting in a wheelchair.

* Option
1)
Only for MAMMOMAT 3000 Nova

2
MAMMOMAT 1000 / 3000 Nova

Technical Data

Basic unit
Universal mammography system for screening, diagnosis and stereotactic biopsies1) on standing, seated or recumbent
patients
The system consists of an examination stand and a microprocessor-controlled, high-frequency generator
Motorized, height-adjustable object table
Motorized, isocentric rotation of the swivel arm with preset rotation angle, other angles preselectable
Opcomp – microprocessor-controlled automatic compression device to optimize compression force for the best image
quality and low dose
Digital stand display in an easy-to-view position showing compression force, compressed breast thickness and projection
angle. LEDs indicate optimized compression and system ready for exposure
Control panel on the generator or optionally on the radiation shield. The control panel displays all relevant exposure
parameters and operator messages
Opdose – Pre-programmable exposure parameters for optimized patient dose, selection of 4 different programs
AEC Automatic Exposure Control
Integrated positioning/collimator light
Object table with Bucky 18 cm x 24 cm

X-ray stand
Height adjustment (motorized) 65 cm to 135 cm
Swivel range (motorized) + 135° to – 180°
Source-to-Image Distance (SID) 65 cm
Compression 3 kg (6.5 Ibs) to 20 kg (44 Ibs)
Grid Reciprocating, grid ratio 4:1, 27 lines/cm
Magnification factors* 1.5 or 1.8, geometric
Radiation shield* 0.1 mm or 0.3 mm lead equivalent lead glass

* Option
1)
Only for MAMMOMAT 3000 Nova

4
X-ray generator
Microprocessor-controlled high-frequency generator (20 kHz)
Power output (IEC 601) 5 kW
kV range 23 kV to 35 kV (adjustable in 1 kV increments)
mAs range
Molybdenum anode 2 mAs to 560 mAs in mAs mode
0 mAs to 600 mAs in AEC mode
Tungsten anode 2 mAs to 710 mAs in mAs mode
0 mAs to 752 mAs in AEC mode
Exposure times, automatic 10 ms to 4 s (large focus), 10 ms to 7 s (small focus)
Exposure times, manual 13 ms to 4 s (large focus), 70 ms to 7 s (small focus)

X-ray tube unit


Mammography X-ray tube unit with quadruple-focus rotating-anode tube (molybdenum / tungsten1))
Focal spot nominal value
Molybdenum focal spot 0.1 / 0.3 (star pattern test)
0.15 / 0.3 (IEC 336)
Tungsten focal spot1) 0.1 / 0.3 (star pattern test)
0.15 / 0.3 (IEC 336)
Maximum tube current
Molybdenum focal spot 150 mA at 25 kV
Tungsten focal spot 1)
188 mA at 25 kV
Nominal voltage (IEC 613) 40 kV
Heat storage capacity (tube unit) 1 100 000 J, 1 500 000 HU
Heat storage capacity (anode) 120 000 J, 162 000 HU
Optical anode angle (IEC 788) 20º
Inherent filtration 1 mm Be
Anode speed 8800 rpm

5
MAMMOMAT 1000 / 3000 Nova

Technical Data

Automatic exposure control


Automatic exposure control (AEC) Microprocessor-controlled, with transparency compensation
Detectors Semiconductor detector
Accuracy of optical density (2 to 6 cm) ± 0.15 OD from the mean optical density for predefined film-screen
combinations (separately for each Bucky table, magnification table and
anode/filter combination at appropriate clinical kV)
Density correction ± 3 exposure points, adjustable in 1 / 8 step increments
Film-screen combinations Selectable at the generator control panel (H/D)

Stereotactic biopsy device*, 1)


Accuracy for ≤ 100 mm x, y, z ± 1.0 mm
compression thickness
Needle position in steps 0.1 mm
Biopsy field size 58 mm x 40 mm
Compression opening 0 to 140 mm
Accommodates needle lengths from 30 mm to 175 mm
Needle holders 0.7 mm (22 G), 0.9 mm (20 G), 1.2 mm (18 G), 1.65 mm (16 G) 2.1 mm (14 G),
2.5 mm (13 G)
Open needle holders (localization) 0.9 mm (20 G), 1.2 mm (18 G)
Film size 18 cm x 24 cm
Cassettes DIN 6832 / IEC 406 (15 mm thickness)
Weight, biopsy unit 12 kg
Weight, evaluation unit 12 kg

Weight
Examination stand 300 kg (660 Ibs)
Generator with radiation shield 141 kg (310 Ibs)
Free-standing radiation shield 46 kg (100 Ibs)

* Option
1)
Only for MAMMOMAT 3000 Nova

6
Options
Compression plate with low edge, 18 cm x 24 cm
Compression plate with high edge, 18 cm x 24 cm
Compression plate with low edge, 24 cm x 30 cm
Compression plate with high edge, 24 cm x 30 cm
Flexible compression plate with high edge, 18 cm x 24 cm
Flexible compression plate with high edge, 24 cm x 30 cm
Detail/spot compression plate
Axilla compression plate
Magnification table 1.5
Magnification table 1.8
Perforated biopsy compression plate
Second object table with Bucky 18 cm x 24 cm
Object table with Bucky, 24 cm x 30 cm
Object table without Bucky, 18 cm x 24 cm
Object table without Bucky, 24 cm x 30 cm
Wall holder for compression plates
Rack for storing cassettes
Rack for the evaluation unit
0.1 or 0.3 mm lead equivalent radiation shield
Biopsy device with shadow cross
Stereotactic biopsy device 1)
Opdima
Monitor desk for Opdima
Exposure release button
Printer interface
Triacon ID network camera

Operating data
Power requirements 110 V, 208 V, 230 V, 240 V, 277 V, ± 10%, single-phase
208 V, 230 V, 240 V, 277 V, 400 V ± 10%, two-phase, 50/60 Hz ± 1 Hz

Environmental operating conditions


Temperature range + 20ºC to + 40ºC
Relative humidity 15% to 75%, non-condensing
Atmospheric pressure 700 hPa to 1060 hPa

7
MAMMOMAT 1000 / 3000 Nova

Accessories

Accessories (optional)

Compression plate with low edge 18 cm x 24 cm # 44 94 378

Standard compression plate with 40 mm high front edge.

Compression plate with high edge 18 cm x 24 cm # 44 94 386

Standard compression plate with 70 mm high front edge.

Compression plate with low edge 24 cm x 30 cm # 44 94 410

Standard compression plate with 40 mm high front edge.

8
Accessories (optional)

Compression plate with high edge 24 cm x 30 cm # 44 94 428

Standard compression plate with 70 mm high front edge.

Flexible compression plate with high edge 18 cm x 24 cm # 44 94 394

This flexible compression plate flexes in two directions:


– chest wall to nipple side for CC views
– axillary-inframammary fold side for lateral views
The self-adjusting tilt depends on the thickness and shape of the breast, and
provides a more uniform compression over the entire breast surface.
The plate can be used for all standard views.

Flexible compression plate with high edge 24 cm x 30 cm # 44 94 436

This flexible compression plate flexes in two directions:


– chest wall to nipple side for CC views
– axillary-inframammary fold side for lateral views
The self-adjusting tilt depends on the thickness and shape of the breast, and
provides a more uniform compression over the entire breast surface.
The plate can be used for all standard views.

9
MAMMOMAT 1000 / 3000 Nova

Accessories

Accessories (optional)

Detail/spot compression plate # 44 94 444

Special plate used with 18 cm x 24 cm object table. The compression with this
plate simplifies the visualization of the suspicious area in spot diagnostics.
External collimator is included.

Axilla compression plate # 44 94 477

Special plate for compression of axillary areas. Can also be used for
male breasts, small breasts, breasts after surgery and implants (push back
views). External collimator is included.

10
Accessories (optional)

Magnification table 1.5 # 44 94 469

The magnification table with magnification factor 1.5 includes two com-
pression plates; one spot compression plate with external collimator plate and
one flat compression plate.

11
MAMMOMAT 1000 / 3000 Nova

Accessories

Accessories (optional)

Magnification table 1.8 # 44 94 451

The magnification table with magnification factor 1.8 includes two com-
pression plates; one spot compression plate with external collimator plate and
one flat compression plate.

12
Accessories (optional)

Second object table with Bucky 18 cm x 24 cm # 44 95 227

Detachable object table with Bucky for 18 cm x 24 cm cassettes. (One object


table with 18 x 24 cm Bucky is included in the scopy of supply).

Object table without Bucky 18 cm x 24 cm # 44 94 501

Object table without Bucky, attaches with quick-action lock, for 18 cm x 24 cm


cassettes.

Object table with Bucky 24 cm x 30 cm # 44 94 402

Detachable object table with Bucky for 24 cm x 30 cm cassettes.

Object table without Bucky 24 cm x 30 cm # 44 94 519

Object table without Bucky, attaches with quick-action lock, for 24 cm x 30 cm


cassettes.

13
MAMMOMAT 1000 / 3000 Nova

Accessories

Accessories (optional)

Stereotactic biopsy device # 44 94 709

The stereotactic biopsy device is mounted on the base plate of the


MAMMOMAT 3000 Nova.
Special two-angle views are taken on the same film and processed in the
evaluation unit.
The coordinates are defined and sent to the biopsy unit where the needles
are inserted. The accuracy is ± 1 mm.
The biopsy attachment is used together with a special compression plate and
external collimator. The stereotactic attachment can be used with fine needles,
core needles and with the Mammotome system from Ethicon.
Different needle holders are required depending on the needle size.
Only for MAMMOMAT 3000 Nova.

14
Accessories (optional)

Perforated biopsy compression plate # 44 94 493

A special plate for biopsy and localization. The plate has several perforations
and a coordinate system. The patient remains in compression while the phy-
sician determines where the lesion is located. The needle is then inserted into
the breast. The position of the lesion can only be defined in the x and y direc-
tion. The depth must be checked with perpendicular views.

Biopsy device / Shadow cross # 44 94 485

The shadow cross attachment is mounted on the tube unit and used with an
external collimator and a fenestrated coordinate plate. The patient remains in
compression while the physician determines where the lesion is located.
The cross is set over the coordinates and the needle is inserted at the center of
the cross. Only X and Y coordinates are defined, but with the use of the cross
the accuracy is higher than with the perforated biopsy compression plate.

15
MAMMOMAT 1000 / 3000 Nova

Accessories

Accessories (optional)

Wall holder for compression plates # 44 94 584

Wall holder for 6 compression plates with corresponding diaphragms and


shadow cross.

Rack for the evaluation unit # 44 94 600

Rack for the evaluation unit or an ID camera, 45 cm width, 47 cm depth.

Rack for storing cassettes # 44 94 592

Rack for storing cassettes, object tables and the biopsy device, 75 cm width,
47 cm depth.

16
Accessories (optional)

Exposure release button # 44 94 360

Remote exposure release button for use with separate generator control
console.

0.1 mm Lead equivalent radiation shield # 44 94 345

Free-standing radiation shield consisting of a frame and 0.1 mm lead


equivalent lead glass.

0.3 mm Lead equivalent radiation shield # 44 94 352

Free-standing radiation shield consisting of a frame and 0.3 mm lead


equivalent lead glass.

17
MAMMOMAT 1000 / 3000 Nova

Technical Data

Dimensions in cm (inches)

MAMMOMAT 3000 Nova

3000 Nova
��������

��
�����
�����

���
���

��������
��

�� ��
�� ���
���

MAMMOMAT 1000

�������������
��
�����
�����

���
���
���

������� �������������
��������
���

���

��

������ �� ������ ���� ��

�� ��� ��

���

18
������

�����
�����

���
���

��������
��

�� ��
Dimensions in cm (inches) �� ���
���

MAMMOMAT 1000/3000 Nova

�������������

��
�����
�����

���
���

���

��������
������� �������������

��
���

���

�� ��
�������� ������ ��� �� ��
��� ��
���

������� �������������
���

�� ��
��

19
The information in this document contains general
descriptions of the technical options available
and may not always apply in individual cases.

The required features should therefore be specified


in each individual case at completion of contract.

Siemens reserves the right to modify the design


and specifications contained herein without prior
notice. Please contact your local Siemens sales
representative for the most current information.

Original images always lose a certain amount of


detail when reproduced.

In the interest of complying with legal requirements


concerning the environmental compatibility of our
Contact addresses: A
products (protection of natural resources, waste
conservation), we recycle certain components. In the USA
Using the same extensive quality assurance measures
as for new components, we guarantee the quality
Siemens Medical Solutions USA, Inc.
of these recycled components. 51 Valley Stream Parkway
Malvern, PA 19355
Siemens AG Telephone: +01 610 448 4500
Wittelsbacherplatz 2 Telefax: +01 610 448 1620
D-80333 München
Siemens Medical
Germany In Germany
Siemens AG, Medical Solutions
Solutions that help
Headquarters Special Systems
Siemens AG, Medical Solutions Allee am Röthelheimpark 2
Henkestr. 127, D-91052 Erlangen D-91052 Erlangen
Germany Germany © 12.2005 Siemens AG
Telephone ++49 9131 84-0 Order No. A91001-M1610-G953-1-7600
Telephone ++49 9131 84-0 Printed in Germany
www.siemens.com/medical siemens.com/medical SP PLM DA 12054